Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242988

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-27

    (274 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.

    One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.

    Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)

    Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.

    The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.

    The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.

    One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus

    Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.

    Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.

    The Clavicle Superior Plates 3.5 are indicated for:

    • Fractures of the clavicle shaft
    • Fractures of the lateral clavicle
    • Malunions of the clavicle
    • Non-unions of the clavicle

    The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.

    The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.

    The Tibia Proximal Lateral Plates 3.5 are indicated for:

    • Split-type fractures of the lateral tibial plateau
    • Lateral split fractures with associated depressions
    • Pure central depression fractures
    • Split or depression fractures of the medial plateau

    The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.

    The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.

    The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for

    • Intraarticular fractures of the distal humerus
    • Supracondylar fractures of the distal humerus
    • Non-unions of the distal humerus

    The Humerus Distal Medial Plates 2.7/3.5 are indicated for

    • Intraarticular fractures of the distal humerus
    • Supracondylar fractures of the distal humerus
    • Non-unions of the distal humerus

    The Ulna Proximal Olecranon Plates 3.5 are indicated for

    • Complex extra- and intra-articular olecranon fractures
    • Non-unions of the proximal ulna
    • Osteotomies
    • Simple olecranon fractures

    Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.

    Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.

    Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.

    T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur

    Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.

    Device Description

    The Genostis Osteosynthesis System comprises of fracture fixation plates, corresponding screws and required surgical instruments. Different shapes and sizes are available to accommodate for different anatomical regions and locations of fractures. The plate screw interface is either locking, non-locking or, for some plates, realized by a combination hole offering both options (locking and non-locking).

    All plates and screws are made of titanium alloy (Ti 6Al 4V ELI) according to ASTM F136 and ISO 5832-3.

    The implants of the Genostis Osteosynthesis System are sterilized by radiation and are delivered sterile.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Genostis Osteosynthesis System does not describe a study involving an AI/software device or human readers. Instead, it describes a traditional medical device (bone fixation plates and screws) and primarily references non-clinical (biomechanical and biocompatibility) testing to demonstrate substantial equivalence to predicate devices.

    Therefore, many of the requested criteria related to AI/software performance, human reader studies, and ground truth establishment in a diagnostic context are not applicable to this document.

    However, I can extract and structure the information that is relevant to the device's acceptance criteria based on the provided text, particularly concerning non-clinical testing.

    Here's the breakdown based on the information available:


    Acceptance Criteria and Study for Genostis Osteosynthesis System

    The Genostis Osteosynthesis System is a traditional medical device (bone fixation plates and screws), not an AI/software diagnostic device. As such, the "acceptance criteria" and "study" proving it meets them are primarily based on non-clinical engineering and biological tests, not AI performance metrics or human-in-the-loop diagnostic studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device Performance
    Material BiocompatibilityCytotoxicity (ISO 10993-5)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Sensitization (ISO 10993-10)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Irritation/Intracutaneous Reactivity (ISO 10993-23)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Chemical Characterization (ISO 10993-18)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Toxicological Risk AssessmentSatisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Bacterial Endotoxin TestingAcceptable endotoxin levelsPerformed; results satisfactory.
    Mechanical PerformanceScrew Performance (ASTM F543)Satisfactory mechanical performancePerformed; results satisfactory.
    Plate Performance (ASTM F382)Satisfactory mechanical performancePerformed; results satisfactory.
    Packaging & SterilizationSterility (Radiation)Device must be sterile at deliveryImplants are sterilized by radiation and delivered sterile. Test conducted on final sterile product.
    GeneralSubstantial Equivalence to PredicatesAll non-clinical testing demonstrates substantial equivalence.All testing indicates substantial equivalence to predicates.

    Note: The document states "All results were satisfactory" for all tests, indicating the device met the pre-defined acceptance criteria for each test.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact number of units or samples used for each biomechanical or biocompatibility test. It generally states that testing was "performed" and "conducted on final sterile product."
    • Data Provenance: Not explicitly stated as "country of origin," but testing would be performed in a controlled laboratory environment. The tests are non-clinical (biomechanical and biocompatibility), not data from patients. The study is by nature prospective in that new manufactured devices were tested to meet the standards, not a retrospective analysis of existing data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This is a medical device (hardware) clearance, not an AI/software diagnostic device that requires expert ground truth for interpretation of medical images or other diagnostic data. The "ground truth" here is the adherence to established engineering and biocompatibility standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is non-clinical mechanical and biocompatibility testing, there is no "adjudication" in the sense of reconciling expert opinions on diagnostic findings. The results are quantitative and objective measurements against established standard thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • Not Applicable. MRMC studies are for evaluating clinical interpretation of diagnostic results, typically with AI assistance. This device is a physical implant, not a diagnostic tool or AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This pertains to AI algorithm performance. The device is a bone fixation system and does not rely on an algorithm for its primary function.

    7. The Type of Ground Truth Used

    • Pre-defined Standards and Controls: The "ground truth" for this device's performance is established by well-defined, internationally recognized (ISO) and American (ASTM) standards for material properties, mechanical strength, and biocompatibility. The device must perform within the specified parameters of these standards. For example, screws must meet certain torque resistance (ASTM F543), and plates must meet specific load-sharing characteristics (ASTM F382).

    8. The Sample Size for the Training Set

    • Not Applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. There is no "training set" for this type of medical device clearance. The design and manufacturing processes are established based on engineering principles and existing predicate device designs, with performance validated against the standards mentioned above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140357

    Validate with FDA (Live)

    Date Cleared
    2014-06-09

    (117 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OIC Variable Angle Small Fragment Locking Plate System is indicated for the fixation of fractures, mal-unions, nonunions or osteolomies for the clavicle, humerus, radius, ulna, metacarpal, tibia, fibula, malleolus and metatarsal.

    Device Description

    The OIC Variable Ande Small Fragment Locking Plate System consists of titanium plates for the Distal Radius. Clavice. Proximal Humerus, Tibia and Distal Fibula, bone screws and instruments for implantation. The platety of sizes and are pre-contoured to match the anatomy of the patient and 3.5mm bone screws. The bone screws are available in two diameters (2.5mm) and range in length from 6mm to 130mm. The bone screws are available with both threaded (locking) and non-threaded (non-locking) heads.

    The OIC Variable Angle Small Fragment Locking Plate System implants are made of titanium alloy or cp titanium in compliance with ASTM F136 or ASTM F67.

    AI/ML Overview

    The provided text describes a medical device submission (K140357) for the OIC Variable Angle Small Fragment Locking Plate System. However, it does not contain information regarding traditional acceptance criteria or a study proving the device meets said criteria in the context of clinical performance or diagnostic accuracy.

    Instead, this document focuses on demonstrating substantial equivalence to existing legally marketed devices, primarily through engineering and material testing, rather than clinical trials with patient outcomes or AI-driven performance studies.

    Therefore, many of the requested categories cannot be filled as the information is not present in the provided text.

    Here's an attempt to answer based on the available information, with clear indications where the information is not present:


    Acceptance Criteria and Device Performance Study for K140357: OIC Variable Angle Small Fragment Locking Plate System

    The provided 510(k) summary for the OIC Variable Angle Small Fragment Locking Plate System does not detail specific acceptance criteria or clinical performance studies related to diagnostic accuracy or human-in-the-loop improvements for an AI device. Instead, the document focuses on demonstrating substantial equivalence through mechanical testing and material compliance.

    The "acceptance criteria" can be inferred from the standards the device conforms to and the evaluation methods used to demonstrate substantial equivalence, which are primarily related to mechanical properties and materials.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Inferred from standards/methods)Reported Device Performance
    Material ComplianceConformance to ASTM F136 or ASTM F67Implants made of titanium alloy or cp titanium in compliance with ASTM F136 or ASTM F67.
    Bone Screw StandardsConformance to ASTM F543-07Device conforms to ASTM F543-07.
    Metallic Bone Plate StandardsConformance to ASTM F382Device conforms to ASTM F382.
    Mechanical CharacteristicsAcceptable mechanical characteristics for intended uses, similar to predicate devices.The screws, plates, locking interfaces, and behavior were evaluated using Finite Element Stress/Strain Analyses (FEA). They were "found to have acceptable mechanical characteristics for the intended uses." Each component and construct showed "similar mechanical performance compared to similar devices used for the same indications as identified via a literature review."
    SterilizationProvided non-sterile, suitable for steam sterilization.The device is provided non-sterile and is steam-sterilized by the medical facility prior to implantation.
    Substantial EquivalenceNo significant differences from predicate devices affecting safety and effectiveness."The new device is substantially equivalent to the predicate devices in regards to intended use, materials, and function. There are no significant differences... Any minor differences have no effect on safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable/not provided. The evaluation was primarily based on engineering analyses (FEA) and literature review, not a test set of data in the context of AI or clinical outcomes.
    • Data Provenance: Not applicable/not provided. The document refers to "literature review" for comparative mechanical performance, but does not specify the origin of this literature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Not applicable. The ground truth for mechanical performance was established via engineering analyses and comparisons to published standards and literature, not expert consensus on a test set of cases.

    4. Adjudication Method for the Test Set

    • Not applicable. No test set requiring expert adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices or AI-assisted interpretation, which is not the nature of this orthopedic implant submission.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm performance study was not done. This device is a physical orthopedic implant, not an algorithm.

    7. The Type of Ground Truth Used

    • The 'ground truth' for this 510(k) submission relates to engineering standards, material specifications, and documented mechanical performance characteristics established through finite element analysis and literature review. It is not clinical pathology, outcomes data, or expert consensus on patient cases.

    8. The Sample Size for the Training Set

    • Not applicable. This device is an orthopedic implant, not an AI algorithm requiring a training set. The "training" for its design and evaluation would come from established engineering principles, material science, and the design of predicate devices.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no "training set" in the context of this orthopedic implant. The benchmarks for performance (comparable mechanical properties, material compliance, etc.) were established through recognized ASTM standards, the performance of predicate devices, and engineering analyses (FEA).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1