K Number
K161494

Validate with FDA (Live)

Date Cleared
2016-11-15

(167 days)

Product Code
Regulation Number
862.1225
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atellica™ CH Creatinine 2 (Crea 2) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer, Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

The Atellica™ CH Chemistry Calibrator (CHEM CAL) is for in vitro diagnostic use in calibrating the Crea 2 assay using the Atellica™ CH Analyzer.

Device Description

The Atellica CH Creatinine_2 (Crea_2) assay is based on the reaction of picric acid with creatinine in an alkaline medium as described in the original procedure of Jaffe. Creatinine reacts with picric acid in an alkaline medium to produce a red-colored creatinine-picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Creatinine 2 (Crea_2) assay is a modification of the Jaffe method using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because nonspecific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 umol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 umol/L) from each result.

The Atellica CH Chemistry Calibrator (CHEM CAL) is a 1 level lyophilized calibrator product prepared from bovine serum base product.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details for the Atellica CH Creatinine 2 (Crea 2) and Atellica CH Chemistry Calibrator (CHEM CAL) devices:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied/Direct)Reported Device Performance (New Device)
Limit of Blank (LoB)95th percentile, non-parametric approachSerum: 0.03 mg/dL, Urine: 0.35 mg/dL
Limit of Detection (LoD)Not explicitly stated as a criterion, but determined per CLSI EP17-A2Serum: 0.08 mg/dL, Urine: 0.51 mg/dL
Limit of Quantitation (LoQ) - SerumTotal Error (TE) ≤ ±0.1 mg/dL for serumMeasured LoQ: 0.13 mg/dL (supports claim of 0.15 mg/dL)
Limit of Quantitation (LoQ) - UrineTotal Error (TE) ≤ ±1.5 mg/dL for urineMeasured LoQ: 2.57 mg/dL (supports claim of 3.00 mg/dL)
Linearity (Serum/Plasma)p-values of nonlinear terms in quadratic and cubic fit equations are nonsignificant (p ≤ 0.05). If p-value > 0.05, allowable bias ≤ 5% or 0.15 mg/dL (whichever is greater).The assay was deemed linear across the measuring interval (details of specific p-values / bias not provided but stated as meeting criteria).
Linearity (Urine)p-values of nonlinear terms in quadratic and cubic fit equations are nonsignificant (p ≤ 0.05). If p-value > 0.05, allowable bias ≤ 5% or 0.15 mg/dL (whichever is greater).The assay was deemed linear across the measuring interval (details of specific p-values / bias not provided but stated as meeting criteria).
PrecisionNot explicitly stated as acceptance criteria, but evaluated per CLSI EP05-A3. (Results in table below are the "reported performance")(See detailed table below)
InterferencesBias exceeding 10% is considered interference.No interference detected at specified high concentrations for various compounds in serum and urine.
Method Comparison (vs. Predicate)Good agreement with predicate device.Serum: y = 0.98x + 0.00 (r=0.999, N=140) Urea: y = 0.95x + 0.07 (r=0.999, N=109)
Method Comparison (vs. IDMS)Good agreement with IDMS.Serum: y = 0.96x + 0.05 (r=0.999, N=49)
Matrix Equivalency (Plasma vs. Serum)Not explicitly stated, but "demonstrated" by high correlation and near 1:1 regression.Plasma: y = 1.00x – 0.01 (r=1.000, N=58)

Detailed Precision Results:

Sample TypeMean mg/dL (µmol/L)Repeatability SDa mg/dL (µmol/L)Repeatability CVb (%)Within-Lab Precision SDa mg/dL (µmol/L)Within-Lab Precision CVb (%)
Serum0.38 (34)0.01 (0.5)1.70.010 (0.9)2.8
Plasma Pool0.66 (58)0.01 (0.7)1.20.018 (1.6)2.8
Serum Pool1.16 (103)0.01 (0.9)0.80.017 (1.5)1.5
Serum QC1.97 (174)0.02 (1.6)0.90.024 (2.1)1.2
Serum QC6.35 (561)0.04 (3.7)0.70.062 (5.5)1.0
Serum Pool19.31 (1707)0.04 (3.4)0.20.117 (10.3)0.6
Serum Pool26.00 (2298)0.05 (4.7)0.20.145 (12.8)0.6
Urine QC59.62 (5270)0.15 (13.5)0.30.376 (33.2)0.6
Urine QC133.31 (11785)0.33 (29.4)0.20.961 (85.0)0.7
Urine188.61 (16673)0.52 (46.1)0.31.779 (157.3)0.9

2. Sample Size Used for the Test Set and Data Provenance

  • Detection Limit (LoB/LoD):
    • LoB: 4 samples with no analyte, tested 5 times a day for 3 days (60 reps total).
    • LoD: 4 low analyte samples, tested 5 times a day for 3 days (60 reps total).
  • Limit of Quantitation (LoQ):
    • Serum: 10 low samples, 5 replicates each, on 3 reagent lots for 3 days, 5 calibrations per day (75 measurements per reagent lot per sample). Total of 2250 determinations.
    • Urine: Similar setup for urine samples. Total of 2250 determinations.
  • Linearity Study:
    • Serum/Plasma: 12 samples (high and low concentration mixes). 5 replicates measured for each sample.
    • Urine: 10 samples (high and low concentration mixes). 5 replicates measured for each sample.
  • Precision Studies:
    • Controls, serum, and plasma pools: 80 replicates (n=2 replicates, 2 times a day for at least 20 days).
  • Interferences:
    • "Fresh sample pools" containing low or high levels of creatinine in serum and urine. Specific count not given, but refers to testing across these pools.
  • Method Comparison (Predicate):
    • Serum: 140 remnant de-identified samples.
    • Urine: 109 remnant de-identified samples.
    • Data Provenance: "Remnant de-identified samples" implies retrospective patient data. "No patient history information was obtained." The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D personnel.
  • Method Comparison (IDMS):
    • Serum: 49 samples.
  • Matrix Equivalency:
    • 58 matched serum and lithium heparin plasma samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (Creatinine assay system) does not typically involve human expert adjudication for its ground truth. The "ground truth" for a chemical assay is established through highly accurate reference methods or certified reference materials.

  • Ground Truth for Method Comparison: The predicate device, ADVIA Chemistry Enzymatic Creatinine_2 (ECRE_2), served as the comparative "ground truth" for the method comparison study. Additionally, "Isotope Dilution Mass Spectrometry (IDMS)" was used as a reference method for a subset of samples, which is a highly accurate and precise method for determining analyte concentrations and is recognized as a gold standard in clinical chemistry.
  • Ground Truth for Calibration/Standardization: The device uses "IDMS Reference Method" for standardization, and the predicate uses "SRM967" (Standard Reference Material 967, which is a NIST standard for creatinine in human serum, often value-assigned by IDMS). These are highly precise and accurate methods, not typically involving human expert consensus in the same way an imaging device would.

4. Adjudication Method for the Test Set

Not applicable for this type of in-vitro diagnostic device. Ground truth is established by quantitative measurement or comparison to a reference method, not by expert adjudication of human interpretations.


5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in-vitro diagnostic device for quantitative chemical analysis, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a standalone, automated analytical system (Atellica CH Analyzer) that performs the creatinine assay. The performance results reported (LoB, LoD, LoQ, Linearity, Precision, Interference, Method Comparison) represent the standalone performance of the assay on the analyzer. There is no human-in-the-loop component for the measurement of creatinine by this system.


7. The Type of Ground Truth Used

  • Reference Methods:
    • The predicate device (ADVIA Chemistry Enzymatic Creatinine_2) was used as a reference for method comparison.
    • Isotope Dilution Mass Spectrometry (IDMS) was used as a highly accurate reference method for a subset of serum samples. This is considered a gold standard for creatinine measurement.
  • Reference Materials: Standardization mentions "IDMS Reference Method," which implies traceability to primary reference materials. The predicate mentions SRM967, a certified reference material.
  • Defined Protocols: CLSI (Clinical and Laboratory Standards Institute) protocols (EP17-A2, EP05-A3, EP06-A, EP7-A2, EP09-A3, EP28-A3c) were followed for various performance evaluations, which define how to establish performance characteristics against expected statistical and analytical metrics.

8. The Sample Size for the Training Set

Not explicitly mentioned in the provided text as a "training set" in the context of machine learning, which is typically what this question implies. For an IVD assay, method development and initial optimization would involve numerous experiments and samples, but these are typically not referred to as a "training set" in the AI sense. The text focuses on the validation studies performed to demonstrate performance.


9. How the Ground Truth for the Training Set Was Established

As above, the concept of a "training set" with established ground truth as used in AI/ML is not directly applicable here. The development of an IVD assay involves extensive R&D, where analytical methods are refined to accurately measure the analyte. The "ground truth" during this development phase would be established by reference methods or gravimetric/volumetric preparations of known concentrations, similar to how the validation ground truth is established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 15, 2016

SIEMENS HEALTHCARE DIAGNOSTICS, INC. LAURA DUGGAN, PH.D. REGULATORY TECHNICAL SPECIALIST 500 GBC DRIVE, PO BOX 6101 MS 514 NEWARK DE 19711

Re: K161494

Trade/Device Name: Atellica CH Creatinine 2 (Crea 2); Atellica™ CH Chemistry Calibrator (CHEM CAL) Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: II Product Code: CGX, JIT Dated: September 26, 2016 Received: September 27, 2016

Dear Dr. Laura Duggan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161494

Device Name

Atellica CH Creatinine 2 (Crea 2);Atellica™ CH Chemistry Calibrator (CHEM CAL)

Indications for Use (Describe)

The Atellica™ CH Creatinine 2 (Crea 2) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer, Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

The Atellica™ CH Chemistry Calibrator (CHEM CAL) is for in vitro diagnostic use in calibrating the Crea 2 assay using the Atellica™ CH Analyzer.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

10. 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

ASSIGNED 510(K) NUMBER

The assigned 510(k) number is

APPLICANT AND DATE

Laura J. Duggan, Ph. D., RAC Siemens Healthcare Diagnostics Inc. 500 GBC Drive, M/S 514 Newark, DE 19714-6101 Email: laura.j.duggan@siemens.com Phone: 302-631-7654 Fax: 302-631-6299

November 11, 2016

MANUFACTURER

Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown, NY 10591 Registration Number: 2432235

REGULATORY INFORMATION

Regulatory Submission for the Atellica™ CH Creatinine_2 (Crea_2) and Atellica™ CH Chemistry Calibrator (CHEM CAL)

Common Name:Alkaline Picrate, Colorimetry,CreatinineCalibrator, Secondary
Proprietary Name:Atellica CH Creatinine_2(Crea_2)Atellica CH Chemistry Calibrator(CHEM CAL)
ClassificationName:Creatinine Test SystemCalibrator
RegulationNumber:21CFR862.122521CFR862.1150
Classification:Class IIClass II
Product Code:CGXJIT
Panel:Clinical ChemistryClinical Chemistry
Predicate Device:ADVIA Chemistry EnzymaticADVIA Chemistry Calibrator

{4}------------------------------------------------

Creatinine_2 (ECRE_2)(K050374)
(K070727)

DEVICE DESCRIPTION

ATELLICA CH CREATININE 2 (CREA 2)

The Atellica CH Creatinine_2 (Crea_2) assay is based on the reaction of picric acid with creatinine in an alkaline medium as described in the original procedure of Jaffe. Creatinine reacts with picric acid in an alkaline medium to produce a red-colored creatinine-picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Creatinine 2 (Crea_2) assay is a modification of the Jaffe method using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because nonspecific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 umol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 umol/L) from each result.

Reaction Equation

OH. Creatinine + Picric acid Creatinine-Picrate

Serum, lithium heparin plasma and urine specimens may be used. The reagent is stored unopened at 2 - 8 °C and is stable for use on system for 17 days. Calibration is performed every 60 days for a reagent lot or every 6 days for an individual pack.

ATELLICA CH CHEMISTRY CALIBRATOR (CHEM CAL)

The Atellica CH Chemistry Calibrator (CHEM CAL) is a 1 level lyophilized calibrator product prepared from bovine serum base product. The product is stored at 2 - 8 °C. The Atellica CH Chemistry Calibrator is stable for 48 hours at 2 - 8 °C after being opened, rehydrated and securely recapped.

INTENDED USE/INDICATIONS FOR USE

ATELLICA CH CREATININE_2 (CREA_2)


The Atellica™ CH Creatinine 2 (Crea 2) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.


{5}------------------------------------------------

The Atellica™ CH Chemistry Calibrator (CHEM CAL) is for in vitro diagnostic use in calibrating the Crea_2 assay using the Atellica™ CH Analyzer.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Below is a features comparison for the Atellica CH Creatinine_2 (Crea_2) assay and the Chemistry calibrator (CHEM CAL) vs. their predicates:

FeaturePredicate Device:ADVIA Chemistry EnzymaticCreatinine_2 (ECRE_2)(K070727)New Device:Atellica CH Creatinine_2(Crea_2)
Intended Use :For in vitro diagnostic use inthe quantitativedetermination of creatininein humanserum, plasma (lithiumheparin and K2EDTA), andurine on ADVIA Chemistrysystems.The Atellica™ CHCreatinine_2 (Crea_2) assayis for in vitro diagnostic use inthe quantitative determinationof creatinine in human serum,plasma (lithium heparin), andurine using the Atellica™ CHAnalyzer.
Indications for Use:Such measurements areused in the diagnosis andtreatment of renal diseases,and in monitoring renaldialysisSame
Device Technology:Enzymatic reaction ofFossati, Prencipe, and BertiReaction of picric acid withcreatinine in an alkalinemedium as described in theoriginalprocedure of Jaffe.
Sample Type:Serum, Lithium HeparinSerum, Lithium Heparin
plasma, K2-EDTA plasmaand urineplasma, and urine
Reference Interval:Serum/PlasmaMales 0.6 – 1.1 mg/dLFemales 0.5 – 0.8 mg/dLUrineMales 800 – 2000 mg/dayFemales 600 – 1800 mg/daySerum/plasmaMales 0.70-1.30 mg/dLFemales 0.55 -1.02 mg/dLUrine :Males 950 – 2490 mg/dayFemales 600 – 1800 mg/day
Standardization:SRM967IDMS Reference Method
CalibrationFrequency:60 daysSame
Analytical MeasuringInterval:Serum/Plasma:0.10 – 30.00 mg/dLUrine:1.0 – 245.00 mg/dLSerum and plasma:0.15–30.00 mg/dLUrine: 3.00–245.00 mg/dL
Interferences:Bilirubin (Unconjugated) –30 mg/dLBilirubin (Conjugated) – 30mg/dLLipemia (Intralipid®) – 1000mg/dLHemoglobin - 500 mg/dLBilirubin (Unconjugated) – 10mg/dLBilirubin (Conjugated) – 20mg/dLLipemia (Intralipid®) – 500mg/dLHemoglobin - 500 mg/dL
Calibrators:ADVIA Chemistry Calibrator(K050374)Atellica CH ChemistryCalibrator

{6}------------------------------------------------

FeaturePredicate Device:New Device:
ADVIA Chemistry Calibrator(K050374)Atellica CH ChemistryCalibrator (CHEM CAL)
Intended Use :For in vitro diagnostic use inthe calibration of chemistryThe AtellicaTM CH ChemistryCalibrator (CHEM CAL) is for

{7}------------------------------------------------

assays on ADVIA®Chemistry systems.in vitro diagnostic use incalibrating the Crea_2 assayusing the Atellica™ CHAnalyzer.
Calibrator Matrix:Bovine Serum BaseSame
Calibrator Form:LyophilizedSame
Number of CalibratorLevels:OneSame

SUMMARY OF PERFORMANCE TESTING

Assay performance comparison results for the Atellica CH Creatinine_2 (Crea_2) with the Atellica CH Chemistry Calibrator (CHEM CAL) were obtained by processing the appropriate body fluids. Summary statistics for each are provided. These data demonstrate substantial equivalency of the Atellica CH Creatinine 2 (Crea 2) and the Atellica CH Chemistry Calibrator (CHEM CAL) versus the predicate devices. The following data represent typical assay performance.

DETECTION LIMIT

The Limit of Blank (LoB) and Limit of Detection (LoD) were evaluated in accordance with CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation: Approved Guideline.

Assessment of LoB was the 95th percentile of all values (sorted from lowest to highest), using non-parametric approach.

LoB Rank Position = 0.5 +0.95*B, where B=total reps=60; Rank = 57.5

Atellica CH Creatinine_2 (Crea_2) - Limit of Detection Results
LimitProtocolResult
LoB4 samples with no analytewere tested (N=5) for 3days, one run per day, 3reagent lots0.03 mg/dL serum0.35 mg/dL urine

{8}------------------------------------------------

LoD4 low analyte sampleswere tested (N=5) for 3days, one run per day, 3reagent lots0.08 mg/dL serum0.51 mg/dL urine
-----------------------------------------------------------------------------------------------------------------------------------------

The nonparametric approach described in EP17-A2 was followed to determine the Limit of Detection.

LOQ

The Limit of Quantitation (LoQ) for serum was determined as described in CLSI Document EP17-A2. Total Error is calculated using: TE = bias + 2 * SD.

Five replicates of ten low samples were processed on three reagent lots for three days on five calibrations per day, on one instrument for a total of 75 measurements per reagent lot per sample. For serum, the measured LoQ was 0.13 mg/dL in support of the LoQ claim of 0.15 mg/dL for serum and plasma samples. For urine, the measured LoQ was 2.57 mg/dL in support of the LoQ claim of 3.00 mg/dL for urine samples. The LoQ claims for serum and urine are each based a total of 2250 determinations with a total error goal of ≤ ±0.1 mg/dL for serum specimens and ≤ ±1.5 mg/dL for urine specimens.

LINEARITY STUDY

Linearity was evaluated with 12 samples which spanned the assay measuring interval for serum and plasma specimens and 10 samples which spanned the assay measuring interval for urine specimens. Each was prepared by mixing high and low concentration samples across the measurement interval as described in CLSI Evaluation of the Linearity of Quantitative Measurement Procedure (EP06-A). The high sample was prepared by spiking native serum or urine pools with purified creatinine. Low pools were created by diluting serum and urine samples with a 6% BSA or 0.9% saline solution respectively. Five replicates were measured for each sample. The mean of these replicates was used for the calculations.

The assay was considered linear across the measuring interval if the p values of nonlinear terms in the quadratic and cubic fit equations are nonsignificant (p ≤ 0.05). If the p-value is > 0.05, then the allowable bias is ≤ 5% or 0.15 mg/dL, whichever is greater.

PRECISION STUDIES

Precision testing was performed in accordance with CLSI EP05-A3 Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline -

{9}------------------------------------------------

Third Edition. Precision was tested n = 2 replicates, two times a day for at least 20 days for a total of 80 replicates with controls, serum and plasma pools on one instrument. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of EP05-A3. The data are summarized in the following table.

RepeatabilityWithin-Lab Precision
Sample TypenMeanmg/dL (µmol/L)SDamg/dL (µmol/L)CVb(%)SDamg/dL (µmol/L)CVb(%)
Serum800.38 (34)0.01 (0.5)1.70.010 (0.9)2.8
Plasma Pool800.66 (58)0.01 (0.7)1.20.018 (1.6)2.8
Serum Pool801.16 (103)0.01 (0.9)0.80.017 (1.5)1.5
Serum QC801.97 (174)0.02 (1.6)0.90.024 (2.1)1.2
Serum QC806.35 (561)0.04 (3.7)0.70.062 (5.5)1.0
Serum Pool8019.31 (1707)0.04 (3.4)0.20.117 (10.3)0.6
Serum Pool8026.00 (2298)0.05 (4.7)0.20.145 (12.8)0.6
Urine QC8059.62 (5270)0.15 (13.5)0.30.376 (33.2)0.6
Urine QC80133.31 (11785)0.33 (29.4)0.20.961 (85.0)0.7
Urine80188.61 (16673)0.52 (46.1)0.31.779 (157.3)0.9

) = standard deviation

CV = coefficient of variation

INTERFERENCES

CLSI EP7-A2 was followed for the interference testing. The interference study was conducted using a "paired difference worst case scenario" approach where these compounds were spiked into fresh sample pools containing either low or high levels of measurand in serum and urine pools.

Bias is the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. Bias exceeding 10% is considered interference. Dilution studies were conducted to determine the level at which the spiked substance no longer displayed significant interference. Dilution studies were conducted at two analyte concentrations, if both sample pools show significant interference. This study was conducted as needed for both serum pools.

{10}------------------------------------------------

Approximate Concentration (within 15%) of Analytes in Test Pools
AnalyteMatrixLowHigh
CreatinineSerum1.50 mg/dL5.00 mg/dL
CreatinineUrine40.00 mg/dL180.00 mg/dL

No interference was detected at the following analyte concentrations.

Compounds tested for interference in Serum

SubstanceSubstance Test ConcentrationCommon Unit (SI Unit)
Glucose182 mg/dL (10.1 mmol/L)
Ascorbate3.0 mg/dL (171 mmol/L)
Total Protein11.7 g/dL (117 g/L)
Cefoxitin3.75 mg/dL (88 mmol/L)
Cephalexin25 mg/dL (720 mmol/L)
Pyruvate7.5 mg/dL (852 µmol/L)
Acetoacetate20 mg/dL (1959 µmol/L)
Potassium Oxalate500 mg/dL (30 mol/L)
Dopamine (LevaDopa)75 mg/dL (3.8 mol/L)
Albumin6000 mg/dL (60 g/L)
Amikacin8 mg/dL (137 µmol/L)
Caffiene6 mg/dL (308 µmol/L))
Carbamezepine3 mg/dL (127 µmol/L)
Cephradine25 mg/dL (769 µmol/L)
Chloramphenicol5 mg/dL (155 µmol/L)
Chlordiazepoxide1 mg/dL (33.3 µmol/L)
Cimetidine2 mg/dL (2652 µmol/L)
Dextran6000 mg/dL (1500 µmol/L)
Diazepam0.51 mg/dL (18 µmol/L)
Digoxin6.1 ng/mL (7.8 nmol/L)
EDTA200 mg/dL (2 g/L)
Erythromycin6 mg/dL (81.6 µmol/L)
Ethanol400 mg/dL (86.8 mmol/L)
Ethosuximide25 mg/dL (1770 µmol/L)
Furosemide6 mg/dL (181 µmol/L)
Gentamicin1 mg/dL (21 µmol/L)
IgG5000 mg/dL (5000 g/L)
Isopropanol1 g/dL (166 mmol/L)
Lidocaine1.2 mg/dL (51.2 µmol/L)
Nicotine0.1 mg/dL (6.2 µmol/L)
Nortriptyline1000 ng/mL (3797 µmol/L)
Penicillin G(1654)25 U/mL (25000 U/L)
Pentobarbital8 mg/dL (354 µmol/L)
Phenobarbitol10 mg/dL (431 µmol/L)
Phenytoin5 mg/dL (198 µmol/L)
Primidone4 mg/dL (183 µmol/L)
Propoxyphene0.16 mg/dL (4.91 µmol/L)
Sodium Fluoride400 mg/dL (4 g/L)
Theophylline4 mg/dL (222 µmol/L)
Triglyceride2500 mg/dL (17 mmol/L
Urea500 mg/dL (28.3 mmol/L)
Uric acid20 mg/dL (1190 µmol/L)
Valproic Acid50 mg/dL (3467 µmol/L)
Vancomycin10 mg/dL (69 µmol/L)
SubstanceSubstance Test ConcentrationCommon Unit (SI Unit)
6N HCL0.01% HCI
Boric Acid1% w/v
pH 4
pH 9
Ascorbate3.0 mg/dL (171 mmol/L)
Glucose2000 mg/dL (111 mmol/L)
Conjugated Bilirubin50 mg/dL (855 umol/L)
Hemoglobin100 mg/dL (62.2 umol/L)
Acetaminophen200 mg/dL (1324 umol/L)
N-Acetyl cysteine2 mg/dL (123 mmol/L)
Ibuprofen500 mg/dL (24 mol/L)
Acetic Acid25 mL/24 hr collection
6N Nitric Acid0.60%
Ethanol1 g/dL (217 mmol/L)
Gamma Globulin0.5 g/dL (5 g/L)
Human Serum Albumin0.5 g/dL (5 g/L)
Oxalic Acid0.1 g/dL (110 mmol/L)
Sodium carbonate5 g/24 hr collection

Compounds tested for interference in Urine

METHOD COMPARISON

The predicate device selected for the method comparison study was the ADVIA Chemistry Enzymatic Creatinine_2 (ECRE_2) (K070727). Remnant de-identified

..............................................................................................................................................................................

{11}------------------------------------------------

samples were tested. No patient history information was obtained on these samples. Inclusion/exclusion data criteria are not applicable. The study included native and spiked samples to properly span the assay intervals.

These studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel. The personnel conducting the study were laboratory technicians with training similar to personnel who would conduct the tests in a hospital laboratory setting. They were trained on the operation of both the device and the predicate device. A split sample method comparison, following EP09-A3, demonstrated good agreement between the Atellica CH Creatinine_2 (Crea_2) and the predicate ADVIA Chemistry Enzymatic Creatinine_2 (ECRE_2) assay with patient samples.

The table below also summarizes the data for comparison between Isotope Dilution Mass Spectrometry (IDMS) and the Atellica Creatinine 2 (Crea 2) assay. These data demonstrate good agreement with IDMS.

The results across the full assay intervals were analyzed using weighted Deming regression. One replicate of each sample was tested and used in the analysis.

SpecimenTypeComparison Assay(x)NrRegressionEquationSample Range
SerumADVIA Ecre_21400.999y = 0.98x + 0.00mg/dL0.12-28.89mg/dL
UrineADVIA Ecre_21090.999y = 0.95x +0.07mg/dL3.57 - 238.85mg/dL
SerumIsotope Dilution MassSpectrometry (IDMS)490.999y = 0.96x + 0.05mg/dL0.41 - 32.09mg/dL

MATRIX EQUIVALENCY

Serum and lithium heparin plasma equivalency was demonstrated by testing fifty eight matched samples. Some samples were spiked with creatinine to obtain samples spanning the assay measuring intervals. The table below summarizes the weighted Deming linear regression statistics. One replicate of each sample was tested and used in the analysis.

Specimen TypeComparison Assay (x)NrRegression EquationSample Range
Plasma(Lithium heparin)Atellica CH Crea_2 – Serum581.000$y = 1.00x – 0.01$ mg/dL0.68 – 24.87 mg/dL

REFERENCE INTERVAL

Reference intervals for healthy adults were verified on the Atellica CH Analyzer in accordance with CLSI Document EP28-A3c. As with all in vitro diagnostic assays, each laboratory should determine its own reference interval for the diagnostic evaluation of patient results. Consider these values as guidance only.

{12}------------------------------------------------

GroupSpecimen typeReference Intervalcommon unit (SI unit)
Adult malesSerum/plasma10.70 - 1.30 mg/dL (62 - 115 µmol/L)
Adult femalesSerum/plasma²0.55 - 1.02 mg/dL (49 - 90 µmol/L)
Adult malesUrine3950 - 2490 mg/day (8.4 - 22.0 mmol/day)
Adult femalesUrine"600 - 1800 mg/day (5.3 - 15.9 mmol/day)
    1. Burtis CA, Ashwood ER, eds. Tietz Fundamentals of Clinical Chemistry. 5th ed. Philadelphia, PA: WB Saunders Co: 2001:23-25, 422.
    1. Ceriotti F, Boyd JC, Klein G, et al. Reference intervals for serum creatinine concentrations: assessment of available data for global application. Clin Chem. 2008;54(3):559-566.
    1. Junge W, Wilke B, Halabi A, Klein G. Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffé method. Clin Chim Acta. 2004;344(1-2):137-48.
    1. Burtis CA, Ashwood ER, eds. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia, PA: WB Saunders Co; 1999:1809.

CONCLUSION

The Atellica CH Creatinine 2 (Crea 2) and the Atellica CH Chemistry Calibrator (CHEM CAL) are substantially equivalent to the ADVIA Chemistry Enzymatic Creatinine_2 (ECRE 2) (K070727) and the ADVIA Chemistry Calibrator (K050374) in principle and performance based on the similarity of device designs and function demonstrated through method comparison and other performance attributes.

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.