Search Filters

Search Results

Found 24 results

510(k) Data Aggregation

    K Number
    K222635

    Validate with FDA (Live)

    Date Cleared
    2023-08-04

    (338 days)

    Product Code
    Regulation Number
    864.7415
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Premier Resolution System is an automated High Performance Liquid Chromatography (HPLC) system which performs the separation of hemoglobin species in venous whole blood samples for the quantitative analysis of normal hemoglobin (A, A2, and F), and the qualitative detection of major variant hemoglobin S, C, D-Los Angeles, and E in adult, adolescent, children and infant populations. The assays are performed on venous whole blood samples collected in tubes containing K2EDTA as anticoagulant.

    The Premier Resolution System is intended for Professional Laboratory Use only.

    The Premier Resolution System is intended for use with analytical components and reagents provided by Trinity Biotech.

    The Premier Resolution System is intended to be used in conjunction with other laboratory and clinical findings.

    For In Vitro Diagnostic Use.

    Device Description

    The Premier Resolution System consists of a high performance liquid chromatographic analyzer, reagents, analytical column and software which allows for the fractionation and quantitation of fetal hemoglobin (Hb F), and hemoglobin A2 (Hb A2), and with fractionation and presumptive identification of abnormal hemoglobin variants. This is accomplished using the principles of ion-exchange (IEX) high performance liquid chromatography (HPLC).

    AI/ML Overview

    This document describes the performance data for the "Premier Resolution System," an automated High Performance Liquid Chromatography (HPLC) system for hemoglobin analysis. The study aims to demonstrate that the device is substantially equivalent to a predicate device, the Bio-Rad Variant II ß-thalassemia (K991127).

    1. Acceptance Criteria and Reported Device Performance:

    The document outlines comparative performance against a predicate device (Bio-Rad Variant II ß-thalassemia) and precision studies. The acceptance criteria are implicitly defined by demonstrating comparability and acceptable precision, rather than explicit thresholds for each metric. The reported device performance includes:

    Acceptance Criteria (Implicit)Reported Premier Resolution System Performance (Quick Scan Assay)Reported Premier Resolution System Performance (High Resolution Assay)
    Correlation (Method Comparison) - Mean Bias vs. Predicate (Bio-Rad Variant II)
    HbF comparability-0.3 bias (1.1 to 48.9% interval) with 160 patient results-0.4 bias (1.1 to 46.6% interval) with 158 patient results
    HbA comparability0.7 bias (2.5 to 89.7% interval) with 682 patient results2.4 bias (3.5 to 90.5% interval) with 586 patient results
    HbA2 comparability0.1 bias (1.6 to 6.1% interval) with 602 patient results0.1 bias (1.6 to 6.0% interval) with 598 patient results
    HbS comparability0.3 bias (6.8 to 67.1% interval) with 106 patient samples1.3 bias (1.9 to 67.9% interval) with 110 patient samples
    HbC comparability-1.1 bias (9.5 to 82.8% interval) with 49 patient results-1.0 bias (10.2 to 82.5% interval) with 49 patient results
    HbD-LA comparability1.6 bias (11.6 to 82.7% interval) with 17 patient results2.7 bias (11.7 to 84.1% interval) with 17 patient results
    HbE comparability-3.0 bias (5.5 to 70.4% interval) with 25 patient results-4.9 bias (5.3 to 66.7% interval) with 25 patient results
    Precision (Single Site) - Within-Laboratory %CV
    HbA (High)3.61%3.63%
    HbA (Mid)0.89%1.37%
    HbA2 (Mid)2.23%7.22%
    HbA2 (Low)5.99%7.15%
    HbF (High)1.05%3.49%
    HbF (Mid)3.26%9.10%
    HbS (High)0.89%1.33%
    HbS (Mid)0.98%1.69%
    HbC (High)0.78%1.15%
    HbC (Mid)1.75%2.03%
    HbD (High)1.88%2.20%
    HbD (Mid)2.04%1.32%
    HbE (High)2.84%4.63%
    HbE (Mid)3.06%3.14%
    LoD/LoQ (Quick Scan)
    HbF LoD0.2%0.1%
    HbF LoQ1.1%1.1%
    HbA LoD0.1%0.7%
    HbA LoQ2.3%2.2%
    HbA2 LoD0.1%0.2%
    HbA2 LoQ1.5%1.5%
    HbS LoD0.1%0.3%
    HbS LoQ1.0%0.9%
    HbC LoD0.1%0.3%
    HbC LoQ1.0%1.7%
    HbD-LA LoD0.1%0.1%
    HbD-LA LoQ1.5%1.4%
    HbE LoD0.1%0.6%
    HbE LoQ1.5%2.7%

    2. Sample Size and Data Provenance (Test Set):

    • Correlation (Method Comparison): A total of 780 unique patient samples were collected and analyzed. The data provenance is described as being collected and analyzed at three (3) professional external laboratory sites. It is implicit that these were retrospective real-world samples, as they are referred to as "patient samples" used for method comparison against an existing device. The country of origin is not explicitly stated, but given the FDA submission, it is likely the US.
    • Precision (Single Site): The sample size for each analyte was 80 data points, generated by a 20x2x2 study design over 20 days, with two runs per day and two replicates per run. These were "samples of varying concentrations" and not explicitly patient samples.
    • Precision (Multisite): The sample size for each analyte was 75 data points (3x5x5 study design across three external sites, over five days with five replicates per day). These were "precision samples."
    • Limits of Detection: 60 determinations of low-level samples for each hemoglobin type. These were "human whole blood samples" with varying levels of hemoglobins.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    This device is an in-vitro diagnostic (IVD) instrument for quantitative and qualitative analysis of hemoglobin species. The ground truth for such devices is typically established by well-characterized reference methods or by comparison to a legally marketed predicate device. In this submission, the primary method for establishing the device's performance is:

    • Correlation (Method Comparison): Comparison against the Bio-Rad Variant II ß-thalassemia, which serves as the "ground truth" or reference for establishing substantial equivalence. No human experts are described as establishing ground truth for the test set, as the ground truth is the measurement from the predicate device.

    4. Adjudication Method for the Test Set:

    Not applicable. The study design involves direct comparison of quantitative measurements from two analytical instruments (device under review vs. predicate device), and precision of the device itself. There is no subjective interpretation requiring adjudication by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is an in-vitro diagnostic device that provides quantitative and qualitative measurements, not an image-based diagnostic system requiring human interpretation with or without AI assistance. Therefore, an MRMC study is not relevant.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop):

    Yes, the entire study focuses on the standalone performance of the "Premier Resolution System" as an automated HPLC system. There is no human-in-the-loop component described for its routine operation or for the performance studies presented. The device performs the analysis and provides results independently.

    7. Type of Ground Truth Used:

    The ground truth for the device's performance is established mainly through:

    • Comparison to a legally marketed predicate device: The Bio-Rad Variant II ß-thalassemia, for method comparison (correlation).
    • Internal consistency and statistical measures: For precision (repeatability, within-laboratory, reproducibility) and limits of detection/quantitation. This relies on the inherent analytical capabilities and controls of the new device itself.

    8. Sample Size for the Training Set:

    The document does not explicitly describe a "training set" in the context of machine learning or AI models. This device is an automated HPLC system, where performance is based on chemical separation principles, not a learning algorithm that requires a separate training phase with labeled data in the AI sense. Its "training" or development would involve chemical and engineering optimization.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention of a "training set" in the context of an AI/ML model for this device. The development of an HPLC system involves instrument design, reagent formulation, and software development, which are validated through analytical performance studies like those presented (precision, linearity, method comparison, etc.).

    Ask a Question

    Ask a specific question about this device

    K Number
    K172254

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-23

    (89 days)

    Product Code
    Regulation Number
    866.3830
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lyme B. burgdorferi (IgM) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgM antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2-EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

    Device Description

    The kit is an immunoblot method to detect IgM antibodies against B. burqdorferi antigens. The test kit contains:
    . Nitrocellulose Test Strips with purified B. burgdorferi antigens (3) and quality control lines (3) present in specific positions
    . Sample Diluent. Provided for specimen dilutions. Contains BSA and PBS
    Positive Control derived from human serum positive for Lyme disease. Contains <0.1% sodium azide
    Negative Control derived from human serum negative for Lyme disease. Contains <0.1% sodium azide
    Conjugate. Antihuman IgM-HRP Conjugate binds reactive antibodies to the Substrate
    TMB Substrate. Provides colorimetric reaction for visual read of bound antibodies
    PBS Wash Buffer concentrate. Removes reagents and unbound antibodies after incubation steps. Must be reconstituted to 1L with distilled or deionized water.

    To perform the test, serum or plasma is individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diluted, ready-to-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 2 (or more) of the 3 test lines are considered positive for specific IgM antibody to B. buradorferi.

    AI/ML Overview

    The provided document describes the Lyme B. burgdorferi (IgM) MarStripe Test, an immunoblot assay for detecting IgM antibodies to Borrelia burgdorferi. The submission aims to establish substantial equivalence to a legally marketed predicate device, the MarDx B. burgdorferi IgM MarBlot Strip Test System (K951709).

    Here's an analysis of the acceptance criteria and the study data:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for performance metrics. Instead, the study aims to demonstrate substantial equivalence to a predicate device. Therefore, the "acceptance criteria" can be inferred as achieving comparable performance to the predicate device and demonstrating acceptable analytical characteristics.

    Performance MetricImplied Acceptance Criteria (based on predicate equivalence)Reported Device Performance (Lyme B. burgdorferi (IgM) MarStripe Test)
    Method Comparison (vs. Predicate IgM WB for EIA-positive/equivocal samples)Substantial equivalence to predicate device's agreement percentages.Positive % Agreement: 93.5% (95% CI: 90.1% - 95.8%) Negative % Agreement: 93.5% (95% CI: 90.2% - 95.7%)
    Analytical Specificity (Normal Individuals)High specificity (e.g., >95%)99.5% (95% CI: 97.1% - 100%)
    Cross-reactivityAll positive results should be confirmed by predicate device. Low overall false positivity rate.All 7 positive specimens (out of 246 potentially cross-reactive) by MarStripe Test were confirmed positive by predicate IgM Western blot device. Overall positive specimens in cross-reactive cohort: 7 (2.8%)
    Sensitivity Comparison (Well-characterized Lyme disease specimens)Comparable sensitivity to predicate device.18.4% (16/87) (95% CI: 11.2% - 28.4%)
    Sensitivity Comparison (CDC Panels)Comparable sensitivity to predicate device.Overall: 23.8% (10/42)
    Precision (Intra-laboratory)High qualitative agreement, especially for negative and strongly positive samples.Low Negative and Moderate Positive samples: 100% agreement. Low Positive: 99.0% positive agreement. Cutoff: 81.3% positive agreement.
    Reproducibility (Inter-laboratory)High qualitative agreement across sites and operators.Low Negative and one Moderate Positive samples: 100% final positive/negative agreement. One High Negative: 99.7% negative agreement. One Low Positive: 99.3% positive agreement. Cutoff: 68.4% positive agreement.
    Interference100% qualitative agreement with and without interfering agents.100% qualitative agreement for all tested specimens.
    Serum vs. Plasma Matrix Comparison100% qualitative agreement between serum and plasma samples.100% qualitative agreement for all 20 pairs.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison: 676 specimens tested (those found positive or equivocal on an FDA-cleared first-step EIA). The study was a prospective study performed at three geographically distinct study sites. The country of origin for the data is not explicitly stated but implied to be the USA given the FDA context and reference to CDC.
    • Analytical Specificity: 220 sera from normal individuals (blood bank donors) representing endemic geographic regions of the United States. Data provenance is from US blood bank donors, retrospective or prospective not specified but typical for such collections.
    • Cross-reactivity: 246 potentially cross-reactive specimens from individuals with various other infectious conditions. Data provenance is not specified.
    • Sensitivity: 87 well-characterized Lyme disease clinical specimens. Samples included early, early disseminated, and late phases. Data provenance is not specified but implied to be clinical.
    • CDC Panels: 10 samples from the Lyme Disease Validation Panel and 32 samples from the Lyme Disease Basic Research Panel (Total 42 samples). These are reference panels from the Centers for Disease Control and Prevention (CDC).
    • Precision: 8 specimens, each tested in 4 replicates, two runs per day over 12 days (total 96 tests per specimen).
    • Reproducibility: 8 specimens, each tested in 4 replicates, two runs per day over 12 days, at each of three laboratory sites (total 192 tests per specimen per site, or 576 read-outs in total).
    • Interference: 2 Lyme IgM negative and 3 IgM positive sera.
    • Serum vs. Plasma Matrix Comparison: 20 pairs of sera/plasma (total 40 samples).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test sets.

    • For the Method Comparison study, the ground truth was established by comparison to results from an "FDA cleared immunoblot results" (the predicate device) which followed "recommended criteria described by the Centers for Disease Control (CDC) and the Second National Conference on Serological Diagnosis of Lyme Disease". This implies standardized interpretation criteria rather than individual expert consensus.
    • For the Sensitivity study, "87 well characterized Lyme disease clinical specimens" were used, implying their disease status was previously determined, likely through a combination of clinical diagnosis and laboratory testing, but specific expert involvement is not detailed.
    • For the CDC Panels, the ground truth is the CDC's characterization of these reference panels.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set interpretation by human readers. For the reproducibility study, results were read by "2 human operators at each site," but it does not specify if or how discrepancies between these operators were resolved (e.g., 2+1, 3+1). The "ground truth" for the overall studies appears to be based on established reference methods or clinical characterization rather than adjudicated interpretations of the MarStripe test itself.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described in the context of human readers improving with AI vs. without AI assistance. The studies performed were primarily focused on the device's analytical performance and its agreement with a predicate device and established reference panels.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself (Lyme B. burgdorferi (IgM) MarStripe Test) is an immunoblot assay that requires visual reading. The results show "Positives", "Negatives", "Band Type" (Pos., Neg., Cut., Wpos.), implying visual interpretation by a human. For example, the reproducibility study explicitly mentions "Results were read by 2 human operators at each site". Therefore, this is not a standalone algorithm without human-in-the-loop performance; human visual interpretation is an integral part of the test.

    7. The Type of Ground Truth Used

    The ground truth varied depending on the study:

    • Method Comparison: Ground truth was the results from an FDA-cleared predicate IgM Western blot device, interpreted according to CDC and Second National Conference on Serological Diagnosis of Lyme Disease criteria.
    • Analytical Specificity: Implied ground truth is "normal individuals" from blood banks, presumed negative for Lyme disease.
    • Cross-reactivity: Ground truth was the clinical diagnosis of other infectious conditions, with validation against the predicate IgM Western blot device for any MarStripe positive results.
    • Sensitivity: "Well characterized Lyme disease clinical specimens," implying clinical diagnosis of Lyme disease stages.
    • CDC Panels: Characterization provided by the CDC for their reference panels.
    • Precision/Reproducibility/Interference/Serum vs. Plasma: The ground truth for these analytical studies was related to the known characteristics of the samples (e.g., known negative, known positive, spiked with interferents, paired serum/plasma) as determined by prior testing (e.g., FDA-cleared B. burgdorferi ELISA results).

    8. The Sample Size for the Training Set

    The document describes performance studies for the Lyme B. burgdorferi (IgM) MarStripe Test. This device is a diagnostic assay, not an AI/ML model that requires a "training set" in the computational sense. The studies presented are validation studies of the finished device. There is no mention of an algorithm development phase with a dedicated training set.

    9. How the Ground Truth for the Training Set Was Established

    Since the device is a diagnostic assay (not an AI/ML model), there is no "training set" as understood in machine learning. Therefore, this question is not applicable to the information provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163095

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-02-01

    (89 days)

    Product Code
    Regulation Number
    866.3830
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lyme B. burgdorferi (IgG) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgG antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2- EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

    Device Description

    The kit is an immunoblot method to detect IgG antibodies against B. burqdorferi antigens. The test kit contains: Nitrocellulose Test Strips with purified B. burgdorferi antigens (10) and quality control lines (3) present in specific positions, Sample Diluent, Positive Control, Negative Control, Conjugate (Antihuman IgG-HRP Conjugate), TMB Substrate, and PBS Wash Buffer concentrate. To perform the test, serum or plasma is added to individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diuted, ready-to-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 5 (or more) of the 10 test lines are considered positive for specific IgG antibody to B. burgdorferi.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Lyme B. burgdorferi (IgG) MarStripe Test, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined acceptance criteria for performance metrics like sensitivity and specificity percentage targets. However, the reported performance against the predicate device and consensus can be considered the de-facto acceptance criteria for equivalence.

    Performance MetricAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance (Lyme B. burgdorferi (IgG) MarStripe Test)
    Reproducibility100% positive or negative agreement for all specimens across sites and operators100% (for overall positive/negative agreement across 8 specimens, 3 sites, 2 operators, 432 readouts)
    Analytical SpecificityNot explicitly stated, but high specificity is expected, ideally comparable to predicate.99.5% (95% CI: 97.1% - 100%) against 219 healthy blood donors
    Cross-reactivityLow positive incidence, and positive specimens should be confirmed by predicate device.4.4% positive incidence across 136 potentially cross-reactive specimens; all positives confirmed by predicate IgG Western blot.
    Interference100% qualitative agreement with and without interfering agents.100% qualitative agreement for all specimens (5 sera spiked with hemoglobin, bilirubin, RF, triglycerides, cholesterol).
    Serum vs. Plasma Equivalence100% qualitative agreement between serum and plasma samples.100% qualitative agreement for all 20 pairs of serum/plasma samples.
    Positive Agreement (vs. Predicate)High agreement (e.g., >90%) deemed substantially equivalent to predicate.92.9% (95% CI: 88.6% - 95.7%)
    Negative Agreement (vs. Predicate)High agreement (e.g., >95%) deemed substantially equivalent to predicate.97.7% (95% CI: 95.9% - 98.7%)
    Sensitivity (Clinical Specimens, vs. Predicate)Similar sensitivity to the predicate device.22.3% (21/94) (95% CI: 14.7%-32.3%) - Predicate: 20.2% (19/94)
    Sensitivity (CDC Panels, vs. Predicate)Similar specificity to the predicate device.16.7% (7/42) - Predicate: 16.7% (7/42)

    2. Sample Size Used for the Test Set and Data Provenance

    • Reproducibility Test: 8 specimens (2 low negative, 2 high negative, 2 low positive, 2 moderate positive). Each specimen tested in 4 replicates, 2 runs per day, for a total of 72 tests per specimen at three laboratory sites.
    • Analytical Specificity: 219 sera from healthy blood donors.
    • Cross-reactivity: 136 potentially cross-reactive specimens from individuals with various other or infectious conditions.
    • Interference: 5 sera (2 Lyme IgG negative, 3 Lyme IgG positive).
    • Serum vs. Plasma Comparison: 10 pairs of native serum/plasma samples, plus 10 pooled serum/plasma samples (total 20 pairs).
    • Method Comparison (Clinical Tests): 753 prospectively collected consecutive specimens, all initially positive on an FDA-cleared first-step EIA.
    • Sensitivity (Clinical Specimens): 94 well-characterized Lyme disease clinical specimens (from the 753 specimens in Method Comparison).
    • CDC Panels: 42 specimens from CDC (Lyme Disease Validation Panel n=10, Lyme Disease Basic Research Panel n=32).

    Data Provenance:

    • Healthy Blood Donors (Analytical Specificity): Representing endemic geographic regions of the United States.
    • Clinical Specimens (Method Comparison & Sensitivity): Prospectively collected consecutive specimens tested at three sites. No further country of origin or specific demographic details are provided.
    • CDC Panels: From the Center for Disease Control and Prevention.

    Retrospective or Prospective:

    • Method Comparison (Clinical Specimens): Explicitly stated as "prospectively collected consecutive specimens." The other studies do not explicitly state but generally, these types of validation studies are prospective or involve carefully selected retrospective samples to meet specific criteria.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    There is no mention of external human experts establishing the ground truth for the test set in the conventional sense of, for example, radiologists interpreting images.

    • Reproducibility: Results were read by 2 human operators at each of the 3 sites. No specific qualifications (e.g., years of experience) are provided for these operators, who are presumably laboratory personnel following a protocol.
    • Method Comparison: The ground truth for comparing the device primarily relies on the predicate device (MarDx B. burgdorferi IgG MarBlot Strip Test System).
    • Discrepant Analysis: For the 29 discrepant specimens, two additional commercially available B. burgdorferi IgG Western blot methods were used as comparators to establish a "consensus result." This implies laboratory methods rather than expert human interpretation.
    • CDC Panels: Ground truth is established by the CDC's characterization of the panels.

    4. Adjudication Method for the Test Set

    • Reproducibility: No formal adjudication method like "2+1" is mentioned for individual band interpretation. The "Final positive or negative agreement was 100% for all specimens" suggests direct agreement between operators or that any differences were resolved internally to reach a unified result for each specimen.
    • Discrepant Analysis (Method Comparison): A "2/3 comparator" consensus was used where the Lyme B. burgdorferi (IgG) MarStripe Test result was compared to results from two additional commercially available B. burgdorferi IgG Western blot methods.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done in the context of human readers improving with AI vs. without AI assistance. This device is a diagnostic test kit (immunoblot assay) and not an AI-powered image analysis system that would typically involve human-in-the-loop performance evaluation. The "human operators" mentioned in reproducibility are interpreting the results of the kit, not an AI output.

    6. Standalone Performance Study

    • Yes, a standalone performance study was done. The entire set of non-clinical and clinical tests (reproducibility batch, analytical specificity, cross-reactivity, interference, serum vs. plasma testing, and the initial method comparison against the predicate) represents the standalone performance of the device/kit itself. The "algorithm" here refers to the immunodetection process and interpretation rules inherent to the diagnostic kit, not a separate AI algorithm.

    7. Type of Ground Truth Used

    • Reproducibility: Based on prior FDA-cleared B. burgdorferi ELISA results (categorizing samples as low/high negative, low/moderate positive).
    • Analytical Specificity: Healthy blood donors (presumed negative for Lyme).
    • Cross-reactivity: Individuals with other infectious or autoimmune conditions (used to challenge specificity).
    • Interference: Artificially spiked samples.
    • Serum vs. Plasma Comparison: Comparison to results from an FDA-cleared Lyme EIA assay and subsequent Western blot IgG positivity/negativity.
    • Method Comparison (Clinical Tests): The primary ground truth was the predicate device's result. For discrepant samples, a consensus of two additional commercial IgG Western blot methods served as the ground truth.
    • Sensitivity (Clinical Specimens): "Well characterized Lyme disease clinical specimens" from the clinical study, with characterization across disease stages (early, early disseminated, late).
    • CDC Panels: Characterized reference panels provided by the Center for Disease Control and Prevention.

    8. Sample Size for the Training Set

    No training set is explicitly mentioned or applicable in the context of this device. This is a diagnostic test kit that functions based on established biochemical reactions and interpretation algorithms, not a machine learning or AI model that requires a separate training phase. The described studies are validation studies for the finished product.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for an AI/ML model, this question is not applicable. The device's "ground truth" (i.e., its intended performance characteristics) is established through the development and analytical/clinical validation studies presented.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140455

    Validate with FDA (Live)

    Date Cleared
    2014-05-22

    (87 days)

    Product Code
    Regulation Number
    866.3520
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Measles IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings.

    Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.

    Device Description

    The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assays (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H2SO4. The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the studies performed for the Trinity Biotech Captia™ Measles IgM ELISA Test Kit, presented in the requested format:

    Acceptance Criteria and Device Performance for Trinity Biotech Captia™ Measles IgM ELISA

    The provided document describes the performance characteristics and clinical studies of the Trinity Biotech Captia™ Measles IgM ELISA. The acceptance criteria are not explicitly stated as pass/fail thresholds in a dedicated table, but are inferred from the reported performance characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Analytical Performance
    High Dose Hook EffectNo dose hook effect (prozone) of antigenAntigen is titrated to ensure no dose hook effect. Positive control in each kit lot is titrated to negative at each kit QC test stage to ensure no prozoning.
    IgM/IgG CompetitionDiminished competing virus-specific IgG and minimized RF interferenceSerum Diluent Plus Solution diminishes competing virus-specific IgG (maximum 1380 mg/dL removed) and minimizes rheumatoid factor interference (highest titer of RF+ tested (1:1280; 500 IU/mL) did not adversely affect performance).
    Precision (Interassay)Acceptable %CV across sites, operators, and daysInterassay (Between All Sites): %CV ranged from 4.3% to 12.0% across 6 samples (averaged across 3 sites, each with 2 operators, 10 days, triplicate runs; N=180 data points/sample). Inter-Lot Reproducibility: %CV ranged from 3.19% to 22.55% across 6 samples (averaged across 3 kit lots, triplicate runs). Intra-run (Between Users): %CV ranged from 0.40% to 5.00% across 6 samples (averaged across 2 operators, 10 days, triplicate runs; N=60 data points/sample).
    Analytical Specificity (Cross-Reactivity)Limited or acceptable cross-reactivity with common interfering agentsObserved cross-reactivity: CMV IgM (3 positive, 1 equivocal out of 7), HSV1 IgM (7 positive out of 16), HSV2 IgM (2 positive, 1 equivocal out of 10). No cross-reactivity observed: Mumps IgM (4/4 negative), VZV IgM (4/4 negative), Rubella IgM (4/4 negative). Not ruled out: Parvovirus, Respiratory Syncytial Virus (RSV), Parainfluenza.
    Analytical Specificity (Interfering Substances)No adverse results or significant change with tested interferentsNo adverse results or significant change observed with Hemoglobin (≥ 500 mg/dL) and Bilirubin (unconjugated, ≥ 20 mg/dL). Not tested: Red cells, cholesterol, gamma globulin, triglycerides, beta carotene, protein, ascorbic acid, anticoagulants.
    Assay Cut-off DeterminationClear distinction between negative, equivocal, and positiveCut-off ISR = 1.00. Values ≤ 0.90 considered negative, values ≥ 1.10 considered positive, and 0.91 to 1.09 considered equivocal. Based on a study of 228 Measles IgM negative sera (mean OD=0.105, SD=0.070), positive threshold determined as mean + 4 SD = 0.385.
    SeroconversionConsistent IgM response with seroconversion panelDemonstrated IgM response consistent with seroconversion panel (BIOMEX SCP-MEA-001) across 4 different lots. Trinity kit showed seroconversion at Day 20.
    Specificity in Normal PopulationLow incidence of positive results in healthy individualsOut of 200 random serum samples, 0.5% (1 sample) was in the 0.91-1.10 equivocal range, 0.5% (1 sample) was in the 1.21-1.40 positive range, and 0.5% (1 sample) was >2.00 positive. The majority (approximately 92.5%) were in the 0.00-0.60 negative range.
    Clinical Performance
    Positive Agreement (Study 1 & 2)High agreement with comparator algorithm for positive samples90.7% Positive Agreement (88/97 samples) with 95% CI (83.3-95.0%) (when indeterminate comparator algorithm results were included in total).
    Negative Agreement (Study 1 & 2)High agreement with comparator algorithm for negative samples80.5% Negative Agreement (70/87 samples) with 95% CI (70.9-87.4%) (when indeterminate comparator algorithm results were included in total).
    Equivocal Rate (Study 1 & 2)Low percentage of equivocal results8.6% (16 of 187) of Captia™ Measles IgM results were equivocal.
    Positive Agreement (Study 3)High agreement with comparator PCR algorithm for positive samples75.00% Positive Agreement (3/4 samples) with 95% CI (30.1-95.4%) (when indeterminate comparator algorithm results were included in total).
    Negative Agreement (Study 3)High agreement with comparator PCR algorithm for negative samples100.00% Negative Agreement (4/4 samples) with 95% CI (39.8-100.0%) (when indeterminate comparator algorithm results were included in total).

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Studies 1 & 2): 188 samples from suspected measles infection cases.
      • Data Provenance: Retrospective, collected from individuals in whom a measles IgM test was ordered by the Texas Department of Health, Austin, Texas (66 samples) and the State Laboratory of Public Health in Raleigh, North Carolina (122 samples), USA.
    • Test Set (Clinical Study 3): 8 samples.
      • Data Provenance: Retrospective, collected from submissions to the Kansas Department of Health and Environment Laboratory (KDHE), Topeka, Kansas, USA.
    • Test Set (Specificity in a Normal Population): 200 random serum samples.
      • Data Provenance: Not explicitly stated, but implied to be from a general population (individuals between ages 18-65 with no known clinically apparent Measles infection). Location not specified.
    • Test Set (Cross-Reactivity): 47 samples.
      • Data Provenance: Confirmed positive for other IgM antibodies (CMV, HSV1, HSV2, Rubella, VZV, Mumps, Parvo-B19) by Trinity Biotech Captia™ test kits. Location not specified.
    • Test Set (Precision/Reproducibility): 6 blinded panel members (4 low-to-moderate positive, 2 negative).
      • Data Provenance: Not specified, likely internal or commercially sourced panels.

    3. Number of Experts and Qualifications for Ground Truth

    • The document does not specify the number of experts used to establish the ground truth for the test sets in the clinical studies.
    • The ground truth in Clinical Studies 1 & 2 was established using a "comparator testing algorithm" which consisted of "the results of other laboratory confirmation methods including comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutionation Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit." This implies multiple laboratory results were used to form a consensus or a defined algorithm, rather than individual expert adjudication of each case.
    • Similarly, in Clinical Study 3, the "comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples."

    4. Adjudication Method for the Test Set

    • For Clinical Studies 1 & 2, the ground truth was established by a "comparator algorithm" based on multiple laboratory confirmation methods (e.g., other ELISAs, IFA, CF, HI). The specific adjudication rule (e.g., 2 out of 3, 3 out of 4) is not explicitly detailed, but it generally implies a form of consensus derived from multiple test results rather than human expert adjudication.
    • For Clinical Study 3, the ground truth was established by a "comparator algorithm" based on PCR results from various sample types. Again, the specific adjudication rules for discrepant PCR results (e.g., "NPS negative at two sites, urine positive at one site and urine indeterminate at one site") are not explicitly detailed for how the final "positive," "indeterminate," or "negative" comparator algorithm result was reached.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not performed. This device is an Enzyme-Linked Immunosorbent Assay (ELISA), which is an automated or semi-automated laboratory test. It does not involve human readers interpreting images or data where AI assistance would be applicable in the traditional sense of an MRMC study for imaging diagnostics. The "readers" are the laboratory technicians operating the ELISA equipment.

    6. Standalone Performance Study

    • Yes, standalone performance was done for the device. All reported performance metrics (analytical performance, seroconversion, cross-reactivity, clinical agreement with comparator algorithms) represent the performance of the Captia™ Measles IgM ELISA as an algorithm/device only, without human "in-the-loop" assistance for interpreting the result of the ELISA itself. While humans perform the test and interpret the final ISR value against predefined cut-offs, the "algorithm" is determining the ISR.

    7. Type of Ground Truth Used

    • The ground truth used was a combination of other laboratory confirmation methods and algorithms.
      • For Clinical Studies 1 & 2: "comparator testing algorithm used at each institution to determine the presence of current or recent measles infection" which included "comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutionation Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit." This represents a form of expert-defined algorithm/consensus from multiple laboratory tests.
      • For Clinical Study 3: "comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples." This is essentially molecular pathology/laboratory results.
      • For Cross-Reactivity: Confirmed positive status based on other specific IgM ELISA tests.
      • For Seroconversion: A commercially available seroconversion panel.

    8. Sample Size for the Training Set

    • The document does not explicitly state a separate "training set" in the context of machine learning model development. This is an ELISA kit, not typically a machine learning algorithm that requires a distinct training phase in the same way. The development and optimization of the assay (e.g., antigen titration, calibrator development, component formulation) would be part of an iterative development process that could be considered analogous to "training," but not in the sense of a labeled dataset for an AI model.
    • The cut-off determination study involved 228 Measles IgM negative sera, which were used to establish the positive threshold (mean + 4SD). This could be considered a form of "training" or "calibration" data for the assay's interpretive criteria.

    9. How the Ground Truth for the Training Set Was Established

    • As noted above, there isn't a traditional "training set" for a machine learning algorithm.
    • For the cut-off determination, the 228 Measles IgM negative sera were explicitly described as "Measles IgM negative sera" and were used to calculate statistical parameters (mean and standard deviation of optical density readings) to define the initial positive threshold. Four samples were disqualified because they were in the equivocal range on Trinity IgM ELISA or indeterminate on a comparator Measles IFA test kit (due to non-specific staining). This implies an initial screening or pre-classification of these samples as "negative" before their use in cut-off calculation. The final cut-off was derived using these empirical observations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120001

    Validate with FDA (Live)

    Device Name
    UNIGOLD GIARDIA
    Manufacturer
    Date Cleared
    2013-03-01

    (423 days)

    Product Code
    Regulation Number
    866.3220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Trinity Biotech Uni-Gold™ Giardia is a single use rapid immunoassay for the qualitative detection of Giardia lamblia ( G. lamblia ) antigens in human stool specimens. This test is intended for use with patients with gastrointestinal symptoms as an aid in the diagnosis of suspected Giardia gastrointestinal infections. As with other Giardia tests, results should be considered in conjunction with the clinical evaluation and medical history. For In-Vitro Diagnostic use.

    Device Description

    The Trinity Biotech Uni-Gold™ Giardia was designed as a single use, rapid, lateral flow immunoassay test device to detect the presence of Giardia lamblia antigen in unpreserved (fresh & frozen), preserved and media containing human stool specimens.

    The Trinity Biotech Uni-Gold™ Giardia test strip (5mm x 60mm) combines a Nitrocellulose Membrane with designated fiber pads (conjugate, sample and absorbant). The test strip is then placed into a plastic housing and is sealed constituting the Test Device.

    The Giardia Nitrocellulose Membrane Test Strip - above consist of

    • A) Mouse anti-Giardia lamblia antibody is coated onto the Test Line region of the test strip.
    • B) Rabbit anti-Goat IgG antibody is coated onto the Control Line region of the Test Strip.
    • C) Goat anti- Giardia lamblia antibodies and Goat IgG antibodies are conjugated to red latex particles and dried onto the inert glass fiber (Conjugate Pad) which is inserted into the test strip below the nitrocellulose zone.

    When Giardia antigens are present in the sample they combine with the antibody/red latex complex. As the complex migrates it binds to the antibodies in the test region forming a visible pink/red band. (Picture B) This forms the basis for the double antibody sandwich assay. Excess conjugate forms a second pink/red band in the control region of the device. The control line should always appear as a visible pink/red band in the control region of the device. This internal control line is to ensure and indicate that the test device is functioning correctly.

    The plastic housing device contains a window where the diluted stool sample is added (Sample Well) and a window above where the results are read in 15 minutes.

    The test concept:

    Mouse anti-Giardia lamblia is coated onto the test line region of the nitrocellulose zone of the test strip. Rabbit anti- Goat IgG is coated onto the control line region.

    Goat anti-Giardia lamblia antibodies are conjugated to red latex particles and dried onto inert glass fiber. This is inserted into the test strip below the nitrocellulose zone.

    A buffered solution is added to a dilution tube followed by the addition of the stool specimen (2 drops) via a disposable pipette. This mixture is then dispensed in total into the sample well of the lateral flow cartridge device with a dropper pipette and migrates through a pad containing red microspheres that have been coated with a antibody specific for the Giardia antigen. If the antigen is present, an immune complex forms. The migration continues along the membrane, which contains a striped down anti Giardia capture antibody. If Giardia antigen is present, the immune complex reacts with the anti-Giardia antibody at the test line on the membrane.

    Thus Giardia antigens present in the sample combine with the antibody/red latex. As this complex migrates it binds to the antibodies in the test region forming a visible pink/red band.

    Excess conjugate forms a second pink/red band in the control region of the device. The control line should always appear as a visible pink/red band in the control region of the device to indicate that the test device is functioning correctly.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the Uni-Gold™ Giardia device based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in a quantitative format (e.g., "Sensitivity must be >90%"). However, based on the studies presented, the implicit acceptance criteria are 100% agreement/sensitivity/specificity in the evaluated studies.

    Performance MetricAcceptance Criteria (Implicit)Reported Device PerformanceStudy TypeN (positive/negative)
    Intra-run Precision/Reproducibility100% reproducibility100% reproducibilityInternal (initial)6 blind panel members (2 Low Pos, 2 High Pos, 2 Neg)
    100% reproducibilityInternal (additional)12 blind panel members (4 Low Pos, 4 High Pos, 4 Neg)
    Agreement vs. Comparator DeviceHigh percent agreementSite 1: 100% Pos Agr, 94.1% Neg AgrRetrospective26 Pos, 48 Neg (Uni-Gold negative & Comparator Negative, 3 Uni-Gold positive & Comparator negative)
    Site 2: 100% Pos Agr, 100% Neg AgrRetrospective51 Pos, 49 Neg
    Site 3: 100% Pos Agr, 83.3% Neg AgrRetrospective54 Pos, 30 Neg (Uni-Gold negative & Comparator Negative, 6 Uni-Gold positive & Comparator negative)
    Sensitivity vs. DFA Microscopy100% Sensitivity100% (91/91) (95% CI 95-100%)Retrospective91 Pos
    Specificity vs. DFA Microscopy100% Specificity100% (150/150) (95% CI 97-100%)Retrospective150 Neg
    Positive Percent Agreement (PPA) vs. Non-Fluorescent Microscopy (Wheatley's Stain)100% PPA100% (22/22)Retrospective22 Pos
    Negative Percent Agreement (NPA) vs. Non-Fluorescent Microscopy (Wheatley's Stain)100% NPA100% (45/45)Retrospective45 Neg
    Positive Percent Agreement (PPA) vs. Non-Fluorescent Microscopy (Iron Hematoxylin Stain)100% PPA100% (60/60)Retrospective60 Pos
    Negative Percent Agreement (NPA) vs. Non-Fluorescent Microscopy (Iron Hematoxylin Stain)100% NPA100% (199/199)Retrospective199 Neg
    Specificity vs. DFA Microscopy (Prospective Study)100% Specificity100% (378/378) (95% CI 99-100%)Prospective378 Neg

    2. Sample Sizes and Data Provenance

    • Test Set for Agreement vs. Comparator Device:

      • Total Samples: 267 retrospective samples.
      • Matrix types: unpreserved frozen (42), C&S (15), SAF (139), and formalin (71).
      • Country of Origin: Not explicitly stated, but the studies were conducted at 3 external laboratories. Given the "US and Canada" mention for overall sample collection, it's likely from these regions.
      • Retrospective/Prospective: Retrospective.
    • Test Set for Sensitivity/Specificity vs. DFA Microscopy (Retrospective):

      • Total Samples: 241 (91 positive, 150 negative).
      • Positive matrix types: formalin (48), SAF (13), unpreserved frozen (17), Cary Blair (3), and C&S (10).
      • Negative matrix types: formalin (42), SAF (70), unpreserved frozen (25), Cary Blair (3), and C&S (10).
      • Country of Origin: Not explicitly stated, but collected from "Hospitals throughout the US and Canada."
      • Retrospective/Prospective: Retrospective.
    • Test Set for Additional Retrospective Studies (Non-Fluorescent Microscopy):

      • Site 2 (Wheatley's Stain): 67 retrospective samples (22 positive, 45 negative).
      • Site 3 (Iron Hematoxylin Stain): 259 retrospective samples (60 positive, 199 negative).
      • Country of Origin: Not explicitly stated, but likely from "US and Canada."
      • Retrospective/Prospective: Retrospective.
    • Test Set for Prospective Study (DFA Microscopy):

      • Total Samples: 378 negative samples.
      • Matrix types: unpreserved fresh (153), unpreserved frozen (45), formalin (45), SAF (45), C&S (45), and Cary Blair (45).
      • Country of Origin: Not explicitly stated, but collected from "Hospitals throughout the US and Canada."
      • Retrospective/Prospective: Prospective.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the "number of experts" used to establish ground truth or their specific qualifications (e.g., "Radiologist with 10 years of experience"). However, the ground truth methods (DFA microscopy, Wheatley's Stain, Iron Hematoxylin Stain) are laboratory diagnostic techniques typically performed by trained medical technologists or laboratory personnel. It implies that these established methods served as the expert-level reference.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The "ground truth" was established by the reference methods (DFA microscopy, Wheatley's Stain, Iron Hematoxylin Stain). In cases where the Uni-Gold device and the initial comparator device disagreed (e.g., in the "Agreement vs. Comparator Device" study), further resolution was sought through DFA microscopy or Iron Hematoxylin Stain, which effectively acted as an adjudicator to determine the true positive/negative status for those discrepant samples.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was done. The study compares the device performance to established laboratory methods or an existing comparator device, not the improvement of human readers with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, standalone performance was evaluated. The studies directly compare the Uni-Gold™ Giardia device's results (algorithm only, as it's a rapid immunoassay) against the chosen established ground truth methods (DFA microscopy, non-fluorescent microscopy, or a comparator device). There is no "human-in-the-loop" aspect to the device's reading mechanism described.

    7. Type of Ground Truth Used

    The types of ground truth used were primarily:

    • DFA Microscopy: Direct Fluorescent Antibody microscopy.
    • Non-Fluorescent Microscopy: Specifically, Wheatley's Stain and Iron Hematoxylin Stain.
    • Comparator Device: Remel Xpect™ Giardia Lateral Flow Assay (510K #: K031942) for initial agreement studies, with discrepant results often adjudicated by DFA or Iron Hematoxylin Stain.

    8. Sample Size for the Training Set

    The document describes a rapid immunoassay device. It does not mention a "training set" in the context of machine learning. The device is a chemical/immunological test, not an algorithm that requires a training set in the AI sense. The development of the assay itself would have involved internal validation and optimization, but not a distinct "training set" as understood in AI studies.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" for an AI algorithm, this question is not applicable to the Uni-Gold™ Giardia immunoassay device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121565

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-02-08

    (255 days)

    Product Code
    Regulation Number
    866.3220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Trinity Biotech Uni-Gold™ Cryptosporidium is a single use rapid immunoassay for the qualitative detection of Cryptosporidium parvum (C. parvum) antigens in human stool specimens. This test is intended for use with patients with gastrointestinal symptoms as an aid in the diagnosis of suspected Cryptosporidium gastrointestinal infections. As with other Cryptosporidium tests, results should be considered in conjunction with the clinical evaluation and medical history. For In-Vitro Diagnostic use.

    Device Description

    The Trinity Biotech Uni-Gold™ Cryptosporidium Test was designed as a single use, rapid, lateral flow immunoassay to detect the presence of Cryptosporidium parvum antigen in unpreserved (fresh & frozen), preserved, and media containing human stool specimens. The Trinity Biotech Uni-Gold™ Cryptosporidium test strip (5mm x 60mm) combines a nitrocellulose membrane with designated fiber pads (conjugate, sample and absorbant). The test strip is then placed into a plastic housing and is sealed constituting the Test Device.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Uni-Gold™ Cryptosporidium device, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state pre-defined "acceptance criteria" in terms of numerical thresholds for sensitivity and specificity. Instead, it presents the results of studies and aims to demonstrate substantial equivalence to a predicate device. For the purpose of this request, I will infer the implicit acceptance is achieving high sensitivity and specificity in comparison to gold standard methods and predicate devices.

    Metric / StudyAcceptance Criteria (Implicit)Reported Device Performance (Uni-Gold™ Cryptosporidium)
    Sensitivity (vs. DFA Microscopy)High agreement with DFA microscopy100% (77/77) 95% CI 94 - 100%
    Specificity (vs. DFA Microscopy)High agreement with DFA microscopy100% (157/157) 95% Cl 97 - 100%
    PPA (vs. Modified Acid-Fast Stain)High positive percent agreement100% (26/26)
    NPA (vs. Modified Acid-Fast Stain)High negative percent agreement100% (21/21)
    PPA (vs. Modified Kinyoun Stain)High positive percent agreement92% (55/60)
    NPA (vs. Modified Kinyoun Stain)High negative percent agreement90% (198/221)
    Specificity (Prospective Study vs. DFA Microscopy)High specificity (as no positives were found)100% (378/378) 95% Cl 99 - 100%
    Inter-run Precision/Reproducibility100% reproducibility across sites and days100% reproducibility observed for all sites, for all days
    Cross-ReactivityNo cross-reactivity with common interfering organisms/substancesNo cross-reactivity observed with listed organisms; No test interference observed by listed compounds

    2. Sample Size Used for the Test Set and Data Provenance

    • Comparator Device Study (Side-by-side with Remel Xpect®): 299 retrospective samples. Data provenance: United States (multiple external sites). Sample matrices included unpreserved frozen, Cary Blair, SAF, and formalin.
    • Sensitivity/Specificity Study (vs. DFA Microscopy): 234 retrospective samples (77 positive, 157 negative). Data provenance: United States (2 external laboratories). Sample matrices included formalin, SAF, unpreserved frozen, Cary Blair, and C&S.
    • Additional Retrospective Studies (vs. Non-fluorescent Microscopy):
      • Site 2: 47 retrospective samples (26 positive, 21 negative) vs. Modified Acid-Fast Stain.
      • Site 3: 281 retrospective SAF samples (60 positive, 221 negative) vs. Modified Kinyoun Stain.
    • Prospective Study (vs. DFA Microscopy): 378 samples (all negative). Data provenance: one external laboratory in the US. Sample matrices included unpreserved fresh, unpreserved frozen, formalin, SAF, C&S, and Cary Blair.
    • Overall Combined Test Set: 940 (prospective and retrospective) samples. Data provenance: Collected from Hospitals throughout the US and Canada.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of individual experts or their exact qualifications (e.g., number of years of experience, specific medical specialty) used to establish the ground truth for the test sets.

    Instead, it refers to "DFA microscopy," "Modified Kinyoun Stain light microscopy," and "Modified Acid-Fast Stain" as the reference methods (ground truth). It is implied that these microscopy methods are performed and interpreted by trained laboratory professionals, but specific expert details are not provided.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1 reader consensus) for resolving discrepancies in the reference methods (DFA microscopy, Modified Kinyoun Stain, Modified Acid-Fast Stain).

    In cases of discrepancy between the Uni-Gold™ device and the comparator/reference method within the studies, some retesting or re-evaluation against a "higher" ground truth was performed:

    • Site 3 (Comparator Device Study): For the 51 samples positive by Uni-Gold™ but negative by comparator, 30 were positive by Modified Kinyoun Stain light microscopy and 3 by DFA microscopy (agreeing with Uni-Gold™). The one sample negative by Uni-Gold™ and positive by comparator was negative by Modified Kinyoun Stain microscopy (agreeing with Uni-Gold™). This indicates that microscopy was used as an adjudicator for samples where the test device and the comparator device disagreed.
    • Site 3 (Non-fluorescent Microscopy Study): For the 23 negative samples (by Modified Kinyoun Stain) that tested positive on Uni-Gold™, three subsequently tested positive by DFA microscopy (agreeing with Uni-Gold™). This implies DFA microscopy acted as a higher-level adjudicator for certain discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This question is not applicable to the Uni-Gold™ Cryptosporidium device. The device is a rapid in vitro diagnostic immunoassay, not an AI-assisted diagnostic tool that involves human readers interpreting images. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed and is not relevant to this type of device.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is also not applicable. The Uni-Gold™ Cryptosporidium device is a standalone lateral flow immunoassay. Its performance is the "algorithm only" performance (the chemical reaction and visual readout) without human intervention in the detection process itself, beyond the initial sample preparation and result interpretation. There is no separate "algorithm without human-in-the-loop" performance as it is the test.

    7. The Type of Ground Truth Used

    The primary types of ground truth used for the test sets were:

    • DFA Microscopy: Direct Fluorescent Antibody microscopy, considered a gold standard for Cryptosporidium detection.
    • Modified Kinyoun Stain Light Microscopy: A non-fluorescent microscopy staining method.
    • Modified Acid-Fast Stain: Another non-fluorescent microscopy staining method.
    • Comparator Device Results: In some comparative studies, the results of the legally marketed predicate device (Remel Xpect® Cryptosporidium) were used for comparison. However, when discrepancies arose in these studies, microscopy methods (DFA or Modified Kinyoun) were often used to adjudicate.
    • Clinical Evaluation and Medical History: The intended use statement notes that "results should be considered in conjunction with the clinical evaluation and medical history," implying that clinical context is part of the broader diagnostic process, but the technical ground truth for device performance was microscopy.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of device development or any machine learning algorithm. This device is a biochemical immunoassay, not an AI or machine learning model that requires a discrete training set. The development of the assay's reagents and parameters would have involved internal validation and optimization, but not in the sense of a machine learning training set.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" for an AI/ML algorithm in this context, this question is not applicable. The ground truth for developing and optimizing the immunoassay would have been established through standard laboratory practices using known positive and negative controls, and possibly characterized clinical samples, during the assay development process, prior to the formal validation studies presented here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083896

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-07-02

    (185 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Destiny Max Coagulation Analyzer is a multipurpose system for in vitro coagulation studies consisting of one automated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

    Device Description

    The Destiny Max instrument performs coagulation testing (clotting, chromogenic and immuno-turbidimetric) using human samples. The system is comprised of an instrument (Destiny Max Analyzer) that performs the tasks necessary to general an assay result together with a Personal Computer (PC) that receives the user's secueeds and provides the results. The assays used with the Destiny Max are generally requences detection of clotting deficiencies or disorders and/or monitoring of anticoagalant therapy on various patient populations.

    AI/ML Overview

    The Trinity Biotech Destiny MAX Coagulation Analyzer is a multipurpose system for in vitro coagulation studies. Its acceptance criteria and performance are detailed through a method comparison study and precision studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided document. However, the performance is reported through method comparison against a predicate device (AMAX Destiny) and precision data, implying an expectation of strong correlation (for method comparison) and low variability (for precision).

    Method Comparison (Slope, Intercept, R/R² values):
    The device performance is reported across various assays (TriniCLOT PT HTF, TriniCLOT Excel S, TriniCLOT APTT S, TriniCLOT Thrombin Time, TriniCLOT Fibrinogen, TriniCLOT FVII, TriniCLOT FIX, TriniCHROM Antithrombin IIa, TriniLIA D-Dimer) in both optical and mechanical modes (where applicable). The reported R or R² values are consistently high, ranging from 0.88 to 1.00. Slopes are generally close to 1 and intercepts close to 0, indicating strong correlation and agreement with the predicate device.

    Linearity (R values):
    For linearity, the reported R values are all very high, ranging from 0.993 to 1.000.

    Precision (%CV values):
    Precision is assessed at different levels for each assay and mode, with "Within Run" and "Total" %CVs reported. The %CV values generally range from 0.6% to 24.2%, depending on the assay, level, and mode. For most clotting assays (PT, APTT, Thrombin Time), the %CVs are often below 5%. For Fibrinogen, FVII, FIX, Antithrombin, and D-Dimer, some %CVs are higher, particularly at lower concentrations or for “Total” precision.

    Since specific acceptance criteria were not explicitly stated (e.g., "R² must be > 0.95" or "%CV must be < 10%"), the determination of "meeting acceptance criteria" is inferred from the high reported correlation and acceptable precision for in vitro diagnostic devices. The submission implies these results demonstrate substantial equivalence and acceptable performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison:
      • The sample sizes (N) for the method comparison study vary by assay and site, ranging from 19 to 117 clinical samples per assay/site combination.
      • The data provenance is described as collected at 3 sites (implying clinical or laboratory settings). The document refers to "a variety of clinical samples," suggesting retrospective or prospectively collected samples from a clinical population. The country of origin is not explicitly stated but implied to be regions where the predicate device (AMAX Destiny) is used.
    • Linearity: The sample size for linearity studies is not explicitly stated in terms of number of patient samples, but rather the range of concentrations tested.
    • Precision: The sample sizes (N) for precision studies vary by assay, level, and site, with most entries showing N values between 45 and 60 runs. These typically refer to repeated measurements of control materials or pooled patient samples over several days/runs. Data provenance is from 3 sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (coagulation analyzer) does not typically involve human expert interpretation for "ground truth" establishment in the same way imaging or pathology devices do. The "ground truth" for the test set is established by the predicate device, the AMAX Destiny Coagulation Analyzer. Therefore, no human experts were involved in establishing the ground truth for this comparison study. The performance of the predicate device itself is implicitly considered the reference standard.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the "ground truth" or reference method for comparison is the predicate device's measurement. There is no human interpretation or classification that would require an adjudication method.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an in vitro diagnostic (IVD) device for quantitative biochemical measurements, not a diagnostic imaging or interpretive device that would typically involve human readers.

    6. Standalone Performance

    Yes, a standalone performance assessment was conducted in the form of linearity and precision studies. These studies evaluate the Destiny MAX's inherent measurement capabilities independent of direct comparison to the predicate device for each individual test.

    • Linearity data (Table 3) shows the analytical measurement range and correlation (R value) of the Destiny MAX for various assays.
    • Precision data (Tables 4-12) reports the within-run and total precision (%CV) of the Destiny MAX for various assays and at different concentration levels, demonstrating its reproducibility and variability.

    7. Type of Ground Truth Used

    • For the method comparison study, the "ground truth" was the measurements obtained from the predicate device, the AMAX Destiny Coagulation Analyzer. This establishes substantial equivalence by demonstrating comparable performance.
    • For linearity and precision studies, the ground truth is analytical standards, control materials, or known concentrations used to characterize the device's inherent performance.

    8. Sample Size for the Training Set

    The document does not provide details on a "training set" for the Destiny MAX in the context of machine learning or AI. This device is a quantitative analyzer, and its development would typically involve engineering, calibration, and internal validation steps, rather than traditional "training sets" as understood in AI/ML contexts. The focus of the 510(k) submission is on performance verification against a predicate and analytical characteristics.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a "training set" in the context of AI/ML for this device. The device's operational parameters and calibration are established through standard laboratory practices using reference materials and established assay principles.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033105

    Validate with FDA (Live)

    Date Cleared
    2004-07-13

    (287 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 1 Type Specific IgG kit is an Enzyme-linked Immunosorbent Assay (ELISA) intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 in human serum. In conjunction with the Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 2 Type Specific IgG kit, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. Due to the implications of positive results, it is recommended they be confirmed in a low prevalence population with Western blot. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment. The user is responsible for establishing assay performanace in these populations and with automated equipment

    Device Description

    The Captia™ HSV 1 IgG Type Specific kit is an Enzyme-Linked Immunosorbent Assav (ELISA) for the qualitative determination of IgG antibodies in human serum to Herpes simplex Type 1 antigen. The Captia™ HSV 1 IgG Type Specific assay may be used as an aid in the diagnoses of Herpes infection.

    For In Vitro Diagnostic Use Only.

    The Captia™ HSV 1 IgG Type Specific test is an Enzyme-Linked Immunosorbent assay to detect IgG antibodies to Herpes simplex 1 antigen. Purified recombinant HSV gG1 antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present, it will bind to the antibody attached to the antigen on the well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period, the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Captia™ HSV 1 IgG Type Specific ELISA Test Kit, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the performance characteristics presented in the study. The device is deemed acceptable if its agreement with the predicate device (Western Blot or other ELISA) falls within certain confidence intervals, demonstrated across various patient populations.

    Acceptance Criteria (Implied)Reported Device Performance95% Confidence Interval (CI)
    Expectant Mothers (vs. HSV 1 Western Blot)
    % Agreement Positive to WB: Acceptably high87.74% (136/155)82.6-92.9%
    % Agreement Negative to WB: Acceptably high98.18% (54/55)90.3-100.0%
    Sexually Active Adults (vs. HSV 1 Western Blot)
    % Agreement Positive to WB: Acceptably high87.93% (102/116)82.0-93.9%
    % Agreement Negative to WB: Acceptably high100.00% (80/80)95.5-100.0%
    Low Prevalence Population (vs. HSV 1 Western Blot)
    % Agreement Positive to WB: Acceptably high79.25% (42/53)65.9-89.2%
    % Agreement Negative to WB: Acceptably high97.71% (128/131)93.5-99.5%
    Culture Positives (vs. Culture)
    % Agreement Positive to Culture: Acceptably high69.81% (37/53)55.7-81.7%
    Culture Positives (vs. HSV 1 Western Blot)
    % Agreement Positive to WB: Acceptably high81.82% (36/44)67.3-91.8%
    Alternate HSV 1 Type Specific IgG ELISA
    Percent Positive Agreement: Acceptably high93.88% (92 / 98)87.2 - 97.7%
    Percent Negative Agreement: Acceptably high97.06% (99 / 102)91.6 - 99.4 %
    Percent Agreement: Acceptably high95.50% (191 / 200)91.6 - 97.9 %
    Type Specificity with HSV 2 Western Blot Positives
    Type-specificity relative to WB: Acceptably high96.4% (54/56)87.7-99.6%
    Type cross-reactivity relative to WB: Acceptably low3.6% (2/56)0.43-12.3%

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Expectant Mothers: n = 210 sera. Data provenance: Consented, coded, unselected, banked, and masked sera. Country of Origin: Not specified beyond "Pacific Northwest university" which references the Western Blot. Retrospective (banked sera).
      • Sexually Active Adults: n = 198 sera. Data provenance: Consented, unselected, and masked sera. Country of Origin: Not specified beyond "Pacific Northwest university". Retrospective (masked sera implies pre-collected).
      • Low Prevalence Population: n = 184 sera. Data provenance: Unselected, banked, and masked sera. Country of Origin: Not specified beyond "Pacific Northwest university". Retrospective (banked sera).
      • Culture Positives: n = 53 sera. Data provenance: Unselected, retrospective, and masked sera from patients at least six weeks but not more than one year post clinical presentation and culture. Country of Origin: Not specified beyond "Pacific Northwest university". Retrospective.
      • Alternate HSV 1 Type Specific IgG ELISA Comparison: n = 200 specimens. Data provenance: Prospective, unselected, sequentially submitted specimens. Country of Origin: Not specified beyond "Pacific Northwest University".
      • Type Specificity with HSV 2 Western Blot Positives: n = 56 sera. Data provenance: Samples from the "above described populations" (expectant mothers, sexually active adults, low prevalence persons, and HSV 2 culture positives) that were HSV 2 Western Blot positive and HSV 1 Western Blot negative. Country of Origin: Not specified beyond "Pacific Northwest University". Retrospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not explicitly state the number or qualifications of experts used to establish the ground truth. The ground truth (reference method) was established by an "HSV 1 Western Blot (WB) from a Pacific Northwest university" or "culture" for some cases. While Western Blot interpretation requires expertise, no details about the experts are provided.

    3. Adjudication method for the test set:
      The document does not mention an adjudication method for the test set. Results from the Trinity ELISA were compared directly against the Western Blot or culture reference method. For the expectant mothers and low prevalence population studies, "equivocal" results from the Trinity ELISA were excluded from the percentage agreement calculations but reported in the raw counts.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic ELISA kit, not an AI-assisted diagnostic tool that would involve human readers interpreting results with or without AI. The evaluation is solely on the performance of the assay itself compared to a reference method.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
      Yes, the studies presented are all standalone performance evaluations of the Captia™ HSV 1 IgG Type Specific ELISA kit. The performance metrics (e.g., % agreement positive, % agreement negative) are direct comparisons between the ELISA results and the reference method (Western Blot or culture) without human interpretation being factored in beyond the initial reading of the ELISA and the reference method itself.

    6. The type of ground truth used:
      The primary type of ground truth used was:

      • Expert Consensus / Established Reference Method: Western Blot (HSV 1 Western Blot, HSV 2 Western Blot). The Western Blot is considered a gold standard for type-specific HSV serology.
      • Pathology / Outcomes Data: Culture (for the "Culture Positives" study), indicating active infection.
    7. The sample size for the training set:
      The document does not provide information on a training set. The presented studies are performance evaluations of the final device, implying that any training and development would have occurred prior to these validation studies. Therefore, the sample sizes mentioned are for the test sets.

    8. How the ground truth for the training set was established:
      Since no training set details are provided, the method for establishing its ground truth is also not mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033106

    Validate with FDA (Live)

    Date Cleared
    2004-07-13

    (287 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 2 Type Specific IgG kit is an Enzyme-linked Immunosorbent Assay (ELISA) intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human serum. In conjunction with the Trinity Biotech Captia™ Herpes Simplex Virus (HSV) I Type Specific IgG kit, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. Due to the implications of positive results, it is recommended they be confirmed in a low prevalence population with Western blot. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment. The user is responsible for establishing assay performance in these populations and with automated equipment.

    Device Description

    The CaptiaTM HSV 2 IgG Type Specific kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgG antibodies in human serum to Herpes simplex Type 2 antigen. The Captia™ HSV 2 IgG Type Specific assay may be used as an aid in the diagnoses of Herpes infection. Purified recombinant HSV gG2 antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present, it will bind to the antibody attached to the antigen on the well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period, the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    The provided document details the performance characteristics of the Captia™ HSV 2 IgG Type Specific ELISA Test Kit. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve X% sensitivity and Y% specificity"). Instead, it presents the performance of the device against a predicate device (Western Blot) and an alternate ELISA. The reported performance is summarized below from different study populations. The predicate device used for comparison is the Western Blot (WB) from a Pacific Northwest university.

    Performance MetricStudy PopulationReported Device Performance (Captia™ HSV 2 IgG ELISA)95% Confidence Interval (CI)
    % Agreement Positive to WBExpectant Mothers100.00% (43/43)91.8-100.0%
    % Agreement Negative to WBExpectant Mothers91.52% (151/165)86.2-95.3%
    % Agreement Positive to WBSexually Active Adults96.72% (59/61)88.7-99.6%
    % Agreement Negative to WBSexually Active Adults90.30% (121/134)84.0-94.7%
    % Agreement Positive to WBLow Prevalence Population100.00% (4/4)39.8-100.0%
    % Agreement Negative to WBLow Prevalence Population91.06% (163/179)85.9-94.8%
    % Agreement Positive to CultureCulture Positives100.00% (56/56)93.6-100.0%
    % Agreement Positive to WBCulture Positives100.00% (55/55)93.5-100.0%
    % Positive Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera97.14% (68/70)90.1-99.7%
    % Negative Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera92.13% (117/127)86.0-96.2%
    % Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera92.50% (185/200)87.9-95.7%
    Type-specificity relative to WBHSV 1 WB Positive, HSV 2 WB Negative Sera90.75% (265/292)86.8-93.8%
    Type cross-reactivity relative to WBHSV 1 WB Positive, HSV 2 WB Negative Sera8.0% (23/292)5.06-11.6%

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes several distinct test sets:

    • Expectant Mothers:
      • Sample Size: n = 210
      • Data Provenance: Consented, coded, unselected, banked, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Sexually Active Adults:
      • Sample Size: n = 198
      • Data Provenance: Consented, unselected, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Low Prevalence Population:
      • Sample Size: n = 184
      • Data Provenance: Unselected, banked, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Culture Positives:
      • Sample Size: n = 56
      • Data Provenance: Unselected, retrospective, and masked sera from patients at least six weeks but not more than one year post clinical presentation and culture HSV 2 positive. Reference methods (culture and Western Blot) were from a Pacific Northwest university.
    • Prospectively Collected, Sequential Sera (vs. Alternate ELISA):
      • Sample Size: n = 200
      • Data Provenance: Prospective, unselected, sequentially submitted specimens. Collected by an outside investigator at a Pacific Northwest University.
    • Type Specificity with HSV 1 Western Blot Positives:
      • Sample Size: n = 292
      • Data Provenance: HSV 1 Western Blot positive and HSV 2 Western Blot negative sera from the above described populations (expectant mothers, sexually active adults, low prevalence persons, and HSV 1 culture positives). Collected by an outside investigator at a Pacific Northwest University. This is a retrospective analysis of previously collected samples.

    In general, the data seems to be from the United States (Pacific Northwest university) and is predominantly retrospective (banked, unselected, masked, retrospective sera), with one prospective study for comparison against an alternate ELISA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly mention "experts" in the context of establishing ground truth. The ground truth for comparative studies is established by reference methods, primarily:

    • Western Blot (WB): From a "Pacific Northwest university." No specific detail on the qualifications of the individuals performing or interpreting the Western Blot.
    • Culture (for Culture Positives set): Implies standard laboratory procedure for viral culture, not explicitly associated with "experts" in this context.

    Therefore, the specific number and qualifications of experts beyond the unspecified standard practices of a university lab for performing Western Blots and cultures are not detailed in the provided information.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1). The comparisons are directly between the Captia™ ELISA and the reference method results. There's no mention of a process where multiple readers or experts review discordant results to reach a consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (ELISA test kit) and does not involve human readers interpreting images or results, nor does it incorporate AI. Its performance is measured directly against laboratory reference standards.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented describe the standalone performance of the Captia™ HSV 2 IgG Type Specific ELISA Test Kit. This is an enzymatic immunoassay (ELISA) performed in a laboratory setting, and its efficacy is measured by comparing its outputs directly against established reference methods (Western Blot, culture, or another ELISA). There is no "human-in-the-loop" component in the performance evaluation described, beyond the technical execution of the tests themselves.

    7. The type of ground truth used

    The primary type of ground truth used across the various studies is:

    • Reference Standard (Western Blot): The Western Blot (WB) method from a Pacific Northwest university served as the gold standard for serological detection of HSV 2 IgG antibodies in most studies.
    • Culture: For the "Culture Positives" group, the ground truth for actual infection was established by HSV 2 culture positivity.
    • Alternate ELISA: In one study, another commercially available HSV 2 type-specific IgG ELISA was used as a comparative reference.

    8. The sample size for the training set

    The document provides performance data based on comparative studies with clinical samples. It does not mention a "training set" in the context of machine learning or algorithm development. This is a traditional IVD device, not an AI-driven algorithm. The performance evaluation focuses on the device's accuracy against established reference methods using specific test populations.

    9. How the ground truth for the training set was established

    As there is no explicit "training set" described for the development of a machine learning algorithm, this question is not applicable. The device's underlying mechanism is a biochemical ELISA, not an algorithm that learns from data. Its "ground truth" for development would relate to the antigen/antibody interactions it's designed to detect.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033059

    Validate with FDA (Live)

    Device Name
    HERPES GROUP IGG
    Date Cleared
    2003-11-26

    (58 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Herpes Group IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative determination of IgG antibodies in human serum to Herpes simplex virus. The Herpes Group IgG ELISA kit may be used to determine serologic status in females of child bearing age, and to evaluate paired sera for the presence of a seroconversion of IgG as an aid in the diagnosis of Herpes simplex virus infection. It is not intended for determining the type of Herpes simplex virus.

    Device Description

    The Herpes Group IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative determination of IgG antibodies in human serum to Herpes simplex virus. The Herpes Group IgG ELISA kit may be used to evaluate paired sera for the presence of seroconversions of IgG as an aid in the diagnosis of Herpes simplex virus infection.

    The Herpes Group IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Herpes simplex virus. Purified Herpes Group antigen is attached to a solid phase microtiter well. Diluted test sera are added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing and indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    The Trinity Biotech Herpes Group IgG ELISA Test Kit was evaluated for its agreement with predicate devices (Clark HSV 1 and HSV 2 ELISA assays) and for precision, cross-reactivity, and paired serum study performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state acceptance criteria in terms of predefined thresholds that the device must meet for approval. Instead, it presents performance characteristics (agreement percentages and precision) observed in comparison to predicate devices and other studies. The approval from the FDA indicates that the device's performance, as demonstrated, was deemed substantially equivalent to existing predicate devices. We can infer that the reported percentages of agreement and precision values were considered acceptable by the regulatory body for demonstrating substantial equivalence.

    Based on the cumulative data from the four comparison studies with the predicate device, the key performance metrics are:

    MetricAcceptance Criteria (Inferred from FDA Approval for Substantial Equivalence)Reported Device Performance (Cumulative across 4 studies)
    % Agreement PositivePerformance demonstrated substantial equivalence to predicate device98.9% (95% CI: 97.9% - 100%)
    % Agreement NegativePerformance demonstrated substantial equivalence to predicate device96.7% (95% CI: 94.2% - 99.1%)
    % Total AgreementPerformance demonstrated substantial equivalence to predicate device98.1% (95% CI: 97.0% - 99.2%)
    Precision (Intersite CV)< 15% (for appropriate technique), as stated for user expectationFrom 9.08% to 15.1% for positive sera (Sera #1-5)
    121% and 150% for negative sera (Sera #6-7)
    Paired Serum Study (Seroconversion)100% agreement expected for evaluable pairs100% agreement (for 11 evaluable pairs)
    Cross-ReactivityNo cross-reactivity with EBV, CMV, VZVNo cross-reactivity observed with tested sera
    CDC Panel Total AgreementPerformance demonstrated substantial equivalence to predicate device96.9% (excluding 2 equivocals)
    CDC Panel Positive AgreementPerformance demonstrated substantial equivalence to predicate device95.7%
    CDC Panel Negative AgreementPerformance demonstrated substantial equivalence to predicate device100%

    2. Sample Sizes Used for the Test Set and Data Provenance

    The primary test set for comparison with the predicate device involved sera tested across four different sites.

    • Study 1 (Maryland R&D lab): 187 frozen sera (normals aged 12-83, various gender and geographical areas). Retrospective.
    • Study 2 (New York R&D lab): 152 frozen sera (normals aged 17-59, various gender and geographical areas). Retrospective.
    • Study 3 (Pennsylvania clinical lab): 176 prospective samples sent for Herpes antibody testing. Prospective.
    • Study 4 (Wisconsin clinical lab): 88 frozen random normal samples. Retrospective.
    • Total Sample Size for Primary Comparison: 603 sera (cumulative across all four studies).
    • CDC Panel: A masked, characterized serum panel (details on total number not explicitly stated, but 72% positive and 28% negative samples, with 2 equivocals excluded). The origin is the Centers for Disease Control (CDC), implying a well-characterized and respected source.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This study does not involve human readers interpreting results or establishing ground truth based on expert consensus. The "ground truth" for the comparative studies is the result obtained from the predicate device, the Clark HSV 1 and HSV 2 ELISA assays. For the CDC panel, the "ground truth" is the CDC's characterization of the panel samples. No human experts in the traditional sense (e.g., radiologist) are described as establishing ground truth in this context; instead, the comparator assays serve this role.

    4. Adjudication Method for the Test Set

    Not applicable. The study compares the new device's results directly against a predicate device. There is no mention of human adjudication for discrepancies between the new device and the predicate device or a clinical outcome.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is typically done for diagnostic imaging tests or similar assessments where human readers interpret results, and the AI's impact on their performance is evaluated. The device in question is an ELISA kit, which produces quantitative results, not interpretations by human readers in the style of imaging.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies presented are standalone performance evaluations of the Trinity Biotech Herpes Group IgG ELISA kit. The results are generated solely by the device (the assay) and compared against a predicate device or a characterized panel. There is no human "in the loop" impacting the diagnostic output of the device itself.

    7. The Type of Ground Truth Used

    • For the primary comparison: The predicate device's results (Clark HSV 1 and HSV 2 ELISA assays) served as the "ground truth" or reference standard against which the new device's qualitative agreement (% agreement positive, negative, and total) was measured.
    • For the paired serum study: The complement fixation (CF) method was used to determine seroconversion, serving as the ground truth.
    • For the cross-reactivity study: The presence of IgG antibodies to specific viruses (EBV, CMV, VZV) was likely determined by established ELISA or other serological methods for those viruses, acting as the ground truth for non-cross-reactivity.
    • For the CDC panel: The CDC's characterization of the masked serum panel served as the ground truth.

    It is important to note that the document explicitly states: "There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease." This clarifies that the ground truth is primarily based on concordance with another assay, not directly on clinical pathology or patient outcomes.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of an algorithm or AI development. ELISA kits are laboratory tests with defined chemical reactions and optical detection methods; they are not typically "trained" in the way AI models are. The development and optimization of the test kit itself would involve internal studies and reagent validation, but these are distinct from a typical AI training/validation set paradigm.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" for an algorithm described, this point is not applicable. The development of the ELISA test kit relies on established biochemical principles and validation procedures to ensure the reagents and protocol accurately detect the target antibodies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3