Search Results
Found 15 results
510(k) Data Aggregation
(90 days)
VA 24060-6358
Re: K171078
Trade/Device Name: Tri-combo Parasite Screen Regulation Number: 21 CFR 866.3220
Cryptosporidium spp., and E. histolytica detection test
1.4 Classification and Regulation
Class II
21 CFR 866.3220
The TECHLAB® TRI-COMBO PARASITE SCREEN test is an enzyme immunoassay for the simultaneous qualitative detection of Giardia spp., Cryptosporidium spp. and/or E. histolytica antigen in human fecal specimens. The test is indicated as an aid in the diagnosis of gastrointestinal infection when giardiasis, cryptosporidiosis and amebiasis is suspected. The test does not differentiate between the three parasites and follow-up testing is required for all positive results to confirm the specific diagnosis.
The TRI-COMBO PARASITE SCREEN test is an enzyme immunoassay for the simultaneous qualitative detection of Giardia spp., Cryptosporidium spp. and/or E. histolytica antigen in human fecal specimens. The test uses monoclonal and polyclonal antibodies to cell-surface antigens of Giardia, Cryptosporidium and E. histolytica. The microassay plate in the kit contains immobilized monoclonal antibodies against the antigens, and the Conjugate consists of polyclonal antibodies against the antigens. In the assay, an aliquot of a diluted fecal specimen is transferred to a microassay well. The immobilized monoclonal antibodies bind the Giardia, Cryptosporidium and/or E. histolytica antigens if they are present. Upon addition, Conjugate then binds to the antigen/ antibody complex. Any unbound materials are removed during the washing steps. Following the addition of Substrate, a color is detected due to the enzyme-antibody-antigen complexes that formed in the presence of antigens and conjugate.
Here's an analysis of the provided text regarding the TECHLAB® TRI-COMBO PARASITE SCREEN test, structured according to your requested points:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve sensitivity of X% and specificity of Y%"). Instead, it presents performance data which implicitly serves as the basis for the FDA's substantial equivalence decision. Therefore, I will present the reported performance data from the prospective clinical study as the proxy for "acceptance criteria met." For the retrospective study, while providing higher values, the prospective study is generally weighted more heavily for regulatory approval as it reflects real-world conditions more closely. The document also provides "95% Confidence Limits" for sensitivity and specificity.
Metric (Prospective Study vs. Microscopy) | Reported Device Performance | 95% Confidence Limits |
---|---|---|
Sensitivity | 92.9% | 68.5% - 98.7% |
Specificity | 98.1% | 96.9% - 98.9% |
Positive Percent Agreement (vs. Microscopy and Molecular) | 62.1% | 44.0% - 77.3% |
Negative Percent Agreement (vs. Microscopy and Molecular) | 98.8% | 97.7% - 99.4% |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Prospective Study:
- Sample Size: 754 (740 negative, 14 microscopy positive - 13 Giardia, 1 E. histolytica)
- Data Provenance: Conducted at 3 independent clinical sites. Specific country of origin is not explicitly stated, but the submission is to the U.S. FDA by a U.S. company, suggesting U.S. clinical sites. The data is prospective.
- Retrospective Study:
- Sample Size: 96 archived specimens (previously collected and frozen)
- Data Provenance: From one clinical site in an E. histolytica endemic area (not specified by country). The data is retrospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
The document primarily refers to "microscopy" and "molecular testing."
- Microscopy: It states that "accuracy of O&P results depends upon the skill of the technician." This implies trained laboratory technicians or medical technologists are performing microscopy. However, the exact number of experts or their specific qualifications (e.g., board certification, years of experience) are not provided in the document.
- Molecular Testing: The document mentions "molecular testing of a commercial FDA cleared device and PCR with sequencing." It does not specify the number or qualifications of experts involved in interpreting or performing these molecular tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
The document does not describe a formal adjudication method (e.g., 2+1, 3+1) involving multiple human readers for discrepancies. Instead, discrepancies between the device and microscopy were resolved via further testing:
- For the prospective study, the 14 TRI-COMBO PARASITE SCREEN positives that were microscopy negative were "confirmed to be positive for Giardia with an alternate FDA cleared antigen test or by PCR with sequencing." This constitutes a form of resolution by a more definitive method rather than human adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not an AI-assisted device for human readers. The device is an in vitro diagnostic (IVD) immunoassay for detecting antigens in fecal specimens. Therefore, an MRMC comparative effectiveness study involving human readers improving with AI assistance is not applicable to this device. The comparison is between the immunoassay and traditional methods (microscopy and molecular testing).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the device is a standalone diagnostic test (an immunoassay). Its performance is evaluated directly against comparator methods (microscopy, PCR, other FDA-cleared antigen tests) without human interpretation of the device's generated data being a variable. The test results are "spectrophotometric" or "visual" interpretations of a color change, but the core performance data reflects the device's ability to detect the analytes by itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the test sets was established using a combination of:
- Microscopy: Primarily light microscopy for identification of Giardia, Cryptosporidium, and E. histolytica.
- Molecular Testing:
- For the prospective study, "composite results from multiple tests that consisted of light microscopy, molecular testing of a commercial FDA cleared device and PCR with sequencing for the identification of Giardia spp., Cryptosporidium spp., in addition to identification and subspeciation of E. histolytica." This "composite" is considered the ground truth.
- For discrepancies, "alternate FDA cleared antigen test or by PCR with sequencing" was used.
- Archived Specimen Characterization: For the retrospective study, specimens were characterized as "microscopy and PCR positive."
8. The sample size for the training set
The document describes performance studies for the device but does not mention a training set because this is an in vitro diagnostic (IVD) immunoassay, not a machine learning or AI-based device that requires a training set. The design of the assay (antibodies, reagents) is based on scientific principles and previous research, not on learning from a specific dataset.
9. How the ground truth for the training set was established
As there is no training set for this type of device, this question is not applicable.
Ask a specific question about this device
(90 days)
Histolytica Ouik Chek Regulation Number: 21 CFR 866.3220 Regulation Name: Entamoeba histolytica serological
Classification Name
E. histolytica detection test
1.4 Classification and Regulation
21 CFR 866.3220
The TECHLAB® E. HISTOLYTICA QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from Entamoeba histolytica in a single use cassette. It is intended for use with human fecal specimens from patients with diarrhea or dysentery as an aid in the diagnosis of E. histolytica gastrointestinal infection. Test results should be considered in conjunction with patient history. FOR IN VITRO DIAGNOSTIC USE
The E. HISTOLYTICA QUIK CHEK™ test uses antibodies to adhesin. The device contains a Reaction Window with two vertical lines of immobilized antibodies. The test line ("T") contains monoclonal antibodies specific for E. histolytica adhesin. The control line ("C") contains antibodies to horseradish peroxidase (HRP). The Conjugate consists of polyclonal antibodies to E. histolytica adhesin coupled to horseradish peroxidase. To perform the test, the sample is added to a tube containing a mixture of Diluent and Conjugate. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, E. histolytica adhesin in the samples bind to the antibodyperoxidase conjugate. The antigen-antibody complexes migrate through a filter pad to a membrane where they are captured by the immobilized antibodies in the Reaction Window is subsequently washed with Wash Buffer, and the test is developed with the addition of Substrate. After a 10 minute incubation period, the "T" reaction is examined visually for the appearance of a vertical blue line on the "T" side of the Reaction Window. A blue line indicates a positive test. A positive "C" reaction, indicated by a vertical blue line on the "C" side of the Reaction Window, monitors/confirms that the sample and reagents were added correctly, the reagents were active at the time of performing the assay, and that the sample migrated properly through the Membrane Device. It also confirms the reactivity of the other reagents associated with the assay.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
E. HISTOLYTICA QUIK CHEK™ Performance Study Summary
The E. HISTOLYTICA QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from Entamoeba histolytica in human fecal specimens.
1. Acceptance Criteria and Reported Device Performance
The study compares the E. HISTOLYTICA QUIK CHEK™ test to a Composite Reference Method (CRM). While explicit "acceptance criteria" are not listed in terms of specific sensitivity/specificity thresholds, the overall performance against the CRM demonstrates the device's efficacy. The reported performance for prospective samples, while showing lower sensitivity, is accompanied by 100% specificity and PPV, indicating no false positives in this dataset. The retrospective data shows 100% sensitivity and specificity.
Table 1: Acceptance Criteria (Implied) and Reported Device Performance
Metric | Implied Acceptance Goal (vs. CRM) | Reported Performance (Prospective Samples, N=755) | Confidence Limits (95%) | Reported Performance (Retrospective Samples, N=96) | Confidence Limits (95%) |
---|---|---|---|---|---|
Sensitivity | High | 40.0% | 7.3% - 83.0% | 100% | 85.9% - 100% |
Specificity | High | 100% | 99.4% - 100% | 100% | 93.1% - 100% |
Predictive Positive Value | High | 100% | 19.8% - 100% | N/A (not explicitly shown in table) | N/A |
Predictive Negative Value | High | 99.6% | 98.7% - 99.9% | N/A (not explicitly shown in table) | N/A |
Note: The three false-negative results in the prospective study were PCR positive but antigen negative, suggesting a limitation in antigen detection for those specific samples, rather than a lack of E. histolytica presence.
2. Sample Size Used for the Test Set and Data Provenance
- Total Sample Size: 851 fecal samples
- Prospective Samples: 755
- Provenance: Not explicitly stated, but collected from "3 independent sites," implying diverse geographical locations likely within the US given the FDA submission. The context suggests these were real-world clinical samples collected over time.
- Retrospective Samples: 96
- Provenance: Not explicitly stated, but referred to as "frozen (retrospective) specimens." Likely from archived clinical samples.
- Prospective Samples: 755
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not explicitly stated. The ground truth (Composite Reference Method) involves a combination of tests, not directly requiring individual expert interpretation per se, beyond the expertise required to perform and interpret the individual tests (Microscopy, FDA cleared multiplex NAAT test, Alternate PCR with sequencing).
4. Adjudication Method for the Test Set
The adjudication method for the ground truth (Composite Reference Method - CRM) is clearly defined by an algorithm:
Microscopy | FDA cleared multiplex NAAT test | Alternate PCR with sequencing | CRM Result |
---|---|---|---|
Pos | Pos | Pos | Pos |
Pos | Pos | Neg | Neg |
Pos | Neg | Pos | Pos |
Pos | Neg | Neg | Neg |
Neg | Pos | Pos | Pos |
Neg | Pos | Neg | Neg |
Neg | Neg | N/A | Neg |
For cases where Microscopy and multiplex NAAT were both negative, no further PCR testing was performed, and the CRM result was deemed negative. This is a rule-based adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned as part of this submission. The study focuses on the standalone performance of the device against a Composite Reference Method.
6. Standalone Performance Study
Yes, a standalone performance study was conducted. The E. HISTOLYTICA QUIK CHEK™ test was evaluated as an "algorithm only (without human-in-the-loop performance)" device, as it is an in vitro diagnostic which provides a visual qualitative result that is then interpreted by a user. The performance metrics (sensitivity, specificity, PPV, NPV) presented in Tables 1 and 2 directly reflect this standalone capability, comparing its results directly to the CRM. The "visual" interpretation by a human is part of the intended use, but the analytical performance is solely dependent on the device's output.
7. Type of Ground Truth Used
The primary ground truth used was a Composite Reference Method (CRM). This CRM combined:
- Microscopy
- An FDA cleared multiplex NAAT test
- Alternate PCR with sequencing
This approach aims to establish a more robust and accurate determination of the true E. histolytica infection status than any single method alone.
8. Sample Size for the Training Set
The document does not explicitly state a separate "training set" sample size or a description of model training, as this device is a rapid membrane enzyme immunoassay (a chemical/biological test), not an AI/machine learning algorithm that requires training data in the traditional sense. The document describes analytical and clinical validation, which ensures the device's inherent functional performance.
9. How the Ground Truth for the Training Set Was Established
As noted above, this device is not an AI/ML algorithm requiring a training set in the conventional sense. Therefore, the concept of establishing ground truth for a training set does not apply here. The device's underlying principles are based on antibody-antigen reactions. Its design and development would have involved internal validation and optimization, but not "training" on a dataset in the way an AI model would.
Ask a specific question about this device
(423 days)
Giardia SPP 866.3220 Code MHI
Carlsbad, CA 92008
Re: K120001
Trade Name: Uni-Gold™ Giardia Regulation Number: 21 CFR 866.3220 Regulation
Trinity Biotech Uni-Gold™ Giardia is a single use rapid immunoassay for the qualitative detection of Giardia lamblia ( G. lamblia ) antigens in human stool specimens. This test is intended for use with patients with gastrointestinal symptoms as an aid in the diagnosis of suspected Giardia gastrointestinal infections. As with other Giardia tests, results should be considered in conjunction with the clinical evaluation and medical history. For In-Vitro Diagnostic use.
The Trinity Biotech Uni-Gold™ Giardia was designed as a single use, rapid, lateral flow immunoassay test device to detect the presence of Giardia lamblia antigen in unpreserved (fresh & frozen), preserved and media containing human stool specimens.
The Trinity Biotech Uni-Gold™ Giardia test strip (5mm x 60mm) combines a Nitrocellulose Membrane with designated fiber pads (conjugate, sample and absorbant). The test strip is then placed into a plastic housing and is sealed constituting the Test Device.
The Giardia Nitrocellulose Membrane Test Strip - above consist of
- A) Mouse anti-Giardia lamblia antibody is coated onto the Test Line region of the test strip.
- B) Rabbit anti-Goat IgG antibody is coated onto the Control Line region of the Test Strip.
- C) Goat anti- Giardia lamblia antibodies and Goat IgG antibodies are conjugated to red latex particles and dried onto the inert glass fiber (Conjugate Pad) which is inserted into the test strip below the nitrocellulose zone.
When Giardia antigens are present in the sample they combine with the antibody/red latex complex. As the complex migrates it binds to the antibodies in the test region forming a visible pink/red band. (Picture B) This forms the basis for the double antibody sandwich assay. Excess conjugate forms a second pink/red band in the control region of the device. The control line should always appear as a visible pink/red band in the control region of the device. This internal control line is to ensure and indicate that the test device is functioning correctly.
The plastic housing device contains a window where the diluted stool sample is added (Sample Well) and a window above where the results are read in 15 minutes.
The test concept:
Mouse anti-Giardia lamblia is coated onto the test line region of the nitrocellulose zone of the test strip. Rabbit anti- Goat IgG is coated onto the control line region.
Goat anti-Giardia lamblia antibodies are conjugated to red latex particles and dried onto inert glass fiber. This is inserted into the test strip below the nitrocellulose zone.
A buffered solution is added to a dilution tube followed by the addition of the stool specimen (2 drops) via a disposable pipette. This mixture is then dispensed in total into the sample well of the lateral flow cartridge device with a dropper pipette and migrates through a pad containing red microspheres that have been coated with a antibody specific for the Giardia antigen. If the antigen is present, an immune complex forms. The migration continues along the membrane, which contains a striped down anti Giardia capture antibody. If Giardia antigen is present, the immune complex reacts with the anti-Giardia antibody at the test line on the membrane.
Thus Giardia antigens present in the sample combine with the antibody/red latex. As this complex migrates it binds to the antibodies in the test region forming a visible pink/red band.
Excess conjugate forms a second pink/red band in the control region of the device. The control line should always appear as a visible pink/red band in the control region of the device to indicate that the test device is functioning correctly.
Here's a summary of the acceptance criteria and the study details for the Uni-Gold™ Giardia device based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in a quantitative format (e.g., "Sensitivity must be >90%"). However, based on the studies presented, the implicit acceptance criteria are 100% agreement/sensitivity/specificity in the evaluated studies.
Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance | Study Type | N (positive/negative) |
---|---|---|---|---|
Intra-run Precision/Reproducibility | 100% reproducibility | 100% reproducibility | Internal (initial) | 6 blind panel members (2 Low Pos, 2 High Pos, 2 Neg) |
100% reproducibility | Internal (additional) | 12 blind panel members (4 Low Pos, 4 High Pos, 4 Neg) | ||
Agreement vs. Comparator Device | High percent agreement | Site 1: 100% Pos Agr, 94.1% Neg Agr | Retrospective | 26 Pos, 48 Neg (Uni-Gold negative & Comparator Negative, 3 Uni-Gold positive & Comparator negative) |
Site 2: 100% Pos Agr, 100% Neg Agr | Retrospective | 51 Pos, 49 Neg | ||
Site 3: 100% Pos Agr, 83.3% Neg Agr | Retrospective | 54 Pos, 30 Neg (Uni-Gold negative & Comparator Negative, 6 Uni-Gold positive & Comparator negative) | ||
Sensitivity vs. DFA Microscopy | 100% Sensitivity | 100% (91/91) (95% CI 95-100%) | Retrospective | 91 Pos |
Specificity vs. DFA Microscopy | 100% Specificity | 100% (150/150) (95% CI 97-100%) | Retrospective | 150 Neg |
Positive Percent Agreement (PPA) vs. Non-Fluorescent Microscopy (Wheatley's Stain) | 100% PPA | 100% (22/22) | Retrospective | 22 Pos |
Negative Percent Agreement (NPA) vs. Non-Fluorescent Microscopy (Wheatley's Stain) | 100% NPA | 100% (45/45) | Retrospective | 45 Neg |
Positive Percent Agreement (PPA) vs. Non-Fluorescent Microscopy (Iron Hematoxylin Stain) | 100% PPA | 100% (60/60) | Retrospective | 60 Pos |
Negative Percent Agreement (NPA) vs. Non-Fluorescent Microscopy (Iron Hematoxylin Stain) | 100% NPA | 100% (199/199) | Retrospective | 199 Neg |
Specificity vs. DFA Microscopy (Prospective Study) | 100% Specificity | 100% (378/378) (95% CI 99-100%) | Prospective | 378 Neg |
2. Sample Sizes and Data Provenance
-
Test Set for Agreement vs. Comparator Device:
- Total Samples: 267 retrospective samples.
- Matrix types: unpreserved frozen (42), C&S (15), SAF (139), and formalin (71).
- Country of Origin: Not explicitly stated, but the studies were conducted at 3 external laboratories. Given the "US and Canada" mention for overall sample collection, it's likely from these regions.
- Retrospective/Prospective: Retrospective.
-
Test Set for Sensitivity/Specificity vs. DFA Microscopy (Retrospective):
- Total Samples: 241 (91 positive, 150 negative).
- Positive matrix types: formalin (48), SAF (13), unpreserved frozen (17), Cary Blair (3), and C&S (10).
- Negative matrix types: formalin (42), SAF (70), unpreserved frozen (25), Cary Blair (3), and C&S (10).
- Country of Origin: Not explicitly stated, but collected from "Hospitals throughout the US and Canada."
- Retrospective/Prospective: Retrospective.
-
Test Set for Additional Retrospective Studies (Non-Fluorescent Microscopy):
- Site 2 (Wheatley's Stain): 67 retrospective samples (22 positive, 45 negative).
- Site 3 (Iron Hematoxylin Stain): 259 retrospective samples (60 positive, 199 negative).
- Country of Origin: Not explicitly stated, but likely from "US and Canada."
- Retrospective/Prospective: Retrospective.
-
Test Set for Prospective Study (DFA Microscopy):
- Total Samples: 378 negative samples.
- Matrix types: unpreserved fresh (153), unpreserved frozen (45), formalin (45), SAF (45), C&S (45), and Cary Blair (45).
- Country of Origin: Not explicitly stated, but collected from "Hospitals throughout the US and Canada."
- Retrospective/Prospective: Prospective.
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the "number of experts" used to establish ground truth or their specific qualifications (e.g., "Radiologist with 10 years of experience"). However, the ground truth methods (DFA microscopy, Wheatley's Stain, Iron Hematoxylin Stain) are laboratory diagnostic techniques typically performed by trained medical technologists or laboratory personnel. It implies that these established methods served as the expert-level reference.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The "ground truth" was established by the reference methods (DFA microscopy, Wheatley's Stain, Iron Hematoxylin Stain). In cases where the Uni-Gold device and the initial comparator device disagreed (e.g., in the "Agreement vs. Comparator Device" study), further resolution was sought through DFA microscopy or Iron Hematoxylin Stain, which effectively acted as an adjudicator to determine the true positive/negative status for those discrepant samples.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study was done. The study compares the device performance to established laboratory methods or an existing comparator device, not the improvement of human readers with AI assistance.
6. Standalone (Algorithm Only) Performance
Yes, standalone performance was evaluated. The studies directly compare the Uni-Gold™ Giardia device's results (algorithm only, as it's a rapid immunoassay) against the chosen established ground truth methods (DFA microscopy, non-fluorescent microscopy, or a comparator device). There is no "human-in-the-loop" aspect to the device's reading mechanism described.
7. Type of Ground Truth Used
The types of ground truth used were primarily:
- DFA Microscopy: Direct Fluorescent Antibody microscopy.
- Non-Fluorescent Microscopy: Specifically, Wheatley's Stain and Iron Hematoxylin Stain.
- Comparator Device: Remel Xpect™ Giardia Lateral Flow Assay (510K #: K031942) for initial agreement studies, with discrepant results often adjudicated by DFA or Iron Hematoxylin Stain.
8. Sample Size for the Training Set
The document describes a rapid immunoassay device. It does not mention a "training set" in the context of machine learning. The device is a chemical/immunological test, not an algorithm that requires a training set in the AI sense. The development of the assay itself would have involved internal validation and optimization, but not a distinct "training set" as understood in AI studies.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" for an AI algorithm, this question is not applicable to the Uni-Gold™ Giardia immunoassay device.
Ask a specific question about this device
(255 days)
Cryptosporidium SPP 866.3220 Code MHJ
CA 92008
Re: K121565
Trade Name: Uni-Gold™ Cryptosporidium Regulation Number: 21 CFR 866.3220 Regulation
Trinity Biotech Uni-Gold™ Cryptosporidium is a single use rapid immunoassay for the qualitative detection of Cryptosporidium parvum (C. parvum) antigens in human stool specimens. This test is intended for use with patients with gastrointestinal symptoms as an aid in the diagnosis of suspected Cryptosporidium gastrointestinal infections. As with other Cryptosporidium tests, results should be considered in conjunction with the clinical evaluation and medical history. For In-Vitro Diagnostic use.
The Trinity Biotech Uni-Gold™ Cryptosporidium Test was designed as a single use, rapid, lateral flow immunoassay to detect the presence of Cryptosporidium parvum antigen in unpreserved (fresh & frozen), preserved, and media containing human stool specimens. The Trinity Biotech Uni-Gold™ Cryptosporidium test strip (5mm x 60mm) combines a nitrocellulose membrane with designated fiber pads (conjugate, sample and absorbant). The test strip is then placed into a plastic housing and is sealed constituting the Test Device.
Here's an analysis of the acceptance criteria and study details for the Uni-Gold™ Cryptosporidium device, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary does not explicitly state pre-defined "acceptance criteria" in terms of numerical thresholds for sensitivity and specificity. Instead, it presents the results of studies and aims to demonstrate substantial equivalence to a predicate device. For the purpose of this request, I will infer the implicit acceptance is achieving high sensitivity and specificity in comparison to gold standard methods and predicate devices.
Metric / Study | Acceptance Criteria (Implicit) | Reported Device Performance (Uni-Gold™ Cryptosporidium) |
---|---|---|
Sensitivity (vs. DFA Microscopy) | High agreement with DFA microscopy | 100% (77/77) 95% CI 94 - 100% |
Specificity (vs. DFA Microscopy) | High agreement with DFA microscopy | 100% (157/157) 95% Cl 97 - 100% |
PPA (vs. Modified Acid-Fast Stain) | High positive percent agreement | 100% (26/26) |
NPA (vs. Modified Acid-Fast Stain) | High negative percent agreement | 100% (21/21) |
PPA (vs. Modified Kinyoun Stain) | High positive percent agreement | 92% (55/60) |
NPA (vs. Modified Kinyoun Stain) | High negative percent agreement | 90% (198/221) |
Specificity (Prospective Study vs. DFA Microscopy) | High specificity (as no positives were found) | 100% (378/378) 95% Cl 99 - 100% |
Inter-run Precision/Reproducibility | 100% reproducibility across sites and days | 100% reproducibility observed for all sites, for all days |
Cross-Reactivity | No cross-reactivity with common interfering organisms/substances | No cross-reactivity observed with listed organisms; No test interference observed by listed compounds |
2. Sample Size Used for the Test Set and Data Provenance
- Comparator Device Study (Side-by-side with Remel Xpect®): 299 retrospective samples. Data provenance: United States (multiple external sites). Sample matrices included unpreserved frozen, Cary Blair, SAF, and formalin.
- Sensitivity/Specificity Study (vs. DFA Microscopy): 234 retrospective samples (77 positive, 157 negative). Data provenance: United States (2 external laboratories). Sample matrices included formalin, SAF, unpreserved frozen, Cary Blair, and C&S.
- Additional Retrospective Studies (vs. Non-fluorescent Microscopy):
- Site 2: 47 retrospective samples (26 positive, 21 negative) vs. Modified Acid-Fast Stain.
- Site 3: 281 retrospective SAF samples (60 positive, 221 negative) vs. Modified Kinyoun Stain.
- Prospective Study (vs. DFA Microscopy): 378 samples (all negative). Data provenance: one external laboratory in the US. Sample matrices included unpreserved fresh, unpreserved frozen, formalin, SAF, C&S, and Cary Blair.
- Overall Combined Test Set: 940 (prospective and retrospective) samples. Data provenance: Collected from Hospitals throughout the US and Canada.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of individual experts or their exact qualifications (e.g., number of years of experience, specific medical specialty) used to establish the ground truth for the test sets.
Instead, it refers to "DFA microscopy," "Modified Kinyoun Stain light microscopy," and "Modified Acid-Fast Stain" as the reference methods (ground truth). It is implied that these microscopy methods are performed and interpreted by trained laboratory professionals, but specific expert details are not provided.
4. Adjudication Method for the Test Set
The document does not describe an explicit adjudication method (e.g., 2+1 reader consensus) for resolving discrepancies in the reference methods (DFA microscopy, Modified Kinyoun Stain, Modified Acid-Fast Stain).
In cases of discrepancy between the Uni-Gold™ device and the comparator/reference method within the studies, some retesting or re-evaluation against a "higher" ground truth was performed:
- Site 3 (Comparator Device Study): For the 51 samples positive by Uni-Gold™ but negative by comparator, 30 were positive by Modified Kinyoun Stain light microscopy and 3 by DFA microscopy (agreeing with Uni-Gold™). The one sample negative by Uni-Gold™ and positive by comparator was negative by Modified Kinyoun Stain microscopy (agreeing with Uni-Gold™). This indicates that microscopy was used as an adjudicator for samples where the test device and the comparator device disagreed.
- Site 3 (Non-fluorescent Microscopy Study): For the 23 negative samples (by Modified Kinyoun Stain) that tested positive on Uni-Gold™, three subsequently tested positive by DFA microscopy (agreeing with Uni-Gold™). This implies DFA microscopy acted as a higher-level adjudicator for certain discrepancies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This question is not applicable to the Uni-Gold™ Cryptosporidium device. The device is a rapid in vitro diagnostic immunoassay, not an AI-assisted diagnostic tool that involves human readers interpreting images. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed and is not relevant to this type of device.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is also not applicable. The Uni-Gold™ Cryptosporidium device is a standalone lateral flow immunoassay. Its performance is the "algorithm only" performance (the chemical reaction and visual readout) without human intervention in the detection process itself, beyond the initial sample preparation and result interpretation. There is no separate "algorithm without human-in-the-loop" performance as it is the test.
7. The Type of Ground Truth Used
The primary types of ground truth used for the test sets were:
- DFA Microscopy: Direct Fluorescent Antibody microscopy, considered a gold standard for Cryptosporidium detection.
- Modified Kinyoun Stain Light Microscopy: A non-fluorescent microscopy staining method.
- Modified Acid-Fast Stain: Another non-fluorescent microscopy staining method.
- Comparator Device Results: In some comparative studies, the results of the legally marketed predicate device (Remel Xpect® Cryptosporidium) were used for comparison. However, when discrepancies arose in these studies, microscopy methods (DFA or Modified Kinyoun) were often used to adjudicate.
- Clinical Evaluation and Medical History: The intended use statement notes that "results should be considered in conjunction with the clinical evaluation and medical history," implying that clinical context is part of the broader diagnostic process, but the technical ground truth for device performance was microscopy.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of device development or any machine learning algorithm. This device is a biochemical immunoassay, not an AI or machine learning model that requires a discrete training set. The development of the assay's reagents and parameters would have involved internal validation and optimization, but not in the sense of a machine learning training set.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for an AI/ML algorithm in this context, this question is not applicable. The ground truth for developing and optimizing the immunoassay would have been established through standard laboratory practices using known positive and negative controls, and possibly characterized clinical samples, during the assay development process, prior to the formal validation studies presented here.
Ask a specific question about this device
(245 days)
|
| Classification | 21 CFR 866.3220
2011
Re: K103673
Trade/Device Name: The GIARDIA/CRYTOSPORIDIUM QUIK CHEK™ Regulation Number: 21 CFR§ 866.3220
The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antiqen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. The test results should be considered in conjunction with the patient history.
The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane immunoassay for the simultaneous detection of Giardia cyst antigen and Cryptosporidium occyst antigen in a single test device. It is performed with a 25 to 30-minute total incubation time. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test uses monoclonal and polyclonal antibodies to cell-surface antigens of the device contains a Reaction Window with three vertical lines of immobilized antibodies. The Giardia test line ("Giar") contains mouse monoclonal antibodies against Giardia. The Crypto test line ("Cryp") contains mouse monoclonal antibodies against Cryptosporidium. The control line ("C") is a dotted line that contains anti-horseradish peroxidase (HRP) antibodies. The Conjugate consists of polyclonal antibodies coupled to horseradish peroxidase. To perform the test, the sample is added to a tube containing a mixture of Diluent and Conjugate. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, cyst antigens in the sample bind the antibody-peroxidase conjuqates. The antigen-antibody-conjugate complexes migrate through a filter pad to a membrane where they are captured by the immobilized Giardia and/or Cryptosporidium-specific antibodies in the test lines. The Reaction Window is subsequently washed with Wash Buffer, followed by the addition of Substrate. After a 10 minute incubation period, the reaction is examined visually for the appearance of a vertical blue line on either side of the Reaction Window. A blue line indicates a positive "control" reaction, indicated by a vertical dotted blue line under the "C" portion of the Reaction Window. confirms that the test is working properly and the results are valid.
Here's a breakdown of the acceptance criteria and study information for the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ device, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for a diagnostic test like this are typically defined by performance metrics such as sensitivity, specificity, and agreement with a predicate device or gold standard. The reported performance for the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test against the Microscopy - IFA gold standard is as follows:
Performance Metric | Acceptance Criteria (Implicit from reported results and regulatory acceptance) | Reported Device Performance (Giardia) | Reported Device Performance (Cryptosporidium) |
---|---|---|---|
Sensitivity | High (e.g., >95%) | 98.9% (95% CI: 95.7 - 99.8%) | 100% (95% CI: 96.7 – 100%) |
Specificity | High (e.g., >95%) | 100% (95% CI: 99.2 - 100%) | 99.8% (95% CI: 99.0 – 100%) |
Overall Correlation | High (e.g., >95%) | 99.7% (95% CI: 99.7 - 99.7%) | 99.9% (95% CI: 100 – 100%) |
The reported performance against Commercial ELISA Predicate Devices is as follows:
Performance Metric | Acceptance Criteria (Implicit from reported results and regulatory acceptance) | Reported Device Performance (Giardia) | Reported Device Performance (Cryptosporidium) |
---|---|---|---|
Percent Positive Agreement | High (e.g., >95%) | 99.1% (95% CI: 96.3 - 99.8%) | 99.2% (95% CI: 95.2 - 100%) |
Percent Negative Agreement | High (e.g., >95%) | 99.7% (95% CI: 98.7 - 99.9%) | 99.6% (95% CI: 98.7 - 99.9%) |
Overall Percent Agreement | High (e.g., >95%) | 99.5% (95% CI: 99.5 - 99.5%) | 99.5% (95% CI: 99.5 - 99.5%) |
The regulatory acceptance of the device (K103673) by the FDA suggests that these reported performance metrics met their criteria for substantial equivalence to legally marketed predicate devices.
Study Information
-
Sample size used for the test set and the data provenance:
- Test set for comparison to Microscopy (IFA): N = 791 (combined from Study Sites #1 and #3). This included 220 fresh, 140 frozen, 216 preserved-formalin, and 215 preserved-SAF fecal specimens.
- Test set for comparison to Commercial ELISAs (predicate devices): N = 849 (combined from Study Sites #1 and #2). This included 349 fresh, 322 frozen, 36 preserved-formalin, and 142 preserved-SAF fecal specimens.
- Data Provenance: The studies were conducted at "3 geographically diverse sites" within the US. The data is prospective in the sense that the test samples were collected and then tested with the new device, but the text doesn't specify if these were newly collected for the study or a collection of existing samples. Given the various preservation methods, it's likely a mix of prospective collection and retrospective analysis of stored samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical specimens. It refers to "Microscopy - IFA (considered the gold standard)" and "two commercially available ELISAs (predicate devices)" as the comparators, implying the expertise lies in the performance of those established methods.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not specify an adjudication method like 2+1 or 3+1 for resolving discrepancies in the test set. The comparisons are presented as direct comparisons between the new device's results and the results from the gold standard/predicate devices.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. This device is a rapid membrane enzyme immunoassay (a lab test), not an AI-assisted imaging device that would involve human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, this was a standalone performance study of the diagnostic device itself. The "device performance" metrics (sensitivity, specificity, agreement) directly reflect the algorithm's (immunoassay's) ability to detect the antigens in the samples. There is no human-in-the-loop scenario described for the device's main function, as it's a qualitative visual membrane assay. The reading of the lines by a technician would be part of the standard operation of this type of IVD, but the fundamental detection is by the assay itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For clinical performance, the primary ground truth reference was Microscopy - IFA (Immunofluorescence Assay), which is explicitly stated as "considered the gold standard" for detecting Giardia cysts and Cryptosporidium oocysts in fecal specimens.
- Additionally, two commercially available ELISA predicate devices were used as comparators for agreement in other parts of the study.
- For analytical sensitivity (LOD) and cross-reactivity studies, the ground truth involved purified Giardia cysts or Cryptosporidium oocysts quantified by immunofluorescent antibody microscopy (IFA) or documented positive specimens for other organisms by microscopy.
-
The sample size for the training set:
- The document does not explicitly mention a "training set" in the context of machine learning or AI. This is a traditional immunoassay. Therefore, there's no training set as understood in AI/ML development. The development of the assay (e.g., antibody selection, reagent concentrations) would be an iterative process, but not in the "training set" paradigm.
-
How the ground truth for the training set was established:
- Not applicable, as there is no specific "training set" mentioned in the context of an AI/ML algorithm for this immunoassay device. The assay development would rely on internal validation and optimization against known positive and negative samples, similar to how the analytical studies for sensitivity and cross-reactivity were performed.
Ask a specific question about this device
(275 days)
|
| Classification name: | Entamoeba histolytica serological reagents
(21 CFR §866.3220
K081064
Trade/Device Name: Giardia Fecal Antigen Detection Lateral Flow Kit Regulation Number: 21 CFR 866.3220
The IVD Research, Inc. Giardia Fecal Antigen Detection Lateral Flow Kit is a qualitative immunoassay for the detection of Giardia antigens in preserved and unpreserved human fecal specimens. This test is indicated as an aid in the clinical laboratory diagnosis of suspected Giardia infections. For in vitro diagnostic use.
The IVD Research, Inc. Giardia Fecal Antigen Detection Lateral Flow Kit is an immunochromatographic assay for the detection of Giardia lamblia antigen in human fecal samples. The test uses sample wicking to capture Giardia antigen on a discrete test line containing antibodies specific for Giardia antigen. A specimen is added to a dilution tube and mixed with a buffer solution. The mixture is dispensed into the sample well of the device which resolubilizes the Giardia specific mouse monoclonal antibody that has been conjugated to colored microparticles. This solution wicks along a membrane containing capture antibodies bound to the membrane at the Test and Control lines. The Giardia immune complex, if present, reacts with antibody at the Test line. Unbound conjugate not captured at the test line is captured at the Control line containing anti-mouse antibody. If Giardia antigens are present in the fecal sample, two pink-to-purple bands (one at the Sample line and one at the Control line) will appear in the test window. If no Giardia antigen is present, or if the level of antigen is below the detection limit of the assay, only one pink-to-purple band at the Control line will appear in the test window. For the test to be valid, a pink-to-purple band must always appear at the Control line position of the device test window regardless of whether the sample is positive or negative. This Control line indicates that the test is working properly.
Acceptance Criteria and Study for IVD Research, Inc. Giardia Fecal Antigen Detection Lateral Flow Kit
This response will detail the acceptance criteria and the study performance for the Giardia Fecal Antigen Detection Lateral Flow Kit, as extracted from the provided 510(k) Pre-market Notification.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity and specificity. Instead, the "acceptance" is implied through a comparison to a predicate device and a reference method, aiming for "substantial equivalence" and acceptable clinical performance.
Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Sensitivity | Substantially equivalent to predicate device and reference method | 97.2% (106/109) [95% CI = 92.2% to 99.4%] |
Specificity | Substantially equivalent to predicate device and reference method | 96% (97/101) [95% CI = 90.2% to 98.9%] |
Reproducibility | 100% agreement with expected results | 100% (54/54 positive, 36/36 negative) |
Analytical Sensitivity (LOD) | A detectable limit suitable for clinical diagnosis | 20 ng/ml |
Analytical Specificity/Cross-reactivity | No cross-reactivity with common related organisms/substances | No cross-reactivity observed with tested organisms/substances. No interference with blood, mucin, Imodium®. |
Percent Positive Agreement (vs. Predicate) | High agreement with predicate device (Xpect™ Giardia Lateral Flow Assay) | 98% (48/49) |
Percent Negative Agreement (vs. Predicate) | High agreement with predicate device (Xpect™ Giardia Lateral Flow Assay) | 100% (61/61) |
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Performance Study:
- Sample Size: A total of 210 human fecal specimens were used.
- Data Provenance: Retrospective. The samples were "well-characterized, archived samples" collected in 10% formalin or SAF. They were submitted to an independent laboratory for testing. The country of origin is not specified but is implicitly the US given the FDA submission.
- Unpreserved Fecal Specimens Study:
- Sample Size: 42 unpreserved fecal samples (15 positive, 27 negative)
- Data Provenance: Retrospective. These samples were from IVD Research's frozen sample bank, stored at -15℃ or lower. The country of origin is implicitly the US.
- Reproducibility Study:
- Sample Size: A masked panel of ten samples (number of positive/negative not specified, but 6 positive and 4 negative in the summary table). These were tested repeatedly.
- Data Provenance: Not explicitly stated, but performed at three sites (one internal, two external), suggesting a multi-center study, likely within the US.
- Predicate Device Comparison Study:
- Sample Size: Total of 110 samples (49 positive by predicate, 61 negative by predicate).
- Data Provenance: Not explicitly stated, but implies existing samples were used for comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Clinical Performance Study (Reference Method: Microscopy or Direct Immunofluorescence Assay): The document does not specify the number of experts or their qualifications for establishing the ground truth using microscopy or direct immunofluorescence assay. It states that the specimens were "submitted to an independent laboratory for testing," implying expertise within that laboratory. For the 4 false positives, these were "re-tested using a direct immunofluorescence assay and shown to be positive," suggesting a re-evaluation by an expert or a highly sensitive method.
- Unpreserved Fecal Specimens Study (Reference Method: Giardia lamblia Antigen Detection Microwell ELISA): The ground truth was established by another IVD Research, Inc. product, the Giardia lamblia Antigen Detection Microwell ELISA. No human experts are explicitly mentioned for this ground truth establishment, as it is an automated assay.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for reconciling discrepancies for the primary clinical performance test set. For the "false positive" samples identified with the Giardia LF test (4 cases), they were "re-tested using a direct immunofluorescence assay and shown to be positive," which implies a secondary, definitive reference method was used to adjudicate these discrepancies and re-classify them as true positives. For the comparison with the predicate device, the agreement was calculated directly, without mention of an adjudication process for discordant results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was mentioned. The device is a Lateral Flow Kit, which generally involves visual interpretation by a single user, not an AI or human reader improvement scenario.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This information is not applicable. The device is a Lateral Flow Kit, which is a diagnostic assay that provides a visual result (bands) requiring human interpretation, not an algorithm running independently.
7. Type of Ground Truth Used for Clinical Performance
- Clinical Performance Study: The ground truth was established by "Microscopy or Direct Immunofluorescence Assay." This represents an expert-determined or highly sensitive laboratory reference method, likely considered the gold standard for clinical diagnosis of Giardia at the time.
- Unpreserved Fecal Specimens Study: The ground truth was established by a Giardia lamblia Antigen Detection Microwell ELISA (IVD Research, Inc.). This is a laboratory assay serving as the reference.
8. Sample Size for the Training Set
The document does not provide information about a "training set" in the context of an algorithm or machine learning model. The studies described are performance evaluations of the completed device. For IVD diagnostic kits, development typically involves internal analytical studies and optimization, rather than a distinct "training set" as understood in AI/ML.
9. How the Ground Truth for the Training Set Was Established
As there is no explicit mention of a "training set" or a machine learning algorithm, this question is not applicable to the provided information.
Ask a specific question about this device
(127 days)
21 CFR 866.3220
21 CFR 866.3220
, VA 24060-6358
K051929 Trade/Device Name: GIARDIA/CRYTOSPORIDIUM CHECK™ Regulation Number: 21 CFR 866.3220
The GIARDIA/CRYPTOSPORIDIUM CHEK™ test is an enzyme immunoassay for the qualitative detection of Giardia cyst and Cryptosporidium oocyst antigen in human fecal specimens. It is indicated for use as an aid in the diagnosis of patients with diarrhea suspected of Giardia and/or Cryptosporidium gastrointestinal infections.
The GIARDIA/CRYPTOSPORIDIUM CHEK™ test uses monoclonal and polyclonal antibodies to cell-surface antigens of Giardia and an oocyst antigen of Cryptosporidium. The Microassay Plate in the kit contains immobilized monoclonal antibodies against the antigens and the Conjugate consists of polyclonal antibodies against the antigens. In the assay, an aliquot of a diluted fecal specimen is transferred to a microassay well. The immobilized monoclonal antibodies bind the Giardia and Cryptosporidium antigens if the antigens are present. Upon addition, Conjugate then binds to the antigen/antibody complex. Any unbound materials are removed during the washing steps. Following the addition of substrate, a color is detected due to the enzyme-antibody-antigen complexes that formed in the presence of Giardia or Cryptosporidium antigens and Conjugate.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of clinical performance thresholds. Instead, it presents the device's performance in comparison to predicate devices (ELISA) and microscopy. The implication is that the performance must be demonstrably comparable or superior to these established methods.
Here's the reported device performance:
Performance Metric | Target/Comparison | Reported Device Performance (GIARDIA/CRYPTOSPORIDIUM CHEK™) |
---|---|---|
Comparison to Commercial ELISA | Agreement with ProSpecT® Giardia/Cryptosporidium Microplate Assay | |
Positive Agreement | - | 98.6% (95% CI: 96.2% - 99.6%) |
Negative Agreement | - | 98.7% (95% CI: 96.4% - 99.6%) |
Overall Agreement | - | 98.6% (95% CI: 98.4% - 98.8%) |
Comparison to Microscopy (IFA-confirmed) | Performance against Microscopic Analysis | |
Sensitivity | - | 97.6% (95% CI: 92.6% - 99.4%) |
Specificity | - | 100% (95% CI: 95.1% - 100%) |
Predictive Positive Value | - | 100% (95% CI: 96.2% - 100%) |
Predictive Negative Value | - | 96.9% (95% CI: 90.5% - 99.2%) |
Correlation | - | 98.6% (95% CI: 98.2% - 98.9%) |
Reproducibility | Consistent results across labs and trials | 100% correlation across 3 sites and 3 trials |
Analytical Sensitivity | Detection limit for Giardia cysts and Cryptosporidium oocysts | 375 Giardia cysts/mL, 6250 Cryptosporidium oocysts/mL |
Analytical Specificity | No cross-reactivity with common intestinal pathogens | No cross-reactivity observed |
Interfering Substances | No effect on assay performance at tested concentrations | No effect on positive or negative results |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Evaluation compared to ELISA): n = 590 fecal specimens.
- Sample Size (Clinical Evaluation compared to Microscopy): n = 217 fecal specimens.
- Data Provenance: The samples were collected from "three clinical study sites" (Table 1) and specifically "Site 1" for the microscopy comparison (Table 2). The country of origin is not explicitly stated, but the submission is to the U.S. FDA, suggesting the studies were likely conducted in the U.S. The data appears to be prospective clinical evaluation data, as it describes "clinical evaluations" and samples "used in the clinical evaluation of the test."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing the ground truth.
- For the comparison to the commercially available ELISA (ProSpecT® Giardia/Cryptosporidium Microplate Assay), the ELISA itself served as a reference.
- For the comparison to microscopy, the ground truth was established by "IFA-confirmed microscopy results." The individuals performing or confirming these IFA microscopy results are not described.
4. Adjudication Method for the Test Set
The document does not describe an explicit adjudication method. The comparisons are made directly against the results of the reference methods (commercial ELISA or IFA-confirmed microscopy).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This device is an enzyme immunoassay (ELISA), not an AI-powered image analysis or diagnostic tool that involves human readers interpreting results with or without AI assistance. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study of human readers with AI assistance is not applicable and was not performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are for the device (GIARDIA/CRYPTOSPORIDIUM CHEK™) in a standalone capacity. The device itself performs the qualitative detection of antigens. The performance metrics (sensitivity, specificity, agreement) are for the assay's ability to detect the target antigens directly, without explicit human-in-the-loop interpretation that would significantly alter its performance.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the clinical evaluations was established by:
- Predicate Device Performance: Comparison to a "commercially available ELISA" (ProSpecT® Giardia/Cryptosporidium Microplate Assay) which itself is an established diagnostic method.
- IFA-Confirmed Microscopy: This is a laboratory diagnostic method where trained personnel visualize parasites in samples, and the "IFA-confirmed" implies an additional, highly specific immunological staining technique to confirm the identity of Giardia cysts and Cryptosporidium oocysts. This can be considered a form of expert-driven laboratory confirmation.
8. The Sample Size for the Training Set
This document describes a diagnostic test kit (ELISA), not a machine learning or AI algorithm. Therefore, the concept of a "training set" in the context of AI development is not applicable. The device's components (antibodies, reagents) are developed and optimized through laboratory research, but not via training on a data set in the same manner as an AI model.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" is not applicable for this type of device. The ground truth for the analytical studies (sensitivity, specificity, cross-reactivity) was established using known positive and negative controls, spiked samples with quantified antigen levels, and samples positive for other specific pathogens.
Ask a specific question about this device
(16 days)
Virginia 24060-6358
K052932 Re:
Trade/Device Name: CRYPTOSPORIDIUM II Regulation Number: 21 CFR § 866.3220
The CRYPTOSPORIDIUM II test is an enzyme immunoassay for the qualitative detection of Cryptosporidium oocyst antigen in human fecal specimens. It is indicated for use as an aid in the diagnosis of patients with diarrhea suspected of Cryptosporidium gastrointestinal infection. FOR IN VITRO DIAGNOSTIC USE.
Not Found
The provided text is a 510(k) premarket notification letter from the FDA, granting market clearance for the "CRYPTOSPORIDIUM II" device. It outlines the device's indications for use and states that it has been deemed substantially equivalent to a legally marketed predicate device.
However, the provided text does not contain the detailed study information required to answer your specific questions regarding acceptance criteria, reported device performance, sample sizes, ground truth establishment, expert qualifications, or MRMC comparative effectiveness studies. This type of information is typically found in the 510(k) summary or the actual study reports submitted to the FDA, not in the clearance letter itself.
The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device we have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA)." This indicates that the device was cleared based on substantial equivalence, not necessarily on a de novo study proving it meets specific acceptance criteria with detailed performance metrics in the way you've outlined for an AI/imaging device.
Therefore, I cannot populate the table or answer the subsequent questions based solely on the provided text. To get this information, you would need to access the full 510(k) submission for K052932, which might be publicly available through FDA databases or by request.
Ask a specific question about this device
(147 days)
Giardia SPP 866.3220 Code: MHI
Intended Use:
REMEL's Xpect™ Giardia kit is an in vitro qualitative
NW Ramsey, MN 55303
Re: K031942
Trade/Device Name: Xpect™ Giardia Regulation Number: 21 CFR 866.3220
REMEL's Xpect™ Giardia kit is an in vitro qualitative immunoassay for the detection of Giardia antigens in preserved and unpreserved fecal specimens. This test is intended as an aid in the laboratory diagnosis of suspected Giardia infections.
The Xpect™ Giardia Lateral Flow Assay is a chromatographic immunoassay that detects the presence of Giardia antigen. The test utilizes sample wicking to capture Giardia antigen on a discrete test line containing antigen-specific antibodies for Giardia. A specimen is added to a dilution tube containing a buffered solution. A conjugate containing colored micro-particles linked to murine monoclonal antibody specific for Giardia is added. The mixture is dispensed into the sample well of the device and wicks along a membrane containing capture antibody stripes. The Giardia immune complex, if present, reacts with anti- Giardia antibody at the test line. Antibody-labeled microparticles not bound at the test line are later captured at the control line containing anti-mouse antibody. A blue line of any intensity (light blue to black) will appear at the Giardia test position if Giardia antigen is present. A complete line at the Control position indicates that the test is working properly.
Here's a breakdown of the acceptance criteria and study information for the Xpect™ Giardia Lateral Flow Assay, based on the provided 510(k) notification:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance (Xpect™ Giardia) |
---|---|
Preamble device comparison: Not substantially equivalent to the predicate device | Substantially equivalent to the predicate device (Becton Dickinson ColorPAC™ Giardia/Cryptosporidium Rapid Assay) |
Sensitivity for Giardia detection: | 97.9% (95/97); 95% Confidence Interval (CI) = 92.8-99.4% |
Specificity for Giardia detection: | 97.1% (464/478); 95% CI = 95.1-98.2% |
Percent Agreement with Predicate Device (Positive Agreement): | 92.3% (24/26) |
Percent Agreement with Predicate Device (Negative Agreement): | 94.2% (114/121) |
Overall Percent Agreement with Predicate Device: | 93.9% (138/147) |
Cross-reactivity: No cross-reactivity with specified organisms | No cross-reactivity observed with a list of 24 organisms/pathogens (excluding Astrovirus and Caliciviruses for which it was not established). |
Interfering Substances: No interference from specified substances | No interference with expected results from blood, mucin, fecal fat, or common anti-diarrheal products (Pepto-Bismol®, Imodium® A-D, and Kaopectate®). |
Reproducibility: Consistent results across sites and days with varying activity. | 100% of 630 samples tested produced the expected result across seven sites (including one in-house) on three separate days with ten blinded samples of varying activity. |
Study Information
-
Sample Size Used for the Test Set and Data Provenance:
- Clinical Performance Study (compared to microscopy):
- Total samples with Giardia present by microscopy: 97
- Total samples with Giardia absent by microscopy: 478
- Total samples in this comparison: 575 (97 positive + 478 negative)
- Data Provenance: The study was evaluated at "six geographically diverse laboratories," indicating a multi-center, likely prospective collection of clinical samples. The country of origin is not explicitly stated but can be inferred to be within the US given the submission to the FDA.
- Comparison to Predicate Device:
- Total samples: 147 (26 positive by predicate, 121 negative by predicate).
- Clinical Performance Study (compared to microscopy):
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
- The primary ground truth for the clinical performance study was microscopy. The document does not specify the number of experts (e.g., medical technologists, parasitologists) involved, nor their specific qualifications (e.g., years of experience). It simply states the comparison was "compared to microscopy."
-
Adjudication Method for the Test Set:
- The document does not explicitly state an adjudication method (like 2+1, 3+1). It implies that the microscopy results served as the definitive ground truth for the comparison.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a rapid diagnostic test (lateral flow immunoassay), not an AI-powered diagnostic imaging or interpretation tool designed to assist human readers.
-
Standalone (Algorithm Only) Performance:
- Yes, the performance presented in the "Sensitivity/Specificity" table is the standalone performance of the Xpect™ Giardia Lateral Flow Assay, as it directly compares the device's results to the microscopy ground truth. It operates independently of human interpretation beyond reading the positive/negative line.
-
Type of Ground Truth Used:
- The primary ground truth used for the clinical performance evaluation was microscopy, which involves expert examination of fecal specimens for the presence of Giardia parasites.
-
Sample Size for the Training Set:
- The document does not specify a separate "training set" or its size. As this is a lateral flow immunoassay, the development process would involve analytical studies and optimization rather than machine learning training on a distinct dataset. The studies described are performance validation studies.
-
How the Ground Truth for the Training Set Was Established:
- Since a separate training set, in the context of machine learning, is not applicable or described for this device, the method for establishing its ground truth is not provided. The ground truth for the validation test set was established by microscopy.
Ask a specific question about this device
(141 days)
Cryptosporidium SPP 866.3220 Code: MHJ
Intended Use:
REMEL's Xpect™ Cryptosporidium kit is an in
Lenexa, KS 66215
Re: K031965
Trade/Device Name: Xpect™ Cryptosporidium Regulation Number: 21 CFR 866.3220
REMEL's Xpect™ Cryptosporidium kit is an in vitro qualitative immunoassay for the detection of Cryptosporidium antigens in preserved and unpreserved fecal specimens. This test is intended as an aid in the laboratory diagnosis of suspected Cryptosporidium infections.
The Xpect™ Cryptosporidium Lateral Flow Assay is a chromatographic immunoassay that detects the presence of Cryptosporidium antigen. The test utilizes sample wicking to capture Cryptosporidium antigen on a test line containing antigen-specific antibody. A specimen is added to a dilution tube containing a buffered solution. A conjugate containing colored micro-particles linked to murine monoclonal antibody specific for Cryptosporidium is added. The mixture is dispensed into the sample well of the device and wicks along a membrane containing capture antibody stripes. The Cryptosporidium immune complex, if present, reacts with anti-Cryptosporidium antibody at the test line. Conjugate not bound at the test line is later captured at the control line containing anti-mouse antibody. A red line of any intensity will appear at the Cryptosporidium test position if Cryptosporidium antigen is present. A line in the Control position indicates that the test is working properly.
Here's a summary of the acceptance criteria and study details for the Xpect™ Cryptosporidium Lateral Flow Assay, based on the provided 510(k) notification:
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Sensitivity | Not explicitly stated (implied to be high enough for clinical utility relative to microscopy) | 96.4% (95% CI: 91.2-98.6%) |
Specificity | Not explicitly stated (implied to be high enough for clinical utility relative to microscopy) | 98.3% (95% CI: 96.6-99.1%) |
Cross-Reactivity | No cross-reactivity with common intestinal parasites and bacteria. | No cross-reactivity observed with 24 specified organisms, including Ascaris lumbricoides, Blastocystis hominis, Campylobacter coli, Giardia lamblia, and others. |
Interfering Substances | No interference with common substances (blood, mucin, fecal fat) and anti-diarrheal products. | No interference observed with blood, mucin, fecal fat, Pepto-Bismol®, and Kaopectate®. Interference observed with Imodium® A-D at 20% (v/v) for low levels of Cryptosporidium antigen. |
Reproducibility | Consistent results across different sites and days. | 100% agreement with expected results for 630 samples tested across seven sites (including one in-house site) over three days. |
Percent Agreement vs. Microscopy (for comparison with predicate) | Not explicitly stated, but typically a high percentage agreement is expected. | 94.5% for Xpect™ Cryptosporidium (vs. microscopy), 84.9% for Predicate Device (vs. microscopy) |
Study Details
-
Sample sizes used for the test set and data provenance:
- Sensitivity/Specificity Study:
- Total Samples: 578 (112 positive for Cryptosporidium by microscopy, 466 negative by microscopy).
- Provenance: Not explicitly stated regarding country of origin; the study was conducted at "six geographically diverse laboratories" (implying multiple sites within a country or potentially internationally), and data was retrospective (compared to microscopy, which would have been performed on collected samples).
- Percent Agreement Study (Comparison with Predicate):
- Total Samples: 146 (30 positive for Cryptosporidium by microscopy, 116 negative by microscopy).
- Provenance: Not explicitly stated regarding country of origin or retrospective/prospective nature.
- Sensitivity/Specificity Study:
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth for the test set was established by microscopy. The document does not specify the number of experts (e.g., microscopists) involved in reading these slides or their qualifications. It's common in diagnostic studies for the comparator method (like microscopy) to be performed by qualified laboratory personnel, but this detail is not provided.
-
Adjudication method for the test set:
- The document does not specify any formal adjudication method for the microscopy results that served as the ground truth. It simply states the comparison was "to microscopy."
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is a standalone in vitro diagnostic (IVD) lateral flow assay, not an AI or imaging device designed to assist human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, this was a standalone performance study. The Xpect™ Cryptosporidium Lateral Flow Assay is an "algorithm-only" in the sense that it provides a direct qualitative result (presence/absence of a line) without human interpretation beyond reading the line. The presented sensitivity and specificity values represent the performance of the device itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth used was microscopy (specifically, for detection of Cryptosporidium).
-
The sample size for the training set:
- The document does not provide information regarding a separate training set for algorithm development. As this is a lateral flow immunoassay, not an AI or machine learning model, a distinct "training set" in the context of data-driven algorithm development is typically not applicable. The assay's performance characteristics are inherent to its design and manufacturing.
-
How the ground truth for the training set was established:
- As there's no explicit mention of a training set in the context of algorithm development, this information is not applicable to this device. Early development and optimization of the assay would involve various characterized positive and negative samples, but these are not typically referred to as a "training set" in the same way as for AI.
Ask a specific question about this device
Page 1 of 2