Search Results
Found 43 results
510(k) Data Aggregation
(83 days)
Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini B
Regulation Number: 21 CFR 866.3990
Panel 2
QIAstat-Dx GI Panel 2 Mini B&V
QIAstat-Dx GI Panel 2 Mini B
Classification: 21 CFR 866.3990
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
**Similarities
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
**Similarities
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
**Similarities
QIAstat-Dx Gastrointestinal Panel 2:
The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2:
- Adenovirus F40/F41
- Astrovirus
- Norovirus GI/GII
- Rotavirus A
- Campylobacter (C. jejuni, C. coli and C. upsaliensis)
- Shigella/Enteroinvasive Escherichia coli (EIEC)
- Enteropathogenic Escherichia coli (EPEC)
- Enterotoxigenic Escherichia coli (ETEC) lt/st
- Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC)
- Salmonella
- Plesiomonas shigelloides
- Yersinia enterocolitica
- Cryptosporidium
- Cyclospora cayetanensis
- Entamoeba histolytica
- Giardia lamblia*
*Also known as Giardia intestinalis and Giardia duodenalis
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B&V:
The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V:
- Norovirus
- Campylobacter
- Shigella
- Shiga-like toxin E. coli (STEC)
- Salmonella
Concomitant culture is necessary for organism recovery and further typing of bacterial agents.The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B:
The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B:
- Campylobacter
- Shigella
- Shiga-like toxin Escherichia coli (STEC)
- Salmonella
- Yersinia enterocolitica
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B ("QIAstat-Dx Gastrointestinal Panels") are multiplexed nucleic acid tests which are designed for use with the QIAstat-Dx Analyzer 2.0. The device modification is to add the QIAstat-Dx Rise as an additional instrument for use with the QIAstat-Dx Gastrointestinal Panels. The QIAstat-Dx Rise is a higher throughput platform, incorporating up to eight QIAstat-Dx Analytical Modules (AM) on a small footprint. The instrument allows queuing up to 18 cartridges, with processing being limited up to 16 cartridges for the QIAstat-Dx Gastrointestinal Panels. The cartridges are scheduled for processing and delivered to the appropriate AM by an integrated robotic handler. The AM used with the QIAstat-Dx Rise is the same AM that can be used with the QIAstat-Dx Analyzer 2.0.
The modified QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are identical to the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B (K252329), respectively, with the exception of the Instructions for Use which were updated to include the assay-specific procedure for use with the QIAstat-Dx Rise.
The QIAstat-Dx Gastrointestinal Panels are intended for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple targets directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection.
All the reagents required for the complete execution of the test are pre-loaded and self-contained in the QIAstat-Dx Gastrointestinal Panels. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators. This eliminates any possibility of exposure of the user or the analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber.
The QIAstat-Dx Gastrointestinal Panels Assay Definition File (ADF) automatically interprets test results and displays a summary on the instrument display screen. The detected analytes are displayed in red. All other tested but not detected analytes are listed in green. The instrument will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").
N/A
Ask a specific question about this device
(226 days)
California 94089
Re: K251721
Trade/Device Name: Xpert GI Panel
Regulation Number: 21 CFR 866.3990
transcription polymerase chain reaction (RT-PCR) and detection test
- Regulation number: 21 CFR 866.3990
| Xpert® GI Panel | BioFire FilmArray GI Panel (K140407) |
| Regulation | Same | 21 CFR 866.3990
The Xpert GI Panel, performed on the GeneXpert® Instrument Systems, is a qualitative multiplexed in vitro diagnostic test that is capable of the simultaneous detection and identification of DNA and RNA from multiple bacteria, parasites and/or virus directly from stool samples in Cary Blair transport media obtained from individuals with signs and symptoms of gastrointestinal infection. The test utilizes automated, qualitative real time polymerase chain reaction (PCR). The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and virus are identified using the Xpert GI Panel:
Pathogens Detected:
- Bacteria: Campylobacter (C. jejuni/C. coli), Shiga toxin-producing Escherichia coli (STEC) stx1/stx2, Salmonella, Shigella/Enteroinvasive Escherichia coli, Yersinia enterocolitica, Vibrio parahaemolyticus, Vibrio cholerae
- Parasites: Giardia (also known as G. intestinalis, G. duodenalis & G. lamblia), Cryptosporidium
- Virus: Norovirus GI/GII
Pathogens Reported:
- Bacteria: Campylobacter, STEC stx1, STEC stx2, Salmonella, Shigella EIEC, Yersinia, V. parahaemolyticus, V. cholerae
- Parasites: Giardia, Cryptosporidium
- Virus: Norovirus
Results are meant to be used in conjunction with other clinical, laboratory and epidemiological data and should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive results do not rule out co-infection with pathogens not included in the Xpert GI Panel. The pathogen detected may not be the definite cause of the disease. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The Xpert Gastrointestinal (GI) Panel test, performed on Cepheid GeneXpert® Instrument Systems equipped with 10-color modules, is an automated in vitro diagnostic test for qualitative detection and differentiation of eleven (11) pathogens in stool in Cary Blair specimens collected from individuals suspected of gastrointestinal infection (GI). The results from the Xpert GI Panel test will be available in approximately 74 minutes.
The Xpert GI Panel is performed on the Cepheid GeneXpert® Instrument Systems equipped with GeneXpert 10-color modules running software version 6.4 and higher (GeneXpert Dx), software version Xpertise 7.1 or higher (GeneXpert Infinity) or software version COS 2.1 or higher (GeneXpert with Touchscreen). The GeneXpert® Instrument Systems automate and integrate sample purification, nucleic acid amplification, and detection of the target sequences from clinical specimens using reverse transcription (conversion of RNA template into DNA) followed by real-time polymerase chain reaction (real-time PCR) and melt curve analysis. The systems consist of an instrument, computer, and preloaded software for running tests and viewing the results. Each test requires the use of a single-use disposable GeneXpert cartridge that contains target-specific reagents and carries out the reverse transcription and PCR processes.
The Xpert GI Panel test includes reagents needed: (i) for sample preparation and (ii) to detect the different bacteria, parasites, and virus. A Sample Processing Control (SPC), an Internal Control (IC) and a Probe Check Control (PCC) are also included in the cartridge. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability. The IC is present to ensure adequate processing of RNA targets and monitor the presence of inhibitor(s) in the PCR reaction. The SPC is present to control for adequate extraction and processing of the target sequences in the PCR reaction. The SPC also acts as a control for functionality of melt curve analysis.
The Xpert GI Panel test results are interpreted by GeneXpert Instrument Systems from measured fluorescent signals and embedded calculation algorithms and will be reported as positive or negative for each of the targets.
N/A
Ask a specific question about this device
(266 days)
BD COR System, and Enteric Bacterial Panel Diluent for BD COR System
Regulation Number: 21 CFR 866.3990
BD EBP for BD COR™ System
Page 7
Regulatory Information
Regulation section: 21 CFR 866.3990
BD Enteric Bacterial Panel for BD COR™ System
BD Enteric Bacterial Panel for BD COR™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens. BD Enteric Bacterial Panel for BD COR™ System detects nucleic acids from:
- Salmonella spp.
- Campylobacter spp. (C. jejuni and C. coli)
- Shigella spp. / Enteroinvasive Escherichia coli (EIEC)
- Shiga toxin 1 (stx1) / Shiga toxin 2 (stx2) genes (found in Shiga toxin-producing Escherichia coli [STEC]) as well as Shigella dysenteriae, which can possess a Shiga toxin gene (stx) that is identical to the stx1 gene of STEC.
Testing is performed utilizing the BD Fecal Collection and Transport Kit. Unpreserved soft to diarrheal stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis are probed using the flocked swab and material is transferred to the BD Fecal Collection and Transport tube containing modified Cary-Blair medium. The test is performed utilizing real-time polymerase chain reaction (PCR) for the amplification of specific sequences of target DNA. The test utilizes fluorogenic sequence-specific hybridization probes for detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Salmonella, Shigella/EIEC, Campylobacter, and Shiga toxin-producing E. coli (STEC). Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
BD Enteric Bacterial Panel plus for BD COR™ System
BD Enteric Bacterial Panel plus for BD COR™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens. BD Enteric Bacterial Panel plus for BD COR™ System detects nucleic acids from:
- Salmonella spp.
- Campylobacter spp. (C. jejuni and C. coli)
- Shigella spp. / Enteroinvasive Escherichia coli (EIEC)
- Shiga toxin 1 (stx1) / Shiga toxin 2 (stx2) genes (found in Shiga toxin-producing Escherichia coli [STEC]) as well as Shigella dysenteriae, which can possess a Shiga toxin gene (stx) that is identical to the stx1 gene of STEC.
- Plesiomonas shigelloides
- Vibrio spp. (V. vulnificus, V. parahaemolyticus, and V. cholerae)
- Enterotoxigenic Escherichia coli (ETEC) heat-labile enterotoxin (LT) / heat-stable enterotoxin (ST) genes
- Yersinia enterocolitica
Testing is performed utilizing the BD Fecal Collection and Transport Kit. Unpreserved soft to diarrheal stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis are probed using the flocked swab and material is transferred to the BD Fecal Collection and Transport tube containing modified Cary-Blair medium. The test is performed utilizing real-time polymerase chain reaction (PCR) for the amplification of specific sequences of target DNA. The test utilizes fluorogenic sequence-specific hybridization probes for detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Salmonella, Shigella/EIEC, Campylobacter, Shiga toxin-producing E. coli (STEC), Plesiomonas shigelloides, Vibrio spp. (V. vulnificus, V. parahaemolyticus, and V. cholerae), Enterotoxigenic Escherichia coli (ETEC) LT/ST, and Yersinia enterocolitica. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
Enteric Bacterial Panel Diluent for BD COR™ System
The Enteric Bacterial Panel Diluent for BD COR™ System is intended to be used in clinical settings according to instructions provided for aliquoting into Molecular Aliquot Tubes by the BD COR™ System. The Enteric Bacterial Panel Diluent for BD COR™ System is only for use with BD Fecal Collection and Transport Kit specimens tested on BD COR™ Systems.
The BD Enteric Bacterial Panel for BD COR™ System (BD EBP for BD COR™ System) simultaneously detects pathogens responsible for gastroenteritis due to Salmonella spp., Campylobacter spp. (C. jejuni and C. coli), Shigella spp./EIEC, stx/stx1/stx2 found in Shiga toxin-producing E. coli and in Shigella dysenteriae, and an internal Sample Processing Control. The BD Enteric Bacterial Panel plus for BD COR™ Systems (BD EBP plus for BD COR™ System) additionally detects Plesiomonas shigelloides, Vibrio (V. vulnificus, V. parahaemolyticus, and V. cholerae), Enterotoxigenic E. coli (ETEC) LT/ST, Yersinia enterocolitica and an internal Sample Processing Control with a second master mix. The assays automate the testing process and minimize operator intervention from the time the sample is placed onto the BD COR™ System until results are available.
The BD COR™ System is comprised of a single BD COR™ PX System attached to a BD COR™ MX Analyzer as described in K210585 and K224653. Once the specimens are loaded, the BD COR™ PX System performs the necessary pre-analytical steps such as vortexing, aliquoting into a diluent filled molecular aliquot tube, sorting/grouping of the secondary samples for testing by assay, pre-warming and cooling of the sample (where required), and transport of the sample into the BD COR™ MX Analyzer, where extraction, amplification and detection take place.
Additionally, the steps of ordering tests on the instrument for specific samples will be managed directly by the user interaction with the Laboratory Information System (LIS), which communicates directly with the instrument.
Once the clinical specimens are received in the laboratory and loaded into the transport racks, the user will not be required to directly handle the specimen again prior to result reporting and removal from the system.
N/A
Ask a specific question about this device
(89 days)
Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini B
Regulation Number: 21 CFR 866.3990
Panel 2
QIAstat-Dx GI Panel 2 Mini B&V
QIAstat-Dx GI Panel 2 Mini B
Classification: 21 CFR 866.3990
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
Similarities
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
Similarities
| 21 CFR 866.3990 |
| Product Code | PCH | PCH |
| Device Class | Class II | Class II |
Similarities
QIAstat-Dx Gastrointestinal Panel 2
The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2:
- Adenovirus F40/F41
- Astrovirus
- Norovirus GI/GII
- Rotavirus A
- Campylobacter (C. jejuni, C. coli and C. upsaliensis)
- Shigella/Enteroinvasive Escherichia coli (EIEC)
- Enteropathogenic Escherichia coli (EPEC)
- Enterotoxigenic Escherichia coli (ETEC) lt/st
- Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC)
- Salmonella
- Plesiomonas shigelloides
- Yersinia enterocolitica
- Cryptosporidium
- Cyclospora cayetanensis
- Entamoeba histolytica
- Giardia lamblia*
*Also known as Giardia intestinalis and Giardia duodenalis
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B&V
The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V:
- Norovirus
- Campylobacter
- Shigella
- Shiga-like toxin E. coli (STEC)
- Salmonella
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B
The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0. for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B:
- Campylobacter
- Shigella
- Shiga-like toxin Escherichia coli (STEC)
- Salmonella
- Yersinia enterocolitica
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are multiplexed nucleic acid tests that are designed for use with the QIAstat-Dx Analyzer 2.0. Modification made to the devices is to:
- Unmask the results for Shiga-like toxin E. coli (STEC) stx1/stx2, E. coli O157, and Enteropathogenic E. coli (EPEC), for the FecalSwab sample type for the QIAstat-Dx Gastrointestinal Panel 2,
- Unmask the results for Shiga-like toxin E. coli (STEC) for the FecalSwab sample type for the QIAstat-Dx GI Panel 2 Mini B&V and the QIAstat-Dx GI Panel 2 Mini B, and
- Remove the QIAstat-Dx Analyzer 1.0 from the indicated amplification and detection instrument systems for all three (3) devices, leaving only the QIAstat-Dx Analyzer 2.0. The QIAstat-Dx Analyzer 1.0 and the QIAstat-Dx Analyzer 2.0 belong to the same instrument family.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are intended for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple targets directly from preserved stool samples (Para-Pak® C&S or FecalSwab™) obtained from individuals with signs and/or symptoms of gastrointestinal infection.
The QIAstat-Dx system is based on single-test cartridges with pre-packaged reagents including both wet and dry chemistry to handle the sample preparation and detection steps for the presence of a range of selected analytes by PCR technology. After insertion of the sample, the QIAstat-Dx assay cartridge is processed by the QIAstat-Dx Analyzer 2.0.
Once the cartridge has been inserted into the instrument, the test starts automatically and runs for approximately 80 minutes. When the test is finished, the cartridge is removed by the user and discarded. The QIAstat-Dx Analyzer 2.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in gray if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail, and no results will be provided (screen will show "FAIL").
All the reagents required for the complete execution of the test are pre-loaded and self-contained in the QIAstat-Dx Gastrointestinal Panels. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators. This eliminates any possibility of exposure of the user or the analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber.
N/A
Ask a specific question about this device
(90 days)
Trade/Device Name: Panther Fusion GI Expanded Bacterial Assay
Regulation Number: 21 CFR 866.3990
** Gastrointestinal microorganism multiplex nucleic acid-based assay
Regulation Number: 21 CFR 866.3990
/Product code** | 21CFR 866.3990PCH, OOI | 21CFR 866.3990PCH, OOI |
| Intended use | The
The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.
The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.
Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.
Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).
Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction. Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.
The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:
| Analyte | Gene Targeted | Instrument Channel |
|---|---|---|
| Yersinia enterocolitica | InvA (Invasive antigen A) | FAM |
| Vibrio parahaemolyticus | gyrB (Gyrase B) | HEX |
| Vibrio vulnificus | gyrB (Gyrase B) | HEX |
| Vibrio cholerae | ompW (Outer Membrane Protein W) | HEX |
| Escherichia coli O157 | rfbE (Perosamine synthase-O-antigen) | ROX |
| Plesiomonas shigelloides | hugA (Heme utilization gene A) | RED647 |
| Internal Control | Not Applicable | RED677 |
Assay Components: The assay components configuration for the Panther Fusion GI Expanded Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Expanded Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Expanded Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:
- Aptima Assay Fluids Kit (303014)
- Aptima Multitest Swab Specimen Collection Kit (PRD-03546)
Table 1: Reagents Required to Perform the Panther Fusion GI Expanded Bacterial Assay
| Box | Components Description |
|---|---|
| Refrigerated Box | Panther Fusion GI Expanded Bacterial Assay Cartridges |
| Room Temperature Box | Panther Fusion Extraction Reagent-B• Panther Fusion Capture Reagent-B bottles• Panther Fusion Enhancer Reagent-B bottles |
| Refrigerated Box | Panther Fusion Internal Control-B |
| Room Temperature Box | Panther Fusion Reconstitution Buffer I |
| Room Temperature Box | Panther Fusion Elution Buffer |
| Room Temperature Box | Panther Fusion Oil |
| Refrigerated Box | Panther Fusion GI Expanded Bacterial Assay Controls• Panther Fusion GI Expanded Bacterial Positive Control• Panther Fusion Negative Control |
Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Expanded Bacterial Assay
| Aptima Assay Fluids Kit |
|---|
| Aptima Multitest Swab Specimen Collection Kit |
Instrumentation: The Panther Fusion GI Expanded Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay.
N/A
Ask a specific question about this device
(99 days)
K251868**
Trade/Device Name: Panther Fusion GI Bacterial Assay
Regulation Number: 21 CFR 866.3990
** Gastrointestinal microorganism multiplex nucleic acid-based assay
Regulation Number: 21 CFR 866.3990
/Product code** | 21CFR 866.3990 PCH, OOI | 21CFR 866.3990 PCH, OOI |
| Intended use | The FilmArray
The Panther Fusion GI Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Salmonella, Shigella/Enteroinvasive Escherichia coli (EIEC), Campylobacter (C. coli, C. jejuni) nucleic acids and Shiga-toxin producing Escherichia coli Shiga toxins 1 and 2 (undifferentiated) genes. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
This assay is intended to aid in the differential diagnosis of Salmonella, Campylobacter, Shigella/Enteroinvasive E. coli (EIEC) and Shigatoxigenic Escherichia coli (STEC) infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.
The Panther Fusion GI Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test developed for use on the fully automated Panther Fusion system to detect nucleic acids from Salmonella, Shigella/Enteroinvasive Escherichia coli (EIEC), Campylobacter (C. coli, C. jejuni) and Shiga-toxin producing Escherichia coli Shiga toxins 1 and 2 (undifferentiated) genes.
The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.
Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.
Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).
Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction. Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.
The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:
| Analyte | Gene Targeted | Instrument Channel |
|---|---|---|
| Salmonella | InvA (Invasive antigen A) | FAM |
| Campylobacter | glyA (serine hydroxymethyl transferase)/cadF (outer membrane fibronectin-binding protein) | HEX |
| Shigella/EIEC | ipaH (Invasion plasmid antigen H) | ROX |
| STEC | stx1 (Shigatoxin 1)/stx2 (Shigatoxin 2) | RED647 |
| Internal Control | Not Applicable | RED677 |
Assay Components
The assay components configuration for the Panther Fusion GI Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:
- Aptima Assay Fluids Kit (303014)
- Aptima Multitest Swab Specimen Collection Kit (PRD-03546)
Table 1: Reagents Required to Perform the Panther Fusion GI Bacterial Assay
| Box | Components Description |
|---|---|
| Refrigerated Box | Panther Fusion GI Bacterial Assay Cartridges |
| Room Temperature Box | Panther Fusion Extraction Reagent-B |
| Refrigerated Box | Panther Fusion Internal Control-B |
| Room Temperature Box | Panther Fusion Reconstitution Buffer I |
| Room Temperature Box | Panther Fusion Elution Buffer |
| Room Temperature Box | Panther Fusion Oil |
| Refrigerated Box | Panther Fusion GI Bacterial Assay Controls |
Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Bacterial Assay
| Kit |
|---|
| Aptima Assay Fluids Kit |
| Aptima Multitest Swab Specimen Collection Kit |
Instrumentation
The Panther Fusion GI Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Bacterial Assay.
N/A
Ask a specific question about this device
(23 days)
08028 Spain
Re: K250324
Trade/Device Name: QIAstat-Dx GI Panel 2 Mini B Regulation Number: 21 CFR 866.3990
| | |
| Classification: | 21 CFR 866.3990
|
| Regulation | 21 CFR 866.3990
| 21 CFR 866.3990
The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the OIAstat-Dx Analyzer 1.0 for the simultaneous in vitro qualitative detection of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B:
- Campylobacter
- Shigella
- Shiga-like toxin Escherichia coli (STEC)*
- Salmonella
- Yersinia enterocolitica
*Only with Para-Pak C&S, not reported for FecalSwab
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QlAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Postive results do not rule out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx® GI Panel 2 Mini B (Cat. no. 691423) assay is a modified device (reduced version) of the OIAstat-Dx Gastrointestinal Panel 2 (Cat. no. 691421). The OIAstat-Dx GI Panel 2 Mini B is identical to the QIAstat-Dx Gastrointestinal Panel 2 (K220062) but uses an Assay Definition File (ADF) which masks all but five pathogens (targets) from the OIAstat-Dx Gastrointestinal Panel 2. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the OIAstat-Dx GI Panel 2 Mini B: Campvlobacter. Shigella, Shiga-like toxin E. coli (STEC), Salmonella and Yersinia enterocolitica. The QIAstat-Dx GI Panel 2 Mini B is part of the QIAstat-Dx system and works with the OIAstat-Dx Analyzer 1.0. It will be available in a separately labeled kit.
The QIAstat-Dx GI Panel 2 Mini B is intended to be used with stool samples in Para-Pak C&S or FecalSwab transport media.
QIAstat-Dx is based on single-test cartridges with pre-packaged reagents including both wet and dry chemistry to handle the sample preparation and detection steps for the presence of a range of selected analytes by PCR technology. After insertion of the sample, the QIAstat-Dx assay cartridge is processed by the QIAstat-Dx Analyzer 1.0.
Once the cartridge is inserted into the instrument, the test starts automatically and runs for about 78 minutes. When the test is finished, the cartridge is removed by the user and discarded. The QIAstat-Dx Analyzer 1.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in grav if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").
All the reagents required for the complete execution of the test are pre-loaded and selfcontained in the QIAstat-Dx GI Panel 2 Mini B cartridge. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber (TC) to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample Pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber.
The provided text does not contain detailed acceptance criteria or a study that explicitly proves the device meets those criteria in a format with numerical results and statistical analyses typical for such studies. It states that the performance data for the QIAstat-Dx GI Panel 2 Mini B is equivalent to the predicate device (QIAstat-Dx Gastrointestinal Panel 2, K220062) for the five specific analytes it detects. It also mentions that the "QIAGEN QIAstat-Dx GI Panel 2 Mini B Instructions for Use" should be consulted for performance tables, which are not included in this document.
However, based on the information provided, I can infer the general nature of the acceptance criteria (qualitative detection of specific pathogens) and describe the comparative nature of the "study" (equivalence to a predicate device).
Here's an attempt to structure the information based on your request, with significant caveats that detailed numerical results, sample sizes, expert qualifications, and study methodologies for the actual performance are not present in the provided text.
Acceptance Criteria and Device Performance for QIAstat-Dx GI Panel 2 Mini B
The QIAstat-Dx GI Panel 2 Mini B is a modified version of the predicate device, QIAstat-Dx Gastrointestinal Panel 2 (K220062). The device performance for the QIAstat-Dx GI Panel 2 Mini B is presented as being equivalent to the predicate device for the five target analytes it identifies.
1. Table of Acceptance Criteria and Reported Device Performance
Since specific numerical acceptance criteria and performance metrics (e.g., sensitivity, specificity, PPV, NPV) are not provided in this document, the table below reflects the qualitative nature of the information given. The "acceptance criteria" are inferred from the demonstrated performance equivalence to the predicate device for the detected analytes.
| Analyte (Bacteria) | Acceptance Criteria (Inferred) | Reported Device Performance (Summary) |
|---|---|---|
| Campylobacter | Qualitative detection equivalent to predicate device (K220062) | Performance equivalent to QIAstat-Dx Gastrointestinal Panel 2 (K220062) for Campylobacter |
| Shigella | Qualitative detection equivalent to predicate device (K220062) | Performance equivalent to QIAstat-Dx Gastrointestinal Panel 2 (K220062) for Shigella |
| Shiga-like toxin Escherichia coli (STEC) | Qualitative detection equivalent to predicate device (K220062) | Performance equivalent to QIAstat-Dx Gastrointestinal Panel 2 (K220062) for STEC |
| Salmonella | Qualitative detection equivalent to predicate device (K220062) | Performance equivalent to QIAstat-Dx Gastrointestinal Panel 2 (K220062) for Salmonella |
| Yersinia enterocolitica | Qualitative detection equivalent to predicate device (K220062) | Performance equivalent to QIAstat-Dx Gastrointestinal Panel 2 (K220062) for Yersinia enterocolitica |
2. Sample Size Used for the Test Set and Data Provenance
The provided document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature). It simply states that the performance data for the modified device is "equivalent" to the predicate device. To find this information, one would need to refer to the 510(k) submission K220062 for the predicate device and/or the "QIAGEN QIAstat-Dx GI Panel 2 Mini B Instructions for Use."
3. Number of Experts and Qualifications
The document does not mention the number of experts used to establish ground truth or their qualifications. This information would typically be found in the detailed performance study report of the predicate device, or IFU for the current device. Given that the device detects nucleic acids for specific bacteria, ground truth is likely established through culture or other confirmed molecular methods, rather than expert consensus on images.
4. Adjudication Method
The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set. This type of method is more common in image-based AI studies where human interpretation can vary; for a nucleic acid test, ground truth is typically established by definitive laboratory methods.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study was not applicable or performed for this device as it is a standalone in vitro diagnostic (IVD) device for nucleic acid detection, not an AI-assisted diagnostic tool for human readers. Therefore, there is no effect size related to human reader improvement with AI assistance.
6. Standalone Performance
Yes, a standalone performance assessment was conducted. The QIAstat-Dx GI Panel 2 Mini B is an in vitro diagnostic device, meaning its performance (analytical and clinical performance) is evaluated independently without human interpretation as part of a human-in-the-loop system. The document states its performance is "equivalent" to the predecessor device, which would have undergone rigorous standalone performance testing.
7. Type of Ground Truth Used
While not explicitly stated for this specific submission, for a nucleic acid-based assay like the QIAstat-Dx GI Panel 2 Mini B, the ground truth would typically be established by:
- Culture: Bacterial culture for viability and identification of the target organisms.
- Reference Molecular Methods: e.g., validated PCR assays or sequencing for definitive detection and identification of microbial nucleic acids.
- Composite Reference Method (CRM): A combination of multiple methods, often including culture, microscopy, and/or multiple molecular methods, to establish the presence or absence of the target pathogen.
8. Sample Size for the Training Set
The document does not provide the sample size for the training set. For an IVD like this, "training set" might refer to the data used for initial assay development and optimization rather than a machine learning context. This information would be in the detailed development reports.
9. How the Ground Truth for the Training Set Was Established
The document does not describe how the ground truth for the training set was established. Similar to the test set, it would likely involve established laboratory methods such as culture or reference molecular assays for the target pathogens.
Ask a specific question about this device
(29 days)
Trade/Device Name: BIOFIRE FILMARRAY Gastrointestinal (GI) Panel Mid Regulation Number: 21 CFR 866.3990
FILMARRAY® Gastrointestinal (GI) Panel Mid
Primary Requlation Number for Device Classification: 21 CFR 866.3990
Regulation Number: 21 CFR 866.3990
Classification Name: Gastrointestinal microorganism multiplex nucleic
The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel Mid is an automated qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems. The BIOFIRE FILMARRAY GI Panel Mid is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria, parasites, and viruses are identified using the BIOFIRE FILMARRAY GI Panel Mid:
- · Campylobacter (C. jejuni/C. coli/C. upsaliensis)
- · Clostridioides (Clostridium) difficile (toxin A/B)
- · Salmonella
- · Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae)
- · Yersinia enterocolitica
- · Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
- · Shigella/ Enteroinvasive Escherichia coli (EIEC)
- Cryptosporidium
- · Cyclospora cayetanensis
- · Giardia lamblia (also known as G. intestinalis and G. duodenalis)
- Norovirus GI/GII
The BIOFIRE FILMARRAY GI Panel Mid is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BIOFIRE FILMARRA Y GI Panel Mid. The agent detected may not be the definite cause of the disease.
Concomitant culture is necessary for organism recovery and further typing of bacterial agents.
This device is not intended to monitor or guide treatment for C. difficile infection.
Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Yersinia enterocolitica, were established primarily with retrospective clinical specimens.
Performance characteristics for Vibrio (V. parahaemolyticus, and Vibrio cholerae) was established primarily using contrived clinical specimens.
Negative BIOFIRE FILMARRAY GI Panel Mid results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis. irritable bowel syndrome, or Crohn's disease.
A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection of acute gastroenteritis in the context of outbreaks.
The BIOFIRE® FILMARRAY® Gastrointestinal Panel Mid is designed to simultaneously identify 11 gastrointestinal pathogens from stool specimens collected in Cary Blair transport medium. The BIOFIRE FILMARRAY GI Panel Mid is compatible with BioFire's PCR-based in vitro diagnostic BIOFIRE® FILMARRAY® 2.0 and BIOFIRE® FILMARRAY® TORCH Systems for infectious disease testing. A panel-specific software module (i.e., BIOFIRE FILMARRAY GI Panel Mid pouch module software) is used to perform BIOFIRE FILMARRAY GI Panel Mid testing on these systems. Results from the BIOFIRE FILMARRAY GI Panel Mid test are available within about one hour.
A test is initiated by loading Hydration into one port of the BIOFIRE pouch and a stool sample (in Cary Blair transport medium) mixed with the provided Sample Buffer into the other port of the BIOFIRE FILMARRAY GI Panel Mid pouch and placing it in a BIOFIRE System. The pouch contains all the reagents required for speciment esting and analysis in a freezedried format; the addition of Hydration and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the BIOFIRE Software guides the user though the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.
The BIOFIRE System contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Pettier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.
Nucleic acid extraction occurs within the BIOFIRE pouch using mechanical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the BIOFIRE system performs a nested multiplex PCR that is executed in two stages. During the first stage, the BIOFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green Plus®, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in single plex fashion in each well of the array. At the end of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.
The BIOFIRE Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.
The BIOFIRE FILMARRAY GI Panel Mid is intended for use by trained medical and laboratory professionals in a laboratory setting or under the supervision of a trained laboratory professional.
The provided text is a 510(k) summary for the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel Mid. This document primarily details the analytical and clinical performance of the device to demonstrate its substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and study data based on the provided text:
Device: BIOFIRE FILMARRAY Gastrointestinal (GI) Panel Mid
Indications for Use: Automated qualitative multiplexed nucleic acid-based in vitro diagnostic test for simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Specific Acceptance Criteria and Study Details:
It's important to note that this is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. The "acceptance criteria" presented are implicitly derived from the performance shown to be equivalent to the predicate, rather than explicit pre-defined pass/fail thresholds in a typical AI/ML study. The data provided are from a clinical and analytical evaluation of the parent device (BIOFIRE FILMARRAY GI Panel), with the "Mid" version being identical in hardware and reagents, only differing in software to mask some analytes.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a diagnostic test and not an AI/ML model for image analysis, the acceptance criteria are typically framed in terms of sensitivity (or positive percent agreement - PPA) and specificity (or negative percent agreement - NPA) compared to a reference method.
| BIOFIRE FILMARRAY GI Panel Mid Analyte | Acceptance Criteria (Implied) - High PPA/NPA | Reported Performance (Prospective Clinical Evaluation) |
|---|---|---|
| Bacteria | ||
| Campylobacter (C. jejuni/C. coli/C. upsaliensis) | High Sensitivity, High Specificity | Sensitivity/PPA: 97.1% (34/35) |
| Specificity/NPA: 98.4% (1497/1521) | ||
| Clostridioides (Clostridium) difficile toxin A/B | High PPA, High NPA | PPA: 98.8% (163/165) |
| NPA: 97.1% (1350/1391) | ||
| Salmonella | High Sensitivity, High Specificity | Sensitivity/PPA: 100% (31/31) |
| Specificity/NPA: 99.6% (1519/1525) | ||
| Shiga-like toxin-producing E. coli (STEC) stx1/stx2 | High Sensitivity, High Specificity | Sensitivity/PPA: 100% (33/33) |
| Specificity/NPA: 99.7% (1518/1523) | ||
| Shigella/Enteroinvasive E. coli (EIEC) | High Sensitivity, High Specificity | Sensitivity/PPA: 95.9% (47/49) |
| Specificity/NPA: 99.9% (1505/1507) | ||
| Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae) | High Sensitivity, High Specificity | Sensitivity/PPA: 0/0 (Not applicable due to no positives) |
| Specificity/NPA: 99.9% (1554/1556) | ||
| Yersinia enterocolitica | High Sensitivity, High Specificity | Sensitivity/PPA: 100% (1/1) |
| Specificity/NPA: 100% (1555/1555) | ||
| Parasites | ||
| Cryptosporidium | High PPA, High NPA | PPA: 100% (18/18) |
| NPA: 99.6% (1532/1538) | ||
| Cyclospora cayetanensis | High PPA, High NPA | PPA: 100% (19/19) |
| NPA: 100% (1537/1537) | ||
| Giardia lamblia | High PPA, High NPA | PPA: 100% (20/20) |
| NPA: 99.5% (1529/1536) | ||
| Viruses | ||
| Norovirus GI/GII (2013 data) | High PPA, High NPA | PPA: 94.5% (52/55) |
| NPA: 98.8% (1483/1501) | ||
| Norovirus GI/GII (2023 data) | High PPA, High NPA | PPA: 97.1% (34/35) |
| NPA: 96.5% (808/837) |
Note: The text explicitly states:
"C. difficile performance is reported as positive percent agreement in contrast to the table headings. The performance measures of sensibility and specificity only refer to those analytes for which the [culture method] was used as the reference method; Campylobacter, Salmonella, Vibrio, and Yersinia enterocolitica. Performance measures of positive percent agreement (NPA) refer to all other analytes, for which PCR/sequencing assays were used as comparator methods." This distinction is reflected in the table.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set (Clinical Evaluation):
- The "Prospective Clinical Evaluation" mentioned in Table 2 was conducted from May through September 2013.
- The total number of specimens that contributed to the sensitivity/PPA and specificity/NPA calculations vary by analyte but are in the range of ~1500-1550 specimens per analyte (e.g., 1556 for Vibrio, 1555 for Yersinia, etc.). These numbers represent the denominator for (TP+FN) and (TN+FP) across different analytes.
- A separate, more recent "Prospective Clinical Evaluation" for Norovirus GI/GII was conducted from April through July 2023, involving 872 specimens (35 positives, 837 negatives).
- Retrospective Clinical Specimens: Used for Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) to establish performance characteristics due to small numbers of positive specimens in the prospective study. The specific number of retrospective specimens is not provided for these.
- Contrived Clinical Specimens: Used for Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) to establish performance. The specific number of contrived specimens is not provided.
- Data Provenance: The document does not explicitly state the country of origin for the clinical data. It describes the studies as "prospective clinical study" and "retrospective clinical specimens." This implies they are real-world clinical samples.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- The process for establishing ground truth is described by the reference methods used for each analyte:
- For Campylobacter, Salmonella, Vibrio, and Yersinia enterocolitica: Traditional culture methods were used as the reference.
- For all other analytes (Clostridioides difficile, STEC, Shigella/EIEC, Cryptosporidium, Cyclospora cayetanensis, Giardia lamblia, Norovirus GI/GII): PCR/sequencing assays were used as comparator methods.
- The document does not mention experts being used to establish a subjective "ground truth" (e.g., expert consensus for image review). This is a molecular diagnostic test, where ground truth is typically established by established laboratory reference methods (culture, PCR/sequencing). Therefore, the concept of "number of experts" is not directly applicable in the way it would be for an AI model interpreting medical images.
4. Adjudication Method for the Test Set
- This concept is not relevant for this type of in vitro diagnostic device study. Adjudication (e.g., 2+1, 3+1) is typically performed when subjective interpretations (e.g., radiologist reads) form the ground truth against which an AI model is compared. Here, the ground truth is established by objective laboratory methods (culture, PCR/sequencing). Discrepancies between the device and the reference method might undergo further investigation (e.g., "re-testing," "bi-directional sequence analysis" mentioned for "false positive" or "false negative" specimens), but this is not an "adjudication" in the MRMC sense.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC study was not done. MRMC studies are typically for medical imaging AI where the human reader performance (with and without AI assistance) is evaluated. This device is an automated molecular diagnostic test; it does not involve human "readers" interpreting results in the same way as an imaging AI. Its performance is evaluated against established laboratory reference methods, not human reader improvement.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, the provided performance data represents standalone performance. The BIOFIRE FILMARRAY GI Panel Mid is an automated system that provides qualitative results (Detected/Not Detected). The clinical performance data (sensitivity/PPA, specificity/NPA) directly reflects the device's ability to detect the target analytes directly from the sample without human interpretation or intervention in the diagnostic call itself, beyond operational steps.
7. The Type of Ground Truth Used
- Laboratory Reference Methods:
- Culture: For Campylobacter, Salmonella, Vibrio, and Yersinia enterocolitica.
- PCR/sequencing assays: For Clostridioides difficile, STEC, Shigella/EIEC, Cryptosporidium, Cyclospora cayetanensis, Giardia lamblia, Norovirus GI/GII.
- The text notes that for some discrepancies, "bi-directional sequence analysis" was used to confirm findings for both false positives and false negatives, suggesting a highly definitive molecular method was used for discordant results.
8. The Sample Size for the Training Set
- The document describes the BIOFIRE FILMARRAY GI Panel Mid as "an identical product to the BIOFIRE® FILMARRAY® Gastrointestinal Panel (K242367) except it uses modified labeling and modified software to mask and report only 11 of the 22 targets normally reported."
- "The performance presented here was established during the original clinical and analytical evaluations for the BIOFIRE FILMARRAY GI Panel."
- This implies that there wasn't a separate "training set" for the "Mid" version as per typical AI/ML development. The underlying "algorithm" (the PCR primer sets, probes, and melt curve analysis interpretation) was developed and validated on the original BIOFIRE FILMARRAY GI Panel.
- The document does not provide details on the training set sizes used for the development of the original BIOFIRE FILMARRAY GI Panel. The data presented are from the test/validation phase for the original panel which is now being leveraged for this "special 510k" submission.
9. How the Ground Truth for the Training Set Was Established
- As mentioned above, specific "training set" details for an AI model are not provided because this is a molecular diagnostic hardware/reagent system, not an AI/ML software. The "ground truth" for the development and optimization of the PCR assays (which are the "algorithm" in this context) would have been established through well-characterized analytical samples (known strains, clinical isolates) and potentially early-stage clinical samples validated by comparator methods (culture, sequencing). The document focuses on the validation data (clinical and analytical performance studies) used for regulatory submission.
Ask a specific question about this device
(109 days)
The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative multiplexed in vitro diagnostic test intended for use with the BioCode MDx 3000 Instrument. The BioCode GPP is capable of the simultaneous detection of nucleic acids from multiple bacteria, viruses, and parasites extracted directly from unpreserved in Cary-Blair transport medium obtained from individuals with simptoms of gastrointestinal infection. The following bacteria, parasites, and viruses are identified using the BioCode Gastrointestinal Pathogen Panel:
- . Campylobacter (C. jejuni/C. coli)
- Clostridium difficile (C. difficile) toxin A/B (Fresh samples only)
- l Salmonella spp
- Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae), including specific identification of Vibrio parahaemolyticus .
- . Yersinia enterocolitica
- . Enteroaggregative Escherichia coli (EAEC)
- Enterotoxigenic Escherichia coli (ETEC) lt/st
- E. coli 0157 serogroup
- Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
- Shigella/ Enteroinvasive Escherichia coli (EIEC)
- Cryptosporidium spp (C. parvum/C. hominis)
- Entamoeba histolytica
- Giardia lamblia (also known as G. intestinalis and G. duodenalis)
- . Adenovirus F 40/41
- Norovirus GI/GII ■
- . Rotavirus A
The BioCode GPP is indicated as an aid in the diagnosis of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. For In Vitro Diagnostic Use Only. For Prescription Use Only.
Positive results do not rule out co-infection with organisms not included in the BioCode GPP. The agent detected may not be the definite cause of the disease. Negative results in the setting of clinical illness compatible with gastroenteriis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection.
Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Adenovinus 40/41, Campylobacter, E. coli 0157, Shigella(EIEC, Yersinia enterocolitica, and Giardia lamblia were established additionally with retrospective clinical specimens. Performance characteristica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. cholerae) were established primarily using contrived clinical specimens.
The BioCode® Gastrointestinal Pathogen Panel (GPP) is a multiplexed nucleic acid-based test designed to be used with the BioCode MDx-3000 system. The BioCode MDx-3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for multiple gastrointestinal pathogens from a single stool specimen, either unpreserved or in Cary Blair. Stool specimens are processed, and nucleic acids extracted with easyMAG, MagNA Pure 96, KingFisher Flex and KingFisher Apex Dx. Once the PCR plate is set up and sealed, all other operations are automated on MDx-3000. The BioCode MDx-3000 Gastrointestinal Infection Panel simultaneously tests for 17 pathogens (see table below) from unpreserved stool specimens or stool collected in Cary-Blair transport medium. Results from the BioCode Gastrointestinal Pathogen Panel (GPP) test are available within less than 4 hours.
The provided text describes the BioCode Gastrointestinal Pathogen Panel (GPP), a diagnostic test for gastrointestinal pathogens, and a study to demonstrate its substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and study data:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for each pathogen are implied by the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) compared to the easyMAG extraction method (used with the predicate device) for each extraction system (KingFisher Flex and KingFisher Apex Dx). While explicit pre-defined acceptance criteria values are not stated, the tables present the achieved performance with 95% Confidence Intervals. Generally, a high PPA and NPA (typically >90-95%) with narrow confidence intervals are expected for substantial equivalence for diagnostic tests.
Here's a summary of the reported performance for the BioCode GPP using KingFisher Flex and KingFisher Apex Dx, focusing on the "All Archived" data as it covers a larger sample size for each target. Targets with PPA < 90% are highlighted as indicated in the document; however, the document notes that these are "Positive agreement <90%" for Archived samples only, and specifically points out individual false positives/negatives that were retested and often resolved. Targets for which "N/A" is reported for PPA/NPA indicate no positive/negative archived samples were available for comparison, and thus performance was primarily established with contrived samples (as indicated in the 'Indications for Use' section).
Table of Device Performance (Archived Samples - All Combined)
| Target (KingFisher Flex) | PPA (Agreement with easyMAG) | 95% CI | NPA (Agreement with easyMAG) | 95% CI |
|---|---|---|---|---|
| Campylobacter spp.a | 100% (43/43) | 91.8 - 100 | 99.3% (410/413) | 97.9 - 99.8 |
| Clostridium difficileb | 94.3% (66/70) | 86.2 - 97.8 | 99.5% (386/388) | 98.1 - 99.9 |
| E. coli O157c | 100% (13/13) | 77.2 - 100 | 100% (442/442) | 99.1 - 100 |
| Enteroaggregative E. coli (EAEC) d | 96.2% (50/52) | 87.0 - 98.9 | 100% (404/404) | 99.1 - 100 |
| Enterotoxigenic E. coli (ETEC) e | 100% (25/25) | 86.7 - 100 | 100% (430/430) | 99.1 - 100 |
| Shiga toxin-producing E. coli (STEC) f | 94.1% (32/34) | 80.9 - 98.4 | 99.8% (421/422) | 98.7 - 100 |
| Salmonella spp.g | 100% (41/41) | 91.4 - 100 | 99.5% (412/414) | 98.3 - 99.9 |
| Shigella/ EIEC h | 100% (31/31) | 89.0 - 100 | 99.8% (424/425) | 98.7 - 100 |
| Vibrio parahaemolyticus i | 100% (2/2) | 34.2 - 100 | 100% (453/453) | 99.2 - 100 |
| Vibrio spp. (not parahaemolyticus) j | 50% (1/2)** | 9.5 - 90.5 | 99.8% (452/453) | 98.8 - 100 |
| Yersinia enterocolitica k | 100% (6/6) | 61.0 - 100 | 99.6% (447/449) | 98.4 - 99.9 |
| Cryptosporidium spp. l | 100% (27/27) | 87.5 - 100 | 100% (430/430) | 99.1 - 100 |
| Entamoeba histolytica m | N/A | N/A | 100% (455/455) | 99.2 - 100 |
| Giardia lamblia n | 100% (15/15) | 79.6 - 100 | 100% (440/440) | 99.1 - 100 |
| Adenovirus 40/41 o | 92.3% (12/13) | 66.7 - 98.6 | 99.5% (440/442) | 98.4 - 99.9 |
| Norovirus (GI/GII) p | 95.3% (41/43) | 84.5 - 98.7 | 99.8% (413/414) | 98.6 - 100 |
| Rotavirus A q | 100% (19/19) | 83.2 - 100 | 97.7% (427/437) | 95.8 - 98.8 |
| Combined Targets | 96.1% (223/232) | 92.8 - 97.9 | 99.6% (3941/3956) | 99.4 - 99.8 |
| Target (KingFisher Apex Dx) | PPA (Agreement with easyMAG) | 95% CI | NPA (Agreement with easyMAG) | 95% CI |
|---|---|---|---|---|
| Campylobacter spp.a | 97.7% (42/43) | 87.9 – 99.6 | 99.3% (413/416) | 97.9 – 99.8 |
| Clostridium difficile b | 91.1% (51/56) | 80.7 – 96.1 | 99.5% (402/404) | 98.2 – 99.9 |
| E. coli O157 c | 92.3% (12/13) | 66.7 – 98.6 | 99.8% (445/446) | 98.7 – 100 |
| Enteroaggregative E. coli (EAEC) d | 100% (52/52) | 93.1 – 100 | 99.8% (407/408) | 98.6 – 100 |
| Enterotoxigenic E. coli (ETEC) e | 100% (25/25) | 86.7 – 100 | 99.3% (431/434) | 98.0 – 99.8 |
| Shiga toxin-producing E. coli (STEC) f | 94.1% (32/34) | 80.9 – 98.4 | 99.8% (424/425) | 98.7 – 100 |
| Salmonella spp. g | 100% (41/41) | 91.4 – 100 | 99.3% (415/418) | 97.9 – 99.8 |
| Shigella/ EIEC h | 96.8% (30/31) | 83.8 – 99.4 | 99.1% (425/429) | 97.6 – 99.6 |
| Vibrio parahaemolyticus i | 100% (2/2) | 34.2 – 100 | 99.8% (456/457) | 98.8 – 100 |
| Vibrio spp. (not parahaemolyticus) j | 50% (1/2)** | 9.5 – 90.5 | 99.8% (456/457) | 98.8 – 100 |
| Yersinia enterocolitica k | 100% (6/6) | 61.0 – 100 | 99.6% (451/453) | 98.4 – 99.9 |
| Cryptosporidium spp.l | 96.3% (26/27) | 81.7 – 99.3 | 99.8% (432/433) | 98.7 – 100 |
| Entamoeba histolytica m | N/A | N/A | 100% (459/459) | 99.1 – 100 |
| Giardia lamblia n | 93.3% (14/15) | 70.2 – 98.8 | 100% (444/444) | 99.1 – 100 |
| Adenovirus 40/41 ° | 92.3% (12/13) | 66.7 – 98.6 | 99.6% (444/446) | 98.4 – 99.9 |
| Norovirus (GI/GII) p | 97.7% (42/43) | 87.9 – 99.6 | 100% (417/417) | 99.1 – 100 |
| Rotavirus A 9 | 100% (19/19) | 83.2 – 100 | 100% (441/441) | 99.1 – 100 |
| Combined Targets | 95.4% (208/218) | 91.8 – 97.5 | 99.7% (4006/4017) | 99.5 - 99.8 |
Note:
- Asterisks (**) in the table indicate that the specific performance measure (PPA) was below 90% for archived samples. The document explicitly notes for Vibrio spp. (not parahaemolyticus) that 1 out of 2 positive concordance was observed for both KingFisher Flex and KingFisher Apex Dx, resulting in 50% PPA.
- Retesting of discordant samples often showed resolution in favor of the predicate or new device, but the summary tables reflect the initial comparison. The footnotes (a-q) provide detailed explanations for discordant cases, including instances where samples were invalid or initial false results were later identified as true negatives/positives upon retesting.
2. Sample Size and Data Provenance
- Test Set (Clinical Method Comparison Study):
- Archived Samples: 468 remnant, de-identified samples (254 frozen unpreserved stool and 214 inoculated Cary-Blair stool). These samples were prospectively collected for the clinical study that resulted in the predicate device (K180041 BioCode® GPP FDA clearance). The country of origin is not explicitly stated, but clinical studies for FDA clearance typically involve multi-center studies within the US or compliant international sites. "Mayo-GI" and "TRI" sample names suggest US clinical sites.
- Fresh Samples (for C. difficile testing): 54 freshly collected leftover samples (unpreserved).
- Contrived Samples: 120 samples (15 at 3x LoD and 15 at 6xLoD for each of three targets: Entamoeba histolytica, Yersinia enterocolitica, and Vibrio spp.).
3. Number of Experts to Establish Ground Truth for Test Set and Qualifications
The ground truth for the clinical method comparison study was established by comparing the results of the new device (BioCode GPP with KingFisher Flex and KingFisher Apex Dx extraction) against the results obtained with the easyMAG extraction, which was used with the predicate device (BioCode GPP K190585). The predicate device itself would have undergone its own clinical validation against a pre-established ground truth. Therefore, the "ground truth" for this specific study is the performance of the BioCode GPP with easyMAG extraction.
The document does not explicitly state the number or qualifications of experts used for establishing the initial ground truth of the predicate device, or for adjudicating discordant results in this study. However, the retesting notes for discordant samples mention "consensus results" were not used in the agreement calculation, implying that the retesting was performed to understand the discrepancy rather than to establish a new ground truth.
4. Adjudication Method for the Test Set
The adjudication method is described in the footnotes for Tables 11 and 16. For discordant results between the easyMAG (reference) and the KingFisher Flex/Apex Dx (new device), samples were retested twice with both original extraction methods. The outcome of these retests served to clarify the discrepancy, though the initial raw comparison was used to calculate the agreement, with retest findings noted in the footnotes. For example, a "false negative" for the new device might become a "true negative" upon retesting if the reference system also failed to detect it consistently. However, "consensus results of the discordant samples were not used in the calculation of the agreements" in the main tables but rather for detailed understanding.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This study focuses on a laboratory diagnostic device, not an imaging device requiring human reader interpretation. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
6. Standalone Performance Study
Yes, a standalone performance study was done for the device in the context of the new extraction platforms. The "Nonclinical testing summary" details a Reproducibility Study and a Limit of Detection (LoD) study. These studies evaluate the algorithm's (BioCode GPP assay's) consistency and sensitivity when integrated with the new KingFisher Flex and KingFisher Apex Dx extraction systems, independent of direct human interpretative input beyond standard lab procedures.
- Reproducibility Study: Assessed intra-assay, inter-assay, day-to-day, and instrument-to-instrument (operator-to-operator) reproducibility using quantitative controls diluted in stool.
- Limit of Detection (LoD) Study: Determined the lowest concentration at which ≥95% detection was achieved in 20 replicates for each target, for both unpreserved and Cary-Blair stool, using the new extraction methods.
7. Type of Ground Truth Used
For the clinical method comparison study (test set), the ground truth was the result generated by the predicate device's established extraction method (easyMAG). This is a comparative effectiveness study where the predicate device's performance established the "truth" for evaluating the new device's substantial equivalence.
For the LoD and reproducibility studies, the ground truth was based on quantified stocks of bacteria, viruses, or parasites (except for Norovirus GI/GII which used positive clinical specimens with serial dilutions) introduced into negative stool matrices.
8. Sample Size for the Training Set
The document does not explicitly describe a "training set" for the BioCode GPP itself, as it's a multiplex nucleic acid-based test rather than a machine learning or AI algorithm in the common sense requiring explicit training data. The assay's design (probes, primers, etc.) would have been developed and optimized against a range of target and non-target organisms.
The 468 remnant clinical samples used in the method comparison study were prospectively collected for the predicate device's clinical study (K180041 BioCode® GPP FDA clearance), suggesting they were part of the validation for the predicate, not a training set for the current device's underlying assay design.
9. How the Ground Truth for the Training Set Was Established
As noted above, an explicit "training set" with ground truth in the AI/ML context isn't specified for this type of diagnostic assay. The development of such an assay involves extensive analytical validation to establish specificity, sensitivity, and inclusivity using characterized microbial strains, clinical isolates, and negative controls. The ground truth for these developmental studies would be based on molecular characterization (e.g., sequencing), culture, or other reference methods for the target organisms. The performance of the predicate device (K190585) would have been established through a clinical study where its results were compared against clinical diagnosis, culture, or other validated reference methods.
Ask a specific question about this device
(28 days)
Maryland 20874
Re: K243813
Trade/Device Name: QIAstat-Dx GI Panel 2 Mini B&V Regulation Number: 21 CFR 866.3990
|
| Classification: | 21 CFR 866.3990
|
| Regulation | 21 CFR 866.3990
| 21 CFR 866.3990
The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the OLAstat-Dx Analyzer 1.0 for the simultaneous in vitro qualitative detection of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagence E. col/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V:
- Norovirus
- · Campylobacter
- · Shigella
- · Shiga-like toxin Escherichia coli (STEC)*
- · Salmonella
*Only with Para-Pak C&S, not reported for FecalSwab
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QlAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of gastrontestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Postive results do not rule-out co-infection with organisms not detected by the QlAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx® GI Panel 2 Mini B&V (Cat. no. 691424) assay is a modified device (reduced version) of the QIAstat-Dx Gastrointestinal Panel 2 (Cat. no. 691421). The QIAstat-Dx GI Panel 2 Mini B&V is identical to the OIAstat-Dx Gastrointestinal Panel 2 (K220062) with the exception of their respective labeling and Assay Definition File (ADF) which masks all but five pathogens (targets) from the OIAstat-Dx Gastrointestinal Panel 2. The following virus and bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the OIAstat-Dx GI Panel 2 Mini B&V: Norovirus, Campvlobacter, Shigella, Shiga-like toxin Escherichia coli (STEC) and Salmonella. The QIAstat-Dx GI Panel 2 Mini B&V is part of the OIAstat-Dx system and works with the OIAstat-Dx Analyzer 1.0.
The QIAstat-Dx GI Panel 2 Mini B&V is intended to be used with stool samples in Para-Pak C&S or FecalSwab transport media.
QIAstat-Dx is based on single-test cartridges with pre-packaged reagents including both wet and dry chemistry to handle the sample preparation and detection steps for the presence of a range of selected analytes by PCR technology. After insertion of the sample, the QIAstat-Dx assay cartridge is processed by the QIAstat-Dx Analyzer 1.0.
Once the cartridge has been inserted into the instrument, the test starts automatically and runs for approximately 78 minutes. When the test is finished, the cartridge is removed by the user and discarded. The QIAstat-Dx Analyzer 1.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in gray if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").
All the reagents required for the complete execution of the test are pre-loaded and selfcontained in the QIAstat-Dx GI Panel 2 Mini B&V cartridge. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically-operated microfluidics without any direct contact with the user or the analyzer actuators.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber (TC) to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample Pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K ●
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix ●
- Transfer of defined aliquots of eluate/master mix to different reaction chambers ●
- Performance of multiplex real-time RT-PCR testing within each reaction ● chamber.
The provided text describes a 510(k) premarket notification for a medical device called the QIAstat-Dx GI Panel 2 Mini B&V. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device, the QIAstat-Dx Gastrointestinal Panel 2 (K220062), rather than detailing original acceptance criteria and a comprehensive study designed to prove the device meets those criteria from scratch.
The core of the submission is that the QIAstat-Dx GI Panel 2 Mini B&V is a "reduced version" of the predicate device. It is identical in hardware, reagents, and underlying PCR technology, with the only difference being a modified "Assay Definition File (ADF)" that masks results for all but five specific pathogens. Because of this, the performance data for the new device is considered "equivalent" to the predicate device's data for these five analytes.
Therefore, many of the typical elements of an acceptance criteria study (like an independent test set, MRMC study, or detailed ground truth establishment for a new study) are not explicitly present for the QIAstat-Dx GI Panel 2 Mini B&V in this document, as the submission relies on the existing clearance of the predicate device for its performance claims.
However, based on the provided text, here's an attempt to extract and infer the information requested:
1. A table of acceptance criteria and the reported device performance
The document does not provide a specific table of acceptance criteria for this specific submission. Instead, it states that "The performance data for the QIAstat-Dx GI Panel 2 Mini B&V is equivalent to the QIAstat-Dx Gastrointestinal Panel 2 (K220062) with the exception that it only includes data for the five analytes detected by the QIAstat-Dx GI Panel 2 Mini B&V (Norovirus, Campylobacter, Shigella, Shiga-like toxin E. coli (STEC) and Salmonella)."
It then directs the reader to "Please see the QIAGEN QIAstat-Dx GI Panel 2 Mini B&V Instructions for Use for performance tables." Since these tables are not included in the provided text, we cannot present a direct table of acceptance criteria and reported device performance from this document. The implication is that the predicate device's performance, as accepted during its original 510(k) clearance (K220062), serves as the de facto "acceptance criteria" for these five analytes for the new device by virtue of its identical underlying technology.
2. Sample size used for the test set and the data provenance
The document does not describe a new, independent test set for the QIAstat-Dx GI Panel 2 Mini B&V. Instead, it relies on the data collected for the predicate device (QIAstat-Dx Gastrointestinal Panel 2, K220062). The sample size, country of origin, and whether the data was retrospective or prospective for the predicate device's studies are not detailed in this document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document, as it refers to the predicate device's data rather than a new study with independent expert ground truth establishment for this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This device is a diagnostic nucleic acid test, not an AI-powered image analysis tool or a device that directly assists human readers in interpreting imaging or other complex data. Therefore, an MRMC study and effects on human reader performance are not applicable to this type of device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This refers to the performance of the assay itself. The performance of the QIAstat-Dx GI Panel 2 Mini B&V (which functions like a standalone test after sample input) is considered "equivalent" to the predicate device's performance for the five detected analytes. The document notes that "The QIAstat-Dx Analyzer 1.0 automatically interprets test results and displays a summary on the analyzer display screen." This implies a standalone (algorithm only) performance for result generation, as long as human intervention refers to the interpretation of the raw data by the user, rather than the final qualitative result presented by the device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For nucleic acid amplification tests like this, the ground truth is typically established by well-characterized reference methods, often involving:
- Culture: For bacterial targets, traditional microbiology culture is a common reference standard.
- Reference PCR/Molecular Methods: Highly sensitive and specific laboratory-developed or validated molecular assays.
- Sequencing: For definitive characterization where applicable.
The document does not explicitly state the specific ground truth methods used for the predicate device's studies, but these are the standard approaches for such assays. It does mention that "Concomitant culture is necessary for organism recovery and further typing of bacterial agents" in its indications for use, suggesting culture is an important complementary method in clinical practice.
8. The sample size for the training set
The document does not describe a training set in the context of an AI/machine learning model. This device is a molecular diagnostic assay (PCR-based), not an AI-driven system. Therefore, the concept of a "training set" as it applies to AI models is not relevant here. Development and validation of such assays involve different types of studies (e.g., analytical validation, clinical validation) rather than "training" an algorithm.
9. How the ground truth for the training set was established
As per point 8, the concept of a training set for an AI model is not applicable. The development and validation of PCR assays involve establishing the analytical and clinical performance through rigorous testing against reference methods (as mentioned in point 7) and clinical samples.
Ask a specific question about this device
Page 1 of 5