Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ABX Pentra Calcium AS CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of calcium in human serum, plasma and urine based on a colourimetric method, using the ABX Pentra 400 Clinical Chemistry analyzer. Measurement of calcium is used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba Medical methods on Horiba Medical clinical chemistry analyzers.

    The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision of Horiba Medical methods on Horiba Medical clinical chemistry analyzers.

    The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision of Horiba Medical methods on Horiba Medical clinical chemistry analyzers.

    The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision of Horiba Medical methods on Horiba Medical clinical chemistry analyzers.

    Device Description

    All the reagent, controls and calibrator included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX Pentra Calcium AS CP is an in vitro diagnostic assay for the quantitative in vitro determination of calcium in human serum, plasma and urine based on colourimetric method. It is composed of a monoreagent cassette (79 mL). The reagent is chemical solution with additives.

    The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer.

    This calibrator is provided in ten vials of 3 ml.

    The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml.

    The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annex, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the ABX PENTRA CALCIUM AS CP device and its associated controls and calibrators, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied/General Industry Practice for IVDs)Reported Device Performance (ABX Pentra Calcium AS CP)
    Detection Limit(Not explicitly stated, but lower is better)Serum/Plasma: 0.28 mg/dl; Urine: 0.23 mg/dl
    Limit of Quantitation(Not explicitly stated, but lower is better)Serum/Plasma: 1.54 mg/dl; Urine: 0.64 mg/dl
    Repeatability (CV%)(Typically low CV% for good precision)Serum/Plasma: • Control 1: 0.49% • Control 2: 0.37% • Specimen 1: 0.71% • Specimen 2: 0.40% • Specimen 3: 0.43% Urine: • Control 1: 0.62% • Control 2: 0.76% • Specimen 1: 0.46% • Specimen 2: 0.56% • Specimen 3: 0.37%
    Reproducibility (CV%)(Typically low CV% implying precision over time)Serum/Plasma: • Control 1: 1.44% • Control 2: 1.49% • Specimen 1: 1.56% • Specimen 2: 1.54% • Specimen 3: 1.54% Urine: • Control 1: 1.45% • Control 2: 1.50% • Specimen 1: 1.57% • Specimen 2: 1.57% • Specimen 3: 1.56%
    Measuring Range (Linearity)(Demonstrate linearity within clinical range)Serum/Plasma: 4.0 mg/dl - 18.05 mg/dl (up to 54.15 mg/dl with post-dilution); Urine: 0.64 mg/dl - 18.05 mg/dl (up to 54.15 mg/dl with post-dilution)
    Method Comparison (Correlation with Reference)(High correlation, ideally slope ≈ 1, intercept ≈ 0)Serum/Plasma: Y = 1.00 x + 0.04 (mg/dl), r² = 0.9903 Urine: Y = 0.98 x -0.03 (mg/dl), r² = 0.993
    Matrix Comparison(High correlation between serum and plasma)Y = 1.006x - 0.0022, r² = 0.996 (for serum vs. lithium plasma)
    Calibration Stability(Period for which calibration holds)10 days (for serum/plasma and urine)
    Reagent Stability(Shelf-life and on-board stability)Closed stability: 24 months at 2-8°C; On-board stability: 60 days at 2-8°C

    2. Sample Size Used for the Test Set and Data Provenance

    • Detection Limit & Limit of Quantitation: Not specified for individual samples, but determined according to CLSI (NCCLS), EP17-A protocol.
    • Repeatability (within-run precision):
      • 3 specimens (low, medium, high concentration)
      • 2 controls
      • Each tested 20 times.
      • Data provenance: Not explicitly stated, but likely from laboratory testing conducted by Horiba ABX SAS, France. Retrospective/Prospective is not specified.
    • Reproducibility (total precision):
      • 3 specimens (low, medium, high levels)
      • 2 controls
      • Each tested in duplicate for 20 days (2 series per day).
      • Data provenance: Not explicitly stated, but likely from laboratory testing conducted by Horiba ABX SAS, France. Retrospective/Prospective is not specified.
    • Method Comparison:
      • Serum/Plasma: n = 145 patient samples
      • Urine: n = 143 patient samples
      • Data Provenance: Not explicitly stated beyond "patient samples," but implied to be from a clinical setting, likely in France given the company's location. Retrospective/Prospective is not fully specified, but patient samples are used to compare against a "commercial reagent taken as reference," suggesting a retrospective collection of samples for comparison.
    • Matrix Comparison:
      • n = 32 paired serum and lithium plasma samples (3 samples were altered)
      • Data Provenance: Not explicitly stated beyond "samples," but implied to be from a clinical setting, likely in France. Retrospective/Prospective is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to in vitro diagnostic (IVD) assays for quantitative chemical measurements. The "ground truth" for these studies is typically established by measurements from a well-characterized reference method or a predicate device, rather than expert consensus on images or clinical assessments. The studies performed here compare the device to a "commercial reagent taken as reference" (predicate device K061575) and uses established laboratory protocols (CLSI, Valtec).

    4. Adjudication Method for the Test Set

    Not applicable. The "ground truth" is derived from quantitative measurements by reference methods, not subjective assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret images, sometimes with AI assistance. This document describes a clinical chemistry analyzer and reagents, which do not involve human interpretation of images in the same way.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies described are standalone performance evaluations of the assay system (reagent and instrument). The performance metrics (detection limits, precision, linearity, method comparison, stability) reflect the algorithm's and the physical assay's performance without direct human interpretation influencing the measurement results. The human input is in setting up the system, running controls, and interpreting the numerical output, but the analytical measurement itself is automated.

    7. Type of Ground Truth Used

    • Reference Method/Predicate Device Comparison: For the method comparison study, the "ground truth" was established by measurements from a "commercial reagent taken as reference" (Olympus Calcium Arsenazo reagent, K061575).
    • CLSI/NCCLS Protocols: Other performance parameters (detection limit, quantitation limit, precision, linearity) were evaluated against established CLSI/NCCLS and Valtec protocols, which define acceptable measurement performance characteristics. These protocols inherently define the "truth" in terms of statistical and analytical performance.

    8. Sample Size for the Training Set

    Not applicable. This device is a quantitative in vitro diagnostic reagent and assay system, not an AI/machine learning model that requires a training set in the conventional sense. The development of the reagents and assay parameters is based on chemical and analytical principles, and the performance is validated through the studies listed.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of IVD device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba Medical methods on Horiba Medical chemistry analyzers.

    The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

    Device Description

    All the reagent, controls and calibrator included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative in vitro determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). It is composed of a bi-reagent cassette (R1= 27.5 mL ; R2= 8 mL). Reagents are chemical solutions with additives.

    The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. This calibrator is provided in ten vials of 3 ml.

    The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml.

    The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annex, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for the ABX PENTRA Creatinine 120 CP reagent and associated calibrators/controls, for use on the ABX PENTRA 400 clinical chemistry analyzer.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    ABX PENTRA Creatinine 120 CP (Reagent)
    Detection limit (Serum/Plasma)N/A (Comparison to predicate implies similar or better performance)0.074 mg/dl
    Detection limit (Urine)N/A1.40 mg/dl
    Limit of Quantitation (Serum/Plasma)N/A0.22 mg/dl
    Limit of Quantitation (Urine)N/A2.90 mg/dl
    Accuracy and Precision (Serum/Plasma)N/A (likely defined by an allowable CV% based on clinical standards or predicate)CV. Total < 4.72%
    Accuracy and Precision (Urine)N/ACV Total < 2.06%
    Measuring range (Serum/Plasma)N/A0.22 mg/dl - 18.08 mg/dl
    Measuring range (Urine)N/A2.90 mg/dl - 282.50 mg/dl
    Upper linearity limit (Serum/Plasma)N/A18.08 mg/dl (54.24 mg/dl with automatic post-dilution)
    Upper linearity limit (Urine)N/A282.5 mg/dl (847.50 mg/dl with automatic post-dilution)
    Correlation (Serum/Plasma)N/A (likely a high correlation coefficient with the predicate method)Y = 0.99 x + 0.03 (mg/dl) with r2 = 0.9984
    Correlation (Urine)N/AY = 1.00 x - 0.60 (mg/dl) with r2 = 0.9984
    Calibration stability (Serum/Plasma)N/A (Standard stability duration for similar reagents)3 days
    Calibration stability (Urine)N/A3 days
    Reagent stability (closed)N/A36 months at 2-8°C
    Reagent stability (on-board)N/A19 days
    ABX PENTRA Multical (Calibrator)
    Stability (Closed)N/A (Typical shelf-life for calibrators)24 months at 2-8°C
    Stability (Open, general)N/A (Typical in-use stability for calibrators)8 hours at 15-25°C; 2 days at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A3 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A6 hours at 15-25°C; 1 day at 2-8°C
    ABX PENTRA N Control (Control)
    Stability (Closed)N/A30 months at 2-8°C
    Stability (Open, general)N/A12 hours at 15-25°C; 5 days at 2-8°C; 1 month at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A4 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A8 hours at 15-25°C; 1 day at 2-8°C; 2 weeks at -25 to -15°C
    ABX PENTRA P Control (Control)
    Stability (Closed)N/A30 months at 2-8°C
    Stability (Open, general)N/A12 hours at 15-25°C; 5 days at 2-8°C; 1 month at -25 to -15°C
    Stability (Open, Direct Bilirubin)N/A4 hours at 15-25°C; 8 hours at 2-8°C; 2 weeks at -25 to -15°C
    Stability (Open, Total Bilirubin)N/A8 hours at 15-25°C; 1 day at 2-8°C; 2 weeks at -25 to -15°C
    ABX PENTRA Urine Control L/H (Control)
    Stability (Closed)N/A2 years at 2-8°C
    Stability (Open)N/A30 days at 2-8°C

    Note: The document explicitly states "The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that the devices are substantially equivalent to their respective predicate devices." However, the exact numerical acceptance criteria values themselves (e.g., maximum allowable CV, minimum r2) for each parameter are not explicitly defined in the provided text, but are implied by the reported performance falling within acceptable ranges for substantial equivalence to the predicate devices.

    2. Sample sizes used for the test set and the data provenance

    • Creatinine 120 CP (Test Set):
      • Serum/Plasma Correlation: n = 165
      • Urine Correlation: n = 117
    • Data Provenance: Not explicitly stated, but the company is Horiba ABX SAS, FRANCE, suggesting the data may originate from France or related study sites. The study is for premarket notification, implying prospective testing for device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the text. For in vitro diagnostic devices like this, ground truth is typically established by comparing performance to an established, often predicate, method or reference method, rather than through expert consensus in the way a medical imaging AI might use radiologists.

    4. Adjudication method for the test set

    • This information is not provided as it's not relevant for this type of in vitro diagnostic device performance study. Performance is assessed quantitatively against a reference method or specified analytical limits.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not applicable/not provided. This is an in-vitro diagnostic assay and associated controls/calibrators, not an AI-assisted diagnostic tool that involves human readers interpreting results.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is a standalone diagnostic assay (reagent, calibrator, control) for use on an automated clinical chemistry analyzer (ABX PENTRA 400). The performance data presented (detection limits, accuracy, precision, measuring range, correlation, stability) reflect the standalone analytical performance of the device on the analyzer. There is no "human-in-the-loop" component for the measurement process itself beyond loading samples and reagents.

    7. The type of ground truth used

    • For the ABX PENTRA Creatinine 120 CP reagent, the ground truth for correlation studies was established by comparing its measurements (Y) against those of the predicate device/method (X), as indicated by the correlation equations (e.g., "Y = 0.99 x + 0.03"). This implies a comparative method ground truth (i.e., comparison against a legally marketed, established method).
    • For the calibrators and controls, the "assigned values" of their components serve as their reference values or ground truth, established through rigorous manufacturing and value assignment processes.

    8. The sample size for the training set

    • This information is not provided. This type of in vitro diagnostic device is chemically based and does not typically involve a "training set" in the machine learning sense. The development likely involves extensive R&D and analytical validation, but not a distinct "training set" of patient data for a learning algorithm.

    9. How the ground truth for the training set was established

    • This information is not applicable as there is no "training set" in the context of this device's development as described.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ABX PENTRA Enzymatic Creatinine CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on an enzymatic method using a multi-step approach ending with a photometric end-point reaction. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba Medical methods on Horiba Medical chemistry analyzers.

    The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

    The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

    Device Description

    All the reagent, controls and calibrator included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX PENTRA Enzymatic Creatinine CP is an in vitro diagnostic assay for the quantitative in vitro determination of creatinine in human serum, plasma and urine based on an enzymatic method using a multi-step approach ending with a photometric end-point reaction. It is composed of a bi-reagent cassette (R1= 22 mL ; R2= 8 mL). Reagent is a chemical solution with additives.

    The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. This calibrator is provided in ten vials of 3 ml.

    The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml.

    The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexe, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

    AI/ML Overview

    The provided text describes performance data for the ABX PENTRA Enzymatic Creatinine CP reagent, along with associated calibrators and controls, for use on the ABX PENTRA 400 clinical chemistry analyzer. The study aims to demonstrate substantial equivalence to predicate devices.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents performance data for the ABX PENTRA Enzymatic Creatinine CP reagent, which implicitly serve as acceptance criteria for its clearance.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Detection LimitNot explicitly stated as acceptance criteria, but reported values are key performance indicators.Serum/Plasma: 0.026 mg/dlUrine: 0.66 mg/dl
    Limit of QuantitationNot explicitly stated as acceptance criteria, but reported values are key performance indicators.Serum/Plasma: 0.11 mg/dlUrine: 1.71 mg/dl
    Accuracy and PrecisionTotal CV for Serum/Plasma < 4.12%Total CV for Urine < 4.84%Serum/Plasma CV Total < 4.12%Urine CV Total < 4.84%
    Measuring RangeNot explicitly stated as acceptance criteria, but reported values are key performance indicators.Serum/Plasma: 0.11 mg/dl - 16.95 mg/dlUrine: 3.56 mg/dl - 175 mg/dl
    Upper Linearity LimitNot explicitly stated as acceptance criteria, but reported values are key performance indicators.Serum/Plasma: 16.95 mg/dl (50.85 mg/dl with auto post-dilution)Urine: 175 mg/dl (525 mg/dl with auto post-dilution)
    CorrelationHigh correlation (r²) close to 1 with predicate (or reference) method.Serum/Plasma (n=153): Y = 1.00 x + 0.00 (mg/dl) with r² = 0.999.Urine (n=105): Y = 0.97 x - 0.33 (mg/dl) with r² = 0.9982.
    Calibration StabilitySerum/Plasma: 14 daysUrine: 14 daysSerum/Plasma: 14 daysUrine: 14 days
    Reagent StabilityClosed: 18 months at 2-8°COn-board: 30 daysClosed: 18 months at 2-8°COn-board: 30 days

    For the calibrators (ABX PENTRA Multical) and controls (ABX PENTRA N Control, ABX PENTRA P Control, ABX PENTRA Urine Control L/H), the acceptance criteria relate to their stability (closed and open vial) and their ability to ensure optimal calibration and control of the methods. The document reports specific stability periods for each.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • For Correlation testing (Creatinine CP reagent):
        • Serum/Plasma: n=153
        • Urine: n=105
      • The sample sizes for other performance metrics (detection limit, limit of quantitation, accuracy, precision, measuring range, linearity, stability) are not explicitly stated in the provided text. They are summarized, but the number of unique samples used to establish these values is not detailed.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable to this type of device and study. The device is a clinical chemistry reagent for quantitative determination, not a diagnostic imaging or AI device requiring expert interpretation for ground truth establishment. The ground truth here would be established by a reference method (the predicate device K070383 for Creatinine) or a recognized laboratory standard.

    4. Adjudication Method for the Test Set

    This information is not applicable to this type of device and study. Adjudication methods like 2+1 or 3+1 are typically used for studies involving human interpretation (e.g., radiologists interpreting images) to resolve discrepancies and establish a consensus ground truth. For a quantitative chemical assay, the comparison is against an established reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable to this type of device and study. An MRMC study is relevant for evaluating the impact of AI on human reader performance in diagnostic interpretation, which is not the function of a clinical chemistry reagent.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable in the context of an algorithm in the AI sense. This device is a reagent system that performs quantitative chemical analysis. The performance data presented (accuracy, precision, correlation, etc.) is the standalone performance of the reagent system on the analyzer, without human interpretation as a performance metric.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the Creatinine CP reagent, the ground truth for performance evaluation (specifically correlation) would have been established by comparing the results obtained using the new device against results from a legally marketed predicate device (K070383) or a recognized reference method. The "Y = 1.00 x + 0.00 (mg/dl)" correlation equations suggest a comparison against a reference method, where 'x' represents the reference method's result.

    For the calibrators and controls, the "assigned values" for their components are the ground truth, ensuring optimal calibration and control. These values are established through rigorous testing and standardization, often traceable to international reference materials.

    8. The Sample Size for the Training Set

    This information is not explicitly provided in the summary. For a medical device like a chemical reagent, the "training set" concept is not directly analogous to machine learning. However, the development and optimization of the reagent formulation and methods would involve extensive internal testing using numerous samples to establish the assay's characteristics before formal validation studies. The provided summary focuses on the validation data.

    9. How the Ground Truth for the Training Set Was Established

    As mentioned in point 8, the concept of a "training set" and its "ground truth" establishment isn't directly applicable in the same way as AI/ML devices. For a chemical reagent, the development process would involve:

    • Reference Methods: Using established, highly accurate reference methods or certified reference materials to determine the true values of analytes in samples used during development.
    • Standardization: Developing and refining the assay chemistry to consistently achieve accurate and precise measurements against these reference values.
    • Specifications: Establishing performance specifications based on clinical requirements and comparison to existing, cleared methods.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1