Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233083

    Validate with FDA (Live)

    Date Cleared
    2024-03-09

    (165 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osteon Precision Milled Suprastructure is indicated for attachments in the treatment of partially or fully edentulous jaws for the purpose of restoring function. The Osteon Precision Milled Suprastructures are intended for attachment to a minimum of two (2) abutments.

    The Osteon Milled Suprastructure is indicated for compatibility with the following abutment systems:

    • · Astra Tech Implant System® Multi Base Abutment EV, 4.8mm, max 30°
    • · BioHorizons Multi Unit Abutment, 4.8mm, max 30°
      • · CONELOG® Implant System
    • · Biomet 3 i Multi Unit Abutments, 4.8mm, max 30°
      • · TSXTM Implants
      • · Tapered Screw-Vent Implant
    • · DESS Dental Multi Unit Abutments, 3.4-5.7 mm, 0°
      • · 3i OSSEOTITE®
      • · Astra Tech OsseoSpeed™
      • · Neodent Grand Morse
      • NobelReplace® Trilobe
      • · NobelReplace® Conical
      • · Nobel Brånemark System®
      • · Straumann BLX Implants
    • · DESS Dental Multi Unit Abutments. Angled. 3.4-6.5 mm, max 30°
      • · NobelActive® NobelParallel Conical
      • Straumann® Bone Level
      • · Zimmer Screw Vent® and Tapered Screw-Vent®
    • · Dentium SuperLine® Abutments, 4.5-5.5 mm, max30°
    • · Genesis ACTIVE™ Multi-Unit Abutments, 4.8mm, max 30°
    • · Implant Direct GPS® Angled Abutment, 5.0mm, max 30°
    • KDG Abutments, 4.8mm, max 30°
    • · Keystone Multi Unit Abutment, 4.8mm, 0°
    • · Medentika Multi Unit Abutments, 4.8mm, max 30°
      • · EV Series Dentsply® Implants Astratech Osseospeed®
      • · F Series Nobel Biocare NobelActive® NobelReplace® Conical
      • · H Series Biomet 3i Certain®
      • L Series Straumann Bone Level
      • · N Series Straumann Soft tissue Level
      • · R Series Zimmer Dental Tapered Screwvent®
    • Medentika Multi Unit Abutments, 4.8mm, 0°
      • E Series Nobel Biocare Replace™ Select
      • · I Series Biomet 3i Osseotite®
      • K Series Nobel Biocare™ Branemark
      • · S Series Astra Tech OsseoSpeedTM
      • · T Series Dentsply Friadent® Frialit/Xive®
    • · MegaGen Multi Unit Abutments, 4.8mm, max 30°
    • · Xpeed® AnyRidge® Internal Implant System
    • AnyOne® Internal Implant System
    • AnyRidge® Octa 1 Implant System
    • AnyOne® External Implant System
    • AnyRidge® Octa 1 Implant System
    • AnyOne® Internal Implant System
    • Rescue Internal Implant System
    • MIS Multi-unit Abutments, 4.8mm
      • · C1 Conical Connection Implant System, max 30°
      • · V3 Conical Connection Implant System, max 30°
      • · Internal Hex Implant System, max 30°
      • · Conical Connection, max 30°
    • · Neodent GM Mini Conical Abutment, 4.8 mm, max 30°
    • · Nobel Biocare™ Brånemark Multi Unit Abutment, 4.8 mm, max 17°
    • · Nobel Biocare™ Multi Unit Abutment Plus, 4.8 mm, max 30°
    • · Nobel Biocare™ Multi Unit Abutment, 4.8 mm, max 30°
    • Nobel Biocare™ Multi Unit Abutments for Straumann and Astra Tech System, 4.8 mm, max 30°
    • Nobel Biocare™ Multi Unit Abutments for Astra Tech, Camlog and Ankylos Implant Systems, 4.8 mm, max 30°
    • · Nobel Biocare Xeal Abutments, 4.8 mm, max 30°
    • · OSSTEM Multi Unit Abutment, 4.8mm, max 30°
      • SS SA Fixture Implants
      • SA Implant System
      • · ET US SSS Prosthetic System
    • · Paltop Multi Unit Abutment, 5.0 mm, max 17°
    • · Southern Compact Conical Abutments, 4.8 mm
      • MAX Implant System, 0°
      • · Provata Implant System, max 30°
      • · Deep Conical (DC) Implants, 0°
      • · Piccolo Implants, 0°
      • · External Hex Implants, max 30°
    • Straumann® BLX Screw Retained Abutment, 4.6 mm, max 30°
    • · Straumann® Screw Retained Abutment, 4.6 mm, max 30°
    • · Zimmer Angled Tapered Abutments, 4.5 mm, max 30°
    Device Description

    The Osteon Precision Milled Suprastructures (also referred to as superstructures) are metallic dental restorative devices that are intended to be attached by screw retention to dental implant abutments to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function. These suprastructures attach to dental implant abutments using the prosthetic screws from the original equipment manufacturers (OEM) and are used to support the final multi-unit restoration.

    The Osteon Precision Milled Suprastructure is designed for an individual patient from scans of the patient's dental impression. The suprastructure is manufactured with the aid of Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) technology. All CAD/CAM fabrication is performed by Osteon Medical.

    Osteon Precision Milled Suprastructures facilitate the attachment of both removable and fixed dental prosthesis and hence are categorized as Type A and Type B.

    • Type A: Intended to act as a supporting structure to facilitate the attachment of a removable dental prosthesis and include Primary Bar and Nexus Removable Bar.
    • . Type B: Intended to act as a supporting structure to facilitate the attachment of a fixed dental prosthesis and include Melbourne Bar and Nexus Fixed Bar (including Nexus Hybrid, Nexus Bridge, Micro Nexus, and Nexus Wraparound).
    AI/ML Overview

    The provided FDA 510(k) summary (K233083) describes the acceptance criteria and study for the Osteon Precision Milled Suprastructure.

    1. Table of Acceptance Criteria and Reported Device Performance

    This submission is a 510(k) for an expansion of an existing device (K221019) to add compatibility with new OEM abutment systems and introduce three new Type B Nexus Fixed Bars. The acceptance criteria are primarily based on demonstrating substantial equivalence to the predicate devices in terms of intended use, technological characteristics, and material composition.

    The tables below synthesize the design specifications (acceptance criteria) and the device's performance (reported as meeting these specifications) by demonstrating substantial equivalence to predicate devices.

    Table 1: Comparison of General Characteristics and Materials (summarized for clarity)

    CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (Subject Device)
    Intended UseDental implant abutment for restoring chewing function in partially or fully edentulous jaws, attached to a minimum of two abutments. Specific compatible OEM abutment systems listed.Substantially Equivalent. The subject device shares the same intended use. The primary difference is an expanded list of compatible OEM abutment systems.
    Device MaterialTitanium alloy Ti-6Al-4V (ASTM F136)Meets/Substantially Equivalent. Manufactured from titanium alloy conforming to ASTM F136.
    Design/TechnologyCAD/CAM milling from single milling blanksMeets/Substantially Equivalent. CAD/CAM milling from single milling blanks.
    Design/ConstructionPatient specific/machinedMeets/Substantially Equivalent. Patient specific/machined.
    SterilitySupplied NonsterileMeets/Substantially Equivalent. Supplied Nonsterile (to be sterilized by end-user).
    Prescription/OTCPrescription onlyMeets/Substantially Equivalent. Prescription only.
    Recommended Cleaning & MaintenanceProper oral hygieneMeets/Substantially Equivalent. Proper oral hygiene.

    Table 2: Design Specifications (Acceptance Criteria vs. Reported Performance)

    Design ParametersType A (Removable Prosthesis) Acceptance Criteria (Predicate)Type B (Fixed Prosthesis) Acceptance Criteria (Predicate)Type A (Removable Prosthesis) Reported Performance (Subject Device)Type B (Fixed Prosthesis) Reported Performance (Subject Device)
    Total Cylinders2-102-102-102-10
    Suprastructure Span Between Cylinders (mm)1-30 mm1-30 mm1-30 mm1-30 mm
    Suprastructure Height (mm)3-12 mm3*-22 mm3-12 mm3*-22 mm
    Suprastructure Width (mm)3.4-12 mm3.4**-12 mm3.4-12 mm3.4**-12 mm
    Distal Cantilever Section (mm)0-15 mm0-15 mm0-15 mm0-15 mm
    Cylinder Height (mm)0-4.6 mm0-4.6 mm0-4.6 mm0-4.6 mm
    Cylinder Diameter (mm)4.5-8 mm4.5-8 mm4.5-8 mm4.5-8 mm

    *Note for Suprastructure Height: The Minimum Suprastructure Height for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3*-22 mm, where * indicates variability for this new bar.
    **Note for Suprastructure Width: The Minimum Suprastructure Width for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3.4**-12 mm, where ** indicates variability for this new bar.

    2. Sample size used for the test set and the data provenance

    The submission is a 510(k) for an expansion of an existing device (K221019). The "study" proving the device meets acceptance criteria primarily relies on non-clinical data and demonstration of substantial equivalence to predicate devices.

    • Test set/Sample size: The document does not describe a "test set" in the traditional sense of a clinical or imaging study with a defined sample size for statistical analysis. Instead, it relies on:
      • Reverse engineering analysis of OEM abutments and OEM abutment screws: This was done to confirm compatibility. The sample size for this analysis is not explicitly stated but would involve the specific OEM abutments listed for compatibility with the subject device.
      • Fatigue testing: The document states, "Fatigue testing was not performed since the Subject devices are abutment-borne and are not intended to compensate for angulation in excess of the maximum angulation of OEM angled abutments..." This indicates that engineering analysis and design specifications, rather than physical fatigue testing, were used to demonstrate performance in this regard.
    • Data provenance: Not directly applicable as there isn't a "test set" from patients. The data relies on engineering analysis, existing predicate device data, and compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study requiring expert readers to establish ground truth. Substantial equivalence claims are based on engineering, material, and design comparisons to legally marketed devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no "test set" or diagnostic performance study that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical dental suprastructure, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical dental suprastructure, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" equivalent for this submission is the established safety and effectiveness of the primary predicate device (K221019) and various reference devices, supported by:

    • Compliance with recognized standards: Biocompatibility (ISO 10993-5, ISO 10993-12), sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 14937).
    • Material specifications: Conformance to ASTM F136 for titanium alloy.
    • Engineering analysis and design specifications: Comparison of design features to predicate devices and confirmation of compatibility through reverse engineering.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1