510(k) Data Aggregation

    K Number
    K032525

    Validate with FDA (Live)

    Date Cleared
    2003-09-09

    (25 days)

    Regulation Number
    862.1117
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bayer ADVIA Centaur is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assay analysis on clinical specimens. The system menu will include endocrine, anemia, allergy, reproductive, cardiovascular, oncology, adrenal, bone metabolism, therapeutic drug, and infectious disease assays. All assays are based on chemiluminescent technology.

    Device Description

    The ADVIA Centaur system is a stand-alone, continuous operation, immunochemistry analyzer. The system performs the following functions:

    • . Aspirates and dispenses samples
    • Performs dilutions
    • . Adds reagents
    • Incubates reaction vessels
    • . Separates solid and liquid wastes
    • Measures photon emissions .
    • Performs data reduction
    • Collects and maintains patient demographics and results ●
    AI/ML Overview

    The provided text is a 510(k) Summary for the Bayer ADVIA Centaur, an automated immunoassay analyzer. This type of document focuses on establishing substantial equivalence to a predicate device based on technological characteristics and intended use, rather than conducting new clinical performance studies with acceptance criteria, ground truth, and expert evaluation as would be seen for novel diagnostic devices.

    Therefore, the provided text does not contain the kind of detailed information requested for a clinical study proving a device meets acceptance criteria. Specifically, it lacks:

    • Explicit acceptance criteria: There is no table detailing specific performance metrics (e.g., sensitivity, specificity, accuracy) that the device must meet.
    • A "study" proving acceptance criteria: The document focuses on demonstrating that the ADVIA Centaur is technologically equivalent to its predicate (ACS Centaur V1.0), implying that its performance will be similar, but it does not present a standalone clinical study with an associated test set, ground truth, or expert review.
    • Sample size for test set, data provenance, expert details, adjudication methods, MRMC studies, standalone performance, or training set details. These are all elements of a clinical validation study, which is not presented here.

    Based on the provided text, I can only provide the following, which is largely descriptive of the device and its predicate equivalence rather than a performance study:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not include specific acceptance criteria with quantitative performance targets (e.g., sensitivity, specificity, accuracy, precision ranges, etc.) for the ADVIA Centaur beyond the general equivalence to the predicate device. The "performance" described is largely a comparison of technological features to the predicate, implying similar performance.

    FeaturePredicate Device (ACS Centaur V1.0)ADVIA Centaur V2.5 (Reported Performance/Characteristics)
    Principles of OperationChemiluminescence using magnetic-particle solid phase and chemiluminescent labelSame
    Optical SystemPMT used in photon counting modeSame
    Temperature ControlReactions are controlled at 37°C; Reagent Storage: 4°C to 8°CSame
    Dispense SystemAutomated pipetting of samples and reagents; Precision syringes; Sample Probe: Air pressure fluid sensing, Air pressure disposable tip sensing, Clog detection mechanism; Reagent Probes: No level sense, fluid monitoring during aspirationSame
    Reagent and Sample HandlingSamples: 5 tube racks, up to 180 samples; Tube size selected by barcode; Assay Reagents: Reagent Tray with 30 positions, Refrigeration, Pack contains Solid Phase and Tracer Reagent; Ancillary Reagents: Reagent Compartment with 15 positions, RefrigerationSame
    Test ProcessingRandom Access and Batch; Sample scheduling optimized for throughput; Continuous OperationSame
    Assay Protocols7.5 min, single step; 20 min, single step; 7.5-20 min, two step; 20-20 min, two stepSame
    Human Interface - Data Output17" Color Monitor with GUI; External printers; Serial bi-directional LIS Interface; Audible beeper; Computer LIS Interface; External Modem for Remote Diagnostics InterfaceSame
    Human Interface - Data Input101 key keyboard; Hand-held barcode reader; Stationary barcode scanners for patient samples; Moving Barcode reader for primary reagents; Computer LIS InterfaceSame, with "LIS Software, additional features" (e.g., enhanced LIS)
    Human Interface - Data AnalysisAutomated data reduction; Assay-specific data reductionSame
    QC SoftwareStored control results; L-J plotting; Statistical enhancements; Compatibility with CCD QC ReportingSame, with "Added New Features to QC Functionality" (e.g., likely enhanced, but specifics not detailed)
    SpecimensSerum or plasma, sample cups or primary tubes; Dilutions allowed on a per-assay basis; Capability of Dilution of Samples Requiring PretreatmentSame
    DisposablesSample cups; Reaction cuvettes; Cuvette loading/unloading during run; Reagent 1 & 2 status tracked/displayed; Time to First Result: 15, 30, 60 min (assay-dependent); On-board supplies for 1000 testsSame
    Additional FeaturesN/AAdded diagnostic tools
    SoftwareUnix based GUI, Multiple distributed real-time computing platforms, LIS/LAS communicationSame, except for "additional features" including "New Reports-maintenance, event log, reagent tracking. Screen Saver added. New Exit queue status window."
    Hardware ImprovementsACS Centaur SystemSame, except for enhancements: "New UI Module with increased capacity. Hitachi/Universal rack option added. High Resolution Barcode Scanner (LS4000i) released."

    2. Sample Size for Test Set and Data Provenance

    Not applicable. The document describes a substantial equivalence submission, not a clinical performance study with a distinct "test set" in the context of diagnostic accuracy. The data provenance and sample sizes for any underlying validation studies that led to the predicate's original clearance or the ADVIA Centaur's internal verification are not detailed here.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This information pertains to a clinical validation study where ground truth is established by experts (e.g., radiologists, pathologists). The 510(k) summary focuses on comparing technical specifications.

    4. Adjudication Method

    Not applicable. This is relevant for expert-adjudicated ground truth in clinical studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study assesses how AI intervention affects human reader performance. The ADVIA Centaur is an automated immunoassay analyzer, not an AI-powered image analysis or diagnostic aid designed to assist human readers. It performs tests on its own.

    6. Standalone Performance Study

    Yes, implicitly. The very nature of an automated immunoassay analyzer means it operates in a standalone fashion, providing results without direct human intervention in the analytical process. Its "performance" is demonstrated by its technical specifications and equivalence to the predicate, which implies its ability to accurately measure analytes in samples independently. However, specific standalone quantitative performance metrics (e.g., analytical sensitivity, specificity, imprecision, accuracy against a reference method) are not detailed in this summary; they would be found in detailed validation reports or product labeling.

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically applied in AI/diagnostic imaging studies (e.g., pathology, outcomes data, expert consensus) is not directly applicable here. For an immunoassay analyzer, the "ground truth" for its analytical performance would be established through:

    • Reference methods: Comparisons to established, highly accurate laboratory methods for measuring analytes.
    • Certified reference materials/standards: Testing samples with known, certified concentrations of the analytes.
    • Clinical correlation: Showing that results from the device correlate well with clinical diagnoses or patient outcomes, although this is more for clinical utility than analytical accuracy.

    This 510(k) summary does not specify which of these were used, nor does it present the results of such studies.

    8. Sample Size for the Training Set

    Not applicable. This device is an immunoassay analyzer and does not utilize a "training set" in the machine learning sense. Its operation is based on established chemiluminescent technology and predefined assay protocols, not trainable AI models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device.