K Number
K093562

Validate with FDA (Live)

Date Cleared
2010-10-14

(330 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zygomatic implant is intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with fully or partially edentulous maxillae.

Device Description

The Zygomatic implant is intended to be implanted in the upper jay arch to provide support for fixed or removable dental prostheses in patients with fully or partially edentulous maxillae.

AI/ML Overview

The provided document details a 510(k) premarket notification for a Zygomatic Implant System, focusing on demonstrating substantial equivalence to predicate devices rather than proving new efficacy. Therefore, much of the information typically found in acceptance criteria for AI/ML devices or clinical trials (e.g., performance metrics like sensitivity/specificity, sample sizes for test/training sets, expert qualifications, MRMC studies) is not present in this medical device submission.

This document describes a traditional medical device (dental implant), not an AI/ML device. The "acceptance criteria" here relate to engineering performance and biocompatibility standards rather than diagnostic accuracy.

Here's a breakdown of the available information, noting the absence of AI/ML-specific criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily based on meeting established international standards for material, manufacturing, sterilization, packaging, and mechanical performance (fatigue). The 'reported device performance' is that the device meets these standards and is technologically comparable to predicate devices.

Acceptance Criteria CategorySpecific Standards / Criteria MetReported Device Performance
MaterialASTM F67-95 Grade IV Titanium (for screw-type implants and abutments)Manufactured from ASTM F67-95 Grade IV Titanium.
SterilizationCo60 irradiation, minimum dose of 25.0 kGy (2.5 m rads), creating a Sterility Assurance Level of 10^-6. Validation per AAMI, ISO 11137, ISO 11137-2, ISO 13409.Sterilization is achieved using Co60 irradiation at a minimum dose of 25.0 kGy, meeting the specified SAL. Validation follows AAMI and ISO standards.
PackagingAdherence to numerous standards including ASTM D 4169-08, ASTM F 88-00, ASTM F 1929-98, ASTM F 1980-07, EN 552, EN226, EN 868-1:1997, EN 868-5:1999, EN 868-9:2000, EN 868-10:2000, ISO 11607. These cover performance testing of shipping containers, seal strength, leak detection, accelerated aging, and general requirements for sterile medical device packaging.Packagingvalidated following all specified standards. Implants are placed in plastic tubing, capped, and then heat-sealed in a blister pack (transparent film and Tyvek backing) which serves as the primary microbial barrier. A sterilization indicator sticker is applied, and the blister is enclosed in a plastic box for sterilization. The packaging method is consistent with their existing cleared dental implant lines.
Mechanical PerformanceAdherence to FDA Class II Special Controls Guidance Document and ISO standard 14801: 2007(E) for fatigue testing.Fatigue studies conducted as per specified guidance and standard, revealing a stable screw joint at the highest forces tested. (Detailed results are not provided in this summary but were enclosed in the full submission).
Technological EquivalenceComparable physical properties and designs to legally marketed predicate devices (K970499 Branemark System Zygomatic Implant; K070182 Nobel Biocare Zygoma Implant). No new issues of safety or effectiveness from any differences.The new implants and accessories have physical properties and designs comparable to predicate devices. A specific comparison table for the Southern 55° angle-corrected tapered implant vs. Nobel Biocare Zygomatic 45° Implant shows similar material (Titanium), exterior geometry (threaded), restorative platform width (4.05 vs 4.0 mm), and internal screw access width (2.0 mm). Differences in specific lengths, maximum diameter, apical end diameter, and screw access angle (55° vs 45°) are acknowledged but deemed not to raise new safety/effectiveness issues.
Surface ModificationConsistent with existing cleared implant lines. Blasting with 100 micron alumina particles, low pressure, visual inspection post-enhancement, and quarterly SEM testing.Same surface modification method as existing cleared devices: 100-micron alumina blasting with low pressure. Visual inspection of every implant and quarterly SEM testing are performed. The S value (roughness) of 1.43 microns is a fraction of the particle size (110 microns), suggesting minimal embedment.

2. Sample Sizes and Data Provenance

  • Test Set Sample Size:
    • For Fatigue Studies: "Fatigue studies were conducted... Testing revealed a stable screw joint at the highest forces tested." The exact number of devices tested is not specified in this summary.
    • For SEM Testing: "a sample implant is sent for SEM testing four times a year." This implies a very small, ongoing sample, not a large test set for a single study.
    • For Visual Inspection: "Each and every implant is visually inspected under a microscope after surface enhancement." This is 100% inspection during manufacturing, not a formal test set for a study.
    • For Packaging Validation: The standards referenced (e.g., ASTM D 4169-08) outline specific testing protocols which would involve a sample of packaged products, but the sample size itself is not stated in this document.
  • Data Provenance: All studies described (fatigue, surface modification checks, packaging validation) appear to be prospective tests conducted by Southern Implants, Inc. as part of their device development and manufacturing quality control. There is no indication of country of origin for the data beyond being generated by the submitter (Southern Implants, Inc., Irvine, CA, USA).

3. Number of Experts and Qualifications for Ground Truth

  • This submission does not involve establishing ground truth from expert interpretation in the way an AI/ML diagnostic device would. There are no radiologists or other clinicians involved in creating a "ground truth" for diagnostic performance.
  • The 'ground truth' here is essentially objective compliance with engineering standards and material specifications, verified through laboratory testing and manufacturing protocols by technical personnel.

4. Adjudication Method for the Test Set

  • Not applicable. There is no expert adjudication process described, as this is a physical medical device proven through engineering tests and comparison to existing devices, not by diagnostic interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (especially those involving human interpretation, often with AI assistance), which this implant is not.
  • The study performed was a comparison of technological characteristics and engineering performance (e.g., fatigue) against established standards and predicate devices.

6. Standalone (Algorithm Only) Performance Study

  • No, this is not applicable. This device is a physical dental implant, not an algorithm. Therefore, there is no "standalone algorithm" performance to report.

7. Type of Ground Truth Used

  • The "ground truth" for this device's acceptance is based on:
    • Objective Engineering Standards & Specifications: Compliance of materials (ASTM Grade IV Titanium), sterilization parameters (Co60 dose, SAL), packaging integrity (numerous ASTM, EN, ISO standards), and mechanical performance (ISO 14801: 2007(E) for fatigue).
    • Manufacturing Quality Control: Visual inspection, SEM testing for surface modification.
    • Benchmarking/Comparison: Demonstration of comparable technological characteristics to legally marketed predicate devices.

8. Sample Size for the Training Set

  • Not applicable. This is not an AI/ML device that requires a training set. The "training" in a manufacturing context would relate to process validation and operator training, not data used to develop an algorithm.

9. How Ground Truth for the Training Set was Established

  • Not applicable, as there is no training set for an algorithm.

{0}------------------------------------------------

K0933662

510(K) PREMARKET NOTIFICATION SUMMARY

Name/Address of Submitter: Southern Implants, Inc. 5 Holland, Bldg. 209 Irvine, CA 92618

OCT 1 4 2010

Establishment Registration Number: 3003845138

Contact Person: Michael A. Kehoe Phone: (866) 700-2100 x 225 Fax: (949) 273-8508

Date Summary Prepared: October 20, 2009

Device Classification Name: Endosseous Implant and Accessories

Device Classification Regulation Number: 21 CFR 872.3640

Device Regulatory Status: Class II Special Controls

Trade Name: Zygomatic Implant

Purpose: The purpose of this 510(k) is to include additional implants and accessories in the Southern Implant System that did not fall within the size range and design shapes identified in prior 510(k) submissions.

Performance Standards: FDA has not established a performance standard applicable to endosseous implants and their accessories. The materials in Southern's Zygomatic Implant System meet applicable voluntary standards. Southern Implant's screw-type implants and abutments are manufactured from ASTM F67-95 Grade IV Titanium.

K970499 Branemark System Zygomatic Implant; K070182 Nobel Biocare Zygoma Predicate Devices: Implant

Device Description and Intended Use: The Zygomatic implant is intended to be implanted in the upper jay arch to provide support for fixed or removable dental prostheses in patients with fully or partially edentulous maxillae.

Sterilization Methods Used: Sterilization of these implants will be achieved using Co60 irradiation, with a minimum dose of 25.0 kGy (2.5 m rads), creating a Sterility Assurance Level of 10 . Validation of sterilization will be done as specified by the Association for the Advancement of Medical Instrumentation (AAMI). Standards utilized include:

180 11137 Sterilization of Health Care Products - Requirements for validation and routine control - Radiation sterilization

150 11137-2 Sterilization of Medical Devices - Microbial Methods - Part 2: Tests of sterility performed in the validation of a sterilization process

Sterilization of Health Care Products - Radiation Sterilization - substantiation of 25kGy as a ISO 13409 sterilization dose for small or infrequent production batches.

Packaging Method: Please note that our implants are packaged the same as our existing line of dental implants that are cleared NSI hexed and non-hexed Implant Systems noted above as predicate devices. Implants are placed into plastic tubing (PT6.1) and capped on both ends. The plastic tube is then heat sealed in a blister pack consisting of a transparent film (P.E.T.) and a porous sheet material backing (Tyvek 1073B). This blister pack is considered the primary pack (that which provides the microbial barrier) for the implants. The Tyvek is coated with an adhesive. A sterilization indicator sticker is placed on the blister with its contents are then enclosed in a clear plastic box and sent for sterilization.

{1}------------------------------------------------

Packaging Validation: All Southern Implants packaging is validated following these standards:
ASTM D 4169-08Standard Practice for Performance Testing of Shipping Containers and Systems
ASTM F 88-00Standard Test Method for Seal strength of Flexible Barrier Materials
ASTM F 1929-98Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by DyePenetration
ASTM F 1980-07Standard Guide for Accelerated Aging of Sterile Medical Device Packages
EN 552Sterilization of medical devise – Validation and routine control of sterilization byirradiation
EN226Sterilization of medical devices - Requirements for medical devices to be labeled "Sterile"
EN 868-1:1997Packaging materials and systems for medical devices with are to be sterilized: Part 1General requirements and test methods
EN 868-5:1999Packaging materials and systems for medical devices which are to be sterilized - Part 5:Heat and self-sealable pouches and reels of paper and plastic film construction -Requirements and test methods
EN 868-9: 2000Packaging materials and systems for medical devices which are to be sterilized - Part 9:Uncoated non-woven materials of polyolefines suitable for use as packaging of medicaldevices which are to be terminally sterilized - Requirements and test methods.
EN 868-10:2000Packaging materials and systems for medical devices which are to be sterilized - Part 10:Adhesive coated nonwoven material of polyolefines for use in the manufacture of heatsealable pouches, reels and lids - Requirement s and test methods
ISO 11607Packaging for terminally sterilized medical devices

Technological Characteristics: The physical properties and designs of the additional implants and accessories in the NSI Endosseous Dental Implant System were compared with legally marketed predicate devices. The technological characteristics were comparable.

Surface Modifications: Please note that this is the same surface modification method currently used with our existing line of dental implants that are cleared NSI hexed and non-hexed Implant Systems noted above as predicate devices. The surface of our implant is blasted using 100 micron alumina (Al2O2) particles. Alumina is a highly biocompatible material and hence if any particles remain embedded in the surface, they will not pose a complication. The other measure taken to reduce the potential of embeddiment is to blast with relatively low pressure. If the indentations caused are significantly smaller than the size of the particles tend to not adhere to the surface. (Our S = 1.43 microns is a fraction of the particle size of 110 microns). Each and every implant is visually inspected under a microscope after surface enhancement as a matter of manufacturing protocol. In addition to visual inspection, a sample implant is sent for SEM testing four times a year for evaluation of the surface as well.

Brief Discussion of Clinical Studies were not conducted, or deemed necessary, for the purpose of this 510(k) submission.

Brief Discussion of Engineering Studies: Fatigue studies were conducted as per FDA Class II Special Controls Guidance Document and ISO standard 14801: 2007(E). Testing revealed a stable screw joint at the highest forces tested. See enclosed study.

Conclusions Drawn: The Southern Zygomatic Implant System has the same intended use as, and technological characteristics similar to, the legally marketed predicate devices. Any differences in the technological characteristics did not raise new issues of safety or effectiveness.

{2}------------------------------------------------

▲ ﺘﻪ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴ

Southern 55°anglecorrected tapered implantNobel Biocare Zygomatic45° Implant
Specification for materialASTM Grade IV titaniumTitanium
Exterior geometryThreadedThreaded
Implant width atrestorative platform (mm)4.054.0
Implant lengths (mm)35, 37.5, 40, 42.5, 45,47.5, 50, 52.5, 5530, 35, 40, 42.5, 45, 47.5,50, 52.5
Maximum diameter (mm)4.054.1
Apical end diameter (mm)3.82.8
Hex width x height (mm)2.7 x 0.72.7 x 0.7
Internal screw accesswidth (mm)2.02.0
Angle of screw accessopening relative to longaxis of implant55°45°

Comparison of Properties and Features of Southern's 4.0 mm 55° angle-corrected implant to Nobel Biocare's Zygomatic 45° implant

Indication for Use: The Zygomatic implant is intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with fully or partially with fully or partially edentulous maxillae.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Michael A. Kehoe President and Chief Executive Officer Southern Implants, Incorporated 5 Holland Building 209 Irvine, California 92618

OCT 1 4 2010

Re: K093562

Trade/Device Name: Zygomatic Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: September 23, 2010 Received: September 24, 2010

Dear Mr. Kehoe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Kehoe

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm_for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number: K093562

OCT 1 4 2010

Device Name: Zygomatic Implant System

.

Indication for Use: The Zygomatic implant is intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae.


Concurrence of CDRH Office of Device Evaluation

Prescription Use (Per 21 CFR801.109)

OR

Over-the-counter Use

Susan Puanor

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesion Devices Division Control, Denta. Devices

510(k) Number: K093562

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.