K Number
K120979

Validate with FDA (Live)

Date Cleared
2012-04-25

(23 days)

Product Code
Regulation Number
882.1870
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).

Evoked Potential (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300, Motor Evoked Potentials (MEP) and Contingent Negative Variation (CNV). The Nicolet EDX with Synergy Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Synergy Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.

The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.

The Nicolet EDX with Synergy Software is intended to be used by a qualified healthcare provider.

Device Description

The CareFusion Nicolet EDX system with Synergy Software (Synergy EDX) is designed for the acquisition, display, analysis, reporting, and management of electrophysiological information from the human nervous and muscular systems. The system is designed to perform nerve conduction studies (NCS), needle electromyography (EMG) testing, evoked potential (EP) testing, and intra-operative monitoring (IOM). Synergy EDX provides a variety of tests spanning the various modalities.

The Synergy EDX consists of the following major components:

  • Nicolet EDX console base unit;
  • Synergy control panel:
  • Nicolet amplifier (there are two types available: 2 channel (AT2) with two non-switched amplifier channels and an 8 channel (AT2 + 6) amplifier with two non-switched and six switched amplifier channels;
  • Desktop or laptop computer with a keyboard and mouse;
  • Display monitor; and
  • Synergy Software

The Synergy EDX optional accessories/components consists of the following:

  • Nicolet HB6 or HB7 Head Box
  • Stimulator probes (RS 10 probe, WR 50 Probe, S403 probe)
  • SP1/SP2 electrical stimulator switching units
  • Footswitches (single and triple)
  • LED goggles
  • Patient response button
  • Photic strobe
  • Headphones or other auditory transducers
  • Cart
  • Isolation transformer
  • Printer
AI/ML Overview

This document is a 510(k) Summary for the CareFusion Nicolet EDX with Synergy Software. It primarily serves to demonstrate substantial equivalence to previously cleared predicate devices, rather than presenting a performance study with detailed acceptance criteria and results from a clinical trial or a structured validation study as one might expect for a novel AI/ML device.

Here's an analysis based on the provided text, addressing your points:

The Device: The CareFusion Nicolet EDX with Synergy Software is a diagnostic electromyograph system designed for acquiring, displaying, analyzing, reporting, and managing electrophysiological information from human nervous and muscular systems. It performs nerve conduction studies (NCS), needle electromyography (EMG), evoked potential (EP) testing, and intra-operative monitoring (IOM).

Overview of Acceptance Criteria and Performance (Based on Substantial Equivalence):

This submission is a 510(k) for substantial equivalence, not a new de novo device with specific clinical performance acceptance criteria against a ground truth. The acceptance criteria essentially revolve around demonstrating that the new device has the same technological characteristics, indications for use, and safety/effectiveness as its predicate devices, or that any differences do not raise new questions of safety or effectiveness.

The document primarily shows that the CareFusion Nicolet EDX with Synergy Software is "Identical" or "Similar" to its predicate device, the CareFusion Nicolet EDX with Viking software (K112052), with one minor addition (CNV test to EP modality) and one minor difference in auditory tone frequency that doesn't change indications for use or safety/effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is a substantial equivalence submission relying on comparison to a predicate, the "acceptance criteria" are implied by the similarity to the predicate, and "performance" is demonstrated by showing these similarities through technical specifications, rather than new quantitative performance metrics.

CharacteristicAcceptance Criteria (Implied by Predicate K112052)Reported Device Performance (CareFusion Nicolet EDX with Synergy Software)Discussion of Differences (from Predicate)
1.1 Indications For UseSame as CareFusion Nicolet EDX with Viking (K112052): Acquisition, display, analysis, storage, reporting, and management of electrophysiological information from human nervous and muscular systems (NCS, EMG, EP, Autonomic Responses, IOM incl. EEG). EP includes VEP, AEP, SEP, ERG, EOG, P300, MEP. Autonomic testing includes GSR, SSR, RR Interval variability. Used to detect physiologic function of nervous system, locate neural structures during surgery, and support diagnosis of neuromuscular disease or condition.Same as described for acceptance criteria.Identical to the Nicolet EDX with Viking software, with the minor addition of the CNV test being a part of the Evoked Potential Modality.
1.2 WarningsItems related to off-label use or misuse.Identical.Identical to the Nicolet EDX with Viking software.
1.3 Contra-indicationsItems related to design and indicated use limitations (e.g., not for use in flammable anesthetics or with defibrillation).Identical.Identical to the Nicolet EDX with Viking software.
2.1 General systems approachComputer-based equipment with dedicated hardware peripherals/components.Identical.Identical to the Nicolet EDX with Viking software.
2.2 User input deviceWindow mouse/keyboard driven graphic interface with dedicated control panel.Window mouse/keyboard driven graphic interface with dedicated control panel.Identical to the Nicolet EDX with Viking software. (Footnote 1 states Synergy control panel is functionally equivalent but differs in number of controls, which is deemed acceptable).
2.12 Safety StandardsCompliance with IEC 60601-1:1998 + A1:1991+A2:1995, IEC 60601-1-1:2000, EN/IEC 60601-1-2: Ed. 2.0+A1:2004, IEC 60601-2-40: 1998, UL 60601-1: 2003-04-25 ED1 Rev:2003/06/30, CAN/CSA-C22.2 no. 601.1-M90.Compliance demonstrated through testing.Identical to the Nicolet EDX with Viking software (as implied by compliance with same standards).
3.1 Number of channels2 to 82 to 8Identical to the Nicolet EDX with Viking software.
3.2 CMMR> 110 dB> 110 dBIdentical to the Nicolet EDX with Viking software.
3.3 Noise< 0.6uV RMS (from 2 Hz to 10 kHz)< 0.6uV RMS (from 2 Hz to 10 kHz)Identical to the Nicolet EDX with Viking software.
4.2.4 Auditory Tone Frequency250 to 8000 Hz125 to 8000 Hz (This Submission); Predicate K965065 has 125 Hz.Similar to the Nicolet EDX with Viking software with the addition of 125 Hz frequency. The additional frequency does not change the indications for use or safety and effectiveness. The Synergy predicate (K965065) has 125 Hz frequency.
Application Modules (EMG, EP, IOM)Functionally equivalent to corresponding modules in predicate device (NCS, EMG, EP, IOM, specific EP tests like SEP, AEP, VEP, P300, ERG, EOG, Processed EEG, MEP, CNV, Free Run Acquisition, Stimulator Triggered, Signal Triggered, SPA, SFEMG, Motor Unit Analysis, F-Wave, H-Reflex, SSR).Functional equivalence demonstrated, e.g., "Yes" for all listed modules. For CNV, "Yes" in this submission versus "No" in Viking predicate.CNV is a new addition to the EP modality (as discussed under 1.1). The document also lists "New Features" like Automatic Report Narrative Generation and Electrical Stimulus Automation as "Yes" for Synergy System and "No" for Viking. These are considered "New Features" but not specifically new performance claims against acceptance criteria beyond functional equivalence.

Summary of the "Study" (Demonstration of Substantial Equivalence):

The "study" here is not a typical clinical trial or performance study with a test set of data and ground truth as one would conduct for an AI/ML device. Instead, it's a demonstration of substantial equivalence to previously cleared devices (predicates) through:

  • Comparison of Indications for Use: Showing they are the same or have minor, non-critical differences.
  • Comparison of Device Description and Technological Characteristics: Detailing hardware components, software approach, electrical characteristics, and application modules, and highlighting that they are identical or very similar to the predicate.
  • Non-Clinical Performance Testing:
    • Biocompatibility: Compliance with ISO 10993-1:2009.
    • Software testing: Designed and developed according to robust software development process (FDA guidance documents cited). Test results demonstrated compliance with predetermined specifications.
    • Electrical Safety & EMC Testing: Compliance with IEC 60601-1, UL 60601-1, and IEC 60601-1-2.
    • Performance Testing - Bench: Compliance with IEC 60601-2-40 (for electromyographs and evoked response equipment).

Specific Answers to Your Questions:

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
This type of information is not applicable or provided in this 510(k) summary. The submission does not describe a performance study with a "test set" of patient data or clinical images/signals. The testing reported is primarily bench testing, software verification/validation against specifications, and safety/EMC compliance. No clinical data (retrospective or prospective) from patient populations is mentioned for performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. There was no "test set" of clinical data requiring expert review for ground truth establishment.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. No clinical test set data was adjudicated.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was done. This device is a diagnostic instrument, not an AI-powered assistive diagnosis tool in the sense that would require an MRMC study to show improvement in human reader performance. Its function is to acquire, display, and analyze electrophysiological signals, not to provide an AI-driven interpretation or assisted read.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable in the typical AI sense. The device's "performance" is its ability to accurately acquire, process, and display biological signals according to its technical specifications, and to function safely and effectively as an electromyograph. Its performance was assessed through engineering bench testing against established standards, not through an algorithm-only standalone performance evaluation against a "ground truth" derived from patient data.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the technical characteristics and safety/EMC, the "ground truth" is adherence to engineering specifications, international safety standards (IEC, UL), and FDA guidance for software development. For biocompatibility, it's compliance with ISO 10993-1. These are widely accepted objective standards and specifications for medical device manufacturing and performance. There is no biological "ground truth" (e.g., pathology, expert diagnosis) used in this submission to establish device accuracy for a diagnostic claim, as the device itself is the measurement tool, comparable to a thermometer or blood pressure cuff.

8. The sample size for the training set:
Not applicable. This document describes a diagnostic device, not an AI/ML system that undergoes a "training" phase with a dataset.

9. How the ground truth for the training set was established:
Not applicable. No training set was used.

{0}------------------------------------------------

1

·

510(k) Summary CareFusion Nicolet EDX with Synergy Software

[A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92]

Submitter / 510(k) Holder 1.

NameCareFusion 209, Inc.
Address1850 Deming Way, Middleton WI 53562
Establishment Reg. No.3008289288
Contact personMr. Curtis Truesdale
Phone number608-829-8684
Fax number608-829-8769
EmailCurtis.truesdale@carefusion.com
Date preparedApril 18, 2012
Device Name
Proprietary nameCareFusion Nicolet EDX with Synergy Software
Common nameDiagnostic Electromyograph
Device ClassClass II
Classification nameEvoked Response Electrical Stimulator [primary]Diagnostic ElectromyographNerve Conduction Velocity Measurement DeviceNon-normalizing Quantitative Electroencephalograph SoftwareEvoked Response Auditory StimulatorEvoked Response Photic StimulatorEvoked Response Mechanical Stimulator
Product Code, RegulationGWF 21 CFR §882.1870 [primary]IKN 21 CFR §890.1375JXE 21 CFR §882.1550OLT 21 CFR §882.1400GWJ 21 CRF §882.1900GWE 21 CFR §882.1890GZP 21 CFR §882.1880

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

{1}------------------------------------------------

3.Predicate Devices510(K)NumberDevice Name
K112052CareFusion Nicolet EDX with Viking Software
K965065CareFusion Synergy Mobile System
K070109NeuroMetrix Advance System

Device Description 4.

The CareFusion Nicolet EDX system with Synergy Software (Synergy EDX) is designed for the acquisition, display, analysis, reporting, and management of electrophysiological information from the human nervous and muscular systems. The system is designed to perform nerve conduction studies (NCS), needle electromyography (EMG) testing, evoked potential (EP) testing, and intra-operative monitoring (IOM). Synergy EDX provides a variety of tests spanning the various modalities.

The Synergy EDX consists of the following major components:

  • Nicolet EDX console base unit; .
  • . Synergy control panel:
  • . Nicolet amplifier (there are two types available: 2 channel (AT2) with two non-switched amplifier channels and an 8 channel (AT2 + 6) amplifier with two non-switched and six switched amplifier channels;
  • . Desktop or laptop computer with a keyboard and mouse;
  • . Display monitor; and
  • . Synergy Software

The Synergy EDX optional accessories/components consists of the following:

  • Nicolet HB6 or HB7 Head Box .
  • . Stimulator probes (RS 10 probe, WR 50 Probe, S403 probe)
  • SP1/SP2 electrical stimulator switching units
  • . Footswitches (single and triple)
  • . LED goggles
  • Patient response button ●
  • . Photic strobe
  • Headphones or other auditory transducers .
  • . Cart
  • Isolation transformer . .
  • Printer ●

Description of Major Components

Nicolet EDX Console Base:

The console base is an AC mains powered electrical device that houses the core Nicolet EDX hardware, and provides interconnection capability with the rest of the Nicolet EDX hardware and PC. This is exactly the same as in the Nicolet EDX with Viking software system. The PC contains the software.

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

Page 2 of 14

{2}------------------------------------------------

Synergy Control Panel:

The control panel along with the mouse is the primary user interface for the Nicolet EDX system. The control panel contains a variety of controls that allow the user to access and use the Nicolet EDX system from the touch of a button or knob. This control panel performs similar functions as the Viking control panel from the Nicolet EDX with Viking software system.

Nicolet Amplifiers (AT2 and AT2 + 6):

The Nicolet amplifiers are DC powered electrical devices that record, amplify and transmit responses from the nerve and/or muscle to the Nicolet EDX console base. The AT2 amplifier collects up to two channels of neurophysiological information and the AT2 + 6 amplifier collects up to 8 channels of neurophysiological information. These are exactly the same amplifiers as in the Nicolet EDX with Viking software system.

Personal Computer (PC) with Keyboard and Mouse:

The Nicolet EDX system hardware is used in conjunction with a personal computer (PC), which is offered in either a desktop or laptop configuration. These are exactly the same PCs as in the Nicolet EDX with Viking software system. The Synergy software is supplied preloaded onto the PC.

Display Monitor:

The desktop PC version requires a display monitor which is provided with the Nicolet EDX system hardware. The laptop version does not require a separate display monitor beyond the built-in laptop display; however, an external display monitor may be used if desired. The additional display monitor is facilitated through the connection to an isolation transformer that is included with a desktop PC and available as an optional accessory for laptop based systems. These are exactly the same monitors and isolation transformer as in the Nicolet EDX with Viking software system.

Description of Optional Accessory Components

HB6 and HB7 Head Boxes:

The HB6 and HB7 Head Boxes are passive devices. These are exactly the same head boxes as in the Nicolet EDX with Viking software system. They can be connected to the AT2+6 Amplifier via a cable to allow the electrodes' receptacles to be in closer proximity to the patient.

Electrical Stimulator Probes:

There are three types of electrical stimulator probes available for use with the Nicolet EDX System Hardware: (1) the WR 50 comfort Plus Probe, (2) the RS 10 Comfort Probe and (3) the S403 Probe. These are exactly the same Stimulus Probes as in the Nicolet EDX with Viking software system. The probes connect to the Nicole EDX console base via a cable. Each probe contains a tip that facilitates direct stimulus contact to the patient when activated. The probes can deliver a stimulus ranging from 0 - 400 V / 0 - 100 mA.

SP-1/SP-2 Electrical Stimulus Switching Units:

The SP-1 and SP-2 Stimulus Switching Units allow the user to connect multiple sets of electrodes on the patient for stimulation at different locations. These are exactly the same Stimulus Switching Pods as in the Nicolet EDX with Viking software system.

Footswitch and Triple Footswitch:

The footswitches allow the user to activate defined functions such as electrical stimulation and

{3}------------------------------------------------

initiate acquisition of trace data. Pressing the footswitch activates or deactivates the user defined function. These are exactly the same footswitches and functions as in the Nicolet EDX with Viking software system.

Patient Response Button:

The patient response button allows the patient to respond to a rare stimulus during specific EP testing. This is exactly the same Patient Response Button and function as in the Nicolet EDX with Viking software system. The patient presses the button when the rare stimulus is detected. The Nicolet EDX senses the button press and increments a count.

Visual Stimulators:

  • (1) LED goggles are available to provide visual stimulation to the patient when the Visual Evoked Potential (VEP) software option is in use. These are exactly the same LED goggles and function as in the Nicolet EDX with Viking software system.
  • (2) A photic strobe is also a visual stimulus option. This is exactly the same photic strobe and function as in the Nicolet EDX with Viking software system. It can be mounted on an optional stand. Similar to the other visual stimulus; it is used with the VEP option. All three visual stimulators have no controls or indicators and are connected to the Nicolet EDX Console Base Unit. This photic strobe was previously cleared under K921927 and K991054.

Headphones or other Auditory Transducers:

Headphones are available, or other auditory transducers may be used to provide auditory stimulation to the patient through the transducer when the Auditory Evoked Potential (AEP) is in use. Headphones and transducers have no controls or indicators and are connected to the Nicolet EDX Console Base unit auditory stimulator output connectors. These are exactly the same transducers and function as in the Nicolet EDX with Viking software system.

Cart and Printer:

The metal cart provides a convenient way to contain all of the components of the Nicolet EDX System Hardware into one mobile location. The cart has lockable wheels and a convenient handle to facilitate movement of the cart. Two articulating arms are provided for mounting the display monitor and amplifier. There are multiple shelves present to accommodate the placement and storage of various Nicolet EDX System Hardware components and supplies. A retractable keyboard shelf and mouse pad is available. In terms of printing capabilities, a printer equivalent to a DeskJet or Laser printer is available for connection to the system to print reports or screen copies. The cart is exactly the same and functions the same as in the Nicolet EDX with Viking software system.

ક. Indications for Use:

The CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).

Evoked Potential (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300. Motor Evoked Potentials (MEP) and Contingent Negative Variation (CNV). The

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

{4}------------------------------------------------

Nicolet EDX with Synergy Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Synergy Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.

The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.

The Nicolet EDX with Synergy Software is intended to be used by a qualified healthcare provider.

Summary of Technical Characteristics Compared to the Predicate Devices

6

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

11

{5}------------------------------------------------

1. General - Indications
CharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
1.1 IndicationsFor UseThe CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management ofelectrophysiological information from the human nervous andmuscular systems including Nerve Conduction (NCS),Electromyography (EMG), Evoked Potentials (EP), AutonomicResponses and Intra-Operative Monitoring includingElectroencephalography (EEG).Evoked Potential (EP) includes Visual Evoked Potentials (VEP),Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials(SEP), Electroretinography (ERG), Electrooculography (EOG), P300,Motor Evoked Potentials (MEP) and Contingent Negative Variation(CNV). The Nicolet EDX with Synergy Software may be used todetermine autonomic responses to physiologic stimuli by measuringthe change in electrical resistance between two electrodes (GalvanicSkin Response and Sympathetic Skin Response). Autonomic testingalso includes assessment of RR Interval variability. The Nicolet EDXwith Synergy Software is used to detect the physiologic function ofthe nervous system, for the location of neural structures duringsurgery, and to support the diagnosis of neuromuscular disease orcondition.The listed modalities do include overlap in functionality. In general,Nerve Conduction Studies measure the electrical responses of thenerve; Electromyography measures the electrical activity of themuscle and Evoked Potentials measure electrical activity from theCentral Nervous System.The Nicolet EDX with Synergy Software is intended to be used by aqualified healthcare provider.The CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management ofelectrophysiological information from the human nervous andmuscular systems including Nerve Conduction (NCS),Electromyography (EMG), Evoked Potentials (EP), AutonomicResponses and Intra-Operative Monitoring includingElectroencephalography (EEG).Evoked Potential (EP) includes Visual Evoked Potentials (VEP),Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials(SEP), Electroretinography (ERG), Electrooculography (EOG), P300,and Motor Evoked Potentials (MEP). The Nicolet EDX with SynergySoftware may be used to determine autonomic responses tophysiologic stimuli by measuring the change in electrical resistancebetween two electrodes (Galvanic Skin Response and SympatheticSkin Response). Autonomic testing also includes assessment of RRInterval variability. The Nicolet EDX with Synergy Software is usedto detect the physiologic function of the nervous system, for thelocation of neural structures during surgery, and to support thediagnosis of neuromuscular disease or condition.The listed modalities do include overlap in functionality. In general,Nerve Conduction Studies measure the electrical responses of thenerve; Electromyography measures the electrical activity of themuscle and Evoked Potentials measure electrical activity from theCentral Nervous System.The Nicolet EDX with Viking Software is intended to be used by aqualified healthcare provider.Identical to the Nicolet EDX with Viking softwarewith the minor addition of the CNV test being a partof the Evoked Potential Modality.
1.2 WarningsItems related to off-label use or misuse.Items related to off-label use or misuse.Identical to the Nicolet EDX with Viking software.
1.3 Contra-indicationsItems related to design and indicated use limitations, such as, notfor use in the presence of flammable anesthetics or in conjunctionwith defibrillation equipmentItems related to design and indicated use limitations, such as, notfor use in the presence of flammable anesthetics or in conjunctionwith defibrillation equipment.Identical to the Nicolet EDX with Viking software
2. General - Design
CharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
2.1 General systems approachComputer based equipment with dedicated hardware peripherals / components.Computer based equipment with dedicated hardware peripherals / components.Identical to the Nicolet EDX with Viking software
2.2 User input deviceWindow mouse/keyboard driven graphic interface with dedicated control panel.Window mouse/keyboard driven graphic interface with dedicated control panel.Identical to the Nicolet EDX with Viking software.(See footnote 1)
2.3 User output deviceDigital color display and commercial printersDigital color display and commercial printersIdentical to the Nicolet EDX with Viking software
2.4 Patient inputs2 to 8 channel amplifier, isolated2 to 8 channel amplifier, isolatedIdentical to the Nicolet EDX with Viking software
2.5 Signal acquisitionAnalog to digital conversion at 48kHz sample rateAnalog to digital conversion at 48kHz sample rateIdentical to the Nicolet EDX with Viking software
2.6 Trigger input (synchronization to external events)YesYesIdentical to the Nicolet EDX with Viking software
2.7 Trigger output(synchronization for external devices)YesYesIdentical to the Nicolet EDX with Viking software
2.8 Footswitch for hands-free operationYesYesIdentical to the Nicolet EDX with Viking software
2.9 Use of standard software platform (Operating System)Yes Microsoft WindowsYes Microsoft WindowsIdentical to the Nicolet EDX with Viking software
2.10 Customization of protocolsVia storage / retrieval of user defined settingsVia storage / retrieval of user defined settingsIdentical to the Nicolet EDX with Viking software
2.11 Application flexibility / expandabilityVia software updateVia software updateIdentical to the Nicolet EDX with Viking software
2.12 Safety StandardsEN/IEC 60601-1:1998 + A1:1991+A2:1995,IEC 60601-1-1:2000EN/IEC 60601-1-2: Ed. 2.0+A1:2004IEC 60601-2-40: 1998, Ed:1,UL 60601-1: 2003-04-25 ED1 Rev:2003/06/30,CAN/CSA-C22.2 по. 601.1-М90 Issue: 1990/01/01Rev:2003/11EN/IEC 60601-1:1998 + A1:1991+A2:1995,IEC 60601-1-1:2000EN/IEC 60601-1-2: Ed. 2.0+A1:2004IEC 60601-2-40: 1998, Ed:1,UL 60601-1: 2003-04-25 ED1 Rev:2003/06/30,CAN/CSA-C22.2 по. 601.1-M90 Issue: 1990/01/01Rev:2003/11Identical to the Nicolet EDX with Viking software
CharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
2.13 Patient circuitry isolationEuropean Community (CE Mark)Optic/transformerEuropean Community (CE Mark)Optic/transformerIdentical to the Nicolet EDX with Viking software
2.14 System componentsEDX base console including 2 electrical stimulators, auditorystimulator, trigger input/output, LED goggle interface;EDX base console including 2 electrical stimulators, auditorystimulator, trigger input/output, LED goggle interface;Identical to the Nicolet EDX with Viking software
Control panel;Amplifier;Control panel;Amplifier;
Computer, monitor, keyboard, mouse, printerComputer, monitor, keyboard, mouse, printer
2.15 System - computer interfaceUSBUSBIdentical to the Nicolet EDX with Viking software
2.16 System power supplyMains (100 – 240 VAC) thru an isolation transformerMains (110 – 240 VAC) thru an isolation transformerIdentical to the Nicolet EDX with Viking software
2.17 Amplifier power supply15 VDC from base console15 VDC from base consoleIdentical to the Nicolet EDX with Viking software
2.18 Size (L/W/D) cm35.6 x 34.3 x 8.6 (base console)35.6 x 34.3 x 8.6 (base console)Identical to the Nicolet EDX with Viking software
2.19 Weight kg3.5 (base console)3.5 (base console)Identical to the Nicolet EDX with Viking software
Design - AcquisitionCharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
3.1Number of channels2 to 82 to 8Identical to the Nicolet EDX with Viking software
3.2CMMR> 110 dB> 110 dBIdentical to the Nicolet EDX with Viking software
3.3Noise< 0.6uV RMS (from 2 Hz to 10 kHz)< 0.6uV RMS (from 2 Hz to 10 kHz)Identical to the Nicolet EDX with Viking software
3.4Input impedance>1000 MΩ>1000 MΩIdentical to the Nicolet EDX with Viking software
3.5Low Filter0.05 Hz to 5 kHz0.05 Hz to 5 kHzIdentical to the Nicolet EDX with Viking software
3.6High filter30 Hz to 20 kHz30 Hz to 20 kHzIdentical to the Nicolet EDX with Viking software
3.7Notch filter50 / 60 selectable50 / 60 selectableIdentical to the Nicolet EDX with Viking software
3.8A/D conversion24 bit24 bitIdentical to the Nicolet EDX with Viking software
3.9Sampling rate (cumulative)384 kHz384 kHzIdentical to the Nicolet EDX with Viking software
3.10Time base range0.01 to 5000 ms0.01 to 5000 msIdentical to the Nicolet EDX with Viking software
3.11Number of Time bases allowedMultipleMultipleIdentical to the Nicolet EDX with Viking software
3.12Trigger modeFree run, internal, externalFree run, internal, externalIdentical to the Nicolet EDX with Viking software
3.13Signal delay (pre/post)-3000 to +500 ms-3000 to +500 msIdentical to the Nicolet EDX with Viking software
3.14Impedance meter500 Ω to 480 kΩ500 Ω to 480 kΩIdentical to the Nicolet EDX with Viking software

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

.

Page 6 of 14

:

{6}------------------------------------------------

. ·

.

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

·

Page 7 of 14

e 7 of 14

:

.

{7}------------------------------------------------

Footnote 1: The Synergy control panel is functionally equivalent to the Viking control panel but the number of controls differs.

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

.

.

Page 8 of 14

{8}------------------------------------------------

·

:

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

Page 9 of 14

i

· .

:

·

.

:

:

:

' .

{9}------------------------------------------------

Stimulators
-
Design
4.

·

CharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
1 Electrical Stimulator
1.1 TypeConstant Current or Constant VoltageConstant Current or Constant VoltageIdentical to the Nicolet EDX with Viking software
1.2 Number1 or 21 or 2Identical to the Nicolet EDX with Viking software
1.3 Maximum Output100mA or 400V100mA or 400VIdentical to the Nicolet EDX with Viking software
1.4 Duration0.01 to 1 ms0.01 to 1 msIdentical to the Nicolet EDX with Viking software
1.5 ModeSingle or TrainSingle or TrainIdentical to the Nicolet EDX with Viking software
1.6 BiphasicYesYesIdentical to the Nicolet EDX with Viking software
2 Auditory Stimulator
2.1 TypeClick, Pip, BurstClick, Pip, BurstIdentical to the Nicolet EDX with Viking software
2.2 Intensity0 to 139 dB pSPL0 to 139 dB pSPLIdentical to the Nicolet EDX with Viking software
2.3 PolarityCondensation, Rarefaction, AlternatingCondensation, Rarefaction, AlternatingIdentical to the Nicolet EDX with Viking software
2.4 Tone Frequency125 to 8000 Hz250 to 8000 HzSimilar to the Nicolet EDX with Viking softwarewith the addition of 125 Hz frequency. Theadditional frequency does not change theindications for use or safety and effectiveness.The Synergy predicate (K965065) has 125 Hzfrequency.
2.5 Click Duration0.05 to 1 ms0.05 to 1 msIdentical to the Nicolet EDX with Viking software
2.6 SideLeft, Right, BothLeft, Right, BothIdentical to the Nicolet EDX with Viking software
2.7 TransducersTDH 39, TIP 300, Bone VibratorTDH 39, TIP 300, Bone VibratorIdentical to the Nicolet EDX with Viking software

ﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

: .

:

Page 10 of 14

{10}------------------------------------------------

Modules
Application
EMGട.

:

·

CharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
6.1 Free Run AcquisitionYesYesIdentical to the Nicolet EDX with Viking software
6.2 Nerve Conduction Study (NCS)YesYesIdentical to the Nicolet EDX with Viking software
6.3 Stimulator TriggeredYesYesIdentical to the Nicolet EDX with Viking software
6.4 Signal Triggered AcquisitionYesYesIdentical to the Nicolet EDX with Viking software
6.5 Spontaneous Activity (SPA)YesYesIdentical to the Nicolet EDX with Viking software
6.6 Single Fiber EMG (SFEMG)YesYesIdentical to the Nicolet EDX with Viking software
6.7 Motor Unit AnalysisYesYesIdentical to the Nicolet EDX with Viking software
6.8 F-WaveYesYesIdentical to the Nicolet EDX with Viking software
6.9 H Reflex (H-Wave)YesYesIdentical to the Nicolet EDX with Viking software
6.10 Sympathetic Skin Response (SSR)YesYesIdentical to the Nicolet EDX with Viking software

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

:

: :

Page 11 of 14

{11}------------------------------------------------

Evoked Potential Application ModulesCharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
1 Somatosensory EP (SEP)YesYesIdentical to the Nicolet EDX with Viking software
2 Auditory EP (AEP)YesYesIdentical to the Nicolet EDX with Viking software
3 Visual EP (VEP)YesYesIdentical to the Nicolet EDX with Viking software
4 P300YesYesIdentical to the Nicolet EDX with Viking software
5 ERGYesYesIdentical to the Nicolet EDX with Viking software
6 EOGYesYesIdentical to the Nicolet EDX with Viking software
7 LED Goggles/Photic (Visual EP Flash)YesYesIdentical to the Nicolet EDX with Viking software
IOM Application ModulesCharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Nicolet EDX with Viking(K112052)Discussion of Differences
1 IOMYesYesIdentical to the Nicolet EDX with Viking software
2 Processed EEGYesYesIdentical to the Nicolet EDX with Viking software
3 MEPYesYesIdentical to the Nicolet EDX with Viking software
3. New FeaturesCharacteristicCareFusion 209, Inc.Nicolet EDX with Synergy System(This Submission)CareFusion 209, Inc.Synergy System(formerly Oxford Instruments Medical)(K965065)NeuroMetrix Advance(K070109)Discussion of Differences

Page 12 of 14 510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979

ldentical to the NeuroMetrix predicate dentical to the NeuroMetrix predicate

dentical to the Synergy predicati

No Yes Yes

.

Yes No No

Yes Yes Yes

8.2 Automatic Report Narrative Generation

8.1CNV

8.3 Electrical Stimulus Automation

{12}------------------------------------------------

Summary of Non-Clinical Performance Testing Conducted for the Determination of Substantial Equivalence

Biocompatibility:

7

CareFusion has demonstrated the biocompatibility of all direct and indirect patient contacting material associated with the CareFusion Nicolet EDX (Synergy EDX) through compliance with ISO 10993-1: 2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Test results have indicated that contacting materials complies with the standard and are safe for its intended use. There are no different materials than in the Nicolet EDX with Viking Software.

Software testing:

The Synergy EDX contains MODERATE level of concern software. The software was designed and developed according to a robust software development process, and was rigorously verified and validated consistent with the following guidelines:

  • FDA guidance: The content of premarket submissions for software contained in medical ● devices, 11 May 05;
  • . FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; and
  • . FDA guidance: General principles of software validation: Final guidance for industry and FDA staff, 11 Jan 02.

The tests results demonstrated that the Synergy EDX complies with its predetermined specifications.

Electrical Safety and Electromagnetic Compatibility (EMC) Testing:

The Synergy EDX was tested for electrical safety. Test results demonstrated that the Synergy EDX complies with the following standards:

  • IEC 60601-1: 1988, Am1: 1991, Am2: 1995, Medical Electrical Equipment, Part 1: . General Requirements for Safety; and
  • UL 60601-1: 2006, Medical Electrical Equipment, Part 1: Particular Requirements for . Safety.

Electromagnetic Compatibility (EMC) testing was conducted on the Synergy EDX according to the applicable standard. Test results indicated that the system complies with the following:

  • . IEC 60601-1-2: 2001, Am1: 2004, Medical Electrical Equipment, Part 1: Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests.

Performance Testing - Bench:

The Synergy EDX was tested to assess performance in accordance with requirements of the applicable performance standard. Test results demonstrated that the Synergy EDX met specifications and complies with the following standard:

  • IEC 60601-2-40: 1998, Medical Electrical Equipment, Part 2-40: Particular Requirements . for the Safety of Electromyographs and Evoked Response Equipment.

Performance Testing - Animal &Clinical:

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979]

14

{13}------------------------------------------------

Animal testing and clinical testing were not needed to demonstrate safety and effectiveness.

8

·

Conclusion

The technological characteristics and performance data for the CareFusion Nicolet EDX System . with Synergy Software demonstrates that it is substantially equivalent to the predicate devices. .

510(k) Summary, CareFusion Nicolet EDX with Synergy Software [K120979] 14

Page 14 of

{14}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is simple, using a black and white color scheme.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AFR 2 5 2012

CareFusion 209, Inc. c/o Mr. Curtis D. Truesdale Manager, Regulatory Assurance 1850 Deming Way Middleton, WI 53562

Re: K120979

Trade/Device Name: CareFusion Nicolet EDX with Synergy Software Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF, IKN, JXE, OLT, GWJ, GWE, GZP Dated: March 28, 2012 Received: April 2, 2012

Dear Mr. Truesdale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{15}------------------------------------------------

Page 2 - Mr. Curtis D. Truesdale

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

E.e. Then MD for

Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K12 0979

Device Name: CareFusion Nicolet EDX with Synergy Software

Indications for Use:

The CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).

Evoked Potential (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300, Motor Evoked Potentials (MEP) and Contingent Negative Licettoooalography (2007) Food) x with Synergy Software may be used to determine autonomic resnonses to physiologic stimuli by measuring the change in electrical resistance between two relectrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Synergy Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.

The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical mousal of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.

The Nicolet EDX with Synergy Software is intended to be used by a qualified healthcare provider.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

KRISTEN BOWSHER

(Division Sign-Off)

(Division Sign-Orr)
Division of Ophthalmic, Neurological and Ear,

Nose and Throat Devices

510(k) Number_k120979

§ 882.1870 Evoked response electrical stimulator.

(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).