K Number
K221836

Validate with FDA (Live)

Date Cleared
2022-12-07

(167 days)

Product Code
Regulation Number
868.5260
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Filter CareStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in pediatric (with a tidal volume between 100 and adult patients, depending on the respective device. Filter SafeStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult patients. Filter/HME TwinStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and for respiratory use, as well as heat and moisture exchanger for humidifying respired gases for the patient. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult, pediatric and neonatal patients, depending on the respective device.

Device Description

The devices are breathing circuit filters used to filter the inhaled and/or the exhaled air of the patient against microbiological and particulate matter from the gases in the breathing circuit. They enclose a filter material in a housing that fits to standard breathing system connectors. Additionally, there are breathing system filters combined with a foam to function as HME (Heat and Moisture Exchangers) for passively humidifying the inspired air. The portfolio contains the following types of breathing circuit filters: - Filter CareStar Plus are electrostatic filters for use against contamination with microorganisms - . Filter SafeStar Plus are mechanical filters for use against contamination with microorganisms - Filter/HME TwinStar Plus are filters for use against contamination with microorganisms and for passive humidification of breathing gases

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The provided document describes three devices: Filter CareStar Plus, Filter SafeStar Plus, and Filter/HME TwinStar Plus. The table below compiles the acceptance criteria and reported performance for these devices based on the "Summary of non-clinical testing" section (pages 18-19).

Test Method & PurposeAcceptance CriteriaReported Performance
ISO 9360-1:2000 - Determination and Evaluation of Pneumatic ComplianceCompliance is less than or equal to 1mL/kPa at 15, 30, 60, and 70 hPaPASSED
ISO 9360-1:2000 - Determination and Evaluation of Pneumatic LeakagePneumatic leakage is less than or equal to 50mL/min at 70hPa.PASSED
ISO 9360-1:2000 - Determination and Evaluation of Pressure Drop (Pneumatic Resistance)Pneumatic resistance/pressure drop is acc. to IfU valuePASSED
ISO 5356-1:2015 - Evaluation of Conical Connectors (ISO 5356-1)Cone dimensions comply with ISO 5356-1.PASSED
ISO 80369-7:2021 - Luer Lock Connector (ISO 80369-7)Luer-Lock connector fulfills the requirements laid out in ISO 80369-7.PASSED
IEC 60601-1:2005 - Product's Ability to Withstand Damage from DroppingWhen dropped, the product should not suffer any damage which influences its function.PASSED
ISTA 3A - Product Durability During Transport, Mechanical Aspects- The packaging shows no or minor damage- The DUT (Device Under Test) shows no signs of damage and retains functionality after simulated transport.PASSED
ISO 23328-1:2003 - Filtration Efficiency (Particulate Matter) incl. Usage Time- Adult and pediatric electrostatic filters achieve a filtration efficiency of >90% before and after the specified usage time- Neonatal filters achieve a filtration efficiency of >75% before and after the specified usage time- Mechanical filters achieve a filtration efficiency of >99% before and after the specified usage time- Mechanical filters achieve a HEPA classification ≥ class H13PASSED
ASTM F2101:2019 - Filtration Efficiency (Viral and Bacterial)- Electrostatic filters achieve 99.99% (bacterial) and 99.9% (viral) filtration efficiency- Electrostatic filters for neo applications achieve 99.98% (bacterial) and 99.9% (viral) filtration efficiency.PASSED
ISO 10993:2018 and ISO 18562-1:2017 - Evaluation of Product's Biological CompatibilityEvaluation according to ISO 10993:2018 and/or ISO 18562-1:2017PASSED
ISO 9360-1:2000 - Evaluation of HME Water Loss, Resistance· pediatric/neonatal: The moisture loss shall be <=11mg/L at VT=50ml· pediatric: The moisture loss shall be <=12mg/L at VT=250ml· adult: The moisture loss shall be <=11mg/L at VT=500mlPASSED

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes used for each non-clinical test. It mentions that "The devices... have undergone extensive testing" but does not detail the number of units tested for each criterion. The data provenance is also not specified; however, given that the submitter is "Drägerwerk AG & Co. KGaA" based in "Lübeck, Germany" and they tested against international standards (ISO, ASTM, IEC), it is likely that the testing was conducted in a controlled laboratory environment, potentially in Germany or by a certified testing facility adhering to these standards. The testing appears to be prospective, specifically designed to verify the devices' adherence to the mentioned standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not mention the use of experts to establish ground truth for the non-clinical test set. The validation relies on adherence to established international and national standards (e.g., ISO, ASTM, IEC) rather than expert consensus on a test set. This implies that the "ground truth" is defined by the objective metrics and methodologies outlined in these standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

There is no mention of an adjudication method in the text for the non-clinical test set. Given the nature of the tests (physical and performance characteristics against standards), it's a pass/fail determination based on quantitative measurements against predefined criteria, not a subjective assessment requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document explicitly states "Summary of Clinical Testing N/A" (page 20). This indicates that no clinical studies, including MRMC comparative effectiveness studies involving human readers or AI assistance, were performed or submitted. The entire submission focuses on non-clinical performance and substantial equivalence based on technical characteristics and adherence to standards.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Since the devices are breathing circuit filters and not software or AI-powered devices, the concept of a "standalone (algorithm only)" performance study is not applicable. The device's performance is intrinsically mechanical and material-based, not algorithmic.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for this submission is based on established international and national consensus standards (e.g., ISO 9360-1, ISO 5356-1, ISO 80369-7, IEC 60601-1, ISTA 3A, ISO 23328-1, ASTM F2101, ISO 10993, ISO 18562-1). The device's performance is measured against the quantitative requirements and methodologies specified within these standards.

8. The sample size for the training set

This question is not applicable. The document describes physical medical devices (filters), not machine learning or AI models that require a "training set."

9. How the ground truth for the training set was established

This question is not applicable, as there is no training set for these devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 7, 2022

Drägerwerk AG & Co. KGaA Luise Lang Regulatory Affairs Manager Moislinger Allee 53-55 Lüebeck, Schleswig-Holstein 23542 Germany

Re: K221836

Trade/Device Name: Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar Plus Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: November 4, 2022 Received: November 4, 2022

Dear Luise Lang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K221836

Device Name

Filter CareStar Plus / Filter SafeStar Plus / Filter/HME TwinStar Plus

Indications for Use (Describe) Filter CareStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines.

They are intended for use in pediatric (with a tidal volume between 100 and adult patients, depending on the respective device.

Filter SafeStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult patients.

Filter/HME TwinStar Plus

Intended use

Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and for respiratory use, as well as heat and moisture exchanger for humidifying respired gases for the patient.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines.

They are intended for use in adult, pediatric and neonatal patients, depending on the respective device.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in Dräger has two dots above it. The word is the logo for the Dräger company, which specializes in medical and safety technology.

Traditional 510(k) - K221836 510(k) Summary

510(k) Premarket Notification Summary
Submitter:Drägerwerk AG & Co. KGaAMoislinger Allee 53-5523542 Lübeck, GermanyEstablishment's registration number: 9611500
Contact Person:Dr. Bettina MoebiusHead of Regulatory Affairs CentralE-Mail: Bettina.moebius@draeger.comTelephone: +49 451 882 4100
Applicant's US Contact Person:Tom HirteHead of Regulatory AffairsE-Mail: tom.hirte@draeger.comTelephone: (978) 3796461
Date prepared:December 6, 2022
Device Name:Trade Name:Filter CareStar Plus,Filter SafeStar Plus,Filter/HME TwinStar Plus
Classification Name:Regulation Number:Product Code:Class:Filter, Bacterial,Breathing Circuit21 CFR § 868.5260CAHII

Predicate Device:

The Filter CareStar Plus is substantially equivalent to the Bact-Trap from Pharma Systems AB (K202459).

The Filter SafeStar Plus is substantially equivalent to the HepaShield Breathing System Filter from Flexicare Medical Limited (K191909)

The Filter/HME TwinStar Plus is substantially equivalent to VR Medical Heat and Moisture Exchanger Filter from VR Medical Technology Co. (K132709).

Reference Device (for Filter/HME TwinStar Plus):

EMS Electra Filter and Filter/HME (K013122)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is likely a logo for the Dräger company, which is a German company that makes medical and safety technology.

Traditional 510(k)

510(k) Summary

Device Description

The devices are breathing circuit filters used to filter the inhaled and/or the exhaled air of the patient against microbiological and particulate matter from the gases in the breathing circuit. They enclose a filter material in a housing that fits to standard breathing system connectors.

Additionally, there are breathing system filters combined with a foam to function as HME (Heat and Moisture Exchangers) for passively humidifying the inspired air.

The portfolio contains the following types of breathing circuit filters:

  • Filter CareStar Plus are electrostatic filters for use against contamination with microorganisms
  • . Filter SafeStar Plus are mechanical filters for use against contamination with microorganisms
  • Filter/HME TwinStar Plus are filters for use against contamination with microorganisms and for passive humidification of breathing gases

Indications for Use

Filter CareStar Plus

Intended use

Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines.

They are intended for use in pediatric (with a tidal volume between 100 and 500 ml) and adult patients, depending on the respective device.

Filter SafeStar Plus:

Intended use

Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult patients.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue. The word is easily readable and the image is clear.

Traditional 510(k)

510(k) Summary

Filter/HME TwinStar Plus:

Intended use

Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and for respiratory use, as well as heat and moisture exchanger for humidifying respired gases for the patient.

Indications

All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult, pediatric and neonatal patients, depending on the respective device.

{7}------------------------------------------------

Traditional 510(k) 510(k) Summary

Comparison to Predicate

Filter CareStar Plus

SpecificationProposed DevicePredicate DeviceComments
Filter CareStar PlusBact-Trap
ManufacturerDrägerwerk AG & Co. KGaAPharma Systems AB-
510(k) NumberK221836K202459-
Regulation Number868.5260 Breathing circuit bacterial filter868.5260 Breathing circuit bacterial filterSame
Product CodeCAHCAHSame
ClassificationClass IIClass IISame
Intended Use/Indications for UseBidirectionally breathing system filteragainst bacterial and viral contaminationfor anesthetic and respiratory use.All devices are intended for single use upto 24 hours and must be used by trainedmedical personnel only. The devices aredesigned for use with ventilators andanesthesia machines.They are intended for use in pediatric (witha tidal volume between 100 and 500 ml)and adult patients, depending on therespective device.Bact Trap filter is a breathing systemfilter which is designed to reducepossible airborne or liquid-borne crosscontamination with microrganisms andparticulate matter via anaesthetic orventilator breathing systems.The Bact Trap filter may either be usedon the patient side or on the device sideof the ventilator anaesthetic device.Similar
Patient PopulationAdult and pediatric patients with a tidalvolume from 100 to 1500 mlAdult patients with a tidal volume from 50to 1500 mlSimilar
Device configurationStraight with luer port--
MaterialsHousing: PolypropyleneFilter media: Polypropylene with syntheticfibersHousing: SBC, Polypropylene, PolyethyleneFilter media: Acrylic and Polypropylene fibersSimilar
Traditional 510(k)510(k) Summary
SterilityNon-sterileNon sterileSame
ApplicationDisposableup to 24 hoursDisposableup to 24 hoursSame
Biocompatibility testingISO 10993 and ISO 18562-1 compliantISO 10993 and ISO 18562-1 compliantSame
Filtration performanceISO 23328-1 compliant and ASTM F2101 compliantISO 23328-1 and ASTM F2101 compliantSame
Compliance, leakageand pressure dropISO 23328-2 and ISO 9360-1 compliantISO 23328-2 and ISO 9360-1 compliantSame
Technological Characteristics
Principle of OperationElectrostatic filtration methodElectrostatic filtration methodSame

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue and are slightly larger than the rest of the letters. The word is centered and takes up most of the image.

Traditional 510(k)

510(k) Summary

SpecificationProposed DeviceFilter CareStar PlusPredicate DeviceBact-TrapComments
ConnectorsStandard conical connectorsLuer lock port for gas samplingStandard conical connectorsLuer lock port for gas samplingSame
General Performance
Resistance@30 l/min: ≤ 1.3 mbar@30 l/min: = 1.1 mbarSimilar
Filtration efficiencyBFE: ≥ 99,99%VFE: ≥ 99,9 %BFE: > 99,999 %VFE: > 99,99 %Similar
Shelf-Life3 years3 yearsSame
Dead Space20 to 35 ml23 to 76 mlSimilar
Leakage at 70 mbar≤ 15 ml / minNot stated-
Compliance≤ 1 ml / kPaNot stated-
SpecificationProposed DevicePredicate DeviceComments
Filter SafeStar PlusHepaShield Breathing System Filter
ManufacturerDrägerwerk AG & Co. KGaAFlexicare Medical Limited
510(k) NumberK221836K191909
Regulation Number868.5260 Breathing circuit bacterial filter868.5260 Breathing circuit bacterial filterSame
Product CodeCAHCAHSame
ClassificationClass IIClass IISame
Intended Use /Indications for UseBidirectionally breathing system filteragainst bacterial and viral contaminationfor anesthetic and respiratory use.All devices are intended for single use upto 24 hours and must be used by trainedmedical personnel only. The devices aredesigned for use with ventilators andanesthesia machines.They are intended for use in adultpatients.Flexicare's HepaShield Bacterial ViralBreathing System Filters are intended toreduce the transmission of bacteria andviruses to/from a patient duringanesthesia. For use with ventilators,anesthesia machines and open flowsystems where filtration of inspiredand/or expired gases is desired.Flexicare's HepaShield Bacterial ViralBreathing System Filters are single usedevices for use on a single patient for upto 24hrs and are available in adult size.Flexicare's HepaShield Bacterial ViralBreathing System Filters are designed tobe used in hospital environments bytrained personnel.Similar
Patient PopulationAdult patients with a tidal volume from 100to 1500 mlAdult patients with a tidal volume of 141to 800 mlSimilar
Device configurationStraight and angular versions with luerportStraight with luer portSimilar

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The background is white.

Traditional 510(k)

510(k) Summary

Filter SafeStar Plus

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

Traditional 510(k)510(k) Summary
MaterialsHousing: PolypropyleneFilter media: Paper with glass fibersCasting compound: PolyolefinUnknown-
SterilityNon-sterileSterile and no-sterile variantssimilar
ApplicationDisposableUp to 24hDisposableUp to 24hSame
Biocompatibility testingISO 10993 and ISO 18562-1 compliantISO 10993 and ISO 18562-1 compliantSame
Filtration performanceISO 23328-1 compliant and ASTM F2101compliantISO 23328-1 compliant and ASTM F2101compliantSame
Compliance, Leakageand pressure dropISO 23328-2 and ISO 9360-1 compliantISO 23328-2 and ISO 9360-1 compliantSame
Technological Characteristics
Principle of OperationMechanical filtration methodMechanical filtration methodSame

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The logo is simple and modern.

Traditional 510(k)

510(k) Summary

SpecificationProposed DevicePredicate DeviceComments
Filter SafeStar PlusHepaShield Breathing System Filter
ConnectorsStandard conical connectorsLuer lock port for gas samplingStandard conical connectorsLuer lock port for gas samplingSame
General Performance
Resistance@30 l/min: ≤ 2 mbar@30 l/min: = 1.7 mbarSimilar
Filtration efficiencyBFE: ≥ 99,999 %VFE: ≥ 99,999 %BFE: 99,99999 %VFE: 99,9999 %Similar
Shelf-Life5 years5 yearsSame
Dead Space55 to 90 ml47 mlDifferent
Leakage at 70 mbar≤ 15 ml / min< 2ml / minDifferent
Compliance≤ 1 ml / kPa0.057 ml / cmH2OSimilar
SpecificationProposed DevicePredicate DeviceReference DeviceComments
Filter/HME TwinStar PlusVR Medical Heat and MoistureExchanger FilterEMS Electra Filter and Filter/HME
ManufacturerDrägerwerk AG & Co. KGaAVR Medical Technology Co.Engineered Medical Systems,Inc.
510(k) NumberK221836K132709K013122
Regulation Number868.5260 Breathing circuitbacterial filter868.5260 Breathing circuitbacterial filter868.5260 Breathing circuitbacterial filterSame
Product CodeCAHCAHCAHSame
ClassificationClass IIClass IIClass IISame
Intended Use /Indications for UseBidirectionally breathing systemfilter against bacterial and viralcontamination for anesthetic andfor respiratory use, as well as heatand moisture exchanger forhumidifying respired gases for thepatient.All devices are intended for singleuse up to 24 hours and must beused by trained medical personnelonly. The devices are designed foruse with ventilators andanesthesia machines.They are intended for use in adult,pediatric and neonatal patients,depending on the respectivedevice.The Heat and MoistureExchanger and Filter is adisposable single-use deviceindicated for patients whorequire humidification during thedelivery of ventilator gases andprovide filtration for reducingpossible cross contaminationbetween patient and equipment.The products mentioned aboveare designed for disposable useand should be changed at leastevery 24 hoursFor use with ventilators,anesthesia machines and openflow systems where filtration ofinspired and/or expired gases isdesired and to add maintain andretain moisture for the exhaledbreathe of the patientSimilar
SpecificationProposed DevicePredicate DeviceReference DeviceComments
Filter/HME TwinStar PlusVR Medical Heat and MoistureExchanger FilterEMS Electra Filter and Filter/HME
Patient PopulationNeonatal, pediatric, and adultpatients with a tidal volume from30 to 1500 ml.Adult patients with a tidal volumefrom 250 to 1500 mlNeonatal, pediatric and adultpatients with a tidal volume from20 cc to > 150 cc.Similar
Device configurationStraight and angular versions withluer port---
MaterialsHousing: PolypropyleneFilter media: Polypropylene withsynthetic fibers (except Filter/HMETwinStar HEPA Plus, which uses paperwith glass fibers)HME-Media: Polyurethaneimpregnated with calcium chlorideHousing: K-ResineFilter media: Polyolefin syntheticfibers (except VR005 which usespaper with glass fibers)HME-Media: Polyurethaneimpregnated with calcium chlorideHousing: PolystyreneFilter media: ElectrostaticpolypropyleneDifferent
SterilityNon-sterileNon-sterileUnknownSame
Traditional 510(k)510(k) Summary
ApplicationDisposableup to 24 hoursDisposableup to 24 hoursDisposableup to 24 hoursSame
BiocompatibilitytestingISO 10993 and ISO 18562-1compliantISO 10993 compliantUnknownSimilar
FiltrationperformanceISO 23328-1 compliant and ASTMF2101 compliantISO 23328-1 compliantUnknownSimilar
Compliance,Leakage andpressure dropISO 23328-2 and ISO 9360-1compliantUnknownUnknown-
Technological Characteristics
Principle ofOperationElectrostatic or mechanicalfiltration method combined with afoam for passive humidificationElectrostatic or mechanicalfiltration method combined with afoam for passive humidificationElectrostatic filtration methodcombined with a foam for passivehumidificationSame
Traditional 510(k)510(k) Summary
ConnectorsStandard conical connectorsLuer lock port for gas samplingStandard conical connectorsLuer lock port for gas samplingStandard conical connectorsLuer lock port for gas samplingSame
General Performance
Resistance
Adult@30 l/min: ≤ 1.6 mbar@30 l/min: ≤ 1.5 mbar@30 l/min: not statedSimilar
Neonatal /Pediatric@15 l/min: ≤ 1.5 mbar-@20 l/min: 1.0 cmH2O
Filtration efficiency
BFEVFE≥ 99,98 %≥ 99,9 %99.999 %99.99%≥ 99.999 %≥ 99.99 %Similar
Moisture loss / Moisture return¹
AdultMoisture loss @ Vt= 500 ml≤ 10.9 mg/l(which results in a moisture returnof ≥ 33,1 mg/l)Moisture loss @ Vt = 500 ml:≤ 10.5 mg/lMoisture return @ Vt = 1000 cc:= 32 mg/lSimilar
PediatricMoisture loss @ Vt = 250 ml:≤ 11.8 mg/l(which results in a moisture returnof ≥ 32,2 mg/l)-Moisture return@ Vt = 250 cc:= 32 mg/lSimilar
NeonatalMoisture loss @ Vt=50ml:≤ 10.3 mg/l(which results in a moisture returnof ≥ 33.7 mg/l)-Moisture return @ Vt= 50 ml:= 30 mg/lSimilar
Shelf-Life3 or 5 years, depending on therespective deviceNot statedNot stated-
Dead space9 to 90 ml60 to 85 ml10 to 65 mlSimilar
Leakage at 70 mbar≤ 15ml / min---
Compliance≤ 1ml / kPa---

{13}------------------------------------------------

Traditional 510(k) 510(k) Summary

Filter/HME TwinStar Plus

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is likely a logo for the Dräger company, which is a German company that makes medical and safety technology.

Traditional 510(k)

510(k) Summary

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "Dräger" in blue font. The word is written in a bold, sans-serif font. The two dots above the "a" are also blue. The background is white.

{17}------------------------------------------------

Traditional 510(k) 510(k) Summary

Image /page/17/Picture/5 description: The image shows the word "Dräger" in blue font. The "ä" in Dräger has two dots above it. The font is bold and sans-serif. The background is white.

1 Moisture return = 44 mg/l – moisture loss

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

Traditional 510(k) 510(k) Summary

Discussion of Non-clinical Testing

The devices Filter CareStar Plus, Filter SafeStar Plus and Filter/HME TwinStar Plus are new products and have undergone extensive testing to qualify it with e.g. national and international consensus standards, technical system requirements and other requirements. The following identified verification and validation activities necessary to establish substantial equivalence to the predicate device were carried out and wellestablished method, the evidence is included in this 510(k) submission.

  • Biocompatibility ●
  • Technical system requirements, including
    • o Technical data
    • o Essential safety and performance
  • Connectors .
  • Human factors engineering .

Summary of non-clinical testing

Test methodPurposeAcceptance criteriaResult
ISO 9360-1:2000Determination andEvaluation of PneumaticCompliancecompliance is less than orequal to 1mL/kPa at 15,30, 60, and 70 hPaPASSED
ISO 9360-1:2000Determination andEvaluation of PneumaticLeakagepneumatic leakage is lessthan or equal to 50mL/minat 70hPa.PASSED
ISO 9360-1:2000Determination andEvaluation of PressureDrop (PneumaticResistance)pneumaticresistance/pressure drop isacc. to IfU valuePASSED
ISO 5356-1:2015Evaluation of ConicalConnectors (ISO 5356-1)cone dimensions complywith ISO 5356-1.PASSED
ISO 80369-7:2021Luer Lock Connector(ISO 80369-7)Luer-Lock connector fulfillsthe requirements laid outin ISO 80369-7.PASSED
IEC 60601-1:2005Product's Ability toWithstand Damage fromDroppingWhen dropped, theproduct should not sufferany damage whichinfluences its function.PASSED
ISTA 3AProduct Durability DuringTransport, MechanicalAspects- the packaging shows noor minor damage- the DUT shows no signsof damage and retainsfunctionality aftersimulated transport.PASSED
ISO 23328-1:2003Filtration Efficiency(Particulate Matter) incl.Usage Time- adult and pediatricelectrostatic filters achievea filtration efficiency of>90% before and after thespecified usage time- neonatal filters achieve afiltration efficiencyof >75% before and afterthe specified usaqe time- mechanical filtersachieve a filtrationefficiency of >99% beforeand after the specifiedusage time- mechanical filtersachieve a HEPAclassification ≥ class H13PASSED
ASTM F2101:2019Filtration Efficiency (Viraland Bacterial)- electrostatic filtersachieve 99.99% (bacterial)and 99.9% (viral) filtrationefficiency- electrostatic filters for neoapplications achieve99.98% (bacterial) and99.9% (viral) filtrationefficiency.PASSED
ISO 10993:2018and ISO 18562-1:2017Evaluation of Product'sBiological Compatibility(ISO 10993:2018 andISO 18562-1:2017)evaluation accordingto ISO 10993:2018 and/orISO 18562-1:2017PASSED
ISO 9360-1:2000Evaluation of HME WaterLoss, Resistance·pediatric/neonatal: Themoisture loss shall be<=11mg/L at VT=50ml·pediatric: The moistureloss shall be <=12mg/L atVT=250ml· adult: The moisture lossshall be <=11mg/L atVT=500mlPASSED

{19}------------------------------------------------

{20}------------------------------------------------

Standard Number and VersionTitle
AAMI ANSI ISO 5356-1:2004Anaesthetic And Respiratory Equipment – Conical Connectors: Part 1: Cones And Sockets
ISO 18652-1:2017Biocompatibility Evaluation Of Breathing Gas Pathways In Healthcare Applications
AAMI / ANSI / IEC 62366-1:2015Medical devices - Part 1: Application of usability engineering to medical devices
ANSI AAMI ISO 10993-1:2018Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ASTM F2101-19:2019Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials
EN ISO 9360-1:2009Anaesthetic and respiratory equipment - Heat and moisture exchangers (HMEs) for humidifying respired gases in humans - Part 1: HMEs for use with minimum tidal volumes of 250 ml
EN ISO 23328-1:2008Breathing system filters for anaesthetic and respiratory use - Part 1: Salt test method to assess filtration performance
EN ISO 23328-2:2009Breathing system filters for anaesthetic and respiratory use - Part 2: Non-filtration aspects

Summary of Clinical Testing

N/A..

Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the Filter CareStar Plus, Filter SafeStar Plus and Filter/HME TwinStar Plus are as safe, as effective and perform as well as or better than the legally marketed devices identified in this section.

  • END -

§ 868.5260 Breathing circuit bacterial filter.

(a)
Identification. A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.(b)
Classification. Class II (performance standards).