K Number
K140455

Validate with FDA (Live)

Date Cleared
2014-05-22

(87 days)

Product Code
Regulation Number
866.3520
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trinity Biotech Captia™ Measles IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings.

Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.

Device Description

The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assays (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H2SO4. The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.

AI/ML Overview

Here's an analysis of the acceptance criteria and the studies performed for the Trinity Biotech Captia™ Measles IgM ELISA Test Kit, presented in the requested format:

Acceptance Criteria and Device Performance for Trinity Biotech Captia™ Measles IgM ELISA

The provided document describes the performance characteristics and clinical studies of the Trinity Biotech Captia™ Measles IgM ELISA. The acceptance criteria are not explicitly stated as pass/fail thresholds in a dedicated table, but are inferred from the reported performance characteristics.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
Analytical Performance
High Dose Hook EffectNo dose hook effect (prozone) of antigenAntigen is titrated to ensure no dose hook effect. Positive control in each kit lot is titrated to negative at each kit QC test stage to ensure no prozoning.
IgM/IgG CompetitionDiminished competing virus-specific IgG and minimized RF interferenceSerum Diluent Plus Solution diminishes competing virus-specific IgG (maximum 1380 mg/dL removed) and minimizes rheumatoid factor interference (highest titer of RF+ tested (1:1280; 500 IU/mL) did not adversely affect performance).
Precision (Interassay)Acceptable %CV across sites, operators, and daysInterassay (Between All Sites): %CV ranged from 4.3% to 12.0% across 6 samples (averaged across 3 sites, each with 2 operators, 10 days, triplicate runs; N=180 data points/sample). Inter-Lot Reproducibility: %CV ranged from 3.19% to 22.55% across 6 samples (averaged across 3 kit lots, triplicate runs). Intra-run (Between Users): %CV ranged from 0.40% to 5.00% across 6 samples (averaged across 2 operators, 10 days, triplicate runs; N=60 data points/sample).
Analytical Specificity (Cross-Reactivity)Limited or acceptable cross-reactivity with common interfering agentsObserved cross-reactivity: CMV IgM (3 positive, 1 equivocal out of 7), HSV1 IgM (7 positive out of 16), HSV2 IgM (2 positive, 1 equivocal out of 10). No cross-reactivity observed: Mumps IgM (4/4 negative), VZV IgM (4/4 negative), Rubella IgM (4/4 negative). Not ruled out: Parvovirus, Respiratory Syncytial Virus (RSV), Parainfluenza.
Analytical Specificity (Interfering Substances)No adverse results or significant change with tested interferentsNo adverse results or significant change observed with Hemoglobin (≥ 500 mg/dL) and Bilirubin (unconjugated, ≥ 20 mg/dL). Not tested: Red cells, cholesterol, gamma globulin, triglycerides, beta carotene, protein, ascorbic acid, anticoagulants.
Assay Cut-off DeterminationClear distinction between negative, equivocal, and positiveCut-off ISR = 1.00. Values ≤ 0.90 considered negative, values ≥ 1.10 considered positive, and 0.91 to 1.09 considered equivocal. Based on a study of 228 Measles IgM negative sera (mean OD=0.105, SD=0.070), positive threshold determined as mean + 4 SD = 0.385.
SeroconversionConsistent IgM response with seroconversion panelDemonstrated IgM response consistent with seroconversion panel (BIOMEX SCP-MEA-001) across 4 different lots. Trinity kit showed seroconversion at Day 20.
Specificity in Normal PopulationLow incidence of positive results in healthy individualsOut of 200 random serum samples, 0.5% (1 sample) was in the 0.91-1.10 equivocal range, 0.5% (1 sample) was in the 1.21-1.40 positive range, and 0.5% (1 sample) was >2.00 positive. The majority (approximately 92.5%) were in the 0.00-0.60 negative range.
Clinical Performance
Positive Agreement (Study 1 & 2)High agreement with comparator algorithm for positive samples90.7% Positive Agreement (88/97 samples) with 95% CI (83.3-95.0%) (when indeterminate comparator algorithm results were included in total).
Negative Agreement (Study 1 & 2)High agreement with comparator algorithm for negative samples80.5% Negative Agreement (70/87 samples) with 95% CI (70.9-87.4%) (when indeterminate comparator algorithm results were included in total).
Equivocal Rate (Study 1 & 2)Low percentage of equivocal results8.6% (16 of 187) of Captia™ Measles IgM results were equivocal.
Positive Agreement (Study 3)High agreement with comparator PCR algorithm for positive samples75.00% Positive Agreement (3/4 samples) with 95% CI (30.1-95.4%) (when indeterminate comparator algorithm results were included in total).
Negative Agreement (Study 3)High agreement with comparator PCR algorithm for negative samples100.00% Negative Agreement (4/4 samples) with 95% CI (39.8-100.0%) (when indeterminate comparator algorithm results were included in total).

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Studies 1 & 2): 188 samples from suspected measles infection cases.
    • Data Provenance: Retrospective, collected from individuals in whom a measles IgM test was ordered by the Texas Department of Health, Austin, Texas (66 samples) and the State Laboratory of Public Health in Raleigh, North Carolina (122 samples), USA.
  • Test Set (Clinical Study 3): 8 samples.
    • Data Provenance: Retrospective, collected from submissions to the Kansas Department of Health and Environment Laboratory (KDHE), Topeka, Kansas, USA.
  • Test Set (Specificity in a Normal Population): 200 random serum samples.
    • Data Provenance: Not explicitly stated, but implied to be from a general population (individuals between ages 18-65 with no known clinically apparent Measles infection). Location not specified.
  • Test Set (Cross-Reactivity): 47 samples.
    • Data Provenance: Confirmed positive for other IgM antibodies (CMV, HSV1, HSV2, Rubella, VZV, Mumps, Parvo-B19) by Trinity Biotech Captia™ test kits. Location not specified.
  • Test Set (Precision/Reproducibility): 6 blinded panel members (4 low-to-moderate positive, 2 negative).
    • Data Provenance: Not specified, likely internal or commercially sourced panels.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not specify the number of experts used to establish the ground truth for the test sets in the clinical studies.
  • The ground truth in Clinical Studies 1 & 2 was established using a "comparator testing algorithm" which consisted of "the results of other laboratory confirmation methods including comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutionation Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit." This implies multiple laboratory results were used to form a consensus or a defined algorithm, rather than individual expert adjudication of each case.
  • Similarly, in Clinical Study 3, the "comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples."

4. Adjudication Method for the Test Set

  • For Clinical Studies 1 & 2, the ground truth was established by a "comparator algorithm" based on multiple laboratory confirmation methods (e.g., other ELISAs, IFA, CF, HI). The specific adjudication rule (e.g., 2 out of 3, 3 out of 4) is not explicitly detailed, but it generally implies a form of consensus derived from multiple test results rather than human expert adjudication.
  • For Clinical Study 3, the ground truth was established by a "comparator algorithm" based on PCR results from various sample types. Again, the specific adjudication rules for discrepant PCR results (e.g., "NPS negative at two sites, urine positive at one site and urine indeterminate at one site") are not explicitly detailed for how the final "positive," "indeterminate," or "negative" comparator algorithm result was reached.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not performed. This device is an Enzyme-Linked Immunosorbent Assay (ELISA), which is an automated or semi-automated laboratory test. It does not involve human readers interpreting images or data where AI assistance would be applicable in the traditional sense of an MRMC study for imaging diagnostics. The "readers" are the laboratory technicians operating the ELISA equipment.

6. Standalone Performance Study

  • Yes, standalone performance was done for the device. All reported performance metrics (analytical performance, seroconversion, cross-reactivity, clinical agreement with comparator algorithms) represent the performance of the Captia™ Measles IgM ELISA as an algorithm/device only, without human "in-the-loop" assistance for interpreting the result of the ELISA itself. While humans perform the test and interpret the final ISR value against predefined cut-offs, the "algorithm" is determining the ISR.

7. Type of Ground Truth Used

  • The ground truth used was a combination of other laboratory confirmation methods and algorithms.
    • For Clinical Studies 1 & 2: "comparator testing algorithm used at each institution to determine the presence of current or recent measles infection" which included "comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutionation Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit." This represents a form of expert-defined algorithm/consensus from multiple laboratory tests.
    • For Clinical Study 3: "comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples." This is essentially molecular pathology/laboratory results.
    • For Cross-Reactivity: Confirmed positive status based on other specific IgM ELISA tests.
    • For Seroconversion: A commercially available seroconversion panel.

8. Sample Size for the Training Set

  • The document does not explicitly state a separate "training set" in the context of machine learning model development. This is an ELISA kit, not typically a machine learning algorithm that requires a distinct training phase in the same way. The development and optimization of the assay (e.g., antigen titration, calibrator development, component formulation) would be part of an iterative development process that could be considered analogous to "training," but not in the sense of a labeled dataset for an AI model.
  • The cut-off determination study involved 228 Measles IgM negative sera, which were used to establish the positive threshold (mean + 4SD). This could be considered a form of "training" or "calibration" data for the assay's interpretive criteria.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, there isn't a traditional "training set" for a machine learning algorithm.
  • For the cut-off determination, the 228 Measles IgM negative sera were explicitly described as "Measles IgM negative sera" and were used to calculate statistical parameters (mean and standard deviation of optical density readings) to define the initial positive threshold. Four samples were disqualified because they were in the equivocal range on Trinity IgM ELISA or indeterminate on a comparator Measles IFA test kit (due to non-specific staining). This implies an initial screening or pre-classification of these samples as "negative" before their use in cut-off calculation. The final cut-off was derived using these empirical observations.

{0}------------------------------------------------

Trinity Biotech Captia™ Measles IgM 510(k) Summary

Trinity Biotech hereby submits this 510(k) summary for the Captia™ Measles IgM Test in accordance with the requirements of 21CFR 807.92(C)

Submitter'sldentification:
Name &Trinity Biotech
Address2823 Girts Road
Jamestown, NY 14701, USA
Contact:Bonnie B. DeJoy, Corporate Vice President of Regulatory Affairs
Email: bonnie.dejoy@trinityusa.com
Phone:716-483-3851 x 1030
Fax:716-488-1990

A. 510(k) Number:

K140455

B. Purpose for Submission:

Traditional 510(k) in accordance with the requirements of 21CFR 807.92(C)

C. Measurand:

Measles-specific IgM antibodies in human serum

D. Type of Test:

Enzyme-Linked Immunosorbent Assay (ELISA)

E. Applicant:

Trinity Biotech

F. Proprietary and Established Names:

Captia™ Measles IgM

G. Regulatory Information:

1. Regulation section:

21 CFR 866.3520 Rubeola (measles) virus serological reagents

{1}------------------------------------------------

    1. Classification:
      Class I
    1. Product code:
      PCL
    1. Panel:
      83 (Microbiology)

H. Intended Use:

  • l. Intended use(s):
    The Trinity Biotech Captia™ Measles IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings.

Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.

    1. Indication(s) for use:
      Same as Intended Use
    1. Special conditions for use statement(s):
      For prescription use only
    1. Special instrument requirements:
      Not applicable

I. Device Description: .

The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assays (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen

{2}------------------------------------------------

complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H2SQ2. The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.

Each kit contains the following components in sufficient quantities to perform the number of tests indicated on the package label.

    1. Purified measles antigen coated microassay plate: 96 wells, configured in twelve 1x8 strips stored in a foil pouch with desiccant. (96T: one plate)
    1. Calibrator: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and pen/strep (0.01%) added as preservatives, with kit specific factor printed on vial label. The Calibrator is used to calibrate the assay to account for day-to-day fluctuations in temperature and other testing conditions. (96T: one vial, 0.4 mL) *
    1. Positive Control: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and pen/strep (0.01%) added as preservatives, with established range printed on vial label. The Positive Control is utilized to control the positive range of the assay.(96T:one vial, 0.4 mL) *
  • Negative Control: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and 4. pen/strep (0.01%) added as preservatives, with established range printed on vial label. The Negative Control is utilized to control the negative range of the assay (96T: one vial, 0.4 mL) *
    1. Horseradish-peroxidase (HRP) Conjugate: Ready to use. Goat anti-human IgM, containing ProClin® (0.1%) and gentamicin as preservatives. (96T: one bottle, 16 mL)
    1. Serum Diluent Plus: Ready to use, Contains goat/sheep anti-human IgG for serum absorption to remove competing IgG, with protein stabilizers and ProClin® (0.1%) as a preservative. (96T: two bottles, 45 mL each)
  • 7 . Wash Buffer Type I (20X concentrate): dilute 1 part concentrate + 19 parts deionized or distilled water. Contains TBS, Tween-20 and ProClin® (0.1%) as a preservative. (96T: one bottle, 50 mL)
    1. Chromogen/Substrate Solution Type I: Tetramethylbenzidine (TMB), ready to use. The reagent should remain closed when not in use. If allowed to evaporate, a precipitate may form in the reagent wells. (96T: one bottle, 15 mL)
    1. Stop Solution: Ready to use, contains a 1N H2SO4 solution. (96T: one bottle, 15 mL)
    1. Package Insert

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      Gull Rubeola IgM ELISA kit
    1. Predicate 510(k) number(s):
      K904083

{3}------------------------------------------------

3. Comparison with predicate:

Similarities
ItemDevicePredicate
Intended UseThe Trinity Biotech Captia™Measles IgM Enzyme-LinkedImmunosorbent Assay(ELISA) is intended for thequalitative detection ofMeasles IgM antibodies inhuman serum of patientssuspected of measles(rubeola) infection. Thisassay is intended for use asan aid in the diagnosis of acurrent or recent measles(rubeola) infection inconjunction with otherclinical information andlaboratory findings.Performance characteristicshave not been evaluated inimmunocompromised orimmunosuppressedindividuals. This test is notintended for use in neonatalscreening or for use at apoint of care. This test is notintended for use inscreening blood and plasmadonors.The Gull Rubeola igMELISA test is intended forthe qualitative detection ofIgM antibody to rubeola(measles) virus in humanserum. When performedaccording to theinstructions, the RubeolaIgM ELISA test is of valuein the determination ofrecent exposure to andinfection with rubeola virusand in the diagnosis ofrubeola virus-associateddisease.
TechnologyEnzyme-LinkedImmunosorbent Assay(ELISA)Same
Specimen TypesHuman SerumSame
1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1Differences
ItemDevicePredicate

4

:

·

{4}------------------------------------------------

K. Standard/Guidance Document Referenced (if applicable):

  • CLSI EP7-A2 Interference in Clinical Chemistry .
  • . CLSI H18-A4 Procedures for Handling & Processing Blood Specimens
  • CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement ● Methods
  • CLSI EP12-A2 User Protocol for Evaluation of Qualitative Test Performance .

L. Test Principle:

The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assays (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H2SQ4, The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. High Dose Hook Effect:

When antigen is evaluated for purchase, it is titrated to ensure there is no dose hook effect (prozone) of antigen on the plate. If prozoning is apparent then the antigen is titrated out further past the prozone point.

Due to the nature of obtaining IgM positive serum in general, positive serum purchased for the Captia Measles IgM generally contains less antibody than would be present to demonstrate dose hook effect (prozone), although the positive control in each kit lot is titrated to negative at each kit QC test stage to ensure the positive control is not prozoning.

b. IgM/IgG Competition:

Specific IgG may compete with the IgM for sites and may result in a false negative. Conversely, rheumatoid factor in the presence of specific IgG may result in a false positive reaction. The Serum Diluent Plus Solution diminishes competing virusspecific IgG and minimizes rheumatoid factor interference in samples. Studies indicate that the maximum amount of IgG which can be removed by the kit Serum Diluent Plus is in excess of the expected high end of the normal range for IgG > 1380 mg/dL. The highest titer of RF+ tested (1: 1280; 500 IU/mL) did not adversely affect the performance of the assay.

{5}------------------------------------------------

Precision/Reproducibility: C.

The Trinity Biotech Captia™ Measles IgM ELISA Test Kit was evaluated for interassay precision. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: Four (4) low to moderate Positive samples and two (2) Negative samples run in triplicate at three (3) separate facilities. Testing at each site was done over 10 days by two (2) operators at each site resulting in a total number of 180 data points. The results are summarized below:

SampleIDISRSite 1N=60ISRSite 2N=60ISRSite 3N=60MEANSTDDEV% CV
11.301.341.231.290.0556784.3%
20.570.680.670.640.0608289.5%
31.161.261.071.160.0950448.2%
41.921.841.761.840.080004.3%
50.670.820.840.780.09291612.0%
61.281.181.011.160.13650411.8%

Precision Between All Sites.

The Trinity Biotech Captia™ Measles IgM ELISA Test Kit was evaluated for lot to lot precision. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: four (4) low to Moderate Positive samples and two (2) Negative samples run in triplicate on three (3) separately manufactured kit lots. The results are summarized below:

SampleIDISR'sLot 1ISR'sLot 2ISR'sLot 3MeanStd Dev% CV
11.221.431.361.340.1207739.02
20.570.420.460.480.10897222.55
31.191.311.241.250.1032538.27
42.131.961.561.880.2792414.83
50.790.750.560.70.10954515.72
61.271.321.31.290.0412653.19

Inter-lot Reproducibility

A separate precision/reproducibility study was performed internally (Trinity Biotech, Jamestown) using one (1) lot of the Captia™ Measles IgM ELISA Test Kit. The study compared the consistency between two (2) Operators. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: Four (4) Low to

{6}------------------------------------------------

moderate Positive samples, and two (2) Negative samples run by each of the Operators over 10 days. The results are summarized below:

SampleIDISR'sOperator1 N=30ISR'sOperator2 N=30MEANISRSTDDEV% CV
11.261.331.30.04953.80%
20.580.550.570.021213.80%
31.141.171.160.021211.80%
41.921.911.920.007070.40%
50.690.650.670.028284.20%
61.321.231.280.063645.00%

Intra-run Precision/Reproducibility Between users: Total Precision at Site 1 with Two Operators Over Ten Days

  • d. Linearity/assay reportable range:

Not Applicable

  • Traceability, Stability, Expected values (controls, calibrators, or methods): ૯.
    Traceability - The Controls and Calibrators are traceable against frozen internal Quality Control standard panels consist of patient samples whose values span the assay range.

Stability - Real time stability studies support the shelf life of the Reagents. Controls and Calibrators at 2 - 8° C up to the stated expiration date on the labeling.

Expected Values - The ISR (Immune Status Ratio) Values for the Positive and Negative Controls should be in their respective ranges printed on the vial labels. If the Control values are not within their respective ranges, the test should be considered invalid and should be repeated.

  • f. Detection limit:
    Not Applicable

  • g. Analytical specificity:
    Cross Reactivity

47 samples were used for establishing the potential cross reactivity of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. The samples were selected as confirmed positive by Trinity Biotech Captia™ CMV IgM, HSV1 IgM, HSV2 IgM, Rubella IgM, VZV IgM and

{7}------------------------------------------------

Mumps IgM test kits, for each of the following potentially cross-reacting agents. CMV, HSV-1. HSV-2. Rubella. VZV. Mumps and Parvo-B19. Fourteen (14) of the suspected cross reactive samples tested as measles IgM antibody tested positive or equivocal (12 positive and 2 equivocal) with the Trinity Biotech Captia™ Measles IgM ELISA test kit. These 14 samples were then run on a comparator Measles IgM IFA test kit and a comparator Measles IgM ELISA test kit. A sample was determined as measles IgM antibody positive if the results on two (2) measles IgM tests were positive. A sample was determined as measles IgM antibody negative if the results on two (2) measles IgM tests were negative. The results are presented in the table below:

CrossReactingAgentsNumberofSamplesTrinity Biotech Captia™Measles IgMResultConsensus ComparatorResult*
NegPosEquivNegPos
Mumps IgM4400NTNT
VZV IgM4400NTNT
Rubella IgM4400NTNT
CMV IgM733131
HSV1 IgM1697052
HSV 2 IgM1072121

NT = Not tested

  • Tested only if Captia™ Measles IgM result was positive.

Potential cross reactivity was observed with CMV IgM, HSV1 and HSV2 IgM. Potential cross-reactivity with Parvovirus, Respiratory Syncytial Virus (RSV) and parainfluenza has not been ruled out, as either specimens were not tested or a limited number were tested.

Interfering Substances

Interfering substance testing was conducted using three (3) characterized samples for measles IgM, spiked with recommended interfering substances according to the quidance in the CLSI document EP7-A2 (Interference Testing in Clinical Chemistry: Approved Guideline 2nd Ed). Results from the three (3) samples tested as spiked with the interfering substance and unspiked (as a control) were favourable. No adverse results or significant change in results occurred with these samples tested with the following:

Potential InterferentConcentration
Hemoglobin$\ge$ 500 mg/dL
Bilirubin (unconjugated)$\ge$ 20 mg/dL

Interference with substances such as red cells, cholesterol, gamma globulin, triglycerides, beta carotene, protein, ascorbic acid and anticoagulants have not been tested.

{8}------------------------------------------------

h. Assay cut-off:

The obiective of this study was to demonstrate the performance of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit; specifically to generate data to demonstrate determination of the cut-off. The results are summarized below:

228 Measles IgM negative sera were assayed on the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. Four (4) samples were disqualified as they were in the equivocal range on Trinity IgM ELISA or indeterminate on a comparator Measles IFA test kit (due to non-specific staining most likely due presences of ANA antibodies). The mean and standard deviation of the optical density readings for the sera were 0.105 and 0.070 respectively. The positive threshold for the assay was determined by adding the mean and 4 standard deviations (0.105 + 4(0.070) = 0.385). Positive sera were titrated to give a constant ratio of the threshold value to obtain a Cut-off Calibrator serum. On all subsequent assays, this serum was run and the assay was calibrated by multiplying the OD value for the calibrator by the ratio to the cut-off calibrator value. This value was then divided into the O.D. for the patient sera to obtain an Immune Status Ratio (ISR). By definition the Cut-off ISR is equal to 1.00. To account for inherent variation in immunoassay, values of 0.91 to 1.09 were considered equivocal. Therefore ≤ 0.90 are considered negative and values ≥ 1.10 are considered positive. values

i. Seroconversion study:

The BIOMEX Rubeola IgG and IgM Antibody Seroconversion Panel (Cat No. SCP-MEA-001) contained 15 samples depicting the onset and decline of IgG and IgM antibodies to Rubeloa (Measles) from one individual over a period of 55 days. The 15 member panel demonstrated an IgM response consistent with an antibody response for seroconversion when tested on four (4) different lots of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit.

The seroconversion panel consisted of one patient with 15 draws during an approximately two month period from May 315 to July 250", 1994. Seroconversion was demonstrated at Day 20 on the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. and remained positive until Day 22Comparator 1 demonstrated seroconversion at Dav 22 and was positive only at that bleed date. Comparator 2 sero-conversion happened earlier than the Trinity kit by 2 days and remained positive for the remainder of timeframe. The results are summarized below:

{9}------------------------------------------------

ComparatorComparator
Trinity Biotech Captia™ Measles IgM ELISA12
LotLotLot
Lot 0272829Lot 030Rubeola IgMRubeola IgM
BleedDayISRResultISRResultISRResultISRResultS/COResultS/COResult
Day10.22Neg0.29Neg0.2Neg0.12NegNTNT0.429Neg
Day 60.29Neg0.33Neg0.24Neg0.15Neg0.02Neg0.449Neg
Day 80.25Neg0.31Neg0.21Neg0.14Neg-0.04Neg0.389Neg
Day 130.26Neg0.34Neg0.27Neg0.21Neg0.28Neg1.101Pos
Day 150.44Neg0.51Neg0.54Neg0.43Neg0.46Neg3.045Pos
Day 201.53Pos1.71Pos1.77Pos1.28Pos0.92Neg7.332Pos
Day 221.36Pos1.4Pos1.49Pos1.24Pos1.09Pos5.919Pos
Day 270.84Neg0.92equiv0.96equiv0.71Neg0.5Neg3.785Pos
Day 290.59Neg0.8Neg0.89Neg0.54Neg0.33Neg3.296Pos
Day 350.54Neg0.63Neg0.62Neg0.47Neg0.61Neg2.449Pos
Day 370.44Neg0.53Neg0.53Neg0.79Neg0.24Neg2.113Pos
Day 430.44Neg0.6Neg0.54Neg0.38Neg0.17Neg2.04Pos
Day 450.37Neg0.5Neg0.5Neg0.31Neg0.15Neg1.802Pos
Day 490.34Neg0.59Neg0.39Neg0.28Neg0.04Neg1.471Pos
Day 540.38Neg0.5Neg0.42Neg0.29Neg0.27Neg1.47Pos

j. Specificity in a Normal Population

A total of 200 random serum samples were tested to establish the expected values in a population of individuals between the ages 18-65 with no known clinically apparent Measles infection. The table below summarizes the distribution of Trinity Biotech Measles IgM assay ISR Values observed for the population.

Distribution of Trinity Biotech Captia™ Measles IgM Assay ISR Values from a Normal Population

Measles IgMISR RangeNumber ofSpecimens(Frequency)Percent ofTotal
0.00-0.206532.5%
0.21-0.409145.5%
0.41-0.602914.5%
0.61-0.8084.0%
0.81-0.9042.0%
0.91-1.1010.5%
1.11-1.2000.0%
1.21-1.4010.5%
>2.0010.5%

1 0

{10}------------------------------------------------

Image /page/10/Figure/0 description: This bar graph shows the specificity in a normal population of 200 people. The x-axis shows the ISR values, and the y-axis shows the frequency. The bar graph shows that the highest frequency is in the 0.21-0.40 range, with a frequency of approximately 90. The 0.00-0.20 range has a frequency of approximately 65.

Comparison studies: 2.

  • a. Method comparison with predicate device:
    See Clinical Studies.

  • b. Matrix comparison:
    Not Applicable. The kit is for use with Human Serum only.

    1. Clinical studies:
    • a. Clinical Sensitivity:

Not Applicable

  • b. Clinical specificity:
    Not Applicable

  • c. Other clinical supportive data (when a. and b. are not applicable):
    Three (3) studies were conducted to evaluate the performance of the Trinity Biotech Captia™ Measles IgM ELISA test kit.

{11}------------------------------------------------

Studies 1 and 2

188 samples were collected from individuals in whom a measles IgM test was ordered (suspected of measles infection in outbreak settings by the Texas Department of Health, Austin, Texas (66 samples) and the State Laboratory of Public Health in Raleigh, North Carolina (122 samples)). All serum samples were tested on the Trinity Biotech Captia™ Measles IgM ELISA test kit. All but one (1) sample was tested by a comparator testing algorithm used at each institution to determine the presence of current or recent measles infection, referred to as "comparator algorithm" in the performance tables. The algorithms used consisted of the results of other laboratory confirmation methods including comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutionation Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit. See Table below for results.

Comparator Algorithm

+Indeterminate
+886
Equivocalನಿ10
70
# AgreeTotalResult95% CI
88તેર્ભ91.67%84.2-96.3
708681.40%71.6-89.0

Percent of Captia M Measles IgM Results that are Equivocal = 8.6% (16 of 187)

Performance with the indeterminate results from the comparator algorithm calculated against the performance of the Captia™ Measles IgM:

Comparator Algorithm

+Indeterminate-
+8816
Trinity BiotechCaptia™ Measles IgMEquivocal3310
-5170
# AgreeTotalResult95% CI
% Positive Agreement889790.7%83.3-95.0%
% Negative Agreement708780.5%70.9-87.4%

Percent of Captia 100 Measles IgM Results that are Equivocal = 8.6% (16 of 187)

{12}------------------------------------------------

Study 3

Eight (8) samples were submitted for testing at the Kansas Department of Health and Environment Laboratory (KDHE), Topeka, Kansas. All samples were tested on the Captia™ Measles IgM and Captia™ Measles IgG ELISA test kits as well ascompared to the method used by the testing institution to determine the presence of current or recent measles infection, referred to as "comparator algorithm" in the performance table. The comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples. The final % agreement is presented in the table below:

Comparator Algorithm
+Indeterminate-
Trinity Biotech CaptiaTMMeasles IgM+300
Equivocal000
-01*4
# AgreeTotalResult95% CI
% Positive Agreement33100.00%29.2-100.0%
% Negative Agreement44100.00%39.8-100.0%

This sample tested NPS negative at two sites, urine positive at one site and urine . indeterminate at one site (all testing done by PCR).

Performance with the indeterminate results from the comparator algorithm calculated against the performance of the Captia™ Measles IgM:

Comparator Algorithm
+Indeterminate-
Trinity Biotech Captia™Measles IgM+300
Equivocal000
-01*4
# AgreeTotalResult95% CI
% Positive Agreement3475.00%30.1-95.4%
% Negative Agreement44100.00%39.8-100.0%
    1. Clinical cut-off:
      Refer to Assay Cut-off above.

{13}------------------------------------------------

5. Expected values/Reference range:

IgM antibodies appear in the first week of infection and usually peak within a month. Although uncommon, low levels of IgM may persist for one (1) year or longer. The annual incidence rates are reported to vary in different geographical areas.

Recent reports (2004-2011) from the Centers for Disease Control and Prevention (CDC) indicate that annual measles incidence is <1 reported case per 1 million population. A large proportion of cases (40%) are international imported or associated with importation. Unknown source cases are not linked to any endemic chain (12%). Most imported cases do not lead to spread in the United States. This is attributed to high levels of vaccination coverage (CDC Report: Documentation and Verification of Measles, Rubella and Congenital Rubella Syndrome Elimination in the Region of the Americas, United States National Report, CDC March 28, 2012).

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three overlapping shapes, resembling an abstract human figure or a symbol representing health and well-being.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 22, 2014

TRINITY BIOTECH BONNIE B. DEJOY 2823 GIRTS RD JAMESTOWN, NY 14701

Re: K140455

Trade/Device Name: CaptiaTM Measles IgM Regulation Number: 21 CFR 866.3520 Regulation Name: Rubeola (measles) virus serological reagents Regulatory Class: I Product Code: PCL Dated: February 21, 2014 Received: February 24, 2014

Dear Ms. DeJoy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{15}------------------------------------------------

Page 2-Ms. DeJoy

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.ida.gov/McdicalDevices/Salety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Stephen J. Lovell -S for

Sally A. Hojvat M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K140455

Device Name Captin'M Meastes IgM ELISA

Indications for Use (Describe)

The Trinty Bioloch Captia™ Meastes Inked Immunosorbent Assay (ELISA) Is Intended for the qualitative desstion of Meastes IgM antibodies in human serum of patients suspected of measies (rubeola) infection. This assay is intended for use as an aid in the diagness of a current or recent measing (rubola) infection with other clinical information and leboratory findings.

Performance characteristics have not been evaluated in immunosuppressed individuals. This test is not intended for use in neonals screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Stephen J. Lovell -S 2014.05.22 12:18:54 -04'00'

FORM FDA 3881 (1/14)

§ 866.3520 Rubeola (measles) virus serological reagents.

(a)
Identification. Rubeola (measles) virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubeola virus in serum. The identification aids in the diagnosis of measles and provides epidemiological information on the disease. Measles is an acute, highly infectious disease of the respiratory and reticuloendothelial tissues, particularly in children, characterized by a confluent and blotchy rash.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.