K Number
K252842

Validate with FDA (Live)

Date Cleared
2026-01-11

(125 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

Device Description

The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, Delta Dilators and Stimulating Probe and Delta Clip contained in various kits.

The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the SafeOp 3: Neural Informatix System (K252842) primarily describe the device's substantial equivalence to a predicate device (K234092) and outline a Predetermined Change Control Plan (PCCP) for future modifications. However, it does not include a detailed study proving the device meets specific acceptance criteria in the context of clinical performance or AI algorithm validation for the initial clearance.

The "Performance Data" section explicitly states: "Determination of substantial equivalence is not based on an assessment of clinical performance data." and refers to nonclinical performance testing (functional, system, software requirements) and usability testing.

Therefore, the following information is extracted and synthesized based on the available document, but a comprehensive study proving acceptance criteria for all aspects of neural informatic performance is not present. The document focuses on the future PCCP for algorithmic changes, which will have performance criteria.


Acceptance Criteria and Reported Device Performance

The core of the documentation for this 510(k) clearance is about demonstrating substantial equivalence to an existing predicate device (SafeOp 3™: Neural Informatix System, K234092). Therefore, the "acceptance criteria" for the cleared device itself are primarily its ability to match the predicate's performance and specifications, which is shown through a direct comparison.

The document does not report specific quantitative performance metrics for the SafeOp 3: Neural Informatix System for its initial clearance, as it relies on substantial equivalence.

However, the Predetermined Change Control Plan (PCCP) outlines future modifications to the SSEP Baseline Classification Algorithm and defines acceptance criteria for these future algorithmic changes. These are the only explicitly stated performance acceptance criteria in the document.

Table 1: Acceptance Criteria and Reported Device Performance (from PCCP for future algorithmic changes)

Acceptance Criteria (for future SSEP algorithm modifications)Reported Device Performance (for future SSEP algorithm modifications)Notes
Modification #1 & #4 (SSEP Baseline Classification Algorithm Retraining/Maintenance):No reported performance data for the current cleared device.The document states these are criteria for future modifications of the SSEP algorithm, not for the initial clearance of the SafeOp 3 System.
Overall Positive Predictive Value (PPV) ≥ 90%To be demonstrated with verification testing of modified algorithm.These criteria are goals for future changes.
Overall Accuracy ≥ 85%To be demonstrated with verification testing of modified algorithm.
Overall F1-score ≥ 85%To be demonstrated with verification testing of modified algorithm.
Performance of each stimulation site-specific classifier to be demonstrated as non-inferior to the current SafeOp 3 System at the 95% confidence level.To be demonstrated with verification testing of modified algorithm.
Modification #2 (Removal of Recalculating Active Trace (RAT) State):No reported performance data for the current cleared device.These are criteria for a future modification involving signal processing.
Average noise content for the reconstructed Ensemble Average (EA) following restart to be demonstrated as non-inferior to that of the preceding EA(s) with 95% confidence.To be demonstrated with internal testing and retrospective analysis of intraoperative data.These criteria are goals for future changes.

Study Details (Based on the document, for the initial clearance)

Because the 510(k) is based on substantial equivalence and explicitly states "Determination of substantial equivalence is not based on an assessment of clinical performance data," there is no detailed "study" proving the device meets acceptance criteria in the way a de novo or PMA submission might. The "study" mentioned below refers to the verification and validation activities conducted to demonstrate substantial equivalence and safety/effectiveness for the initial clearance, as described in the "Performance Data" section.

  1. Sample size used for the test set and the data provenance:

    • Not explicitly stated for clinical performance data for the initial clearance. The document mentions "nonclinical performance testing" for functional, system, and software requirements, and "usability testing." These typically involve controlled test cases and user interactions rather than a large clinical test set.
    • For the future PCCP modifications of the SSEP algorithm, it mentions "an independent, firewalled dataset" and "additional labeled data" for re-training and testing, but no sample size or provenance is specified within this document.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not explicitly stated for the initial clearance. Since the clearance is based on substantial equivalence and does not assess clinical performance data, detailed expert ground truth establishment for a clinical test set is not described.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not explicitly stated for the initial clearance. Not applicable given the reliance on substantial equivalence and lack of clinical performance data assessment.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. The document does not describe an MRMC study. The device is a neural informatic system, and while it involves signal processing, it's not positioned as an AI-assisted diagnostic tool that human readers would interpret with assistance in the traditional sense of an MRMC study. The SSEP algorithm mentioned in the PCCP appears to be a standalone classification, not a human-in-the-loop assistance tool subject to such a study.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • For the initial clearance, the document refers to "nonclinical performance testing" demonstrating the system meets functional, system, and software requirements. This implies standalone testing of device functions.
    • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), the acceptance criteria (PPV, Accuracy, F1-score) clearly indicate standalone algorithm performance is intended to be evaluated. However, no such standalone performance data is provided for the initially cleared device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not explicitly stated for the initial clearance. For the "nonclinical performance testing" and "usability testing," ground truth would typically be established based on engineering specifications and validated operational procedures.
    • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," which implies a ground truth established by experts or other definitive methods, but the specific type is not detailed in this document.
  7. The sample size for the training set:

    • Not explicitly stated for the initial clearance. No AI algorithms are reported as being evaluated for the initial clearance.
    • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it mentions the algorithm "will be re-trained using site-specific data" and "periodically re-trained as additional labeled data become available," but no sample size is specified within this document.
  8. How the ground truth for the training set was established:

    • Not explicitly stated for the initial clearance.
    • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," but the method of labeling (establishing ground truth) is not detailed in this document.

FDA 510(k) Clearance Letter - SafeOp 3: Neural Informatix System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

January 11, 2026

Alphatec Spine, Inc.
Garima Shrivastava
Senior Regulatory Affairs Specialist
1950 Camino Vida Roble
Carlsbad, California 92008

Re: K252842
Trade/Device Name: SafeOp 3: Neural Informatix System
Regulation Number: 21 CFR 882.1870
Regulation Name: Evoked response electrical stimulator
Regulatory Class: Class II
Product Code: GWF, GXY, GXZ, IKN, PDQ, ETN
Dated: December 12, 2025
Received: December 12, 2025

Dear Garima Shrivastava:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an

Page 2

K252842 - Garima Shrivastava
Page 2

established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Page 3

K252842 - Garima Shrivastava
Page 3

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JAY R. GUPTA -S

Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K252842

Please provide the device trade name(s).

SafeOp 3: Neural Informatix System

Please provide your Indications for Use below.

The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

Please select the types of uses (select one or both, as applicable).

Page 5

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 1 of 7

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

A. SUBMITTER:

Alphatec Spine, Inc.
1950 Camino Vida Roble
Carlsbad, CA 92008
Phone: (760) 431-9286
Fax: (760) 431-0289

Contact Person: Garima Shrivastava
Senior Regulatory Affairs Specialist
Alphatec Spine, Inc.
Email: gshrivastava@atecspine.com
Contact Phone: (760) 356-6783

Date Summary Prepared: September 5, 2025

B. DEVICE

Trade or Proprietary Name: SafeOp 3: Neural Informatix System
Common or Usual Name: Intraoperative Neuromonitoring
Classification Name: Stimulator, Electrical, Evoked Response
Regulation Number: 21 CFR 882.1870
Classification: Class II
Product Codes: GWF, GXY, GXZ, IKN, PDQ, ETN

C. LEGALLY MARKETED PREDICATE DEVICES

Primary Predicate:

510(k)Product NameClearance Date
K234092SafeOp 3™: Neural Informatix System04/19/2024

D. DEVICE DESCRIPTION

The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, Delta Dilators and Stimulating Probe and Delta Clip contained in various kits.

The subject device is intended for use by trained healthcare professionals, clinical

Page 6

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 2 of 7

neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.

The subject device records the following modalities:

  • Somatosensory evoked potentials (SSEP)
  • Motor evoked potentials (MEP),
  • Train-of-four neuromuscular junction (TO4),
  • Triggered electromyography (tEMG) and
  • Free run electromyography (sEMG)

E. INTENDED USE AND INDICATIONS FOR USE

Intended Use

The SafeOp 3: Neural Informatix System is intended for use by trained healthcare professionals as a diagnostic device that provides intraoperative neuromonitoring during various surgical procedures.

Indications for Use

The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

F. TECHNOLOGICAL COMPARISON TO PREDICATES

The subject device was compared to the predicate device in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, does not impact substantially equivalence, or safety and effectiveness. Table 1 below provides a detailed analysis of the substantial equivalence of the following device characteristics: indications for use and device functionalities.

Page 7

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 3 of 7

Table 1: Comparison for Substantial Equivalence

Specification/PropertyPredicate DeviceSubject DeviceSE Rationale (if not identical)
510(k)SafeOp 3: Neural Informatix System (K234092)SafeOp 3: Neural Informatix System
Intended Use/ Indications for UseThe SafeOp 3: Neural Informatix System is intended for use by trained healthcare professionals as a diagnostic device that provides intraoperative neuromonitoring during various surgical procedures. The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures. SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.The SafeOp 3: Neural Informatix System is intended for use by trained healthcare professionals as a diagnostic device that provides intraoperative neuromonitoring during various surgical procedures. The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures. SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.The indications for use for the subject SafeOp 3 System are identical to the predicate SafeOp 3 System (K234092).
Device ClassIIIIIdentical to predicate, SafeOp 3 System (K234092)
Product CodeGWF, GXY, GXZ, IKN, PDQ, ETNGWF, GXY, GXZ, IKN, PDQ, ETNIdentical to predicate, SafeOp 3 System (K234092)
Regulation Number (21 CFR)§882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820§882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820Identical to predicate, SafeOp 3 System (K234092)
Device Classification NameStimulator, Electrical, Evoked Response, Surgical nerve stimulator/locatorStimulator, Electrical, Evoked Response, Surgical nerve stimulator/locatorIdentical to predicate, SafeOp 3 System (K234092)
Monitoring Modalities/Operating Modes• Electrode Test • Electromyography (EMG) • Somatosensory Evoked Potentials (SSEP) • Transcranial Motor Evoked Potential (TcMEP or MEP) • Neuromuscular Junction Testing – Train of Four (TO4)• Electrode Test • Electromyography (EMG) • Somatosensory Evoked Potentials (SSEP) • Transcranial Motor Evoked Potential (TcMEP or MEP) • Neuromuscular Junction Testing – Train of Four (TO4)Identical to predicate, SafeOp 3 System (K234092).
System Components• Patient Interface • Tablet • Docking station• Patient Interface • Tablet • Docking stationSubstantially equivalent to predicate, SafeOp 3 System (K234092). The subject device incorporates minor hardware updates to the Patient

Page 8

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 4 of 7

Specification/PropertyPredicate DeviceSubject DeviceSE Rationale (if not identical)
510(k)SafeOp 3: Neural Informatix System (K234092)SafeOp 3: Neural Informatix System
Interface and the Tablet that do not alter the device's intended use, technological characteristics, or clinical performance. Performance testing confirmed equivalent performance and no new or increased risks.
60601-1 CompliantYesYesIdentical to predicate, SafeOp 3 System (K234092)
Power Supply100 to 240 VAC, 50-60 Hz (input); 12 VDC, 2.5 A (output)100 to 240 VAC, 50-60 Hz (input); 12 VDC, 2.5 A (output)Identical to predicate, SafeOp 3 System (K234092)
Mode of OperationContinuousContinuousIdentical to predicate, SafeOp 3 System (K234092)
DimensionsPatient Interface: 14"H. x 8.5"W x 5"D Tablet with Docking Station: 11.5"H x 13.3"W x 6.5"DPatient Interface: 14"H. x 8.5"W x 5"D Tablet with Docking Station: 11.5"H x 13.3"W x 6.5"DIdentical to predicate, SafeOp 3 System (K234092)
WeightPatient Interface: < 7 lbs Tablet with Docking Station: 5.45 lbsPatient Interface: < 7 lbs Tablet with Docking Station: 5.45 lbsIdentical to predicate, SafeOp 3 System (K234092)
Total Amplifier ChannelsUp to 16Up to 16Identical to predicate, SafeOp 3 System (K234092)
Stim WaveformMonophasic, RectangularMonophasic, RectangularIdentical to predicate, SafeOp 3 System (K234092)
Stim Pulse Duration100 to 500 μsec (SSEP) 200 μsec (tEMG) 200 to 400 μsec (TO4) 50 to 150 μsec (MEP)100 to 500 μsec (SSEP) 200 μsec (tEMG) 200 to 400 μsec (TO4) 50 to 150 μsec (MEP)Identical to predicate, SafeOp 3 System (K234092)
Stim Frequency (Pulse Rate)0.1 to 50 Hz0.1 to 50 HzIdentical to predicate, SafeOp 3 System (K234092).
Stim Current Range0 to 100 mA 0 to 1500 mA for MEP only0 to 100 mA 0 to 1500 mA for MEP onlyIdentical to predicate, SafeOp 3 System (K234092).
Stim Voltage Range0-1000 V for MEP0-1000 V for MEPIdentical to predicate, SafeOp 3 System (K234092).
Input Impedance> 50 MΩ> 50 MΩIdentical to predicate, SafeOp 3 System (K234092)
Low Frequency Filter1 - 30 Hz1 - 30 HzIdentical to predicate, SafeOp 3 System (K234092).
High Frequency Filter1.5 – 3kHz1.5 – 3kHzIdentical to predicate, SafeOp 3 System (K234092).
Operating SystemWindows 10 powered tabletWindows 11 powered tabletSubstantially equivalent to predicate, SafeOp 3 System (K234092). For more information, see Section-9, Performance Testing.

Page 9

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 5 of 7

Specification/PropertyPredicate DeviceSubject DeviceSE Rationale (if not identical)
510(k)SafeOp 3: Neural Informatix System (K234092)SafeOp 3: Neural Informatix System
Remote AccessNoNoIdentical to predicate, SafeOp 3 System (K234092).
Voltage Range50-1000V50-1000 VIdentical to predicate, SafeOp 3 System (K234092).
Current Range50 μsec - 500 μsec50 μsec - 500 μsecIdentical to predicate, SafeOp 3 System (K234092).
Impedance MeasurementAll outputsAll outputsIdentical to predicate, SafeOp 3 System (K234092).
MEP Stim OutputsFour outputs, arranged in two pairs, polarity selectableFour outputs, arranged in two pairs, polarity selectableIdentical to predicate, SafeOp 3 System (K234092).
MEP Repetition rate1 Hz1 HzIdentical to predicate, SafeOp 3 System (K234092).
Maximum energy into 1 kohm impedance50mJ/pulse50mJ/pulseIdentical to predicate, SafeOp 3 System (K234092).

Page 10

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 6 of 7

G. PREDETERMINED CHANGE CONTROL PLAN (PCCP)

The PCCP for the SafeOp 3 System includes the following planned software and algorithmic modifications, which will be implemented under documented design control procedure. Each proposed change will be implemented in accordance with detailed modification protocols and must meet defined performance and safety criteria prior to implementation.

• Modification #1 - Stimulation site specific SSEP Baseline Classification Algorithm

The system's SSEP baseline classification algorithm will be re-trained using site-specific data from the posterior tibial, saphenous, and ulnar nerves. Dedicated classifiers for each stimulation site are expected to improve robustness and performance while maintaining the algorithm's existing inputs, outputs, and intended use. Performance verification will include testing against predefined accuracy, positive predictive value (PPV), and F1-score criteria using an independent, firewalled dataset. The modified SSEP baseline classification algorithm must meet overall PPV ≥ 90%, overall accuracy and F1-score ≥ 85%. Performance of each stimulation site-specific classifier must be demonstrated to be non-inferior to the current SafeOp 3 System at the 95% confidence level.

• Modification #2 - Removal of the Recalculating Active Trace (RAT) State

The planned modification will remove the RAT state from the SSEP modality. Following restart of SSEP acquisition, new single trials will be incorporated into the existing ensemble average rather than resetting it. Internal testing and retrospective analysis of intraoperative data will be used to confirm non-inferiority in signal quality relative to the current implementation. The average noise content for the reconstructed EA following restart will be demonstrated to be non-inferior to that of the preceding EA(s) with 95% confidence.

• Modification #3 - tEMG Using Only the SafeOp Delta Stimulating Clip

The system will be modified to enable both traditional and Delta triggered EMG (tEMG) functionality using only the SafeOp Delta Stimulating Clip. An impedance-based detection method will determine the type of surgical instrument connected and automatically initiate the appropriate workflow. Verification activities will confirm system behavior, stimulation accuracy, and UI performance across clinical workflows.

• Modification #4 – Ongoing Maintenance and Update of the SSEP Baseline Classification Algorithm

The stimulation site–specific SSEP classifiers may be periodically re-trained as additional labeled data become available. The model type, inputs, outputs, and intended use will remain unchanged, with all configuration parameters constrained to predefined bounded ranges in the PCCP. The frequency of updates is expected to be four releases per year or fewer and will be

Page 11

Traditional 510(k) Premarket Notification SafeOp™ 3 System

Page 7 of 7

implemented under established design-control procedures. Verification testing will assess accuracy, PPV, and F-1 score using an independent dataset, and validation will confirm non-inferiority to the cleared classifier. The modified SSEP baseline classification algorithm must meet overall PPV ≥ 90%, overall accuracy and F1-score ≥ 85%. Performance of each stimulation site-specific classifier must be demonstrated to be non-inferior to the current SafeOp 3 System at the 95% confidence level.

Additionally, software workflow and system integration testing will be completed for each of the planned modifications described above.

Each planned modification will be executed in accordance with detailed modification protocols defined in the PCCP. All modifications will undergo appropriate verification and validation activities—including performance testing, design verification, and user validation (when applicable)—to confirm compliance with established acceptance criteria. Authorized modifications will be documented in the device's labeling, as applicable. The implementation of the predetermined change control plan demonstrates continued substantial equivalence to the predicate device.

H. PERFORMANCE DATA

Nonclinical performance testing demonstrates that the subject SafeOp 3 System meets the functional, system, and software requirements.

Usability testing was performed to demonstrate that the subject SafeOp 3 System presents no adverse effect within the intended environment, and the subject device was therefore found to be substantially equivalent to the predicate. A use-related risk analysis (URRA) was conducted as part of the usability testing to identify potential use-related hazards and confirm that implemented mitigations reduce residual risks to acceptable levels.

Clinical Information

Determination of substantial equivalence is not based on an assessment of clinical performance data.

I. CONCLUSION

Based upon the information provided in this 510(k) submission, it has been determined that the subject device, SafeOp 3 System, is substantially equivalent to the legally marketed predicate device in regard to indications for use, intended use, design, technology, and performance.

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).