Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242661

    Validate with FDA (Live)

    Date Cleared
    2025-01-08

    (126 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K151916, K212785, K230108, K192221

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TRI-matrix® X-Force implants are intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI-matrix® X-Force implants allow for one and two stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading. TRI-matrix® X-Force implants with lengths of 18, 20, or 22 mm, when placed in the maxilla, are indicated only for multiple unit restorations in splinted applications that utilize at least two implants.

    Device Description

    The purpose of this submission is to obtain marketing clearance for TRI-matrix X-Force implants, a line extension of the TRI-matrix implant line previously cleared in K203660. TRI-matrix X-Force implants are tissue-level endosseous dental implants that are an adaptation of the TRI-matrix Tissue Level implants cleared in K203660, with the same prosthetic interface, but with modified external features. They are intended for use in the mandible or maxilla to restore chewing function. This submission includes only implants, and they are compatible with prosthetic components previously cleared in K203660 as part of the TRI-matrix implant line, including the TRI-matrix Crown Abutment.

    TRI-matrix X-Force implants have a tapered body, double-lead threads and self-cutting flutes. A gritblasted and acid-etched surface, named the TRI SBA Surface, is applied to the endosseous portion of the implant. TRI-matrix X-Force implants incorporate a pink anodized, machined, 1.8 mm transgingival collar that is not treated with the TRI SBA Surface. Compared with TRI-matrix implants previously cleared in K203660, TRI-matrix X-Force implants have a modified thread design more suited to immediate placement and restoration, a more tapered apical portion (smaller apical diameter for a given coronal diameter), and more aggressive apical cutting flutes.

    TRI-matrix X-Force 3.3 mm diameter implants have a platform diameter of 3.7 mm. TRI-matrix X-Force 3.7 mm and 4.1 mm diameter implants are available in two (2) platform diameters: 3.7 mm and 4.5 mm. TRI-matrix X-Force 4.7 mm diameter and 5.7 mm diameter implants have a platform diameter of 4.5 mm.

    AI/ML Overview

    This document describes the marketing clearance for medical devices and does not contain information about acceptance criteria or a study proving that a device meets such criteria in terms of performance metrics or clinical outcomes.

    The document discusses the substantial equivalence of the TRI-matrix® X-Force Implants to legally marketed predicate devices. This means that the FDA determined the device is as safe and effective as a predicate device and does not raise different questions of safety and effectiveness. This type of submission (510(k)) generally does not require proof of meeting specific performance acceptance criteria through the kind of studies you've outlined.

    Specifically, the "PERFORMANCE DATA" section states:

    • "Non-clinical data submitted to demonstrate substantial equivalence included: Referenced from the primary predicate device K203660: sterilization validation according to ISO 17665-1 and ISO 17665-2; . Provided in this submission: biocompatibility evaluation according to ISO 10993-5 and ISO 10993-12 worst-case analysis and static and dynamic testing according to ISO 14801 on subject devices to demonstrate that static and dynamic testing referenced from the primary predicate device K203660 demonstrates the subject implants do not create a new worst-case as compared to this predicate device with identical connection platform. No clinical data were included in this submission."

    Therefore, I cannot provide the requested information for the following reasons:

    1. Table of acceptance criteria and reported device performance: This document does not specify general "acceptance criteria" for clinical performance or report clinical performance data. It focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing and design comparisons.
    2. Sample size used for the test set and data provenance: No clinical test set was used or described. The non-clinical testing involved "worst-case analysis and static and dynamic testing according to ISO 14801," but specific sample sizes for these engineering tests are not detailed.
    3. Number of experts used to establish ground truth and their qualifications: Not applicable, as no clinical test set or ground truth determination by experts is mentioned.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable, as no clinical studies were included.
    6. Standalone performance (algorithm only without human-in-the-loop performance): Not applicable, as this is a physical medical device (dental implant), not an AI algorithm.
    7. Type of ground truth used: Not applicable, as no clinical data or ground truth determination is described.
    8. Sample size for the training set: Not applicable, as no training set for an algorithm is mentioned.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232726

    Validate with FDA (Live)

    Date Cleared
    2024-04-01

    (208 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212785, K160119, K050406, K222457

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Standard Length IBR36d Implant Range:

    Southern Implants' External Hex Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading. Southern Implants' External Hex Implants are intended for immediate function when good primary stability with appropriate occlusal loading is achieved.

    When using Southern Implants' Standard Length IBR36D Implants with angulation of 36° a minimum of 4 implants must be used and splinted

    The angled Co-Axis External Hex Implants are intended to be used with straight multiple-unit abutments (Compact Conical abutments) only with no additional angulation allowable on the restoration.

    For Extra Length IBR36d Implant Range:

    Southern Implants' External Hex Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading. Southern Implants' External Hex Implants are intended for immediate function when good primary stability with appropriate occlusal loading is achieved.

    Extra Length IBR36d Implants can be placed bicortically in cases of reduced bone density. Extra Length IBR36d Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Extra Length IBR36d Implants are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.

    The angled Co-Axis External Hex Implants are intended to be used with straight multiple-unit abutments (Compact Conical abutments) only with no additional angulation allowable on the restoration.

    For Extra Length IBR24d Implant Range:

    Southern Implants' External Hex Implants are intended for surgical placement in the upper jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading. Southern Implants' External Hex Implants are intended for immediate function when good primary stability with appropriate occlusal loading is achieved.

    Southern Implants' Extra Length IBR24d Implant Range when placed in the maxilla are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    The angled Co-Axis External Hex Implants are intended to be used with straight multiple-unit abutments (Compact Conical abutments) only with no additional angulation allowable on the restoration.

    Device Description

    External Hex implants are fully-threaded, tapered, root-form dental implants with an external hexagonal abutment interface, and are threaded internally for attachment of mating multiple-unit abutments, cover screws, or healing abutments. The External Hex implants have a Co-Axis design with the prosthetic platform angled at 36° and 24° (inclined) from orthogonal to the long axis of the implant.

    Additionally, the External Hex implants are provided in two configurations, regular surface and MSC surface implants. The regular surface implants are fully roughened excluding a machined collar at the coronal aspect of the implant. The MSC surface implants have an extended length of machined area at the coronal aspect of the implant, with the remaining implant length being roughened.

    The implants subject to this submission are the External Hex IBR36d and IBR24d implant range. The reduced platform MSC-IBR24d implants are only provided as Co-Axis implants, in both the regular surface and MSC surface configuration. The reduced platform IBR36d implants are only provided as Co-Axis implants, in regular surface configuration. The IBR36d implants are available in two body configurations depending on the implant length.

    The IBR36d implants of overall lengths 8.5 mm to 18 mm have a major body diameter of 4.20 mm tapering to 2.60 mm. The IBR36d implants of overall lengths 20 mm to 24 mm have a major body diameter of 4.20 mm tapering to 2.00 mm. Both the IBR24d and MSC-IBR24d implants of overall lengths 20 mm have a major body diameter 4.07 mm tapering to 2.60 mm. All of the subject device implants utilize the same prosthetic connection, previously cleared in K163634.

    All External Hex implants are manufactured from unalloyed titanium (cold-worked, grade 4, UTS ≥ 900 MPa). The MSC-IBR24d implants have a smooth machined surface of length 3 mm extending parallel from the implant prosthetic platform for all implant lengths. The remainder of the implant is grit-blasted with aluminum oxide particles to provide a surface roughness of 1-2 um. The subject device implant material and surface are identical to those of the implants cleared in K163634.

    The IBR36d and IBR24d External Hex implants are provided with a dedicated pre-mounted fixture mount of 36° and 24 respectively, similar to that provided with the other Co-Axis External Hex implants cleared in K163634.

    All of the subject device implants utilize the same prosthetic connection, previously cleared in K163634 for the External Hex IBR24d and MSC-IBR24d implants. The External Hex implants are used in conjunction with the same abutments cleared for use with the External Hex IBR24d implants (implants cleared in K163634). These abutments are cleared in K053478, K070841, K093562, and K163634. The components that are compatible with the External Hex implants include Cover Screw, Healing Abutment, and Compact Conical Abutments. The Compact Conical Abutments are compatible with Temporary Titanium Abutment Cylinders, Gold Abutment Cylinders, and Passive Abutments.

    The subject device Co-Axis implants are indicated for use with straight multiple-unit abutments with a 0° allowable restoration angle, only. This includes restricting straight abutments, that can be restored on a multi-unit abutment, that have an allowable restoration angle of up to 20° associated with them, so that they may only have a restoration angle of 0°. Thus, all the possible compatible abutments will have a 0° allowable restoration angle.

    All External Hex implants are provided sterile to the end-user in a single-unit package, and are for single-patient, single-use only.

    AI/ML Overview

    The provided text describes the 510(k) summary for the "External Hex Implant System" and references performance data to support its substantial equivalence to legally marketed predicate devices. However, it does not contain explicit "acceptance criteria" for the device, nor a study design with specific performance metrics against those criteria in the way typically seen for AI/ML device evaluations.

    Instead, the document focuses on demonstrating substantial equivalence based on non-clinical performance data (mechanical testing and biocompatibility) and clinical literature review. The clinical literature review is used to show that similar devices with comparable characteristics (e.g., angulation, length) have proven safe and effective.

    Therefore, I cannot directly provide a table of acceptance criteria and reported device performance with specific metrics like sensitivity, specificity, or AUC, as these are not defined or reported in the provided text. The "performance" discussed relates to the device's mechanical strength and the success rates of similar implants documented in clinical literature, not an AI algorithm's diagnostic performance.

    However, I can extract the relevant information regarding the studies referenced to support the device's safety and effectiveness, interpreting "acceptance criteria" as the demonstration of substantial equivalence through various assessments:

    Interpretation of "Acceptance Criteria" for this device:

    For this device, "acceptance criteria" are not quantitative performance metrics in the traditional sense for diagnostic AI. Instead, they are met by demonstrating the device's substantial equivalence to legally marketed predicate devices through:

    • Biocompatibility: Conformance to ISO 10993-1.
    • Mechanical Performance: Conformance to ISO 14801 (static and dynamic compression-bending fatigue test), with results showing sufficient strength for intended use.
    • Sterilization: Conformance to ISO 11137-1, ISO 11137-2, and USP 39-NF 34 for bacterial endotoxin.
    • Dimensional Analysis: Verification of critical dimensions.
    • Clinical Safety and Performance (via Literature Review): Demonstrated favorable survival rates and clinical indices for comparable implants under similar conditions and specifications (length, angulation, loading protocols). This implicitly acts as an "acceptance criterion" for clinical effectiveness by showing established safety and effectiveness of similar devices.
    • MR Safety: Conformance to FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (ASTM F2503, ASTM F2013, ASTM F2182, ASTM F2119).

    Summary of Information from the Provided Text:

    1. Table of "Acceptance Criteria" (interpreted as equivalence demonstration points) and "Reported Device Performance":

    | Area of "Acceptance Criteria" (Demonstration of Equivalence) | Reported Device Performance / Evidence |
    | Biocompatibility | Evaluated according to ISO 10993-1 (referenced from K163634). |
    | Mechanical Performance (Static & Dynamic Compression-Bending) | Testing to ISO 14801 was performed on representative worst-case constructs for angulation and abutment/screw materials. Twelve (12) samples for each subject device group were subjected to fatigue testing. The fatigue limit was determined to be where a total of three (3) samples endured 5 x 10^6 cycles with no failures. Results confirmed that the strength of the subject device is sufficient for its intended use. |
    | Sterilization & Shelf Life | Sterilization according to ISO 11137-1, ISO 11137-2. Bacterial endotoxin according to USP 39-NF 34. Sterile barrier shelf life (referenced from K222457). |
    | Dimensional Analysis | Performed. (No specific values reported in this section, but implied to meet design specifications). |
    | Clinical Safety & Performance (Standard Length IBR36d) | Retrospective analysis of real-world clinical data (literature review). The literature review demonstrated that implants with lengths of 7-18mm and angulation range of 20-50° as part of a splinted approach showed similar success rates to standard implants. Survival rates and clinical indices were favorable in both maxillary and mandibular applications, with immediate and delayed loading protocols. The subject device (8.5-18mm lengths, 36° angle) falls within this "window of successful implants." |
    | Clinical Safety & Performance (Extra Length IBR36d) | Retrospective analysis of real-world clinical data (literature review). The literature review supported the use of long implants (10-25mm) placed at angles of 15-90° in the pterygoid region, showing similar success rates to standard implants. The subject device (20-24mm lengths, 36° angle) falls within this "window of successful implants." |
    | MR (Magnetic Resonance) Safety | MR safety testing (ASTM F2503, F2013, F2182, F2119) was performed on previously cleared devices (referenced from K222457). The subject devices are considered equally MR Safe as the predicate devices because they are not worst-case components in terms of material, size, or shape. |

    2. Sample Size and Data Provenance for Test Set (Clinical Literature Review):

    • Standard Length IBR36d: A retrospective analysis of real-world clinical data was conducted by reviewing 14 clinical literature articles.

      • Sample Size (Patients/Implants, per article):
        • Article 1: Not specified, but follow-up up to 17 years.
        • Article 2: Not specified (100% survival rate), follow-up 3 years (42 months).
        • Article 3: Not specified (98.7% survival rate), follow-up 3 years.
        • Article 4: Not specified (96.1% and 95.7% survival rates), follow-up 5 years.
        • Article 5: "High level of implant failures (10%), however; the same number of tilted implants and straight implants failed." Follow-up 1 year.
        • Article 6: Not specified (96.6% survival rate), follow-up 5 years.
        • Article 7: Not specified (94.5% survival rate), follow-up 7 years.
        • Article 8: Not specified (94.8% survival rate), follow-up 10 years.
        • Article 9: Not specified (98.1% for tilted implants), follow-up 3 years.
        • Article 10: Not specified (93.9% cumulative survival rate), follow-up 5-13 years.
        • Article 11: Not specified (93% survival rate), follow-up 18 years.
        • Article 12: Not specified (100% survival rate for tilted implants), follow-up 13 months.
        • Article 13: Not specified, follow-up 3 years (42 months).
        • Article 14: Not specified (100% survival rate), follow-up 3 years (42 months).
        • Note: Specific sample sizes (number of patients or implants) for each referenced article are generally not provided in the summary, which is common for literature reviews used in 510(k) submissions to support equivalence.
      • Data Provenance: Retrospective analysis of real-world clinical data (published literature). The country of origin of the data is not specified for individual studies but can be inferred from the journal names and author affiliations in the references (e.g., Clinical Oral Implants Research, Odontology, Journal of Oral Implantology, Clin Implant Dent Rel Res, European Journal of Oral Implantology, The Journal of the American Dental Association, Dentistry Journal).
    • Extra Length IBR36d: A retrospective analysis of real-world clinical data was conducted by reviewing 5 clinical literature articles.

      • Sample Size (Patients/Implants, per article):
        • Article 1: "Many authors have reported success rates of pterygoid implants ranging from 90% to 100% after follow-up period ranging from 1 to 12 years."
        • Article 2: "Thirteen articles were included, reporting a total of 1053 pterygoid implants in 676 patients."
        • Article 3: "121/125 of the implants were considered successful" after 2 years.
        • Article 4: "A total of 634 patients received 1,893 pterygoid implants."
        • Article 5: Not specified (93.9% survival rate), follow-up 3 years.
      • Data Provenance: Retrospective analysis of real-world clinical data (published literature). The country of origin of the data is not specified for individual studies but can be inferred from the journal names and author affiliations in the references (e.g., International Journal of Contemporary Dental and Medical Reviews, Journal of Oral Implantology, Dent Adv Res, Journal of Craniomaxillofacial Surgery, International Journal of Oral Maxillofacial Implants).

    3. Number of Experts and Qualifications for Ground Truth: No direct "ground truth" establishment by experts for specific cases in the context of an AI algorithm is mentioned in this document. The clinical studies cited in the literature review represent real-world clinical outcomes and expert clinical practice by the authors of those papers. Their original reports would have involved clinical expertise to determine success/failure, bone loss, etc.

    4. Adjudication Method: Not applicable. This document relies on published clinical literature, not an internal test set requiring adjudication for ground truth. The adjudication method, if any, would have been part of each individual study in the literature review.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No. This document does not pertain to an AI algorithm or human-in-the-loop performance. It describes a physical dental implant device.

    6. Standalone (Algorithm Only) Performance Study: No. This document describes a physical dental implant device, not an algorithm.

    7. Type of Ground Truth Used (for Clinical Performance): The "ground truth" for the clinical performance claims is derived from real-world clinical outcomes data reported in the referenced scientific literature. This includes:
    * Implant survival rates
    * Marginal bone loss (MBL)
    * Clinical indices (e.g., pain, prosthetic mobility, bleeding, discomfort)
    * Success rates of implants under various loading protocols (immediate/delayed) and angulations.

    8. Sample Size for the Training Set: Not applicable. This document describes a physical dental implant system, not a machine learning model that requires a training set. The "training" for the device's design would be engineering principles, material science, and clinical experience/literature.

    9. How the Ground Truth for the Training Set was Established: Not applicable, as it's not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232418

    Validate with FDA (Live)

    Date Cleared
    2023-12-01

    (112 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the implants:

    The Single Platform SP1 Implant System is intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading. The Single Platform SP1 Implant System is intended for immediate function when good primary stability with appropriate occlusal loading is achieved.

    The Single Platform SP1 implants in lengths 20, 22 and 24 mm when placed in the maxilla are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    For the conventional abutment and screws:

    The Conventional Abutments and Prosthetic Screws are premanufactured prosthetic components directly connected to endosseous dental implants and intended for use in fully edentulous maxilla and/or maxilla and/or mandible to provide support for crowns, bridges or overdentures.

    For the Titanium Abutment Bases and Passive Abutments:

    The TIB and Passive Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The TIB and Passive abutments consist of two major parts. Specifically, the titanium base and mesostructure components make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: Scan files from desktop scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

    For the Temporary Titanium Cylinders:

    The Southern Implants Temporary Titanium Abutments are prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.

    Device Description

    This submission includes fully threaded root-form dental implants with an internal cone and hexagon interface and mating abutments. The implants are provided in three diameters: Ø3.5, Ø4.0 and Ø5.0 mm. The implants are provided in a straight (0° angled) configuration only, with a 3.0 mm extended machined coronal section. The implants are provided in one prosthetic diameter (2.95 mm implants are provided in overall leneths of 8, 10, 11.5, 13, 16, 18 and 20 mm. The Ø4.0 mm implants are provided in overall lengths of 8, 10, 11.5, 13, 16, 18, 20, 22 and 24 mm. The Ø5.0 mm implants are provided in overall lengths of 8, 10, 11.5, 13, 16, and 18 mm.

    This submission also includes: a Cover Screw, Healing Abutments in four diameters and multiple gingival heights, Titanium Cylinder Abutments for temporary restorations, Passive Abutments with a plastic burn-out component, Equator Overdenture Abutments in multiple gingival heights, Compact Conical Abutments in straight (0) and angled (17° and 30°) designs, TIB Abutment Bases, and abutment screws.

    The Passive Abutments may be restored using either traditional cast-on workflows or digital CAD/CAM workflows using milled zirconia restorations (similar use to the TIB Abutment Bases).

    Using the traditional cast-on workflow, the Passive Abutments function as UCLA castable abutments which interface with a plastic, burn-out sleeve used to fabricate a prosthesis that is bonded directly to the top of the abutment, limited to a straight (0°) restoration when utilizing precious metal cast-on restorations.

    Using digital CAD/CAM workflows, the TIB Abutment Bases and Passive Abutments function as two-piece abutment designs, consisting of standard premanufactured titanium alloy abutments for supporting the second half (or top-half) of the abutment, a hybrid/crown dental restoration and mesostructure (SageMaxx Zirconia) bonded with cement (Ivoclar Vivadent Multilink Hybrid cement). The dental laboratory is to fabricate the mesostructure restoration by CAD/CAM technique out of zirconia. The TIB Abutment Bases and Passive Abutments then serves as the interface between the endosseous implant and the Zirconia restoration. The TIB Abutment Bases and Passive Abutments are designed to support the restoration on an endosseous implant in order to restore chewing function for the patient.

    The mesostructured restoration is a CAD/CAM designed prosthesis milled out of Zirconia (SageMaxx Zirconia), which is designed to fit the abutment base in order to restore chewing function for the patient. Each restoration is custom designed using 3Shape Abutment Designer Software in order to meet the requirements of each patient on a case-by-case basis. Limitations have been put in place in 3Shape Abutment Designer in order to prevent malfunctioning of the restoration and a maximum allowable angulation of 20° for the mesostructured component.

    All Single Platform SP1 implants are manufactured from unalloyed titanium conforming to ASTM F67, with a smooth machined collar (extended machined surface of 3 mm). The remainder of the entire endosseous threaded surface) is grit-blasted. The subject device implant material and surface is identical to the implants cleared in K222457 and K163060. The Cover Screw and Healing Abutments are manufactured from unalloyed titanium conforming to ASTM F67. The remaining Abutments are manufactured from titanium alloy conforming to ASTM F136. The abutment screws are manufactured from titanium alloy conforming to ASTM F136. All subject device components are manufactured in the same facilities using the same materials and manufacturing processes as used for the Southern Implants devices previously cleared in K222457 and K163060.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Single Platform SP1 Implant System." This document focuses on demonstrating substantial equivalence to existing, legally marketed predicate devices, rather than proving that the device meets specific acceptance criteria through a clinical study or a standalone algorithm performance test.

    Therefore, the requested information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in terms of clinical performance with statistical measures, expert reading, and ground truth establishment is not present in the provided document. The 510(k) pathway for medical devices typically relies on demonstrating that a new device is as safe and effective as a predicate device, rather than requiring new clinical efficacy studies unless there are significant technological differences or new intended uses that raise new safety or effectiveness questions.

    The "Performance Data" section explicitly states: "No clinical data were included in this submission."

    However, I can extract information related to the non-clinical performance data provided to support substantial equivalence and the device's characteristics compared to predicates.

    Here's a breakdown of what is available:

    1. Table of Acceptance Criteria and Reported Device Performance:
    The document does not define specific "acceptance criteria" for clinical performance (e.g., sensitivity, specificity, accuracy) or report device performance against such metrics. Instead, it demonstrates through non-clinical testing that the device's mechanical, material, and functional properties are substantially equivalent to marketed predicate devices.

    Here's a table summarizing the non-clinical performance data and findings:

    Acceptance Criteria (Proxy: Demonstration of Substantial Equivalence via Non-Clinical Performance)Reported Device Performance (Findings)
    Engineering AnalysisDemonstrated equivalence.
    Dimensional AnalysisDemonstrated equivalence.
    Sterilization Validation (ISO 11137-1, ISO 17665-1, ISO 17665-2)Validated.
    Bacterial Endotoxin (USP 39-NF 34)Tested according to standard.
    Static and Dynamic Compression-Bending (ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants)Results demonstrated fatigue performance of the subject device that exceeds its indication.
    Scanning and Milling Validation for Passive AbutmentsValidated.
    MR Safety Testing (FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" - ASTM F2503, ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119)Subject devices "can be considered equally MR Safe as the predicate devices" (as tested on previously cleared devices and compared).
    Biocompatibility Evaluations (ISO 10993-1)Supported by identical materials, processing, and storage conditions as predicate devices (referenced from K222457 and K163060).
    Sterile Barrier Shelf LifeReferenced from K222457.
    Validated Steam Sterilization Instructions (ISO 17665-2)Referenced from K193084 (for non-sterile components intended for end-user sterilization).
    Software Validation Testing (FDA Guidance Document for Off-The-Shelf Software Use in Medical Devices)Software verification and validation testing provided for abutment design library, including restrictions to prevent designs outside allowable limits and locked design limitations/specifications. (referenced from K193084, applicable to TIB Abutment Bases and Passive Abutments for digital CAD/CAM workflows).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable in the context of clinical efficacy; the "test set" here refers to the non-clinical testing of the physical device. For example, "dynamic compression-bending testing was performed on worst-case subject device constructs." The exact number of constructs tested is not specified but would be dictated by the ISO standard.
    • Data Provenance: The document does not specify the country of origin for the non-clinical testing data directly, but the manufacturer is Southern Implants (Pty) Ltd from Irene, Gauteng, SOUTH AFRICA. The data is retrospective in the sense that it supports a submission for a new device, often using internal lab data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • This information is not applicable as no clinical study with human readers or ground truth established by experts for diagnostic performance was conducted or reported.

    4. Adjudication Method for the Test Set:

    • Not applicable, as there was no human reader interpretation or clinical assessment requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done. The submission explicitly states: "No clinical data were included in this submission." Therefore, no effect size of human readers improving with AI vs. without AI assistance is reported.

    6. Standalone (Algorithm-Only) Performance:

    • Not applicable. This is a physical dental implant system, not an AI algorithm. The only "software" mentioned is for design (CAD/CAM workflow), and its validation is to ensure it prevents designs outside of allowable limitations.

    7. Type of Ground Truth Used:

    • For the non-clinical performance data, the "ground truth" is established by adherence to recognized international standards (ISO, ASTM, USP) and internal engineering specifications. For instance, the "ground truth" for fatigue performance is the passing criteria defined by ISO 14801.

    8. Sample Size for the Training Set:

    • Not applicable. This product does not involve machine learning or AI models that require a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable as there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1