K Number
K251993

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-09-25

(90 days)

Product Code
Regulation Number
866.3990
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.

This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.

Device Description

The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.

The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.

Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.

Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).

Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction. Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.

The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:

AnalyteGene TargetedInstrument Channel
Yersinia enterocoliticaInvA (Invasive antigen A)FAM
Vibrio parahaemolyticusgyrB (Gyrase B)HEX
Vibrio vulnificusgyrB (Gyrase B)HEX
Vibrio choleraeompW (Outer Membrane Protein W)HEX
Escherichia coli O157rfbE (Perosamine synthase-O-antigen)ROX
Plesiomonas shigelloideshugA (Heme utilization gene A)RED647
Internal ControlNot ApplicableRED677

Assay Components: The assay components configuration for the Panther Fusion GI Expanded Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Expanded Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Expanded Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:

  1. Aptima Assay Fluids Kit (303014)
  2. Aptima Multitest Swab Specimen Collection Kit (PRD-03546)

Table 1: Reagents Required to Perform the Panther Fusion GI Expanded Bacterial Assay

BoxComponents Description
Refrigerated BoxPanther Fusion GI Expanded Bacterial Assay Cartridges
Room Temperature BoxPanther Fusion Extraction Reagent-B• Panther Fusion Capture Reagent-B bottles• Panther Fusion Enhancer Reagent-B bottles
Refrigerated BoxPanther Fusion Internal Control-B
Room Temperature BoxPanther Fusion Reconstitution Buffer I
Room Temperature BoxPanther Fusion Elution Buffer
Room Temperature BoxPanther Fusion Oil
Refrigerated BoxPanther Fusion GI Expanded Bacterial Assay Controls• Panther Fusion GI Expanded Bacterial Positive Control• Panther Fusion Negative Control

Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Expanded Bacterial Assay

Aptima Assay Fluids Kit
Aptima Multitest Swab Specimen Collection Kit

Instrumentation: The Panther Fusion GI Expanded Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Panther Fusion GI Expanded Bacterial Assay

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 25, 2025

Hologic Inc.
Patricia Villarreal
Regulatory Affairs Specialist
10210 Genetic Center Drive
San Diego, California 92121

Re: K251993
Trade/Device Name: Panther Fusion GI Expanded Bacterial Assay
Regulation Number: 21 CFR 866.3990
Regulation Name: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assay
Regulatory Class: Class II
Product Code: PCH, OOI
Dated: July 26, 2025
Received: August 26, 2025

Dear Patricia Villarreal:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

Sincerely,

Bryan M. Grabias -S [Digitally signed by Bryan M. Grabias -S Date: 2025.09.25 11:10:47 -04'00']

Bryan Grabias, Ph.D.
Acting Branch Chief
Bacterial Respiratory and Medical Countermeasures Branch
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251993

Device Name: Panther Fusion GI Expanded Bacterial Assay

Indications for Use (Describe)

The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.

This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY

Panther Fusion GI Expanded Bacterial Assay

I. SUBMITTER:

Hologic, Inc.
10210 Genetic Center Drive
San Diego, CA 92121

Contact Person: Patricia Villarreal
Regulatory Affairs Specialist
Patricia.Villarreal@hologic.com
Phone: 562.833.7906
Fax: N/A

Date Prepared: June 26, 2025

II. DEVICE

Proprietary Name: Panther Fusion GI Expanded Bacterial Assay

Classification Name: Gastrointestinal microorganism multiplex nucleic acid-based assay

Regulation Number: 21 CFR 866.3990 and 21 CFR 862.2570

Regulatory Class: Class II

Product Code: PCH and OOI

III. PREDICATE DEVICE

The predicate device for the Panther Fusion GI Expanded Bacterial Assay is the BioFire FilmArray Gastrointestinal (GI) Panel (K160459, K143005 & K140407).

Page 6

IV. DEVICE DESCRIPTION

The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.

The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.

Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.

Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).

Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction.

Page 7

Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.

The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:

AnalyteGene TargetedInstrument Channel
Yersinia enterocoliticaInvA (Invasive antigen A)FAM
Vibrio parahaemolyticusgyrB (Gyrase B)HEX
Vibrio vulnificusgyrB (Gyrase B)HEX
Vibrio choleraeompW (Outer Membrane Protein W)HEX
Escherichia coli O157rfbE (Perosamine synthase-O-antigen)ROX
Plesiomonas shigelloideshugA (Heme utilization gene A)RED647
Internal ControlNot ApplicableRED677

Assay Components

The assay components configuration for the Panther Fusion GI Expanded Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Expanded Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Expanded Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:

  1. Aptima Assay Fluids Kit (303014)
  2. Aptima Multitest Swab Specimen Collection Kit (PRD-03546)

Page 8

Table 1: Reagents Required to Perform the Panther Fusion GI Expanded Bacterial Assay

BoxComponents Description
Refrigerated BoxPanther Fusion GI Expanded Bacterial Assay Cartridges
Room Temperature BoxPanther Fusion Extraction Reagent-B• Panther Fusion Capture Reagent-B bottles• Panther Fusion Enhancer Reagent-B bottles
Refrigerated BoxPanther Fusion Internal Control-B
Room Temperature BoxPanther Fusion Reconstitution Buffer I
Room Temperature BoxPanther Fusion Elution Buffer
Room Temperature BoxPanther Fusion Oil
Refrigerated BoxPanther Fusion GI Expanded Bacterial Assay Controls• Panther Fusion GI Expanded Bacterial Positive Control• Panther Fusion Negative Control

Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Expanded Bacterial Assay

Aptima Assay Fluids Kit
Aptima Multitest Swab Specimen Collection Kit

Instrumentation

The Panther Fusion GI Expanded Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay.

Page 9

V. INDICATIONS FOR USE

The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.

This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.

Page 10

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A comparison of the Panther Fusion GI Expanded Bacterial Assay to the predicate BioFire FilmArray Gastrointestinal (GI) Panel (K160459, K143005 & K140407) is summarized in Table 3.

Table 3: Comparison of Similarities and Differences Between the Predicate Device (BioFire FilmArray Gastrointestinal (GI) Panel) and the Subject Device (Panther Fusion GI Bacterial Assay)

PredicateBioFire FilmArray Gastrointestinal (GI) PanelSubjectHologic Panther Fusion GI Expanded Bacterial Assay
ClassificationClass IIClass II
Regulation no./Product code21CFR 866.3990PCH, OOI21CFR 866.3990PCH, OOI
Intended useThe FilmArray Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the FilmArray GI Panel:Campylobacter (C. jejuni/C. coli/C. upsaliensis)Clostridium difficile (C. difficile) toxin A/BPlesiomonas shigelloidesSalmonellaThe Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical

Page 11

PredicateBioFire FilmArray Gastrointestinal (GI) PanelSubjectHologic Panther Fusion GI Expanded Bacterial Assay
Intended use (continued)Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae), including specific identification of Vibrio choleraeYersinia enterocoliticaEnteroaggregative Escherichia coli (EAEC)Enteropathogenic Escherichia coli (EPEC)Enterotoxigenic Escherichia coli (ETEC) lt/stShiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli O157 serogroup within STEC)Shigella/ Enteroinvasive Escherichia coli (EIEC)CryptosporidiumCyclospora cayetanensisEntamoeba histolyticaGiardia lamblia (also known as G. intestinalis and G. duodenalis)Adenovirus F 40/41AstrovirusNorovirus GI/GIIRotavirus ASapovirus (Genogroups I, II, IV, and V)The FilmArray GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out coinfection with organisms not included in the FilmArray GI Panel. The agent detected may not be the definite cause of the disease.Concomitant culture is necessary for organism recovery and further typing ofpresentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.

Page 12

PredicateBioFire FilmArray Gastrointestinal (GI) PanelSubjectHologic Panther Fusion GI Expanded Bacterial Assay
Intended use (continued)bacterial agents.This device is not intended to monitor or guide treatment for C. difficile infection.Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens.Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.Negative FilmArray GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.
TechnologyFully automated nucleic acid amplification, detection and result interpretation (multiplex PCR)Fully automated nucleic acid amplification, detection and result interpretation (multiplex PCR)
AnalyteRNA/DNADNA
Organism DetectedCampylobacter (C. jejuni/C. coli/C.upsaliensis), Clostridium difficile (C.difficile) toxin A/B, Plesiomonas shigelloides, Salmonella, Vibrio (V.Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides

Page 13

PredicateBioFire FilmArray Gastrointestinal (GI) PanelSubjectHologic Panther Fusion GI Expanded Bacterial Assay
Organism Detected (continued)parahaemolyticus/V. vulnificus/V. cholerae) including specific identification of Vibrio cholera, Yersinia enterocolitica, Enteroaggregative Escherichia coli (EAEC), Enteropathogenic Escherichia coli (EPEC), Enterotoxigenic Escherichia coli (ETEC) lt/st, Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli O157 serogroup within STEC), Shigella/Enteroinvasive Escherichia coli (EIEC),Cryptosporidium, Cyclospora cayetanensis, Entamoeba histolytica, Giardia lamblia (also known as G. intestinalis and G. duodenalis), Adenovirus F 40/41, Astrovirus, Norovirus GI/GII, Rotavirus A, and Sapovirus (Genogroups I,II, IV, and V).
Specimen TypesHuman stool in Cary Blair transport mediumHuman stool in Cary Blair transport medium
Sample PreparationAutomated sample processingAutomated sample processing
Intended UsersProfessional useProfessional use
Amplification modeNested multiplex PCRReal-time PCR
Detection modeFluorogenic double stranded DNA binding dyeFluorogenic target-specific hybridization
InstrumentationBioFire FilmArray SystemsPanther Fusion System
Time to ResultApproximately 1 hourApproximately 2.4 hours
ControlsTwo controls are included in each reagent pouch to control for sample processing and both stages of PCR and melt analysis.Internal control in each sample. External control processed at periodic interval

Page 14

VII. PERFORMANCE DATA

The following performance data (analytical and clinical) were provided in support of the substantial equivalence determination.

Brief Description of Analytical (Non-Clinical) Studies

The following analytical studies (non-clinical) were conducted to support the clearance of the Panther Fusion GI Expanded Bacterial Assay on the Panther Fusion System.

Analytical Sensitivity - Limit of Detection (LoD)

The analytical sensitivity (Limit of Detection or LoD) of the Panther Fusion GI Expanded Bacterial Assay was determined by testing dilutions of preserved negative Cary-Blair stool processed with STM (negative CBS matrix) spiked with bacterial cultures of Yersinia (2 strains), Vibrio (3 strains), Plesiomonas (2 strains), and STEC O157 (2 strains). A minimum of 24 replicates were tested with each of the 3 reagent lots. The LoD for each analyte was determined by Probit analysis for each reagent lot and was confirmed with an additional 24 replicates using a single reagent lot in single analyte and multi- analyte configuration. Analytical sensitivity is defined as the lowest concentration at which ≥95% of all replicates tested positive, as summarized in Table 4.

Table 4. Analytical Sensitivity

StrainLoD Concentration (CFU/mL)ᵃ
Aptima Multitest TubePreserved Stool
Yersinia enterocolitica, 33114911,820
Yersinia enterocolitica, 1375, O:8941,880
Vibrio parahaemolyticus, EB101901,800
Vibrio vulnificus, B962910200
Vibrio cholerae, 802133660
STEC O157:H7, EDL 931531,060
O157:NM, CDC 92-30733577,140
Plesiomonas shigelloides, CDC 3085-55651,300
Plesiomonas shigelloides, GNI 1434680

CFU = colony forming units.
ᵃ Analyte concentrations in Aptima Multitest tube are ~ 20X dilute compared to preserved stool (~150μL preserved stool in ~3 mL

Page 15

Inclusivity/Reactivity – Wet Testing

The inclusivity/reactivity of the Panther Fusion GI Expanded Bacterial Assay was determined by testing bacterial strains in negative CBS matrix. Each strain was tested in triplicate at 3X LoD with 1 reagent lot in single or multi-analyte configuration. For strains not detected at 3X LoD, additional testing at higher concentrations was performed until 100% positivity was observed. Table 5 shows the lowest concentration of each strain at which 100% positivity was observed.

Table 5. Inclusivity/Reactivity Summary for the GI Expanded Bacterial Assay Analytes

OrganismATCC# or SourceStrain/ serovar/ serotype/ antigenic propertiesTest Concentration (3X LoD) (CFU/mL)
Aptima Multitest TubePreserved Stool
Yersinia enterocoliticaBEI NR-207CDC 497-70, O:82825,640
BEI NR-212NCTC 11175, O:32825,640
23715Billups-1803-68, O:82825,640
493971375, O:8c2825,640
NCTC 10463P 77, O:5, 272825,640
CCUG 4588Type 2, O:92825,640
CCUG 8050N/A2825,640
CCUG 8232Type 5, O:1, 2, 3 | O:2, 3 | O:3/XI2825,640
CCUG 8234Type 42825,640
55075O:92825,640
27729WA, Type 1, O:82825,640
Vibrio parahaemolyticusBEI NR-2199048057, O4: K122705,400
BEI NR-21992KXV 755, O4: K412705,400
BAA-242VP250, O1:KUT2705,400
27969FC 10112705,400
BAA-241VP232, O4:K682705,400
33845117 [CDC KC830]2705,400
43996NCTC 10884 [70/116655]2705,400
33846205 [9302]2705,400
49529MDL 3875-7-83, O4:K122705,400
CCUG 34902N/A2705,400
CCUG 67711N/A2705,400
Vibrio vulnificus33847279 [11590]2705,400
33817329 [CDC B3547], Biotype 233660
BAA-86CDC 9505-9533660
CCUG 38297N/A33660

Page 16

OrganismATCC# or SourceStrain/ serovar/ serotype/ antigenic propertiesTest Concentration (3X LoD) (CFU/mL)
Aptima Multitest TubePreserved Stool
Vibrio vulnificus (continued)CCUG 47321N/A33660
29306CDCA1402 [P. Baumann 328]33660
43382VVL133660
29307CDCA869433660
CCUG 38297N/Aᵇ551,110
Vibrio choleraeBEI NR-147N16961, O:1991,980
BEI NR-148CVD 101, O:1991,980
BEI NR-149Nanking 32/123, O:2991,980
BEI NR-152Nanking 32/124 (NCTC 8042), O:7991,980
14033NCTC 8457 [R. Hugh 1092], O1, Inaba991,980
9459AMC 20-A-10 [R. Hugh 583], Inaba991,980
CCUG 2573NAG/NCV991,980
CCUG 2569NAG/NCV991,980
CCUG 4070Non O-1991,980
CCUG 2158918991,980
CCUG 56875N/A991,980
CCUG 53725O1/O139991,980
CCUG14542NA991,980
9458AMC 20-A-41 [R. Hugh 582], Ogawa991,980
25870569B991,980
STEC O157: H743890CDC C984 [CDC 3526-87], H71593,180
43895CDC EDL 933, H71593,180
43894CDC EDL 932, H71593,180
700927EDL 933, H7:K-1593,180
STEC O157: NM700375CDC 94-G7771, NM1,19723,940
700377CDC 92-3099, NM1,19723,940
700378CDC 92-3073, NM1,19723,940
AR Bank # 427ᵃN/A1,19723,940
AR Bank # 428ᵃN/A1,19723,940
AR Bank # 429ᵃN/A1,19723,940
AR Bank # 430ᵃN/A1,19723,940
Plesiomonas shigelloides14030CDC 16408 [Ferguson and Henderson C27, RH 864], O:171953,900

Page 17

OrganismATCC# or SourceStrain/ serovar/ serotype/ antigenic propertiesTest Concentration (3X LoD) (CFU/mL)
Aptima Multitest TubePreserved Stool
Plesiomonas shigelloides (continued)51903GNI 14ᶜ1953,900
51572CIP 69.35 [2886]1953,900
CCUG 7041AO17: H21953,900
CCUG 9221O171953,900
CCUG 14309O17: H21953,900
CCUG 14597N/A1953,900

CFU = colony forming units.
ᵃ These strains were evaluated using the higher LoD of the 2 serotypes which is the NM serotype.
ᵇ For this strain 100% positivity was observed at ~ 5X LoD. In silico analysis showed 100% homology to amplification region.
ᶜ Strains used to establish LoD.

Inclusivity/Reactivity - In silico Analysis

The inclusivity of the Panther Fusion GI Expanded Bacterial Assay was evaluated using in silico inclusivity analysis for each analyte. In silico analysis was performed using analyte sequences available in the NCBI database and the whole genome shotgun sequence database. For each analyte, corresponding oligonucleotide sequences (primers and probes) were evaluated against the database sequences. Any sequences with insufficient lengths (not covering the entire amplicon region) were excluded from the analysis.

Based on in silico analysis of all sequences available up to May 30, 2023 in the databases, the Panther Fusion GI Expanded Bacterial Assay is predicted to detect 99.9% of 1,054 Yersinia Enterocolitica, 99.5% of 1,337 Vibrio parahaemolyticus, 99.1% of 1,180 Vibrio vulnificus, 98.0% of 1,189 Vibrio cholerae, 100% of 2,004 STEC O157, and 91.5% of 47 Plesiomonas shigelloides sequences evaluated.

Analytical Specificity: Cross Reactivity and Microbial Interference - Wet Testing

Analytical specificity (cross-reactivity) and microbial interference for the Panther Fusion GI Expanded Bacterial Assay were evaluated in the presence of non-targeted microorganisms that are either phylogenetically related to the assay analytes or potentially found in clinical specimens. Panels consisting of 109 bacteria, viruses, parasites, and yeast listed in Table 6 were tested in negative CBS matrix in the absence and in the presence of GI Expanded Bacterial Assay analytes at 3X LoD. Except where noted, bacteria, yeast, and parasites were

Page 18

evaluated at 10⁶ CFU/mL or 10⁶ rRNA copies/mL or 10⁶ cells/mL; viruses were evaluated at 10⁵ TCID₅₀/mL. If cross-reactivity or interference was observed in the initial testing, then the organism was tested at lower concentrations until the expected result was observed. No cross-reactivity or microbial interference was observed with any of the organisms tested on the Panther Fusion GI Expanded Bacterial Assay at the indicated concentrations.

Table 6. Microorganisms Tested for Cross-Reactivity and Microbial Interference

MicroorganismTest ConcentrationMicroorganismTest Concentration
Arcobacter cryaerophilus10⁶ CFU/mLEntercoccus faecalis10⁶ CFU/mL
Neisseria gonorrhoeae10⁶ CFU/mLEnterobacter aerogenes10⁶ CFU/mL
Streptococcus pyogenes10⁶ CFU/mLEnterobacter cloacae10⁶ CFU/mL
Trabulsiella guamensis10⁶ CFU/mLEscherichia fergusonii10⁶ CFU/mL
Faecalibacterium prausnitzii10⁶ rRNA copies /mLEscherichia hermanii10⁶ CFU/mL
Escherichia coli (non-shigatoxigenic)10⁶ CFU/mLEscherichia vulneris10⁶ CFU/mL
Giardia lamblia BG-Aᵃ10⁶ copies/mLGardnerella vaginalis10⁶ CFU/mL
Cyclosporaᵃ10⁶ copies/mLHelicobacter pylori10⁶ CFU/mL
Cryptosporidiumᵃ10⁶ copies/mLKlebsiella oxytoca10⁶ CFU/mL
Norovirus (Noro GII)ᵃ10⁶ copies/mLKlebsiella ozaenae10⁶ CFU/mL
Astrovirusa10⁶ copies/mLKlebsiella pneumoniae10⁶ CFU/mL
Sapovirus (GII)ᵃ10⁵ copies/mLLactobacillus acidophilus10⁶ CFU/mL
Enterovirus (Ent V)ᵃ10⁵ copies/mLLactobacillus crispatus10⁶ CFU/mL
Rhinovirusᵃ10⁵ copies/mLLactococcus lactis10⁶ CFU/mL
Coronavirus 229E10⁵ TCID₅₀/mLListeria grayi10⁶ CFU/mL
Coxsakeivirus Type B410⁵ TCID₅₀/mLListeria monocytogenes10⁶ CFU/mL
Adenovirus Type 7A10⁵ TCID₅₀/mLMorganella morganii10⁶ CFU/mL
Rotavirusᵃ10⁵ copies/mLPeptostreptococcus anaerobius10⁶ CFU/mL
Anaerococcus tetradius10⁶ CFU/mLPeptostreptococcus micros10⁶ rRNA copies /mL
Abiotrophia defectivia10⁶ CFU/mLPhotobacterium damselae10⁶ CFU/mL
Acinetobacter baumannii10⁶ CFU/mLPrevotella bivia10⁶ CFU/mL

Page 19

MicroorganismTest ConcentrationMicroorganismTest Concentration
Acinetobacter lwoffii10⁶ CFU/mLPrevotella melaninogenica10⁶ CFU/mL
Aeromonas hydrophila10⁶ CFU/mLProteus mirabilis10⁶ rRNA copies /mL
Alcaligenes faecalis10⁶ CFU/mLProteus penneri10⁶ CFU/mL
Campylobacter upsaliensis10⁶ CFU/mLProteus vulgaris10⁶ CFU/mL
Anaerococcus vaginalis10⁶ CFU/mLProvidencia alcalifaciens10⁶ CFU/mL
Arcobacter butzleri10⁶ CFU/mLProvidencia rettgeri10⁶ CFU/mL
Bacillus cereus10⁶ CFU/mLProvidencia stuartii10⁶ CFU/mL
Bacteriodes fragilis10⁶ CFU/mLPseudomonas aeruginosa10⁶ CFU/mL
Bacteroides thetaiotaomicron10⁶ CFU/mLPseudomonas fluorescens10⁶ CFU/mL
Bacteroides vulgatus10⁶ CFU/mLSerratia liquefaciens10⁶ CFU/mL
Bifidobacterium adolescentis10⁶ CFU/mLSerratia marcescens10⁶ CFU/mL
Bifidobacterium longum10⁶ rRNA copies /mLStaphylococcus aureus10⁶ CFU/mL
Campylobacter fetus10⁶ CFU/mLStaphylococcus epidermidis10⁶ CFU/mL
Campylobacter hyointestinalis10⁶ CFU/mLStenotrophomonas maltophilia10⁶ CFU/mL
Campylobacter rectus10⁶ CFU/mLStreptococcus anginosus10⁶ CFU/mL
Campylobacter sputorum10⁶ CFU/mLStreptococcus dysgalactiae10⁶ CFU/mL
Candida albicans10⁶ CFU/mLYersinia bercovieri10⁶ CFU/mL
Citrobacter freundii10⁶ CFU/mLYersinia pseudotuberculosis10⁶ CFU/mL
Citrobacter koseri10⁶ CFU/mLYersinia rohdei10⁶ CFU/mL
Clostridium difficile10⁶ CFU/mLCampylobacter lari10⁶ CFU/mL
Clostridium perfringens10⁶ CFU/mLEntamoeba histolytica10⁴ cells/mL
Clostridium ramosum10⁶ CFU/mLMegasphaeara elsdenii10⁶ CFU/mL
Clostridium sordellii10⁶ CFU/mLChlamydia trachomatis10⁵ IFU/mL
Clostridium tertium10⁶ CFU/mLLeptotrichia buccalis10⁶ CFU/mL
Collinsella aerofaciens10⁶ CFU/mLCytomegalovirus10⁵ TCID₅₀/mL
Corynebacterium genitalium10⁶ CFU/mLSalmonella enterica10⁶ CFU/mL
Cronobacter sakazakii10⁶ CFU/mLCampylobacter jejuni10⁶ CFU/mL
Edwardsiella tarda10⁶ CFU/mLShigella sonnei10⁶ CFU/mL
Egglerthella lenta10⁶ rRNA copies /mLSTEC - stx110⁶ CFU/mL
STEC - stx210⁶ CFU/mLVibrio mimicus10⁶ CFU/mL
Vibrio fluvialis10⁶ CFU/mLYersinia frederiksenii10⁶ CFU/mL
Vibrio furnissii10⁶ CFU/mLYersinia kristensenii10⁶ CFU/mL
Vibrio metschnikovii10⁶ CFU/mLVibrio alginolyticusᵇ10⁴ CFU/mL

CFU = colony forming units, IFU = inclusion forming units, rRNA copies = ribosomal ribonucleic acid copies, TCID₅₀ = Median Tissue Culture Infectious Dose.
ᵃ In vitro transcripts were used to evaluate cross-reactivity and microbial interference as cultured virus or whole genome purified nucleic acid are not readily available.
ᵇ Cross reactivity was observed at concentrations ≥10⁵ CFU/mL.

Page 20

Coinfection/Competitive Interference

Competitive interference in the Panther Fusion GI Expanded Bacterial Assay was evaluated in triplicate using pairs of assay analytes at low/high concentrations in negative CBS matrix. The low concentration analyte was tested at 3X LoD against a high concentration analyte at 10⁶ CFU/mL. Additionally, analytes were also tested in the absence of a second analyte. If less than 100% positivity was observed for the low concentration analyte, the high concentration analyte was diluted until a concentration was reached where 100% positivity was achieved for the low concentration analyte. The highest concentration of competing analyte at which the low concentration analyte maintained a 100% positivity is shown in Table 7. When the analytes were tested at high concentration, all results for other analytes maintained expected positivity; no competitive interference was observed.

Table 7. Summary of Coinfection Results

Analyte 1Analyte 2Yersinia % PosVibrio % PosSTEC O157 % PosPlesiomonas % Pos
Name3X LoD (CFU/mL)ᵃNameHigh Conc (CFU/mL)ᵇ
NegativeNANegativeNA0%0%0%0%
None0100%0%0%0%
Vibrioᵇ10⁴100%100%0%0%
Yersinia282STEC O15710⁶100%0%100%0%
Plesiomonas10⁶100%0%0%100%
None00%100%0%0%
Yersinia10⁶100%100%0%0%
Vibrio282STEC O15710⁶0%100%100%0%
Plesiomonas10⁶0%100%0%100%

CFU = colony forming units, Pos = positive.
ᵃ Analyte concentration in Aptima Multitest tube.
ᵇ Less than 100% positive results were observed for analyte 1 with Vibrio at ≥10⁵ CFU/mL.

Page 21

Interference

Potential inhibitory effects of endogenous and exogenous substances that may be present in a specimen were evaluated in the Panther Fusion GI Expanded Bacterial Assay. Clinically relevant concentrations of potentially interfering substances were added to negative CBS matrix and tested in the absence and in the presence of GI Expanded Bacterial Assay analytes at 3X LoD. Tests were performed in triplicate. The substances and test concentrations are shown in Table 8. No impact on the performance of the Panther Fusion GI Expanded Bacterial Assay was observed for any of the substances at the concentrations tested.

Page 22

Table 8. Substances Tested for Interference

Substance TypeGeneric NameActive Ingredient(s)Test Concentrationᵃ' ᵇ' ᶜ
AntibioticsAmoxicillinAmoxicillin0.7 μg/mL
AmpicillinAmpicillin0.9 μg/mL
DoxycyclineDoxycycline0.2 μg/mL
MetronidazoleMetronidazole1.5 μg/mL
Neosporin®Polymyxin B sulfate, bacitracin zinc, neomycin sulfate1.3% w/v
Anti-microbial and anti-fungalBZK Antiseptic TowelettesBenzalkonium chloride1.3% v/v
NystatinNystatin1.3% v/v
Dulcolax® suppositoryBisacodyl75 ng/mL
Colace®Docusate sodium3.0 μg/mL
Fleet® mineral oil enemaMineral oil1.3% v/v
Laxatives and stool softenersEx-Lax®Sennosides0.8 μg/mL
Miralax®Polyethylene glycol 33500.1 mg/mL
Milk of MagnesiaMagnesium hydroxide, Aluminum hydroxide1.3% v/v
Visicol®Sodium phosphate53 ng/mL
Anti-diarrhealImodiumLoperamide hydrochloride0.1 μg/mL
Anti-ItchVagisil®Benzocaine1.3% w/v
Preparation H®Hydrocortisone1.3% w/v
Phenylephrine hydrochloride (for hemorrhoids)Phenylephrine hydrochloride0.4 ng/mL
Anti-inflammatoryMesalazine (Rx only, for Crohns disease/ ulcerative colitis)Salicylic acid0.4 μg/mL
Aleve®Naproxen sodium4.5 μg/mL
AntacidPepto-Bismol®Bismuth subsalicylate1.3% v/v
Tums®Calcium carbonate55 μg/mL
Radiopaque contrast materialBarium sulfateBarium sulfate0.1 mg/mL
Lubricants and skin protectantsK-Y® Personal Lubricant Jelly GlycerinGlycerin1.3% w/v
Vaseline® Original 100% Pure Petroleum Jelly WhitePetrolatum1.3% w/v
Desitin®Zinc oxide1.3% w/v

Page 23

Substance TypeGeneric NameActive Ingredient(s)Test Concentrationᵃ' ᵇ' ᶜ
SpermicideOptions Conceptrol®Vaginal Contraceptive GelNonoxynol-91.3% w/v
EndogenousCholesterolCholesterol50 μg/mL
Fatty acidsPalmitic acid16 μg/mL
Fatty acidsStearic acid34 μg/mL
Triglycerides, total (Fecal fat, Intralipid)Triglycerides1.3% v/v
Human BileBilirubin, conjugated5.0 μg/mL
UrineHuman Urine1.3% v/v
Human Whole BloodBlood hemoglobin1.3% v/v
MucinPurified mucin protein0.05% w/v

ᵃ Analyte concentration in Aptima Multitest tube.
ᵇ v/v: volume by volume.
ᶜ w/v: weight by volume.

Stool specimens prepared in various preservative media were evaluated for potential impact on the Panther Fusion GI Expanded Bacterial Assay performance. The preservative media evaluated include 7 different types of Cary-Blair transport media from different vendors and preservative media containing fixatives shown in Table 9. All media were tested with GI Expanded Bacterial Assay analytes at 3X LoD. Comparable performance was seen with all Cary-Blair media. Interference was observed when specimens were processed in media containing fixative.

Page 24

Table 9. Stool Preservative Media Tested for Interference

Cary-Blair Media

  • Culture & Sensitivity (C&S) Medium
  • Cary Blair Transport Medium w/ Indicator
  • Para-Pak® C&S
  • Para-Pak® Enteric Plus
  • Cardinal Health™ C&S Stool Transport Vial
  • Protocol Cary Blair Medium
  • Enteric Transport Media (ETM)

Fixative Media (interference was observed)

  • Fisher® 10% Buffered Formalin
  • Para-Pak® 10% Buffered Formalin
  • Para-Pak® LV-PVA

Carryover Contamination

Panther Fusion GI Bacterial Assay and GI Expanded Bacterial Assay belong to the same family of assays that both utilize Cary Blair Stool as the sample type and follow identical assay processing steps. Carryover contamination was evaluated using Panther Fusion GI Bacterial Assay as a representative assay and demonstrated a 0% carryover rate.

Within Laboratory Precision/Repeatability

Panther Fusion GI Expanded Bacterial Assay within laboratory precision was evaluated with a 5-member panel consisting of assay analytes in negative CBS matrix. The 5-member panel included 1 negative, 2 single analyte (Yersinia), and 2 multi-analyte (with Vibrio, STEC O157, and Plesiomonas) panel members. The panels were tested by 3 operators on 2 runs per day, using 3 reagent lots on 3 Panther Fusion Systems over 9 days.

The panel members are described in Table 10, along with a summary of the agreement with the expected results, mean Ct, variability analysis between reagent lots, operators, instruments, days, between and within runs, and overall (total).

Page 25

Table 10. Ct Variability Analysis Summary

PanelDescriptionAnalyteAgreement %ᵃMean CtBetween LotsBetween InstrumentsBetween OperatorsBetween DaysBetween RunsWithin RunTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
1Negative (Internal Control)162/16210034.60.070.20.080.230.040.120.000.000.000.000.481.390.51.43
2Low Pos (1.5X LoD)Yersinia162/16210033.70.030.080.090.260.000.000.000.000.000.000.411.230.42
3Mod Pos (1.5X LoD)Yersinia162/16210033.70.120.350.070.210.010.040.000.000.170.520.230.690.32
Vibrio162/16210032.70.070.210.120.370.000.000.000.000.190.570.20.60.3
4Low Pos (1.5X LoD)STEC O157162/16210032.40.020.080.040.130.000.000.000.000.110.340.280.870.31
Plesiomonas162/16210031.30.020.080.060.20.000.000.030.10.000.000.210.680.22
Vibrio162/16210033.80.080.250.050.140.000.000.000.00<0.010.030.250.730.26
3Mod Pos (3X LoD)STEC O157162/16210033.10.050.17<0.010.030.010.030.060.170.000.000.190.560.2
Plesiomonas162/162100280.110.390.321.150.000.000.000.000.120.420.140.510.39

Ct = cycle threshold, CV = coefficient of variation, Mod = Moderate, N = sample size, Pos = positive, SD = standard deviation.
ᵃ Agreement to expected panel positivity result.

Reproducibility

Panther Fusion GI Expanded Bacterial Assay reproducibility was evaluated at 3 US sites using 1 negative panel member and 4 panel members positive for 1 or 3 targets. Testing was performed for 5 days by 6 operators (2 at each site) using 1 lot of assay reagents. Each run included 3 replicates of each panel member.

A negative panel member was created using a matrix comprised of stool specimens negative for all assay targets preserved in Cary-Blair media processed into STM. Positive panel members were created by spiking 1.5X LoD (low positive) or 3X LoD (moderate positive) concentrations of the target analytes into the negative matrix.

The agreement with expected results was 100% for all panel members components for Yersinia, Vibrio, STEC O157, and Plesiomonas (Table 11).

Page 26

Table 11. Agreement of Panther Fusion GI Expanded Bacterial Assay Results with Expected Results

DescriptionAnalyteN% (95% CI)
NegInternal Control89/89100 (95.9-100)
Low PosᵃYersiniaᶜ90/90100 (95.9-100)
Vibrioᶜ90/90100 (95.9-100)
STEC O15790/90100 (95.9-100)
Plesiomonasᶜ90/90100 (95.9-100)
Mod PosᵇYersiniaᶜ'ᵈ90/90100 (95.9-100)
Vibrioᶜ90/90100 (95.9-100)
STEC O15790/90100 (95.9-100)
Plesiomonasᶜ90/90100 (95.9-100)

CI = score confidence interval, Mod = moderate, N = sample size, Neg = negative, Pos = positive.
ᵃ Low Pos = All targets are 1.5X LoD.
ᵇ Mod Pos = All targets are 3X LoD.
ᶜ Yersinia enterocolitica, Vibrio parahaemolyticus, STEC O157, and Plesiomonas shigelloides strains were used to build the positive panels.
ᵈ One (1) false positive Vibrio result was obtained for a moderate positive Yersinia panel member.

Signal variability was measured as %CV of the Ct values. The total signal variability was ≤1.61% (SD ≤0.55) for all panel components (Table 12). For the sources of variation except the 'within run' factor, %CV values were ≤1.03% for all panel components. The signal variability was ≤1.01% (SD ≤0.33) for the Panther Fusion GI Expanded Bacterial Assay positive controls (Table 13).

Page 27

Table 12. Signal Variability of the Panther Fusion GI Expanded Bacterial Assay by Target and Concentration

DescriptionAnalyteNMean CtBetween SiteBetween Operator/RunᶜBetween DayWithin RunTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
Low PosᵃYersinia9034.70.170.500.210.610.090.270.441.250.521.51
Vibrio9033.70.160.490.080.250.000.000.260.770.320.95
STEC O1579032.40.170.530.130.410.000.000.300.920.371.14
Plesiomonas9033.90.160.470.060.170.000.000.320.940.361.06
Mod PosᵇYersinia9033.80.351.030.190.580.070.210.371.080.551.61
Vibrio9032.70.200.600.090.260.110.350.220.680.331.01
STEC O1579031.40.240.750.080.270.070.210.260.810.361.16
Plesiomonas9033.20.220.670.120.370.000.000.260.780.361.09

Ct = cycle threshold, CV = coefficient of variation, Mod = moderate, N = sample size, Pos = positive, SD = standard deviation.
Note: The analysis was performed using the SAS MIXED procedure, which applies a lower boundary of 0 to all variance components in the model by default. If a variance component is 0, SD, and %CV are displayed as 0.00
ᵃ Low Pos = All targets are 1.5X LoD.
ᵇ Mod Pos = All targets are 3X LoD.
ᶜ Between Operator may be confounded with Between Run; therefore, Between Operator and Between Run estimates are combined in Between Operator/Run.

Table 13. Signal Variability of the Panther Fusion GI Expanded Bacterial Assay Positive Controls

ControlAnalyteNMean CtBetween SiteBetween OperatorBetween DayWithin DayTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
PosYersinia3032.70.220.660.000.000.000.000.250.750.331.01
Vibrio3033.40.000.000.000.000.000.000.290.860.290.86
STEC O1573031.50.110.350.000.000.070.230.260.830.290.93
Plesiomonas3032.00.050.160.080.230.120.370.240.740.290.87

Ct = cycle threshold, CV = coefficient of variation, N = sample size, Pos = positive, SD = standard deviation.
Note: The analysis was performed using the SAS MIXED procedure, which applies a lower boundary of 0 to all variance components in the model by default. If a variance component is 0, SD, and %CV are displayed as 0.00.

Page 28

Brief Description of Clinical Studies

A multicenter study was conducted using remnant stool specimens in Cary-Blair preservative medium collected as part of routine patient care at 10 US clinics from pediatric or adult patients suspected of acute gastroenteritis. All specimens were tested with the Panther Fusion GI Expanded Bacterial Assay and with comparator assays: a PCR plus bidirectional sequencing (run in duplicate) for STEC O157 and an FDA-cleared Nucleic Acid Amplification Test (NAAT) for all other targets. An alternate FDA-cleared NAAT was used for discordant resolution testing, if applicable. Positive (PPA) and negative (NPA) percent agreement, with corresponding 2-sided 95% Score CIs, were calculated relative to comparator results, by target and by specimen category.

A total of 1,548 prospective specimens and 251 retrospective specimens were enrolled in the study; 94 specimens were excluded from the performance analyses (for example, duplicate individuals, invalid Panther Fusion or comparator results for all targets). An additional 189 contrived specimens were assessed to supplement the prospective and retrospective data for all targets. Of the 1,919 specimens tested in valid Panther Fusion GI Expanded Bacterial Assay runs, 36 (1.9%) had initial invalid results. Upon retest, 25 of the 36 specimens yielded valid results, for a total of 11 (0.6%) specimens with final invalid results. The final data set consisted of 1,894 evaluable specimens (1,523 prospective specimens, 182 retrospective specimens, and 189 contrived specimens); not all were evaluable for all analytes. Demographic information for the 1,705 evaluable prospective and retrospective specimens is provided in Table 14.

Page 29

Table 14. Summary of Subject Demographics

Total N (%)Prospective N (%)Retrospective N (%)
Total Specimens1,7051,523182
Sex
Female898 (52.1)793 (52.1)95 (52.2)
Male817 (47.9)730 (47.9)87 (47.8)
Age Group
0 to 28 days7 (0.4)7 (0.5)0 (0)
29 days to <2 years74 (4.3)67 (4.4)7 (3.8)
2 to 5 years55 (3.2)50 (3.3)5 (2.7)
6 to 11 years68 (4.0)66 (4.3)2 (1.1)
12 to 17 years73 (4.3)71 (4.7)2 (1.1)
18 to 21 years47 (2.8)44 (2.9)3 (1.6)
22 to 64 years825 (48.4)724 (47.5)101 (55.5)
≥65 years556 (32.6)494 (32.4)62 (34.1)

N = population size

Performance characteristics for detection of Yersinia, Vibrio, STEC O157, and Plesiomonas are shown in Table 15 through Table 18.

Table 15. Clinical Performance - Yersinia spp.

Specimen OriginNTPFPTNFNPrevalenceᵃ (%)PPA % (95% CI)ᵇNPA % (95% CI)ᵇ
Prospective (Fresh)1,507105ᶜ1,4871ᵈ0.790.9 (62.3, 98.4)99.4 (98.9, 99.7)
Retrospective (Frozen)182153ᵉ1640N/Aᶠ100 (79.6, 100)98.2 (94.9, 99.4)
Contrived (Frozen)1896301260N/Aᶠ100 (94.3, 100)100 (97.0, 100)

CI = confidence interval, FN = false negative, FP = false positive, N = sample size, NPA = negative percent agreement, PPA = positive percent agreement, TN = true negative, TP = true positive.
ᵃ Study prevalence reported based on comparator testing.
ᵇ Score CI.
ᶜ 5 of 5 discordant false positive prospective specimens were positive for Yersinia by the alternate NAAT.
ᵈ The discordant false negative prospective specimen was negative for Yersinia by the alternate NAAT.
ᵉ The 3 discordant false positive retrospective specimens were positive for Yersinia by the alternate NAAT.
ᶠ Calculation of prevalence is not applicable.

Page 30

Table 16. Clinical Performance - Vibrio spp.

Specimen OriginNTPFPTNFNPrevalenceᵃ (%)PPA % (95% CI)ᵇNPA % (95% CI)ᵇ
Prospective (Fresh)1,507101,5051ᶜ0.150.0 (9.5, 90.5)100 (99.7, 100)
Retrospective (Frozen)18296ᵈ1670N/Aᶠ100 (70.1, 100)96.5 (92.6, 98.4)
Contrived (Frozen)189631ᵉ1250N/Aᶠ100 (94.3, 100)99.2 (95.6, 99.9)

CI = confidence interval, FN = false negative, FP = false positive, N = sample size, NPA = negative percent agreement, PPA = positive percent agreement, TN = true negative, TP = true positive.
ᵃ Study prevalence reported based on comparator testing.
ᵇ Score CI.
ᶜ The discordant false negative prospective specimen was positive for Vibrio by the alternate NAAT.
ᵈ All 6 discordant false positive retrospective specimens were positive for Vibrio by the alternate NAAT.
ᵉ The discordant false positive contrived specimen was negative for Vibrio by the alternate NAAT.
ᶠ Calculation of prevalence is not applicable.

Table 17. Clinical Performance – STEC O157

Specimen OriginNTPFPTNFNPrevalenceᵃ (%)PPA % (95% CI)ᵇNPA % (95% CI)ᵇ
Prospective (Fresh)1,52212ᶜ1,51900.1100 (20.7, 100)99.9 (99.5, 100)
Retrospective (Frozen)18231ᵈ1780N/Aᵍ100 (43.9, 100)99.4 (96.9, 99.9)
Contrived (Frozen)189621ᵉ1251ᶠN/Aᵍ98.4 (91.5, 99.7)99.2 (95.6, 99.9)

CI = confidence interval, FN = false negative, FP = false positive, N = sample size, NPA = negative percent agreement, PPA = positive percent agreement, TN = true negative, TP = true positive.
ᵃ Study prevalence reported based on comparator testing.
ᵇ Score CI.
ᶜ 1 of 2 discordant false positive prospective specimens was negative for STEC O157 by the alternate NAAT. The other discordant was positive for O157 but negative for stx1/stx2 by the alternate NAAT.
ᵈ The discordant false positive retrospective specimen was positive for STEC O157 by the alternate NAAT.
ᵉ The discordant false positive contrived specimen was negative for STEC O157 by the alternate NAAT.
ᶠ The discordant false negative contrived specimen was not retested by the alternate NAAT.
ᵍ Calculation of prevalence is not applicable.

Page 31

Table 18. Clinical Performance - Plesiomonas

Specimen OriginNTPFPTNFNPrevalenceᵃ (%)PPA % (95% CI)ᵇNPA % (95% CI)ᵇ
Prospective (Fresh)1,50711ᶜ1,50500.1100 (20.7, 100)99.9 (99.6, 100)
Retrospective (Frozen)18281ᵈ1730N/Aᶠ100 (67.6, 100)99.4 (96.8, 99.9)
Contrived (Frozen)1896201261ᵉN/Aᶠ98.4 (91.5, 99.7)100 (97.0, 100)

CI = confidence interval, FN = false negative, FP = false positive, N = sample size, NPA = negative percent agreement, PPA = positive percent agreement, TN = true negative, TP = true positive.
ᵃ Study prevalence reported based on comparator testing.
ᵇ Score CI.
ᶜ The discordant false positive prospective specimen was positive for Plesiomonas by the alternate NAAT.
ᵈ The discordant false positive retrospective specimen was positive for Plesiomonas by the alternate NAAT.
ᵉ The discordant false negative contrived specimen was not retested by the alternate NAAT.
ᶠ Calculation of prevalence is not applicable.

No coinfections were detected by the Panther Fusion GI Expanded Bacterial Assay or by the comparator methods in prospective and retrospective specimens.

VIII. CONCLUSIONS

The analytical and clinical study results demonstrate that the Panther Fusion GI Expanded Bacterial Assay on the Panther Fusion system performs comparably to the predicate device that is currently marketed for the same intended use. Hardware and software verification and validation demonstrate that the Panther Fusion GI Expanded Bacterial Assay on the Panther Fusion system will perform as intended.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).