Search Results
Found 351 results
510(k) Data Aggregation
(121 days)
FGE
The EndoFix™Tissue Fixation System is an endoscopic tissue approximation system intended for anchoring the gallbladder to the stomach or duodenum to aid placement of a luminal apposing metal stent (LAMS) for EUS-guided gallbladder drainage.
The EndoFix™ Tissue Fixation System (TFS) is an endoscopic (through the scope) tissue approximation device consisting of two main components:
- EndoFix Tissue Fixation Device – Enables the approximation of soft tissue in the gastrointestinal (GI) tract using a suture-based implant delivered through an 18 gauge echogenic needle.
- Secure Suture Locking Device – Secures the implant assembly with an implant grade suture lock and cuts the suture once approximation is complete.
The EndoFix TFS is designed to place an implant for endoscopic tissue approximation to anchor the gallbladder to the stomach or duodenum to aid placement of a luminal metal apposing stent (LAMS). During use, the EndoFix Tissue Fixation Device is advanced through a ≥3.2 mm working channel of a commercially available echoendoscope (EUS). Under ultrasound guidance, the EndoFix TFS needle is advanced through the target tissues and the implant subassembly is deployed to approximate the luminal tissues (either transgastric or transduodenal based on clinical requirements). With applied suture tension, a Suture Lock is deployed to secure the tissues using the Secure Suture Locking Device. The implant subassembly and Suture Lock are considered permanent implants.
Each of the respective delivery systems used to deliver the implant assembly/Suture Lock are short term contact devices.
This FDA 510(k) clearance letter for the EndoFix™ Tissue Fixation System does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a study proving those criteria are met. This document primarily focuses on establishing substantial equivalence to predicate devices for regulatory clearance.
Specifically, it does not include:
- A table of acceptance criteria with reported device performance metrics (e.g., success rates, tensile strength, specific measurements).
- Details about a "test set" for a performance study (sample size, data provenance, ground truth establishment, expert qualifications, or adjudication methods).
- Information on a multi-reader multi-case (MRMC) comparative effectiveness study or any effect size for human reader improvement.
- Distinction between standalone algorithm performance and human-in-the-loop performance, as this is a medical device, not an AI/software-only device in the context of typical AI performance criteria.
- Training set sample size or how its ground truth was established.
However, based on the provided text, I can infer and extract some relevant information regarding performance testing.
Inferable Acceptance Criteria and Reported Device Performance
The document lists categories of performance testing that were conducted with "acceptable results." While specific numerical acceptance criteria and reported performance values are not given, the implication is that the device met internal, pre-defined criteria for these aspects.
A table of acceptance criteria and reported performance cannot be fully constructed from the provided text because specific numerical values for criteria and performance are absent. However, based on the categories mentioned under "Performance Data," we can infer the types of criteria and the general statement of performance:
Acceptance Criteria Category | General Reported Device Performance |
---|---|
Visual Inspection | Acceptable |
Dimensional Verification | Acceptable |
Endoscope Compatibility | Acceptable |
Functionality (approximate/suture tissue) | Acceptable |
Destructive Testing (Product Integrity, Tensile of joints, withstand minimum forces) | Acceptable |
Packaging Verification | Acceptable |
Biocompatibility | Acceptable (per ISO 10993-1) |
Usability Evaluation | Acceptable |
Sterilization Validation | Acceptable (per ISO 11135, SAL 10-6) |
Safety & Performance in Animal Model | Safe and appropriately designed as a pre-stenting tissue fixation device for EUS-guided gallbladder drainage. |
Study Details (as far as extractable):
-
Sample size used for the test set and the data provenance:
- The document mentions "bench testing" and a "preclinical animal study." For bench testing, sample sizes are not specified.
- For the animal study, the model used was a "porcine model," indicating an animal test subject. The number of animals or specific cases within that study is not provided.
- Data provenance: Porcine model (animal study), laboratory/bench for other tests. Retrospective/prospective is not specified, but animal studies are typically prospective experimental designs.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number or qualifications of experts for establishing ground truth, as this device's performance assessment relies on physical and functional tests, and an animal study, rather than expert interpretation of medical images or data.
-
Adjudication method for the test set:
- Not applicable/Not specified. The performance testing described (bench, animal study) does not involve adjudication processes typically used in clinical imaging or diagnostic studies.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a medical device, not an AI-assisted diagnostic tool subject to MRMC studies. The clearance is for the physical device itself.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm, so "standalone" performance in the context of AI software is not relevant here.
-
The type of ground truth used:
- For bench testing: Engineering specifications, physical measurements, functional demonstrations.
- For the animal study: Observation of safety and performance (e.g., successful tissue approximation, LAMS deployment, absence of adverse events) within the porcine model. This could be considered "outcomes data" in an experimental animal model context.
-
The sample size for the training set:
- Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning algorithms. Its design would be informed by engineering principles and possibly prior animal/cadaver studies, but not a formal 'training set'.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set mentioned for this type of device.
Ask a specific question about this device
(257 days)
FGE
The ARCHIMEDES biodegradable pancreatic stent is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting.
The ARCHIMEDES Biodegradable Pancreatic Stent is intended to be delivered through the working channel of a therapeutic endoscope using a guidewire with a maximum outer diameter of 0.035 inches and a pushing catheter of between 5 to 10 French.
The ARCHIMEDES™ Biodegradable Pancreatic Stent is a single use, sterile, biodegradable stent implant intended for use in the pancreatic duct. The device is intended to be delivered endoscopically through the working channel of a duodenoscope.
The ARCHIMEDES™ has a fluted cross-sectional profile and provides three drainage channels, two of which spiral in a double helical pattern down the length of the stent and one of which is the hollow inner lumen of the device tube. This unique design permits fluids to flow both around and through the stent. It has a curved form design to fit the natural curvature of the pancreatic duct, and contains flap and split-end features to minimize spontaneous migration of the stent while in situ. The tips of the stent are tapered to facilitate atraumatic insertion through the papilla. The stent is visible under fluoroscopy. The 8F and 10F stent design contains flaps on each end. The 6F stent design has a flap on one end and a split-end feature on the other end.
The stent is offered in outer diameters of 2.0mm (6F), 2.6mm (8F), or 3.4mm (10F), and in various lengths to accommodate variations in pancreatic anatomy across individuals (40, 60, 80, 100, 125, 150, 175, 200, and 225 mm). The ARCHIMEDES™ is designed to be an alternative to plastic stents that maintains the same clinical purpose. In particular, it is used in a manner similar to current FDA-cleared plastic pancreatic stents. The primary difference is that the ARCHIMEDES™ biodegrades in-situ rather than requiring an additional procedure for removal needed of plastic stents.
The Minimal Strength Retention (MSR) time for the material - i.e., the minimum amount of time for which the device retains at least 10% of an initial strength parameter and remains intact with no breaks is 12 days. Full-degradation or no stent presence is reached within
The provided FDA 510(k) clearance letter and summary for the ARCHIMEDES™ Biodegradable Pancreatic Stent do not contain the specific acceptance criteria and detailed performance results for the device. While it states that "All tests met required acceptance criteria," the actual criteria and reported numerical performance are not explicitly listed.
Furthermore, the document alludes to "eight clinical studies" but provides no details whatsoever about their methodology, sample sizes, data provenance, expert qualifications, adjudication methods, or specific outcomes. It's a high-level statement without any actionable data.
Therefore, many of your requested fields cannot be filled from the provided text.
Here's a breakdown of what can be extracted and what cannot, based on the input:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Non-clinical Performance | |
Visual Inspection | Met required acceptance criteria (no specific details provided) |
Outer diameter, Length | Met required acceptance criteria (no specific details provided) |
Stent Swelling Characterization | Met required acceptance criteria (no specific details provided) |
Flow Rate | Met required acceptance criteria (no specific details provided) |
Inherent Viscosity | Met required acceptance criteria (no specific details provided) |
Guidewire Compatibility | Met required acceptance criteria (no specific details provided) |
Endoscope Compatibility | Met required acceptance criteria (no specific details provided) |
Introducer Sleeve Compatibility | Met required acceptance criteria (no specific details provided) |
Simulated Use | Met required acceptance criteria (no specific details provided) |
Trackability | Met required acceptance criteria (no specific details provided) |
Pushability | Met required acceptance criteria (no specific details provided) |
Flexibility/Kink Resistance | Met required acceptance criteria (no specific details provided) |
Retraction Force | Met required acceptance criteria (no specific details provided) |
Tensile Strength | Met required acceptance criteria (no specific details provided) |
Fluoroscopic Visibility | Met required acceptance criteria (no specific details provided) |
Degradation | Met required acceptance criteria (no specific details provided) |
Biocompatibility (per ISO 10993-1:2009) | |
MEM Elution Cytotoxicity | Met required acceptance criteria (no specific details provided) |
Implantation | Met required acceptance criteria (no specific details provided) |
Guinea Pig Maximization Sensitization | Met required acceptance criteria (no specific details provided) |
28 Day Dual Route IV/IP Systemic Toxicity | Met required acceptance criteria (no specific details provided) |
Acute Systemic Injection | Met required acceptance criteria (no specific details provided) |
Intracutaneous Reactivity | Met required acceptance criteria (no specific details provided) |
Material Mediated Pyrogen | Met required acceptance criteria (no specific details provided) |
Chemical Extractables Studies & Toxicological Risk Assessment | Met FDA's Guidance document (no specific details provided) |
Sterilization (per ISO 11135-1) | Met Sterilization requirements (no specific details provided) |
Ethylene Oxide Sterilization Residuals (per ISO 10993-7) | Met Ethylene Oxide Sterilization Residuals requirements (no specific details provided) |
Minimal Strength Retention (MSR) | The device retains at least 10% of an initial strength parameter and remains intact with no breaks for 12 days. |
Full Degradation | Full-degradation or no stent presence is reached within |
Ask a specific question about this device
(69 days)
FGE
The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD):
• Used to drain pancreatic ducts
The Advanix™ Pancreatic Stent and NaviFlex™ RX Pancreatic Delivery System and Pushers is a plastic pancreatic stent designed for the delivery of the stent to the pancreatic duct and used to drain pancreatic ducts.
The pancreatic stents are provided in straight or single pigtail shape. The straight shape stents have trailing barbs and/or leading barbs depending on application, in rounded or tapered leading end tip to facilitate access through papilla, and a rounded trailing end or about the push catheter portion of the delivery system or stent pusher. The single pigtail stent may or may not have leading end barbs depending on application. Some stents have lateral drainage holes in the pigtails, a tapered or rounded leading end tip, and a rounded trailing end. All stents have either an endoscopic or fluoroscopic marker, or both on the trailing or leading end of the stent to assist with depth of placement in the pancreatic duct. The location and presence of an endoscopic or fluoroscopic marker is dependent on the length and diameter size of the stent. Some codes have side port holes in the body of the stent.
I regret to inform you that the provided text is a U.S. FDA 510(k) clearance letter and summary for a medical device (Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers).
This type of document primarily focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance bench testing and biocompatibility.
Crucially, this document does not contain information about clinical studies with human patients, nor does it establish device performance against specific acceptance criteria in a clinical setting.
Therefore, I cannot extract the information you requested regarding:
- A table of acceptance criteria and the reported device performance: These are not defined in the context of clinical outcomes within this document. The "performance" mentioned refers to bench testing.
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): No clinical test set is described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as there is no clinical test set with ground truth established by experts.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is a physical medical device (a stent and delivery system), not an AI/software device. Therefore, MRMC studies and AI assistance are not relevant.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as it's not an algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable for the clinical performance criteria you are asking about. The "ground truth" for the bench tests would be the established engineering specifications and standards.
- The sample size for the training set: Not applicable, as there is no training set for a clinical algorithm.
- How the ground truth for the training set was established: Not applicable.
The "Non-Clinical and/or Clinical Tests Summary & Conclusions" section explicitly states that the submission includes Performance Bench Testing and Biocompatibility Testing. These are standard tests for physical devices to ensure their safety and physical functionality, not their clinical effectiveness in terms of patient outcomes or diagnostic accuracy.
In summary, the provided document is not a study report that demonstrates clinical performance against acceptance criteria for patient outcomes or diagnostic accuracy.
Ask a specific question about this device
(278 days)
FGE
The EGIS Biliary Double Bare Stent is indicated for the palliation of malignant strictures in the biliary tree.
EGIS Biliary Double Bare Stent has straight and round cylinder form made of nitinol wire. The double bare type is composed of two structures, an inner stent and an outer stent, and has a double-layer form. Each stent has the same structure as the single bare type of the product. A double layer is formed by overlapping a separately manufactured inner stent and an outer stent, and both ends are physically fixed using medical sutures. No additional bonding material in this process. This manufacturing method also allows the product to have more conformability and a smaller cell size.
The provided FDA 510(k) clearance letter for the EGIS Biliary Double Bare Stent does not contain any information about a study involving acceptance criteria for device performance with respect to AI or human reader assistance.
The document primarily focuses on demonstrating the substantial equivalence of the EGIS Biliary Double Bare Stent to a predicate device (EGIS Biliary Single Bare Stent) through non-clinical bench testing and biocompatibility evaluation. The letter explicitly states: "Clinical testing was not required for this submission." This means no human-in-the-loop or standalone AI performance studies were conducted or reported in this clearance.
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets those criteria based on the provided input. The information requested regarding AI performance, human reader studies, ground truth establishment, expert adjudication, and training/test set details is entirely absent from this 510(k) clearance letter, as it's not relevant to the type of device (a biliary stent) or the regulatory pathway chosen (510(k) based on substantial equivalence through non-clinical data).
If you have a document pertaining to an AI device or a study involving human reader performance, please provide that document, and I would be happy to analyze it according to your requested criteria.
Ask a specific question about this device
(175 days)
FGE
This device is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract. It is also indicated in adults for endoscopic dilatation of Sphincter of Oddi with or without prior sphincterotomy.
According to the structure of the device, it can be divided into three types, RX type, OTW type and OTW with Stainless Wire-guided type. The Rx proposed device mainly consists of tip, balloon, marker band, shaft, sleeve, stress diffusion tube and hub. The OTW proposed device mainly consists of tip, balloon, marker band, shaft, stress diffusion tube and hub. The structure of the OTW with Stainless Wire-guided model same with the structure of OTW model. The difference between OTW with Stainless Wire-guided type catheter and OTW type catheter is that OTW with Stainless Wire-guided type catheter is preloaded with a guide wire and a guide wire locking component. The difference between the RX type catheter and the OTW series catheter is that the guidewire exit position is different. The guide wire of OTW series catheter passed through the guide wire lumen of the catheter. The guidewire of the Rx type catheter will pass through the guidewire exchange port on the shaft.
Single-use Balloon Dilatation Catheter is divided into different specifications. Shafts are available in two different diameters. The effective length of the catheter is 180cm and 240cm. Balloon diameters are available in six different diameters 6-7-8mm, 8-9-10 mm, 12-13.5-15 mm, 15-16.5-18 mm, 18-19-20 mm, and the balloon lengths are available in 30 mm,55 mm and 80 mm three different length.
The provided FDA 510(k) summary for the "Single-use Balloon Dilatation Catheter" (K241888) details a non-clinical study to demonstrate substantial equivalence to predicate devices, rather than an AI/ML software performance study. Therefore, most of the requested information regarding AI/ML device acceptance criteria, human reader studies, and ground truth establishment for AI/ML models is not applicable to this document.
The document describes performance testing for a physical medical device. Here's a breakdown of the applicable information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states: "All the test results demonstrate proposed device meet the requirements of its pre-defined acceptance criteria and intended uses." However, it does not provide a specific table detailing the exact quantitative acceptance criteria for each test and the corresponding measured performance values. It lists the types of tests performed.
Test Category | Acceptance Criteria (Generic Statement) | Reported Device Performance (Generic Statement) |
---|---|---|
Dimension Test | Met predefined dimensional specifications | Met specifications |
Appearance | Conformed to visual quality standards | Conformed to standards |
Compatibility Test | Compatible with intended accessories/systems | Demonstrated compatibility |
Delivery and Retrieval Force | Within acceptable force limits for safe delivery/retrieval | Met force limits |
Peak Tensile Force | Withstood specified tensile forces without failure | Withstood forces |
Burst Pressure | Withstood specified internal pressures without bursting | Met burst pressure requirements |
Kink Stability | Demonstrated resistance to kinking | Demonstrated good kink stability |
Corrosion Resistance Test | Resisted corrosion in specified environments | Showed resistance to corrosion |
Air Leakage | No detectable air leakage at specified pressures | No air leakage detected |
Liquid Leakage | No detectable liquid leakage at specified pressures | No liquid leakage detected |
Luer Connector | Conformed to ISO 80369-7 standards for luer connectors | Complied with ISO 80369-7 |
Radiopacity | Met ASTM F640-20 standards for radiopacity | Complied with ASTM F640-20 |
Balloon Burst Pressure | Withstood specified pressures before rupturing | Met burst pressure requirements |
Balloon Compliance | Exhibited expected compliance characteristics | Demonstrated expected compliance |
Balloon Deflation Time | Deflated within specified time limits | Deflated within limits |
Balloon Fatigue | Withstood specified fatigue cycles without failure | Passed fatigue testing |
Sterilization (SAL) | SAL of 10^-6 (ISO 11135:2014) | Achieved 10^-6 SAL |
EO & ECH Residuals | Below limits specified in ISO 10993-7:2008 | Below specified limits |
Bacterial Endotoxins | Below 20 EU/device (USP ) | Below 20 EU/device |
Shelf-life | Demonstrated performance for proposed 3-year shelf-life (ASTM F1980-16) | Validated 3-year shelf-life |
Package Integrity (Visual Insp.) | Conform to ASTM F1886/F1886M-16 | Conformed |
Package Integrity (Seal Strength) | Conform to ASTM F88/F88-21 | Conformed |
Package Integrity (Dye Penetration) | Conform to ASTM F1929-15 | Conformed |
Biocompatibility (Cytotoxicity) | No cytotoxicity (ISO 10993-5:2009) | No cytotoxicity |
Biocompatibility (Sensitization) | No sensitization (ISO 10993-10:2021) | No sensitization |
Biocompatibility (Intracutaneous Reactivity) | No intracutaneous reactivity (ISO 10993-10:2021) | No intracutaneous reactivity |
Biocompatibility (Systemic Toxicity) | No acute systemic toxicity (ISO 10993-11:2017) | No acute systemic toxicity |
Biocompatibility (Pyrogen) | No pyrogen (USP) | No pyrogen |
2. Sample sizes used for the test set and data provenance:
The document states "The performance testing conducted on subject device and predicate device are listed below." and mentions "ASTM F3172 Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices". This suggests that standard device testing sample sizes were used in accordance with this guideline, but the specific sample sizes for each test are not explicitly stated in this summary.
Data provenance: This is a physical device, so "data provenance" would refer to the testing conditions and results. The tests were performed in a lab setting to verify design specifications and compliance with international standards (e.g., ISO, ASTM, USP). The country of origin for the testing is not specified, but the applicant company is Leo Medical Co., Ltd. in China. The data is prospective as it's generated through device testing.
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
This question is not applicable as this is a physical device performance study, not a study involving human reader interpretation of images or AI/ML model output requiring "ground truth" established by experts in a clinical context. The "ground truth" for these tests are the physical measurements and compliance with engineering and material standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This is not applicable as it's a physical device performance study. Adjudication methods are typically for clinical consensus on ground truth in image interpretation or diagnosis, not for engineering performance tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. No MRMC study was performed as this is not an AI/ML-assisted diagnostic device. The document explicitly states: "No clinical study is included in this submission."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable. This is not an algorithm, but a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For this physical device, the "ground truth" for the non-clinical tests is based on established engineering and material science standards, measurements, and pre-defined acceptance criteria. For example, burst pressure is measured against a standard, biocompatibility is assessed against ISO 10993, and sterility against ISO 11135.
8. The sample size for the training set:
This is not applicable. No training set was used as this is a physical device, not an AI/ML model being "trained."
9. How the ground truth for the training set was established:
This is not applicable. No training set was used.
Ask a specific question about this device
(179 days)
FGE
This device is used to dilate strictures in the pancreatobiliary systems and to dilate openings via the transgastric or transduodenal wall. This device is indicated for adult use only.
The Tornus ES is a rotary-operated dilator designed for use in bile duct/pancreatic duct strictures and via transgastric/transduodenal dilation of openings. The Tornus ES is comprised of stainless-steel coils and polymeric materials. The Dilation and Shaft segments are composed of coils materials that are welded together. The Tornus ES is available for prescription use only.
The provided text describes the performance data for the Tornus ES device, specifically focusing on non-clinical testing and biocompatibility assessments. It does not detail a study involving human subjects or AI assistance, which would typically involve acceptance criteria related to accuracy, sensitivity, specificity, or reader performance metrics. Therefore, several points of your request cannot be fulfilled as they are not applicable to the information provided.
Based on the provided text, primarily pages 8 and 9, here's the information regarding the device's acceptance criteria and the study that proves it meets those criteria:
1. A table of acceptance criteria and the reported device performance
The document provides acceptance criteria specifically for biocompatibility testing, and a general "Pass" for non-clinical performance tests.
Table 1: Non-Clinical Testing Performance
Test Item | Reported Device Performance |
---|---|
Appearance | Pass |
Tensile strength | Pass |
Guidewire pass-through ability | Pass |
Dilation ability | Pass |
Slide durability | Pass |
Radio-detectability | Pass |
Corrosion resistance | Pass |
Polyurethane strength | Pass |
Guide wire trackability | Pass |
Kink resistance | Pass |
Dimension measurement | Pass |
Simulated use and (torsional strength) | Pass |
Table 2: Biocompatibility Testing Acceptance Criteria and Results
Test Method | Standard | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Cytotoxicity MEM Elution Test | ISO 10993-5 (No deviations) | The test system is considered suitable if no signs of cellular reactivity (Grade 0) are noted for both the negative control article and the medium control. | Non-cytotoxic |
Sensitization KLIGMAN Maximization Test | ISO 10993-10 (No deviations) | The extracts should show no evidence of causing delayed dermal contact sensitization in the guinea pig. | Non-Sensitizing |
Irritation Intracutaneous Injection Test | ISO 10993-10 (No deviations) | The test extract and the negative control must exhibit similar edema and erythema scores. | Non-Irritant |
Systemic Toxicity Acute System Toxicity Test | ISO 10993-11 (No deviations) | The test article must not show significantly greater biological activity than the control. | Non-toxic |
Systemic Toxicity Rabbit Pyrogen Test (material mediated) | ISO 10993-11 (No deviations) | The test article should not increase the rectal temperature of any of the animals by more than 0.5 degrees Celsius. | Non-pyrogenic |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document refers to "Non clinical laboratory testing" and "in vitro bench tests". This indicates that the testing was performed on units of the device itself and not on patient data. No specific sample sizes for these tests are provided, nor is the country of origin of the data or whether it was retrospective or prospective, as these terms are generally applicable to clinical trials or studies involving patient data, which is not the case here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. The testing was non-clinical and benchtop, not requiring human expert interpretation or ground truth establishment in the context of imaging or clinical diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this refers to adjudication of ground truth in clinical data, not bench testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document describes a medical device (a catheter) and its non-clinical performance and biocompatibility. It does not involve AI or human image readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical instrument, not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical tests, the "ground truth" would be the engineering specifications and performance standards established for the device. For biocompatibility, the ground truth is established by the methods and criteria defined in the ISO 10993 series of standards.
8. The sample size for the training set
Not applicable. This device is not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(74 days)
FGE
The WallFlex Biliary PLUS RX Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms and relief of malignant biliary obstruction prior to surgery.
The WallFlex™ Biliary PLUS RX Stent System is an implantable biliary self-explaining metal stent that is pre-loaded onto a Delivery System with a working length of 194cm, which allows delivery of the stent into the Biliary system endoscopically. The self-expanding metal stent consists of Platinum cored Nitinol wires wound together to form a cylinder including flares on both the hilar and duodenal ends. WallFlex™ Biliary PLUS RX Stent is available fully covered with a Permalume™ covering.
I am sorry, but the provided text does not contain the information required to describe the acceptance criteria and the study proving the device meets these criteria. The document is an FDA 510(k) clearance letter for a medical device (WallFlex Biliary PLUS RX Stent System) and its accompanying 510(k) summary.
While it mentions that "Performance testing for the proposed WallFlex™ Biliary PLUS RX Stent System was completed in accordance with the following FDA Guidance documents to support substantial equivalence," it does not provide:
- A table of acceptance criteria and reported device performance.
- Details on sample size, data provenance, number/qualifications of experts, or adjudication methods for any test set (it mentions non-clinical tests, not a clinical trial or AI model validation).
- Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
- Stand-alone (algorithm only) performance studies.
- Type of ground truth used or how it was established.
- Training set details (sample size or ground truth establishment).
The document is primarily a regulatory communication confirming substantial equivalence to predicate devices based on non-clinical performance data and design comparisons, not a detailed report of a clinical efficacy/effectiveness study or performance validation for an AI/CADe device.
Ask a specific question about this device
(196 days)
FGE
The WallFlex Biliary RX Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms and relief of malignant biliary obstruction prior to surgery.
The WallFlex Biliary RX Fully Covered Stent System RMV is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms, relief of malignant biliary obstruction prior to surgery and for indwell up to 12 months in the treatment of benign biliary strictures secondary to chronic pancreatitis.
The Epic Biliary Endoscopic Stent System is indicated for palliation of malignant neoplasms in the biliary tree.
The WallFlex™ Biliary RX Stent System is available in the following stent configurations: Uncovered, Fully Covered. The WallFlex Bliary RX Stent System consists of a self-expanding metal stent and a delivery system. The self-expanding metal stent consists of Platinum cored Nitinol wires wound together to form a cylinder including flares on both the proximal end and distal end.
WallFlex Fully Covered (FC) and Partially Covered (PC) Stents have a Permalume™ Coating, which is a translucent silicone polymer. The coating is used to reduce the potential for tumor in growth through the stent. The coating is used to reduce the potential for tumor in growth through the stent. The FC and PC stents have a retrieval loop for removal during the initial stent placement procedure, to be used in the event of incorrect placement.
The WallFlex Biliary RX Fully Covered Stent System RMV consist of a flexible delivery system preloaded with a self-expanding billiary metal stent. The stents are available in a fully covered configuration only, and have a Permalume™ Covering, which consists of a translucent silicone polymer, to reduce potential for tumor in growth through the stents also have a retrieval loop for removal during the initial stent placement procedure where the retrieval loop is used in the event of incorrect placement, or for removal following indwell up to 12 months.
The Epic "Billary Stent is a sterile laser cut sent composed of a nickel titanium alloy. The stents are available uncovered.
On both the proximal and distal ends of the stent, radiopaque markers increase visibility of the stent to aid in placement.
Since the provided text is a 510(k) summary for medical devices (biliary stents), it does not contain any information about acceptance criteria or a study proving that an AI/device meets acceptance criteria. This type of document is about demonstrating substantial equivalence to a predicate device, primarily through non-clinical testing and comparison of technical characteristics, rather than through clinical studies with specific performance metrics and acceptance criteria for a novel algorithm's clinical performance.
Therefore, I cannot provide the requested information. The document focuses on regulatory approval based on equivalence for physical medical devices, not on the performance of a diagnostic or assistive AI system.
Ask a specific question about this device
(262 days)
FGE
Zimmon® Pancreatic Stents/Stent sets (SPSOF, SPSOS, ZEPDF, ZPSOF, ZPSOS); Geenen® Pancreatic Stents/ Stent Sets (GEPD, GPDS, GPSO, GPSOS); Geenen® Sof-Flex® Pancreatic Stent (GPSO-SF. GPSOS-SF); Endoscopic pancreatic stent placement is used for pancreatic drainage that could be caused by pancreatitis, stricture, pancreatic cancer, anatomic anomalies of the pancreas, pancreatic fluid collection, pancreatic stones, disrupted duct, fistula/pancreatic leak. Pancreatic stents are also used prophylactically for prevention of post-ERCP pancreatitis.
Johlin® Pancreatic Wedge Stent and Introducer Set (JPWS): Endoscopic pancreatic stent placement for pancreatic drainage of obstructed ducts that could be caused by pancreatitis, stricture, pancreatic cancer, anatomic anomalies of the pancreas, pancreatic fluid collection, pancreatic stones, and disrupted duct.
Pushing Catheter and Guiding Catheter (GC, PC): These devices are indicated for use with biliary and pancreatic stents for the following indications. For endoscopic biliary stent placement for biliary drainage of obstructed ducts that could be caused by common bile duct stones, malignant biliary obstruction and benign or malignant strictures. For endoscopic pancreatic stent placement for pancreatic duct drainage that could be caused by pancreatitis, stricture, pancreatic cancer, anatomic anomalies of the pancreatic fluid collection, pancreatic stones, disrupted duct, fistula / pancreatic leak. Pancreatic stents are also used prophylactically for prevention of post-ERCP pancreatitis.
The intended use of all Cook pancreatic stents and sets is to drain pancreatic ducts. A variety of stents in different sizes are available across the device range to accommodate various patient anatomies, the size and location of the obstruction and physician preference. They are offered in French sizes of between 3Fr and 11.5Fr, and in labelled lengths of between 2cm and 25cm.The subject devices and their components can be supplied as stent only, introducer only (guiding or pushing catheter) or as stent sets combining stent and introducers/introducer systems.
The stent sets can contain one or several of the following stent placement components; a flap protector, a guiding catheter, a pushing catheter or a dedicated introducer system. The flap protector is provided with stents that have duodenal flap-type anti-migration features, and is used to collapse the flap(s) on the device as it is introduced into the working channel of the endoscope. The function of the guiding catheter is to guide the pancreatic stent as part of its introduction to its intended location. The guiding catheter also has a Hub that allows for contrast injection. The function of the Pushing Catheter is to advance the stent, over a pre-positioned wire guide or Guiding Catheter, to its intended location within the anatomy, and to maintain the position of the Stent as it being deployed. The stents are polymeric and some of the stents have radiopaque bands. The stent designs include anti-migrational features such as duodenal pigtails, duodenal bends and ductal and duodenal flaps. To facilitate stent insertion and removal the stent ends are tapered or buffed. Side-ports on the pancreatic stents assist in drainage. All stents are deployed endoscopically over a guide wire in the same manner under fluoroscopic and endoscopic monitoring.
These Cook pancreatic stents and sets are all for professional use and are provided sterile. They are all intended for short-term use and have an indicated indwell of up to 3 months.
The provided document (K233079) describes the Cook Ireland Ltd. Zimmon and Geenen Pancreatic Stents/Stent sets, Johlin Pancreatic Wedge Stent and Introducer Set, and Pushing and Guiding Catheters. This submission is for medical devices, not an AI/ML powered device, therefore, the information requested in the prompt related to acceptance criteria and studies that prove the device meets these criteria is not applicable in the context of AI/ML performance.
Specifically, the document focuses on demonstrating substantial equivalence to a predicate device (Pancreatic Stents & Sets cleared under K172057) through comparisons of technological characteristics and non-clinical performance data. There is no mention of acceptance criteria related to AI/ML device performance metrics like sensitivity, specificity, or AUC, nor any studies involving test sets, ground truth established by experts, or human reader performance with or without AI assistance.
The performance data mentioned in the document are:
- Biocompatibility evaluation: Conducted in accordance with ISO 10993-1: 2018 and FDA's biocompatibility guidance.
- Performance testing: Included simulated use, dimensional and visual testing, tensile strength testing, MRI conditional testing, radiopacity, flow rate, and shelf-life testing.
These tests aim to ensure the physical and material integrity, safety, and functionality of the pancreatic stents and catheters.
Therefore, I cannot provide the requested information in the format of the table or answer the specific questions related to AI/ML acceptance criteria and studies based on the provided text.
Ask a specific question about this device
(202 days)
FGE
It is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract.
The 3-Stage Balloon Dilation Catheter can provide 3 different sizes and gradually increasing diameters through controlled radial expansion. The specific balloon size is printed on the label of each package and catheter. The 3-Stage Balloon Dilation Catheter is designed to pass through the working channel of the endoscope, and its guidewire cavity can accommodate a 0.035inches (0.89mm diameter) guidewire. In the package, a 0.035inches (0.89mm diameter) guidewire is pre-installed in the guidewire cavity. There is a guidewire locking device connected to the guide wire hole of the catheter. The locking device will be in the "OFF" (closed) position when it leaves the factory. Only when the switch of the locking device is in the "ON" position can it be pushed or moved forward. The guide wire is removed from the catheter. After turning the switch to the "OFF" position, the guide wire will be fixed in the catheter.
This document details the FDA's 510(k) clearance for the VedDilator™ 3-Stage Balloon Dilation Catheter. It does not describe a study involving an AI/Machine Learning device or a diagnostic device. Therefore, the requested information regarding acceptance criteria and performance of an AI/ML-based device cannot be extracted from this document.
The document focuses on the substantial equivalence of this medical device (a physical catheter) to a predicate device based on non-clinical performance testing and biocompatibility.
Here's why the requested information cannot be provided from this document:
- No AI/Machine Learning: The device is a physical catheter, not a software or AI-based diagnostic tool.
- No Diagnostic Performance Study: The document lists non-clinical bench testing for physical characteristics (e.g., burst pressure, maneuverability, sterility) and biocompatibility, not diagnostic performance metrics like accuracy, sensitivity, or specificity.
- No Human Reader Study: Since it's a physical device, there's no concept of human readers or MRMC studies.
- No Ground Truth Establishment: The "ground truth" in this context refers to the physical properties of the device, measured through standard engineering tests, not a clinical diagnosis established by experts.
Summary of available information (not directly addressing the prompt's request for AI/ML device performance):
- Device: VedDilator™ 3-Stage Balloon Dilation Catheter
- Purpose: To endoscopically dilate strictures of the alimentary tract.
- Study Type: Non-clinical bench testing and biocompatibility testing for substantial equivalence to a predicate device (K112994).
- No Clinical Study: Section 7 explicitly states, "No clinical study is included in this submission."
- Test Performed (Non-Clinical): Rated burst pressure test, Balloon tightness test, Connection firmness test, Maneuverability test, X-ray development performance test, Sterility test, Residual amount of ethylene oxide Test, Balloon fatigue Test, Burst Test of Guidewire, Bend Test of Guidewire, Connection Strength Test of Guidewire, Corrosion resistance Test of Guidewire.
- Biocompatibility Tests: Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Material mediated pyrogenicity.
- Conclusion: The device meets design specifications and demonstrates substantial equivalence to the predicate device based on these non-clinical tests.
Ask a specific question about this device
Page 1 of 36