510(k) Data Aggregation

    K Number
    K171250

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-02

    (157 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BRIDALVEIL Occipital Cervical Thoracic System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (Cl to C7) and the thoracic spine from TI-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions ( e.g., pseudarthrosis ); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The BRIDALVEIL Occipital Cervical Thoracic System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    In order to achieve additional levels of fixation, the BRIDALVEIL Occipital Thoracic System may be connected to the OLYMPIC Posterior Spinal Fixation System rods and connectors. Transition rods with differing diameters may also be used to connect the BRIDALVEIL Occipital Cervical Thoracic System to the OLYMPIC Posterior Spinal Fixation System. Refer to the OLYMPIC Posterior Spinal Fixation System package insert for instructions for use and indications for use.

    Device Description

    The BRIDALVEIL Occipital Cervical Thoracic System is a spinal fixation system intended to stabilize the uppermost portion of the spine during the fusion process. The system contains a wide variety of implants and instruments which allows for the transition across multiple spinal segments: Occipital Plate with Screws, Cervical Polyaxial Screws, Laminar Hooks, Cross Connectors, Rod Connectors, and Rods manufactured from Ti6Al4V ELI (ASTM F136) and cobalt chrome alloy (ASTM F1537).

    AI/ML Overview

    The provided text describes the regulatory clearance for the BRIDALVEIL Occipital Cervical Thoracic System, a spinal fixation system. It does not detail acceptance criteria and a study that proves the device meets those criteria in the context of an AI/ML device.

    This document is a 510(k) premarket notification for a traditional medical device (spinal fixation system), not an AI/ML device. Therefore, the details requested in the prompt, such as reported device performance, sample size for test sets, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment for the training set, are not applicable in the context of this submission.

    Instead, the document focuses on demonstrating substantial equivalence to existing predicate devices based on:

    • Intended Use: Providing immobilization and stabilization of spinal segments as an adjunct to fusion for various conditions (traumatic fractures/dislocations, instability, tumors, degenerative disease, etc.) in the craniocervical junction, cervical spine (C1-C7), and thoracic spine (T1-T3). It also mentions restoring spinal column integrity for a limited time in advanced stage tumor patients.
    • Design: The system includes occipital plates with screws, cervical polyaxial screws, laminar hooks, cross connectors, rod connectors, and rods.
    • Materials: Ti6Al4V ELI (ASTM F136), cobalt chrome alloy (ASTM F1537), Elgiloy CoCrNi alloy (ASTM F1058), and Nitinol #1 (ASTM E2063).
    • Mechanical Safety and Performance: Evaluated through non-clinical bench testing.

    Here's the information extracted from the document, framed in the context of a traditional medical device submission, rather than an AI/ML one:


    1. A table of acceptance criteria and the reported device performance

    For this traditional medical device (BRIDALVEIL Occipital Cervical Thoracic System), "acceptance criteria" revolve around demonstrating substantial equivalence to predicate devices through conformity to recognized standards for mechanical performance and material properties. "Reported device performance" is derived from the results of these non-clinical tests meeting the requirements of those standards.

    Acceptance Criterion (Demonstrating Substantial Equivalence)Reported Device Performance (Summary from Non-Clinical Tests)
    Mechanical Performance:The BRIDALVEIL Occipital Cervical Thoracic System demonstrated mechanical performance equivalent to the predicate devices across various tests.
    Static Compression Bending (ASTM F1717)Results indicated equivalence to predicate devices.
    Dynamic Compression Bending (ASTM F1717)Results indicated equivalence to predicate devices.
    Static Torsion (ASTM F1717)Results indicated equivalence to predicate devices.
    Static Compression Bending (ASTM F2706)Results indicated equivalence to predicate devices.
    Dynamic Compression Bending (ASTM F2706)Results indicated equivalence to predicate devices.
    Static Torsion (ASTM F2706)Results indicated equivalence to predicate devices.
    Dynamic Torsion (ASTM F2706)Results indicated equivalence to predicate devices.
    Static Axial Grip (ASTM F1798)Results indicated equivalence to predicate devices.
    Static Torsional Grip (ASTM F1798)Results indicated equivalence to predicate devices.
    Static Transverse Moment (ASTM F1798)Results indicated equivalence to predicate devices.
    Material Biocompatibility:Materials used (Ti6Al4V ELI, CoCrMo alloy, Elgiloy CoCrNi alloy, Nitinol #1) are commonly accepted in spinal implants and are equivalent to those used in predicate devices, implying established biocompatibility. (No specific biocompatibility tests are listed, but material equivalence is claimed.)
    Sterility:(Not explicitly detailed in the summary, but implicit for an implantable device that it would meet relevant sterility standards.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable in the context of human data. The "test set" here refers to physical specimens of the device components tested in a laboratory setting. No human patients or retrospective/prospective data were used for performance evaluation.
    • Data Provenance: The data provenance is from non-clinical bench testing conducted in a laboratory. The specific country of origin of the lab is not stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. "Ground truth" in this context refers to the defined parameters and performance standards derived from the ASTM standards for mechanical testing. These standards are established by expert committees in engineering and materials science, but individual experts are not "adjudicating" a test set as they would for clinical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There's no human adjudication process described for the mechanical test results; rather, the results are compared against predefined criteria within the ASTM standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, and no MRMC studies or human reader performance evaluations were conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML device. The "standalone performance" is the mechanical testing of the device itself against engineering standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For this device, the "ground truth" (or reference for evaluation) is established through recognized industry standards for mechanical testing of spinal implants (e.g., ASTM F1717, ASTM F2706, ASTM F1798). The device's performance is compared against the requirements and typical performance of predicate devices as measured under these standardized conditions.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151885

    Validate with FDA (Live)

    Date Cleared
    2015-09-03

    (56 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132332, K042508

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SUMMIT® SI OCT Spinal Fixation System and MOUNTAINEER® OCT Spinal System are intended to provide immobilization and stabilization of spinal segments as an adjunct to following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3):

    • · traumatic spinal fractures and/or traumatic dislocations;
    • · instability or deformity;
    • · failed previous fusions (e.g. pseudarthrosis);
    • · tumors involving the cervical/thoracic spine;

    • and degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and

    · degenerative disease of the facets with instability.

    The SUMMIT and MOUNTAINEER Systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    The SONGER® Wire/Cable System to be used with the SUMMIT and MOUNTAINEER Systems allows for wire/cable attachment to the posterior cervical spine.

    The SUMMIT and MOUNTAINEER Systems can also be linked to the ISOLA®, MONARCH®, MOSS® MIAMI, VIPER® and EXPEDIUM® Spine Systems using the dual wedding band and axial connectors, and via dual diameter rods.

    Device Description

    The SUMMIT and MOUNTAINEER Systems are posterior spinal fixation systems intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3). The systems are composed of multiple components to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The systems consist of bone anchors including hooks, SONGER® cables and screws, longitudinal members including rods and occipital plates, transverse connectors including cross connectors and interconnection mechanisms including lateral offset connectors, cable connectors, set screws, nuts, axial and wedding band connectors. The system components are implanted using class I exempt manual surgical instruments. This premarket notification expands the indications for use of the SUMMIT and MOUNTAINEER Systems to include posterior cervical screw fixation.

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria in the context of AI/ML device performance. This document is a 510(k) premarket notification for a spinal fixation system, focusing on its substantial equivalence to predicate devices based on technological characteristics and mechanical performance testing, rather than an AI/ML driven device and its performance metrics.

    Therefore, I cannot extract the requested information from the provided text.