K Number
K163095

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-02-01

(89 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lyme B. burgdorferi (IgG) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgG antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2- EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

Device Description

The kit is an immunoblot method to detect IgG antibodies against B. burqdorferi antigens. The test kit contains: Nitrocellulose Test Strips with purified B. burgdorferi antigens (10) and quality control lines (3) present in specific positions, Sample Diluent, Positive Control, Negative Control, Conjugate (Antihuman IgG-HRP Conjugate), TMB Substrate, and PBS Wash Buffer concentrate. To perform the test, serum or plasma is added to individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diuted, ready-to-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 5 (or more) of the 10 test lines are considered positive for specific IgG antibody to B. burgdorferi.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Lyme B. burgdorferi (IgG) MarStripe Test, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined acceptance criteria for performance metrics like sensitivity and specificity percentage targets. However, the reported performance against the predicate device and consensus can be considered the de-facto acceptance criteria for equivalence.

Performance MetricAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance (Lyme B. burgdorferi (IgG) MarStripe Test)
Reproducibility100% positive or negative agreement for all specimens across sites and operators100% (for overall positive/negative agreement across 8 specimens, 3 sites, 2 operators, 432 readouts)
Analytical SpecificityNot explicitly stated, but high specificity is expected, ideally comparable to predicate.99.5% (95% CI: 97.1% - 100%) against 219 healthy blood donors
Cross-reactivityLow positive incidence, and positive specimens should be confirmed by predicate device.4.4% positive incidence across 136 potentially cross-reactive specimens; all positives confirmed by predicate IgG Western blot.
Interference100% qualitative agreement with and without interfering agents.100% qualitative agreement for all specimens (5 sera spiked with hemoglobin, bilirubin, RF, triglycerides, cholesterol).
Serum vs. Plasma Equivalence100% qualitative agreement between serum and plasma samples.100% qualitative agreement for all 20 pairs of serum/plasma samples.
Positive Agreement (vs. Predicate)High agreement (e.g., >90%) deemed substantially equivalent to predicate.92.9% (95% CI: 88.6% - 95.7%)
Negative Agreement (vs. Predicate)High agreement (e.g., >95%) deemed substantially equivalent to predicate.97.7% (95% CI: 95.9% - 98.7%)
Sensitivity (Clinical Specimens, vs. Predicate)Similar sensitivity to the predicate device.22.3% (21/94) (95% CI: 14.7%-32.3%) - Predicate: 20.2% (19/94)
Sensitivity (CDC Panels, vs. Predicate)Similar specificity to the predicate device.16.7% (7/42) - Predicate: 16.7% (7/42)

2. Sample Size Used for the Test Set and Data Provenance

  • Reproducibility Test: 8 specimens (2 low negative, 2 high negative, 2 low positive, 2 moderate positive). Each specimen tested in 4 replicates, 2 runs per day, for a total of 72 tests per specimen at three laboratory sites.
  • Analytical Specificity: 219 sera from healthy blood donors.
  • Cross-reactivity: 136 potentially cross-reactive specimens from individuals with various other or infectious conditions.
  • Interference: 5 sera (2 Lyme IgG negative, 3 Lyme IgG positive).
  • Serum vs. Plasma Comparison: 10 pairs of native serum/plasma samples, plus 10 pooled serum/plasma samples (total 20 pairs).
  • Method Comparison (Clinical Tests): 753 prospectively collected consecutive specimens, all initially positive on an FDA-cleared first-step EIA.
  • Sensitivity (Clinical Specimens): 94 well-characterized Lyme disease clinical specimens (from the 753 specimens in Method Comparison).
  • CDC Panels: 42 specimens from CDC (Lyme Disease Validation Panel n=10, Lyme Disease Basic Research Panel n=32).

Data Provenance:

  • Healthy Blood Donors (Analytical Specificity): Representing endemic geographic regions of the United States.
  • Clinical Specimens (Method Comparison & Sensitivity): Prospectively collected consecutive specimens tested at three sites. No further country of origin or specific demographic details are provided.
  • CDC Panels: From the Center for Disease Control and Prevention.

Retrospective or Prospective:

  • Method Comparison (Clinical Specimens): Explicitly stated as "prospectively collected consecutive specimens." The other studies do not explicitly state but generally, these types of validation studies are prospective or involve carefully selected retrospective samples to meet specific criteria.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

There is no mention of external human experts establishing the ground truth for the test set in the conventional sense of, for example, radiologists interpreting images.

  • Reproducibility: Results were read by 2 human operators at each of the 3 sites. No specific qualifications (e.g., years of experience) are provided for these operators, who are presumably laboratory personnel following a protocol.
  • Method Comparison: The ground truth for comparing the device primarily relies on the predicate device (MarDx B. burgdorferi IgG MarBlot Strip Test System).
  • Discrepant Analysis: For the 29 discrepant specimens, two additional commercially available B. burgdorferi IgG Western blot methods were used as comparators to establish a "consensus result." This implies laboratory methods rather than expert human interpretation.
  • CDC Panels: Ground truth is established by the CDC's characterization of the panels.

4. Adjudication Method for the Test Set

  • Reproducibility: No formal adjudication method like "2+1" is mentioned for individual band interpretation. The "Final positive or negative agreement was 100% for all specimens" suggests direct agreement between operators or that any differences were resolved internally to reach a unified result for each specimen.
  • Discrepant Analysis (Method Comparison): A "2/3 comparator" consensus was used where the Lyme B. burgdorferi (IgG) MarStripe Test result was compared to results from two additional commercially available B. burgdorferi IgG Western blot methods.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done in the context of human readers improving with AI vs. without AI assistance. This device is a diagnostic test kit (immunoblot assay) and not an AI-powered image analysis system that would typically involve human-in-the-loop performance evaluation. The "human operators" mentioned in reproducibility are interpreting the results of the kit, not an AI output.

6. Standalone Performance Study

  • Yes, a standalone performance study was done. The entire set of non-clinical and clinical tests (reproducibility batch, analytical specificity, cross-reactivity, interference, serum vs. plasma testing, and the initial method comparison against the predicate) represents the standalone performance of the device/kit itself. The "algorithm" here refers to the immunodetection process and interpretation rules inherent to the diagnostic kit, not a separate AI algorithm.

7. Type of Ground Truth Used

  • Reproducibility: Based on prior FDA-cleared B. burgdorferi ELISA results (categorizing samples as low/high negative, low/moderate positive).
  • Analytical Specificity: Healthy blood donors (presumed negative for Lyme).
  • Cross-reactivity: Individuals with other infectious or autoimmune conditions (used to challenge specificity).
  • Interference: Artificially spiked samples.
  • Serum vs. Plasma Comparison: Comparison to results from an FDA-cleared Lyme EIA assay and subsequent Western blot IgG positivity/negativity.
  • Method Comparison (Clinical Tests): The primary ground truth was the predicate device's result. For discrepant samples, a consensus of two additional commercial IgG Western blot methods served as the ground truth.
  • Sensitivity (Clinical Specimens): "Well characterized Lyme disease clinical specimens" from the clinical study, with characterization across disease stages (early, early disseminated, late).
  • CDC Panels: Characterized reference panels provided by the Center for Disease Control and Prevention.

8. Sample Size for the Training Set

No training set is explicitly mentioned or applicable in the context of this device. This is a diagnostic test kit that functions based on established biochemical reactions and interpretation algorithms, not a machine learning or AI model that requires a separate training phase. The described studies are validation studies for the finished product.

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned for an AI/ML model, this question is not applicable. The device's "ground truth" (i.e., its intended performance characteristics) is established through the development and analytical/clinical validation studies presented.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures connected by a flowing line. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

February 1, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Immco Diagnostics, Inc. Kevin Lawson Regulatory Officer 60 Pineview Dr. Buffalo, New York 14031

Re: K163095

Trade/Device Name: Lyme B. Burgdorferi (IgG) MarStripe Test Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LSR Dated: September 10, 2016 Received: November 4, 2016

Dear Kevin Lawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Steven R. Gitterman -S

Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationIndications for UseForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
510(k) Number (if known)K163095
Device NameLyme B. burgdorferi (IgG) MarStripe Test
Indications for Use (Describe)Lyme B. burgdorferi (IgG) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human IgG antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K2- EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi .

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office

of Chief Information Officer Paperwork

Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14) Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green dots on the left, followed by the company name "immco" in blue, and "DIAGNOSTICS" in green below it. Underneath the company name, it says "A Trinity Biotech Company" in a smaller font.

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

1. Submitter Name:Immco Diagnostics, Inc. a Trinity Biotech Company
Address:60 Pineview Dr., Buffalo, NY 14228
Phone Number:716-691-0091 ext. 110
Contact Person:Kevin Lawson
Summary Prepared:1-4-2017
  • Lyme B. burgdorferi (IgG) MarStripe Test 2. Device Name: Common Name: Lyme B. burgdorferi (IgG) Immunoblot Product Code: LSR
    1. Substantially equivalent to: MarDx B. burgdorferi IgG MarBlot Strip Test System (K950829)
    1. Description of device: The kit is an immunoblot method to detect IgG antibodies against B. burqdorferi antigens. The test kit contains:
    • . Nitrocellulose Test Strips with purified B. burgdorferi antigens (10) and quality control lines (3) present in specific positions
    • . Sample Diluent. Provided for specimen dilutions. Contains BSA and PBS
    • Positive Control derived from human serum positive for Lyme disease. Contains <0.1% sodium azide
    • Negative Control derived from human serum negative for Lyme disease. Contains <0.1% sodium azide
    • . Conjugate. Antihuman IgG-HRP Conjugate binds reactive antibodies to the Substrate
    • TMB Substrate. Provides colorimetric reaction for visual read of bound antibodies
    • PBS Wash Buffer concentrate. Removes reagents and unbound antibodies after incubation steps. Must be reconstituted to 1L with distilled or deionized water.

To perform the test, serum or plasma is individual B. burgdorferi Test Strips. In positive sera antibodies specifically bind to one or more of the test lines on the strips are washed according to the protocol, and then the pre-diuted, readyto-use Conjugate is added to the test strips. After incubation and wash steps, the ready-to-use Substrate is added to the strips. During a 10 minute (±4 min) incubation, conjugate and substrate binding produces visible blue/purple lines for Serum Addition Control (SAC), Conjugate Addition Control (CAC) and Cut-Off Control lines. If the sample is positive for any of the antigen coated test lines, it will show a reaction more intense than the Cut-Off line. Reactions are read visually and reported as positive or equivocal (comparable to Cut-Off line). Strips which have 5 (or more) of the 10 test lines are considered positive for specific IgG antibody to B. burgdorferi.

  • Intended Use: Lyme B. burqdorferi (IgG) MarStripe Test is an immunoblot assay for the in vitro qualitative detection of human 5. lgG antibody to individual proteins of Borrelia burgdorferi in human serum or plasma (K- EDTA) in samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.
    1. Similarities and Differences: The Lyme B. burgdorferi (IgG) MarStripe Test was compared to a commercially marketed kit by Trinity Biotech the B. burgdorferi (lgG) MarBlot Strip Test System (K950829). Both kits have the same intended use and use the same methodology except that the MarStripe Test has been validated in plasma (K- EDTA) as well as serum. Both immunoblot kits detect antibodies to B. burgdorferi antigens using the standard immunoblot IgG algorithm. Both assays are qualitative. The MarStripe uses a Horseradish Peroxidase Conjugate and Tetramethylbenzidine (TMB) Substrate/Chromogen in comparison to The MarBlot Alkaline Phosphatase Conjugate and BCIP/NBT Substrate/Chromogen. The MarStripe uses a Cutoff Control line incorporated in the strip in comparison to the 41kD band of Weakly Reactive Control.

7. Non-clinical Tests:

60 Pineview Drive Buffalo, NY 14228-2120 USA Toll free (800) 537-8378  •  tel. (716) 691-0091  •  fax (716) 691-0466

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Immco Diagnostics, a Trinity Biotech Company. The logo features a cluster of blue and green circles on the left, followed by the word "immco" in blue, bold letters. Below "immco" is the word "DIAGNOSTICS" in smaller, green letters, and below that is the phrase "A Trinity Biotech Company" in a smaller, lighter font.

Reproducibility

Eight specimens were tested by Lyme B. burgdorferi (lgG) MarStripe Test in 4 replicates, two runs per days for a total of 72 tests for each specimen at three laboratory sites. Results were read by 2 human operators at each site, equaling a total of 432 read-outs. Samples were selected based on FDA cleared B. burgdorferi ELISA results, including 2 low negative samples, 2 high negative samples, 2 low positive samples. Final positive samples. Final positive or negative agreement was 100% for all specimens.

Samplep93p66p58p45p41p39p30p28p23p18
1n=432
Low NegNeg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.
Band Type0000000000
Positives432432432432432432432432432432
Negatives0%0%0%0%0%0%0%0%0%0%
2% Positive
Low NegNeg.Neg.Neg.Neg.Pos.Neg.Neg.Neg.Neg.Neg.
Band Type000043200000
Positives4324324324320432432432432432
Negatives0%0%0%0%100%0%0%0%0%0%
3% Positive
High NegNeg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.Neg.
Band Type0000000000
Positives432432432432432432432432432432
Negatives0%0%0%0%0%0%0%0%0%0%
4% Positive
High NegNeg.Neg.Neg.Neg.Pos.Neg.Neg.Neg.Neg.Neg.
Band Type000043200000
Positives4324324324320432432432432432
Negatives0%0%0%0%100%0%0%0%0%0%
5% Positive
Low PosNeg.Neg.Neg.Neg.Pos.Neg.Neg.Neg.Pos.Neg.
Band Type00004320004320
Positives43243243243204324324320432
Negatives0%0%0%0%100%0%0%0%100%0%
6% Positive
Low PosNeg.Neg.Neg.Neg.Pos.Neg.Neg.Neg.Pos.Neg.
Band Type00004320004320
Positives43243243243204324324320432
Negatives0%0%0%0%100%0%0%0%100%0%
7% Positive
Moderate PosWpos.Pos.Pos.Neg.Pos.Pos.Cut.Pos.Pos.Pos.
Band Type4124324320432432412432432432
Positives20004320020000
Negatives95%100%100%0%100%100%95%100%100%100%
8% Positive
Moderate PosPos.Neg.Pos.Neg.Pos.Pos.Neg.Neg.Pos.Neg.
Band Type43204320432432004320
Positives04320432004324320432
Negatives100%0%100%0%100%100%0%0%100%0%
% Positive

Pos = positive band. Neg = negative band. Wpos = weak positive band. Cut = equivocal band.

Analytical Specificity: 219 sera from healthy blood donors representing endemic geographic regions of the United States were tested. Analytical specificity was determined to be 99.5% (95% Cl: 97.1% - 100%).

Normal Individuals
Lyme IgGMarStripeTestPositive1
Negative218
Total219

60 Pineview Drive ● ● Buffalo, NY 14228-2120 ● ● USA Toll free (800) 537-8378  •  tel. (716) 691-0091  •  fax (716) 691-0466

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a cluster of blue and green dots on the left, followed by the company name in blue and green text. Below the company name, it says "A Trinity Biotech Company" in smaller, gray text.

Cross-reactivity: A total of 136 potentially cross-reactive specimens from individuals with other or infectious conditions were tested on this assay. Positive incidence was 4.4% and all positives specimens by the Lyme B. burgdorferi (IgG) MarStripe Test were confirmed positive when tested by the predicate IgG Western blot device.

Populationnn Positive% Positive
Ehrlichia chafeensis1000
Babesia microti10440
Leptospira interrogans1000
Helicobacter pylori10220
Syphilis1000
Influenza1000
Rocky Mountain Spotted Fever1000
Parvovirus B19900
Systemic lupus erythematosus1500
Cytomegalovirus1000
Rheumatoid arthritis1600
Celiac1600
Total13664.4

Interference: Two Lyme IgG negative and three Lyne IgG positive sera were spiked with hemoglobin (2g/L), unconjugated bilirubin (342 µmol/L), RF (100 IU/ml), triglycerides (3.7 mmol/L) and total cholesterol (13 mmol/L) and tested using this assay. Samples were tested with and without interfering agents. Qualitative agreement was 100% for all specimens.

Serum vs. Plasma Matrix Comparison Studies: To establish equivalence of serum vs. plasma (Kg-EDTA) matrix, 10 pairs of sera/plasma (samples A-J below) were sourced from specimens tested on an FDA cleared Lyme EIA assay. Two of these were Western blot lgG positive, 8 were negative. These specimens were selectively pooled to create another set of 10 samples (samples 1-10 below), including 3 Western blot IgG positives. These samples were assyed on the Lyme B. burgdorferi (IgG) MarStripe Test. Qualitative agreement for all pairs was 100%.

Image /page/5/Picture/6 description: The image displays the text "www.immco.com" in a blue, sans-serif font. The text is underlined with a solid blue line. The background of the image is plain white, providing a clean and simple presentation of the website address.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a cluster of blue and green circles to the left of the word "immco" in blue, with a small blue circle above the "o". Below "immco" is the word "DIAGNOSTICS" in green. Underneath that, it says "A Trinity Biotech Company" in a smaller font.

Sample ID1Type2p93p66p58p45p41p39p30p28p23p18ResultBand % PosAgrmtBand % NegAgrmt
ASerum0000100010NEG100
APlasma0000100010NEG100
BSerum0000000000NEG100
BPlasma0000000000NEG100
CSerum1010111011POS100
CPlasma1010111011POS100
DSerum0010111011POS100
DPlasma0010111011POS100
ESerum0000000001NEG100
EPlasma0000000001NEG100
FSerum0000000000NEG100100
FPlasma0000000000NEG
GSerum0000000000NEG100100
GPlasma0000000000NEG
HSerum0000000000NEG100100
HPlasma0000000000NEG
ISerum0000000000NEG100100
IPlasma0000000000NEG
JSerum0000000000NEG100100
JPlasma0000000000NEG
Pool 1 (A+B)Serum0000000010NEG100100
Pool 1 (A+B)Plasma0000000010NEG
Pool 2 (B+C)Serum1010110011POS100100
Pool 2 (B+C)Plasma1010110011POS
Pool 3 (C+D)Serum1010111011POS100100
Pool 3 (C+D)Plasma1010111011POS
Pool 4 (D+E)Serum0010110011POS100100
Pool 4 (D+E)Plasma0010110011POS
Pool 5 (A+E)Serum0000100010NEG100100
Pool 5 (A+E)Plasma0000100010NEG
Pool 6 (F+G)Serum0000000000NEG100
Pool 6 (F+G)Plasma0000000000NEG100
Pool 7 (G+H)Serum0000000000NEG100100
Pool 7 (G+H)Plasma0000000000NEG
Pool 8 (H+I)Serum0000000000NEG100100
Pool 8 (H+I)Plasma0000000000NEG
Pool 9 (I+J)Serum0000000000NEG100100
Pool 9 (I+J)Plasma0000000000NEG
Pool 10 (F+J)Serum0000000000NEG100100
Pool 10 (F+J)Plasma0000000000NEG
  1. Pool 1 to 10 contain a 1:1 mixture of the native samples as indicated. 2. 0 = negative band result. 1 = positive band result.

8. Clinical Tests:

Method Comparison: Three sites tested utilized prospectively collected consecutive specimens testing positive on a FDA cleared first-step ElA in the study. 753 specimens were tested on the predicate MarDx B. burgdorferi lgG MarBlot Test device as well as the Lyme B. burgdorferi (IgG) MarStripe Test

Predicate IgG WB
Lyme IgGMarStripe TestPositiveNegativeTotal
Positive22112233
Negative17503520
Total238515753

Positive Agreement: 92.9% (95% CI: 88.6% - 95.7%) Negative Agreement: 97.7% (95% CI: 95.9% - 98.7%)

60 Pineview Drive ● Buffalo, NY 14228-2120 ● Toll free (800) 537-8378  •  tel. (716) 691-0091  •  fax (716) 691-0466

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Immco Diagnostics. The logo features a series of blue and green dots on the left side, followed by the word "immco" in blue, with the word "DIAGNOSTICS" in green underneath. Below that, it says "A Trinity Biotech Company" in a smaller font.

The 29 discrepant specimens were tested on two additional commercially available B. burgdorferi IgG Western blot methods. Results of the Lyme B. burgdorferi (IgG) MarStripe Test were compared to the consensus (2/3 comparator) results. The 12 positives by MarStripe remained positive by consensus testing and the 17 negatives by MarStripe were found to be positives.

Consensus Result
PositiveNegative
Lyme IgGMarStripe TestPositive120
Negative170

Sensitivity: 94 well characterized Lyme disease clinical specimens from the above study were B. burgdorferi (IgG) MarStripe Test. Specimens included samples from early, early disseminated, and late phase. The sensitivity obtained was compared with that of the predicate device.

IntervalnLyme IgG MarStripeTestPredicate IgG WB
positive%positive%
Early Lyme (stage 1)22418.2313.6
Early disseminated(stage 2)441227.31125.0
Late Lyme (stage 3)28517.9517.9
Overall942119.51920.2

Sensitivity Comparison:

Lyme B. burgdorferi (IgG) MarStripe Test: Predicate device: Difference in proportion:

22.3% (21/94) (95% Cl: 14.7%-32.3%) 20.2% (19/94) (95% Cl: 12.9%-30.0%) 2.1%

CDC Panels: Reference panels from the Center for Disease Control and Prevention (Lyme Disease Validation Panel n=10, Lyme Disease Basic Research Panel n=32) were tested on the Lyme B. burgdorferi (lgG) MarStripe Test and the predicate device.

IntervalnLyme IgGMarStripe TestPredicate IgG WB
positive%positive%
Controls250000
Early Lyme (stage 1)10110110
Early disseminated (stage 2)331003100
Late Lyme (stage 3)441004100
Overall42716.7716.7

Note: The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.

    1. Conclusion: From the performance data and kit comparison above, it is our conclusion that the Lyme B. burgdorferi (IgG) MarStripe Test is substantially equivalent to the B. burgdorferi (IgG) MarBlot Strip Test System (K950829) commercially marketed by Trinity Biotech
      Klaus

Kevin J. Lawson VP Regulatory Affairs

60 Pineview Drive Buffalo, NY 14228-2120 537-8378 · fax (716) 691-0466

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).