Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALYARD® SMART-FOLD Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    • Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 30 minutes.
    • 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 110° F/43.3ºC.

    STERIS V-PRO® Low Temperature Sterilization Systems.

    • · STERIS® V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® 1 (Lumen Cycle)

    • · STERIS® V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS® V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS® V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    • Advanced Sterilization Products STERRAD® Sterilization System

    • STERRAD® 100S

    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)

    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)

    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GL Cycle)

    • STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)

    HALYARD* ONE-STEP, QUICK CHECK and SEQUENTIAL Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    · Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for Models 100 and 200, and for 30 minutes for Models 300, 400, 500, and 600.

    • · 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 43.3℃.
      · STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles.

    • · STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® 1 (Lumen Cycle)

    • · STERIS V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    · Gravity steam at 250°F/121°C for 30 minutes (25 minute dry time for Models 100, 200 and 30 minute dry time for Models 400, 500 and 600)

    • · Advanced Sterilization Products STERRAD® Sterilization System
    • · STERRAD® 50, 100S, and 200
    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)
    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERILUCENT® HC80TT Sterilization Cycles (Lumen and Flexible Cycles)
    • · Stryker Sterizone® VP4 Sterilizer Cycle 1
    Device Description

    HALYARD SEQUENTIAL Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD ONE-STEP, HALYARD QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* SEQUENTIAL Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously.

    The HALYARD ONE-STEP, QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider.

    AI/ML Overview

    The document describes the regulatory submission for HALYARD sterilization wraps. It focuses on demonstrating substantial equivalence to a predicate device through performance testing. No AI/ML algorithm is involved in the device. Therefore, a complete response is not possible for all aspects of your request.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    PurposeTestAcceptance CriteriaResults
    Sterilant Penetration/EfficacyANSI/AAMI ST79, ANSI/AAMI/ISO 11138-7Achieving a 10^-6 sterility assurance level following processing in a worst-case half-cyclePassed
    Performance Testing (Non-sterile/Sterile)ANSI/AAMI/ISO 11607-1 Annex B, ISO 13938-2, ASTM D4966-12, CPSC 1610Complies with the selected physical propertiesPassed
    Maintenance of Package IntegrityANSI/AAMI/ISO 11607-1, ANSI/AAMI ST79Maintain sterility for 1 year, except for gravity steam (30 days), STERRAD ULTRA GI (3 months) and Sterilucent (6 months).Passed
    BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-7Non-cytotoxic, Non-irritating, Non-sensitizingPassed
    ResidualsSteris V-PRO Low Temperature Sterilization System H2O2 residuals, Sterizone VP4 Cycle 1 residual H2O2, ISO 14937 (ULTRA GI Cycle only), ISO 10993-7 EtO and ECHH2O2 ≤ 0.000218 mg/cm2, H2O2 ≤ 0.56 µg/cm2, H2O2 ≤ 9100 µg/cm2, EO < 4 mg, ECH < 9 mgPassed

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the test sets of each specific test. It refers to validation loads and specific lumen configurations used in testing, but not the total number of items tested for each criterion. The data provenance is not specified, but the testing would typically be conducted in a laboratory setting under controlled conditions. It's prospective testing as it's for a new or modified device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable as the device is a sterilization wrap, not an AI/ML diagnostic tool requiring expert interpretation for ground truth. The "ground truth" for this type of device is established through scientific and engineering standards for sterility and material properties.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable since there is no human interpretation or subjective assessment involved that would require an adjudication method. The tests are objective and based on measurable physical and chemical properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a sterilization wrap, not an AI/ML software or a device that impacts human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

    This is not applicable. The device is a physical product (sterilization wrap), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device is based on established scientific and regulatory standards for sterilization assurance, material performance, biocompatibility, and residual levels. This includes:

    • Sterility Assurance Level (SAL): A 10^-6 SAL is a universally accepted standard for sterilization, meaning there is a 1 in 1,000,000 chance of a non-sterile item.
    • Physical Properties: Standards like ISO 11607-1, ISO 13938-2, ASTM D4966-12, and CPSC 1610 define acceptable limits for the physical integrity and strength of packaging materials.
    • Biocompatibility: ISO 10993 series standards define acceptable levels for cytotoxicity, irritation, and sensitization to ensure the materials are safe for patient contact.
    • Residuals: Specific thresholds for residual sterilants (e.g., H2O2, EtO, ECH) are defined by ISO standards and validated for safety.

    8. The sample size for the training set

    This is not applicable as the device is a physical product and does not involve AI/ML requiring a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is a physical product and does not involve AI/ML requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240330

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-07

    (184 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140963 (SMART-FOLD Sterilization Wrap)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALYARD SMART-FOLD Sterilization Wrap (H450 and H650) is intended to enclose another medical device that is to be sterilized by a healthcare provider using the Advanced Sterilization Products STERRAD 100NX Sterilizer ULTRA GI Cycle.

    The H450 and H650 models of the HALYARD SMART-FOLD Sterilization Wrap have been validated for use with the STERRAD 100NX ULTRA GI Cycle as follows:
    • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions of ≥1mm inner diameter x ≤1500mm in length, or ≥2mm inner diameter x ≤1630mm in length.
    • One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle

    Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf.
    Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes

    The HALYARD SMART-FOLD Sterilization Wrap (H450, H650) is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used.

    Device Description

    Halyard Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* Sequential Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously. The HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Advanced Sterilization Products, Inc. (ASP) STERRAD 100NX Sterilizer ULTRA™ GI Cycle.

    AI/ML Overview

    The provided text describes the performance testing for the HALYARD* SMART-FOLD* Sterilization Wrap (H450 and H650).

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    PurposeTestAcceptance CriteriaResult
    Sterilant Penetration/EfficacyISO 14937AAMI TIR No 12-2010Achieving a 10$^{-6}$ sterility assurance level following processing in a worst-case half-cyclePass
    Performance Testing (Pre-Sterilization and Post-Sterilization)ANSI/AAMI/ISO 11607-1 Annex BISO 13938-2ASTM D4966-12CPSC 1610Complies with the selected physical propertiesPass
    Maintenance of Package IntegrityANSI/AAMI/ISO 11607-1ANSI/AAMI ST79AAMI TIR No 12-2010ANSI/AAMI/ISO 14937Maintain sterility for up to 90 daysPass
    BiocompatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-12ISO 14937Non-cytotoxicNon-irritatingResidual level ≤ 9100 µg/cm$^2$Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for each test. The data provenance is not explicitly mentioned as a country of origin, nor is it stated whether the data is retrospective or prospective. These are non-clinical (laboratory) tests by their nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the tests performed are non-clinical, performance-based, and biocompatibility assessments, which rely on defined technical standards and measurements rather than expert human interpretation of complex data (like image analysis).

    4. Adjudication Method

    This section is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. This submission is for a medical device (sterilization wrap), not an AI/software as a medical device (SaMD). Therefore, no MRMC study or AI-related effectiveness assessment was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable for the same reasons as point 5.

    7. The Type of Ground Truth Used

    For the tests conducted, the "ground truth" is defined by the technical standards and specifications outlined in the referenced ISO, AAMI, ASTM, and CPSC documents. For instance:

    • Sterilant Penetration/Efficacy: The ground truth is a sterility assurance level (SAL) of 10$^{-6}$. This is an objective measurement based on microbiological testing.
    • Performance Testing: The ground truth is compliance with specified physical properties (e.g., tensile strength, tear resistance as defined by the standards).
    • Maintenance of Package Integrity: The ground truth is the ability to maintain sterility for up to 90 days, which is determined through microbial barrier testing and accelerated aging studies.
    • Biocompatibility: The ground truth is material properties meeting non-cytotoxic, non-irritating, and specific residual level criteria, evaluated through chemical extractions and biological assays.

    8. The Sample Size for the Training Set

    This section is not applicable as the device is a physical sterilization wrap and does not involve machine learning or AI models with training sets.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181959

    Validate with FDA (Live)

    Date Cleared
    2019-03-22

    (242 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K140963

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALYARD SMART-FOLD* Sterlization Wap (H450, H650) is intended to enclose and is to be sterlized by a healthare provider using Steris V-PRO Low Temperature Sterilization system. The wrap was validated in the following pre-programmed cycles:

    • STERIS V-PRO 1 low temperature sterilization system (lumen cycle),
    • STERIS V-PRO 1 Plus low temperature sterilization system (Lumen and Non-lumen cycle),
    • STERIS V-PRO maX low temperature sterilization system (Lumen cycle, flexible cycle), .
    • STERIS V-PRO 60 sterilizer (Lumen, Non-Lumen cycle, flexible cycle)
    • STERIS V-PRO maX 2 low temperature sterilization system (Lumen cycle, flexible cycle)
      HALYARD SMART-FOLD* Sterlization Wrap (H450, H650) is intended to allow terminal sterlization of medical devices using vaporized hydrogen peroxide (VHP) and to maintain sterilly of the enclosed within the period of time for which performance data demonstrating maintenance of sterility has been provided.
    Device Description

    HALYARD SMART-FOLD* Sterlization Wrap is supplied in bulk to the customer as pre-shaped sterilization wap which is then used to wrap a medical device or a collection of medical devices for sterilization Wrap is compised of two preshaped sheets of HAL YARD* Sequential Sterilization Wrap. Each sheet is composed of a three-laver SMS (spunbond-meltblownspunbond) polypropylene fabric treated with an antistatic treatment.
    The SMART-FOLD* Sterilization Wrap features reinforcement reference line, a white inner layer, sidetabs with closure strips and pull-tabs which allow for aseptic presentation of the sterilized medical device. The white same material composition but contains no blue pigment.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (HALYARD SMART-FOLD* Sterilization Wrap), and it describes the device's acceptance criteria and the studies conducted to prove it meets those criteria. However, it is not an AI/ML device, so many of the requested fields related to AI/ML model validation (e.g., sample size for training set, number of experts for ground truth, MRMC studies) are not applicable and therefore not present in the document.

    Here's a breakdown of the information that is available, formatted to address your specific questions where possible.


    Device Name: HALYARD SMART-FOLD* Sterilization Wrap (Models: H450, H650)
    Device Type: Sterilization Wrap (not an AI/ML device)
    Intended Use: To enclose medical devices for sterilization using Steris V-PRO Low Temperature Sterilization systems and maintain sterility until used.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several performance tests and their results, which essentially serve as the acceptance criteria and proof of performance for this non-AI medical device. The "acceptance criteria" are implied by the "passing results" of these tests.

    Acceptance Criteria (Implied by Test Objective)Reported Device Performance (Results)
    Sterilant Penetration/Efficacy: Demonstrate effective sterilant penetration into instrument trays wrapped with the device, under half-cycle and worst-case lumen conditions, achieving a 10⁻⁶ SAL.The Halyard SMART-FOLD* (H450, H650) Sterilization wrap demonstrated effective sterilant penetration into instrument trays wrapped with Halyard Health SMART-FOLD* (H450, H650) Sterilization Wrap after processing in the V-PRO maX low temperature sterilizer under half cycle condition in the V-PRO maX low temperature sterilizer, worst-case lumen cycle. (And achieved a 10⁻⁶ SAL following processing in a worst-case half-cycle, as noted in the Technological Characteristics Comparison section for both proposed and predicate devices).
    Maintenance of 365-Day Package Integrity: Demonstrate that the device is an effective barrier for maintaining sterility of various sterilization loads for 30, 180, and 365 days after processing in the worst-case V-PRO 60 Low Temperature Sterilizer Lumen cycle, followed by an event-related shelf life study.The Halyard SMART-FOLD* (H450, H650) Sterilization wrap demonstrated effective barrier using the simulated test pack for the best and worst-case weight wraps for maintaining the sterility of their contents for 30, 180, 365 days after processing in the worst-case V-PRO 60 Low Temperature Sterilizer Lumen cycle and followed by an event-related shelf life study. (Also explicitly stated 365 days passing results for all V-PRO models in Technological Characteristics Comparison section).
    Material Biocompatibility: Demonstrate non-irritating and non-cytotoxic properties in accordance with ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).The Halyard SMART-FOLD* (H450, H650) Sterilization wrap was found to be non-cytotoxic and non-irritant based on the conditions of these tests.
    Physical Properties: Demonstrate passing results for various physical properties (bursting, air permeability, abrasion resistance, mass per unit area, flame resistance, water resistance, linting/cleanliness, tearing strength, breaking strength, seam failure, bacterial filtration efficiency).The physical properties testing met the acceptance criteria and demonstrated passing results for the Halyard SMART-FOLD* (H450, H650) Sterilization wrap.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes (e.g., number of sterilization wraps or test packages) used for each test. For example, for "Maintenance of 365-Day Package Integrity," it mentions "simulated test pack for the best and worst-case weight wraps" but not the quantity of these packs.

    Data Provenance: The studies were conducted by the manufacturer, Owens & Minor Halyard, Inc., based in Alpharetta, GA (USA). The document does not specify if the testing was performed retrospectively or prospectively, but standard medical device validation studies are typically prospective.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not Applicable. This is not an AI/ML device that requires human expert annotation for ground truth. The acceptance criteria are based on physical and biological laboratory tests.


    4. Adjudication Method for the Test Set

    Not Applicable. As this is not an AI/ML device, there are no "interpretations" that would require adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    Not Applicable. This is not an AI/ML device, and no human reader effectiveness is being evaluated.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    Not Applicable. This is a physical medical device, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Laboratory Testing Standards: Adherence to established scientific and regulatory standards (e.g., ISO 10993 for biocompatibility, various ASTM and ISO standards for physical properties, and internal validation protocols for sterilization efficacy).
    • Biological Indicators: For sterilization efficacy tests, the "ground truth" of sterility is typically determined by the absence of microbial growth in biological indicators after processing.
    • Physical Measurements: For physical properties, the ground truth is direct measurement against pre-defined specifications.

    8. The Sample Size for the Training Set

    Not Applicable. This is not an AI/ML device and does not have a "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As there is no training set for an AI/ML model, this question is irrelevant to the provided document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1