QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini B

K254032 · QIAGEN GmbH · PCH · Mar 9, 2026 · Microbiology

Device Facts

Record IDK254032
Device NameQIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini B
ApplicantQIAGEN GmbH
Product CodePCH · Microbiology
Decision DateMar 9, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3990
Device ClassClass 2

Intended Use

The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2: Adenovirus F40/F41 Astrovirus Norovirus GI/GII Rotavirus A Campylobacter (C. jejuni, C. coli and C. upsaliensis) Shigella/Enteroinvasive Escherichia coli (EIEC) Enteropathogenic Escherichia coli (EPEC) Enterotoxigenic Escherichia coli (ETEC) lt/st Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC) Salmonella Plesiomonas shigelloides Yersinia enterocolitica Cryptosporidium Cyclospora cayetanensis Entamoeba histolytica Giardia lamblia* *Also known as Giardia intestinalis and Giardia duodenalis Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V: Norovirus Campylobacter Shigella Shiga-like toxin E. coli (STEC) Salmonella Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B: Campylobacter Shigella Shiga-like toxin Escherichia coli (STEC) Salmonella Yersinia enterocolitica Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Device Story

Multiplexed nucleic acid test for gastrointestinal pathogens; uses QIAstat-Dx Analyzer 2.0 or QIAstat-Dx Rise instruments. Input: preserved stool samples (Para-Pak C&S or FecalSwab). Process: automated sample pre-treatment, cell lysis, nucleic acid purification, and multiplex real-time RT-PCR within self-contained cartridges. Output: qualitative detection/identification of specific pathogens displayed on instrument screen. Used in clinical laboratory settings by trained personnel. Results aid diagnosis of GI illness; clinicians use output alongside clinical/epidemiological data to guide patient management. Benefits: rapid, automated identification of causative agents, reducing time to diagnosis.

Clinical Evidence

Bench testing only. Reproducibility study conducted at three sites with 90 replicates per analyte level demonstrated acceptable performance. Instrument equivalency study compared performance on QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise, confirming equivalent hit rates and Ct means for all targets. Carry-over study on QIAstat-Dx Rise showed no cross-contamination.

Technological Characteristics

Multiplexed nucleic acid test; silica membrane-based nucleic acid extraction; real-time RT-PCR amplification; fluorescence-based detection. Self-contained cartridges with pre-loaded reagents; pneumatic microfluidics for sample/fluid transfer. QIAstat-Dx Rise platform supports up to 8 analytical modules and 18 cartridges.

Indications for Use

Indicated for individuals with signs/symptoms of gastrointestinal infection to aid in diagnosis of specific viral, bacterial, and parasitic agents. Requires concomitant culture for bacterial typing. Not for ruling out co-infections or non-infectious causes of gastroenteritis.

Regulatory Classification

Identification

A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

Special Controls

*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 9, 2026 QIAGEN GmbH % Sonia Pablo Senior Manager, Regulatory Affairs STAT-Dx Life, S.L. (A QIAGEN Company) Carrer Baldiri Reixac, 4 Barcelona, 08028 Spain Re: K254032 Trade/Device Name: QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V QIAstat-Dx GI Panel 2 Mini B Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal microorganism multiplex nucleic acid-based assay Regulatory Class: Class II Product Code: PCH Dated: December 16, 2025 Received: December 16, 2025 Dear Sonia Pablo: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K254032 - Sonia Pablo Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K254032 - Sonia Pablo Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, BRYAN M. GRABIAS -S 2026.03.09 14:05:53 -04'00' Bryan Grabias, Ph. D. Acting Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K254032 Device Name QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V QIAstat-Dx GI Panel 2 Mini B Indications for Use (Describe) QIAstat-Dx Gastrointestinal Panel 2: The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2: - Adenovirus F40/F41 - Astrovirus - Norovirus GI/GII - Rotavirus A - Campylobacter (C. jejuni, C. coli and C. upsaliensis) - Shigella/Enteroinvasive Escherichia coli (EIEC) - Enteropathogenic Escherichia coli (EPEC) - Enterotoxigenic Escherichia coli (ETEC) lt/st - Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC) - Salmonella - Plesiomonas shigelloides - Yersinia enterocolitica - Cryptosporidium - Cyclospora cayetanensis - Entamoeba histolytica - Giardia lamblia* *Also known as Giardia intestinalis and Giardia duodenalis Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. QIAstat-Dx GI Panel 2 Mini B&V: The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V: FORM FDA 3881 (8/23) Page 1 of 2 PSC Publishing Services (301) 443-6740 {4} - Norovirus - Campylobacter - Shigella - Shiga-like toxin E. coli (STEC) - Salmonella Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. ## QIAstat-Dx GI Panel 2 Mini B: The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B: - Campylobacter - Shigella - Shiga-like toxin Escherichia coli (STEC) - Salmonella - Yersinia enterocolitica Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 2 of 2 PSC Publishing Services (301) 443-6740 {5} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 1 of 14 # 510(k) Summary ## General Information Submitted by: QIAGEN GmbH QIAGEN Strasse 1 Hilden, Germany 40724 Contact Person: Sonia Pablo Senior Manager, Regulatory Affairs STAT-Dx Life, S.L. (A QIAGEN Company) Carrer Baldiri Reixac, 4 08028 Barcelona Mobile: +34 696 85 81 85 Email: sonia.pablo@qiagen.com Date Prepared: December 15, 2025 Device Name: QIAstat-Dx Gastrointestinal Panel 2 QIAstat-Dx GI Panel 2 Mini B&V QIAstat-Dx GI Panel 2 Mini B Classification: 21 CFR 866.3990 - Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System Product Code: PCH Predicate Device: QIAstat-Dx Gastrointestinal Panel 2, K252329 QIAstat-Dx GI Panel 2 Mini B&V, K252329 QIAstat-Dx GI Panel 2 Mini B, K252329 December 15, 2025 {6} QIAGEN QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary 510(k) Premarket Notification Page 2 of 14 # Device Description The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B ("QIAstat-Dx Gastrointestinal Panels") are multiplexed nucleic acid tests which are designed for use with the QIAstat-Dx Analyzer 2.0. The device modification is to add the QIAstat-Dx Rise as an additional instrument for use with the QIAstat-Dx Gastrointestinal Panels. The QIAstat-Dx Rise is a higher throughput platform, incorporating up to eight QIAstat-Dx Analytical Modules (AM) on a small footprint. The instrument allows queuing up to 18 cartridges, with processing being limited up to 16 cartridges for the QIAstat-Dx Gastrointestinal Panels. The cartridges are scheduled for processing and delivered to the appropriate AM by an integrated robotic handler. The AM used with the QIAstat-Dx Rise is the same AM that can be used with the QIAstat-Dx Analyzer 2.0. The modified QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are identical to the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B (K252329), respectively, with the exception of the Instructions for Use which were updated to include the assay-specific procedure for use with the QIAstat-Dx Rise. The QIAstat-Dx Gastrointestinal Panels are intended for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple targets directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. All the reagents required for the complete execution of the test are pre-loaded and self-contained in the QIAstat-Dx Gastrointestinal Panels. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators. This eliminates any possibility of exposure of the user or the analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges. Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially: - Sample pre-treatment for PCR Inhibitors removal - Resuspension of Internal Control and Proteinase K - Cell lysis using mechanical and/or chemical means - Membrane-based nucleic acid purification - Rehydration of Master Mix - Transfer of defined aliquots of eluate/master mix to different reaction chambers - Performance of multiplex real-time RT-PCR testing within each reaction chamber. December 15, 2025 {7} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 3 of 14 The QIAstat-Dx Gastrointestinal Panels Assay Definition File (ADF) automatically interprets test results and displays a summary on the instrument display screen. The detected analytes are displayed in red. All other tested but not detected analytes are listed in green. The instrument will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show “FAIL”). ## Intended Use ## QIAstat-Dx Gastrointestinal Panel 2 The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic *E. coli*/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2: - Adenovirus F40/F41 - Astrovirus - Norovirus GI/GII - Rotavirus A - *Campylobacter* (*C. jejuni*, *C. coli* and *C. upsaliensis*) - Shigella/Enteroinvasive *Escherichia coli* (EIEC) - Enteropathogenic *Escherichia coli* (EPEC) - Enterotoxigenic *Escherichia coli* (ETEC) lt/st - Shiga-like toxin-producing *Escherichia coli* (STEC) stx1/stx2 (including specific identification of *E. coli* O157 serogroup within STEC) - Salmonella - *Plesiomonas shigelloides* - Yersinia enterocolitica - Cryptosporidium - *Cyclospora cayetanensis* - Entamoeba histolytica - Giardia lamblia* *Also known as Giardia intestinalis and Giardia duodenalis* Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease. December 15, 2025 {8} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 4 of 14 Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. ## QIAstat-Dx GI Panel 2 Mini B&V The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V: - Norovirus - Campylobacter - Shigella - Shiga-like toxin E. coli (STEC) - Salmonella Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstatDx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. ## QIAstat-Dx GI Panel 2 Mini B The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B: - Campylobacter - Shigella - Shiga-like toxin Escherichia coli (STEC) December 15, 2025 {9} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 5 of 14 - Salmonella - Yersinia enterocolitica Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstatDx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. December 15, 2025 {10} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary # Comparison of the QIAstat-Dx Gastrointestinal Panels with QIAstat-Dx Rise and the Predicate Devices Similarities and differences between the QIAstat-Dx Gastrointestinal Panel 2, QIAstat-Dx GI Panel 2 Mini B&V, QIAstat-Dx GI Panel 2 Mini B with QIAstat-Dx Rise and the predicate device are shown in Table 1, Table 2, and Table 3, respectively. Table 1: Comparison of the QIAstat-Dx Gastrointestinal Panel 2 with QIAstat-Dx Rise with the predicate device | Characteristic | Subject Device | Predicate | | --- | --- | --- | | Name | QIAstat-Dx Gastrointestinal Panel 2 | QIAstat-Dx Gastrointestinal Panel 2 | | 510(k) No. | TBD | K252329 | | Regulation | 21 CFR 866.3990 | 21 CFR 866.3990 | | Product Code | PCH | PCH | | Device Class | Class II | Class II | | Similarities | | | | Intended Use/Indications for Use | The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2: • Adenovirus F40/F41 • Astrovirus • Norovirus GI/GII • Rotavirus A • Campylobacter (C. jejuni, C. coli and C. upsaliensis) | The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2: • Adenovirus F40/F41 • Astrovirus • Norovirus GI/GII • Rotavirus A • Campylobacter (C. jejuni, C. coli and C. upsaliensis) • Shigella/Enteroinvasive Escherichia coli (EIEC) | December 15, 2025 {11} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary | Characteristic | Subject Device | Predicate | | --- | --- | --- | | | • Shigella/Enteroinvasive Escherichia coli (EIEC) • Enteropathogenic Escherichia coli (EPEC) • Enterotoxigenic Escherichia coli (ETEC) It/st • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC) • Salmonella • Plesiomonas shigelloides • Yersinia enterocolitica • Cryptosporidium • Cyclospora cayetanensis • Entamoeba histolytica • Giardia lamblia* *Also known as Giardia intestinalis and Giardia duodenalis | • Enteropathogenic Escherichia coli (EPEC) • Enterotoxigenic Escherichia coli (ETEC) It/st • Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC) • Salmonella • Plesiomonas shigelloides • Yersinia enterocolitica • Cryptosporidium • Cyclospora cayetanensis • Entamoeba histolytica • Giardia lamblia* *Also known as Giardia intestinalis and Giardia duodenalis | | | Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease. | Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease. | | | | Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis | December 15, 2025 {12} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 8 of 14 | Characteristic | Subject Device | Predicate | | --- | --- | --- | | | Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | | Specimen Type | Same | Preserved stool in Para-Pak C&S or FecalSwab transport media | | Amplification and Detection Technology | Same | PCR | | Assay Controls | Same | One internal control in each cartridge to control for sample processing that is subjected to all nucleic acid extraction and amplification steps similar to patient samples. Labeling recommends use of negative and positive external controls regularly. Use transport media as the external Negative Control and previously characterized positive samples or negative sample spiked with well characterized target organisms as external positive controls. | | Nucleic Acid Extraction | Same | Extraction of nucleic acids using silica membrane | | Technology | Same | Detection of amplified targets uses an increase in fluorescence due to specific probe binding to generate the assay results. | | Operational | Same | The sample is loaded straight into the cartridge. | | Assay Targets | Same | Detects sixteen (16) targets | December 15, 2025 {13} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Table 2: Comparison of the QIAstat-Dx GI Panel 2 Mini B&V with QIAstat-Dx Rise with the predicate device | Characteristic | Subject Device | Predicate | | --- | --- | --- | | Name | QIAstat-Dx GI Panel 2 Mini B&V | QIAstat-Dx GI Panel 2 Mini B&V | | 510(k) No. | TBD | K252329 | | Regulation | 21 CFR 866.3990 | 21 CFR 866.3990 | | Product Code | PCH | PCH | | Device Class | Class II | Class II | | Similarities | | | | Intended Use/Indications for Use | The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V: • Norovirus • Campylobacter • Shigella • Shiga-like toxin E. coli (STEC) • Salmonella Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The | The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V: • Norovirus • Campylobacter • Shigella • Shiga-like toxin E. coli (STEC) • Salmonella Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The | December 15, 2025 {14} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary | Characteristic | Subject Device | Predicate | | --- | --- | --- | | | typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstatDx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | | Specimen Type | Same | Preserved stool in Para-Pak C&S or FecalSwab transport media | | Amplification and Detection Technology | Same | PCR | | Assay Controls | Same | One internal control in each cartridge to control for sample processing that is subjected to all nucleic acid extraction and amplification steps similar to patient samples. Labeling recommends use of negative and positive external controls regularly. Use transport media as the external Negative Control and previously characterized positive samples or negative sample spiked with well characterized target organisms as external positive controls. | December 15, 2025 {15} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 11 of 14 | Characteristic | Subject Device | Predicate | | --- | --- | --- | | Nucleic Acid Extraction | Same | Extraction of nucleic acids using silica membrane | | Technology | Same | Detection of amplified targets uses an increase in fluorescence due to specific probe binding to generate the assay results. | | Operational | Same | The sample is loaded straight into the cartridge. | | Assay Targets | Same | Detects five (5) targets | | Differences | | | | Amplification and Detection Instrument System | QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise | QIAstat-Dx Analyzer 2.0 | Table 3: Comparison of the QIAstat-Dx GI Panel 2 Mini B with QIAstat-Dx Rise with the predicate device | Characteristic | Subject Device | Predicate | | --- | --- | --- | | Name | QIAstat-Dx GI Panel 2 Mini B | QIAstat-Dx GI Panel 2 Mini B | | 510(k) No. | TBD | K252329 | | Regulation | 21 CFR 866.3990 | 21 CFR 866.3990 | | Product Code | PCH | PCH | | Device Class | Class II | Class II | | Similarities | | | | Intended Use/Indications for Use | The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B: • Campylobacter | The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0. for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B: • Campylobacter | December 15, 2025 {16} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary | Characteristic | Subject Device | Predicate | | --- | --- | --- | | | • Campylobacter • Shigella • Shiga-like toxin Escherichia coli (STEC) • Salmonella • Yersinia enterocolitica Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | • Shigella • Shiga-like toxin Escherichia coli (STEC) • Salmonella • Yersinia enterocolitica Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease. Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | | Specimen Type | Same | Preserved stool in Para-Pak C&S or FecalSwab transport media | | Amplification and Detection Technology | Same | PCR | December 15, 2025 {17} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 13 of 14 | Characteristic | Subject Device | Predicate | | --- | --- | --- | | Assay Controls | Same | One internal control in each cartridge to control for sample processing that is subjected to all nucleic acid extraction and amplification steps similar to patient samples. Labeling recommends use of negative and positive external controls regularly. Use transport media as the external Negative Control, and previously characterized positive samples or negative sample spiked with well characterized target organisms as external positive controls. | | Nucleic Acid Extraction | Same | Extraction of nucleic acids using silica membrane | | Technology | Same | Detection of amplified targets uses an increase in fluorescence due to specific probe binding to generate the assay results. | | Operational | Same | The sample is loaded straight into the cartridge. | | Assay Targets | Same | Detects five (5) targets | | **Differences** | | | | Amplification and Detection Instrument System | QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise | QIAstat-Dx Analyzer 2.0 | December 15, 2025 {18} QIAGEN 510(k) Premarket Notification QIAstat-Dx® Gastrointestinal Panels with QIAstat-Dx Rise 510(k) Summary Page 14 of 14 # Summary of Performance Data The performance for the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B with QIAstat-Dx Rise is equivalent to the performance for the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B (K252329). The following analytical studies were performed on the QIAstat-Dx Rise and successfully demonstrated the equivalent performance of the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B: - Equivalence at Low Analyte Concentration - Carryover - Reproducibility # Conclusions The technological characteristics and the indications for use of the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B remain the same. Validation data and information submitted in this premarket notification to add QIAstat-Dx Rise demonstrate that the modified QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are substantially equivalent to the predicate devices. December 15, 2025
Innolitics FDA 510(k) Search
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%