Search Filters

Search Results

Found 28 results

510(k) Data Aggregation

    K Number
    K233060
    Manufacturer
    Date Cleared
    2024-06-17

    (266 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Binding assay for the in vitro quantitative determination of folate in erythrocytes (red blood cells, RBC). Folate measurements are used in the diagnosis and treatment of anemia. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

    Device Description

    Elecsys Folate III is a binding assay that makes use of a competitive test principle using a ruthenium labeled folate-binding assay.
    Elecsys Folate III is a binding assay for the in vitro quantitative determination of folate in erythrocytes (red blood cells, RBC). Folate measurements are used in the diagnosis and treatment of anemia. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
    Whole blood treated with anticoagulants (heparin or EDTA) is mixed with ascorbic acid solution and incubated for approximately 90 minutes at 20-25 °C. Lysis of the erythrocytes takes place, with liberation and stabilization of the intracellular folate. The resulting hemolysate sample is then used for subsequent measurement.
    Results are determined via a calibration curve, which is instrument-specifically generated by 2point calibration, and a master curve provided via the cobas link.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Elecsys Folate III device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied / Stated)Reported Device Performance
    Precision"All predefined acceptance criteria was met for the precision experiments." (Specific numerical criteria not explicitly stated in this document but implied to be within acceptable limits as per CLSI guideline EP05-A3.)Repeatability (within-run precision) & Intermediate Precision (within-laboratory precision) (cobas e 801 analyzer)
    SampleMean (ng/mL)Repeatability SD (ng/mL)
    -------------------------------------------------------
    Hemolysate 11525.73
    Hemolysate 22066.14
    Hemolysate 32526.70
    Hemolysate 43638.01
    Hemolysate 560510.7
    Lot-to-lot Reproducibility: "All predefined acceptance criteria was met for the lot-to-lot reproducibility experiment." (Specific data not provided, but confirmed to meet criteria.)
    Analytical SensitivityBased on CLSI EP17-A2 guidelines for Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ).LoB: 45 ng/mL
    LoD: 70 ng/mL
    LoQ: 120 ng/mL
    LinearityMeasurements across the claimed measuring range (120 - 620 ng/mL) must be linear as assessed per CLSI EP06-Ed2.Linearity confirmed to support the measuring range of 120 - 620 ng/mL.
    DilutionData must support instruction for use for samples diluted 1:2.Data supports instruction for use.
    Endogenous Interferences"All predefined acceptance criteria were met" for various endogenous substances including Bilirubin, Intralipid, Biotin, Rheumatoid factors, IgG, IgA, IgM, at specified concentrations, confirming no significant interference. (Specific thresholds for non-interference not provided in text, but implied to be within acceptable limits).No significant interference for:
    Bilirubin: ≤ 29 mg/dL
    Intralipid: ≤ 1500 mg/dL
    Biotin: ≤ 1200 ng/mL
    Rheumatoid factors: ≤ 1000 IU/mL
    IgG: ≤ 1.6 g/dL
    IgA: ≤ 0.4 g/dL
    IgM: ≤ 1 g/dL
    Analytical Specificity/Cross-ReactivityExpected low cross-reactivity with specified compounds. (Specific thresholds for cross-reactivity not provided in text, but implied to be within acceptable limits).Low cross-reactivity:
    Amethopterin: (750 ng/mL) 1.7%
    Aminopterin: (750 ng/mL) 2.0%
    Folinic acid: (750 ng/mL) 2.6%
    Exogenous InterferencesNo interference from 17 commonly used pharmaceuticals and erythropoietin. (Specific thresholds for non-interference not provided in text, but implied to be within acceptable limits).No interference found from 17 commonly and 1 specially used pharmaceutical (erythropoietin) compounds.
    Sample Matrix ComparisonResults within specification, supporting the use of hemolysate prepared from whole blood treated with Na-heparin or K3-EDTA.Results were within specification and support the use of hemolysate prepared from whole blood and treated with Na-heparin or K3-EDTA.
    Method Comparison to PredicateHigh correlation and agreement with the predicate device (Elecsys Folate RBC).Number of samples: 119 (concentrations 132-618 ng/mL)
    Passing/Bablok: y = 1.04x - 14.6, τ = 0.913
    Linear regression: y = 1.03x - 11.0, r = 0.991
    Reagent Stability (On-board)Reagent kits can be stored on-board for up to 16 weeks.Tested on one cobas e 801 analyzer; Elecsys Folate III reagent kits can be stored on-board for up to 16 weeks. (Note: The product comparison table states "2 weeks" for predicate; this indicates an improvement for the candidate device.)
    Calibration Stability (Lot)Calibration for a lot is recommended every 12 weeks; during this period, fresh kits of same lot can be used without re-calibration using the day 0 curve.Tested on one cobas e 801 analyzer. Calibration of an Elecsys Folate III reagent lot is recommended every 12 weeks.
    Calibration Stability (On-board)Reagent epacks can be stored on-board for up to 28 days without a new calibration.Tested on one cobas e 801 analyzer. Elecsys Folate III epacks can be stored on board of the analyzers for up to 28 days without a new calibration.

    2. Sample Sizes and Data Provenance

    • Precision (Repeatability & Intermediate Precision): Not explicitly stated, but typically involves multiple replicates over several days/runs with multiple instruments. Specific sample types are "Hemolysate 1" to "Hemolysate 5".
    • Lot-to-lot Reproducibility: "three reagent lots" were used.
    • Analytical Sensitivity (LoB, LoD, LoQ): Not explicitly stated, but determined according to CLSI EP17-A2, which involves specific numbers of blank and low-concentration samples.
    • Linearity and Dilution: "At least seven concentrations using hemolysate samples" for linearity. Dilution study used "high concentration hemolysate samples".
    • Endogenous Interferences: Not explicitly stated, but "various endogenous substances" were evaluated.
    • Analytical Specificity/Cross-Reactivity: Not explicitly stated.
    • Exogenous Interferences: "17 commonly and 1 specially used pharmaceutical" compounds.
    • Sample Matrix Comparison: "Whole blood samples were drawn into Na-Heparin and K3-EDTA tubes." (Number not specified).
    • Method Comparison to Predicate: 119 samples with concentrations between 132 and 618 ng/mL.
    • Reagent Stability (On-board), Lot Calibration Stability, On-board Calibration Stability: Tested on "one cobas e 801 analyzer".

    Data Provenance: The document states "NON-CLINICAL PERFORMANCE EVALUATION" and refers to CLSI (Clinical and Laboratory Standards Institute) guidelines, which are standard for in-vitro diagnostic device performance studies. The data is internal to Roche Diagnostics, a company with global operations. The specific country of origin for the studies is not stated, but given the submission is to the U.S. FDA, the studies are expected to meet international and U.S. regulatory standards. These are retrospective studies in the context of device development and validation.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This device is an in-vitro diagnostic (IVD) assay that produces quantitative measurements of folate. It does not involve human interpretation of images or other subjective data. Therefore, the "ground truth" for its performance is established by reference methods, calibrated standards, and accurate measurement principles, rather than human expert consensus. No human experts were used to establish ground truth in the way one would for an AI imaging device.

    4. Adjudication Method (Test Set)

    Not applicable, as this is an IVD device producing quantitative results, not an AI imaging device requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an in-vitro diagnostic device, not an AI device assisting human readers with interpreting cases.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies described are standalone performance evaluations of the Elecsys Folate III assay (the "algorithm" in a broad sense for an IVD) without human intervention in the measurement process. The device performs the quantitative determination of folate using its defined binding assay and electrochemiluminescence immunoassay (ECLIA) method.

    7. Type of Ground Truth Used

    The ground truth for evaluating the Elecsys Folate III's performance is established through:

    • Reference materials/standards: For analytical sensitivity (LoB, LoD, LoQ) and linearity, where known concentrations are used.
    • Performance against a predicate method: For method comparison, the results are compared to a previously cleared, established method (Elecsys Folate RBC).
    • CLSI guidelines: Adherence to established scientific and statistical methodologies for validating assay performance, implying well-defined benchmarks for accuracy, precision, and interference.

    8. Sample Size for the Training Set

    Not explicitly stated. For an IVD like the Elecsys Folate III, a traditional "training set" as understood in machine learning is not directly applicable. The "training" for such a system would involve the development and optimization of the reagent formulations, assay parameters, and calibration curve algorithms, using various samples and experiments during the research and development phase. However, a specific training set size with corresponding ground truths is not documented in the same way as an AI algorithm.

    9. How the Ground Truth for the Training Set was Established

    As above, a formal "training set ground truth" isn't directly applicable in the machine learning sense. The "ground truth" during the development and optimization (analogous to training) would have been established through:

    • Known concentrations: Using purified folate standards or spiked samples.
    • Reference methods: Comparing early-stage assay performance against established, often more labor-intensive or gold-standard methods for folate determination.
    • Clinical correlation: (Though "clinical testing" is marked Not Applicable for this 510(k), early development might involve correlation with clinical status or outcomes).
    • Statistical optimization: Adjusting reagent ratios, incubation times, and instrument settings to achieve optimal analytical performance characteristics (sensitivity, specificity, precision, linearity) based on these known values and reference method comparisons.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223590
    Date Cleared
    2023-08-23

    (265 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, lithium heparin plasma, and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

    Folate levels in serum, lithium heparin plasma, and red blood cells are used to assess folate status. The serum folate levels is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency.

    Device Description

    The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum and lithium heparin plasma or red blood cells using the Access Immunoassay Systems.

    AI/ML Overview

    The provided text describes the Beckman Coulter Access Folate Assay, a chemiluminescent immunoassay for the quantitative determination of folic acid levels. The submission is a 510(k) premarket notification for demonstrating substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Study TypeAcceptance CriteriaReported Device Performance
    Method ComparisonR² ≥ 0.90 and slope 1.00 ± 0.12. Estimated bias at concentration corresponding to reference limits suggestive that values have not changed appreciably.R² = 0.99, Slope = 1.04 (95% CI: 1.01 - 1.07), Intercept = 0.081 (95% CI: -0.074 - 0.19). The study met the acceptance criteria. The estimated bias at concentration corresponding to reference limits defined on the predicate system suggest that such values have not changed appreciably on the DxI 9000 analyzer.
    LinearityLinear throughout the analytical measuring interval (2.0 - 24.8 ng/mL).The study met the acceptance criterion, indicating linearity on the DxI 9000 Immunoassay Analyzer throughout the analytical measuring interval (2.0 - 24.8 ng/mL).
    Serum ImprecisionNot explicitly stated as acceptance criteria, but based on typical standards for imprecision: Expected low %CV for higher concentrations and low SD for lower concentrations.Within-laboratory (total) % CV: between 2.2% and 4.4% for Folate concentrations > 2.0 ng/mL. Within-laboratory (total) SD: between 0.10 - 0.21 for Folate concentrations ≤ 2.0 ng/mL. Repeatability (within-run) % CV: between 1.6% and 2.7% for Folate concentrations > 2.0 ng/mL. Repeatability (within-run) SD: between 0.08 - 0.09 for Folate concentrations ≤ 2.0 ng/mL.
    RBC ImprecisionNot explicitly stated as acceptance criteria, but based on typical standards for imprecision: Expected low %CV.Within-laboratory (total) % CV: ranged from 1.9% to 4.9%.
    Limit of Blank (LoB)Claimed LoB of 0.80 ng/mL (1.81 nmol/L).The assay is designed to meet the claimed LoB of 0.80 ng/mL.
    Limit of Detection (LoD)Claimed LoD of 1.0 ng/mL (2.27 nmol/L).The assay is designed to meet the claimed LoD of 1.0 ng/mL.
    Limit of Quantitation (LoQ)Claimed LoQ of
    Ask a Question

    Ask a specific question about this device

    K Number
    K192586
    Device Name
    LIAISON® Folate
    Manufacturer
    Date Cleared
    2020-05-06

    (230 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaSorin LIAISON® Folate assay uses chemiluminescent immunoassay (CLIA) technology for the quantitative determination of Folic acid in human serum. Folic acid measurements are used in the diagnosis and treatment of anemias. Assay results should be used in conjunction with other clinical or laboratory data to assist the clinician in making individual patient management decisions.

    The assay must be performed on the LIAISON® XL Analyzer.

    Device Description

    The LIAISON® Folate assay is a competitive chemiluminescence immunoassay (CLIA) for quantitative determination of Folic acid in serum. During the first incubation, Folic acid is dissociated from its binding protein. After five (5) minutes, a high pH buffer is added to prevent re-association to the binding protein. After five (5) minutes, Folic acid binds to a Folate Binding Protein on the solid phase, which competes with a Folic acid linked to an isoluminol derivative. After a third incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added to initiate a flash chemiluminescent reaction. The light signal is measured by a photomultiplier as relative light units (RLU) and is inversely proportional to the concentration of Folic acid present in calibrators, controls, or samples.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for each performance characteristic in a pass/fail format. Instead, it presents study results which implicitly demonstrate the device's acceptable performance. For clarity, I've inferred common acceptance standards for such in vitro diagnostic assays where explicit criteria aren't given and formatted the reported performance against these.

    Performance CharacteristicAcceptance Criteria (Inferred for IVDs)Reported Device Performance
    Method ComparisonGood agreement with a legally marketed predicate device (e.g., R-value > 0.9, acceptable bias at medical decision levels).Passing & Bablok regression: LIAISON® Folate = (y = 0.96x - 0.61); R = 0.948.
    Bias at 4.41 ng/mL medical decision level: -0.772 ng/mL (95% CI: -1.550 to -0.130 ng/mL).
    Interpretation: The R-value of 0.948 indicates strong correlation with the predicate. The bias at a medical decision level is quantified.
    PrecisionLow within-run, run-to-run, day-to-day, and total variability (low %CV).Combined Lots (Reproducibility): %CVs range from 4.3% to 7.0%.
    Single Lot (Total Within-Lot): %CVs range from 4.6% to 7.6%.
    Interpretation: All reported %CVs are within generally accepted limits for quantitative immunoassays, indicating good precision.
    LinearityThe device should demonstrate linearity across its stated measuring range with a strong correlation (R-value close to 1).Regression equation: Observed Folate = 1.011 (Expected) - 0.147; R = 0.999.
    Interpretation: An R-value of 0.999 demonstrates excellent linearity across the tested range (up to 20 ng/mL).
    Recovery%Recovery typically within 90-110% (or similar range) of the expected value.% Recovery: Values ranged from 97.6% to 110.4% across 7 diluted samples.
    Interpretation: All recovery values fall within the generally accepted 90-110% range, indicating accurate recovery after dilution.
    Analytical Specificity (Cross-Reactivity)Minimal or no significant cross-reactivity with structurally similar compounds or metabolites.Aminopterin: 0.488%
    Phenytoin: 0.000%
    Methotrexate: 0.781%
    Folinic Acid: 1.730%
    Interpretation: Low percentages of cross-reactivity indicate good specificity.
    Analytical Specificity (Interfering Substances)No significant interference from common endogenous or exogenous substances at tested concentrations.No interference observed for all listed substances (Hemoglobin, Bilirubin, Triglycerides, Cholesterol, Albumin, IgG, HAMA, Rheumatoid Factor, various drugs) at the respective tested concentrations.
    Interpretation: The device is robust against common interferents.
    Limit of Blank (LoB)Very low, representing the highest concentration likely to be observed in a blank sample.≤1.2 ng/mL
    Interpretation: A low LoB indicates effective distinction from zero analyte.
    Limit of Detection (LoD)The lowest concentration that can be reliably detected.1.4 ng/mL
    Interpretation: Satisfactory detection capability for the intended use.
    Limit of Quantitation (LoQ)The lowest concentration at which analyte can be accurately quantified with acceptable precision and accuracy.1.6 ng/mL
    Interpretation: The assay can reliably quantify Folate at concentrations starting from 1.6 ng/mL.
    Stability (Reagent Cartridge)Maintains performance over the claimed storage period.Open vial: 6 weeks at 2-8°C
    Interpretation: The device maintains performance for the stated duration.
    Stability (Calibration Curve)Maintains accuracy over the claimed period.Calibration curve: 21 days
    Interpretation: Calibration is stable for 21 days, suitable for routine use.
    TraceabilityTraceable to an internationally recognized standard.Traceable to the WHO IS 03/178 (pg/mL).
    Interpretation: Ensures accuracy and comparability of results with other standardized methods.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set (Method Comparison): 157 human serum samples, spanning the assay range.
    • Test Set (Expected Values/Reference Range): 166 prospectively collected serum samples from apparently healthy U.S. adults (21-59 years old) of mixed ethnic backgrounds (30% Caucasian, 32% African American, 38% Hispanic).
    • Data Provenance: The method comparison study and the expected value study used human serum samples. The expected values study explicitly mentions "United States" for its population, implying the data is from the US and prospectively collected. The nature of the other analytical performance studies (precision, linearity, recovery, etc.) generally involves contrived or spiked samples and do not typically draw from specific patient populations or geographical locations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • General Context: For an in vitro diagnostic (IVD) device like the LIAISON® Folate assay, the "ground truth" for the test set is established by the predicate device (Abbott Laboratories, ARCHITECT Folate, K092740) for method comparison, or by the inherent concentration of the analyte in the samples for analytical performance characteristics (like precision, linearity, etc.).
    • No human "experts" established ground truth for the test set in the way radiologists might agree on an image diagnosis. Instead, the predicate device (an established, cleared medical device) serves as the reference standard for comparative effectiveness.

    4. Adjudication Method for the Test Set:

    • This is not applicable to the analytical performance studies of an in vitro diagnostic assay. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies, particularly for diagnostic imaging or pathology, where human interpretation of results is involved and consensus among experts is needed to define the "true" diagnosis or finding for a given sample/case. Here, the comparison is against an instrument's measurement (predicate device) or against known concentrations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable to this type of device. The LIAISON® Folate assay is an automated in vitro diagnostic device, not an AI-assisted diagnostic tool that humans interpret. There are no "human readers" involved in interpreting the results from this specific device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this entire submission describes the standalone performance of the LIAISON® Folate assay. It is an automated chemiluminescent immunoassay (CLIA) system (LIAISON® XL Analyzer) that quantitatively determines Folic acid in human serum. Its performance characteristics (precision, linearity, accuracy, etc.) are evaluated intrinsically as a standalone device, without human intervention in the result determination process once the sample is loaded.

    7. The Type of Ground Truth Used:

    • Method Comparison: The "ground truth" was established by measurements from the predicate device, the Abbott Laboratories, ARCHITECT Folate (K092740).
    • Other Analytical Performance (e.g., Precision, Linearity, Recovery, Specificity): The "ground truth" was established internally through various experimental designs:
      • Precision: By repeated measurements of samples with known or target concentrations.
      • Linearity: By testing serial dilutions from a high-concentration sample, where the "expected" concentration is mathematically derived from the initial concentration and dilution factor.
      • Recovery: By comparing measured concentrations of diluted samples to their calculated expected concentrations.
      • Analytical Specificity: By comparing results of samples spiked with potential interferents/cross-reactants to unspiked samples.
      • Limits (LoB, LoD, LoQ): Through statistical analysis of repeated measurements of blank and low-level samples.

    8. The Sample Size for the Training Set:

    • For an IVD like the LIAISON® Folate assay, there isn't a "training set" in the context of machine learning. The technology is based on established biochemical principles (chemiluminescence immunoassay) and reagents, rather than an AI/ML algorithm that learns from data. Therefore, this question is not applicable.

    9. How the Ground Truth for the Training Set Was Established:

    • As mentioned above, there is no "training set" for this type of device. The design and validation are based on chemical and biological principles and rigorous experimental testing, not machine learning model training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172201
    Date Cleared
    2018-04-12

    (265 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica™ Folate assay is for in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the Atellica IM Analyzer. Folic acid measurements are used in the diagnosis and treatment of anemias.

    Device Description

    The Atellica™ IM Folate Assay is an immunoassay with the following components: Atellica™ IM Folate Primary Reagent ReadyPack (including Lite Reagent, Solid Phase Reagent, and Folate Binding Protein), Atellica™ IM Folate Calibrator (including Low and High Calibrators), Atellica IM Folate DTT/Releasing Agent (sold separately, including Dithiothreitol and Sodium hydroxide), and Atellica IM RBC Folate (sold separately, including Lyophilized ascorbic acid and RBC Folate Ascorbic Acid Diluent).

    AI/ML Overview

    The provided text is a 510(k) summary for an in vitro diagnostic device (Atellica™ IM Folate Assay), not an AI/ML medical device. Therefore, much of the requested information regarding AI/ML device acceptance criteria and study design (such as multi-reader multi-case studies, expert adjudication, training set ground truth, etc.) is not applicable to this document.

    However, I can extract information related to the acceptance criteria for this diagnostic assay and how its performance was proven.

    Here's a summary based on the provided document, addressing the applicable points:

    Device Name: Atellica™ IM Folate Assay

    Indications for Use: For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the Atellica IM Analyzer. Folic acid measurements are used in the diagnosis and treatment of anemias.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a singular table, but rather presents performance characteristics of the device, implying that the observed performance met internal or regulatory (CLSI) standards. The predicate device (ADVIA Centaur Folate Assay, K010050) serves as the benchmark for substantial equivalence.

    Here's a compilation of key performance characteristics, acting as de-facto acceptance criteria for a new in vitro diagnostic assay, and the reported performance. The "Acceptance Criteria" here are implicitly derived from CLSI guidelines and comparison to the predicate.

    Performance CharacteristicImplicit/Stated Acceptance Criteria (often based on CLSI guidelines or predicate performance)Reported Device Performance (Atellica™ IM Folate Assay)
    Precision (Repeatability CV)Acceptable Coefficient of Variation (CV) for different analyte levels (e.g., typically 0.95) with predicate device.Serum: r = 0.99
    RBC hemolysate: r = 0.93
    Method Comparison (Regression Equation)Slope and intercept close to 1 and 0, respectively, when compared to predicate.Serum: y = 0.94x - 0.01 ng/mL
    RBC hemolysate: y = 1.06x – 2.52 ng/mL
    Detection Limits (LoB, LoD, LoQ)Values demonstrated to be fit for clinical purpose.Serum: LoB 0.19 ng/mL, LoD 0.38 ng/mL, LoQ 0.56 ng/mL
    RBC: LoB 0.00 ng/mL, LoD 0.21 ng/mL, LoQ 0.56 ng/mL
    InterferenceNo significant interference (
    Ask a Question

    Ask a specific question about this device

    K Number
    K161646
    Date Cleared
    2016-11-04

    (142 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diazyme Folate Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of folate in human serum. Folic acid measurements are used in the diagnosis and treatment of anemias. For in-vitro diagnostic use only.

    The Diazyme Folate Calibrator Set is intended for use in the calibration of the Diazyme Folate Assay. For in vitro diagnostic use only.

    The Diazyme Folate Control Set is intended for use as quality controls for the Diazyme Folate Assay. For in vitro diagnostic use only.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes the FDA 510(k) clearance for the Diazyme Folate Assay, Diazyme Folate Calibrator Set, and Diazyme Folate Control Set. It focuses on the regulatory approval process and includes an "Indications for Use" statement.

    Crucially, the document is a regulatory approval letter and not a study report. Therefore, it does not contain the detailed performance data, study design, or methodology information required to answer most of your questions about acceptance criteria verification and study details.

    The information provided only allows for a very limited and generic answer:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Implied): The device must be "substantially equivalent" to legally marketed predicate devices for the stated indications for use.
    • Reported Device Performance: The FDA determined that the device is "substantially equivalent" to legally marketed predicate devices.

    2. Sample size used for the test set and the data provenance
    NOT PROVIDED IN THE DOCUMENT. This document does not detail the specific performance studies that Diazyme Laboratories conducted to demonstrate substantial equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    NOT PROVIDED IN THE DOCUMENT.

    4. Adjudication method for the test set
    NOT PROVIDED IN THE DOCUMENT.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    NOT APPLICABLE. This is an in vitro diagnostic (IVD) assay for measuring folate levels, not an AI-assisted diagnostic imaging device. Therefore, MRMC studies and human reader improvement with AI are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    NOT APPLICABLE. As an IVD assay, its performance is inherently "standalone" in a laboratory setting, meaning it measures a biomarker without human interpretation of complex images or AI algorithms. However, the document does not contain the specific performance data (e.g., accuracy, precision, linearity) that would typically be generated in such a study.

    7. The type of ground truth used
    NOT PROVIDED IN THE DOCUMENT. For an IVD assay, ground truth would typically be established through reference methods, clinical diagnosis, or patient outcomes for the condition being diagnosed/monitored (anemia in this case).

    8. The sample size for the training set
    NOT PROVIDED IN THE DOCUMENT.

    9. How the ground truth for the training set was established
    NOT PROVIDED IN THE DOCUMENT.

    In summary, the provided FDA 510(k) clearance letter confirms regulatory approval based on substantial equivalence but does not contain the detailed study data, methodologies, or acceptance criteria that would typically be found in a clinical study report or a 510(k) submission's performance section. To answer your questions comprehensively, you would need to review the actual 510(k) submission materials, which are not included in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141426
    Manufacturer
    Date Cleared
    2014-10-17

    (140 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Binding assay for the in vitro quantitative determination of folate in human serum. The binding assay is intended for use on Elecsys and cobas e immunoassay analyzers. Folic acid measurements are used in the diagnosis and treatment of anemias.

    Device Description

    The Folate III Assay employs a competitive test principle using natural folate binding protein (FBP) specific for folate. Folate in the sample competes with the added folate (labeled with biotin) for the binding sites on FBP (labeled with a ruthenium complex). Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.

    AI/ML Overview

    Roche Diagnostics' Elecsys Folate III Assay is a quantitative in vitro diagnostic device designed for measuring folate levels in human serum, primarily for the diagnosis and treatment of anemias. The following outlines its acceptance criteria and the studies performed to demonstrate its performance.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Elecsys Folate III Assay are primarily established through comparison to a predicate device, the Roche Elecsys Folate III (K082340). The device demonstrates comparable or improved analytical performance characteristics.

    FeaturePredicate Device: Roche Elecsys Folate III (K082340)Candidate Device: Elecsys Folate III AssayAcceptance Criteria (Implied by equivalence)
    Measuring Range1.50 - 20.0 ng/mL2.0 - 20.0 ng/mLSimilar or improved range
    Analytical Sensitivity
    Limit of Blank (LoB)0.640 ng/mL0.6 ng/mLEqual to or lower than predicate
    Limit of Detection (LoD)1.50 ng/mL1.2 ng/mLEqual to or lower than predicate
    Limit of Quantitation (LoQ)2.0 ng/mL2.0 ng/mLEqual to or lower than predicate
    Analytical Specificity (Cross-reactivity)
    Amethopterin2.3% at 750 ng/mL2.5% at 1500 ng/mLComparable or improved
    Aminopterin2.7% at 750 ng/mL4.4% at 1500 ng/mLComparable or improved
    Folonic acid2.3% at 750 ng/mL0.7% at 1500 ng/mLComparable or improved
    Precision (cobas e 411)
    Within-run (Repeatability)%CV range: 3.7% - 5.8%%CV range: 2.2% - 6.8%Comparable to predicate
    Total (Intermediate)%CV range: 6.1% - 9.4%%CV range: 3.7% - 10.8%Comparable to predicate
    Limitations (Interference)
    Bilirubinunaffected
    Ask a Question

    Ask a specific question about this device

    K Number
    K103716
    Manufacturer
    Date Cleared
    2011-05-12

    (142 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    (1) Elecsys Folate RBC Assay: Binding assay for the in vitro quantitative determination of folate erythrocytes (red blood cells, RBCs). The binding assay is intended for use on Elecsys and cobas e immunoassay analyzers. Measurements obtained by this assay are used in the diagnosis and treatment of anemias.
    (2) Elecsys Folate RBC CalSet: The Elecsys Folate RBC CalSet is used for calibrating the quantitative Elecsys Folate RBC assay on the Elecsys and cobas e immunoassay analyzers.
    (3) Elecsys Folate RBC CalCheck: Elecsys Folate RBC CalCheck is used for the verification of the calibration established by the Elecsys Folate RBC reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostic use.

    Device Description

    (1) The Elecsys Folate RBC assay employs a competitive test principle using natural folate binding protein specific for folate. Manually prepared hemolysate samples are used with this assay. Folate in the sample competes with the added folate (labeled with biotin) for the binding sites on folate binding protein (labeled with ruthenium complex). Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent barcode encoded on the reagent packaging.
    (2) The Elecsys Folate RBC CalSet is a lyophilized product consisting of human serum with folate in two concentration ranges. During manufacture, the analyte is spiked into the matrix. This calibrator is used to calibrate the Elecsys Folate RBC assay.
    (3) The Elecsys Folate RBC CalCheck is a lyophilized product consisting of a hemolysate with folic acid. During manufacture, the analyte is spiked into the matrix. This solution is used to verify the calibration established with the Elecsys Folate RBC CalSet.

    AI/ML Overview

    The provided text describes the Elecsys Folate RBC Test System, but it is a device for in vitro quantitative determination of folate in red blood cells and not an AI or imaging device. Therefore, many of the requested categories related to AI performance, such as MRMC studies, ground truth establishment by experts for image data, or training/test set sample sizes for AI models, are not applicable.

    However, I can extract information related to the device's analytical performance and the studies conducted to demonstrate its substantial equivalence to a predicate device, focusing on the available information.

    Here's the breakdown based on the provided text, adapted for a non-AI diagnostic device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in a table format. Instead, the submission demonstrates substantial equivalence by comparing the performance characteristics of the new Elecsys Folate RBC Assay to the predicate device (Elecsys RBC Folate III Assay) in a feature-by-feature comparison. The reported performance of the new device is shown alongside the predicate's performance. The "acceptance criteria" here are implicitly the predicate device's performance, as the goal is to show substantial equivalence.

    FeatureAcceptance Criteria (Predicate Device Performance)Reported Device Performance (Elecsys Folate RBC Assay)
    Measuring Range46.5 - 620 ng/mL120 ng/mL - 620 ng/mL
    Analytical Sensitivity (LoB)≤ 19.84 ng/mL≤ 20 ng/mL
    Analytical Sensitivity (LoD)≤ 46.5 ng/mL≤ 46.5 ng/mL
    Analytical Sensitivity (LoQ)≤ 62.0 ng/mL≤ 120.0 ng/mL
    Dilution (Min. diluted conc.)The concentration of the diluted sample must be > 310 ng/mL.The concentration of the diluted sample must be >265 ng/mL.
    Precision (Repeatability, Elecsys 2010/cobas e411, Sample 1/Hemolysate 2)Mean 229 ng/mL: SD 12.2 ng/mL; CV 5.3%Hemolysate 2, mean 155 ng/mL: SD 7.73 ng/mL; CV 5.0%
    Precision (Repeatability, Elecsys 2010/cobas e411, Sample 2/Hemolysate 3)Mean 350 ng/mL: SD 17.0 ng/mL; CV 4.9%Hemolysate 3, mean 272 ng/mL: SD 11.2 ng/mL; CV 4.1%
    Precision (Repeatability, Elecsys 2010/cobas e411, Sample 3/Hemolysate 4)Mean 481 ng/mL: SD 25.7 ng/mL; CV 5.3%Hemolysate 4, mean 527 ng/mL: SD 17.1 ng/mL; CV 3.3%
    Precision (Repeatability, Elecsys Modular E170/cobas e 601/602, Hemolysate 2)Not applicable (predicate not tested on these platforms)Hemolysate 2, mean 191 ng/mL: SD 11.5 ng/mL; CV 6.0%
    Precision (Repeatability, Elecsys Modular E170/cobas e 601/602, Hemolysate 3)Not applicableHemolysate 3, mean 258 ng/mL: SD 14.1 ng/mL; CV 5.5%
    Precision (Repeatability, Elecsys Modular E170/cobas e 601/602, Hemolysate 4)Not applicableHemolysate 4, mean 580 ng/mL: SD 12.8 ng/mL; CV 2.2%
    Precision (Intermediate Precision, Elecsys 2010/cobas e411, Sample 1/Hemolysate 2)Mean 229 ng/mL: SD 16.1 ng/mL; CV 7.0%Hemolysate 2, mean 155 ng/mL: SD 12.2 ng/mL; CV 7.9%
    Precision (Intermediate Precision, Elecsys 2010/cobas e411, Sample 2/Hemolysate 3)Mean 350 ng/mL: SD 25.2 ng/mL; CV 7.2%Hemolysate 3, mean 272 ng/mL: SD 16.9 ng/mL; CV 6.2%
    Precision (Intermediate Precision, Elecsys 2010/cobas e411, Sample 3/Hemolysate 4)Mean 481 ng/mL: SD 34.6 ng/mL; CV 7.2%Hemolysate 4, mean 527 ng/mL: SD 24.8 ng/mL; CV 4.7%
    Precision (Intermediate Precision, Elecsys Modular E170/cobas e 601/602, Hemolysate 2)Not applicableHemolysate 2, mean 191 ng/mL: SD 12.5 ng/mL; CV 6.5%
    Precision (Intermediate Precision, Elecsys Modular E170/cobas e 601/602, Hemolysate 3)Not applicableHemolysate 3, mean 258 ng/mL: SD 15.1 ng/mL; CV 5.9%
    Precision (Intermediate Precision, Elecsys Modular E170/cobas e 601/602, Hemolysate 4)Not applicableHemolysate 4, mean 580 ng/mL: SD 19.7 ng/mL; CV 3.4%
    Analytical Specificity / Cross-ReactivityAminopterin 2.7%, Folinic acid 2.3%, Amethopterin 2.3%Same (Implies similar or acceptable cross-reactivities, though specific percentages are not re-listed for the new device)
    InterferencesUnaffected by icterus, lipemia, biotin (up to 86.1 nmol/L or 21 ng/mL), IgG, IgA. No interference from rheumatoid factors up to 1000 IU/mL. No interference from 18 common pharmaceuticals or human erythropoietin.Similar interference study reported ("Same") with an added precautionary statement for high protein samples causing protein gel. Specific interference limits are not re-listed but implied to be similar or improved.
    Expected Values (Whole Blood Folate)American Journal of Clinical Nutrition Expected = 4.6 - 34.8 ng/mL (all ages & male/female)Expected = 209-640 (2.5th - 97.5th percentile) (from hemolysate sample)
    Expected Values (RBC Folate)Not separately specified for predicate in this sectionExpected = 499-1504 ng/mL (2.5th - 97.5th percentile) (folate in erythrocyte fraction)

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a substantial equivalence comparison study against a legally marketed predicate device, the Elecsys RBC Folate III Test System (K082340). This type of study demonstrates that the new device is as safe and effective as the predicate device by showing that it has similar technological characteristics and performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a single "test set" sample size in the way an AI model validation might. Instead, it details various analytical performance studies with different "samples" for each attribute:

    • Precision (Repeatability & Intermediate Precision): Tested on multiple "Hemolysate" or "Sample" controls at different concentration levels. For each level, measurements were taken to determine SD and CV. The exact number of replicates or runs is not specified in the summary but is typically substantial for precision studies (e.g., multiple runs over several days).
    • Analytical Sensitivity (LoB, LoD, LoQ): Determined using analytical methods, likely involving dilutions of folate-containing samples. The specific sample count for these determinations is not provided.
    • Interferences: Tested using various interfering substances and pharmaceuticals. The number of samples/donors tested for each interference is not specified, but typically involves spiking known concentrations into samples and assessing recovery.
    • Expected Values: Determined from a "Whole Blood Folate (from hemolysate sample)" study and an "RBC Folate (folate in erythrocyte fraction)" study. The sample size for these population studies is not provided in the summary.

    Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. For in vitro diagnostic device submissions, such data is typically generated in a controlled laboratory setting (prospective) and may involve samples from different geographical regions, but this is not specified here.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is not an AI/imaging device where expert consensus is used to establish ground truth for a test set. This device quantifies a biomarker (folate in RBCs). The "ground truth" for its analytical performance is established through reference methods and internal validation procedures, not human interpretation or adjudication by experts.

    4. Adjudication Method for the Test Set

    Not applicable. As this device measures a quantitative biomarker, there is no "adjudication" in the sense of resolving discrepancies in human interpretation or labeling. Analytical performance is typically verified against established laboratory standards and reference methods.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device used for interpretation or diagnosis by human readers, so an MRMC study is irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone device performance study. The Elecsys Folate RBC Test System performs the quantitative measurement of folate in RBCs automatically on immunoassay analyzers (Elecsys and cobas e platforms) without human-in-the-loop interpretative performance. The performance metrics (measuring range, sensitivity, precision, interference) are all measures of the algorithm and system's analytical capability.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance relies on analytical reference methods and established laboratory principles. For example:

    • For analytical sensitivity (LoB, LoD, LoQ): Determined using standard statistical methods and known concentrations of analyte.
    • For precision: Assessed by running replicate measurements on control samples.
    • For accuracy/comparison to predicate: The predicate device itself (Elecsys RBC Folate III Assay) serves as the "reference" for demonstrating substantial equivalence. The traceability of the new device is stated as "Reference method is Folate III (Application on the E2010)", meaning its measurements are compared to the predicate's measurements. The predicate's traceability is to the "Elecsys Folate II assay."

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a "training set." The system's underlying principles are based on known biochemistry (competitive binding assay using folate binding protein) and not data-driven learning from a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of in vitro diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092740
    Date Cleared
    2010-03-05

    (178 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARCHITECT Folate assay is a chemiluminescent microparticle Folate Binding Protein assay for the quantitative determination of folate in human serum and plasma on the ARCHITECT i System. Folate measurements are used in the diagnosis and treatment of megaloblastic anemia.
    The ARCHITECT Folate Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of folate in human serum and plasma.
    The ARCHITECT Folate Controls are for the verification of the accuracy and precision of the ARCHITECT i System when used for the quantitative determination of folate in human serum and plasma.

    Device Description

    The ARCHITECT Folate assay is a two-step assay for the quantitative determination of folate in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology. Two pre-treatment steps mediate the release of folate from endogenous folate binding protein.
    In Pre-Treatment Step 1, the sample and Pre-treatment Reagent 2 (Dithiothreitol or DTT) are aspirated and dispensed into a reaction vessel (RV). In Pre-Treatment Step 2, an aliquot of sample/Pre-Treatment Reagent 2 mixture is aspirated and dispensed into a second RV. Pre-Treatment Reagent 1 (potassium hydroxide or KOH) is then added. An aliquot of the pre-treated sample is transferred into a third RV, followed by the addition of Folate Binding Protein (FBP) coated paramagnetic microparticles and assay specific diluent. Folate present in the sample binds to the FBP coated microparticles. After washing, pteroic acid-acridinium labeled conjugate is added and binds to unoccupied sites on the FBP coated microparticles. Pre-Trigger and Trigger Solutions are then added to the reaction mixture; the resulting chemiluminescent reaction is measured as relative light units (RLUs). An inverse relationship exists between the amount of folate in the sample and the RLUs detected by the ARCHITECT i optical system.

    AI/ML Overview

    The provided text describes a 510(k) submission for the ARCHITECT Folate assay, a medical device for quantitatively determining folate in human serum and plasma. The study conducted is a non-clinical performance comparison between the new ARCHITECT Folate assay and its predicate device, the AxSYM Folate assay.

    Here's an analysis of the acceptance criteria and study as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    PrecisionSubstantially equivalent to predicateDemonstrated substantially equivalent performance
    LinearitySubstantially equivalent to predicateDemonstrated substantially equivalent performance
    InterferencesSubstantially equivalent to predicateDemonstrated substantially equivalent performance
    Method ComparisonGood correlation with predicateCorrelation coefficient of 0.898 for serum samples
    BiasUnderstandable and attributable to standardization differencesGeneral bias due to standardization differences (predicate uses gravimetric preparations, investigational assay standardized to WHO Serum Folate International Standard (IS) 03/178)

    Note: The document states "The ARCHITECT Folate assay demonstrated substantially equivalent performance to the AxSYM Folate assay with correlation coefficients of 0.898 for serum samples." The acceptance criteria for "substantially equivalent" in areas like precision, linearity, and interferences are not explicitly quantified with numerical thresholds in the provided text. They are implied by the claim of substantial equivalence to the predicate device.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample size used for the test set in the comparison study.

    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. It is a non-clinical performance study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this type of study. The ground truth for a quantitative diagnostic assay is typically established through a reference method or standardization to an internationally recognized standard. In this case, the investigational assay was standardized to the WHO Serum Folate International Standard (IS) 03/178. Human expert consensus is not the method for establishing ground truth for quantitative laboratory tests like this.

    4. Adjudication Method for the Test Set

    This information is not applicable to this type of study. Adjudication methods (like 2+1 or 3+1) are typically used in studies where human readers are interpreting images or making subjective diagnoses, which is not the case for a quantitative immunoassay comparison. The comparison is likely done statistically between measured values.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a non-clinical performance evaluation of a quantitative diagnostic assay, not an AI-assisted human reading study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, this was a standalone performance study of the ARCHITECT Folate assay. The document describes the assay's mechanism and then presents its performance characteristics (precision, linearity, interferences, method comparison) against a predicate device. There is no mention of human-in-the-loop performance testing.

    7. The Type of Ground Truth Used

    The ground truth for the investigational assay was established by standardization to the WHO Serum Folate International Standard (IS) 03/178. For the comparison with the predicate, the predicate device's results serve as a comparative "truth," which were themselves based on gravimetric preparations of folic acid.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning. This is a traditional immunoassay device, not an AI/ML-based device. Therefore, a training set as understood in AI/ML is not applicable. The assay is "standardized" rather than "trained."

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no "training set" in the machine learning sense for this device. The assay was standardized to accurately recover WHO Serum Folate International Standard (IS) 03/178. This standardization process involves calibrating the assay's response to known concentrations of the analyte based on this international standard.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091332
    Date Cleared
    2009-10-02

    (149 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tosoh AIA-PACK RBC FOLATE is designed for in vitro diagnostic use only for the quantitative measurement of red blood cell folate (RBC FOLATE) in whole blood (Heparin or EDTA) samples on TOSOH AIA system analyzer. Measurement of RBC Folate is used in the diagnosis and treatment of anemia.

    Device Description

    The Tosoh AIA-PACK RBC FOLATE is a competitive enzyme immunoassay which, after sample hemolysis and sample pretreatment, is performed entirely within the AIA-PACK. First, sample hemolysis is performed to achieve complete hemolysis of erythrocytes and deconjugation to monoglutamate in the whole blood sample using sample hemolyzing reagents (containing ascorbic acid). After sample hemolysis, sample pretreatment is performed to release folate from endogenous binding proteins in the hemolysate sample using sample pretreatment reagents (containing sodium hydroxide and dithiothreitol). Folate present in the pretreated test sample competes with enzyme-labeled folate for a limited number of binding sites on a fluorescein labeled bovine folate binding protein which then binds to anti-FITC (fluorescein isothiocyanate) antibody immobilized on magnetic beads. The beads are washed to remove the unbound enzyme-labeled folate and are then incubated with a fluorogenic substrate, 4-methylumbelifery| phosphate (4MUP). The amount of enzyme labeled folate that binds to the beads is inversely proportional to the folate concentration in the test sample. A standard curve using a range of known standard concentrations is constructed and unknown folate concentrations are calculated using this curve.

    AI/ML Overview

    The Tosoh AIA-PACK RBC FOLATE is a device designed for the quantitative measurement of red blood cell folate (RBC FOLATE) in whole blood samples, used in the diagnosis and treatment of anemia. The device's performance was evaluated through various studies to demonstrate substantial equivalence to a predicate device (Bayer K010050, ADVIA Centaur and ACS: 180 Folate Immunoassay).

    The document provides performance data for several criteria: Recovery, Dilution, Linearity, Precision, Correlation, Specificity, and Limit of Detection (LoD)/Limit of Quantitation (LoQ). Acceptance criteria are not explicitly stated as distinct numerical targets but are implied by the reported performance and the overall conclusion of substantial equivalence based on established CLSI protocols.

    Here's a breakdown of the acceptance criteria and the study results:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied by CLSI Protocols and Performance)Reported Device Performance
    RecoveryPercent RecoveryGenerally 90-110% (standard for spiked recovery studies)Ranged from 94.2% to 102.7% (demonstrated good recovery)
    DilutionPercent Recovery (after serial dilution)Generally 90-110% (standard for dilution linearity studies)Ranged from 100.0% to 114.7% (demonstrated good linearity upon dilution)
    LinearityLinear range with ±10% differenceLinear from 0.5 to 40.0 ng/mL, within ±10%Demonstrated linearity from 0.5 to 40.0 ng/mL, within ±10% difference. The claimed linearity was 0.62 to 24.0 ng/mL.
    PrecisionWithin-run CV (%)Typically 0.95 for strong correlation0.99
    Slope (Deming Regression)Close to 1 (e.g., 0.9-1.1)0.959 (0.922 to 0.996)
    Y-intercept (Deming Regression)Close to 0- 3.890 (-13.512 to 5.732)
    Correlation (vs. Alternate Method)Correlation Coefficient (r)Typically > 0.90 for good correlation0.87 (Deming Regression), 0.86 (Linear Regression with hematocrit correction)
    Slope (Deming Regression)Close to 1 (e.g., 0.9-1.1)1.06 (0.950 to 1.168), 1.00 (0.897 to 1.107 with hematocrit correction)
    Y-intercept (Deming Regression)Close to 0-19.554 (-44.640 to 5.532), -26.60 (with hematocrit correction)
    SpecificityCross-reactivity (%) for common interferencesLow percentage (e.g.,
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    CGN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    (1) Elecsys Folate III: Binding assay for the in vitro quantitative determination of folate in human serum and plasma. The binding assay is intended for use on Elecsys and cobas e immunoassay analyzers. Measurements obtained by these devices are used in the diagnosis and treatment of anemias. For in vitro diagnostic use.

    (2) Elecsys RBC Folate Hemolyzing Reagent: Elecsys RBC Folate Hemolyzing Reagent is used together with the Elecsys Folate III assay for the quantitative determination of folate in erthrocytes (RBC (red blood cell) folate). Binding assay for the in vitro quantitative determination of folate in human serum and plasma. The binding assay is intended for use on Elecsys and cobas e immunoassay analyzers. Measurements obtained by these devices are used in the diagnosis and treatment of anemias.

    (3) Elecsys Folate III CalSet: Elecsys Folate III CalSet is used for calibrating the quantitative Elecsys Folate III assay on the Elecsys and cobas e immunoassay analyzers.

    (4) Elecsys Folate III CalCheck: Elecsys Folate III CalCheck is used for the verification of the calibration established by the Elecsys Folate III reagent on the indicated Elecsys and cobas e immunoassay analyzers.

    (5) Elecsys PreciControl Anemia: Elecsys PreciControl Anemia is used for the quality control of specified Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers.

    Device Description

    (1) The Elecsys Folate III assay employs a competitive test principle using natural folate binding protein (FBP) specific for folate. Folate in the sample competes with the added folate (labeled with biotin) for the binding sites on FBP (labeled with ruthenium complex). Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent barcode.

    (2) The Elecsys RBC Folate Hemolyzing Reagent is a 0.2% ascorbic acid solution with which whole blood treated with anticoagulants (heparin or EDTA) is diluted. After incubation the erythrocytes are lysed and intracellular folate is liberated and stabilized. The hemolysate is then used as a prediluted sample for subsequent measurement in the Elecsys Folate III assay.

    (3) The Elecsys Folate III CalSet is a lyophilized product consisting of human serum with folic acid in two concentration ranges. During manufacture, the analyte is spiked into the matrix. This calibrator is used to calibrate the Elecsys Folate III assay.

    (4) Elecsys Folate III CalCheck is a lyophilized product consisting of human serum with folic acid. During manufacture, the analyte is spiked into the matrix. This solution is used to verify the calibration established with the Elecsys Folate III CalSet.

    (5) The Elecsys PreciControl Anemia is a lyophilized control serum based on human serum matrix in three concentration ranges. The controls are used for monitoring the accuracy and precision of the Elecsys Ferritin, Folate III, and Vitamin B12 immunoassays.

    AI/ML Overview

    The Elecsys Folate III Test System, including Elecsys Folate III, Elecsys RBC Folate Hemolyzing Reagent, Elecsys Folate III CalSet, Elecsys Folate III CalCheck, and Elecsys PreciControl Anemia, is designed for the in vitro quantitative determination of folate in human serum, plasma, and erythrocytes. This system is intended for use in the diagnosis and treatment of anemias on Elecsys and cobas e immunoassay analyzers.

    The acceptance criteria for the Elecsys Folate III and Elecsys RBC Folate Hemolyzing Reagent assays were established through comparison studies against their respective predicate devices (Elecsys Folate II Assay and Elecsys RBC Folate Hemolyzing Reagent (K051292)). The primary method for proving the device meets these criteria is through Method Comparison studies using Passing-Bablok and Linear Regression analysis.

    Here's the breakdown of the acceptance criteria and performance:

    Acceptance Criteria CategoryAcceptance Criteria (Predicate Device)Reported Device Performance (Elecsys Folate III)
    Elecsys Folate III Assay
    Measuring Range0.600 - 20.00 ng/mL1.5 - 20.0 ng/mL
    Precision (Total CV for Human Samples 10 ng/mL)7.0%5.0%
    Analytical Sensitivity (Lower Detection Limit)0.6 ng/mLLimit of Detection = ≤1.5 ng/mL
    Interferences (Icterus)recovery within ± 10% of initial value for bilirubin
    Ask a Question

    Ask a specific question about this device

    Page 1 of 3