510(k) Data Aggregation

    K Number
    K252206

    Validate with FDA (Live)

    Date Cleared
    2026-01-09

    (179 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nova Allegro urine albumin creatinine ratio (UACR) assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy.

    The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.

    Device Description

    Nova Allegro UACR Assay

    The Nova Allegro UACR Assay is a completely automated assay that functions with the Nova Allegro Analyzer for the determination of albumin and creatinine and the calculation of UACR in human urine. The Nova Allegro UACR Assays are performed sequentially using a single Allegro UACR test cartridge. The Allegro UACR Test Cartridge contains all of the reagents for measuring albumin and creatinine. Allegro UACR Controls are used to monitor the performance of the Nova Allegro UACR Test Cartridge and Allegro analyzer.

    Nova Allegro Analyzer

    The Nova Allegro Analyzer is a compact, point-of-care analyzer that features a clinically important menu of measured and calculated tests. All tests are measured with disposable, ready-to-use cartridges. The analyzer supports multiple wavelengths that are used to measure the assay of interest. The analyzer consists of the following key systems/components that the user interacts with:

    • Two analytical bays where the single use test cartridges are analyzed
    • Color Touchscreen Display
    • Barcode Scanner
    • Printer
    • Data Export Options
    • Ethernet Connection
    • USB Port
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251058

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-26

    (236 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Creatinine (CRE) and Blood Urea Nitrogen (BUN) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the CRE determination, Estimated Glomerular Filtration Rate (eGFR) is calculated by the analyzer.

    The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in diagnosing the following disorders in adults 18 years of age or older:

    • Creatinine (CRE): Renal disease and monitoring of renal dialysis.
    • Blood Urea Nitrogen (BUN): Certain renal and metabolic diseases.

    The Tru Analyzer is an automated, multi-assay integrated system, containing an absorbance reader, a bead scanner, and a cell imager for in vitro diagnostic use. This analyzer is intended for the quantitative determination of various analyte concentrations found in lithium-heparinized venous whole blood. It is for clinical laboratory or point-of-care use.

    Use only TruWellness Panel™ test kits with the Tru Analyzer.

    Device Description

    The TruSystem is an automated, integrated in vitro diagnostic platform consisting of the Tru Analyzer and the Tru Kidney Health Test Panel (part of the TruWellness Panel™), a Single-Use Consumable Kit that includes a Disc and a Support Pack. Designed for point-of-care and clinical laboratory use, the system enables the simultaneous measurement of clinical chemistry, immunoassay, and hematology parameters from a lithium-heparinized venous whole blood sample in a single run. The TruSystem delivers quantitative results for routine clinical chemistry and immunoassay analytes as well as a complete blood count (CBC) with a 3-part differential, all without the need for specialized operating skills, external calibration, or complex infrastructure.

    The Tru Analyzer is a benchtop instrument that fully automates sample processing, assay execution, and result reporting. Its touchscreen interface allows operators to control workflows, initiate tests, and review results with minimal training. The Single-Use Consumable Kit and blood sample are loaded into the analyzer drawer, where all necessary processing takes place. Internally, the Tru Analyzer integrates:

    • A high precision pipettor for automated sample and reagent handling,
    • An onboard centrifuge to separate whole blood into plasma,
    • A closed-loop thermal control system to maintain precise assay temperatures,
    • A motion control system to fully automate sample processing,
    • Dedicated detection modules for clinical chemistry, immunoassay, and hematology testing,
    • A high-definition camera used to collect assay readings, image capture, and instrument quality control checks,
    • An electronics board to manage individual module boards and associated firmware,
    • An integrated computer running the instrument software, accessible via the touchscreen user interface.

    The Tru Analyzer is factory-calibrated and continuously monitors its performance using optical sensing and electronic feedback mechanisms. Every time the analyzer is powered on or a Single-Use Consumable Kit (TruWellness Panel™) is loaded, the system runs an automated self-test to verify that it remains within calibration and is functioning properly. Internal quality control (QC) checks occur throughout the testing process to ensure the integrity of the analyzer, Disc, and Support Pack.

    If an internal QC check does not meet specifications, the instrument will automatically suppress result reporting to prevent the release of potentially inaccurate results. Any impacted tests will display associated codes in the footnotes of the patient report. The Tru Analyzer also applies result flags to highlight conditions that may indicate abnormalities in the patient sample or require attention. These flags appear in-line with impacted assay results for clear visibility. If a system operation failure occurs, the Tru Analyzer will halt the run, and no results will be generated. In these cases, the instrument displays an on-screen error message with a unique code and step-by-step instructions to guide resolution. These measures ensure the system provides accurate results, while requiring no special user knowledge, training, or on-site calibration.

    The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process and analyze samples on the Tru Analyzer, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards at the end of the run.

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) provides in vitro quantitative determination of Creatinine (CRE), Blood Urea Nitrogen (BUN), and calculated estimated Glomerular Filtration Rate (eGFR) in lithium-heparinized venous whole blood samples.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242685

    Validate with FDA (Live)

    Date Cleared
    2024-12-04

    (89 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica® CH Creatinine_3 (Crea3) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin, dipotassium EDTA, and sodium heparin), and urine using the Atellica® CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

    Device Description

    The Atellica CH Crea3 assay is based on the reaction of picrate with creatinine in an alkaline medium to produce a red chromophore creatinine picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Crea3 assay is a modification of the Jaffe method, using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because non-specific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 µmol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 µmol/L) from each result.

    AI/ML Overview

    The provided text describes the performance characteristics and studies for the Atellica® CH Creatinine_3 (Crea3) assay, a new in vitro diagnostic device for quantitative determination of creatinine. It compares this new device to a predicate device, the Atellica® CH Creatinine_2 (Crea_2) assay.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: The document focuses on establishing substantial equivalence for an in vitro diagnostic (IVD) test, which primarily relies on analytical performance characteristics rather than clinical outcome studies or multi-reader multi-case (MRMC) comparative effectiveness studies typically seen with imaging AI devices. Therefore, some of your requested information (like number of experts for ground truth, adjudication methods, MRMC studies, and training set details for an AI model) are not directly applicable or provided in this type of submission.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are established through various analytical performance studies, primarily comparing it to a legally marketed predicate device (Atellica® CH Creatinine_2). The acceptance criteria are implicitly defined by the successful demonstration of equivalence or meeting pre-defined performance goals for each characteristic.

    Here's a table summarizing the acceptance criteria (inferred from the "designed to have" or "determined in accordance with" statements and the reported results meeting these) and the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Atellica® CH Creatinine_3 (Crea3))
    Detection CapabilityLoB: $\le$ LoD for serum and urine samples. LoD: $\le$ 0.15 mg/dL for serum/plasma; $\le$ 3.00 mg/dL for urine. LoQ: $\le$ 0.15 mg/dL for serum/plasma with $\le$ 0.10 mg/dL total analytical error; $\le$ 3.00 mg/dL for urine with $\le$ 1.50 mg/dL total analytical error.Serum/plasma: LoB: 0.05 mg/dL LoD: 0.10 mg/dL LoQ: 0.15 mg/dL Urine: LoB: 0.50 mg/dL LoD: 1.00 mg/dL LoQ: 3.00 mg/dL (All results meet the stated design goals/acceptance criteria).
    PrecisionDetermined in accordance with CLSI Document EP05-A3 (indicates adherence to specific statistical targets for repeatability and within-lab precision, implicitly accepted if within CLSI guidelines for the assay's use).Serum Samples (n=80 each): - Serum 1 (0.38 mg/dL): Repeatability SD 0.006, CV 1.6%; Within-Lab SD 0.012, CV 3.2% - Serum 2 (0.73 mg/dL): Repeatability SD 0.023, CV 3.2%; Within-Lab SD 0.029, CV 4.0% - Serum 3 (0.73 mg/dL): Repeatability SD 0.006, CV 0.8%; Within-Lab SD 0.019, CV 2.6% - Serum 4 (1.18 mg/dL): Repeatability SD 0.007, CV 0.6%; Within-Lab SD 0.019, CV 1.6% - Serum QC 1 (1.85 mg/dL): Repeatability SD 0.007, CV 0.4%; Within-Lab SD 0.024, CV 1.3% - Serum QC 2 (6.21 mg/dL): Repeatability SD 0.011, CV 0.2%; Within-Lab SD 0.067, CV 1.1% - Serum 5 (17.39 mg/dL): Repeatability SD 0.035, CV 0.2%; Within-Lab SD 0.189, CV 1.1% - Serum 6 (28.54 mg/dL): Repeatability SD 0.056, CV 0.2%; Within-Lab SD 0.317, CV 1.1% Urine Samples (n=80 each): - Urine 1 (56.74 mg/dL): Repeatability SD 0.102, CV 0.2%; Within-Lab SD 0.746, CV 1.3% - Urine 2 (135.80 mg/dL): Repeatability SD 0.206, CV 0.2%; Within-Lab SD 1.601, CV 1.2% - Urine QC 1 (195.79 mg/dL): Repeatability SD 0.253, CV 0.1%; Within-Lab SD 2.376, CV 1.2% (All results demonstrate low CVs, indicating good precision).
    ReproducibilityDetermined in accordance with CLSI Document EP05-A3 (implies meeting specific statistical targets for variability components across different days, lots, and instruments).Serum Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.0% to 5.0%. Urine Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.4% to 1.6%. (All results demonstrate good reproducibility across conditions).
    Assay ComparisonSerum: Correlation coefficient $\ge$ 0.950 and slope of 1.00 $\pm$ 0.05, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression. Urine: Correlation coefficient $\ge$ 0.950 and slope of 0.000 $\pm$ 3.00, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression.Serum (n=151): Regression equation y = 1.00x - 0.04 mg/dL, correlation coefficient (r) = 1.000. Sample range 0.44 to 28.64 mg/dL. Urine (n=113): Regression equation y = 1.00x + 0.14 mg/dL, correlation coefficient (r) = 1.000. Sample range 12.60 to 237.06 mg/dL. (Both serum and urine results meet the acceptance criteria for correlation and slope).
    Specimen EquivalenceDetermined using Weighted Deming regression (implicitly, the regression line should demonstrate equivalence, i.e., close to y=x, with high correlation coefficient).Sodium Heparin (n=50): y = 1.00x + 0.00 mg/dL, r=0.999. Lithium Heparin (n=50): y = 0.99x + 0.06 mg/dL, r=0.999. Dipotassium EDTA (n=50): y = 0.98x + 0.04 mg/dL, r=0.998. (All demonstrate strong equivalence to serum reference).
    Interferences (HIL)$\le$ 10% interference from hemoglobin, bilirubin, and lipemia. Bias > 10% or 0.15 mg/dL (whichever is greater for serum/plasma) is considered interference.Reported biases for Hemoglobin (1000 mg/dL), Conjugated Bilirubin (40-45 mg/dL), Unconjugated Bilirubin (45-60 mg/dL), and Lipemia (2250-3000 mg/dL) are all within the $\pm$10% or $\pm$0.15 mg/dL threshold for the tested analyte concentrations, demonstrating acceptable interference profiles.
    Interfering SubstancesBias $\le$ 10% or $\pm$0.15 mg/dL for Serum/plasma samples. Bias $\le$ 10% for Urine samples (for listed substances).Most tested substances (e.g., Acetaminophen, Ascorbic Acid, etc.) show negligible bias, meeting the criteria. Substances showing bias beyond acceptance criteria for Serum: - Cefoxitin: Significant interference (e.g., 243.6% and 947.9% bias at high concentrations). - Cephalothin: Shows significant bias (e.g., 44.0% bias at 180 mg/dL). - Glucose: Shows bias beyond 10% at higher concentrations (e.g., 11.5% at 500 mg/dL and 22.5% at 1000 mg/dL). - Total Protein: Shows bias beyond 0.15 mg/dL at 15 g/dL (0.45 mg/dL). - Acetohexamide: Shows bias beyond 10% at 2.0 mg/dL (10.4%). - Hydroxocobalamin (Cyanokit): Shows significant bias (e.g., 14.5% and 49.3% at higher concentrations). Substances showing bias beyond acceptance criteria for Urine: - Cefoxitin: Shows bias beyond 10% at higher concentrations (e.g., 11.3% and 15.4%). (The document explicitly lists these substances under "Interference beyond $\pm$10% for Serum" and "Interference beyond $\pm$10% for Urine," indicating that they failed the non-interference criteria at the tested concentrations. This is typical for IVD submissions, where known interferences are identified for labeling purposes).
    StandardizationThe assay shall be traceable to the reference material SRM967, from the National Institute of Standards and Technology (NIST).Statement confirms the assay is traceable to NIST SRM967.

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set Sample Sizes:
        • Detection Capability: Not explicitly stated as "sample size" but data points obtained according to CLSI EP17-A2.
        • Precision: 80 data points per serum/urine sample type (duplicate runs for 20 days, 2 runs/day).
        • Reproducibility: 225 data points per serum/urine sample type (n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots).
        • Assay Comparison: 151 serum samples and 113 urine samples.
        • Specimen Equivalence: 50 samples for each plasma type (Sodium Heparin, Lithium Heparin, Dipotassium EDTA) compared to serum.
        • Interference (HIL & Non-Interfering Substances): Not explicitly stated as a total sample size, but experiments are designed to test specific analyte concentrations with and without interferents, following CLSI EP07-ED3.
      • Data Provenance: Not explicitly stated in terms of country of origin. Given the manufacturer (Siemens Healthcare Diagnostics Inc. in Tarrytown, New York, USA) and FDA submission, it's highly probable the studies were conducted in the US or in compliance with US regulatory standards. The studies described are retrospective in the sense that they use pre-collected or prepared samples to assess the analytical performance of the device under controlled conditions, not prospective in tracking patient outcomes in a clinical trial.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • For an in vitro diagnostic (IVD) device measuring a quantitative analyte like creatinine, "ground truth" is typically established by reference methods or established laboratory standards and calibrators, not by human expert consensus or labeling of medical images.
      • The "ground truth" for creatinine concentration in this context is based on traceable reference materials (NIST SRM 967) and established laboratory measurement principles, and the performance is compared against a legally marketed predicate device.
      • Therefore, this question (relevant for AI/imaging devices) does not directly apply to this type of IVD submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical trials or image labeling pipelines where there's human interpretation involved and a need to resolve disagreements among multiple readers; this is an analytical performance study of an IVD assay.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/imaging device. It's an in vitro diagnostic assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is an automated IVD assay performed on the Atellica® CH Analyzer. Its intended use is quantitative determination of creatinine. Therefore, the performance described (precision, accuracy, interference, etc.) is its standalone performance without a human in the loop for the analytical measurement itself, though a human still interacts with the instrument and interprets the results in a clinical context.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary "ground truth" for the Atellica® CH Creatinine_3 assay's performance is traceability to NIST SRM 967 (a certified reference material for creatinine) and comparison to a legally marketed predicate device (Atellica® CH Creatinine_2) using method comparison validated against CLSI guidelines. This is a form of analytical reference standard and comparative performance to an established method.
    7. The sample size for the training set:

      • This device is an analytical chemistry assay, not a machine learning/AI algorithm that requires a "training set" in the computational sense. The "development" or "optimization" of the assay would involve various experimental data, but it's not codified as a "training set" for an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the AI/ML context for this type of device. The assay development would rely on scientific principles of analytical chemistry, reagent formulation, and instrument calibration against known standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221813

    Validate with FDA (Live)

    Date Cleared
    2024-11-19

    (881 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nova Allegro urine albumin creatinine ratio (UACR) Assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy.

    The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.

    Device Description

    Nova Allegro UACR Assay: The Nova Allegro UACR Assay is a completely automated assay for the determination of albumin and creatinine in human urine. The results are used to calculate the UACR (Urine Albumin to Creatinine Ratio). Nova Allegro UACR Test Cartridges are the key element a user interacts with to determine the albumin and creatinine levels in a human urine sample. The main components of the Test Cartridge are the Capillary, the reaction chamber, and the barcode label.

    Nova Allegro Analyzer: The Nova Allegro Analyzer is a compact, point-of-care analyzer that features a clinically important menu of measured and calculated tests. All tests are measured with disposable, ready-to-use cartridges. The analyzer supports multiple wavelengths that are used to measure the assay of interest. The analyzer consists of the following key systems/components that the user interacts with:

    • Two analytical bays where the single use test cartridges are analyzed
    • Color Touchscreen Display
    • Barcode Scanner
    • Printer
    • Data Export Options
    • Ethernet Connection
    • USB Port
    AI/ML Overview

    The provided text is a 510(k) Summary for the Nova Allegro UACR Assay and Analyzer. It details the device's performance testing results. According to the prompt's request, I will extract relevant information to describe the acceptance criteria and the study proving the device meets these criteria.

    It's important to note that this document describes a medical device, specifically an in-vitro diagnostic test, not an AI/ML-based device. Therefore, many of the requested fields related to AI/ML specific quality evaluation criteria, such as "number of experts used to establish ground truth," "adjudication method," "MRMC study," "effect size of human readers improving with AI," "standalone performance," and details about training sets/ground truth for training, are not applicable in this context. The study performed is a performance validation of a diagnostic assay, comparing its results to a comparative method (CM), not a reader study.

    Here's a breakdown of the requested information based on the provided document:


    Device: Nova Allegro UACR Assay, Nova Allegro Analyzer
    Indications for Use: The Nova Allegro urine albumin creatinine ratio (UACR) Assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy. The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful conclusion of each test and the statement that the data "meets the acceptance criteria" or "demonstrated no significant interference." Specific quantitative acceptance criteria are sometimes stated (e.g., <10% difference for interference).

    Performance MetricAcceptance Criteria (Implied/Stated)Reported Device Performance
    LinearityData indicates the system reports linear albumin and creatinine test results across the reportable range when compared to the Comparative Method (CM). Expected to "meet the acceptance criteria."The linearity data met the acceptance criteria when compared to the CM. The data indicates that the Nova Allegro System reports linear albumin and creatinine test results using urine specimens across the reportable range.
    Hook EffectOccurrence above 5,000 mg/L (10 times higher than previously stated criteria in the document).The "Hook Effect" was determined to occur above 50,000 mg/L, meeting the acceptance criteria. This is 10 times higher than the acceptance criteria of 5,000 mg/L and approximately 167 times greater than the upper reportable measurement range for the assay.
    Detection LimitLOB, LOD, and LOQ for Albumin and Creatinine measurements must be below the lower limit of the claimed measurement range.The study demonstrates that the LOB, LOD, and LOQ results for Albumin and Creatinine measurements by the Nova Allegro UACR Assay are all below the lower limit of the claimed measurement range.
    InterferenceA test substance was considered to interfere if the absolute difference between the mean test value and the mean control value was greater than 10%.Most tested substances demonstrated no significant interference (greater than 10% effect). Specific interfering substances (Table 2) identified with their highest non-interfering concentrations. Urines with pH between 4.3 – 10.0 exhibited no interference (except low Albumin at pH 4.0). Specific gravity between 1.003 and 1.048 exhibited no clinical interference.
    Method ComparisonIndicated by "data supports the 510(k) substantial equivalence approval decisions." For albumin, creatinine, and UACR, linear regression analysis "r" values (correlation coefficients) close to 1 (e.g., ≥0.99) and slopes/intercepts indicative of close agreement with the CM.Albumin (ALL Combined): Slope = 0.983, Intercept = 0.858, r = 0.998. Sample range Allegro 5-300 mg/L.
    Creatinine (ALL Combined): Slope = 0.949, Intercept = 3.775, r = 0.997. Sample range Allegro 16-498 mg/dL.
    UACR (ALL Combined): Slope = 0.995, Intercept = 1.101, r = 0.997. Sample range Allegro 1-1592.
    Precision (Repeatability)Low %CV values (e.g., <5-10%) for mean concentrations across sites, indicating consistent results within a laboratory.Microalbumin: Overall %CV: Urine 1 (3.6%), Urine 2 (4.0%), Urine 3 (3.6%)
    Creatinine: Overall %CV: Urine 1 (4.0%), Urine 2 (3.1%), Urine 3 (3.4%)
    UACR: Overall %CV: Urine 1 (2.5%), Urine 2 (2.5%), Urine 3 (2.4%)
    Precision (Between Run/Operator)Low %CV values.Microalbumin: Overall %CV: Urine 1 (2.6%), Urine 2 (0.3%), Urine 3 (1.5%)
    Creatinine: Overall %CV: Urine 1 (2.3%), Urine 2 (0.5%), Urine 3 (2.7%) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    UACR: Overall %CV: Urine 1 (1.2%), Urine 2 (0.6%), Urine 3 (1.8%)
    Precision (Between Lot/Day)Low %CV values.Microalbumin: Overall %CV: Urine 1 (1.3%), Urine 2 (1.3%), Urine 3 (0.5%)
    Creatinine: Overall %CV: Urine 1 (1.0%), Urine 2 (0.5%), Urine 3 (0.5%)
    UACR: Overall %CV: Urine 1 (0.5%), Urine 2 (1.1%), Urine 3 (0.2%)
    Precision (Within Laboratory)Low %CV values across sites.Microalbumin: Overall %CV: Urine 1 (4.6%), Urine 2 (4.1%), Urine 3 (4.2%) ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Creatinine: Overall %CV: Urine 1 (4.8%), Urine 2 (3.4%), Urine 3 (4.5%)
    UACR: Overall %CV: Urine 1 (2.9%), Urine 2 (3.0%), Urine 3 (3.2%)
    Precision (Between Site)Low %CV values.Microalbumin: Overall %CV: Control 1 (1.0%), Control 2 (1.6%)
    Creatinine: Overall %CV: Control 1 (1.0%), Control 2 (0.9%) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    UACR: Overall %CV: Control 1 (1.2%), Control 2 (1.2%)

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:

      • Linearity Testing: Not explicitly stated as a number of patient samples, but involved "UACR control solutions (Level 1 & Level 2)" and "two (2) urine pools containing low and very high albumin and creatinine concentrations."
      • Hook Effect: Involved "both pooled urine specimens and calibrator matrix."
      • Detection Limit: Sixty (60) replicate measurements for LOB, thirty-six (36) replicate measurements for LOQ.
      • Interference Testing: Ten (10) replicate UACR tests were performed on prepared urine specimens per interfering substance.
      • Method Comparison: A total of 535 samples within the measurement range for Albumin (41 altered) and 653 samples within the measurement range for Creatinine (0 altered). For UACR, 531 samples were tested (41 altered). (Note: the number of altered vs. unaltered samples is specified. The altered samples are used to ensure coverage of the analytical measurement range).
      • Precision: Urine samples (3 levels) were tested in replicates over multiple days/runs at 4 sites. Specific numbers of replicates are indicated by the CLSI guidance used, but not explicitly enumerated per site/level in the summary beyond general study setup.
    • Data Provenance:

      • Country of Origin: Not explicitly stated in the summary.
      • Retrospective or Prospective: The "Point-of-Care (POC) Method Comparison studies on fresh urine specimens were conducted within four (4) different POC sites." This implies a prospective collection of samples during the study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Not applicable for this type of diagnostic device performance study. Ground truth is established by a "Comparative Method" (CM), which is another established, legally marketed, and often more robust laboratory assay.
    • The "CM method" for albumin was the "Siemens Dimension EXL 200 Integrated Chemistry System utilizing Dimension® Flex® reagent cartridge MALB, Dimension calibrator cartridge MALB CAL (Traceable to IFCC International Reference Preparation for Plasma Proteins)."
    • The "CM method" for creatinine was "Dimension Flex Reagent Cartridge CRE2, and Dimension CHEM 1 CAL (Traceable to NIST SRM 914 and verified with NIST SRM 967)." These are reference laboratory methods, not human expert interpretations.

    4. Adjudication Method for the Test Set

    • None. This is a quantitative chemical assay, not an interpretative AI/ML system requiring human adjudication of ground truth labels. Comparison is made against a validated laboratory reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs. without AI assistance

    • Not applicable. This is not an AI/ML-based device that assists human readers. It is a standalone diagnostic test for laboratory and point-of-care use.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, this is a standalone device. The Nova Allegro Analyzer and UACR Assay provides quantitative results directly. Its performance is validated against a comparative method in a standalone manner. The "human-in-the-loop" here is the operator running the test, not an interpretive role.

    7. The Type of Ground Truth Used

    • Comparative Method / Reference Method: The ground truth for the "Method Comparison" study was established using the Siemens Dimension EXL 200 Integrated Chemistry System, which served as the reference standard (CM method). This system's calibration is traceable to international reference preparations (IFCC) and NIST standards.

    8. The Sample Size for the Training Set

    • Not applicable. As an in-vitro diagnostic device, this product does not involve machine learning or AI models that require a "training set." Its analytical performance is based on established chemical and immunoturbidimetric principles, validated through traditional analytical and clinical performance studies, not algorithmic training data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210452

    Validate with FDA (Live)

    Device Name
    Creatinine2
    Date Cleared
    2022-03-22

    (399 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Creatinine2 assay is used for the quantitation of creatinine in human serum, plasma, or urine on the ARCHITECT c System.

    The Creatinine2 assay is to be used as an aid in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    Device Description

    The Creatinine2 assay is an automated clinical chemistry assay. At an alkaline pH, creatinine in the sample reacts with picric acid to form a creatinine-picrate complex that absorbs at 500 nm. The rate of increase in absorbance is directly proportional to the concentration of creatinine in the sample.

    AI/ML Overview

    The provided document describes the Abbott Creatinine2 assay, an in vitro diagnostic device, and its performance relative to a predicate device. The information needed to answer the request is primarily found in Section 5: 510(k) Summary (Summary of Safety and Effectiveness), specifically subsections VIII (Summary of Nonclinical Performance) and VII (Comparison of Technological Characteristics).

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a separate, pre-defined column. Instead, it presents the results of various performance studies. The "Reported Device Performance" below is extracted directly from the study results presented in the document. The comparable predicate device values are included for context where available.

    Creatinine2 Assay - Performance Summary

    MetricAcceptance Criteria (Implied/Predicate Performance)Reported Device Performance (Creatinine2)
    Analytical Measuring Interval (AMI)Serum/Plasma: 0.09 – 37.34 mg/dL Urine: 2.54 – 740 mg/dL
    Reportable IntervalSerum/Plasma: 0.04 – 37.34 mg/dL Urine: 1.24 – 740 mg/dL
    Precision (Within-Laboratory %CV)Predicate (Serum/Plasma): 3.18 - 4.95% (at 1.20-4.66 mg/dL) Predicate (Urine): 1.27 - 1.34% (at 61.95-145.48 mg/dL)Serum/Plasma: ≤ 4.5% (Panel A: 0.25 mg/dL), ≤ 2.3% (Panel B: 26.00 mg/dL), ≤ 2.1% (Panel C: 36.36 mg/dL) Urine: ≤ 5.5% (Panel A: 5.37 mg/dL), ≤ 1.8% (Panel B: 278.12 mg/dL), ≤ 1.8% (Panel C: 701.12 mg/dL)
    Accuracy (Bias)Serum/Plasma (vs. NIST SRM 967a): -4.1% to 0.4% bias Urine (vs. NIST SRM 914a): -4.8% to 3.3% bias
    Lower Limits of Measurement (LoB, LoD, LoQ)Predicate (Serum/Plasma LoD): 0.05 mg/dL Predicate (Serum/Plasma LoQ): 0.10 mg/dL Predicate (Urine LoD): 4.00 mg/dL Predicate (Urine LoQ): 5.00 mg/dLSerum/Plasma: LoB: 0.02 mg/dL, LoD: 0.04 mg/dL, LoQ: 0.09 mg/dL Urine: LoB: 0.93 mg/dL, LoD: 1.24 mg/dL, LoQ: 2.54 mg/dL
    LinearityPredicate (Serum/Plasma): 0.20 – 37.00 mg/dL Predicate (Urine): 5.00 – 740.00 mg/dLSerum: 0.09 to 37.34 mg/dL Urine: 2.54 to 740 mg/dL
    Method Comparison (Correlation Coefficient)Predicate: Not explicitly stated as a target, but K083809 (Abbott Creatinine) is the predicate and a correlation of 1.00 for both serum and urine suggests excellent agreement.Serum: 1.00 (between Creatinine2 and predicate Creatinine) Urine: 1.00 (between Creatinine2 and predicate Creatinine)
    Method Comparison (Intercept)Serum: -0.01 Urine: -1.23
    Method Comparison (Slope)Serum: 0.96 Urine: 1.01
    Interference (within ±10%)No significant interferenceSerum/Plasma Endogenous: No interference from listed substances at specified levels (e.g., Acetoacetate 20 mg/dL, Hemoglobin 1000 mg/dL) Serum/Plasma Exogenous: No interference from listed substances at specified levels (e.g., Acetaminophen 160 mg/L) Urine Endogenous: No interference from listed substances at specified levels (e.g., Acetoacetate 480 mg/dL) Urine Exogenous: No interference from listed substances at specified levels (e.g., Acetaminophen 16 mg/dL)

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes several nonclinical laboratory studies.

    • Precision (Within-Laboratory): For both Serum/Plasma and Urine, the studies tested 80 replicates per sample type for each of the two controls and three human panels (5 samples total). This was done in duplicate, twice per day, on 20 days. The provenance of the human panels (e.g., country of origin, retrospective/prospective) is not specified, but they are referred to as "human serum panels" and "human urine panels." This data is ex vivo laboratory testing.
    • Accuracy: No specific sample size of "patient samples" is given. The study was performed using "material standardized to the Certified Reference Material NIST SRM 967a" for serum/plasma and "material standardized to the Certified Reference Material NIST SRM 914a" for urine.
    • Lower Limits of Measurement (LoB, LoD, LoQ): n ≥ 60 replicates for zero-analyte and low-analyte level samples for LoB/LoD, and for low-analyte level samples for LoQ.
    • Linearity: The sample size for linearity is not explicitly stated in terms of number of unique samples, but it covers the analytical measuring interval by spiking and dilution.
    • Method Comparison:
      • Serum: 128 samples
      • Urine: 129 samples
        The provenance of these clinical samples (e.g., country of origin, retrospective or prospective) is not explicitly stated.
    • Interference: "Each substance was tested at 2 levels of the analyte." No specific sample size (n) for the number of replicates per interference test is given beyond this, nor is the provenance of the base samples used.
    • Tube Type: "Samples were collected from a minimum of 40 donors." The provenance is not explicitly stated.

    The studies described are nonclinical laboratory studies, primarily involving analytical performance evaluation rather than clinical patient studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable (N/A) to this specific device (Creatinine2 assay). The device is an in vitro diagnostic for quantitative measurement of creatinine, not an imaging device or a device requiring expert interpretation of complex clinical data to establish ground truth for its performance evaluation (e.g., a diagnosis of a disease from imaging). The "ground truth" for its analytical accuracy is typically established against certified reference materials (NIST SRM 967a for serum/plasma and NIST SRM 914a for urine) or reference methods, not by human experts adjudicating cases for a test set.

    4. Adjudication Method for the Test Set

    This information is not applicable (N/A) for the same reasons as #3. Clinical adjudication by multiple experts (like 2+1, 3+1) is typically used for devices that rely on human interpretation of outputs (e.g., medical images, pathology slides) where consensus or expert opinion defines the ground truth for diagnostic accuracy. This device measures a biochemical analyte.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for medical devices where human readers or interpreters are part of the diagnostic workflow, such as imaging-based AI tools. The Creatinine2 assay is an automated clinical chemistry assay that directly measures creatinine levels in biological samples and does not involve human image interpretation or a "human-in-the-loop" effectiveness study as typically understood in the context of MRMC studies.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The Creatinine2 assay is a standalone (algorithm only) device in the sense that it performs a quantitative measurement without a human-in-the-loop for interpreting the output of the device to arrive at the creatinine value. The performance metrics described (precision, accuracy, linearity, lower limits of measurement, interference, method comparison) are all tests of the device's performance directly, independent of human interpretation or assistance in generating the result.

    7. The Type of Ground Truth Used

    The ground truth for the performance evaluation of the Creatinine2 assay was primarily established using:

    • Certified Reference Materials: NIST SRM 967a for serum/plasma and NIST SRM 914a for urine were used for accuracy studies.
    • Predicate Device/Reference Method: The Creatinine assay (K083809; List No. 3L81) was used as a comparator for the method comparison study to assess substantial equivalence.
    • Defined Standards/Controls: For precision and lower limits of measurement, studies used control materials and low-analyte level samples with known or established concentrations to determine repeatability, detection, and quantitation limits.

    8. The Sample Size for the Training Set

    This information is not applicable (N/A). The Creatinine2 assay is a chemical assay, not a machine learning or AI-based device that requires a "training set" in the computational sense. Its performance is based on the chemical reaction and analytical methods described.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable (N/A) for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161494

    Validate with FDA (Live)

    Date Cleared
    2016-11-15

    (167 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica™ CH Creatinine 2 (Crea 2) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer, Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

    The Atellica™ CH Chemistry Calibrator (CHEM CAL) is for in vitro diagnostic use in calibrating the Crea 2 assay using the Atellica™ CH Analyzer.

    Device Description

    The Atellica CH Creatinine_2 (Crea_2) assay is based on the reaction of picric acid with creatinine in an alkaline medium as described in the original procedure of Jaffe. Creatinine reacts with picric acid in an alkaline medium to produce a red-colored creatinine-picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Creatinine 2 (Crea_2) assay is a modification of the Jaffe method using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because nonspecific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 umol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 umol/L) from each result.

    The Atellica CH Chemistry Calibrator (CHEM CAL) is a 1 level lyophilized calibrator product prepared from bovine serum base product.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study details for the Atellica CH Creatinine 2 (Crea 2) and Atellica CH Chemistry Calibrator (CHEM CAL) devices:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied/Direct)Reported Device Performance (New Device)
    Limit of Blank (LoB)95th percentile, non-parametric approachSerum: 0.03 mg/dL, Urine: 0.35 mg/dL
    Limit of Detection (LoD)Not explicitly stated as a criterion, but determined per CLSI EP17-A2Serum: 0.08 mg/dL, Urine: 0.51 mg/dL
    Limit of Quantitation (LoQ) - SerumTotal Error (TE) ≤ ±0.1 mg/dL for serumMeasured LoQ: 0.13 mg/dL (supports claim of 0.15 mg/dL)
    Limit of Quantitation (LoQ) - UrineTotal Error (TE) ≤ ±1.5 mg/dL for urineMeasured LoQ: 2.57 mg/dL (supports claim of 3.00 mg/dL)
    Linearity (Serum/Plasma)p-values of nonlinear terms in quadratic and cubic fit equations are nonsignificant (p ≤ 0.05). If p-value > 0.05, allowable bias ≤ 5% or 0.15 mg/dL (whichever is greater).The assay was deemed linear across the measuring interval (details of specific p-values / bias not provided but stated as meeting criteria).
    Linearity (Urine)p-values of nonlinear terms in quadratic and cubic fit equations are nonsignificant (p ≤ 0.05). If p-value > 0.05, allowable bias ≤ 5% or 0.15 mg/dL (whichever is greater).The assay was deemed linear across the measuring interval (details of specific p-values / bias not provided but stated as meeting criteria).
    PrecisionNot explicitly stated as acceptance criteria, but evaluated per CLSI EP05-A3. (Results in table below are the "reported performance")(See detailed table below)
    InterferencesBias exceeding 10% is considered interference.No interference detected at specified high concentrations for various compounds in serum and urine.
    Method Comparison (vs. Predicate)Good agreement with predicate device.Serum: y = 0.98x + 0.00 (r=0.999, N=140) Urea: y = 0.95x + 0.07 (r=0.999, N=109)
    Method Comparison (vs. IDMS)Good agreement with IDMS.Serum: y = 0.96x + 0.05 (r=0.999, N=49)
    Matrix Equivalency (Plasma vs. Serum)Not explicitly stated, but "demonstrated" by high correlation and near 1:1 regression.Plasma: y = 1.00x – 0.01 (r=1.000, N=58)

    Detailed Precision Results:

    Sample TypeMean mg/dL (µmol/L)Repeatability SDa mg/dL (µmol/L)Repeatability CVb (%)Within-Lab Precision SDa mg/dL (µmol/L)Within-Lab Precision CVb (%)
    Serum0.38 (34)0.01 (0.5)1.70.010 (0.9)2.8
    Plasma Pool0.66 (58)0.01 (0.7)1.20.018 (1.6)2.8
    Serum Pool1.16 (103)0.01 (0.9)0.80.017 (1.5)1.5
    Serum QC1.97 (174)0.02 (1.6)0.90.024 (2.1)1.2
    Serum QC6.35 (561)0.04 (3.7)0.70.062 (5.5)1.0
    Serum Pool19.31 (1707)0.04 (3.4)0.20.117 (10.3)0.6
    Serum Pool26.00 (2298)0.05 (4.7)0.20.145 (12.8)0.6
    Urine QC59.62 (5270)0.15 (13.5)0.30.376 (33.2)0.6
    Urine QC133.31 (11785)0.33 (29.4)0.20.961 (85.0)0.7
    Urine188.61 (16673)0.52 (46.1)0.31.779 (157.3)0.9

    2. Sample Size Used for the Test Set and Data Provenance

    • Detection Limit (LoB/LoD):
      • LoB: 4 samples with no analyte, tested 5 times a day for 3 days (60 reps total).
      • LoD: 4 low analyte samples, tested 5 times a day for 3 days (60 reps total).
    • Limit of Quantitation (LoQ):
      • Serum: 10 low samples, 5 replicates each, on 3 reagent lots for 3 days, 5 calibrations per day (75 measurements per reagent lot per sample). Total of 2250 determinations.
      • Urine: Similar setup for urine samples. Total of 2250 determinations.
    • Linearity Study:
      • Serum/Plasma: 12 samples (high and low concentration mixes). 5 replicates measured for each sample.
      • Urine: 10 samples (high and low concentration mixes). 5 replicates measured for each sample.
    • Precision Studies:
      • Controls, serum, and plasma pools: 80 replicates (n=2 replicates, 2 times a day for at least 20 days).
    • Interferences:
      • "Fresh sample pools" containing low or high levels of creatinine in serum and urine. Specific count not given, but refers to testing across these pools.
    • Method Comparison (Predicate):
      • Serum: 140 remnant de-identified samples.
      • Urine: 109 remnant de-identified samples.
      • Data Provenance: "Remnant de-identified samples" implies retrospective patient data. "No patient history information was obtained." The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D personnel.
    • Method Comparison (IDMS):
      • Serum: 49 samples.
    • Matrix Equivalency:
      • 58 matched serum and lithium heparin plasma samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (Creatinine assay system) does not typically involve human expert adjudication for its ground truth. The "ground truth" for a chemical assay is established through highly accurate reference methods or certified reference materials.

    • Ground Truth for Method Comparison: The predicate device, ADVIA Chemistry Enzymatic Creatinine_2 (ECRE_2), served as the comparative "ground truth" for the method comparison study. Additionally, "Isotope Dilution Mass Spectrometry (IDMS)" was used as a reference method for a subset of samples, which is a highly accurate and precise method for determining analyte concentrations and is recognized as a gold standard in clinical chemistry.
    • Ground Truth for Calibration/Standardization: The device uses "IDMS Reference Method" for standardization, and the predicate uses "SRM967" (Standard Reference Material 967, which is a NIST standard for creatinine in human serum, often value-assigned by IDMS). These are highly precise and accurate methods, not typically involving human expert consensus in the same way an imaging device would.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device. Ground truth is established by quantitative measurement or comparison to a reference method, not by expert adjudication of human interpretations.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic device for quantitative chemical analysis, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone, automated analytical system (Atellica CH Analyzer) that performs the creatinine assay. The performance results reported (LoB, LoD, LoQ, Linearity, Precision, Interference, Method Comparison) represent the standalone performance of the assay on the analyzer. There is no human-in-the-loop component for the measurement of creatinine by this system.


    7. The Type of Ground Truth Used

    • Reference Methods:
      • The predicate device (ADVIA Chemistry Enzymatic Creatinine_2) was used as a reference for method comparison.
      • Isotope Dilution Mass Spectrometry (IDMS) was used as a highly accurate reference method for a subset of serum samples. This is considered a gold standard for creatinine measurement.
    • Reference Materials: Standardization mentions "IDMS Reference Method," which implies traceability to primary reference materials. The predicate mentions SRM967, a certified reference material.
    • Defined Protocols: CLSI (Clinical and Laboratory Standards Institute) protocols (EP17-A2, EP05-A3, EP06-A, EP7-A2, EP09-A3, EP28-A3c) were followed for various performance evaluations, which define how to establish performance characteristics against expected statistical and analytical metrics.

    8. The Sample Size for the Training Set

    Not explicitly mentioned in the provided text as a "training set" in the context of machine learning, which is typically what this question implies. For an IVD assay, method development and initial optimization would involve numerous experiments and samples, but these are typically not referred to as a "training set" in the AI sense. The text focuses on the validation studies performed to demonstrate performance.


    9. How the Ground Truth for the Training Set Was Established

    As above, the concept of a "training set" with established ground truth as used in AI/ML is not directly applicable here. The development of an IVD assay involves extensive R&D, where analytical methods are refined to accurately measure the analyte. The "ground truth" during this development phase would be established by reference methods or gravimetric/volumetric preparations of known concentrations, similar to how the validation ground truth is established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132638

    Validate with FDA (Live)

    Date Cleared
    2014-01-27

    (157 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension® clinical chemistry system. Creatinine measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    Device Description

    The CRE2 method uses a modified kinetic Jaffe technique. In the presence of a strong base such as sodium hydroxide, picrate reacts with creatinine to form a red chromophore. The rate of increasing absorbance at 510 nm due to the formation of this chromophore is directly proportional to the creatinine concentration in the sample and is measured using a bichromatic (510, 600nm) rate technique. Bilirubin is oxidized by potassium ferricyanide to prevent interference.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a new in vitro diagnostic device, the Dimension® Creatinine (CRE2) Flex® reagent cartridge. This document is a summary of the safety and effectiveness information, comparing the new device to a predicate device already on the market.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria for performance parameters in a pass/fail format. Instead, it provides the results of various performance studies and implies that these results demonstrate substantial equivalence to the predicate device and/or meet established industry guidelines (CLSI). The acceptance criteria are implicitly defined by what is considered "substantially equivalent" or within acceptable limits as per CLSI guidelines.

    Below is a table summarizing the reported device performance, categorized by the studies conducted. It's important to note that specific numerical acceptance criteria (e.g., "slope must be between 0.98 and 1.02") are not explicitly stated, but the reported values passed the FDA's substantial equivalence review.

    Performance CharacteristicReported Device Performance (CRE2 Flex®)Implied Acceptance Criteria / Context
    Method Comparison (vs. Predicate CREA Assay)Agreement with predicate device (CREA).
    Serum Sample (n=191)Range: 0.4 - 19.8 mg/dLGood agreement.
    Slope: 1.00Ideal slope is 1.0.
    Intercept: -0.08 mg/dLIdeal intercept is 0.
    Correlation Coefficient (r): 0.999Close to 1.0 indicates strong correlation.
    Urine Sample (n=113)Range: 13.5 - 372.7 mg/dLGood agreement.
    Slope: 1.04Close to 1.0.
    Intercept: -3.58 mg/dLClose to 0.
    Correlation Coefficient (r): 0.996Close to 1.0 indicates strong correlation.
    Method Comparison (vs. IDMS Reference Method)Agreement with a recognized gold standard.
    Serum Sample (n=48)Range: 0.18 - 6.32 mg/dLGood agreement.
    Slope: 1.04Close to 1.0.
    Intercept: 0.02 mg/dLClose to 0.
    Correlation Coefficient (r): 0.997Close to 1.0 indicates strong correlation.
    Serum Plasma EquivalencyAgreement between serum and plasma samples.
    Lithium Heparin Plasma (n=56)Range: 0.50 - 17.35-
    Slope: 1.05Close to 1.0.
    Intercept: -0.02Close to 0.
    Correlation Coefficient (r): 0.998Close to 1.0 indicates strong correlation.
    Precision (Repeatability & Within-Lab CV)Within acceptable limits for clinical chemistry assays.
    Serum Pool 1 (Mean 1.32 mg/dL)Repeatability SD: 0.04, %CV: 3.0Low %CV indicates high precision.
    Within-Lab SD: 0.04, %CV: 3.2Low %CV indicates high precision.
    Serum Pool 2 (Mean 15.79 mg/dL)Repeatability SD: 0.19, %CV: 1.2Low %CV.
    Within-Lab SD: 0.19, %CV: 1.2Low %CV.
    BioRad Multiqual Level 1 (Mean 0.71 mg/dL)Repeatability SD: 0.03, %CV: 4.7Low %CV.
    Within-Lab SD: 0.04, %CV: 5.1Low %CV.
    ... (other samples shown in table, similar low CVs)
    Limit of Blank (LoB)Meet or exceed claimed LoB.
    Serum LoB0.05 mg/dL (consistent with claim)Claimed LoB: 0.05 mg/dL
    Urine LoB0.87 mg/dL (consistent with claim)Claimed LoB: 1.0 mg/dL
    Limit of Detection (LoD)Meet or exceed claimed LoD.
    Serum LoD0.08 mg/dL (consistent with claim)Claimed LoD: 0.1 mg/dL
    Urine LoD1.51 mg/dL (consistent with claim)Claimed LoD: 2.0 mg/dL
    Limit of Quantitation (LoQ)Defined by allowable total error.
    Serum LoQ0.15 mg/dL (based on 0.15 mg/dL allowable total error)-
    Urine LoQ5.00 mg/dL (based on 3.00 mg/dL allowable total error)-
    Linearity (Measurement Range)Demonstrate linearity across the claimed measuring range.
    Serum (0.15* - 22.00 mg/dL)Slope: 1.01Close to 1.0.
    Intercept: -0.01Close to 0.
    Correlation Coefficient (r): 1.0Indicates perfect linearity.
    Urine (3.00* - 425.34 mg/dL)Slope: 0.998Close to 1.0.
    Intercept: -0.41Close to 0.
    Correlation Coefficient (r): 1.0Indicates perfect linearity.
    Analytical Specificity (Interference)Interference less than 10% bias, or disclosed.
    Non-interfering Substances (e.g., Acetaminophen, Caffeine, Glucose, etc.)% Diff typically between -2% to 6%Bias < 10% (threshold for non-interference). Some exceptions like Total Protein (6% at low pool, -1% at high pool) are also acceptable.
    Interfering Substances (e.g., Acetone, Bilirubin (unconjugated), Cefoxitin, Glucose, Intralipid 20%, Pyruvate, Triglycerides, Ascorbic Acid)% Diff > 10% (e.g., Acetone up to 11% bias)Identified as interfering; implications for use would be in labeling (not detailed in this section for the new device, but assumed to be consistent with predicate/known interferences).
    HIL Interference (Hemolysis, Icterus, Lipemia)Some substances show >10% bias (e.g., Hemoglobin at 1000 mg/dL, Bilirubin (conj) at 40 mg/dL, Bilirubin (unconj) at 20 mg/dL, Intralipid 20% at 1500 mg/dL)Identified as interfering; implications for use would be in labeling. The predicate's stated interferences (e.g., Hemoglobin at 1000 mg/dL, Bilirubin (unconjugated) at 5 mg/dL, Lipemia (Intralipid) at 200 mg/dL) are less tolerant than the new device in some cases for Bilirubin and Lipemia.

    2. Sample sizes used for the test set and the data provenance

    • Method Comparison (vs. Predicate CREA Assay):
      • Serum: 191 remnant de-identified patient samples.
      • Urine: 113 patient samples.
      • Provenance: Remnant de-identified samples. "No patient history information was obtained on these samples." The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel, implying a US-based setting. The data is retrospective as it used "remnant de-identified serum/urine samples."
    • Method Comparison (vs. IDMS Reference Method):
      • Serum: 48 remnant de-identified patient samples.
      • Provenance: Remnant de-identified samples. "No patient history information was obtained on these samples." Conducted internally by Siemens Healthcare Diagnostic Inc. R&D. Retrospective.
    • Serum Plasma Equivalency:
      • Samples: 56 matched serum and lithium heparin plasma samples.
      • Provenance: "All samples in the study were fresh and never frozen." The eight spiked sample sets were prepared. Conducted internally by Siemens Healthcare Diagnostic Inc. R&D. Retrospective, likely sourced from a local population.
    • Precision:
      • Samples: Two serum pools, three levels of BioRad Multiqual material, two levels of BioRad Liquicheck material, and two urine pools. Testing involved 20 days with 2 separate runs and 2 test samples for each material. The specific number of individual patient samples is not given, as this study uses control materials and pools.
      • Provenance: Not specified, but likely prepared internally or from commercial sources.
    • Limit of Blank/Detection (LoB/LoD):
      • Samples: 4 samples with no analyte (for LoB), 4 low patient serum/urine samples (for LoD). Each tested for 3 days, one run per day, across 2 reagent lots.
      • Provenance: Not specified, but generally prepared internally with blank/low-level samples.
    • Linearity (Measurement Range):
      • Samples: 12 equally spaced serum samples and 12 urine samples.
      • Provenance: Prepared by mixing high and low creatinine concentration samples.
    • Analytical Specificity (Interference):
      • Samples: Not specified in terms of number of patient samples, but involved "fresh sample pools containing either low or high levels of creatinine analyte in both serum pools and urine pools." These pools were spiked with various interfering substances.
      • Provenance: Likely developed from internal pools and commercially available substances.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable to this type of device (in vitro diagnostic for quantitative measurement of creatinine). The "ground truth" for clinical chemistry assays like creatinine is established by:

    • Reference methods (e.g., IDMS, considered the "gold standard" for creatinine measurement in some contexts).
    • Predicate devices (already cleared and established to perform reliably).
    • Known concentrations in control materials or spiked samples.

    There are no human "experts" (like radiologists interpreting images) involved in establishing ground truth for the test set in this context. The "personnel carrying out the study were laboratory technicians with training similar to personnel who would conduct the tests in a hospital laboratory setting."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This concept of "adjudication" is typical for studies involving human interpretation (e.g., medical imaging, pathology slides) where there might be disagreement among readers. For an automated in vitro diagnostic device, the measurement is quantitative and typically objective. Therefore, an adjudication method for the test set is not applicable. The "ground truth" is determined by instrumental readings and reference methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable here. This device is a fully automated in vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers. Its performance is evaluated against chemical reference methods and a predicate device, not in terms of how it assists human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, this entire submission and its performance data represent a standalone (algorithm only) performance evaluation. The device is an automated clinical chemistry system that measures creatinine levels directly from biological samples (serum, plasma, urine) without human intervention in the interpretation of the core measurement. The reported results (slope, intercept, correlation, precision, LoB/LoD, linearity, interference) directly reflect the device's inherent analytical performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for evaluating the Dimension® Creatinine (CRE2) Flex® reagent cartridge includes:

    • Reference Method: The IDMS (Isotope Dilution Mass Spectrometry) method for creatinine measurement. This is considered a highly accurate and precise reference method.
    • Predicate Device: The Creatinine Method for Use on the Dimension® Clinical Chemistry System (CREA, K925668). The predicate device's cleared performance serves as a benchmark for demonstrating substantial equivalence.
    • Known Concentrations: For studies like precision, LoB/LoD, linearity, and interference, known concentrations of analytes in control materials, spiked samples, or blank samples serve as the ground truth.

    8. The sample size for the training set

    This document describes a 510(k) submission for an in vitro diagnostic reagent cartridge that uses a modified kinetic Jaffe technique (a traditional chemical assay), not a machine learning or AI-based device. Therefore, the concept of a "training set" for an algorithm is not applicable. The "training" for such a device would refer to its development and optimization based on chemical principles, rather than data-driven machine learning.

    9. How the ground truth for the training set was established

    As answered in point 8, the device does not employ a training set in the context of an AI/ML algorithm. Its "ground truth" for development and calibration would be rooted in established chemical and analytical principles, reference measurement procedures (like IDMS), and potentially existing characterized samples. The manufacturing process of reagents, calibrators, and the instrument's detection capabilities are based on fundamental scientific understanding rather than a data training process as understood in AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133728

    Validate with FDA (Live)

    Date Cleared
    2014-01-24

    (49 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CRE2 method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista® System. Creatinine measurements are used in the diagnosis and treatment of certain renal disease, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    Device Description

    The Dimension Vista® Creatinine (CRE2) Flex® reagent cartridge uses a modified kinetic Jaffe technique. In the presence of a strong base such as sodium hydroxide, picrate reacts with creatinine to form a red chromophore. The rate of increasing absorbance at 510 nm due to the formation of this chromophore is directly proportional to the creatinine concentration in the sample and is measured using a bichromatic (510, 577 nm) rate technique.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dimension Vista® System Creatinine (CRE2) Flex® reagent cartridge, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Method Comparison (Serum)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to predicate device (Vista CREA Assay)Vista CREA Assay (Serum): Range 0.38 – 18.93 mg/dL, Slope = 1.02, Intercept = -0.11 mg/dL, Correlation Coefficient = 1.000 (n=140)
    Method Comparison (Urine)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to predicate device (Vista CREA Assay)Vista CREA Assay (Urine): Range 8.39 – 299.60 mg/dL, Slope = 1.05, Intercept = -4.29 mg/dL, Correlation Coefficient = 0.999 (n=112)
    Method Comparison (IDMS)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 when compared to IDMS Reference MethodIDMS Reference Method (Serum): Range 0.18–6.32 mg/dL, Slope = 1.06, Intercept = -0.03 mg/dL, Correlation Coefficient = 0.997 (n=48)
    Serum Plasma EquivalencySlope close to 1, Intercept close to 0, Correlation Coefficient close to 1 for matched serum vs. lithium heparin plasma samplesSlope = 1.03, Intercept = -0.002, Correlation Coefficient (r) = 0.998 (Range 0.418 - 17.9, n=56)
    PrecisionWithin-Lab %CV values for various serum and urine pools and control materials. (Implicitly, these should be within acceptable limits for a diagnostic test, though specific numerical criteria for acceptance are not explicitly stated as a single value)Reported %CVs: Serum Pool 1 (2.7%), Serum Pool 2 (0.7%), BioRad Multiqual Level 1 (4.0%), Level 2 (2.5%), Level 3 (1.1%), Urine Pool 1 (3.6%), Urine Pool 2 (1.0%), BioRad Liquicheck Level 1 (2.2%), Level 2 (2.0%). (All values are low, indicating good precision.)
    Limit of Blank (LoB)Consistent with claim of 0.05 mg/dL (serum), 1.0 mg/dL (urine)LoB (Serum): 0.030 mg/dL (consistent with 0.05 mg/dL claim) LoB (Urine): 0.438 mg/dL (consistent with 1.0 mg/dL claim)
    Limit of Detection (LoD)Consistent with claim of 0.1 mg/dL (serum), 2.0 mg/dL (urine)LoD (Serum): 0.056 mg/dL (consistent with 0.1 mg/dL claim) LoD (Urine): 0.828 mg/dL (consistent with 2.0 mg/dL claim)
    Limit of Quantitation (LoQ) Serum/PlasmaAllowable total error for creatinine published by CLIA, CAP, AAB, NYS, and WSLH is +/- 0.3 mg/dL or +/- 15%. (LoQ should meet these recommendations).LoQ (Serum/Plasma): 0.15 mg/dL based on allowable total error of 0.15 mg/dL. This meets the recommendations.
    Limit of Quantitation (LoQ) UrineNo recognized allowable total error limit for urine creatinine measurements. (Implicit acceptance is that the determined LoQ is appropriate and justified).LoQ (Urine): 5.00 mg/dL based on allowable total error of 3.00 mg/dL.
    Linearity (Serum)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 across the assay rangeLinearity (Serum): Range 0.00 – 21.9 mg/dL, Slope = 0.996, Intercept = -0.05, Correlation Coefficient = 1.0 (N=12)
    Linearity (Urine)Slope close to 1, Intercept close to 0, Correlation Coefficient close to 1 across the assay rangeLinearity (Urine): Range 1.71 – 338 mg/dL, Slope = 0.995, Intercept = 0.47, Correlation Coefficient = 1.0 (N=12)
    Interfering SubstancesBias exceeding 10% is considered interference. Dilution studies determine the level at which spiked substances no longer display significant interference.Demonstrated concentrations where various substances cause <10% bias or, if >10% bias, the concentration at which interference is no longer significant through dilution. For example, Acetone interferes at 150 mg/dL (18% bias) but not at 75 mg/dL (9% bias). Bilirubin (unconjugated) at 40mg/dL shows -29.6% bias at low creatinine, but 20mg/dL shows -0.2% bias.
    HIL InterferenceBias exceeding 10% is considered interference.Intralipid 20% interferes at 2000 mg/dL and 1500 mg/dL (at low creatinine concentration) (15.5% and 15.1% bias, respectively), but not at 1500 mg/dL (at high creatinine concentration) (9.5% bias) or 1000 mg/dL (7.0% bias). Hemoglobin shows no significant interference at 1000 mg/dL (~7% and 3% bias). Bilirubin (unconjugated) at 40 mg/dL shows -29.6% interference at low creatinine levels. Bilirubin (conjugated) shows 16.1% interference at 40 mg/dL at low creatinine levels.
    Expected Values (Serum & Plasma)≤10% of samples within the study should fall outside of established range.Males: 5 (9.4%) out of 53 samples outside range (meets ≤10%). Females: 2 (5.0%) out of 40 samples outside range (meets ≤10%). (Supports published ranges).
    Expected Values (Urine)≤10% of samples within the study should fall outside of established range.Males: 2 (9.1%) out of 22 samples outside range (meets ≤10%). Females: 1 (5.0%) out of 20 samples outside range (meets ≤10%). (Supports published ranges).
    Reportable Range Dilution (Serum)Mean percent recovery of 90-110% for autodiluted samples.Mean of N=5 reps: 16.5 µL (normal volume) = 16.5; 6.6 µL (auto-dilute volume) = 16.0; % Bias = -3.0%. (Implicitly meets recovery criteria, as bias is very low). Supports extended range to 40.0 mg/dL.
    Reportable Range Dilution (Urine)Mean percent recovery of 90-110% for autodiluted samples.Mean of N=5 reps: undiluted (normal) = 268; auto-dilute 3x with water = 245; % Bias = -8.6%. (Implicitly meets recovery criteria, as bias is very low). Supports extended range to 900 mg/dL.

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Method Comparison (Predicate Device):
        • Serum: 140 patient samples. Remnant de-identified serum samples. No patient history. Inclusion/exclusion criteria not applicable. Both native and spiked samples.
        • Urine: 112 patient samples. Remnant de-identified urine samples (implied). No patient history. Inclusion/exclusion criteria not applicable.
      • Method Comparison (IDMS Reference Method):
        • Serum: 48 patient samples. Remnant de-identified serum samples. No patient history. Inclusion/exclusion criteria not applicable. All native samples.
      • Serum Plasma Equivalency: 56 matched serum and lithium heparin plasma samples. All samples fresh and never frozen. 8 spiked sample sets.
      • Precision: Not a direct "test set" in the sense of patient samples, but testing involved: 2 serum pools, 3 levels of BioRad Multiqual material, 2 levels of BioRad Liquicheck material, and 2 urine pools. Each tested multiple times (over 20 days, 2 runs, 2 samples per run).
      • Limit of Blank/Detection (Serum): 4 samples with no analyte (N=5 reps for 3 days), 4 low patient serum samples (N=5 reps for 3 days).
      • Limit of Blank/Detection (Urine): 4 samples with no analyte (N=5 reps for 3 days), 4 low patient urine samples (N=5 reps for 3 days).
      • Linearity: 12 equally spaced samples for serum, 12 for urine.
      • Analytical Specificity (Interference): Low (~1.5 mg/dL) and High (~5.0 mg/dL) creatinine serum pools; Low (~40 mg/dL) and High (~175 mg/dL) creatinine urine pools. Number of individual samples/tests per interferent is not explicitly stated but implied to be sufficient for creating means and controls.
      • Expected Values:
        • Serum and Plasma: 53 male donors, 40 female donors (all normal healthy adults).
        • Urine: 22 male donors, 20 female donors (all normal healthy adults).
      • Reportable Range Dilution: N=5 replicates for each dilution study (serum and urine).

      Data Provenance: The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel. Remnant de-identified samples were used for method comparison studies, implying retrospective or archived samples. Patient history was not obtained. The origin of the patient samples (e.g., country) is not specified, but the internal R&D location would presumably be Newark, DE, USA based on the applicant's address.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not mention the use of experts to establish ground truth for the test set in the context of clinical interpretation. For method comparison, the "ground truth" for the new device was established by comparison to a legally marketed predicate device (Dimension Vista® Creatinine (CREA) Flex® reagent cartridge) and the IDMS reference method. These are analytical "gold standards" rather than expert interpretation.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      Not applicable. The reported studies are analytical performance evaluations of a quantitative laboratory test, not diagnostic interpretations requiring adjudication.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This is a quantitative in vitro diagnostic test for creatinine, not an imaging or interpretive device that would typically involve human readers or AI assistance in the way an MRMC study would assess.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
      Yes, these are all standalone (algorithm only) performance studies. The device itself performs the quantitative measurement of creatinine from samples without human interpretation in the loop of the measurement.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Method Comparison studies: The ground truth was established by two primary methods:
        • Comparison to a predicate device (Dimension Vista® Creatinine (CREA) Flex® reagent cartridge).
        • Comparison to the IDMS reference method (Isotope Dilution Mass Spectrometry), which is a highly accurate and precise analytical method often considered a "gold standard" for creatinine measurements.
      • For Expected Values studies: Published reference ranges in scientific literature (e.g., Tietz 1999, Clin Chem 54:3 (2008), Clin Chem Acta 344 (2004) 137-148) served as the basis for validating the established ranges for the new device.
      • For LoQ validation: Allowable total error limits published by regulatory and accreditation bodies (CLIA, CAP, AAB, NYS, WSLH) were used. The reference values are traceable to IDMS.
    7. The sample size for the training set:
      Not applicable in the context of typical AI/machine learning studies. This device is a reagent cartridge for an automated clinical chemistry system, not an AI or machine learning algorithm that undergoes 'training' in the conventional sense with labeled data. The development and internal refinement of the reagent formulation and measurement parameters would be analogous to "training" but is not quantified in terms of a discrete sample set size like this.

    8. How the ground truth for the training set was established:
      Not applicable for the same reasons as above. The "ground truth" for developing such a system would involve rigorous analytical chemistry principles, extensive experimentation, and calibration using traceable standards (e.g., NIST SRM 914a (IDMS assigned crystalline creatinine standard)).

    Ask a Question

    Ask a specific question about this device

    K Number
    K123586

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-15

    (175 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyRA Creatinine (CREA) Reagent is for the measurement of creatinine in serum and plasma using the "EasyRA chemistry analyzer". Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis. For in vitro diagnostic use only.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the EasyRA Creatinine Reagent. It states that the device is substantially equivalent to a predicate device for measuring creatinine in serum and plasma.

    However, this document does not contain the specific information required to answer your request about acceptance criteria and study details. The 510(k) clearance letter itself is a summary of the FDA's decision, not the full submission that would detail the performance studies and acceptance criteria.

    To answer your questions, one would need access to the actual 510(k) premarket notification submission for K123586, which would include the performance data and the study design.

    Therefore, I cannot provide the requested table and study details based solely on the provided text. The document confirms the device's indications for use: "The EasyRA Creatinine (CREA) Reagent is for the measurement of creatinine in serum and plasma using the "EasyRA chemistry analyzer". Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis." but does not elaborate on the performance characteristics or the studies conducted to establish substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110530

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-04-30

    (431 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba Medical methods on Horiba Medical chemistry analyzers.

    The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

    Device Description

    All the reagent, controls and calibrator included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative in vitro determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). It is composed of a bi-reagent cassette (R1= 27.5 mL ; R2= 8 mL). Reagents are chemical solutions with additives.

    The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. This calibrator is provided in ten vials of 3 ml.

    The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml.

    The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annex, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for the ABX PENTRA Creatinine 120 CP reagent and associated calibrators/controls, for use on the ABX PENTRA 400 clinical chemistry analyzer.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    ABX PENTRA Creatinine 120 CP (Reagent)
    Detection limit (Serum/Plasma)N/A (Comparison to predicate implies similar or better performance)0.074 mg/dl
    Detection limit (Urine)N/A1.40 mg/dl
    Limit of Quantitation (Serum/Plasma)N/A0.22 mg/dl
    Limit of Quantitation (Urine)N/A2.90 mg/dl
    Accuracy and Precision (Serum/Plasma)N/A (likely defined by an allowable CV% based on clinical standards or predicate)CV. Total < 4.72%
    Accuracy and Precision (Urine)N/ACV Total < 2.06%
    Measuring range (Serum/Plasma)N/A0.22 mg/dl - 18.08 mg/dl
    Measuring range (Urine)N/A2.90 mg/dl - 282.50 mg/dl
    Upper linearity limit (Serum/Plasma)N/A18.08 mg/dl (54.24 mg/dl with automatic post-dilution)
    Upper linearity limit (Urine)N/A282.5 mg/dl (847.50 mg/dl with automatic post-dilution)
    Correlation (Serum/Plasma)N/A (likely a high correlation coefficient with the predicate method)Y = 0.99 x + 0.03 (mg/dl) with r2 = 0.9984
    Correlation (Urine)N/AY = 1.00 x - 0.60 (mg/dl) with r2 = 0.9984
    Calibration stability (Serum/Plasma)N/A (Standard stability duration for similar reagents)3 days
    Calibration stability (Urine)N/A3 days
    Reagent stability (closed)N/A36 months at 2-8°C
    Reagent stability (on-board)N/A19 days
    ABX PENTRA Multical (Calibrator)
    Stability (Closed)N/A (Typical shelf-life for calibrators)24 months at 2-8°C
    Stability (Open, general)N/A (Typical in-use stability for calibrators)8 hours at 15-25°C; 2 days at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A3 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A6 hours at 15-25°C; 1 day at 2-8°C
    ABX PENTRA N Control (Control)
    Stability (Closed)N/A30 months at 2-8°C
    Stability (Open, general)N/A12 hours at 15-25°C; 5 days at 2-8°C; 1 month at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A4 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A8 hours at 15-25°C; 1 day at 2-8°C; 2 weeks at -25 to -15°C
    ABX PENTRA P Control (Control)
    Stability (Closed)N/A30 months at 2-8°C
    Stability (Open, general)N/A12 hours at 15-25°C; 5 days at 2-8°C; 1 month at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A4 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A8 hours at 15-25°C; 1 day at 2-8°C; 2 weeks at -25 to -15°C
    ABX PENTRA Urine Control L/H (Control)
    Stability (Closed)N/A2 years at 2-8°C
    Stability (Open)N/A30 days at 2-8°C

    Note: The document explicitly states "The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that the devices are substantially equivalent to their respective predicate devices." However, the exact numerical acceptance criteria values themselves (e.g., maximum allowable CV, minimum r2) for each parameter are not explicitly defined in the provided text, but are implied by the reported performance falling within acceptable ranges for substantial equivalence to the predicate devices.

    2. Sample sizes used for the test set and the data provenance

    • Creatinine 120 CP (Test Set):
      • Serum/Plasma Correlation: n = 165
      • Urine Correlation: n = 117
    • Data Provenance: Not explicitly stated, but the company is Horiba ABX SAS, FRANCE, suggesting the data may originate from France or related study sites. The study is for premarket notification, implying prospective testing for device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the text. For in vitro diagnostic devices like this, ground truth is typically established by comparing performance to an established, often predicate, method or reference method, rather than through expert consensus in the way a medical imaging AI might use radiologists.

    4. Adjudication method for the test set

    • This information is not provided as it's not relevant for this type of in vitro diagnostic device performance study. Performance is assessed quantitatively against a reference method or specified analytical limits.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not applicable/not provided. This is an in-vitro diagnostic assay and associated controls/calibrators, not an AI-assisted diagnostic tool that involves human readers interpreting results.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is a standalone diagnostic assay (reagent, calibrator, control) for use on an automated clinical chemistry analyzer (ABX PENTRA 400). The performance data presented (detection limits, accuracy, precision, measuring range, correlation, stability) reflect the standalone analytical performance of the device on the analyzer. There is no "human-in-the-loop" component for the measurement process itself beyond loading samples and reagents.

    7. The type of ground truth used

    • For the ABX PENTRA Creatinine 120 CP reagent, the ground truth for correlation studies was established by comparing its measurements (Y) against those of the predicate device/method (X), as indicated by the correlation equations (e.g., "Y = 0.99 x + 0.03"). This implies a comparative method ground truth (i.e., comparison against a legally marketed, established method).
    • For the calibrators and controls, the "assigned values" of their components serve as their reference values or ground truth, established through rigorous manufacturing and value assignment processes.

    8. The sample size for the training set

    • This information is not provided. This type of in vitro diagnostic device is chemically based and does not typically involve a "training set" in the machine learning sense. The development likely involves extensive R&D and analytical validation, but not a distinct "training set" of patient data for a learning algorithm.

    9. How the ground truth for the training set was established

    • This information is not applicable as there is no "training set" in the context of this device's development as described.
    Ask a Question