Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K040830

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-06-18

    (79 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multi-Track Angiographic Catheter (K952984, K003902) - Recommended for use in Multi-Track Anglography of cardiovascular vessels and/or chambers. It can be used for catheterization for anglogiap. In and pressure measurement in any chamber or vessel.

    Ghost II PTA (K003972, K011557) - This catheter is recommended for Percutaneous Gransbuminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Z-MED Catheter (K991977, K003114, K003643, K011557) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that 사 does not require surgical intervention. And;

    (K931009, K011557) This catheter is recommended for Percutaneous Transluminal Angioplasty (ISSIN67) Internoral, ilias, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Tyshak Mini Pediatric PTV Catheter (K003276, K011557, K032591) - This catheter is I your recuration for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve in Pediatric applications.

    • A patient with isolated pulmonary stenosis. u
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that l does not require surgical intervention.

    Tyshak Catheter (K991977, K003114, K011557) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis u
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that ■ does not require surgical intervention. And;

    (K931009, K011557) This catheter is recommended for Percutaneous Transluminal Angioplasty (RSS1007, Norths of I'ms cancenal arteries. These catheters are not designed to be used in the coronary arteries.

    Tyshak II Catheter (K003052, K011557, K030589) - This catheter is recommended for I Yshak II Catherer in Catherer (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis i
    • A patient with isolated pulmonary stenosis
      A patient with valvular pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention. And;

    Z-MED II Catheter (K003052, K011557, K030589) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with isolated pulmonary stenosis with other minor congenital heart disease that l does not require surgical intervention. And;

    High Pressure PTA (Marauder) (K010880, K011557) - This catheter is recommended for right Fressure I 17x (maraduct) (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Z-5 Atrioseptostomy (K960070, K011557) - Used for the palliation of several congenital heart 2-5 All loseptosition of the great arteries, total anomalous pulmonary venous drainage without delects. transposition of the Ereal arcess, mitral stenosis, mitral atresia, and pulmonary atresia with intact ventricular septum.

    Z-5 Braided Atrioseptostomy (K001804, K011557) - Used for the palliation of several 2-3 Draided Attrioseptostom. (2001 of the great arteries, total anomalous pulmonary venous congential licar delected transportion, tricuspid atresia, mitral stenosis, mitral atressa, and pulmonary atresia with intact ventricular septum.

    Tyshak X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention.

    Tyshak II X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that 1 does not require surgical intervention.

    Z-MED X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis a
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that . does not require surgical intervention.

    Z ·MED II X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that . does not require surgical intervention.

    COEfficient Catheter (K014124) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis ﺎ
    • A patient with isolated pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention.

    Ghost PTA (K931009, K011557) - This catheter is recommended for Percutaneous Gross ITTA (KSJ1007) RW115872 - The femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Mullins PTA (K013601) - This catheter is recommended for Percutaneous Transluminal Millins ITA (INVL001) - This Satileter and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Mullins X PTA (K022722)_- This catheter is recommended for Percutaneous Transluminal Mullins A F FA (XN22122) - Frid Sains and renal arteries. These catheters are not designed to be used in the coronary arteries.

    PTS (K003320, K011557) - For use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) summary for NuMED Catheters, primarily focusing on extending the shelf life of their existing devices from 3 to 5 years, and outlining their various intended uses. It does not contain information about acceptance criteria or a study proving device performance in the context of diagnostic or therapeutic efficacy.

    The document details:

    • K040830 - All NuMED Catheters: A submission for various catheters (angiographic, PTA, PTV, atrioseptostomy, etc.).
    • Purpose of Submission: To extend the shelf life validation from 3 to 5 years, indicating a focus on stability rather than primary diagnostic or therapeutic performance.
    • Predicate Devices: Numerous existing NuMED catheters.
    • Regulatory Information: Class II devices, product codes, and relevant CFR sections.
    • Indications for Use: Detailed indications for each specific catheter type (e.g., Multi-Track Angiographic Catheter for angiography, Ghost II PTA for Percutaneous Transluminal Angioplasty of femoral, iliac, and renal arteries, Z-MED Catheter for Percutaneous Transluminal Valvuloplasty of the pulmonary valve).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria for diagnostic or therapeutic performance because this information is not present in the provided text. The document is a regulatory submission for shelf-life extension and outlines intended uses, not a clinical performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013601

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-01-03

    (64 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K931009, K991977, K011557

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Device Description

    The NuMED, Inc. MULLINS™ PTA Catheter is a co-axial over-the-wire catheter with a balloon near the distal tip. It is available in sizes of 12-20mm diameter and 3-4cm in length. One lumen permits guidewire insertion to facilitate advancement of the catheter, while the other lumen is for balloon inflation and deflation. The balloons of the NuMED, Inc. MULLINS™ PTA Catheter is made of a noncompliant polymeric material and are the same balloons that are used on the approved Z-MED Catheter (K931009, K991977, and K011557). The two laminate balloon system is designed to inflate to a specific diameter at a given pressure. The change in diameter is minimal over the range of inflation pressures. The balloons are heat bonded to the shaft. The outer body and inner body tubing are made of polymeric tubing. The area under the balloon is enhanced with five radiopaque platinum image bands. Two that are 5mm on each side of the balloon center and two more under the balloon shoulders. An additional image band is imbedded into the tip of the catheter as an additional safety measure.

    AI/ML Overview

    This document is a 510(k) premarket notification for the NuMED Mullins PTA Catheter. It claims substantial equivalence to the NuMED Z-MED PTA Catheter. The information provided is primarily focused on the device description, biocompatibility, and bench testing, rather than explicit acceptance criteria and a detailed study report with performance metrics.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the document, direct, quantifiable acceptance criteria with corresponding performance results are not explicitly stated in a table format. The submission relies on demonstrating substantial equivalence to a predicate device by comparing design, materials, and intended use. The "acceptance criteria" can be inferred as meeting the standards and performance of the predicate device and relevant biocompatibility/bench testing.

    Feature/TestAcceptance Criteria (Inferred)Reported Device Performance
    Intended UseSame as predicate (PTA of femoral, iliac, renal arteries)Same as predicate (PTA of femoral, iliac, renal arteries)
    BiocompatibilityCompliance with Tripartite Biocompatibility Guidance for Medical Devices"Test results indicate that all materials demonstrate the biocompatibility of the NuMED catheter and are on file at NuMED, Inc."
    Laboratory (Bench) TestingPerformed in accordance with GMP's"All bench testing was performed in accordance with GMP's and the results are kept on file at NuMED, Inc."
    MaterialsSame as predicate (Shaft: PES3, Balloon: PES2, Image Band: Platinum)Same as predicate (Shaft: PES3, Balloon: PES2, Image Band: Platinum)
    ConstructionSame as predicate (Coaxial, distally mounted non-compliant balloons)Same as predicate (Coaxial, distally mounted non-compliant balloons)
    Balloon Performance (Diameter/Pressure)Maintain specific diameter at given pressure with minimal change"The two laminate balloon system is designed to inflate to a specific diameter at a given pressure. The change in diameter is minimal over the range of inflation pressures." (This is a design characteristic, not a direct test result proving it met a specific deviation tolerance)

    Missing specific quantifiable acceptance criteria and performance data. The document states that test results are "on file at NuMED, Inc.", indicating that the detailed data was submitted to the FDA but is not included in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The document only mentions "Biocompatibility Testing" and "Laboratory (Bench) Testing" without indicating the number of devices or components tested.
    • Data Provenance: The tests were conducted internally by NuMED, Inc. ("on file at NuMED, Inc."). The document does not specify the country of origin of the data beyond "NuMED, Inc.", which is based in Hopkinton, NY, USA. The testing appears to be retrospective in the sense that the results were submitted for a premarket notification for a device that has undergone testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The device is a medical catheter, and its evaluation relies on physical and material properties, mechanical testing, and biocompatibility, not on expert interpretation of medical images or clinical data to establish a ground truth for a diagnostic output.

    4. Adjudication Method for the Test Set

    This is not applicable for the reasons stated above. There's no "ground truth" established by experts or adjudication involved in the type of testing described (biocompatibility, bench testing).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. The device is a medical catheter, not an AI-powered diagnostic tool requiring human readability or interpretation. Therefore, there's no comparison of human readers with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable. The device is a physical medical instrument, not an algorithm or AI system.

    7. Type of Ground Truth Used

    The "ground truth" for this device's safety and effectiveness relies on:

    • Biocompatibility Standards: Compliance with the "Tripartite Biocompatibility Guidance for Medical Devices."
    • Engineering Standards: Meeting the device's design specifications for dimensions, material integrity, and functionality (e.g., balloon inflation properties).
    • Good Manufacturing Practices (GMPs): Adherence to quality system requirements during testing.

    There is no "ground truth" derived from expert consensus, pathology, or outcomes data in the context of diagnostic interpretation, as this is an interventional device.

    8. Sample Size for the Training Set

    This is not applicable. The device is a physical medical instrument, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1