Search Results
Found 8 results
510(k) Data Aggregation
(127 days)
Luminescence immunoassay for the in vitro diagnostic quantitative measurement of active free Estradiol, an estrogenic steroid, in saliva and serum. Measurements obtained by this device may be used in the diagnosis and treatment of various hormonal sexual disorders and can be used to evaluate ovarian function. This test is not intended for assessing placental function in complicated pregnancy.
Luminescence immunoassay (LIA) based on the competition principle. An unknown amount of antigen present in the sample and a fixed amount of enzyme labeled antigen compete for the binding sites of the antibodies coated onto the wells. After incubation the wells are washed to stop the competition reaction. After addition of the luminescence substrate solution the intensity of the luminescence measured is inversely proportional to the amount of the antigen in the sample. Results of samples can be determined directly using the standard curve.
The IBL Estradiol LIA is a luminescence immunoassay for the in vitro diagnostic quantitative measurement of active free Estradiol in saliva and serum.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are not explicitly stated as distinct pass/fail thresholds in the provided document. Instead, the document presents performance characteristics and comparisons to established methods. The "acceptance" is implied by the FDA's substantial equivalence determination. We can infer the functional performance criteria from the reported values and comparisons.
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Normal Ranges in Saliva | Establish distinct ranges for different physiological states. | Premenopausal (n=28): Follicular phase 0.6 - 10.4 pg/mL; Mid-cycle Peak 4.5 - 21.2 pg/mL; Luteal phase 0.5 - 10.8 pg/mL.Postmenopausal (n=5): < 3.2 pg/mL. Males (n=40): < 3.4 pg/mL. |
| Comparison to Published Saliva RIA | High correlation and close agreement with established methods. | r² = 0.97 with a regression formula of Y = 0.96 * RIA + 0.55 (n=50 saliva samples). Range measured: 0-51 pg/mL (RIA), 0-55 pg/mL (LIA). |
| Comparison to Commercial Serum RIA | High correlation and close agreement with established methods. | r² = 0.95 with a regression formula of Y = 1.00 * RIA - 4.83 (n=60 serum samples). Range measured: 22 - 448 pg/mL (RIA), 2 - 444 pg/mL (LIA). |
| Serum to Saliva Comparison | Demonstrate a relationship between serum and saliva levels. | R² = 0.712 with a regression formula of serum (pg/mL) = 43.491 * (saliva) - 4.680. |
| Cross-Reactivity | Minimal cross-reactivity with structurally similar or common interfering substances. | 17β-Estradiol: 100.0%Estrone: 0.222%Estriol: 0.138%Other substances (Corticosterone, Dexamethasone, Cortisone, Progesterone, Testosterone, Prednisolone): ≤ 0.01% |
| Analytical Sensitivity (LoD) | Low limit of detection for accurate measurement of low concentrations. | Saliva: 0.3 pg/mL (Mean signal (Zero-Standard) - 2SD)Serum: 7.3 pg/mL (Mean signal (Zero-Standard) - 2SD) |
| Functional Sensitivity | Concentration at which CV is <20%. | Saliva: 0.78 pg/mL (Mean Conc. <20% CV)Serum: 8.0 pg/mL (Mean Conc. <20% CV) |
| Precision (Intra-Assay & Inter-Assay) | Low coefficient of variation (CV) at different concentrations. | Saliva Intra-Assay: CVs from 3.7% to 7.9% (n=10)Saliva Inter-Assay: CVs from 5.9% to 13.9% (n=10)Serum Intra-Assay: CVs from 3.1% to 7.4% (n=10)Serum Inter-Assay: CVs from 4.3% to 9.3% (n=10) |
| Linearity | Measured values should be proportional to expected concentrations across a range of dilutions. | Saliva: Recoveries from 80% to 114% for various dilutions.Serum: Recoveries from 83% to 117% for various dilutions. |
| Recovery | Measured values should be close to added values. | Saliva: Recoveries from 92% to 128% for various spiked concentrations in 3 different saliva samples.Serum: Recoveries from 80% to 96% for various spiked concentrations in 3 different serum samples. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Normal Ranges in Saliva:
- Premenopausal Women: 28 women (month profiles, so multiple samples per woman), age 19-43.
- Postmenopausal Women: 5 women, age 42-62.
- Males: 40 males, age 20-63.
- Data Provenance: Not explicitly stated, but likely prospective for the normal range establishment as saliva samples were collected daily "until first day of bleeding" or as single samples for postmenopausal women and males. No country of origin is specified, but the manufacturer is based in Germany, suggesting European or international data.
-
Comparison of Serum and Saliva (to published/commercial RIA):
- Saliva Comparison: 50 saliva samples.
- Serum Comparison: 60 serum samples.
- Data Provenance: Not explicitly stated, but likely retrospective or a combination of prospective/retrospective for convenience from adult healthy individuals. No country of origin is specified.
-
Serum to Saliva Comparison (using IBL Estradiol LIA):
- Paired Samples: Not explicitly stated, but derived from "saliva and serum pairs collected at the same time." The number of pairs is not provided.
- Data Provenance: Not explicitly stated, but likely prospective for this specific comparison.
-
Cross-Reactivity, Analytical Sensitivity, Functional Sensitivity, Precision, Linearity, Recovery:
- The sample sizes for these analytical performance studies are embedded within the tables (e.g., n=10 for precision at each level). These are typically laboratory studies using spiked samples, controls, and calibration materials rather than patient samples for the direct "test set" definition often associated with diagnostic device clinical trials.
- Data Provenance: Laboratory validation studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Normal Ranges: The "ground truth" here is the observed ranges in the healthy populations studied. There is no mention of external experts establishing a ground truth for these ranges; they are empirical observations from the study participants themselves.
- Comparison Studies: The "ground truth" for the comparison studies are the results obtained from the "published procedure that used a modification in the handling of saliva for a typical RIA test" and a "commercially available RIA test kit." These are established reference methods. There is no mention of individual expert adjudicators for these data points.
4. Adjudication Method for the Test Set
Not applicable in the context of this device. The comparisons are against established laboratory methods, not subjective diagnoses requiring adjudication by human experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is an in vitro diagnostic immunoassay, not an imaging or diagnostic device requiring interpretation by human readers. Therefore, an MRMC study and effects of AI assistance are not relevant to this submission.
6. Standalone Performance Study
Yes, the entire submission describes the standalone performance of the IBL Estradiol LIA. The comparisons to RIA methods ("Comparison of Serum and Saliva," "Serum to Saliva Comparison") are designed to demonstrate its performance relative to existing technologies, but the listed metrics (Normal Ranges, Cross-reactivity, Sensitivity, Precision, Linearity, Recovery) represent the intrinsic performance of the IBL Estradiol LIA device itself.
7. Type of Ground Truth Used
- Normal Ranges: Empirical observation from defined healthy populations.
- Comparison Studies: Results from established, commercially available, or published reference immunoassay methods (RIA).
- Analytical Performance (Cross-Reactivity, Sensitivity, Precision, Linearity, Recovery): These are based on established laboratory analytical methods using controlled samples, calibrators, and known concentrations (e.g., spiked samples).
8. Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning or AI development. For an immunoassay, the equivalent would be the samples used to develop and optimize the assay's reagents, protocols, and calibration curves. This information is not provided in detail. The normal range studies and comparison studies would be considered part of the validation and testing, rather than a "training set" in the modern AI sense.
9. How the Ground Truth for the Training Set Was Established
As there is no distinct "training set" discussed in the context of AI/ML, this question is not directly applicable. However, for the development of the immunoassay itself, the "ground truth" for calibrators and controls would be established through rigorous analytical chemistry techniques, often traceable to international standards, and verified by expert analytical chemists during the assay development phase. The document implicitly relies on standard immunoassay development and validation practices.
Ask a specific question about this device
(41 days)
The IBL Estradiol ELISA test kit is intended for the in vitro determination of estradiol, an estrogenic steroid, in human serum and plasma. Estradiol measurements are used in the diagnostics and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy.
Not Found
This FDA 510(k) clearance letter dated November 24, 1997, for the KMI Diagnostics, Inc. 17 ß-Estradiol (ELISA) test kit, does not contain the detailed information necessary to complete a comprehensive table of acceptance criteria, device performance, and study characteristics.
The document primarily focuses on establishing "substantial equivalence" to a legally marketed predicate device, allowing the manufacturer to market the device. It states the "Indications for Use" but does not provide specific performance metrics, study designs, or ground truth methodologies.
Therefore, I cannot fully address the request with the information provided.
However, I can extract the following limited information:
Indications for Use (from the document):
"The IBL Estradiol ELISA test kit is intended for the in vitro determination of estradiol, an estrogenic steroid, in human serum and plasma. Estradiol measurements are used in the diagnostics and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy."
Based on the typical regulatory requirements for ELISA kits at the time, and the "substantial equivalence" claim, the acceptance criteria would generally refer to demonstrating comparable performance to an existing, legally marketed device (the predicate device). This would involve analytical performance studies rather than clinical efficacy studies on patient outcomes.
Here is what a hypothetical response might look like if the requested information were present in the document, along with an explanation of why I cannot fill it based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric (Hypothetical) | Acceptance Criteria (Hypothetical) | Reported Device Performance (Not Available in Document) |
|---|---|---|
| Analytical Sensitivity (Detection Limit) | Should be ≤ X pg/mL | [Information not provided in the document] |
| Analytical Specificity (Cross-reactivity) | Cross-reactivity to interfering substances ≤ Y% | [Information not provided in the document] |
| Precision (Intra-assay CV) | CV ≤ Z% | [Information not provided in the document] |
| Precision (Inter-assay CV) | CV ≤ A% | [Information not provided in the document] |
| Accuracy (Correlation with reference method) | R² ≥ 0.95 with established reference method | [Information not provided in the document] |
| Linearity of Dilution | % Recovery within B% of expected value | [Information not provided in the document] |
| Measuring Range | X to Y pg/mL | [Information not provided in the document] |
Explanation: The provided 510(k) summary letter does not include specific analytical performance metrics or the acceptance criteria for these metrics. The letter confirms "substantial equivalence" but does not detail the data supporting this claim.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not available in the document.
- Data Provenance: Not available in the document (e.g., country of origin, retrospective/prospective).
Explanation: The document does not describe the specific study data used to demonstrate equivalence, including the number of samples or their origin. For ELISA kits, this would typically involve analytical samples (spiked samples, patient samples compared to a reference method).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not applicable/Not available in the document.
- Qualifications of Experts: Not applicable/Not available in the document.
Explanation: For an in vitro diagnostic (IVD) device like an ELISA kit, ground truth is typically established through a reference method (e.g., GC-MS, LC-MS/MS, or a well-established, previously cleared ELISA). It usually does not involve expert readers or adjudication in the same way an imaging or pathology device might.
4. Adjudication method for the test set
- Adjudication Method: Not applicable/Not available in the document.
Explanation: As mentioned above, this type of device's "ground truth" is analytical, not based on expert review requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study Done?: No.
- Effect Size: Not applicable.
Explanation: This is an immunoassay (ELISA) kit, not an AI-powered diagnostic imaging or pathology device that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance Done?: Yes, the device itself is a standalone laboratory test.
Explanation: The ELISA kit is a laboratory test run by trained technicians. Its performance is measured directly, without a human-in-the-loop interaction in the diagnostic interpretation once the assay is complete.
7. The type of ground truth used
- Type of Ground Truth: Not specified in the document.
Explanation: Typically for an ELISA, the ground truth would be established by comparing results from the investigational device to a recognized "gold standard" or reference method (e.g., Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS/MS), or a well-validated predicate immunoassay). The document does not specify which was used.
8. The sample size for the training set
- Sample Size for Training Set: Not relevant or available in the document.
Explanation: ELISA kits are developed using biochemical principles and reagents, not machine learning algorithms that require "training sets" in the AI sense. While method development involves optimization, it's not analogous to an AI training set.
9. How the ground truth for the training set was established
- How Ground Truth for Training Set Established: Not relevant or available in the document.
Explanation: As above, this concept of a "training set" and its "ground truth" is not directly applicable to the development and validation of an ELISA assay in the context of this 1997 regulatory submission.
Ask a specific question about this device
(30 days)
The Access Estradiol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of estradiol levels in human serum, using the Access Immunoassay System.
The Access® Estradiol reagents and the Access® Immunoassay Analyzer comprise the Access® Immunoassay System for the quantitative determination of Estradiol levels in human serum.
This document describes the validation of the ACCESS® Estradiol assay, an immunoassay for determining estradiol levels in human serum.
Acceptance Criteria and Reported Device Performance
The acceptance criteria for the device are implied by the "Summary of Studies" section, where specific performance metrics are reported. There are no explicit "acceptance criteria" stated as thresholds; rather, the study presents the observed performance and concludes that the device is substantially equivalent to a predicate device.
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Precision (Within Run) | CV within acceptable clinical ranges for estradiol measurement | Ranges from 19.7% CV (at 39 pg/ml) to 3.1% CV (at 2167 pg/ml) |
| Precision (Total) | CV within acceptable clinical ranges for estradiol measurement | Ranges from 20.0% CV (at 39 pg/ml) to 5.0% CV (at 2918 pg/ml) |
| Accuracy (Spiking Recovery) | Recovery between 80% and 120% | Ranging from 80% to 115% |
| Accuracy (Dilution Recovery) | Mean recovery close to 100% | Mean recoveries of 108% to 112%. Average dilution recovery at 1:2 was 106%. For samples >500 pg/ml, average % recovery was 107%. |
| Correlation with Predicate Device | Strong positive correlation (e.g., r > 0.95) | r = 0.98 |
| Analytical Sensitivity (Lowest Detectable Level) | Detectable level at or below clinically relevant thresholds | 20 pg/ml |
| Analytical Specificity (Cross-Reactivity with Estrone Sulfate) | Negligible or no detectable cross-reactivity | Not detectable when spiked at 3600 pg/ml |
| Analytical Specificity (Cross-Reactivity with Estriol Sulfate) | Low percentage apparent cross-reactivity | Apparent cross-reactivity of 0.004% when spiked at 10,000,000 pg/ml |
Study Details
-
Sample sizes used for the test set and the data provenance:
- Precision studies: The number of samples for precision studies is not explicitly stated, but typically involves multiple replicates of several control levels or patient samples. The data provenance is not specified (e.g., country of origin, retrospective/prospective).
- Accuracy (Spiking recovery): Ten serum samples were used. Data provenance not specified.
- Accuracy (Dilution recovery): 2 patient samples and 2 patient pools were used, along with 10 patient samples with estradiol >500 pg/ml. Data provenance not specified.
- Correlation: 103 samples were run. Data provenance not specified.
- Analytical Sensitivity: Implies samples were used to determine the lowest detectable level. Number of samples not specified.
- Analytical Specificity: S0 calibrator was spiked with estriol sulfate and estrone sulfate. Number of samples/spiked conditions not explicitly detailed beyond the two substances.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is an in vitro diagnostic (IVD) device (immunoassay). The "accuracy" and "sensitivity" of such devices are typically established against reference methods or through spiked samples with known concentrations, rather than through expert consensus on images or clinical conditions. Therefore, the concept of "experts establishing ground truth for the test set" in the context of interpretation of results (like in image analysis) does not directly apply here. The "ground truth" for quantitative measurements like estradiol is the actual concentration determined by highly accurate methods (e.g., mass spectrometry, or certified reference materials) or by the known concentration of spiked analytes. The document does not specify details about these reference methods or expert panel for ground truth. -
Adjudication method for the test set:
Not applicable (N/A) for this type of quantitative immunoassay study. Adjudication methods like 2+1 or 3+1 are typically used for qualitative or interpretive assessments, such as in image reading where multiple experts might disagree on findings. Here, results are quantitative values compared to known or reference values. -
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable (N/A). This is an immunoassay, not an AI-assisted diagnostic tool involving human readers interpreting cases. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, this is implicitly a standalone study. The performance characteristics (precision, accuracy, sensitivity, specificity, correlation) are reported for the ACCESS® Estradiol assay system itself (reagents + analyzer) as a device, independent of human interpretation or intervention in the measurement process beyond operating the instrument and following procedures. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth used here consists of:- Known concentrations: For spiking recovery studies (estradiol spiked into serum samples) and analytical specificity studies (estrone sulfate, estriol sulfate spiked into calibrator).
- Reference measurements: The Abbott IMx® Estradiol Kit served as a comparator for the correlation study, implying it was considered a widely accepted and validated method for measuring estradiol. For precision and dilution recovery, the ground truth is the inherent concentration of estradiol in the samples, measured multiple times or diluted to known proportions.
-
The sample size for the training set:
Not applicable (N/A) as this is a traditional immunoassay validation report, not one describing a machine learning or AI-driven device that requires a distinct "training set." The development of the assay itself involves optimization and characterization over many samples, but these are not formally called "training sets" in the context of this type of document. -
How the ground truth for the training set was established:
Not applicable (N/A). See point 7.
Ask a specific question about this device
(121 days)
The Chiron Diagnostics ACS:180 Estradiol-6 II Assay is for the quantitative determination of estradial in serum and to aid in assessing ovarian function and in the monitoring of follicular development using the Chiron Diagnostics ACS:180 automated chemiluminescence systems.
Not Found
This is a difficult request to fulfill accurately. The provided text is a 510(k) clearance letter from the FDA for a medical device (Chiron Diagnostics ACS:180 Estradiol-6 II Assay). While it mentions that the device is "substantially equivalent" to previously marketed devices and permits marketing, it does not contain the detailed study information, acceptance criteria, or performance data that would be typically found in a clinical study report or a more comprehensive summary of safety and effectiveness.
The document primarily focuses on the regulatory clearance process and confirms that the device can be marketed. It does not provide the specifics of the underlying studies that demonstrated performance or met acceptance criteria.
Therefore, I cannot extract the requested information directly from the provided text. To answer your questions, I would need a different type of document, such as:
- A Clinical Study Report (CSR)
- A Summary of Safety and Effectiveness Data (SSED), often found in FDA device approvals (though less common for a 510(k) without major new technology).
- A detailed 510(k) submission summary that includes performance data.
Based on the provided text, I can only state that:
- Acceptance criteria and reported device performance: Not explicitly stated in terms of quantitative values or specific assays. The letter only broadly implies that performance was deemed "substantially equivalent" to predicate devices.
- Sample size, data provenance: Not mentioned.
- Number of experts and qualifications: Not mentioned.
- Adjudication method: Not mentioned.
- MRMC comparative effectiveness study: Not mentioned. This device is an immunoassay, so an MRMC study is highly unlikely to be relevant.
- Standalone performance study: The nature of this immunoassay suggests it would primarily be evaluated in a "standalone" fashion (i.e., algorithm/device only, producing a quantitative result). However, the specific details of such a study are not provided.
- Type of ground truth: For an immunoassay, the "ground truth" would typically be established through highly accurate reference methods or other well-validated laboratory methods. This is not detailed in the provided text.
- Sample size for training set: Not mentioned. This device is a pre-AI era immunoassay; the concept of a "training set" in the modern AI sense is not applicable.
- How ground truth for training set was established: Not applicable in the AI sense.
In summary, the provided document is a regulatory clearance letter and does not contain the detailed technical and clinical study information required to answer your questions.
Ask a specific question about this device
(26 days)
Ask a specific question about this device
(42 days)
The ACCESS® Estradiol assay is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of Estradiol levels in human serum using the ACCESS® Immunoassay System.
The ACCESS® Estradiol reagents and the ACCESS® Immunoassay Analyzer comprise the ACCESS® Immunoassay System for the quantitative determination of Estradiol levels in human serum.
The provided document is a 510(k) summary for the ACCESS® Estradiol assay, which is an in-vitro diagnostic device (IVDD). The questions appear to be tailored for AI/Machine Learning (ML) medical devices, which typically require different types of studies and acceptance criteria than IVDDs.
Given the nature of the device (an immunoassay for quantifying Estradiol levels) and the document (510(k) submission from 1997), many of the requested fields are not applicable or the information is not provided in the same format as for AI/ML device studies.
However, I will extract and describe the available information as best as possible, indicating when a question is not applicable to this type of device or when the information is not present.
1. Table of Acceptance Criteria and Reported Device Performance
For an IVDD product like the ACCESS® Estradiol assay, "acceptance criteria" are typically defined by performance specifications and comparisons to a predicate device, rather than explicit thresholds of diagnostic accuracy (e.g., AUC, sensitivity, specificity) against a ground truth in a clinical setting. The "performance" refers to the analytical performance of the assay itself.
| Acceptance Criteria (Implied from Study Goals) | Reported Device Performance (ACCESS® Estradiol assay) |
|---|---|
| Precision: Demonstrate acceptable reproducibility of results. | Within-run CV: 11.6% (at 117 pg/ml) to 2.7% (at 3041 pg/ml) Total Imprecision CV: 21.1% (at 72 pg/ml) to 5.2% (at 1456 pg/ml) |
| Accuracy (Spiking Recovery): Recover spiked analyte within an acceptable range. | Recovery: 92% to 106% (mean recovery of 105% ± 7% SD for dilution recovery) |
| Accuracy (Dilution Recovery): Recover diluted analyte within an acceptable range. | Mean recovery: 105% with a standard deviation of 7% |
| Correlation with Predicate Device: Strong linear correlation with the legally marketed predicate device. | Correlation (r): 0.958 Linear Regression: y = -39 + 0.983x (compared to Abbott IMx® Estradiol Kit) |
| Analytical Sensitivity (LOD): Lowest detectable level distinguishable from zero. | Lowest detectable level: 23 pg/ml (with 95% confidence) |
| Analytical Specificity: Minimal cross-reactivity with structurally similar compounds. | Estrone sulfate: Not detectable at 3600 pg/ml Estriol sulfate: Apparent cross-reactivity of 0.04% at 10,000,000 pg/ml |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Correlation Study: 214 samples were used for comparison between the ACCESS® Estradiol assay and the Abbott IMx® Estradiol Kit.
- Spiking Recovery: 10 serum samples.
- Dilution Recovery: 18 patient samples.
- Analytical Sensitivity: Implied use of Calibrator S0 (zero calibrator) multiple times to determine the limit of detection, but a specific number of samples/replicates is not given.
- Analytical Specificity: Not specified, but involved spiking into S0 calibrator.
- Precision Studies: Number of samples not explicitly stated, but typically involves multiple replicates of control materials or pooled patient samples run over several days/runs.
- Data Provenance: The document does not explicitly state the country of origin for the samples. Given the applicant's address in Chaska, MN, USA, it is highly probable the samples were from the USA.
- Retrospective or Prospective: Unspecified, but for analytical performance studies of IVDDs, these are typically laboratory-based studies using samples collected specifically for method validation or archived samples, rather than prospective clinical trials in the sense of AI/ML devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable in the context of diagnostic accuracy for an AI/ML device. For an IVDD, the "ground truth" for analytical studies is typically established by:
- The known concentration of an analyte in a spiked sample.
- The calculated expected concentration in a diluted sample.
- The results from a well-established, legally marketed predicate device (as in the correlation study).
- Known pure reference materials for specificity.
- There are no "experts" in the sense of human readers/interpreters establishing a ground truth for imaging or clinical diagnosis here.
4. Adjudication Method for the Test Set
- Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are common in AI/ML studies where multiple human readers interpret data to reach a consensus ground truth. For an IVDD's analytical performance studies, this concept does not apply.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study is not relevant for an immunoassay kit. This device directly measures a biomarker concentration and does not involve human readers interpreting complex data that could be augmented by AI.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- Yes, in spirit. The performance described (precision, accuracy, correlation, sensitivity, specificity) is the standalone performance of the assay system (reagents + analyzer). There is no "human-in-the-loop" component in the direct measurement process of this IVDD; the analyzer outputs a quantitative result. The human then interprets that numerical result, which is distinct from human readers interpreting images or clinical data where AI might assist.
7. The Type of Ground Truth Used
- Reference Method Comparisons / Known Concentrations:
- Correlation Study: The results from the Abbott IMx® Estradiol Kit, a legally marketed predicate device, served as the comparative reference (analogous to a ground truth for comparison).
- Spiking Recovery: Known amounts of estradiol spiked into serum.
- Dilution Recovery: Expected concentrations based on dilution of patient samples.
- Analytical Sensitivity/Specificity: Defined concentrations of the analyte or interferent.
8. The Sample Size for the Training Set
- Not Applicable / Not Provided for this type of device. Immunoassays are not "trained" like AI/ML algorithms. The assay's parameters (e.g., antibody concentrations, incubation times, calibrator values) are developed and optimized during product development. The concept of a "training set" for an AI algorithm does not apply here. Calibrators are used to establish a standard curve for each run, which is different from a "training set" for an AI model.
9. How the Ground Truth for the Training Set was Established
- Not Applicable. As explained in point 8, there is no "training set" in the AI/ML sense. For the calibrators included with the assay, their "ground truth" (i.e., their assigned estradiol concentration) would have been established through a rigorous value assignment process, often involving reference methods or gravimetric preparation of highly pure standards. This is part of the manufacturing process for IVDDs.
Ask a specific question about this device
(60 days)
Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma.
Competition principle. Total duration of assay: 18 minutes, 37 °C.
· 1st incubation (9 minutes): By incubating the sample (50 uL) with an estradiol-specific biotinylated antibody (65 uL), an immunocomplex is formed, the amount of which is dependent upon the analyte concentration in the sample.
·2nd incubation (9 minutes): After addition of streptavidin-coated microparticles (35 µL) and an estradiol derivative labeled with a ruthenium complex* (65 uL), the still-vacant sites of the biotinylated antibodies become occupied, with the formation of an antibody-hapten complex. The entire complex becomes bound to the solid phase via interaction of biotin and streptavidin.
· The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
·Results are determined via a calibration curve which is instrumentspecifically generated by 2-point calibration and a master curve provided via the reagent bar code.
** Tris(2,2'-bipyridy))ruthenium(II) complex (Ru(bpy)" 3)
The provided text describes a 510(k) summary for the Boehringer Mannheim Elecsys® Estradiol Assay, comparing it to a predicate device, the Enzymun-Test® Estradiol Assay (K916132). This submission focuses on establishing substantial equivalence based on performance characteristics.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state formal "acceptance criteria" with pass/fail thresholds. Instead, it presents performance characteristics of the new device (Elecsys® Estradiol) and directly compares them to the predicate device (Enzymun-Test® Estradiol). The implicit acceptance criterion is that the performance of the Elecsys® Estradiol Assay is comparable to or better than the predicate device.
Given this, I will construct a table comparing the performance of the Elecsys® Estradiol Assay with the Enzymun-Test® Estradiol Assay, as this constitutes the 'reported device performance' against an implicit 'acceptance' of being equivalent to the predicate.
| Feature | Elecsys® Estradiol Performance | Enzymun® Estradiol Performance (Predicate) | Implicit Acceptance Criteria (Comparison to Predicate) |
|---|---|---|---|
| Precision | Comparable precision (within-run and total) | ||
| Within-Run (Mean) | Low: 7.7, Mid: 44.0, High: 277.6, High: 1217.0 (pg/mL) | Low: 8.2, Mid: 53.3, High: 3021 (pg/mL) (Note: Different levels reported for Elecsys vs. Enzymun) | - |
| Within-Run (%CV) | Low: 3.0, Mid: 2.7, High: 3.3, High: 2.7 (Note: Different levels reported for Elecsys vs. Enzymun) | Low: 14.3, Mid: 16.1, High: 9.1 | - |
| Total (Mean) | Low: 9.0, Mid: 44.0, High: 277.6, High: 1217.0 (pg/mL) | Low: 10.3, Mid: 53.3, High: 3021 (pg/mL) | - |
| Total (%CV) | Low: 9.0, Mid: 9.0, High: 10.0, High: 10.0 (Note: Different levels reported for Elecsys vs. Enzymun) | Low: 16.1, Mid: 16.1, High: 10.3 | - |
| Lower Detection Limit | 10 pg/mL | 10 pg/mL | Equal or lower detection limit |
| Linearity | 10-4600 pg/mL (with a deviation from a linear line of ±10%) | 10-1300 pg/mL (with a deviation from a linear line of ±10%) | Comparable linearity range, potentially wider |
| Method Comparison (Vs Enzymun-Test Estradiol) | High correlation (r value > 0.97) and small bias | ||
| Least Squares | y = 1.027x + 6.32; r = 0.995; SEE = 31.03; N = 64 | y = 0.902x + 42.94; r = 0.976; SEE = 91.96; N = 74 | - |
| Passing/Bablok | - | y = 0.962x + 12.51; r = 0.976; SEE = 38.71; N = 74 | - |
| Interfering Substances (No interference at) | Bilirubin: 64.5 mg/dL; Hemoglobin: 1.0 g/dL; Lipemia: 1250 mg/dL; Biotin: 20 ng/mL | Bilirubin: 25 mg/dL; Hemoglobin: 200 mg/dL; Lipemia: 1500 mg/dL; Biotin: 24 ng/mL | Comparable or improved resistance to interference |
| Specificity (% Cross-reactivity) | Low cross-reactivity, comparable to predicate or better | ||
| 17 hydroxy-Progesterone | < 0.5 | < 0.001 | - |
| Ethisterone | < 0.5 | --- | - |
| Norethindrone-Acetate | < 0.5 | --- | - |
| 2-Methoxy-estradiol | < 1.0 | 0.07 | - |
| 6-Hydroxy-estradiol | < 600 | 62 | - |
| Estriol | < 1.0 | 0.3 | - |
| Prednisolone | < 0.25 | --- | - |
| Danazol | <0.1 | 0.0006 | - |
| Testosterone | <1.0 | 0.03 | - |
| Di-Hydro-Testosterone | < 1.0 | --- | - |
| 5-Androsten-3B-17B-Diol | < 0.4 | --- | - |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Method Comparison:
- Least Squares: N = 64
- Passing/Bablok: N = 74
- Precision: The "N" value for precision testing is 10 for each reported level (Low, Mid, High). This represents the number of replicates used to calculate the mean and %CV for precision for both within-run and total precision.
- Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is a clinical chemistry assay, not an imaging device requiring expert interpretation. The "ground truth" for method comparison and performance evaluation in such assays is typically established by comparing the new method to a reference method (in this case, the predicate Enzymun-Test Estradiol Assay) or a more definitive method like ID-GC/MS for standardization.
- No experts (like radiologists) were used to establish ground truth in the context of human interpretation, as the device doesn't involve image analysis.
- Assay Standardization: The document mentions "Assay Standardization: ID-GC/MS." This indicates that Isotope Dilution Gas Chromatography/Mass Spectrometry (ID-GC/MS), a highly accurate and precise reference method, was used to standardize both the Elecsys® Estradiol Assay and the predicate Enzymun-Test® Estradiol Assay. While not explicitly called "ground truth" for the test set, it serves as the foundational "truth" for calibrating and establishing the accuracy of the assays.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for this type of in vitro diagnostic device. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation of medical images where disagreements among readers need to be resolved. This device directly measures a biomarker concentration.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an automated in vitro diagnostic assay. It does not involve human readers interpreting images or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies presented (precision, linearity, method comparison, interfering substances, specificity) represent the standalone performance of the Elecsys® Estradiol Assay as an automated in vitro diagnostic device. There is no human-in-the-loop performance aspect beyond standard laboratory procedures for sample handling and operating the instrument.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance evaluation can be considered multifaceted:
- Method Comparison: The predicate device, Enzymun-Test® Estradiol Assay, served as the primary comparative "ground truth" for showing substantial equivalence.
- Assay Standardization: ID-GC/MS (Isotope Dilution Gas Chromatography/Mass Spectrometry) was used for establishing the foundational accuracy of the assay values, acting as a higher-level "ground truth" for calibration.
- Interference and Specificity: These are determined by spiking known concentrations of potential interferents/cross-reactants and observing their effect on the measured estradiol value, with the known absence of interference/cross-reactivity being the "ground truth."
8. The sample size for the training set
The document does not specify a "training set" in the context of machine learning. This is an assay based on established biochemical principles (immunoechemistry) and calibration rather than a machine learning algorithm that requires a separate training set. The calibration curve is generated by 2-point calibration and a master curve provided via the reagent bar code. This master curve and the 2-point calibration define the device's operational parameters, which could be loosely analogous to what a "training set" provides in AI, but it's not a direct comparison.
9. How the ground truth for the training set was established
As there is no distinct "training set" for a machine learning algorithm, the concept of "ground truth for the training set" as typically understood in AI/ML is not directly applicable.
However, if we consider the calibration of the assay as analogous to "training":
- The calibration curve is generated using calibrators of known estradiol concentrations. The "ground truth" for these calibrators would be established through careful quantitative analysis using highly accurate reference methods, again likely involving methods traceable to ID-GC/MS or similar definitive techniques. The reagent bar code also provides a "master curve," implying pre-determined, highly characterized calibration data.
Ask a specific question about this device
(56 days)
Ask a specific question about this device
Page 1 of 1