Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K140482

    Validate with FDA (Live)

    Date Cleared
    2014-04-08

    (41 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Mandrel Guidewires are intended to facilitate the introduction of other diagnostic and treatment devices used in gastroenterology and urology procedures.

    Device Description

    The Mandrel wire family is made of a coated (PTFE or Silicone) or uncoated Nitinol or Stainless Steel core wire that is tapered at the distal tip where a coil is secured to the distal end. The distal coil can be anywhere from 2cm to 30cm depending on specific design and can consist of Stainless Steel, Palladium, Platinum, or the proposed Tungsten materials. The guidewire may contain proximal core markers. The Mandrel guidewire family is bound by the following parameters: Lengths: 20cm to 500cm, Outside Diameter: 0.014" To 0.35", Tips: Straight or shaped with various flexibilities, Coil Length: 2cm to 30cm.

    AI/ML Overview

    This document describes a 510(k) submission for Mandrel Guidewires with the addition of Tungsten as a material option. The submission focuses on demonstrating substantial equivalence to previously marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it states that "all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met" and that "the conclusions drawn from bench testing and biocompatibility testing demonstrate compliance with the design input summary."

    The document lists the following tests performed to ensure compliance with the design input summary:

    Acceptance Criteria (Test)Reported Device Performance
    VisualComplies with design input summary
    RadiopacityComplies with design input summary
    LubricityComplies with design input summary
    J-Memory TestComplies with design input summary
    Body StiffnessComplies with design input summary
    Adhesion / DurabilityComplies with design input summary
    Guidewire Pull testComplies with design input summary
    Torque StrengthComplies with design input summary
    Torque ControlComplies with design input summary
    DimensionalComplies with design input summary
    Linear StiffnessComplies with design input summary
    Lateral StiffnessComplies with design input summary
    Particulate TestComplies with design input summary
    ISO Strength of UnionComplies with design input summary
    ISO Flex TestComplies with design input summary
    ISO Corrosion Resistance TestComplies with design input summary
    ISO Fracture TestComplies with design input summary
    HydrodurabilityComplies with design input summary
    Cytotoxicity (Biocompatibility)Complies with pre-defined acceptance criteria
    Hemolysis (Biocompatibility)Complies with pre-defined acceptance criteria

    Note: The specific numerical or descriptive acceptance criteria for each test are not provided in this summary, nor are the detailed results. The device performance is generally stated as "complies with design input summary" or "complies with pre-defined acceptance criteria."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Test devices were manufactured and inspected according to established requirements." However, it does not provide the specific sample size used for the bench testing or biocompatibility testing. The data provenance is internal to Lake Region Medical through their "formal quality systems." There is no mention of country of origin for data or whether the study was retrospective or prospective in the standard sense (as this is a device modification submission, not a clinical trial). The data is generated from laboratory testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this device submission. The tests conducted are primarily physical, mechanical, and biological evaluations of the device itself, rather than diagnostic interpretations requiring expert consensus for ground truth. The "ground truth" for these tests would be the established engineering specifications and ISO standards.

    4. Adjudication Method for the Test Set

    This information is not applicable for the type of testing described (bench and biocompatibility). Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving human interpretation or subjective assessments.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This submission is for a material modification to a guidewire, and the testing focuses on the device's physical, mechanical, and biological properties, not its interpretive performance in a diagnostic context.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    This question is not applicable. The device is a physical medical guidewire, not an algorithm or AI software. Therefore, there is no "standalone" algorithm performance to evaluate.

    7. The Type of Ground Truth Used

    The ground truth used for this submission is based on established engineering design input summaries, internal specifications, and relevant ISO standards for material properties, mechanical performance, and biocompatibility.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" in the context of this device submission. The device is not an AI model requiring a dataset for training.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140485

    Validate with FDA (Live)

    Date Cleared
    2014-03-27

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Mandrel guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.

    Device Description

    The Mandrel wire family is made of a coated (PTFE or Silicone) or uncoated Nitinol or Stainless Steel core wire that is tapered at the distal tip where a coil is secured to the distal end. The distal coil can be anywhere from 2cm to 30cm depending on specific design and can consist of Stainless Steel, Palladium, Platinum, or the proposed Tungsten materials. The guidewire may contain proximal core markers.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Mandrel Guidewire with Tungsten, focusing on demonstrating substantial equivalence to previously cleared predicate devices. It does not contain information about a study proving the device meets specific acceptance criteria in the context of AI/ML performance, human reader studies, or ground truth establishment for a diagnostic device.

    The document primarily discusses bench testing for physical and material properties, and biocompatibility testing, to show that the modified device (with Tungsten) is as safe and effective as the predicate devices.

    Therefore, many of the requested categories for AI/ML device evaluation are not applicable to this submission.

    Here's a breakdown based on the information provided, highlighting the absence of AI/ML specific criteria:

    Acceptance Criteria and Device Performance (Non-AI/ML Context)

    Acceptance Criteria CategoryReported Device Performance (Non-AI/ML Context)
    Visual InspectionTest devices manufactured and inspected according to established requirements for visual/tactile attributes, demonstrating compliance with design input summary.
    Dimensional ConformanceTest devices manufactured and inspected according to established requirements for dimensional attributes, demonstrating compliance with design input summary.
    RadiopacityBench testing performed to show compliance with design input summary.
    LubricityBench testing performed to show compliance with design input summary.
    J-Memory TestBench testing performed to show compliance with design input summary.
    Body StiffnessBench testing performed to show compliance with design input summary.
    Adhesion / DurabilityBench testing performed to show compliance with design input summary.
    Guidewire Pull testBench testing performed to show compliance with design input summary.
    Torque StrengthBench testing performed to show compliance with design input summary.
    Torque ControlBench testing performed to show compliance with design input summary.
    Linear StiffnessBench testing performed to show compliance with design input summary.
    Lateral StiffnessBench testing performed to show compliance with design input summary.
    Particulate TestBench testing performed to show compliance with design input summary.
    ISO Strength of UnionBench testing performed to show compliance with design input summary.
    ISO Flex TestBench testing performed to show compliance with design input summary.
    ISO Corrosion Resistance TestBench testing performed to show compliance with design input summary.
    ISO Fracture TestBench testing performed to show compliance with design input summary.
    HydrodurabilityBench testing performed to show compliance with design input summary.
    Biocompatibility (Cytotoxicity)Biocompatibility testing per design input summary requirements showed the addition of Tungsten does not affect biocompatibility and is in compliance with pre-defined acceptance criteria.
    Biocompatibility (Hemolysis)Biocompatibility testing per design input summary requirements showed the addition of Tungsten does not affect biocompatibility and is in compliance with pre-defined acceptance criteria.

    Study Details (Non-AI/ML Context)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated as a number of devices. The document mentions "Test devices were manufactured and inspected according to established requirements."
    • Data Provenance: Not applicable in the context of clinical data. This refers to bench testing performed by Lake Region Medical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as this is not a diagnostic device relying on expert interpretation for ground truth. Ground truth for physical properties (e.g., dimensions, strength) would be defined by engineering specifications and measurement standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable; no clinical adjudication process is described for this type of device and testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable; this is not an AI/ML diagnostic device, and no MRMC study or human reader improvement is discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable; this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For bench testing, the "ground truth" is based on established engineering specifications, published standards (e.g., ISO), and physical measurements. For biocompatibility, it's based on established biological safety evaluation criteria.

    8. The sample size for the training set

    • Not applicable; this is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable; no training set described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K080508

    Validate with FDA (Live)

    Date Cleared
    2008-03-21

    (25 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To facilitate the introduction of other diagnostic and treatment devices used in gastroenterology and urology procedures.

    Device Description

    Nitinol or Stainless Steel core wire with Palladium coil secured to the ground, flexible (distal) end of the core. The core wire can be either uncoated or coated with PTFE (Polytetrafluoroethylene). Portions of the Guidewire can also be subsequently coated with silicone MDX4-4159 Fluid. The Guidewires are bound by the following parameters: OUTSIDE DIAMETER: .014" - .035", LENGTHS: 60 cm - 500 cm, TIPS: Straight or shaped with various tip flexibilities., COIL LENGTH: 2cm - 30 cm

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device called "Mandrel Guidewire or M-Wires™." The submission aims to demonstrate the substantial equivalence of the new device iterations to a predicate device.

    1. Acceptance Criteria and Reported Device Performance:

    The document explicitly states that "all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met." However, the specific quantitative acceptance criteria for each test and the detailed reported device performance against those criteria are not provided in the given text. It only vaguely mentions "established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes."

    The general performance claim is: "The results of these tests demonstrated the functionality and performance characteristics of these Guidewires are comparable to the currently marketed devices."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states, "Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes." However, it does not specify the sample size used for the test set.

    Data provenance (e.g., country of origin, retrospective or prospective) is not mentioned.

    3. Number of Experts and their Qualifications for Ground Truth Establishment:

    This information is not applicable as the document describes non-clinical testing of a physical medical device (guidewire) for substantial equivalence, not an AI/ML-based diagnostic device requiring human expert ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3. Testing involved physical device characteristics and biocompatibility, not expert adjudication of diagnostic outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This information is not applicable as this submission is for a physical medical device, not an AI-assisted diagnostic tool. No AI component is described.

    6. Standalone (Algorithm Only) Performance Study:

    This information is not applicable as this submission is for a physical medical device, not an algorithm.

    7. Type of Ground Truth Used:

    For the non-clinical tests, the "ground truth" was established by "established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes." This implies objective, measurable standards for physical properties. For biocompatibility, it was based on ISO 10993 series standards.

    8. Sample Size for the Training Set:

    This information is not applicable as this is a physical medical device, not an AI/ML system requiring a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as this is a physical medical device, not an AI/ML system requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080144

    Validate with FDA (Live)

    Date Cleared
    2008-02-26

    (35 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.

    Device Description

    Nitinol or Stainless Steel core with Palladium coil secured to the ground, flexible (distal) end of the core. The core wire can be either uncoated or coated with PTFE (Polytetratlyoroethylene). Portions of the Guidewire can also be subsequently coated with silicone MDX4-4159 Fluid. The Guidewires are bound by the following parameters: Outside Diameter: .014" - .035" Lengths: Straight or shaped with various tip flexibilities. Tips: Coil Length: 2cm - 30 cm

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Mandrel Guidewires (M-Wires), seeking substantial equivalence to existing predicate devices. However, the document does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and the specifics of a study proving the device meets those criteria.

    The document states that "Risk analysis was completed by means of a Failure Mode and Effect Analysis (FMEA) and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met." It also mentions "LRM performed testing to established requirements. Test pieces were tested and inspected according to established specific inspection criteria requirements for visualitactile, dimensional and mechanical attributes." And "Biocompatibility testing per ISO 10993 series has been performed on the Mandrel device and has been found to be acceptable."

    However, it does not explicitly list the predetermined acceptance criteria themselves or the detailed results of the tests against these criteria in a quantitative manner. It only states that the results "demonstrated the functionality and performance characteristics of these Guidewires are comparable to the currently marketed devices" and "support the claim of substantial equivalence to the predicate devices."

    Therefore, much of the requested information cannot be extracted from this specific filing.

    Here's a breakdown of what can and cannot be answered based on the provided text:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Visual/Tactile Attributes (Established specific inspection criteria)"Supported the claim of substantial equivalence to the predicate devices." "Functionality and performance characteristics... are comparable to the currently marketed devices."
    Dimensional Attributes (Established specific inspection criteria)"Supported the claim of substantial equivalence to the predicate devices." "Functionality and performance characteristics... are comparable to the currently marketed devices."
    Mechanical Attributes (Established specific inspection criteria)"Supported the claim of substantial equivalence to the predicate devices." "Functionality and performance characteristics... are comparable to the currently marketed devices."
    Biocompatibility (Per ISO 10993 series)"Found to be acceptable."
    Risk analysis (FMEA)Predetermined acceptance criteria were met.

    Note: The document states that the predetermined acceptance criteria were met, but it does not specify the quantitative values for these criteria. The "Reported Device Performance" is a summary statement of comparability and acceptability, rather than specific performance metrics against defined thresholds.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided text. The document only mentions "Test pieces were tested and inspected."
    • Data Provenance: Not specified, but likely internal testing conducted by Lake Region Manufacturing, Inc. in Chaska, MN, USA. There is no mention of external data or patient data.
    • Retrospective/Prospective: The testing described is prospective, in the sense that the new Mandrel Guidewire iterations were specifically tested to demonstrate equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable and not provided. The "ground truth" for this type of device (guidewire) is typically established through engineering specifications, material science, and performance testing against established physical and biological standards, not through expert clinical consensus on images or patient outcomes in the context of a diagnostic AI.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human interpretation of data where consensus is needed for ground truth. This filing discusses non-clinical engineering and bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This is a medical device (guidewire), not a diagnostic AI algorithm that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable. The device is a physical medical guidewire, not an algorithm. The testing described is "standalone" in the sense that it evaluates the physical device's properties (visual/tactile, dimensional, mechanical attributes, biocompatibility).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The ground truth for the non-clinical tests appears to be based on established engineering specifications, material standards, and performance requirements relevant to guidewires, as well as the ISO 10993 series for biocompatibility. This is considered objective, measurable physical properties and material safety standards rather than expert clinical consensus or pathology.

    8. The sample size for the training set

    • This is not applicable. The Mandrel Guidewires are physical medical devices, not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    • This is not applicable as there is no training set for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K011084

    Validate with FDA (Live)

    Date Cleared
    2001-05-10

    (30 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1