510(k) Data Aggregation

    K Number
    K231566

    Validate with FDA (Live)

    Date Cleared
    2024-09-05

    (463 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ICX-Implant System is indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. ICX- Implant System is indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7.0 mm are indicated for delayed loading only.

    ICX-Implant System CAD CAM abutments are intended for use with dental implants as a support for single unit or multiple unit prostheses in the maxilla of a partially or fully edentulous patient. All digitally designed abutments for use with ICX-Implant System CAD CAM abutments are intended to be manufactured at a medentis medical GmbH validated milling center.

    Device Description

    The purpose of this submission is to obtain marketing clearance for an endosseous dental implant and abutment system. ICX-Implant System, from medentis medical GmbH. The ICX-Implant System includes a range of endosseous dental implants and prosthetic components. All implants have a selftapping apical thread with a tapered body and root-form designs with an internal hex implant/abutment connection. The implant body surface is blasted and acid-etched.

    Abutments are available in multiple designs, including straight and angled abutments intended for single tooth and multi-unit restorations.

    AI/ML Overview

    The provided text is a 510(k) summary for the "ICX-Implant System." This document is a premarket notification for a medical device and, as such, focuses on demonstrating substantial equivalence to previously cleared devices rather than proving the device meets specific acceptance criteria through a standalone performance study.

    Therefore, the document does not contain the detailed information required to fully answer all aspects of your request, especially regarding specific acceptance criteria for AI/algorithm performance, multi-reader multi-case studies, or detailed ground truth establishment for a test/training set in the context of an AI device.

    The document does describe non-clinical performance data for the dental implant system itself (e.g., sterilization, biocompatibility, mechanical testing) to show its safety and effectiveness, but not for an AI component.

    Here's an analysis based on the provided text, highlighting what is present and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    • Missing: The document does not define "acceptance criteria" in the context of an AI/algorithm's performance (e.g., sensitivity, specificity, AUC). Instead, it discusses the substantial equivalence of the ICX-Implant System and its components to predicate devices based on technological characteristics, materials, and indications for use.

    • The "Performance Data" section (Page 7) lists non-clinical tests performed, such as:

      • Validation of gamma irradiation sterilization to a sterility assurance level (SAL) of 10⁻⁶.
      • Bacterial endotoxin testing: < 20 EU/device.
      • Validation of moist heat sterilization cycle to SAL of 10⁻⁶.
      • Shelf life testing confirming seal integrity and sterility after 5 years.
      • Biocompatibility testing (in vitro cytotoxicity).
      • Characterization of implant surface (SEM, surface chemistry, hydrophilicity).
      • Mechanical testing according to ISO 14801 (static and dynamic loading).
      • Analysis of surface area and initial bone to implant contact area.
      • MR compatibility testing (supporting MR Conditional labeling).

      These are performance requirements for a dental implant and are confirmed to have been met, but they are not "acceptance criteria" for an AI algorithm's diagnostic performance.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Missing: No information on test set size or data provenance for an AI/algorithm is provided. The document explicitly states: "No clinical data were included in this submission" (Page 8). The non-clinical tests listed above would have had specific sample sizes for those tests (e.g., number of implants for mechanical testing), but these are not detailed in the summary beyond stating the tests were performed "accordance with" relevant standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Missing: Not applicable as no AI/algorithm performance test set is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Missing: Not applicable as no AI/algorithm performance test set is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Missing: No MRMC study was done, as no AI component is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Missing: No standalone algorithm performance study was done, as no AI component is mentioned.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Missing: Not applicable as no AI/algorithm performance is described. The "ground truth" for the implant itself is its conformance to material standards, mechanical properties, and biocompatibility, as verified by laboratory tests.

    8. The sample size for the training set:

    • Missing: No training set for an AI/algorithm is mentioned.

    9. How the ground truth for the training set was established:

    • Missing: No training set for an AI/algorithm is mentioned.

    In summary:

    This 510(k) pertains to a dental implant system (a physical device) and its components (e.g., abutments), not a software as a medical device (SaMD) or an AI-powered diagnostic/assistance tool. The entire submission is built on demonstrating substantial equivalence to existing, legally marketed predicate devices through a comparison of their technological characteristics, materials, and non-clinical performance data relevant to the physical implant and its function. It explicitly states that "No clinical data were included in this submission," further confirming that no human-in-the-loop or standalone diagnostic performance studies (which would be typical for AI/ML devices) were conducted or are required for this type of submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231434

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-14

    (89 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K142082, K013227, K072589, K220978, K063286, K111216, K212538, K170588, K22269, K213063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments compatible with a variety of original equipment manufacturers (OEM) of dental implants. This submission adds abutments, designated DESS Multi-Unit Abutments, for implant lines from Zimmer Dental and Biomet 31 (now collectively, ZimVie Dental). The subject device abutments include Multi-Unit Abutments in straight, 17° angled, and 30° angled designs, which are compatible with implants having internal hex connections. This submission is also to change how previously cleared devices are provided; the change is from previously provided nonsterile to now provided sterile. All abutments are provided with the appropriate abutment screw (if applicable) for attachment to the corresponding implant. The subject device is only intended for multi-unit restorations such as bridges and bars.

    AI/ML Overview

    It appears there might be a misunderstanding of the provided text. The document is a 510(k) premarket notification for a dental implant abutment, not an AI/ML medical device. Therefore, the information typically required for describing an AI/ML device's acceptance criteria and study data (like sample sizes for test and training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone performance) is not present.

    The document discusses the substantial equivalence of the DESS Dental Smart Solutions abutment to legally marketed predicate devices. The "performance data" section refers to non-clinical testing to demonstrate the device's physical properties and compatibility, not a study evaluating an AI algorithm's diagnostic or predictive performance.

    Let's break down what can be extracted from the provided text relevant to the device's acceptance, and then explain why the AI/ML specific questions cannot be answered.

    Understanding the Device and its "Acceptance":

    The "acceptance" in this context refers to the FDA's determination that the new DESS Dental Smart Solutions abutments are "substantially equivalent" to predicate devices already on the market. This determination is based on the device having the same intended use and similar technological characteristics (design, materials, performance via non-clinical testing).

    Here's a summary of the information available in the document:

    1. Table of Acceptance Criteria (or rather, Equivalence Criteria) and Reported Device Performance:

    The document doesn't present acceptance criteria in a quantitative table with performance metrics like sensitivity, specificity, or accuracy, as would be common for an AI/ML device. Instead, "acceptance" is based on demonstrating substantial equivalence to predicate devices across several characteristics. The performance is shown through non-clinical testing to ensure safety and functionality.

    CharacteristicSubject Device (DESS Dental Smart Solutions, K231434)Primary Predicate (K230143, DESS Dental Smart Solutions)Reference Device (K213063, TLX SRAs and TLX Gold Abutments)Reported Performance (Non-Clinical Data)
    Intended UseDESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.DESS Multi-Unit Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.Functional and esthetic rehabilitation of the edentulous mandible or maxillaStated as "same intended use" as predicates.
    Abutment Designs/TypesMulti-Unit, straight (0°), 17° angled, 30° angledMulti-Unit, straight (0°), 17° angled, 30° angledMulti-Unit, TLX SRA: 17°, 30°; TLX Gold: 0°, and up to 30°Similar to predicates.
    Prosthesis AttachmentScrew RetainedScrew RetainedScrew RetainedSimilar to predicates.
    RestorationMulti-unitMulti-unitMulti-unitSimilar to predicates.
    Prosthetic Interface ConnectionsInternal hex, External hexInternal hex, Internal conicalInternal conicalSimilar to predicates, with some variations noted and justified.
    Abutment/Implant Platform Diameter2.9 - 5.7 mm3.0 - 4.5 mmTLX SRA: 6 mm; TLX Gold: 4.0 (NT), 5.0 (RT), and 7.0 (WT)Within similar ranges or justified compatibility.
    Prosthetic Platform Diameter4.8 mm4.8 mmTLX SRA: 4.6 mm; TLX Gold: not provided in 510(k) SummarySimilar to predicates.
    Gingival Height1 mm - 5 mm1 mm - 5 mmNot provided in 510(k) SummarySimilar to predicates.
    Abutment MaterialTi-6Al-4V ELI (Titanium alloy)Ti-6Al-4V ELI (Titanium alloy)TLX SRA: Ti-6Al-7Nb; TLX Gold: Ceramicor®Similar (titanium alloys). Biocompatibility tested (ISO 10993-5 referenced).
    Abutment Screw MaterialTi-6Al-4V ELI (Titanium alloy)Ti-6Al-4V ELI (Titanium alloy)Ti-6Al-7NbSimilar (titanium alloys).
    How Provided (Sterilization)Non-sterile, and sterile by gamma irradiationNon-sterileTLX SRA: Sterile by gamma irradiation; TLX Gold: non-sterileBoth non-sterile and sterile options justified via reference devices/testing (e.g., K212538 for gamma, K22269 for moist heat).
    UsageSingle patient, single useSingle patient, single useSingle patient, single useIdentical.
    Mechanical PerformanceEvaluated according to ISO 14801 (mechanical testing); Engineering analysis for new worst-case constructs.Referenced-Demonstrated to not create a new worst-case or met ISO 14801.
    MR SafetyEvaluated according to ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119.--Non-clinical analysis provided to evaluate metallic devices in MR environment.
    Sterility & Shelf LifeRefer to K22269 (moist heat), K212538 (gamma irradiation, bacterial endotoxin, shelf life).Referenced-Validated to SAL of 10^-6, met bacterial endotoxin limits, and 5-year shelf life after accelerated aging.

    Important Note: The "acceptance criteria" here are qualitative statements of substantial equivalence, backed by quantitative non-clinical engineering and materials testing, not performance metrics from a clinical study or an AI model's accuracy.


    Regarding the AI/ML Specific Questions:

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not applicable. This document is for a physical medical device (dental implant abutment), not an AI/ML algorithm. There is no "test set" in the context of data for an AI/ML model. The "testing" refers to mechanical, material, and sterilization validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth establishment by experts is relevant for AI/ML diagnostic devices. This device's performance is assessed through engineering standards and material properties. Compatibility with implant systems is established through design agreements and reverse engineering data (referenced from K170588).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are used in clinical studies or for establishing ground truth labels for AI/ML datasets.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. MRMC studies are for evaluating diagnostic efficacy, particularly with AI assistance. This device is a passive prosthetic component.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. The "ground truth" for this device's performance is its adherence to engineering standards (e.g., ISO 14801 for mechanical properties, ASTM for materials/MR safety) and its demonstrated compatibility with specified implant systems, rather than a clinical ground truth like disease presence.

    8. The sample size for the training set:

    • Not applicable. There is no AI model or training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no AI model or training set.

    Conclusion from the document:

    The provided document describes a 510(k) submission for a Class II medical device, DESS Dental Smart Solutions abutments. The "acceptance criteria" revolve around demonstrating substantial equivalence to existing legally marketed predicate devices. This is achieved through:

    • Identical intended use.
    • Similar technological characteristics: This includes materials (Titanium alloy), design (multi-unit, screw-retained), and geometric parameters (diameters, gingival heights).
    • Performance via non-clinical testing: This includes mechanical testing (ISO 14801), MR safety testing (ASTM standards), sterilization validation (ANSI/AAMI/ISO standards), bacterial endotoxin testing (ANSI/AAMI ST72), shelf-life testing (ASTM F1980), and biocompatibility testing (ISO 10993-5).
    • Compatibility with specified implant systems: This is supported by contractual agreements with OEMs (ZimVie/Terrats Medical SL) and referenced reverse engineering data.

    The document explicitly states: "No clinical data were included in this submission." This further confirms that the evaluation framework is entirely non-clinical and does not involve AI/ML performance metrics, studies with human readers, or expert ground truthing as one would see for an AI-powered diagnostic tool.