K Number
K190585

Validate with FDA (Live)

Date Cleared
2019-06-05

(91 days)

Product Code
Regulation Number
866.3990
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the BioCode MDx 3000 Instrument. The BioCode GPP is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites extracted directly from unpreserved stool samples or stool preserved in Cary-Blair transport medium obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria, parasites, and viruses are identified using the BioCode Gastrointestinal Pathogen Panel:

  • · Campylobacter (C. jejuni/C. coli)
  • · Clostridium difficile (C. difficile) toxin A/B (Fresh samples only)
  • · Salmonella spp
  • · Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae), including specific identification of Vibrio parahaemolyticus
  • · Yersinia enterocolitica
  • · Enteroaggregative Escherichia coli (EAEC)
  • · Enterotoxigenic Escherichia coli (ETEC) lt/st
  • · E. coli 0157 serogroup
  • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
  • Shigella/ Enteroinvasive Escherichia coli (EIEC)
  • · Cryptosporidium spp (C. parvum/C. hominis)
  • Entamoeba histolytica
  • · Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Adenovirus F 40/41
  • Norovirus GI/GII
  • Rotavirus A

The BioCode GPP is indicated as an aid in the diagnosis of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data.

Positive results do not rule out co-infection with organisms not included in the BioCode GPP. The agent detected may not be the definite cause of the disease. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Adenovirus 40/41, Campylobacter, E. coli 0157, Shigella/EIEC, Yersinia enterocolitica, and Giardia lamblia were established additionally with retrospective clinical specimens. Performance characteristics for Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V. cholerae) were established primarily using contrived clinical specimens.

Device Description

The BioCode Gastrointestinal Pathogen Panel (GPP) is a multiplexed nucleic acid-based test designed to be used with the BioCode MDx 3000 system. The BioCode MDx 3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for multiple gastrointestinal pathogens from a single stool specimen, either unpreserved or in Cary Blair. Stool specimens are processed and nucleic acids extracted with the easyMAG and MagNa Pure. Once the PCR plate is set up and sealed, all other operations are automated on MDx 3000. The BioCode MDx 3000 Gastrointestinal Infection Panel simultaneously tests for 17 pathogens (see table below) from unpreserved stool specimens or stool collected in Cary-Blair transport medium. Results from the BioCode Gastrointestinal Pathogen Panel (GPP) test are available within less than 4 hours.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the BioCode Gastrointestinal Pathogen Panel (GPP) as described in the provided document:

Acceptance Criteria and Device Performance for BioCode GPP (with MagNA Pure 96 extraction)

The document focuses on demonstrating the substantial equivalence of the BioCode GPP when used with the MagNA Pure 96 extraction system, compared to its previously cleared version with the easyMAG extraction system (K180041, the predicate device). Therefore, the acceptance criteria and performance are largely derived from the comparison to the existing predicate and the analytical studies.

General Acceptance Criteria (Implied from Clinical Performance Tables):
The acceptance criteria are generally implied to be high positive agreement (PPA) and negative agreement (NPA) with reference methods, typically in the range of 80-100% for PPA and 90-100% for NPA, within a 95% Confidence Interval. For analytical studies, reproducibility is expected to be high (>95%), and the limit of detection (LoD) should be comparable between extraction methods.

Table of Acceptance Criteria and Reported Device Performance (Summary):

Target Pathogen/ToxinSpecimen TypeAcceptance Criteria (Implied) - PPA (95% CI)Acceptance Criteria (Implied) - NPA (95% CI)Reported Device Performance (MP96) - PPA (95% CI)Reported Device Performance (MP96) - NPA (95% CI)
Clinical Performance (Archived Specimens - N=464)
Campylobacter spp.Inoculated/Cary-BlairHigh (>90%)High (>90%)95.65% (79.0–99.2)98.20% (94.9–99.4)
Unpreserved (Frozen)High (>90%)High (>90%)100% (87.5-100)98.80% (96.5-99.6)
All ArchivedHigh (>90%)High (>90%)98.0% (89.5-99.6)98.60% (96.9–99.3)
Clostridium difficileInoculated/Cary-BlairHigh (>90%)High (>90%)90.91% (62.3–98.4)99.44% (96.9–99.9)
Unpreserved (Frozen)High (>90%)High (>90%)95.45% (78.2–99.2)99.20% (97.2–99.8)
All ArchivedHigh (>90%)High (>90%)93.94% (80.4-98.3)99.30% (98.0-99.8)
E. coli O157Inoculated/Cary-BlairHigh (100%)High (100%)100% (43.9-100)100% (98.0-100)
Unpreserved (Frozen)High (100%)High (100%)100% (78.5-100)100% (98.5-100)
All ArchivedHigh (100%)High (100%)100% (81.6-100)100% (99.1-100)
Enteroaggregative E. coli (EAEC)Inoculated/Cary-BlairHigh (>85%)High (>95%)88.24% (65.7-96.7)98.8% (95.9-99.7)
Unpreserved (Frozen)High (>95%)High (>95%)100% (88.3-100)99.59% (97.7-99.9)
All ArchivedHigh (>90%)High (>95%)95.65% (85.5-98.8)99.50% (98.3-99.9)
Enterotoxigenic E. coli (ETEC)Inoculated/Cary-BlairModerate (>60%)High (100%)60.00% (23.1-88.2)100% (98.0-100)
Unpreserved (Frozen)High (100%)High (100%)100% (77.2-100)100% (98.5-100)
All ArchivedHigh (>85%)High (100%)88.89% (67.2-96.9)100% (99.1-100)
Shiga toxin-producing E. coli (STEC)Inoculated/Cary-BlairHigh (>90%)High (100%)92.31% (66.7-98.6)100% (97.9-100)
Unpreserved (Frozen)High (>95%)High (>95%)96.67% (83.3-99.4)99.60% (97.7-99.9)
All ArchivedHigh (>90%)High (>95%)95.35% (84.5-98.7)99.80% (98.7-100)
Salmonella spp.Inoculated/Cary-BlairHigh (100%)High (>95%)100% (81.6-100)97.70% (94.2-99.1)
Unpreserved (Fresh)High (>90%)High (>95%)92.59% (76.6-97.9)99.60% (97.7-99.9)
All ArchivedHigh (>90%)High (>95%)95.45% (84.9-98.7)98.80% (97.2-99.5)
Shigella/EIECInoculated/Cary-BlairHigh (100%)High (100%)100% (70.1-100)100% (97.9-100)
Unpreserved (Frozen)High (>90%)High (100%)90.91% (72.2-97.5)100% (98.5-100)
All ArchivedHigh (>90%)High (100%)93.55% (79.3-98.2)100% (99.1-100)
Vibrio parahaemolyticusInoculated/Cary-BlairHigh (100%)High (100%)100% (20.7-100)100% (98.0-100)
Unpreserved (Frozen)High (100%)High (>99%)100% (20.7-100)99.6% (98.0-99.9)
All ArchivedHigh (100%)High (>99%)100% (34.2-100)99.8% (99.8-100)
Vibrio spp. (not parahaemolyticus)Inoculated/Cary-BlairN/AHigh (100%)N/A100% (98.0-100)
Unpreserved (Frozen)Low (0%)High (>99%)0% (N/A)99.6% (N/A)
All ArchivedLow (0%)High (>99%)0% (N/A)99.8% (98.8-100)
Yersinia enterocoliticaInoculated/Cary-BlairHigh (100%)High (100%)100% (43.9-100)100% (98.0-100)
Unpreserved (Frozen)High (100%)High (>99%)100% (43.9-100)99.26% (97.3-99.8)
All ArchivedHigh (100%)High (>99%)100% (61.0-100)99.6% (98.4-99.9)
Cryptosporidium spp.Inoculated/Cary-BlairHigh (>90%)High (100%)91.67% (64.6-98.5)100% (97.9-100)
Unpreserved (Frozen)High (>90%)High (>99%)91.67% (74.2-97.7)99.20% (97.1-99.8)
All ArchivedHigh (>90%)High (>99%)91.67% (78.2-97.1)99.5% (98.3-99.9)
Entamoeba histolyticaInoculated/Cary-BlairN/AHigh (100%)N/A100% (98.0-100)
Unpreserved (Frozen)N/AHigh (100%)N/A100% (98.6-100)
All ArchivedN/AHigh (100%)N/A100% (99.2-100)
Giardia lambliaInoculated/Cary-BlairHigh (100%)High (100%)100% (43.90-100)100% (98.0-100)
Unpreserved (Frozen)High (100%)High (>98%)100% (78.5-100)98.1% (95.6-99.2)
All ArchivedHigh (100%)High (>98%)100% (81.6-100)98.9% (97.4-99.5)
Adenovirus 40/41Inoculated/Cary-BlairModerate (>70%)High (>98%)70.00% (39.7-89.2)98.32% (95.2-99.4)
Unpreserved (Frozen)Moderate (>75%)High (>96%)78.60% (52.4-92.4)96.90% (94.0-98.4)
All ArchivedModerate (>85%)High (>97%)87.50% (69.0-95.7)97.5% (95.6-98.6)
Norovirus (GI/GII)Inoculated/Cary-BlairHigh (100%)High (>98%)100% (83.2-100)98.20% (95.0-99.4)
Unpreserved (Frozen)High (>95%)High (>98%)95.45% (78.2-99.2)98.40% (96.0-99.4)
All ArchivedHigh (>90%)High (>98%)90.24% (77.5-96.1)98.3% (96.6-99.2)
Rotavirus AInoculated/Cary-BlairHigh (>90%)High (>99%)92.31% (66.7-98.6)99.44% (96.9-99.9)
Unpreserved (Frozen)High (100%)High (>98%)100% (79.6-100)98.5% (96.1-99.4)
All ArchivedHigh (>95%)High (>98%)96.43% (82.3-99.4)98.9% (97.3-99.5)
Clinical Performance (Fresh Specimens - N=53)
Clostridium difficileUnpreserved (Fresh)High (100%)High (100%)100% (88.6-100)100% (85.1-100)
Salmonella spp.Unpreserved (Fresh)High (100%)High (100%)100% (20.7-100)100% (92.3-100)
Shigella/EIECUnpreserved (Fresh)High (100%)High (100%)100% (20.7-100)100% (92.3-100)
Norovirus (GI/GII)Unpreserved (Fresh)High (100%)High (100%)100% (20.7-100)100% (92.3-100)
Analytical Performance (Reproducibility)
All targets (various concentrations)In Vitro>95% agreement with expectedN/A>99% (most 100%)N/A

2. Sample Size for the Test Set and Data Provenance:

  • Clinical Test Set:

    • Archived Samples: 466 leftover, de-identified samples (275 frozen unpreserved and 191 inoculated Cary-Blair).
      • Provenance: Prospectively collected for the clinical study that resulted in the K180041 BioCode GPP clearance. This suggests the samples were originally collected for diagnostic purposes in a real-world setting, then archived and de-identified for the current study. The country of origin is not explicitly stated but is implied to be within the scope of previous FDA clearance.
    • Fresh Samples: 53 freshly collected leftover samples (specifically for C. difficile testing initially, but used for other targets as well).
      • Provenance: Freshly collected, though the specific clinical sites or countries are not mentioned.
    • Contrived Samples: 120 samples
      • Provenance: Primarily used to establish performance characteristics for Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V. cholerae) due to small numbers of naturally positive clinical specimens. These are laboratory-prepared samples.
  • Analytical Test Set (Reproducibility Study): Consisted of 7 contrived samples, with combinations of 12 representative targets at 1.5x LoD (Low) and 3x LoD (Medium). Each sample was extracted in triplicate and assayed in singlet. This involved testing across 3 instruments by 3 operators, 2 runs per day for 5 days (total of 30 runs).

  • Analytical Test Set (LoD Study):

    • Initial screening: 4 replicates of each concentration (near LoD) in negative stool and Cary-Blair, extracted on both easyMAG and MagNA Pure 96, and tested in singlet.
    • Confirmation: 20 replicates of each sample type/extraction method, tested in singlet, at or near presumptive LoD.
    • Norovirus GI and GII LoD used positive clinical specimens with serial dilutions.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

The document mentions "Reference methods" which include culture, FDA cleared NAT, PCR/sequencing, and enrichment culture/cleared antigen test. While reference methods are used, the document does not explicitly state the number of experts used to establish the ground truth or their qualifications. It refers to "historical Reference results (K180041)" and "Reference assay". This implies that the ground truth was established by laboratory testing using established diagnostic methods, rather than clinical experts' consensus.

4. Adjudication Method for the Test Set:

The document describes discordant analysis for clinical samples. Specifically, "Sixty-four (64) archived samples with discordant results were retested twice with both easyMag and/or MagNA Pure 96 systems." The retesting appears to be the primary adjudication method. There is no mention of a particular adjudication method like 2+1 or 3+1 involving human experts. The "Comment" section for discordant results often indicates further molecular testing (PCR/Seq) to resolve discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic test that provides qualitative detection of pathogens. Its performance is evaluated based on its agreement (sensitivity and specificity) with established reference methods, not on how it assists human readers in interpreting images or making diagnoses. Therefore, there is no discussion of human reader improvement with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done:

Yes, the studies are standalone performance evaluations of the BioCode GPP diagnostic test. The device itself is an automated system that performs nucleic acid extraction, amplification, detection, and decoding to provide qualitative results. The "Clinical Performance" and "Analytical Performance" sections directly assess the accuracy of the device's output against reference methods. There is no human-in-the-loop component in the direct operation or interpretation of the assay results in the way it would apply to, for example, an AI-powered image analysis tool. The results are interpreted by the MDx 3000 system.

7. The Type of Ground Truth Used:

The ground truth for the clinical studies was established using reference laboratory methods, including:

  • Culture: For Campylobacter (C. jejuni, C. coli), Escherichia coli (E. coli) 0157, Salmonella, Shigella, Vibrio spp., Yersinia enterocolitica.
  • FDA cleared NAT (Nucleic Acid Test): For Clostridium difficile toxin A/B.
  • PCR/sequencing: For Adenovirus 40/41, Cryptosporidium (C. parvum, C. hominis), Entamoeba histolytica, Enteropathogenic E. coli (EPEC), Enterotoxigenic E. coli (ETEC) LT/ST, Enteroaggregative E. coli (EAEC), Giardia lamblia /intestinalis, Norovirus GI/GII, Rotavirus A.
  • Enrichment culture/cleared antigen test: For Shiga-like Toxin producing E. coli (STEC) stx1/stx2.

Some ground truth for less common organisms (Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio) was established with contrived clinical specimens, where the pathogen presence and concentration are precisely known.

8. The Sample Size for the Training Set:

The document does not explicitly mention a "training set" in the context of machine learning for an AI device. This is a molecular diagnostic assay, not an AI/ML algorithm that requires training data in the typical sense. The studies presented are primarily for validation and verification of the device's performance. The "K180041 BioCode GPP clearance" mentioned as the source of archived samples suggests previous validation, but that wouldn't necessarily be considered a "training set" for the current device.

9. How the Ground Truth for the Training Set Was Established:

Since there is no explicit "training set" for an AI/ML algorithm mentioned, this question is not directly applicable. The performance is validated against clinical and analytical studies using reference methods as described in point 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 5, 2019

Applied Biocode. Inc. Robert Tullio Regulatory Consultant 10020 Pioneer Blvd. Suite 102 Santa Fe Springs, California 90067-0

Re: K190585

Trade/Device Name: Biocode Gastrointestinal Pathogen Panel (GPP) Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal microorganism multiplex nucleic acid-based assay Regulatory Class: Class II Product Code: PCH, OOI Dated: March 4, 2019 Received: March 6, 2019

Dear Robert Tullio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190585

Device Name

BioCode Gastrointestinal Pathogen Panel (GPP)

Indications for Use (Describe)

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the BioCode MDx 3000 Instrument. The BioCode GPP is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites extracted directly from unpreserved stool samples or stool preserved in Cary-Blair transport medium obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria, parasites, and viruses are identified using the BioCode Gastrointestinal Pathogen Panel:

  • · Campylobacter (C. jejuni/C. coli)
  • · Clostridium difficile (C. difficile) toxin A/B (Fresh samples only)
  • · Salmonella spp
  • · Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae), including specific identification of Vibrio parahaemolyticus
  • · Yersinia enterocolitica
  • · Enteroaggregative Escherichia coli (EAEC)
  • · Enterotoxigenic Escherichia coli (ETEC) lt/st
  • · E. coli 0157 serogroup
  • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
  • Shigella/ Enteroinvasive Escherichia coli (EIEC)
  • · Cryptosporidium spp (C. parvum/C. hominis)
  • Entamoeba histolytica
  • · Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Adenovirus F 40/41
  • Norovirus GI/GII
  • Rotavirus A

The BioCode GPP is indicated as an aid in the diagnosis of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data.

Positive results do not rule out co-infection with organisms not included in the BioCode GPP. The agent detected may not be the definite cause of the disease. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Adenovirus 40/41, Campylobacter, E. coli 0157, Shigella/EIEC, Yersinia enterocolitica, and Giardia lamblia were established additionally with retrospective clinical specimens. Performance characteristics for Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V. cholerae) were established primarily using contrived clinical specimens.

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

Introduction: According to the requirements of 21 CFR 807.92, the following provides sufficient information to understand the basis for a determination of substantial equivalence.

Submitted by:

Applied BioCode, Inc. 10020 Pioneer Blvd. Suite 102 Santa Fe Springs, CA 90670

Contact:

Robert Di Tullio Regulatory Consultant rditullio@apbiocode.com Telephone: 310 801 1235 Fax: 323 372 3816

Date Submitted:

March 5, 2019

Trade Name:

BioCode Gastrointestinal Pathogen Panel (GPP)

Classification Name and Regulation Number:

Gastrointestinal microorganism multiplex nucleic acid-based assay (21 CFR 866.3990)

Predicate Device:

K180041 – BioCode Gastrointestinal Pathogen Panel (GPP)

Intended Use:

The BioCode Gastrointestinal Pathogen Panel (GPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with the BioCode MDx 3000 Instrument. The BioCode GPP is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites extracted directly from unpreserved stool samples or stool preserved in Cary-Blair transport medium obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria, parasites, and viruses are identified using the BioCode Gastrointestinal Pathogen Panel:

  • " Campylobacter (C. jejuni/C. coli)
  • Clostridium difficile (C. difficile) toxin A/B (Fresh samples only)
  • l Salmonella spp
  • " Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae), including specific identification of Vibrio parahaemolyticus
  • Yersinia enterocolitica
  • " Enteroaggregative Escherichia coli (EAEC)

{5}------------------------------------------------

  • " Enterotoxigenic Escherichia coli (ETEC) It/st
  • י E. coli 0157 serogroup
  • . Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
  • . Shigella/ Enteroinvasive Escherichia coli (EIEC)
  • " Cryptosporidium spp (C. parvum/C. hominis)
  • . Entamoeba histolytica
  • " Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • " Adenovirus F 40/41
  • " Norovirus GI/GII
  • . Rotavirus A

The BioCode GPP is indicated as an aid in the diagnosis of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data.

Positive results do not rule out co-infection with organisms not included in the BioCode GPP. The agent detected may not be the definite cause of the disease. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Adenovirus 40/41, Campylobacter, E. coli 0157, Shigella/EIEC, Yersinia enterocolitica, and Giardia lamblia were established additionally with retrospective clinical specimens. Performance characteristics for Entamoeba histolytica, Giardia lamblia, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V. cholerae) were established primarily using contrived clinical specimens.

Device Description:

The BioCode Gastrointestinal Pathogen Panel (GPP) is a multiplexed nucleic acid-based test designed to be used with the BioCode MDx 3000 system. The BioCode MDx 3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for multiple gastrointestinal pathogens from a single stool specimen, either unpreserved or in Cary Blair. Stool specimens are processed and nucleic acids extracted with the easyMAG and MagNa Pure. Once the PCR plate is set up and sealed, all other operations are automated on MDx 3000. The BioCode MDx 3000 Gastrointestinal Infection Panel simultaneously tests for 17 pathogens (see table below) from unpreserved stool specimens or stool collected in Cary-Blair transport medium. Results from the BioCode Gastrointestinal Pathogen Panel (GPP) test are available within less than 4 hours.

{6}------------------------------------------------

Bacteria, Viruses, Diarrheagenic E. coli/Shigella, and Parasites Detected by the BioCode MDx Gastrointestinal Pathogen Panel (GPP)

BacteriaParasites
■ Campylobacter spp. (C. jejuni, C. coli)■ Cryptosporidium spp.
■ Clostridium difficile toxin A/B (Fresh samples only)■ Entamoeba histolytica
■ Enteroaggregative E. coli (EAEC)■ Giardia lamblia
■ Enterotoxigenic E. coli (ETEC): LT/STViruses
■ Shiga-toxin producing E. coli (STEC): stx1/stx2■ Adenovirus 40/41
■ E.coli O157■ Norovirus GI/GII
■ Shigella spp. /Enteroinvasive E.coli (EIEC)■ Rotavirus A
■ Salmonella spp.RNA Internal Control
■ Vibrio parahaemolyticus
■ Vibrio spp (not parahaemolyticus)
■ Yersinia enterocolitica

Device Comparison:

Comparison of the Applied BioCode GPP with the Predicate Device

CharacteristicProposed DevicePredicate
NameBioCode Gastrointestinal Pathogen Panel (GPP)BioCodeGastrointestinalPathogen Panel(GPP)
CommonNameGastrointestinal Microorganism Multiplex Nucleic acid-basedassayGastrointestinalMicroorganismMultiplex Nucleicacid-based assay
510(k) No.N/AK180041
Regulation21CFR 866.399021CFR 866.3990
Product CodePCH, OOIPCH, OOI
Device ClassIIII
Similarities
Intended UseThe BioCode Gastrointestinal Pathogen Panel (GPP) is aqualitative, multiplexed nucleic acid-based in vitro diagnostic testintended for use with the BioCode MDx 3000 Instrument. TheBioCode GPP is capable of the simultaneous detection andidentification of nucleic acids from multiple bacteria, viruses, andparasites extracted directly from unpreserved stool samples orstool preserved in Cary-Blair transport medium obtained fromindividuals with signs and/or symptoms of gastrointestinalinfection. The following bacteria, parasites, and viruses areidentified using the BioCode Gastrointestinal Pathogen Panel:Campylobacter ( C. jejuni/C. coli ) Clostridium difficile ( C. difficile ) toxin A/B (Fresh samples only) Salmonella spp Vibrio ( V. parahaemolyticus/V. vulnificus/ V. cholerae ),including specific identification of Vibrio parahaemolyticus Yersinia enterocolitica Enteroaggregative Escherichia coli (EAEC) Enterotoxigenic Escherichia coli (ETEC) lt/st E. coli O157 serogroup Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 Shigella/ Enteroinvasive Escherichia coli (EIEC) Cryptosporidium spp ( C. parvum/C. hominis ) Entamoeba histolytica Giardia lamblia (also known as G. intestinalis and G. duodenalis ) Adenovirus F 40/41 Norovirus GI/GII Rotavirus A The BioCode GPP is indicated as an aid in the diagnosis ofspecific agents of gastrointestinal illness and results are meant tobe used in conjunction with other clinical, laboratory, andepidemiological data. For In Vitro Diagnostic Use Only. ForPrescription Use Only.Positive results do not rule out co-infection with organismsnot included in the BioCode GPP. The agent detected may not bethe definite cause of the disease. Negative results in the settingof clinical illness compatible with gastroenteritis may be due toinfection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowelsyndrome, or Crohn's disease. Concomitant culture is necessaryfor organism recovery and further typing of bacterial agents. Thisdevice is not intended to monitor or guide treatment for C. difficile infection.Same

{7}------------------------------------------------

Premarket Notification 510(k)

{8}------------------------------------------------

Premarket Notification 510(k)
-----------------------------------
CharacteristicProposed DevicePredicate
InstrumentNucleic Acid Purification SystemBioCode MDx 3000Same
Sample TypeUnpreserved stool and stool in Cary Blair MediumSame
ControlsExternally Sourced - ZeptometrixSame
MethodologyMultiplex RT-PCR and probe hybridization to biotinylated PCRproduct(s) followed by fluorescence detection and decoding ofbarcoded magnetic beads (BMB) that are coupled to target-specific probesSame
CalibratorsInternal CalibrationSame
Difference
SampleExtractionRoche MagNA Pure 96easyMAG

Summary of Performance Characteristics of the Biocode GPP.

Clinical Performance

A clinical investigational study was performed in which a total of 466 leftover, de-identified samples (275 frozen unpreserved and 191 inoculated Cary-Blair) that were prospectively collected for the clinical study that resulted in the K180041 BioCode GPP clearance were extracted using the MagNA Pure 96 and the easyMAG, and tested on the MDx 3000 system. Fifty-three (53) freshly collected leftover samples were used for the C. difficile testing. In addition, a total of 120 samples were contrived and tested to determine the performance characteristics for Entamoeba histolytica, Yersinia enterocolitica and Vibrio (V. parahaemolyticus, V. vulnificus, and V. cholerae). The results of the clinical investigational study follow.

Archived Samples
Total Specimen Count466
Gender
Male248/466 (53.22%)
Female211/466 (45.28%)
Unknown7/466 (1.50%)
Age Category
≤ 5 yrs85/466 (18.24%)
6-21 yrs91/466 (19.53%)
22-59 yrs194/466 (41.63%)
60+yrs89/466 (19.10%)
Unknown7/466 (1.50%)

Table. Demographic data for archived specimens (frozen unpreserved and inoculated Cary-Blair)

{9}------------------------------------------------

Fresh Samples
Total Specimen Count53
Gender
Male9/53 (16.98%)
Female44/53 (83.02%)
Age Category
≤ 5 yrs1/53 (1.89%)
6-21 yrs5/53 (9.43%)
22-59 yrs23/53 (43.40%)
60+yrs24/53 (45.28%)

Table. Demographic data for fresh specimens (unpreserved)

Clinical sensitivity (positive agreement) was calculated as TP/(TP + FN). TP = true positive by both the EasyMag and the MagNA Pure 96; FN = false negative by the MagNA Pure 96 only. Clinical specificity (negative agreement) was calculated as TN/(TN + FP). TN = true negative or negative by the EasyMag and the MagNA Pure 96; FP = false positive by the MagNA Pure 96 only. The exact binomial two-sided 95% confidence interval was calculated. The results stratified by sample type and storage method are presented in the table below.

Table. Performance compared to historical Reference results. Stratified by sample type and extraction method. N=290 samples. 1 Invalid for EasyMAG (Cary-Blair sample); 3 Invalid for MP96 (1 Cary-Blair, 2 Unpreserved).

EasyMAGMP96
TargetSpecimen TypePositiveAgreementPPA (%)NegativeAgreementNPA (%)PositiveAgreementPPA (%)NegativeAgreementNPA (%)
Campylobacter spp.InoculatedCary-Blair3/3 (100%)135/140 (96%)3/3 (100%)134/140 (96%)
Unpreserved1/1 (100%)139/145 (96%)1/1 (100%)135/143 (94%)
All Archived4/4 (100%)274/285 (96%)4/4 (100%)269/283 (95%)
Clostridium difficileInoculatedCary-Blair9/12 (75%)131/131 (100%)9/12 (75%)131/131 (100%)
Unpreserved12/13(92%)131/133 (98%)12/13 (92%)127/131 (97%)
All Archived21/25(84%)262/264 (99%)21/25 (84%)258/262 (98%)
E. coli 0157InoculatedCary-Blair1/1 (100%)141/142 (99%)0/1 (N/A)141/142 (99%)
Unpreserved0/0 (N/A)144/146 (99%)0/0 (N/A)142/144 (99%)
All Archived1/1 (100%)285/288 (99%)0/1 (N/A)283/286 (99%)
Enteroaggregative E.coli (EAEC)InoculatedCary-Blair14/14(100%)128/129 (99%)13/14 (93%)127/129 (98%)Unpreserved13/14(93%)131/132 (99%)13/14 (93%)130/131 (99%)
Enteroaggregative E.coli (EAEC)InoculatedCary-Blair14/14(100%)128/129 (99%)13/14 (93%)127/129 (98%)
Unpreserved13/14(93%)131/132 (99%)13/14 (93%)130/131 (99%)

{10}------------------------------------------------

EasyMAGMP96
TargetSpecimen TypePositiveAgreementPPA (%)NegativeAgreementNPA (%)PositiveAgreementPPA (%)NegativeAgreementNPA (%)
All Archived27/28(96%)259/261 (99%)26/28 (93%)257/260 (99%)
Enterotoxigenic E. coli(ETEC)InoculatedCary-Blair3/6 (50%)137/137 (100%)3/6 (50%)137/137 (100%)
Unpreserved6/7 (86%)139/139 (100%)6/7 (86%)138/138 (100%)
All Archived9/13 (69%)276/276 (100%)9/13 (69%)275/275 (100%)
Shiga toxin-producingE. coli (STEC)InoculatedCary-Blair1/3 (33%)139/140 (99%)1/3 (33%)139/140 (99%)
Unpreserved0/0 (N/A)144/146 (99%)0/0 (N/A)142/144 (99%)
All Archived1/3 (33%)283/286 (99%)1/3 (33%)281/284 (99%)
Salmonella spp.InoculatedCary-Blair8/10 (80%)129/133 (97%)9/10 (90%)126/133 (95%)
Unpreserved10/10(100%)132/136 (97%)10/10 (100%)131/134 (98%)
All Archived18/20(90%)261/269 (97%)19/20 (95%)257/267 (96%)
Shigella/ EIECInoculatedCary-Blair3/3 (100%)134/140 (96%)3/3 (100%)134/140 (96%)
Unpreserved2/2 (100%)136/144 (94%)2/2 (100%)135/142 (95%)
All Archived5/5 (100%)270/284 (95%)5/5 (100%)269/282 (95%)
VibrioparahaemolyticusInoculatedCary-Blair0/0 (N/A)142/143 (99%)0/0 (N/A)142/143 (99%)
Unpreserved0/0 (N/A)145/146 (99%)0/0 (N/A)143/144 (99%)
All Archived0/0 (N/A)287/289 (99%)0/0 (N/A)285/287 (99%)
Vibrio spp. (notparahaemolyticus)InoculatedCary-Blair0/0 (N/A)143/143 (100%)0/0 (N/A)143/143 (100%)
Unpreserved0/0 (N/A)146/146 (100%)0/0 (N/A)144/144 (100%)
All Archived0/0 (N/A)289/289 (100%)0/0 (N/A)287/287 (100%)
Yersinia enterocoliticaInoculatedCary-Blair0/0 (N/A)141/143 (99%)0/0 (N/A)141/143 (99%)
Unpreserved0/0 (N/A)146/146 (100%)0/0 (N/A)144/144 (100%)
All Archived0/0 (N/A)287/289 (99%)0/0 (N/A)285/287 (99%)
Cryptosporidium spp.InoculatedCary-Blair3/4 (75%)139/139 (100%)2/4 (50%)139/139 (100%)
Unpreserved3/4 (75%)142/142 (100%)3/4 (75%)140/140 (100%)
All Archived6/8 (75%)281/281 (100%)5/8 (63%)279/279 (100%)
Entamoeba histolyticaInoculatedCary-Blair0/0 (N/A)143/143 (100%)0/0 (N/A)143/143 (100%)
TargetSpecimen TypeEasyMAGPositiveAgreementPPA (%)EasyMAGNegativeAgreementNPA (%)MP96PositiveAgreementPPA (%)MP96NegativeAgreementNPA (%)
Giardia lambliaUnpreserved0/0 (N/A)146/146 (100%)0/0 (N/A)144/144 (100%)
All Archived0/0 (N/A)289/289 (100%)0/0 (N/A)287/287 (100%)
InoculatedCary-Blair1/1 (100%)141/142 (99%)1/1 (100%)141/142 (99%)
Giardia lambliaUnpreserved1/1 (100%)142/145 (98%)1/1 (100%)140/143 (98%)
Giardia lambliaAll Archived2/2 (100%)283/287 (99%)2/2 (100%)281/285 (99%)
Adenovirus 40/41InoculatedCary-Blair1/2 (50%)142/142 (100%)1/2 (50%)138/141 (98%)
Unpreserved5/6 (83%)137/140 (98%)5/6 (83%)134/138 (97%)
All Archived6/8 (75%)279/282 (99%)6/8 (75%)272/279 (97%)
Norovirus (GI/GII)InoculatedCary-Blair19/19(100%)124/124 (100%)19/19 (100%)122/124 (98%)
Unpreserved19/19(100%)125/127 (98%)19/19 (100%)124/125 (99%)
All Archived38/38(100%)249/251 (99%)38/38 (100%)246/249 (99%)
Rotavirus AInoculatedCary-Blair11/11(100%)130/132 (98%)11/11 (100%)132/133 (99%)
Unpreserved9/9 (100%)135/137 (99%)9/9 (100%)130/135 (96%)
All Archived20/20(100%)264/269 (99%)20/20 (100%)262/268 (98%)

{11}------------------------------------------------

Footnotes (D = Detected, N = Not Detected)

HistoricalResults(K180041)MP96Extraction(K190585)
TargetSpecimenIDReferenceBioCodeGPPeasyMAGMP96Comment
01-0027NDDDReference assay (culture) not as sensitive, PCR/seq C. jejuni, but contig too short;concordant with historical GPP result
Campylobacterspp01-0048NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
01-0137NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
01-0237NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
01-0352NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
02-0159NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
02-0183NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
02-0367NDDDReference assay (culture) not as sensitive, PCR/Seq not detected, concordant withhistorical GPP result
03-0153NDDDReference assay (culture) not as sensitive, PCR/Seq not detected, concordant withhistorical GPP result
HistoricalResults(K180041)MP96Extraction(K190585)
TargetSpecimenIDReferenceBioCodeGPPeasyMAGMP96Comment
Clostridiumdifficile03-0308NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
03-0334NDDDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; concordant withhistorical GPP result
02-0069NDNDReference assay (culture) not as sensitive, PCR/Seq C. jejuni; MP96 concordantwith historical GPP result
02-0193NNNDPossible Contamination; see repeat results for additional information
03-0364NNNDPossible Contamination; see repeat results for additional information
01-0031NDDDReference Ct beyond cut-off (38.7); concordant with historical GPP Result
01-0063NDDDReference Ct beyond cut-off with ref (38.3); concordant with historical GPP Result
02-0201DNDNLate Ct with ref (36.6) tested fresh, low positive; see repeat results for additionalinformations
02-0350NNNDPossible Contamination; see repeat results for additional information
03-0053DDNDLate Ct with ref (36.2) tested fresh; low positive; see repeat results for additionalinformations
03-0351NNNDPossible Contamination; see repeat results for additional information
01-0002DNNNLate Ct with ref (36.6) tested fresh; concordant with historical GPP Result (frozen)
02-0378DNNNLate Ct with ref (36.6) tested fresh; concordant with historical GPP Result (fresh)
03-0353DDNNLate Ct with ref (35.5) tested frozen; possible sample degradation
E. coli 015702-0080NDDDReference assay (culture) not as sensitive, PCR/seq O157; concordant withhistorical GPP result
02-0082NDDDReference assay (culture) not as sensitive, PCR/seq O157; concordant withhistorical GPP result
02-0206NDDDReference assay (culture) not as sensitive, PCR/seq 0157; concordant withhistorical GPP result
02-0279DDNNPossible sample degradation; current testing concordant
03-0140NDDDLate Ct with ref (37.1) on repeat testing; concordant with historical GPP Result
02-0183NNDNPossible low positive; see repeat results for additional information
Enteroaggregative E. coli (EAEC)03-0004DDDNPossible low positive; see repeat results for additional information
02-0367NNNDPossible Contamination; see repeat results for additional information
03-0046NNNDPossible Contamination; see repeat results for additional information
EnterotoxigenicE. coli (ETEC)02-0203DNNNReference result could not be confirmed by 2 additional rounds of PCR/Seq;concordant with historical GPP Result
02-0203DNNNReference result could not be confirmed by 2 additional rounds of PCR/Seq;concordant with historical GPP Result
02-0207DNNNLate Ct (35.2) Reference result could not be confirmed by 2 additional rounds ofPCR/Seq; concordant with historical GPP Result
03-0004DNNNReference result could not be confirmed by 2 additional rounds of PCR/Seq;concordant with historical GPP Result
02-0391DDNNLate Ct (37.8) with ref tested fresh; possible sample degradation
Salmonella spp.01-0091NDDDReference assay (culture) not as sensitive, PCR/seq S.enterica; concordant withhistorical GPP result
HistoricalResults(K180041)MP96Extraction(K190585)
TargetSpecimenIDReferenceBioCodeGPPeasyMAGMP96Comment
01-0137NNDNPossible Contamination; see repeat results
01-0278DDNDSee repeat results for additional information
01-0346DDNNPossible sample degradation; current testing concordant
02-0195NNNDPossible Contamination; see repeat results for additional information
02-0213NDDDReference assay (culture) not as sensitive, PCR/seq S.enterica; concordant withhistorical GPP result
02-0260NNNDPossible Contamination; see repeat results for additional information
02-0378NNNDPossible Contamination; see repeat results for additional information
03-0116NDDDReference assay (culture) not as sensitive, PCR/seq S.enterica; concordant withhistorical GPP result
03-0147NDDDReference assay (culture) not as sensitive, PCR/seq S.enterica; concordant withhistorical GPP result
03-0173NDDDReference assay (culture) not as sensitive, PCR/seq S.typhi; concordant withhistorical GPP result
03-0189NDDDReference assay (culture) not as sensitive, PCR/seq S.enterica; concordant withhistorical GPP result
03-0206NDDDReference assay (culture) not as sensitive, PCR/seq S.typhi; concordant withhistorical GPP result
Shiga toxin-producing E. coli(STEC)01-0098NDDDReference assay (culture) not as sensitive, PCR/seq STEC; concordant with historicalGPP result
02-0367NDDDReference assay (culture) not as sensitive, PCR/seq STEC; concordant with historicalGPP result
02-0393NDDDReference assay (culture) not as sensitive, PCR/seq STEC; concordant with historicalGPP result
02-0279DDDNSee repeat results for additional information
02-0081DDNNPossible sample degradation; current testing concordant
01-0019NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0020NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0027NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0099NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0199NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
Shigella/EIEC01-0201NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0202NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0259NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
02-0176NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
03-0056NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
03-0090NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
03-0238NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
HistoricalResults(K180041)MP96Extraction(K190585)
TargetSpecimenIDReferenceBiocodeGPPeasyMAGMP96Comment
03-0380NDDDReference assay (culture) not as sensitive, PCR/seq Shigella spp; concordant withhistorical GPP result
01-0319NNDNPossible low positive; see repeat results for additional information
Vibrioparahaemolyticus03-0194NDDDReference assay (culture) not as sensitive, PCR/seq V.parahaemolyticus; concordantwith historical GPP result
03-0351NDDDReference assay (culture) not as sensitive, PCR/seq V.parahaemolyticus; concordantwith historical GPP result
Yersiniaenterocolitica01-0024NDDDReference assay (culture) not as sensitive, PCR/seq V.parahaemolyticus; concordantwith historical GPP result
03-0307NDDDReference assay (culture) not as sensitive, PCR/seq not detected; concordant withhistorical GPP result
01-0212DDDNSee repeat results for additional information
Cryptosporidiumparvum01-0265DDNNPossible sample degradation; current testing concordant
01-0285DDNNPossible sample degradation; current testing concordant
01-0002NDDDReference assay (PCR/Seq) not as sensitive; concordant with historical GPP result
Giardiaintestinalis01-0300NDDDReference assay (PCR/Seq) not as sensitive; concordant with historical GPP result
02-0258NDDDReference assay (PCR/Seq) not as sensitive; concordant with historical GPP result
03-0140NDDDReference assay (PCR/Seq) not as sensitive; concordant with historical GPP result
Humanadenovirus40/4101-0128NNDDconcordant with historical GPP result
01-0021NNDNPossible Contamination; see repeat results for additional information
02-0288NNDNPossible Contamination; see repeat results for additional information
01-0137NNNDPossible Contamination; see repeat results for additional information
02-0161NNNDPossible Contamination; see repeat results for additional information
02-0183NNNDPossible Contamination; see repeat results for additional information
02-0379NNNDPossible Contamination; see repeat results for additional information
02-0391NNNDPossible Contamination; see repeat results for additional information
03-0400NNNDPossible Contamination; see repeat results for additional information
02-0137DNNNReference result could not be confirmed by 2 additional rounds of PCR/Seq;concordant with historical GPP Result
02-0147DDNNPossible sample degradation; current testing concordant
Norovirus GI/GII01-0004NDDDconcordant with historical GPP result
01-0036NNDNLow positive based on MFI (3150)
02-0376NNNDLow positive based on MFI (1265)
02-0394NNNDLow positive based on MFI (2249)
Human rotavirusA01-0128NNDDCurrent testing concordant
TargetSpecimen IDReferenceHistorical Results(K180041)BioCode GPPMP96 Extraction(K190585)easy MAGMP96MP 96Comment
01-0337NDDDconcordant with historical GPP result
02-0175NDDDconcordant with historical GPP result
02-0246NDDNPossible low positive; see repeat results for additional information
01-0098NDNDPossible low positive; see repeat results for additional information
01-0202NNNDPossible Contamination/low positive; see repeat results for additional information
03-0400NNNDPossible Contamination; see repeat results for additional information

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Comparator (Reference) methods from BioCode GPP original submission

Target Pathogen/Toxin

Target Pathogen/ToxinReference Method
Adenovirus 40/41Composite result of PCR/sequencing
Campylobacter (C. jejuni, C. coli)Culture
Clostridium difficile (C. difficile) toxin A/BFDA cleared NAT
Cryptosporidium (C. parvum, C. hominis)PCR/sequencing
Entamoeba histolyticaPCR/sequencing
Escherichia coli (E. coli) 0157Enrichment culture
Enteropathogenic E. coli (EPEC)Composite result of PCR/sequencing
Enterotoxigenic E. coli (ETEC) LT/STComposite result of PCR/sequencing
Enteroaggregative E. coli (EAEC)Composite result of PCR/sequencing
Giardia lamblia /intestinalisComposite result of PCR/sequencing
Norovirus GI/GIIComposite result of PCR/sequencing
Rotavirus AComposite result of PCR/sequencing
SalmonellaEnrichment culture
Shiga-like Toxin producing E. coli (STEC) stx1/stx2Enrichment culture/cleared antigen test
Shigella (S. boydii, S. sonnei, S. flexneri, S. dysenteriae)/EIECEnrichment culture
Vibrio spp. (V. cholerae, V. parahaemolyticus, V. vulnificus)Culture
Yersinia enterocoliticaCulture
TargetSpecimenType(n)Positive AgreementNegative Agreement
PPA (%)95% CINPA (%)95% CI
Campylobacterspp.aInoculatedCary-Blair19022/23 (95.65)79.0 – 99.2164/167 (98.20)94.9 – 99.4
Unpreserved(Frozen)27427/27 (100)87.5 - 100244/247 (98.80)96.5 - 99.6
All Archived46449/50 (98.0)89.5 - 99.6408/414 (98.60)96.9 – 99.3
Clostridiumdifficile bInoculatedCary-Blair19010/11 (90.91)62.3 – 98.4178/179 (99.44)96.9 – 99.9
Unpreserved(Frozen)27421/22 (95.45)78.2 – 99.2250/252 (99.20)97.2 – 99.8
TargetSpecimen Type(n)Positive AgreementNegative Agreement
All Archived46431/33 (93.94)80.4 - 98.3428/431 (99.30)98.0 - 99.8
InoculatedCary-Blair1903/3 (100)43.9 - 100187/187 (100)98.0 - 100
E. coli O157Unpreserved(Frozen)27414/14 (100)78.5 - 100260/260 (100)98.5 - 100
All Archived46417/17 (100)81.6 - 100447/447 (100)99.1 - 100
InoculatedCary-Blair19015/17 (88.24)65.7 - 96.7171/173 (98.8)95.9 - 99.7
EnteroaggregativeE. coli (EAEC) cUnpreserved(Frozen)27429/29 (100)88.3 - 100244/245 (99.59)97.7 - 99.9
All Archived46444/46 (95.65)85.5 - 98.8416/418 (99.50)98.3 - 99.9
InoculatedCary-Blair1903/5 (60.00)23.1 - 88.2185/185 (100)98.0 - 100
EnterotoxigenicE. coli (ETEC) dUnpreserved(Frozen)27413/13 (100)77.2 - 100261/261 (100)98.5 - 100
All Archived46416/18 (88.89)67.2 - 96.9446/446 (100)99.1 - 100
Shiga toxin-producing E. coli(STEC) eInoculatedCary-Blair19012/13 (92.31)66.7 - 98.6177/177 (100)97.9 - 100
Unpreserved(Frozen)27429/30 (96.67)83.3 - 99.4243/244 (99.60)97.7 - 99.9
All Archived46441/43 (95.35)84.5 - 98.7420/421 (99.80)98.7 - 100
InoculatedCary-Blair19017/17 (100)81.6 - 100169/173 (97.70)94.2 - 99.1
Salmonella spp.fUnpreserved(Fresh)27425/27 (92.59)76.6 - 97.9246/247 (99.60)97.7 - 99.9
All Archived46442/44 (95.45)84.9 - 98.7415/420 (98.80)97.2 - 99.5
Shigella/ EIEC gInoculatedCary-Blair1909/9 (100)70.1 - 100181/181 (100)97.9 - 100
Unpreserved(Frozen)27420/22 (90.91)72.2 - 97.5252/252 (100)98.5 - 100
All Archived46429/31 (93.55)79.3 - 98.2433/433 (100)99.1 - 100
Vibrioparahaemolyticus hInoculatedCary-Blair1901/1 (100)20.7 - 100189/189 (100)98.0 - 100
Unpreserved(Frozen)2741/1 (100)20.7 - 100272/273 (99.6)98.0 - 99.9
All Archived4642/2 (100)34.2 - 100461/462 (99.8)99.8 - 100
Vibrio spp. (notparahaemolyticus)iInoculatedCary-Blair190N/AN/A190/190 (100)98.0 - 100
Unpreserved(Frozen)2740/1 (0%)N/AN/A273/274 (99.6)
All Archived4640/1 (0%)N/A463/464 (99.8)98.8 - 100
Yersiniaenterocolitica jInoculatedCary-Blair1903/3 (100)43.9 - 100187/187 (100)98.0 - 100
Unpreserved(Frozen)2743/3 (100)43.9 - 100269/271 (99.26)97.3 - 99.8
All Archived4646/6 (100)61.0 - 100456/458 (99.6)98.4 - 99.9
SpecimenType(n)Positive AgreementNegative Agreement
TargetPPA (%)95% CINPA (%)95% CI
Cryptosporidiumspp.kInoculatedCary-Blair19011/12 (91.67)64.6 – 98.5178/178 (100)97.9 - 100
Unpreserved(Frozen)27422/24 (91.67)74.2 - 97.7248/250 (99.20)97.1 - 99.8
All Archived46433/36 (91.67)78.2 – 97.1426/428 (99.5)98.3 - 99.9
EntamoebahistolyticaInoculatedCary-Blair190N/AN/A190/190 (100)98.0 - 100
Unpreserved(Frozen)274N/AN/A274/274 (100)98.6 - 100
All Archived464N/AN/A464/464 (100)99.2 - 100
Giardia lamblialInoculatedCary-Blair1903/3 (100)43.90 - 100187/187 (100)98.0 - 100
Unpreserved(Frozen)27414/14 (100)78.5 - 100255/260 (98.1)95.6 - 99.2
All Archived46417/17 (100)81.6 - 100442/447 (98.9)97.4 - 99.5
Adenovirus 40/41mInoculatedCary-Blair1907/10 (70.00)39.7 – 89.2177/180 (98.32)95.2 - 99.4
Unpreserved(Frozen)27411/14 (78.60)52.4 - 92.4252/260 (96.90)94.0 - 98.4
All Archived46421/24 (87.50)69.0 - 95.7429/440 (97.5)95.6 - 98.6
Norovirus(GI/GII)nInoculatedCary-Blair19019/19 (100)83.2 - 100168/171 (98.20)95.0 - 99.4
Unpreserved(Frozen)27421/22 (95.45)78.2 – 99.2248/252 (98.40)96.0 - 99.4
All Archived46437/41 (90.24)77.5 - 96.1416/423 (98.3)96.6 - 99.2
Rotavirus AoInoculatedCary-Blair19012/13 (92.31)66.7 – 98.6176/177 (99.44)96.9 - 99.9
Unpreserved(Frozen)27415/15 (100)79.6 - 100255/259 (98.5)96.1 - 99.4
All Archived46427/28 (96.43)82.3 - 99.4431/436 (98.9)97.3 - 99.5

Table. Summary of Clinical Investigational Study Results (Archived Specimens) stratified by sample type and storage

{16}------------------------------------------------

{17}------------------------------------------------

Sixty-four (64) archived samples with discordant results were retested twice with both easyMag and/or MagNA Pure 96 systems.

a - Campylobacter spp. One (1) false negative retested became true positives retested, 2 were true negative, one became true positive, two remained false positive not retested due to insufficient volume to retest.

b - Clostridium difficile: Of the 2 false negatives retested, 1 became true positive, the other one was true negative. Of the 3 false positives retested, 2 were true negative and 1 was true positive.

c - EAEC: Of the 2 false negatives retested, 1 was false positive. Of the 3 false positive. Of the 3 false positives retested, 2 were true negative and 1 was true positive.

d - ETEC: Two (2) false negatives tested were true negative.

e - STEC: Of 3 false positives retested, 2 were true positive and 1 was true negative.

f - Salmonella spp. Of the 2 false negatives retested, one was false positive and the other one was true negative. Of the 5 false positives retested, 4 were true negative and 1 was true positive.

g - Shigella/EIEC: Of the 2 false negatives retested, 1 was true positive and the other remained false negative.

h - Vibrio parahaemolyticus: The 1 false positive retested was true negative.

i - Vibrio spp. The 1 false negative retested remained false negative. Detected as Vibrio parahaemolyticus by MP96.

j - Yersinia enterocolitica: Two (2) false positives retested were true negative.

k - Cryptosporidium spp: Of the 3 false negative and 2 were false negative and 2 were false negative. The 2 false positives retested remained false positive.

{18}------------------------------------------------

l - Giardia lamblia: Of the 5 false positives retested, 1 became true negative. The remaining 2 were not retested due insufficient volume to retest.

m- Adenovirus 40/41: All 6 false negatives. Of the 11 false positives, 9 were retested and became true negative, and the remaining 2 were not retested due insufficient volume to retest.

n - Norovirus G1/G2: The 1 false negative retested was true negatives retested, 6 were true negative and one became true positive.

o - Rotavirus: The 1 false negative retested was true negatives retested, 1 was true positive, 3 were true negative, and 1 remained false positive.

TargetSample NameSampleTypeExtractionOriginal RunReflex TestingResultFinal Result
01-0079InoculatedCary-BlairEasyMagIC InvalidIC ValidValid
Internal Control(MS2)01-0073InoculatedCary-BlairMagNAPureIC InvalidIC InvalidInvalid
02-0211UnpreservedMagNaPureIC InvalidIC InvalidInvalid
02-0228UnpreservedEasyMagIC InvalidIC ValidValid
Mayo-G1_124UnpreservedMagNAPureIC InvalidIC ValidValid

Table. Results of Reflex Testing of Invalid Samples (Archived Samples)

Note. Five (5) archived samples were invalid. After relflex testing, two samples (1 inoculated Cary-Blair and 1 unpreserved) were still invalid.

TargetSpecimen(n)Positive AgreementNegative Agreement
TypePPA (%)95% CINPA (%)95% CI
Campylobacter spp.Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Clostridium difficileUnpreserved(Fresh)5230/30 (100)88.6-10022/22 (100)85.1-100
E. coli 0157Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Enteroaggregative E.coli (EAEC)Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
EnterotoxigenicE. coli (ETEC)Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Shiga toxin-producingE. coli (STEC)Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Salmonella spp.Unpreserved(Fresh)471/1 (100)20.7-10046/46 (100)92.3-100
Shigella/ EIECUnpreserved(Fresh)471/1 (100)20.7-10046/46 (100)92.3-100
VibrioparahaemolyticusUnpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Vibrio spp. (notparahaemolyticus)Unpreserved(Fresh)47N/AN/A47/47(100)92.4-100
Yersinia enterocoliticaUnpreserved(Fresh)47N/AN/A47/47(100)92.4-100

Table. Summary of Clinical Investigational Study Results of Fresh Specimens

{19}------------------------------------------------

TargetSpecimen(n)Positive AgreementNegative Agreement
TypePPA (%)95% CINPA (%)95% CI
Cryptosporidium spp.Unpreserved(Fresh)47N/AN/A47/47(100)92.4- 100
Entamoeba histolyticaUnpreserved(Fresh)47N/AN/A47/47(100)92.4- 100
Giardia lamblia¹Unpreserved(Fresh)47N/AN/A47/47(100)92.4- 100
Adenovirus 40/41Unpreserved(Fresh)47N/AN/A47/47(100)92.4- 100
Norovirus(GI/GII)Unpreserved(Fresh)471/1 (100)20.7 -10046/46 (100)92.3- 100
Rotavirus AUnpreserved(Fresh)47N/AN/A47/47(100)92.4- 100

Note. Fifty-three (53) fresh unpreserved samples were tested. The Internal Control (MS2) result of the 5 C. difficile positives and 1 C. difficile negative were invalid. Therefore, one sample for C. difficile target was invalid. For the remaining targets, 6 samples were invalid.

Table. Summary of Contrived Specimen Results

TargetSpecimen Type(n)Positive AgreementNegative Agreement
PPA (%)95% CINPA (%)95% CI
Vibrio parahaemolyticus aUnpreserved (Frozen)11928/30 (93.3)78.7 – 98.289/89 (100)95.9 – 100
Vibrio spp. (not parahaemolyticus) bUnpreserved (Frozen)11829/29 (100)88.3 - 10089/89 (100)95.9 - 100
Yersinia enterocoliticaUnpreserved (Frozen)11830/30 (100)88.6 – 10088/88 (100)95.8 – 100
Entamoeba histolyticaUnpreserved (Frozen)11830/30 (100)88.6 – 10088/88 (100)96.8 – 100
All other targetsUnpreserved (Frozen)118N/AN/A118/118 (100)96.8 – 100

Note. One hundred twenty (120) contrived samples were tested. ª The Internal control (MS2) result of one of the Vibrio parahaemolyticus positives was invalid. * One (1) of the Vibrio spp. positives was negative and invalid (IC negative) with both the easyMag and the MagNA Pure 96.

Table. Clinical Co-infection Summary

Co-infection summary
+1 Target246/466 (52.8%)
+2 Targets78/466 (16.7%)
+3 Targets13/466 (2.8%)

Analytical Performance

{20}------------------------------------------------

The results of the analytical studies summarized in the following paragraphs met the acceptance criteria and successfully demonstrated the analytical performance characteristics of the proposed BioCode GPP using the MagNA Pure 96 extraction system.

REPRODUCIBILITY STUDY

A study was performed to assess the Reproducibility of the BioCode GPP using samples extracted with the MagNA Pure 96. This study was designed to assess intra-assay (within run), Inter-assay (run-to-run), day-to-day and instrument-to-instrument (operator) reproducibility. One lot of reagents was assayed at Applied BioCode on 3 instruments by 3 operators, 2 runs per day per operator for 5 days (total of 30 runs). The reproducibility panel consisted of 7 contrived samples (sample 7 a negative control) extracted in triplicate and each assayed in singlet. The samples consisted of combinations of 12 representative targets at 1.5x LoD (Low) and 3x LoD (Medium). Reproducibility was > 99%.

All results are as expected with the exception of two false negative results for Cryptosporidium parvum (one low positive and one medium positive).

% Agreement with Expected Result
Organism TestedTargetProbeConcentration TestedExpectedResultsInstrument1- Operator 1Instrument2- Operator 2Instrument3- Operator 3AllInstruments/Operators
Campylobacterjejuni spp. jejuniATCC 33292CampyMedium Positive 3xLoD1.05 x 103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD5.25 x 102 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Clostridiumdifficile(toxinotype III;Nap1)Zeptometrix0801619cftcdBMedium Positive 3xLoD1.25 x 103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD6.23 x 102 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
EnteroaggregativeE. coli 092:H33(EAEC)STEC TW04440EAECMedium Positive 3xLoD2.10 x 103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD1.05 x 103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Organism TestedTargetProbeConcentration TestedExpectedResultsInstrument1- Operator 1Instrument2- Operator 2Instrument3- Operator 3AllInstruments/Operators
Enterotoxigenic E.coli 078:H11H10407 (ETEC)ATCC 35401ST-1aNoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Medium Positive 3xLoD8.40 x 10² CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD4.20 x 10² CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Salmonellaenterica ssp.entericaATCC 14028SalmNoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Medium Positive 3xLoD3.30 x 10³ CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD1.65 x 10³ CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Shiga-like toxinproducing E. coli(STEC)ATCC BAA-2217stx2NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Medium Positive 3xLoD3.75 x 10³ CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD1.88 x10³ CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Shigella sonneiATCC 29930ShigNoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Medium Positive 3xLoD6.60 x 10² CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Low Positive 1.5xLoD3.30 x 10² CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
VibrioparahaemolyticusATCC 17802V.paraNoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Medium Positive 3xLoD1.95 x10¹ CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
% Agreement with Expected Result
Organism TestedTargetProbeConcentration TestedExpectedResultsInstrument1- Operator 1Instrument2- Operator 2Instrument3- Operator 3AllInstruments/Operators
YersiniaenterocoliticaATCC 23715Low Positive 1.5xLoD9.75x10° CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
YersiniaenterocoliticaATCC 23715NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
YersiniaenterocoliticaATCC 23715Medium Positive 3xLoD2.25 x103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
YersiniaenterocoliticaATCC 23715Y.entLow Positive 1.5xLoD1.13 x103 CFU/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
YersiniaenterocoliticaATCC 23715NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Cryptosporidiumparvumwaterborne P102Medium Positive 3xLoD9.30 x 103 oocysts/mLDetected30/30(100%)30/30(100%)29/30(97%)89/90 (99%)
CryptoLow Positive 1.5xLoD4.65 x 103 oocysts/mLDetected30/30(100%)30/30(100%)29/30(97%)89/90 (99%)
NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
Giardia intestinalis(aka G. lamblia)waterborne P101Medium Positive 3xLoD2.70 x 103 cysts/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
G.lamLow Positive 1.5xLoD1.35 x 103 cysts/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Giardia intestinalis(aka G. lamblia)waterborne P101NoneNotDetected150/150(100%)150/150(100%)150/150(100%)450/450(100%)
RotavirusATCC VR-2018Medium Positive 3xLoD3.75 x 103 TCID50/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
RotaLow Positive 1.5xLoD1.88 x 102 TCID50/mLDetected30/30(100%)30/30(100%)30/30(100%)90/90 (100%)
Organism TestedTarget ProbeConcentration TestedExpected Results% Agreement with Expected Result
Instrument1 - Operator 1Instrument2 - Operator 2Instrument3 - Operator 3All Instruments/ Operators
NoneNot Detected150/150(100%)150/150(100%)150/150(100%)450/450(100%)

Table. Reproducibility of BioCode GPP with MP96 extractions- Qualitative results.

{21}------------------------------------------------

{22}------------------------------------------------

{23}------------------------------------------------

Table. Reproducibility of BioCode GPP with MP96 extractions- Quantitative results

OrganismTargetProbeTest LevelInstrument/OperatorMFI Reproducibility
MeanStDev%CVMinMax
Campylobacter jejuni spp. jejuniATCC 33292CampyModeratePositive3xLoD$1.05 x 10^3$CFU/mLC2- Operator 126128401315826529801
C3- Operator 228562265292218733021
C4- Operator 3241542714111889028822
All Instruments/Operators26281363714826533021
Low Positive1.5xLoD$5.25 x 10^2$CFU/mLC2- Operator 1248313563141432630367
C3- Operator 2258314285171440631649
C4- Operator 3216554326201056630156
All Instruments/Operators241064407141056631649
Clostridium difficile(toxinotype III;Nap1)Zeptometrix0801619cftcdBModeratePositive3xLoD$1.25 x 10^3$CFU/mLC2- Operator 127589227682248731505
C3- Operator 2226723726161711232047
C4- Operator 3227802838111692627598
All Instruments/Operators243473762151692632047
Low Positive1.5xLoD$6.23 x 10^2$CFU/mLC2- Operator 125007176172043627552
C3- Operator 2201163513171375227461
C4- Operator 3209022561111560827662
All Instruments/Operators220083438161375227662
Enteroaggregati ve E. coli092:H33 (EAEC)STEC TW04440EAECModeratePositive3xLoD$2.10 x 10^3$CFU/mLC2- Operator 130474167252561633694
C3- Operator 2284243220142010834967
C4- Operator 325129219891878728489
All Instruments/Operators280093281121878734967
Low Positive1.5xLoD$1.05 x 10^3$CFU/mLC2- Operator 129161149252634331716
C3- Operator 2254023630141872332890
C4- Operator 3228362538111530027002
All Instruments/Operators258003734151530032890
Enterotoxigenic E. coli 078:H11H10407 (ETEC)ATCC 35401ST-1aModeratePositive3xLoD$8.40 x 10^2$CFU/mLC2- Operator 135769100933367738306
C3- Operator 2343584831142539842656
C4- Operator 3287303638132225435493
All Instruments/Operators329524648142225442656
Low Positive1.5xLoD$4.20 x 10^2$CFU/mLC2- Operator 132795224072853636716
C3- Operator 2307263040102421935644
C4- Operator 3249072765112023630170
All Instruments/Operators294764291152023636716
SalmC2- Operator 1152911496101258818918
OrganismTargetProbeTest LevelInstrument/OperatorMeanStDev%CVMinMax
Salmonellaenterica ssp.entericaATCC 14028ModeratePositive3xLoD3.30 x 103CFU/mLC3- Operator 213931264519628418489
C4- Operator 315701145691330619184
All Instruments/Operators14974206814628419184
Low Positive1.5xLoD1.65 x 103CFU/mLC2- Operator 117247110861515319682
C3- Operator 2148801682111084719127
C4- Operator 316234145571330119674
All Instruments/Operators161201721101084719682
Shiga-like toxinproducing E. coli(STEC)ATCC BAA-2217stx2ModeratePositive3xLoD3.75 x 103CFU/mLC2- Operator 127217111242360630023
C3- Operator 231888131642918334448
C4- Operator 326310151562297030028
All Instruments/Operators284722785102297034448
Low Positive1.5xLoD1.88 x103CFU/mLC2- Operator 127467154762391430339
C3- Operator 231460150152800234121
C4- Operator 325293209782034029403
All Instruments/Operators280733089112034034121
Shigella sonneiATCC 29930ShigModeratePositive3xLoD6.60 x 102CFU/mLC2- Operator 117531117371532719469
C3- Operator 215369378125757321561
C4- Operator 3164542056131191021114
All Instruments/Operators16451269716757321561
Low Positive1.5xLoD3.30 x 102CFU/mLC2- Operator 113919186213923317022
C3- Operator 212242250320643516434
C4- Operator 312832218417835217489
All Instruments/Operators12998228318643517489
Vibrioparahaemolyticus ATCC 17802V.paraModeratePositive3xLoD1.95 x101CFU/mLC2- Operator 1236052305101775029798
C3- Operator 2214583112151313827526
C4- Operator 317486282116854322789
All Instruments/Operators20850373918854329798
Low Positive1.5xLoD9.75x100CFU/mLC2- Operator 1226282402111714526762
C3- Operator 218000606734604327526
C4- Operator 316263306019698923597
All Instruments/Operators18964492226604327526
YersiniaenterocoliticaATCC 23715Y.entModeratePositive3xLoD2.25 x103CFU/mLC2- Operator 1143651847131020917892
C3- Operator 2153031786121081317893
C4- Operator 314256232516952319982
All Instruments/Operators14641203314952319982
Low Positive1.5xLoD1.13 x103CFU/mLC2- Operator 1147411694111185319628
C3- Operator 2155922014131297820765
C4- Operator 3147322253161075019994
All Instruments/Operators150222019131075020765
OrganismTargetProbeTest LevelInstrument/OperatorMeanStDev%CVMinMax
Cryptosporidium parvumwaterborneP102CryptoModeratePositive3xLoD$9.30 x 10^3$oocysts/mLC2- Operator 118731432323990428443
C3- Operator 2208433978191152628678
C4- Operator 3*14867445731566926611
All Instruments/Operators18222487326566928678
Low Positive1.5xLoD$4.65 x 10^3$oocysts/mLC2- Operator 113816517037257224148
C3- Operator 215402399626816622933
C4- Operator 3*10785340933441318053
All Instruments/Operators13363463135257224148
Giardiaintestinalis (akaG. lamblia)waterborneP101G.lamModeratePositive3xLoD$2.70 x 10^3$cysts/mLC2- Operator 114327448731769630050
C3- Operator 221743500623893231538
C4- Operator 3169633780221152725036
All Instrument/Operators17677537730769631538
Low Positive1.5xLoD$1.35 x 10^3$cysts/mLC2- Operator 1173403668211133825626
C3- Operator 2247514701191247831282
C4- Operator 3198213772201267725949
All Instruments/Operators206375081251133831282
RotavirusATCC VR-2018RotaModeratePositive3xLoD$3.75 x 10^3$TCID50/mLC2- Operator 139523208653583544023
C3- Operator 239162340092886344274
C4- Operator 332899295682510738361
All Instruments/Operators371954170132510744274
Low Positive1.5xLoD$1.88 x 10^2$TCID50/mLC2- Operator 1365155114141612740941
C3- Operator 2363123863112649141547
C4- Operator 3319253280102111436158
All Instruments/Operators349174628111612741547

{24}------------------------------------------------

{25}------------------------------------------------

*The one replicate with 'Not detected' result was not included in the calculations.

SINGLE VS. MULTI-SPIKE

To demonstrate that the LoD is equivalent when spiking with single versus multiple organisms, LoD for representative organisms (S. enterica, C. difficile, G. lamblia, and Rotavirus) were performed as single targets and with all representative organisms combined, using both extraction systems. The results of spiking single versus multiple organisms showed equivalent LoD (see Table below). LoD for the remaining organisms were performed in pairs, except for norovirus which was tested from positive clinical specimens.

Single and multi-spiked samples both achieved ≥95% detection of 20 replicates (≥19 out of 20) at same concentrations for the challenge organisms (concentrations indicated in the tables below).

Table. Comparison of results for limit of detection testing with single or multiple spiked samples for easyMAG extractions assayed with BioCode GPP.

{26}------------------------------------------------

StrainSingleSpike/MultipleSpikeSourceTargetProbeUnpreservedStool LoDUnpreserved StoolDetectionCary-Blair StoolLoDCary-BlairStoolDetection
Clostridiumdifficile(toxinotype 0)SingleSpikeATCC 9689tcdB$9.50 x 10^1$CFU/mL20/20$9.50 x 10^1$CFU/mL20/20
MultiSpike20/2019/20
Salmonellaenterica ssp.entericaSingleSpikeATCC 14028Salm$1.10 x 10^3$CFU/mL20/20$1.10 x 10^3$CFU/mL20/20
MultiSpike20/2020/20
Giardiaintestinalis(aka G. lamblia)SingleSpikewaterborneP101G.lam$9.00 x 10^2$cysts/mL20/20$9.00 x 10^2$cysts/mL20/20
MultiSpike20/2020/20
Rotavirus ASingleSpikeATCC VR-2018Rota$2.50 x 10^3$TCID50/mL20/20$1.25 x 10^3$TCID50/mL20/20
MultiSpike20/2020/20

Table. Comparison of results for limit of detection testing with single or multiple spiked samples for MP96 extractions assayed with BioCode GPP.

StrainSingleSpike/MultipleSpikeSourceTargetProbeUnpreservedStool LoDUnpreserved StoolDetectionCary-Blair StoolLoDCary-BlairStoolDetection
Clostridiumdifficile(toxinotype 0 )SingleSpikeMultiSpikeATCC 9689tcdB$9.50 x 10^1$CFU/mL20/2020/20$9.50 x 10^1$CFU/mL20/2020/20
Salmonellaenterica ssp.entericaSingleSpikeMultiSpikeATCC 14028Salm$1.10 x 10^3$CFU/mL20/2020/20$1.10 x 10^3$CFU/mL20/2020/20
Giardiaintestinalis(aka G. lamblia)SingleSpikeMultiSpikewaterborneP101G.lam$9.00 x 10^2$cysts/mL20/2020/20$9.00 x 10^2$cysts/mL20/2020/20
Rotavirus ASingleSpikeMultiSpikeATCC VR-2018Rota$1.25 x 10^3$TCID50/mL20/2020/20$1.25 x 10^3$TCID50/mL20/2020/20

LIMIT OF DETECTION (LoD)

{27}------------------------------------------------

A study was performed to assess the performance of the BioCode GPP on the BioCode MDx 3000 at the Limit of Detection (LoD) for both unpreserved Stool and Cary-Blair specimens. In this study the Gl Panel was tested with quantified bacteria, virus or parasite stocks (except norovirus which used clinical samples). For initial screening, four replicates of each concentration (near LoD for the predicate) in negative stool and Cary-Blair were extracted on the easyMAG and MagNA Pure 96 Systems and tested in singlet with the BioCode GPP on the BioCode MDx 3000 system. The LoD was confirmed by extracting 20 replicates of each sample type/extraction method and testing each in singlet for a total of 20 replicates at or near presumptive LoD. LoD for each stock was defined as the lowest concentration with ≥95% detection of 20 replicates (19 out of 20), and was determined separately unpreserved stool and Cary-Blair preserved stool. The LoDs determined from the quantitated stocks or clinical samples are presented below and in the labeling. The LoDs were the same or within 2-fold for each extraction system.

EasyMagMagNA Pure 96
StrainSourceUnpreserved StoolLoDDetectionUnpreserved StoolLoDDetection
Campylobacter coliATCC 335592.81 x 101 CFU/mL19/202.81 x 101 CFU/mL20/20
Campylobacter jejunispp. jejuniATCC 332923.50 x 102 CFU/mL19/203.50 x 102 CFU/mL19/20
Clostridium difficile(toxinotype 0 )ATCC 96899.50 x 101 CFU/mL20/209.50 x 101 CFU/mL20/20
Clostridium difficile(toxinotype III; Nap1)Zeptometrix0801619cf4.15 x 102 CFU/mL20/204.15 x 102 CFU/mL20/20
Enteroaggregative E.coli 092:H33 (EAEC)STEC TW044407.00 x 102 CFU/mL20/207.00 x 102 CFU/mL20/20
Enteroinvasive E. coli029:NM (EIEC)ATCC 438923.60 x 102 CFU/mL20/201.80 x 102 CFU/mL20/20
Enterotoxigenic E. coli078:H11 H10407(ETEC)ATCC 354012.80 x 102 CFU/mL20/202.80 x 102 CFU/mL20/20
Salmonella bongoriSGSC 49001.40 x 103 CFU/mL20/201.40 x 103 CFU/mL20/20
Salmonella entericassp. entericaATCC 140281.10 x 103 CFU/mL20/201.10 x 103 CFU/mL20/20
Shiga-like toxinproducing E. coli (STEC)ATCC BAA-22171.25 x 103 CFU/mL20/201.25 x 103 CFU/mL20/20
E. coli 0157ATCC 7003761.65 x 103 CFU/mL20/201.65 x 103 CFU/mL20/20
Shigella sonneiATCC 299302.20 x 102 CFU/mL20/202.20 x 102 CFU/mL20/20
Vibrio choleraeATCC 258702.45 x 102 CFU/mL20/202.45 x 102 CFU/mL20/20
VibrioparahaemolyticusATCC 178026.50 x 100 CFU/mL20/206.50 x 100 CFU/mL20/20

Table. Comparison of results for limit of detection testing for Unpreserved Stool extracted with the easyMAG or MP96 and assayed with BioCode GPP.

{28}------------------------------------------------

EasyMagMagNA Pure 96
StrainSourceUnpreserved StoolLoDDetectionUnpreserved StoolLoDDetection
Yersinia enterocoliticaATCC 23715$7.50 x 10^2$ CFU/mL20/20$7.50 x10^2$ CFU/mL20/20
CryptosporidiumparvumwaterborneP102C$3.10 x 10^3$ oocysts/mL20/20$3.10 x10^3$ oocysts/mL20/20
Entamoeba histolyticaHB-301:NIHBEI NR-178$1.55 x 10^{-1}$ cysts/mL20/20$1.55 x10^{-1}$ cysts/mL20/20
Giardia intestinalis (akaG. lamblia)waterborneP101$9.00 x 10^2$ cysts/mL20/20$9.00 x10^2$ cysts/mL20/20
Adenovirus 40 (dugan)Zeptometrix0810084$2.00 x 10^{-1}$ TCID50/mL20/20$1.00 x10^{-1}$ TCID50/mL20/20
Adenovirus 41 (TAK)Zeptometrix0810085$9.4 x 10^{-2}$ TCID50/mL20/20$4.70 x10^{-2}$ TCID50/mL20/20
Rotavirus AATCC VR-2018$2.5 x 10^3$ TCID50/mL20/20$1.25 x10^3$ TCID50/mL20/20

Table. Comparison of results for limit of detection testing for Cary-Blair Stool extracted with the easyMAG or MP96 and assayed with BioCode GPP.

StrainSourceCary-Blair Stool LoDDetectionCary-Blair Stool LoDDetection
Campylobacter coliATCC 33559$5.61 x 10^1$ CFU/mL20/20$2.81 x 10^1$ CFU/mL20/20
Campylobacter jejunispp. jejuniATCC 33292$3.50 x10^2$ CFU/mL20/20$7.00 x10^2$ CFU/mL20/20
Clostridium difficile(toxinotype 0 )ATCC 9689$9.50 x 10^1$ CFU/mL20/20$9.50 x 10^1$ CFU/mL20/20
Clostridium difficile(toxinotype III; Nap1)Zeptometrix0801619cf$4.15 x10^2$ CFU/mL20/20$4.15 x10^2$ CFU/mL20/20
Enteroaggregative E.coli 092:H33 (EAEC)STEC TW04440$7.00 x10^2$ CFU/mL20/20$7.00 x10^2$ CFU/mL20/20
Enteroinvasive E. coli029:NM (EIEC)ATCC 43892$3.60 x10^2$ CFU/mL20/20$1.80 x10^2$ CFU/mL20/20
Enterotoxigenic E. coli078:H11 H10407(ETEC)ATCC 35401$2.80 x10^2$ CFU/mL20/20$2.80 x10^2$ CFU/mL19/20
Salmonella bongoriSGSC 4900$1.40 x10^3$ CFU/mL19/20$1.40 x10^3$ CFU/mL20/20
Salmonella entericassp. entericaATCC 14028$1.10 x10^3$ CFU/mL20/20$1.10 x10^3$ CFU/mL20/20
Shiga-like toxinproducing E. coli (STEC)ATCC BAA-2217$1.25 x10^3$ CFU/mL20/20$1.25 x10^3$ CFU/mL20/20
E. coli 0157ATCC 700376$3.30 x10^3$ CFU/mL20/20$1.65 x10^3$ CFU/mL20/20

{29}------------------------------------------------

Premarket Notification 510(k)
-----------------------------------
EasyMagMagNA Pure 96
StrainSourceCary-Blair Stool LoDDetectionCary-Blair Stool LoDDetection
Shigella sonneiATCC 299302.20 x102 CFU/mL20/202.20 x102 CFU/mL20/20
Vibrio choleraeATCC 258702.45 x102 CFU/mL20/202.45 x102 CFU/mL20/20
VibrioparahaemolyticusATCC 178026.50 x100 CFU/mL20/206.50 x100 CFU/mL20/20
Yersinia enterocoliticaATCC 237157.50 x102 CFU/mL20/207.50 x102 CFU/mL20/20
CryptosporidiumparvumwaterborneP102C3.10 x103 oocysts/mL20/203.10 x103 oocysts/mL20/20
Entamoeba histolyticaHB-301:NIHBEI NR-1781.55 x10-1 cysts/mL20/201.55 x10-1 cysts/mL20/20
Giardia intestinalis (akaG. lamblia)waterborneP1019.00 x102 cysts/mL20/209.00 x102 cysts/mL20/20
Adenovirus 40 (dugan)Zeptometrix08100842.00 x10-1 TCID50/mL20/201.00 x10-1 TCID50/mL20/20
Adenovirus 41 (TAK)Zeptometrix08100854.70 x10-2 TCID50/mL20/204.70 x10-2 TCID50/mL20/20
Rotavirus AATCC VR-20181.25 x103 TCID50/mL20/201.25 x103 TCID50/mL20/20

For Norovirus GI and GII targets, positive clinical specimens were used, and serial dilutions (initial 10-fold dilution series followed by finer dilutions) were performed. Four replicates of each dilution in negative unpreserved stool and Cary-Blair stool were extracted with the easyMAG and MagNA Pure 96 Systems and tested with the BioCode GPP on the BioCode MDx 3000 system. The LoD was confirmed by extracting 20 replicates of each sample type with each extraction method and testing at or near presumptive LoD. For unpreserved stool, LoD with the easyMag extraction was 2-fold and 8.3-fold lower than the MagNA Pure 96 extraction for Norovirus GI and Norovirus GII, respectively. For Cary-Blair stool, LoD with the easyMag extraction was less than 2-fold lower than the MagNA Pure 96 extraction for both Norovirus Gl and GII.

Table. Norovirus - Comparison of results for limit of detection testing for Unpreserved Stool extracted with the easyMAG or MP96 and assayed with BioCode GPP.

TargetSourceTargetProbeEasyMagMagNA Pure 96
UnpreservedStool DilutionDetectionUnpreservedStool DilutionDetection
Norovirus GIClinical Sample ID#60NoVG11:10,00020/201:5,00019/20
Norovirus GIIClinical Sample ID#54NoVG21:250,00020/201:30,00020/20

Table. Norovirus - Comparison of results for limit of detection testing for Cary-Blair Stool extracted with the easyMAG or MP96 and assayed with BioCode GPP.

{30}------------------------------------------------

TargetSourceTargetProbeEasyMagMagNA Pure 96
Cary-BlairStool DilutionDetectionCary-BlairStool DilutionDetection
Norovirus GIClinical Sample ID#60NoVG11:50,00020/201:80,00019/20
Norovirus GIIClinical Sample ID#54NoVG21:100,00020/201:80,00020/20

Conclusion:

The intended use and fundamental scientific technology of the BioCode GPP using the MagNA Pure 96 is substantially equivalent to the predicate device. Clinical and non-clinical studies have established that the BioCode GPP with the MagNA Pure 96 is substantially equivalent to the predicate device.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).