(243 days)
The VuePoint® OCT System is intended to promote fusion of the cervical spine and occipito-thoracic junction (Occiput-T3), and is indicated for: (1) Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) Degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) Fracture/Dislocation, (4) Spinal Stenosis, (5) Atlantoaxial fracture with instability, (6) Occipitocervical dislocation, (7) Spinal tumor and/or (8) Revision of previous cervical spine surgery.
The occipital bone screws are limited to occipital fixation only.
The use of polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.
The VuePoint OCT System can also be linked to the NuVasive SpheRx Spinal System via the rod to rod connectors or transition rods.
The NuVasive VuePoint OCT System consists of a variety of shapes and sizes of screws, rods, offset connectors, set screws, and cross connectors which can be rigidly locked in a variety of configurations to accommodate patient anatomy.
Here's a breakdown of the acceptance criteria and study information for the NuVasive® VuePoint® OCT System based on the provided 510(k) summary:
This device did not involve AI or algorithms; it is a spinal fixation system. Therefore, many of the requested categories related to AI performance, ground truth establishment, expert adjudication, MRMC studies, and training sets are not applicable. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on demonstrating substantial equivalence to predicate devices through biomechanical testing.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Performance Measure) | Reported Device Performance (Goal/Outcome) |
|---|---|
| Biomechanical Testing | |
| Static and dynamic torsion | Met or exceeded the performance of the predicate device (per ASTM draft WK455-Z9592Z). |
| Static and dynamic compression | Met or exceeded the performance of the predicate device (per ASTM draft WK455-Z9592Z). |
| Interconnection strength | Met or exceeded the performance of the predicate device (per ASTM F1798). |
| Overall Substantial Equivalence | Demonstrated to be substantially equivalent to legally marketed predicate devices in terms of safety and effectiveness for its intended use. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes (number of constructs or test specimens) used for each biomechanical test.
- Data Provenance: Not applicable in the context of clinical data. The tests are nonclinical (laboratory-based biomechanical tests) performed to demonstrate mechanical equivalence. Country of origin for data is not specified, but the submission is to the U.S. FDA, implying adherence to U.S. recognized standards. The study is nonclinical (in-vitro testing).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. This device is a spinal fixation appliance, and the "ground truth" for its performance is established by standardized biomechanical tests demonstrating its physical and mechanical properties, not by human expert opinion or clinical outcomes in the traditional sense. The "ground truth" is adherence to established engineering standards for strength and durability.
4. Adjudication Method for the Test Set
- Not applicable. There was no human adjudication of "cases" or "ground truth" as this was a nonclinical biomechanical study. The adjudication is against the performance of predicate devices and relevant ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- No. This is not an AI-powered device. It is a physical spinal implant system. Therefore, MRMC studies and AI assistance are not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- No. This is not an algorithm or AI device.
7. The Type of Ground Truth Used
- Nonclinical Biomechanical Performance. The "ground truth" for this device's performance is its ability to meet or exceed the biomechanical performance (static and dynamic torsion, compression, interconnection strength) of legally marketed predicate devices, as measured by established ASTM (American Society for Testing and Materials) standards.
8. The Sample Size for the Training Set
- Not applicable. This non-AI device does not have a "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. This non-AI device does not have a "training set" or "ground truth" established for training in the context of machine learning.
{0}------------------------------------------------
JUN 2 3 2010
NUVASIVE
Creative Spine Technology
510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
A. Submitted by:
Sheila Bruschi Senior Regulatory Affairs Associate NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 320-4515 Fax: (858) 320-4615
Date Prepared: June 14, 2010
B. Device Name
| Trade or Proprietary Name: | NuVasive® VuePoint® OCT System |
|---|---|
| Common or Usual Name: | Spinal Fixation Appliances |
| Classification Name: | Appliance, Fixation, Spinal Interlaminal and Spinal Pedicle Fixation Orthosis |
| Device Class: | Class II |
| Classification: | §888.3050 |
| Product Code: | KWP |
C. Predicate Devices
The subject VuePoint OCT System is substantially equivalent to the following predicate devices currently distributed commercially in the U.S .:
- . K071435 - NuVasive OCT System
- K092287 NuVasive SpheRx® II Pedicle Screw System ●
- K080828 DePuy Spine™ MOUNTAINEER™ OCT Spinal System .
D. Device Description
The NuVasive VuePoint OCT System consists of a variety of shapes and sizes of screws, rods, offset connectors, set screws, and cross connectors which can be rigidly locked in a variety of configurations to accommodate patient anatomy.
E. Intended Use
The VuePoint® OCT System is intended to promote fusion of the cervical spine and occipito-thoracic junction (Occiput-T3), and indicated for:
- I. Degenerative disc disease (as defined by back pain' of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
-
- Degenerative spondylolisthesis with objective evidence of neurologic impairment,
-
- Fracture/Dislocation,
{1}------------------------------------------------
-
- Spinal Stenosis,
-
- Atlantoaxial fracture with instability,
-
- Occipitocervical dislocation,
-
- Spinal tumor and/or
-
- Revision of previous cervical spine surgery.
The occipital bone screws are limited to occipital fixation only,
The use of polyaxial screws is limited to placement in the upper thoracic spine (TI-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.
The VuePoint OCT System can also be linked to the NuVasive SpheRx Spinal System via the rod to rod connectors or transition rods.
F. Technological Characteristics
As was established in this submission, the subject VuePoint OCT System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.
Performance Data G.
Nonclinical testing was performed to demonstrate that the subject VuePoint OCT System is substantially equivalent to other predicate devices. The following testing was performed:
- . Static and dynamic torsion per ASTM draft standard WK455-Z9592Z
- Static and dynamic compression per ASTM draft standard WK455-Z9592Z .
- . Interconnection testing per ASTM F1798
The results of these studies showed that the subject VuePoint OCT System meets or exceeds the performance of the predicate device, and the device was therefore found to be substantially equivalent.
H. Conclusions
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject VuePoint OCT System has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 23 2010
NuVasive, Inc. % Ms. Sheila Bruschi Senior Regulatory Affairs Associate 7475 Lusk Boulevard San Diego, California 92121
Re: K093319
Trade/Device Name: VuePoint® OCT System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: May 19, 2010 Received: May 20, 2010
Dear Ms. Bruschi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Ms. Sheila Bruschi
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Barbara Budim
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K093319
Device Name: VuePoint® OCT System
Indications For Use:
The VuePoint® OCT System is intended to promote fusion of the cervical spine and occipito-thoracic junction (Occiput-T3), and is indicated for: (1) Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc by patient history and radiographic studies), (2) Degenerative confirmed spondylolisthesis with objective evidence of neurologic impairment, (3) Fracture/Dislocation, (4) Spinal Stenosis, (5) Atlantoaxial fracture with instability, (6) Occipitocervical dislocation, (7) Spinal tumor and/or (8) Revision of previous cervical spine surgery.
The occipital bone screws are limited to occipital fixation only.
The use of polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.
The VuePoint OCT System can also be linked to the NuVasive SpheRx Spinal System via the rod to rod connectors or transition rods.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KO93319 510(k) Number_
Page 1 of 1
N/A