Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221836

    Validate with FDA (Live)

    Date Cleared
    2022-12-07

    (167 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K013122

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Filter CareStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in pediatric (with a tidal volume between 100 and adult patients, depending on the respective device. Filter SafeStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult patients. Filter/HME TwinStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and for respiratory use, as well as heat and moisture exchanger for humidifying respired gases for the patient. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult, pediatric and neonatal patients, depending on the respective device.

    Device Description

    The devices are breathing circuit filters used to filter the inhaled and/or the exhaled air of the patient against microbiological and particulate matter from the gases in the breathing circuit. They enclose a filter material in a housing that fits to standard breathing system connectors. Additionally, there are breathing system filters combined with a foam to function as HME (Heat and Moisture Exchangers) for passively humidifying the inspired air. The portfolio contains the following types of breathing circuit filters: - Filter CareStar Plus are electrostatic filters for use against contamination with microorganisms - . Filter SafeStar Plus are mechanical filters for use against contamination with microorganisms - Filter/HME TwinStar Plus are filters for use against contamination with microorganisms and for passive humidification of breathing gases

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The provided document describes three devices: Filter CareStar Plus, Filter SafeStar Plus, and Filter/HME TwinStar Plus. The table below compiles the acceptance criteria and reported performance for these devices based on the "Summary of non-clinical testing" section (pages 18-19).

    Test Method & PurposeAcceptance CriteriaReported Performance
    ISO 9360-1:2000 - Determination and Evaluation of Pneumatic ComplianceCompliance is less than or equal to 1mL/kPa at 15, 30, 60, and 70 hPaPASSED
    ISO 9360-1:2000 - Determination and Evaluation of Pneumatic LeakagePneumatic leakage is less than or equal to 50mL/min at 70hPa.PASSED
    ISO 9360-1:2000 - Determination and Evaluation of Pressure Drop (Pneumatic Resistance)Pneumatic resistance/pressure drop is acc. to IfU valuePASSED
    ISO 5356-1:2015 - Evaluation of Conical Connectors (ISO 5356-1)Cone dimensions comply with ISO 5356-1.PASSED
    ISO 80369-7:2021 - Luer Lock Connector (ISO 80369-7)Luer-Lock connector fulfills the requirements laid out in ISO 80369-7.PASSED
    IEC 60601-1:2005 - Product's Ability to Withstand Damage from DroppingWhen dropped, the product should not suffer any damage which influences its function.PASSED
    ISTA 3A - Product Durability During Transport, Mechanical Aspects- The packaging shows no or minor damage- The DUT (Device Under Test) shows no signs of damage and retains functionality after simulated transport.PASSED
    ISO 23328-1:2003 - Filtration Efficiency (Particulate Matter) incl. Usage Time- Adult and pediatric electrostatic filters achieve a filtration efficiency of >90% before and after the specified usage time- Neonatal filters achieve a filtration efficiency of >75% before and after the specified usage time- Mechanical filters achieve a filtration efficiency of >99% before and after the specified usage time- Mechanical filters achieve a HEPA classification ≥ class H13PASSED
    ASTM F2101:2019 - Filtration Efficiency (Viral and Bacterial)- Electrostatic filters achieve 99.99% (bacterial) and 99.9% (viral) filtration efficiency- Electrostatic filters for neo applications achieve 99.98% (bacterial) and 99.9% (viral) filtration efficiency.PASSED
    ISO 10993:2018 and ISO 18562-1:2017 - Evaluation of Product's Biological CompatibilityEvaluation according to ISO 10993:2018 and/or ISO 18562-1:2017PASSED
    ISO 9360-1:2000 - Evaluation of HME Water Loss, Resistance· pediatric/neonatal: The moisture loss shall be <=11mg/L at VT=50ml· pediatric: The moisture loss shall be <=12mg/L at VT=250ml· adult: The moisture loss shall be <=11mg/L at VT=500mlPASSED

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each non-clinical test. It mentions that "The devices... have undergone extensive testing" but does not detail the number of units tested for each criterion. The data provenance is also not specified; however, given that the submitter is "Drägerwerk AG & Co. KGaA" based in "Lübeck, Germany" and they tested against international standards (ISO, ASTM, IEC), it is likely that the testing was conducted in a controlled laboratory environment, potentially in Germany or by a certified testing facility adhering to these standards. The testing appears to be prospective, specifically designed to verify the devices' adherence to the mentioned standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for the non-clinical test set. The validation relies on adherence to established international and national standards (e.g., ISO, ASTM, IEC) rather than expert consensus on a test set. This implies that the "ground truth" is defined by the objective metrics and methodologies outlined in these standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    There is no mention of an adjudication method in the text for the non-clinical test set. Given the nature of the tests (physical and performance characteristics against standards), it's a pass/fail determination based on quantitative measurements against predefined criteria, not a subjective assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document explicitly states "Summary of Clinical Testing N/A" (page 20). This indicates that no clinical studies, including MRMC comparative effectiveness studies involving human readers or AI assistance, were performed or submitted. The entire submission focuses on non-clinical performance and substantial equivalence based on technical characteristics and adherence to standards.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Since the devices are breathing circuit filters and not software or AI-powered devices, the concept of a "standalone (algorithm only)" performance study is not applicable. The device's performance is intrinsically mechanical and material-based, not algorithmic.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this submission is based on established international and national consensus standards (e.g., ISO 9360-1, ISO 5356-1, ISO 80369-7, IEC 60601-1, ISTA 3A, ISO 23328-1, ASTM F2101, ISO 10993, ISO 18562-1). The device's performance is measured against the quantitative requirements and methodologies specified within these standards.

    8. The sample size for the training set

    This question is not applicable. The document describes physical medical devices (filters), not machine learning or AI models that require a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for these devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151498

    Validate with FDA (Live)

    Date Cleared
    2015-12-18

    (197 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K013122

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired and to add, maintain, retain moisture for the exhaled breathe of the patient. Use up to 24 hours.

    The Models of the filters are A501-1, A501-2, A501-3, and A501-4. The filters are intended for patient population that weighs greater than 10Kgs.

    Device Description

    Breathing circuit bacterial/viral filter has into two models: - 1) Bacterial/ viral filters; 2) Bacterial/viral filters HME. Breathing circuit bacterial/viral filters are available in multiple sizes and shapes, rectangular and round, and incorporate standard 15/22 mm connectors with a gas sampling luer port. The depth filter use electrostatic media for filtration and a foam media for the HME media.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: a Breathing Circuit Bacterial/Viral Filter manufactured by Zhejiang Haisheng Medical Device Co., Ltd. The document asserts that the device is substantially equivalent to a legally marketed predicate device (EMS Electra Filter and Filter/HME, K013122).

    Based on the provided text, here's information regarding the acceptance criteria and the study that proves the device meets those criteria:


    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is compared directly to the predicate device, implying that the predicate's performance metrics serve as the acceptance criteria for achieving substantial equivalence.

    Acceptance Criteria (from Predicate)Reported Device Performance (Zhejiang Haisheng Breathing Circuit Bacterial/Viral Filter)Met/Not Met
    Resistance to Flow:
    <3cm H2O@60Lpm - Adult<3cm H2O@60Lpm - AdultMet
    <1cm H2O@20Lpm - Pediatric1cm H2O@20Lpm - PediatricMet
    Humidification Output (mg H2O/L):
    32mg H2O/L at TV of 1000cc - Adult32mg H2O/L at TV of 1000cc - AdultMet
    32mg H2O/L at TV of 250cc - Pediatric32mg H2O/L at TV of 250cc - PediatricMet
    Bacterial Filtration Test:99.999%Met
    99.999%99.99%Met
    Viral Filtration Test:99.99%Met
    Ethylene Oxide (EO) Residual:<10µg/gMet
    <10µg/g3 yearsMet
    Shelf Life:3 yearsMet
    Standards Conformance:
    ISO 5356-1YesMet
    ISO 594-2YesMet
    ISO 9360YesMet

    Additionally, other characteristics are compared for substantial equivalence:

    • Indications For Use: Same
    • Product Structure: Same (top cover, filter media, lower cover, sampling port)
    • Materials (Top Cover, Filter Media, Lower Cover, Gas Sampling Port, Package): Same (polystyrene, electrostatic polypropylene, PE, dialyzing paper and PE film)
    • Color Additive: None (Same)
    • Sterilization: EO (Same)

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions that "Test data and reports were provided as part of the 510(k) submission," but does not specify the sample sizes of devices tested for each performance metric (e.g., how many filters were tested for bacterial filtration efficiency).

    The data provenance is not explicitly stated in terms of country of origin for the testing, nor is it classified as retrospective or prospective. It is implied that the testing was conducted specifically for this 510(k) submission, suggesting it was prospective testing on the Zhejiang Haisheng device. The document states that some tests were performed by "Nelson Lab," which is a known independent laboratory (likely Nelson Laboratories, LLC in the USA), but this is not explicitly stated as the sole source of all test data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This type of information (number and qualifications of experts) is not applicable to this device and submission. The device is a physical breathing circuit filter, and its performance (e.g., filtration efficiency, resistance) is measured objectively through standardized laboratory tests, not through expert interpretation or clinical judgment that would require establishing a "ground truth" by experts.


    4. Adjudication Method for the Test Set

    This is not applicable. As mentioned above, the performance is determined by objective laboratory measurements and does not involve human interpretation or adjudication in the context of, for example, image analysis or diagnostic tasks.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    This is not applicable as the device is a physical medical device (a filter) and not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study and effects of AI assistance are not relevant.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable as the device is a physical filter, not an algorithm or AI system.


    7. The Type of Ground Truth Used

    The "ground truth" for verifying the device's performance is established through objective laboratory testing against recognized international standards and specific performance thresholds. For example:

    • Bacterial/Viral Filtration: Determined by laboratory tests (e.g., Nelson Lab) measuring the percentage removal of specific microorganisms.
    • Resistance to Flow: Measured using flow meters and pressure transducers.
    • Humidification Output: Measured in a laboratory setting under controlled conditions (e.g., per ISO 9360).
    • Material Compatibility/Biocompatibility: Determined by standardized material testing.
    • Sterilization: Verified by ethylene oxide residual testing.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical product, not a machine learning model, so there is no "training set" in the context of artificial intelligence or algorithm development. Production samples are tested for performance.


    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033008

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-13

    (230 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K013089, K013122, K002201, K964540, K984379

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Anesthesia / Respiratory Filters
    For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired.
    Model – HEPA filter 3000/04 – Single patient use for exhalation limb of circuit on NPB 700 series ventilators
    Models – HEPA filters - 6500/01, 6888/01, 8222/01, 8444/01 Model – Non-HEPA filters - 4000/01 Single patient use up to 24 hours. Patient tidal volumes > 150 ml, when applicable.

    Device Description

    The Air Safety HEPA and Non-HEPA filters are available in multiple sizes and shapes, and incorporate standard 15 / 22 mm connectors with or without a gas sampling luer port. Some models adapt to fit ventilator exhalation limb only. The depth (HEPA filtration) filter uses a pleated paper fiber for filtration. Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance. The "HEPA" performance was also tested in accordance to DOE-3025-99, DOE-3020-97 and ASTM D2986 - DOP. The electrostatic (non-HEPA filtration) filters are tested by Nelson Laboratories for BFE and VFE.

    AI/ML Overview

    This document describes the safety and effectiveness of Air Safety HEPA and Non-HEPA Filters. However, it does not fit the typical format of a study designed to prove a device meets acceptance criteria, especially in the context of an AI/ML powered device. This is a 510(k) premarket notification for traditional medical devices (filters), and therefore, many of the requested categories are not applicable.

    Here's an attempt to answer the questions based on the provided text, with clear indications where the information is not available or not relevant to this type of device submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present "acceptance criteria" in a pass/fail format with direct device performance comparison in a single table, as might be found in a study for an AI/ML device. Instead, it lists technical characteristics and performance standards that the device meets.

    However, based on the General Technical Characteristics table, we can infer performance criteria and compare them to the device's reported performance:

    Attribute/Performance StandardAcceptance Criteria (Inferred from Predicate/Standards)Reported Device Performance (Air Safety Filters)
    Dead Space (ml)Comparable to predicate / within acceptable range for intended use45 to 84 ml (various models), 209 mm for Model 3000/04
    Resistance to flow< 3.4 cm H2O @ 60 Lpm (Likely a standard for breathing circuits)< 3.4 cm H2O @ 60 Lpm
    HEPA Filtration Efficiency (for listed HEPA models)Specified by standards (e.g., DOE-3025-99, DOE-3020-97, ASTM D2986 - DOP for > 99.97% of 0.3 micron DOP particle @ 60 Lpm)99.99999% (Models 6500/01, 6888/01, 8222/01, 8444/01) Model 3000/04 - 99.9999%
    Bacterial Filtration Efficiency (BFE) - Nelson (for listed HEPA models)Standard requirement for bacterial filters (implied by predicate and testing)99.99975% (Models 6500/01, 6888/01, 8222/01, 8444/01) Model 3000/04 - 99.9999%
    Viral Filtration Efficiency (VFE) - Nelson (for listed HEPA models)Standard requirement for viral filters (implied by predicate and testing)99.99996% (Models 6500/01, 6888/01, 8222/01, 8444/01)
    Bacterial Filtration Efficiency (BFE) - Nelson (for Non-HEPA Model 4000/01)Standard requirement for bacterial filters (implied)99.99996%
    Viral Filtration Efficiency (VFE) - Nelson (for Non-HEPA Model 4000/01)Standard requirement for viral filters (implied)99.99925%
    Filtration Performance (General)According to EN 13328 Salt for Breathing System filtration performance."Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance." (Implied compliance)
    DOP Test Performance (for HEPA)> 99.97% of 0.3 micron DOP particle @ 60 Lpm> 99.97% of 0.3 micron DOP particle @ 60 Lpm
    Standard 15/22 mm connectorsRequired for compatibilityYes
    Gas sampling portOptionalOptional
    Single patient use up to 24 hoursYesYes
    ISO 5356-1 Conical 15/22Compliance with standardYes
    ISO 594-2 Luer FittingsCompliance with standardYes

    Summary of "Study" to Prove Device Meets Criteria:

    The document details the device characteristics and states that the devices were tested to specific performance standards. This is not a "study" in the clinical trial sense but rather a series of engineering and laboratory performance tests.

    • Testing Methodologies:
      • Filtration Performance: Tested according to EN 13328 Salt for Breathing System filtration performance.
      • HEPA Performance (Specific models): Tested in accordance with DOE-3025-99, DOE-3020-97, and ASTM D2986 - DOP.
      • Electrostatic (non-HEPA) Filtration (BFE and VFE): Tested by Nelson Laboratories.
    • Conclusion: The submission concludes that "The data within the submission demonstrates that the proposed devices when compared to the legally marketed predicate devices are safe and effective and substantially equivalent." This implies that the performance characteristics met or exceeded those of the predicate devices and applicable standards.

    Information Not Applicable / Not Provided for this Device Type:

    The following requested information is primarily relevant for AI/ML-powered diagnostic or predictive devices, or those requiring clinical studies with human assessors. The provided document concerns a physical filter, for which these criteria are generally not applicable or not detailed in this type of 510(k) submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable / Not provided in this context. For physical filters, testing typically involves laboratory samples of the filters themselves, not patient data in the sense of AI/ML. The number of filter units tested for each specific performance claim (e.g., BFE, VFE, resistance) is not specified. Data provenance like "country of origin" for test data is not a typical requirement for device performance testing in this manner.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. "Ground truth" established by experts (like radiologists for an image analysis AI) is not relevant for a breathing circuit filter. Performance is measured objectively through standardized laboratory tests (e.g., particle counts, pressure measurements).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are used to resolve discrepancies in expert opinions, typically for establishing ground truth in clinical datasets. This is not relevant for the objective performance testing of a physical filter.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a physical filter, not an AI/ML-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical filter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable. The "ground truth" for a filter's performance is derived from standardized physical and biological challenge tests (e.g., aerosolized particles, bacterial/viral suspensions) and measurements, not clinical expert consensus or pathology.

    8. The sample size for the training set:

    • Not Applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1