Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K140593
    Manufacturer
    Date Cleared
    2014-12-15

    (283 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K072749,K122940,K011084

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Advance® Salivary Balloon Catheter is intended for the dilation of lesions and strictures in the submandibular and parotid salivary ducts.

    Device Description

    The Advance® Salivary Balloon Catheter is an over-the-wire catheter available with an inflated balloon diameter of 1.5. 2.0, 2.5 and 3.0 millimeters in a balloon length of 2 centimeters. The catheter is available in an outer diameter of 2.5 French and in a length of 30 centimeters. Hydrophilic coating is applied to the distal 15 centimeters of the catheter shaft, which provides a lubricious surface when wetted. The balloon catheter is supplied with a 0.014 inch (0.36 millimeter) diameter stainless steel mandrel wire guide in a length of 50 centimeters. It is sterile and intended for one-time use.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study

    The provided text describes the Advance® Salivary Balloon Catheter with Wire Guide and the studies conducted to demonstrate its reliable design and performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Bond StrengthTensile force during proper clinical use should not fracture or rupture the balloon catheter bonds. Conforms to ISO 10555-1.Predetermined acceptance criteria were met.
    Balloon Minimum Burst StrengthBalloons will burst at or above the minimum rated burst pressure, with all failure modes being linear tears.Predetermined acceptance criterion was met.
    Balloon FatigueBalloons are free from leakage and damage on inflation after 10 cycles of inflation and deflation.Predetermined acceptance criteria were met.
    Inflation and Deflation TimeBalloon will inflate to rated burst pressure within 60 seconds and fully deflate within 60 seconds.Predetermined acceptance criteria were met.
    Resistance to CorrosionAfter being subjected to a corrosive environment, the wire guide withstands a tensile force greater than or equal to 5 N.Predetermined acceptance criterion was met.
    Visual and Dimensional AnalysisVisually and dimensionally, the wire guide and balloon catheter meet the design requirements.Predetermined acceptance criteria were met.
    Simulated UseWire guide and balloon catheter pass through a simulated stricture, and the force required to remove the balloon catheter is less than 0.95 lbf.Predetermined acceptance criteria were met.
    Balloon Working Length and Catheter IntegrityAfter simulated use testing, the catheter shows no signs of damage, and the balloon length is 2.0 centimeters ± 0.1 centimeter.Predetermined acceptance criteria were met.
    Simulated Clinical Use (Human cadaver model)Balloon catheter has an atraumatic tip, safely and reliably introduces and tracks to the intended location, is compatible with the wire guide, can be removed from the salivary duct and/or sheath, and its location is verifiable within the salivary duct.Predetermined acceptance criteria were met.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample sizes for each of the performance tests. It refers to "testing shows" and implies that sufficient samples were tested to meet the predetermined acceptance criteria for each test. The data provenance is not specified beyond the tests being conducted for regulatory submission for a device to be marketed in the USA. Given the nature of these engineering and performance tests, they are typically conducted in a controlled laboratory environment. The "Simulated Clinical Use" test specifically mentions a "human cadaver model," implying prospective testing in a controlled lab setting, likely in the country of manufacture (USA, given the applicant is Cook Incorporated in Bloomington, IN).

    3. Number of Experts and Qualifications for Ground Truth for Test Set

    Not applicable for most of the tests. The tests are engineering and performance-based, not reliant on expert interpretation for ground truth. For the "Simulated Clinical Use" test using a human cadaver model, while clinical utility is being assessed, the document does not specify the number or qualifications of experts who established the "ground truth" or evaluated the performance of the device in this context. It simply states that the "predetermined acceptance criteria were met," which would likely align with successful navigation and deployment as observed by the testing personnel.

    4. Adjudication Method for Test Set

    Not applicable. The tests described are objective, quantitative, or pass/fail engineering and performance tests, not requiring an adjudication method akin to clinical studies where multiple readers interpret results.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. The studies described are focused on the device's inherent design and performance characteristics, not on a comparison of human reader performance with or without AI assistance.

    6. Standalone Performance Study

    Yes, a standalone performance study was performed. The entire section "Technological Characteristics" details various tests conducted on the device itself (e.g., bond strength, burst strength, fatigue, inflation/deflation time, corrosion resistance, visual/dimensional analysis, simulated use, balloon working length/catheter integrity, simulated clinical use). These tests assess the algorithm/device's performance in isolation from a human-in-the-loop scenario.

    7. Type of Ground Truth Used

    The ground truth used for these tests is based on engineering specifications, predetermined quantitative thresholds, and observable physical properties. For example:

    • Bond Strength: Conformance to ISO 10555-1 standards.
    • Balloon Minimum Burst Strength: Bursting at or above a specified pressure.
    • Inflation and Deflation Time: Within 60 seconds.
    • Resistance to Corrosion: Withstanding a tensile force greater than or equal to 5 N.
    • Simulated Use: Catheter passing through a simulated stricture, removal force less than 0.95 lbf.
    • Balloon Working Length: 2.0 cm ± 0.1 cm.
    • Simulated Clinical Use: Observable successful introduction, tracking, compatibility, removal, and verifiable location in a cadaver model.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument (catheter, balloon, wire guide), not an AI algorithm or a device that uses a "training set" in the machine learning sense. The "training set" concept is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132545
    Device Name
    MEDIVALVE ACWIRE
    Manufacturer
    Date Cleared
    2013-11-08

    (87 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K062192, K011084

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ac Wire is intended for use in peripheral vascular and heart catheterization procedures to introduce and assist in positioning diagnostic and interventional devices. The ac Wire may also function as an alignment tool by providing a reference plane of anatomical structures of interest (e.g., the aortic valve).

    Device Description

    The ac Wire device is a 0.035"" outside diameter, single use, disposable guidewire. The ac Wire device consists of a flexible guidewire that incorporates a low profile, radiopaque alignment element consisting of three loops with radiopaque markers that when open, form a "tulip" configuration designed to assist the physician in acquiring a reference plane under fluoroscopy.

    AI/ML Overview

    The MediValve acWire is a guidewire intended for use in peripheral vascular and heart catheterization procedures to introduce and assist in positioning diagnostic and interventional devices. It may also function as an alignment tool by providing a reference plane of anatomical structures of interest (e.g., the aortic valve).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided 510(k) summary does not explicitly list quantitative acceptance criteria with numerical targets. Instead, it states that "descriptive information, laboratory bench testing, and biocompatibility testing were provided to demonstrate the device meets its design specifications, performs as intended, and is safe for its intended use." The performance is evaluated against the predicate devices for substantial equivalence.

    CategoryAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance
    Functional TestingThe guidewire should exhibit equivalent mechanical and functional performance to predicate devices (Ostial Pro Stent Positioning System and Lake Region Manufacturing Catheter Guidewire) in areas such as tip flexibility, torque strength, fracture resistance, coating integrity, particulate generation, tensile strength, repeated use durability, corrosion resistance, shelf-life, radiopacity, loop deflection, usability, stiffness, and compatibility with other interventional devices. This implies meeting industry standards or demonstrating comparable performance to the predicate devices in these specific tests. "Performs as intended" means it facilitates delivery and acts as an alignment tool.Demonstrated safety through design verification testing for: Tip Flexibility, Torque Testing, Torque Strength, Fracture Test, Resistance to Coating Damage, Coating Adhesion, Particulate Residual, Tensile Testing, Repeated Use Test, Corrosion Test, Shelf-Life Test, Radiopacity Test, Loop Deflection Test, Usability Test, Stiffness Test, Compatibility Test, Austenite Finish Test, Ex-vivo Performance Test. The submission asserts that the "performance as evaluated in bench tests, demonstrates that the ac Wire device is as safe and effective as the predicate devices."
    BiocompatibilityThe guidewire materials should be biocompatible and not elicit adverse biological reactions when in contact with human tissue and blood, as per ISO 10993-1. This includes acceptable levels for cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, thrombus formation, and complement activation.Biocompatibility testing performed in accordance with ISO 10993-1, including: Cytotoxicity Study, Maximization Sensitization, Intracutaneous Study, Systemic Toxicity Study, Pyrogen Study, Material Mediated Hemolysis, In Vivo Thromboresistance Study, SC5b-9 Complement Activation Assay, C3a Complement Activation Assay. The collective results "demonstrated that the ac Wire device is safe and is substantially equivalent to the respective predicate devices."
    Substantial EquivalenceThe ac Wire device should be substantially equivalent to the predicate devices (Ostial Pro Stent Positioning System (K062192) and Lake Region Manufacuring (LRM) Catheter Guidewire (K011084)) in terms of design, intended use, principles of operation, and technological characteristics, without raising new issues of safety or effectiveness. This is the overarching "acceptance criteria" for a 510(k) submission.The submission concludes, "The design, intended use, principles of operation, and technological characteristics of the ac Wire are substantially equivalent to those of the predicate devices cited above. Substantial equivalence is based upon descriptive characteristics of the various cited predicate devices and upon the testing conducted to demonstrate that the subject device performs as intended and is substantially equivalent to the predicate devices in terms of its ability to safely perform as a cardiovascular guide wire. Any differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes laboratory bench testing and biocompatibility testing. It does not specify a "test set" in the context of patient data or clinical images. The testing involved physical samples of the device or its components. The document does not provide details on the specific number of units tested for each functional test (e.g., how many guidewires were used for the tensile test, or how many loops for the loop deflection test).

    • Data Provenance: The testing appears to be primarily laboratory-based and ex-vivo (as indicated by "Ex-vivo Performance Test" and "In Vivo Thromboresistance Study" which refers to animal studies rather than human clinical data). There is no mention of country of origin for any human data, as no human data is presented in the summary.

    3. Number of Experts and their Qualifications for Ground Truth:

    This information is not applicable as the summary describes bench and biocompatibility testing of a physical medical device, not an AI/software device that requires expert-established ground truth from medical images or clinical data.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as above. There is no "test set" requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI or imaging devices to compare human performance with and without AI assistance. The MediValve acWire is a physical guidewire that does not involve AI assistance for human readers.

    6. Standalone (Algorithm Only) Performance:

    No, a standalone performance evaluation (algorithm only without human-in-the-loop performance) was not done. This concept applies to AI algorithms and is not relevant to a physical guidewire device like the MediValve acWire.

    7. Type of Ground Truth Used:

    The "ground truth" for the device's performance in this context is defined by:

    • Engineering specifications and design requirements: Functional tests aim to verify that the device meets predetermined mechanical, physical, and performance specifications.
    • Biocompatibility standards (ISO 10993-1): The "ground truth" for biocompatibility is conformance to established biological safety standards as determined by specific assay results.
    • Predicate device characteristics: The ultimate "ground truth" for safety and effectiveness in a 510(k) submission is substantial equivalence to a legally marketed predicate device. This means the new device performs as well as, or better than, the predicate in relevant tests without raising new safety or effectiveness concerns.

    8. Sample Size for the Training Set:

    This information is not applicable. The MediValve acWire is a physical medical device and does not involve AI algorithms that require a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for a physical guidewire device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040592
    Manufacturer
    Date Cleared
    2004-12-17

    (284 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K934474, K011084, K011968, K881787, K994358

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cordis STEER-IT Deflecting Tip Guidewires are intended for use in angiographic and interventional devices within the coronary and peripheral vasculature.

    Device Description

    The Cordis STEER-IT Deflecting Tip Guidewire is similar in design and materials to the Cordis Wizdom Steerable Guidewire, Lake Region Manufacturing, Inc. Mandrel Guidewire, and the Lake Region Manufacturing, Inc. Steerable PTCA Guidewire. The difference is that the STEER-IT Deflecting Tip Guidewire employs a handle mechanism, which deflects the distal tip in two directions. The handle mechanism is butted into a stainless steel hypotube, which is soldered to a nitinol hypotube. The handle mechanism is similar to the USCI Commander™ Deflection Handle, an accessory of the USCI PilotWire™ Deflectable Guide Wire. The distal end of the STEER-IT Deflecting Tip Guidewire (3mm or 7mm) of the nitinol inner core wire is flattened, looped around and joined to the nitinol hypotube. The handle mechanism operates the movement of the distal tip by gripping the inner moveable core and allowing limited movement in two directions.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Cordis STEER-IT Deflecting Tip Guidewire. This type of submission relies on demonstrating substantial equivalence to a predicate device, rather than conducting new clinical studies to prove safety and effectiveness. Therefore, the information requested in the prompt, which is typically derived from clinical trials or performance studies, is largely not present in this document.

    However, based on the text, I can infer and state what is available regarding acceptance criteria and performance:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical/Physical Properties (e.g., joint strength, torqueability, coating integrity, linear stiffness, turns to failure, dimensional testing)"The Cordis STEER-IT Deflecting Tip Guidewire passed all the established acceptance criteria."
    Biocompatibility"All materials were determined to be biocompatible." / "Biocompatibility... demonstrates no significant change to product characterization when compared to the predicate device."
    Functional EquivalenceSimilar in design, materials, and function to predicate steerable guidewires, specifically featuring a handle mechanism to deflect the distal tip in two directions, similar to the USCI Commander™ Deflection Handle.

    Study Information Pertaining to Acceptance Criteria:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a sample size for any "test set" in the context of clinical or large-scale performance studies. The testing mentioned (joint strength, torqueability, etc.) would have involved a sufficient number of samples to demonstrate compliance with internal specifications, but these numbers are not disclosed.
      • Data provenance is not specified beyond "in vitro testing." No information on country of origin or whether testing was retrospective or prospective is provided.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The submission focuses on in vitro engineering tests and substantial equivalence claims, not clinical efficacy or diagnostic accuracy requiring expert ground truth for a test set.
    3. Adjudication method for the test set:

      • Not applicable. There is no mention of a test set requiring adjudication in this technical review.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a medical guidewire, not an AI-powered diagnostic or assistive tool for human readers. Therefore, MRMC studies or AI performance metrics are irrelevant.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an algorithmic or AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the mechanical and physical tests, the "ground truth" would be established engineering specifications and industry standards for guidewire performance. For biocompatibility, it would be established ISO standards for medical device biocompatibility. There is no biological or clinical "ground truth" in the sense of disease diagnosis or outcome for this type of device submission.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1