510(k) Data Aggregation

    K Number
    K251160

    Validate with FDA (Live)

    Date Cleared
    2026-01-08

    (268 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sterile temperature probe is used with Mindray BeneVision N17 patient monitor to monitor body or skin surface temperature. The device is for use by qualified healthcare personnel.

    The probe is offered in the following three configurations:

    • Body cavity temperature probe for monitoring of the core temperature in adult and pediatric patients by insertion into the esophageal; and for monitoring of the body temperature in adult and pediatric patients by insertion into the rectal or nasopharyngeal cavity.
    • Skin contact temperature probe for monitoring of skin temperature to an adult and pediatric patient’s skin surface.
    • Ear cavity temperature probe for monitoring of body temperature by insertion of the foam into pediatric or adult’s auditory canal.

    For pediatric population, the subpopulations are children and adolescents.

    Device Description

    The sterile temperature probe is a compatible sensor for use with Mindary BeneVision N17 patient monitor, as an accessory of the legally marketed patient monitor on the US market, the sterile temperature probe is indicated for continuous monitoring of body temperature in hospital settings. It mainly consists of plug, cable, and probe at patient side. When in use, the probe should be connected to the compatible device, and the temperature change of the patient's measured part is sensed through the built-in NTC thermistor at the probe end and the patient's body temperature is measured. The subject devices are used with Mindray BeneVision N17 patient monitor, which was cleared under K182075.

    The temperature probes are packed individually into a paper plastic pouch in sterile condition and are single use only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K220448

    Validate with FDA (Live)

    Date Cleared
    2023-01-19

    (337 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Non-invasive EEG Sensor is applied directly to the patient's skin to enable recordings of electrophysiological (such as EEG) signals.

    Device Description

    Disposable Non-invasive EEG Sensor is a sensor assembly with pre-gelled electroencephalogram (EEG) electrodes. The Sensor is applied to the skin of the patient to record electrophysiological (such as EEG) signals. It is a low impedance, single patient use, non-sterile disposable electrode sensor that is designed for application to areas including forehead, temporal, above eyebrow and mastoid process. The Sensor is designed to provide ease of use and electrode placement accuracy. It is used in conjunction with a monitor with an EEG module.

    The sensor design is constructed using FPC (Flexible Printed Circuit Board) and medical foam. Place the foam plate and conductive gel on the FPC board. The bottom of the foam is covered with medical grade pressure sensitive adhesive for skin adhesion. A printed circuit board, which is part of the FPC board and printed with conductive silver chloride, is pasted on the other side of the foam. The flexible foam wrapped with conductive gel is used to contact the skin. When the flexible foam contacts the skin, the conductive gel will connect the skin with the FPC board to form a conductive bridge. The silver/silver chloride circuit provides signal continuity from each electrode to the monitor. The FPC board is attached to site like forehead and transmits EEG signals.

    AI/ML Overview

    The company, Shenzhen Med-link Electronics Tech Co., Ltd., is seeking 510(k) clearance for their Disposable Non-invasive EEG Sensor. The device is a non-sterile, single-patient use electrode sensor designed to record electrophysiological (EEG) signals. The indications for use state that the sensor is "applied directly to the patient's skin to enable recordings of electrophysiological (such as EEG) signals."

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / Performance AspectPredicate Device StandardsProposed Device Performance
    Electrical SafetyAdherence to 60601-1 standards (GE Entropy Sensor)Complies with ANSI/AAMI ES60601-1
    PerformanceComplies with ANSI/AAMI EC12Complies with ANSI/AAMI EC12
    BiocompatibilityAdherence to ISO10993-5, ISO 10993-10 (GE Entropy Sensor)All patient-contacting materials evaluated by ISO10993-5, ISO 10993-10
    Package Integrity & Functional PerformanceNot explicitly stated for predicatesCompleted on subject device following Accelerated aging test to support proposed shelf life.

    2. Sample size used for the test set and the data provenance:

    The document mentions "bench testing," "electrical safety and performance tests," "biocompatibility tests," and "package integrity and functional performance testing." However, specific sample sizes for these tests are not provided in the summary. The provenance of the data (country of origin, retrospective/prospective) is also not explicitly stated beyond the manufacturer being in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to the data provided. The studies described are bench tests and engineering performance evaluations of a physical medical device (EEG sensor), not studies that require expert-established ground truth for diagnostic accuracy or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable as the described tests are technical performance evaluations, not studies involving human interpretation or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is an EEG sensor, a hardware component for recording signals, not an AI-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This is not applicable. The device is a sensor, not an algorithm. The performance tests are focused on the hardware's ability to accurately record signals and meet safety standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the tests performed is based on established industry standards and regulatory requirements for electrical safety, performance, biocompatibility, and material properties. For example, ANSI/AAMI ES60601-1 for electrical safety, ANSI/AAMI EC12 for EEG electrode performance, and ISO10993 series for biocompatibility.

    8. The sample size for the training set:

    This is not applicable. The device is a hardware sensor, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220447

    Validate with FDA (Live)

    Date Cleared
    2022-06-28

    (132 days)

    Product Code
    Regulation Number
    870.2900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable ECG Cable and Leadwires are intended to be used with ECG monitoring devices to measure a patient's ECG for general monitoring and/or diagnostic evaluation by heath care professional.

    Disposable ECG Cable and Leadwires meets the requirements of the standard ANSI/AAMI EC53, ECG TRUNK CABLES and PATIENT LEADWIRES. This leadwire set is intended for single-patient use.

    Device Description

    Disposable ECG Cable and Leadwires with specific various lengths are the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors.

    These devices consist of connectors on each cable end and a shielded bulk cables are used to transfer the signals from the electrodes to the patient monitor.

    These devices use the same type of constructions and have the same technological characteristics as the predicate devices. They use a medical grade PVC and TPU cable jacket with medical grade PVC and ABS over molded connectors with integral relief.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the "Med-link Disposable ECG Cable and Leadwires," but it does not contain information about a study proving the device meets acceptance criteria in the way you've outlined.

    The document is a 510(k) premarket notification summary, which aims to demonstrate "substantial equivalence" to a legally marketed predicate device, rather than proving performance against specific acceptance criteria through clinical or detailed standalone studies.

    Here's a breakdown of why this document doesn't fit your request and what information it does provide:

    • Type of Device: The device is a Disposable ECG Cable and Leadwires, which is an accessory to ECG monitoring devices, not a diagnostic algorithm or AI system. Therefore, the concepts of "sensitivity," "specificity," "AI improvement," "training sets," "test sets," "ground truth," or "expert adjudication" as typically applied to AI/ML medical devices are not relevant here.
    • Purpose of 510(k): The primary goal of a 510(k) submission is to show that a new device is "substantially equivalent" to a predicate device already on the market, meaning it has the same intended use and technological characteristics, or if there are differences, those differences do not raise new questions of safety and effectiveness.

    Given this context, I cannot fill out your requested table and information points as they pertain to a different type of device evaluation (e.g., for an AI/ML diagnostic software).

    However, I can extract the information that is relevant to the document provided:

    Summary of Device Evaluation (as per the 510(k) Summary):

    Acceptance Criteria and Device Performance (based on regulatory standards rather than clinical endpoints):

    Acceptance Criteria (Standard / Test)Reported Device Performance
    Electrical Safety: IEC 60601-1 (General requirements for basic safety and essential performance)Complied with IEC 60601-1. The proposed device was tested according to IEC 60601-1. The safety performances are demonstrated by the Third party, CTI through testing following IEC 60601-1.
    ECG Cable and Leadwires Performance: ANSI/AAMI EC53:2013 (ECG Trunk Cables And Patient Leadwires)Meets the requirements of the standard ANSI/AAMI EC53. The proposed device was tested according to EC53. The safety performances are demonstrated by the Third party, CTI through testing following ANSI AAMI EC53. The results of bench testing & type testing provide reasonable assurance that the proposed device has been designed and validated to assure conformance to the requirements for its indication for use.
    Biocompatibility: ISO 10993-1, ISO 10993-5, ISO 10993-10 (Biological evaluation of medical devices) (Specifically for material difference: ISO 10993-10 for irritation and skin sensitization)Complied with ISO 10993-1, ISO 10993-5, ISO 10993-10. For the material difference, the proposed device was specifically tested according to ISO 10993-10. The biocompatibility performance equivalence evidence can be demonstrated.
    Shelf Life / Package Integrity: (Functional performance testing following real-time aging test)Package integrity and functional performance testing were completed on the subject device following real time aging test to support the proposed shelf life.

    Here's why the other points you requested are not available in this document:

    1. Sample size used for the test set and data provenance: Not applicable. This device is an accessory (cable/leadwire) and its performance is evaluated against engineering/safety standards (IEC, AAMI, ISO), not clinical data sets with "test sets" or "data provenance" in the context of an algorithm. The testing involves bench and type testing of the physical device according to standards.
    2. Number of experts used to establish the ground truth... and their qualifications: Not applicable. There is no "ground truth" to establish through experts for an ECG cable's performance in the way a diagnostic algorithm requires. Compliance to standards is verified through laboratory testing.
    3. Adjudication method: Not applicable. No human adjudication is mentioned or relevant for electrical and biocompatibility testing of a cable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This is not an AI/ML diagnostic device, so MRMC studies are irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    6. The type of ground truth used: Not applicable. The "ground truth" here is compliance with established engineering and safety standards, verified through bench testing.
    7. The sample size for the training set: Not applicable. There is no "training set" for an ECG cable.
    8. How the ground truth for the training set was established: Not applicable.

    In conclusion, the provided document details the "substantial equivalence" of an ECG cable to a predicate device by demonstrating compliance with relevant international and national standards for electrical safety, performance, and biocompatibility through non-clinical (bench and type) testing, not through a clinical study with acceptance criteria for diagnostic accuracy metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212752

    Validate with FDA (Live)

    Date Cleared
    2022-05-20

    (263 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K193338

    Validate with FDA (Live)

    Date Cleared
    2020-08-25

    (267 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Med-link Reusable Temperature Probes:

    Med-link Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitors of Philips Model IntelliVueMP50, Mindray Model PM-9000 and Drager Model Infinity Gamma XL.

    These devices are used by qualified medical professional only.

    Med-link Disposable Temperature Probes:

    Med-link Disposable Temperature Probes are intended to be used for monitoring temperature for single patient use. The temperature probes are non-sterile and designed for use with monitors of GE Model B20 and Philips Model IntelliVueMP50.

    These devices are used by qualified medical professional only.

    Device Description

    The proposed devices are used for patient temperature measurement. The probes are reusable depending on models. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The proposed devices are designed to be used in healthcare facilities like hospital.

    AI/ML Overview

    The provided text describes the 510(k) summary for Med-link Reusable and Disposable Temperature Probes. While it outlines the device's characteristics and comparison to predicate devices, it does not contain information about acceptance criteria or a study that proves the device meets specific performance acceptance criteria in the context of AI/ML algorithm evaluation.

    The document states:

    • "No clinical studies were performed to demonstrate substantial equivalence." (Section 10)
    • The non-clinical tests focused on safety, essential performance, and biocompatibility in accordance with relevant IEC and ISO standards for medical electrical equipment and biocompatibility. The performance standard cited is ISO 80601-2-56 for body temperature measurement.

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI/ML algorithm's performance, as this document is for a traditional medical device (temperature probes) and does not involve AI/ML.

    However, I can extract the reported device performance and testing information that is present in the document for the temperature probes themselves, which is related to the accuracy of the temperature measurement.


    Reported Device Performance (for the temperature probes, not an AI/ML algorithm):

    Acceptance Criteria / Performance MetricReported Device Performance
    Measurement Range25-45°C
    Accuracy±0.1°C
    BiocompatibilityComplies with ISO 10993-5 (in vitro cytotoxicity) and ISO 10993-10 (irritation and skin sensitization)
    Electrical SafetyComplies with IEC 60601-1
    Performance StandardComplies with ISO 80601-2-56
    SterilizationNon-sterile
    Operational TypeContinual

    Regarding the other requested information for AI/ML algorithms:

    1. Sample size used for the test set and the data provenance: Not applicable, as this is not an AI/ML algorithm. Non-clinical tests were performed to assess safety and effectiveness against standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Device calibration and accuracy testing typically refer to established laboratory methods and reference standards, rather than expert interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device itself is standalone in its function as a temperature probe; there is no AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For temperature accuracy, the ground truth would typically be established by calibrated reference thermometers in a controlled environment, adhering to the requirements of the relevant performance standards (e.g., ISO 80601-2-56).
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182667

    Validate with FDA (Live)

    Date Cleared
    2019-06-18

    (266 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Med-link Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitors of Philips Model Intelli Vue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL.

    These devices are used by qualified medical professional only.

    Device Description

    The proposed devices are used for patient temperature measurement for multi-patient use. The temperature probes are reusable and consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement.

    AI/ML Overview

    The Med-link Temperature Probes are intended for monitoring temperature for multi-patient use. They are reusable and designed for use with specific monitors (Philips Model IntelliVue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL).

    Here's an analysis of the acceptance criteria and supporting study information:

    1. Acceptance Criteria and Reported Device Performance

    The core performance acceptance criterion for the Med-link Temperature Probes appears to be accuracy.

    Acceptance CriteriaReported Device Performance
    Accuracy: ±0.1 °CAccuracy: ±0.1 °C
    Measurement Range: 25-45°CMeasurement Range: 25-45°C

    Other relevant criteria and performance:

    • Type of Use: Prescription Use
    • Operating Principle: Resistance of thermistor based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement.
    • Measurement Site: Skin, Esophageal, and Rectal
    • Component: Reusable
    • Thermistor Resistance: 2.25K@25°C
    • Sterilization: Non-sterile
    • Electrical Safety: Comply with IEC 60601-1
    • Performance Standard Compliance: ISO 80601-2-56

    2. Sample Size and Data Provenance for Test Set

    The document does not explicitly state the specific sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for any clinical studies.

    The document mentions "performance bench testing in accordance with ISO 80601-2-56 for all models of the subject device with compatible monitors was conducted and the results met the requirements." This indicates that the testing was likely laboratory-based bench testing rather than clinical trials with human subjects.

    3. Number of Experts and their Qualifications for Ground Truth

    The document does not indicate the involvement of experts in establishing ground truth for a test set, as the testing described is primarily bench testing against recognized standards rather than clinical validation requiring expert interpretation or diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable, as the document describes bench testing against technical standards, not a study requiring adjudication of expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on device equivalence based on technical standards and performance specifications, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study in the form of bench testing was conducted. The document states: "performance bench testing in accordance with ISO 80601-2-56 for all models of the subject device with compatible monitors was conducted and the results met the requirements." This testing evaluates the device's technical performance against a recognized standard without human interpretation in the measurement process.

    7. Type of Ground Truth Used

    The ground truth used was established by recognized international technical standards and specifications. Specifically, the device's performance was compared against the requirements of:

    • ISO 80601-2-56: Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
    • The stated accuracy of ±0.1°C and measurement range of 25-45°C are the "ground truth" against which the device's measured performance was assessed during bench testing.

    8. Sample Size for the Training Set

    The document does not describe any machine learning or AI components that would require a "training set." Therefore, this question is not applicable. The device is a traditional electronic thermometer.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set for machine learning or AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160530

    Validate with FDA (Live)

    Date Cleared
    2016-11-28

    (277 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Med-link Disposable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The proposed device, Med-link Disposable Blood Pressure Cuff is rectangle soft inelastic sleeve. There is a single-tube or twin-tube connected to a sealed gas chamber. It is wrapped around the patient's limb/thigh and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in various sizes for different limb range. It is provided non-sterile.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Med-link Disposable Blood Pressure Cuff. It aims to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on the FDA recognized consensus standard ISO 81060-1:2007/ AAMI/ANSI/ISO81060-1:2007/(R)2013 Non-Invasive Sphygmomanometers -Part 1: Requirements And Test Methods For Non-Automated Measurement Type, and ISO 10993 (Biological evaluation of medical devices) series for biocompatibility.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Standard Reference)Reported Device Performance (Summary)
    PerformanceCompliance with ISO 81060-1:2007/ AAMI/ANSI/ISO81060-1:2007/(R)2013, Part 1: Requirements And Test Methods For Non-Automated Measurement Type"Performance testing was performed on the subject device and results were compared with predicate device. Tests were conducted following applicable procedures outlined in the FDA recognized consensus standard of AAMI/ANSI/ISO81060-1:2007/(R)2013, and results met all relevant requirements in the test standard."
    BiocompatibilityISO 10993-1: Biological evaluation of medical devices-Part 1: Evaluation and testing"A series of safety, essential performance and in vitro biocompatibility tests were performed... The tests listed below were conducted in accordance with ISO 10993-1, ISO 10993-5, ISO 10993-10."
    ISO 10993-5: Test for in vitro cytotoxicity"Cytotoxicity Test" performed. Data not explicitly detailed in summary but implied to meet criteria.
    ISO 10993-10: Tests for Irritation And Skin Sensitization"Skin irritation Test" and "Skin Sensitization Test (the Guinea Pig maximization test)" performed. Data not explicitly detailed in summary but implied to meet criteria.
    Pressure Limit0-300mmHg matching predicate devicesProposed device has a pressure limit of "0-300mmHg," which is identical to the primary and secondary predicate devices.
    SterilityNon-sterile matching predicate devicesProposed device is "Non-sterile," which is identical to the primary and secondary predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the performance testing. It mentions "Performance testing was performed on the subject device," but does not specify the number of devices or measurements.

    The data provenance is not specified regarding country of origin or specific study design (retrospective or prospective). The tests are described as "non-clinical testing," which suggests laboratory-based evaluations rather than human subject trials for the core performance and biocompatibility aspects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes the testing of a basic medical accessory (blood pressure cuff). The "ground truth" for its performance is established by objective measurements against a recognized standard (ISO 81060-1). There is no mention of experts establishing a "ground truth" in the way one might for diagnostic imaging or a subjective clinical assessment. The ground truth for biocompatibility is established through standardized laboratory tests.

    4. Adjudication Method for the Test Set

    Not applicable. The performance and biocompatibility tests are objective measurements against established standards, not subjective assessments requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is a 510(k) submission for a blood pressure cuff, which is a physical accessory, not an AI or imaging-based diagnostic device. Therefore, an MRMC comparative effectiveness study, which typically evaluates human readers' performance with and without AI assistance, is not relevant or applicable here.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a passive accessory and does not involve an algorithm. Its function is to interface with a blood pressure measurement system.

    7. Type of Ground Truth Used

    The ground truth used for performance testing (accuracy and reliability of pressure transmission, cuff inflation/deflation, etc.) is the adherence to the ISO 81060-1 standard's objective requirements and test methods.

    For biocompatibility testing, the ground truth is established by the results of standardized laboratory tests (Cytotoxicity, Skin Irritation, Skin Sensitization) as defined by the ISO 10993 series.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning algorithm and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113727

    Validate with FDA (Live)

    Date Cleared
    2013-03-21

    (458 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shenzhen Med-link disposable probe, model S0136J-L, is indicated for single patient use when continuous, non-invasive arterial oxygen saturation and pulse rate monitoring are required for adult patients or pediatric patients weighing more than 30 kg. The S0136J-L is contraindicated for use with patients during motion condition.

    Device Description

    Shenzhen Med-link Pulse Oximeter Probes are electro-optical probes which function without skin penetration, electrical contact, or heat transfer. The probes use optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the patient monitor or oximeter device. The probe contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The LED and photodiode are contained in silicon rubber pads. These SpO2 probes not made with natural rubber latex.

    AI/ML Overview

    The Shenzhen Med-Link Pulse Oximeter Probe, model S0136J-L, was tested to ensure its performance and safety, primarily focusing on SpO2 accuracy and pulse rate accuracy.

    1. Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (from ISO 9919:2005)Reported Device Performance (Shenzhen Med-Link S0136J-L)
    SpO2 Accuracy (70-100%):SpO2 Accuracy (70-100%):
    70-100%: ±2% (Predicate device standard)70-100%: ±3%
    70-80%: ±2%70-80%: ±3%
    80-90%: ±2%80-90%: ±2%
    90-100%: ±2%90-100%: ±2%
    Pulse Rate Accuracy (20-250 bpm):20 to 250 bpm ±3 digits
    20 to 250 bpm ±3 digits
    Excessive Temperature:
    Does not exceed 41°C at probe-tissue interface at 35°C ambient temperatureDoes not exceed 41°C at probe-tissue interface at 35°C ambient temperature

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "clinical testing to validate the performance and accuracy of the probes under controlled hypoxia versus arterial oxygen saturation as determined by co-oximetry." However, it does not specify the exact sample size of subjects used for this clinical test set.

    The data provenance is prospective, as indicated by the description of the clinical study, which involved "all testing performed under an institutionally approved protocol with subject informed consent." The country of origin of the data is not explicitly stated, but it can be inferred to be from the location where the clinical study was conducted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the number of experts used or their qualifications for establishing the ground truth. The ground truth for SpO2 accuracy was determined by "arterial oxygen saturation as determined by co-oximetry." This implies instrumental measurement as the "ground truth" rather than expert consensus on interpretation.

    4. Adjudication Method for the Test Set:

    The document does not describe an adjudication method for the test set, as the ground truth was established by co-oximetry, which is an objective measurement rather than a subjective interpretation requiring adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed as this device is a pulse oximeter probe, not an AI-assisted diagnostic tool that requires human interpretation of images or data. Therefore, the concept of human reader improvement with AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, a standalone performance assessment was conducted for the device. The clinical study evaluated the performance of the Shenzhen Med-link Probe itself in determining SpO2 accuracy under controlled hypoxia, compared to co-oximetry as the reference standard. This is a direct measure of the device's algorithmic performance.

    7. The Type of Ground Truth Used:

    The ground truth for SpO2 accuracy was arterial oxygen saturation as determined by co-oximetry. This is an objective, instrumental measurement of the true physiological value.

    8. The Sample Size for the Training Set:

    The document does not specify a separate training set or its sample size. For a medical device like a pulse oximeter, the "training" typically refers to the development and calibration of the internal algorithms based on physiological principles and empirical data, rather than a distinct machine learning training set in the conventional sense. The "clinical testing" described serves as the validation or test set against which the device's performance is measured.

    9. How the Ground Truth for the Training Set Was Established:

    As no distinct training set for an AI/ML algorithm is described in the context of this traditional medical device, the concept of establishing ground truth for a training set in that manner is not applicable. The device's internal algorithms would have been developed based on established physiological models of light absorption by oxygenated and deoxygenated hemoglobin, and calibrated using data that correlates pulse oximetry readings with gold standard co-oximetry measurements during the device's development process. Details of this development process are not provided in the summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120010

    Validate with FDA (Live)

    Date Cleared
    2012-01-19

    (16 days)

    Product Code
    Regulation Number
    870.2900
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shenzhen Med-link Cable / lead-wire are intended to be used with ECG, EKG, SpO2 and Invasive Blood Pressure monitoring devices. The Cable / lead-wire are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by heath care professional.

    Device Description

    Med-link Cable / Lead-wire with specific various lengths are the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors. These cables consist of connectors on each cable end and a shielded bulk cable. The cables are used to transfer the signals from the electrodes to the patient monitor. The Med-link cables use the same type of constructions and have the same technological characteristics as the predicate devices. They use a medical grade PVC and TPU cable jacket with medical grade PVC and ABS over molded connectors with integral relief.

    AI/ML Overview

    Here's the analysis of the provided text regarding acceptance criteria and the supporting study for the medical device:

    Device: Cable / lead-wire (ECG, EKG, SpO2 and Invasive Blood Pressure) manufactured by Shenzhen Med-link Electronics Tech Co., Ltd.


    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on demonstrating substantial equivalence to predicate devices through various safety and performance standards. The acceptance criteria are defined by compliance with these standards, and the reported performance confirms this compliance.

    Acceptance Criteria (Standard)Reported Device Performance
    Biocompatibility:
    ISO 10993-5:1999 (Cytotoxicity)Non-toxic
    ISO 10993-10:2002/Amd1:2006 (Irritation & Delayed-Type Hypersensitivity)Non-sensitizing and non-irritating
    Electrical Safety:
    IEC 60601-1:1998; Am1; A2:1995 (General requirements for safety)Complies with applicable clauses
    ECG Cable Performance:
    ANSI/AAMI EC53:1995/(R) 2001 (ECG cables and leadwires)Complies with applicable clauses
    ANSI/AAMI EC53:1995/(R) 2001 (for trunk cable lead wires)Complies with applicable clauses
    DIN 42-802 (Interconnection between patient lead-wires and trunk cable)Complies with requirements
    FDA Regulation:
    FDA 21CFR Part 898 Final ruleComplies
    Bench Testing (General Safety and Performance)Met requirements for safety and performance standards and intended use

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a numerical sample size for the test set. It states that "Med-Link Cable / Lead-wire" were tested according to various standards and "Performance Bench testing was conducted on the Shenzhen Med-link cable / lead-wire according with established protocols." This implies testing of a representative number of cable/lead-wire units to ensure compliance with each standard.
    • Data Provenance: The data is internally generated by the manufacturer (Shenzhen Med-link Electronics Tech Co., Ltd.) through "Performance Bench testing." No information about the country of origin of the data itself (beyond the manufacturer's location in China) or whether it's retrospective or prospective is explicitly stated. However, given it's bench testing for a premarket submission, it would be prospective data generated for the purpose of demonstrating compliance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable to this submission. The "ground truth" here is defined by compliance with established international and national medical device standards (ISO, IEC, ANSI/AAMI, FDA regulations). The testing involves objective measurements against these predefined criteria, not subjective expert judgment that would require establishing a ground truth via consensus.


    4. Adjudication Method for the Test Set

    This section is not applicable. As the testing involves objective measurements against predefined standards, there is no need for an adjudication method by experts. The results are either compliant or non-compliant with the specified technical requirements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices where human readers interpret results, often with and without AI assistance. This submission describes an ECG/EKG/SpO2/IBP cable and lead-wire device, which does not involve human interpretation of diagnostic images or AI assistance in that context.


    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This section is not applicable. The device is a passive accessory (cable/lead-wire) for monitoring devices, not an algorithm or AI system. Therefore, the concept of "standalone (algorithm only)" performance does not apply. The tests performed (biocompatibility, electrical safety, bench testing) assess the device's inherent physical and electrical properties.


    7. The Type of Ground Truth Used

    The "ground truth" used for this device is compliance with established international and national medical device standards and regulations. These include:

    • ISO 10993-5:1999 (Cytotoxicity)
    • ISO 10993-10:2002/Amd1:2006 (Irritation & Delayed-Type Hypersensitivity)
    • IEC 60601-1:1998; Am1; A2:1995 (Medical electrical equipment - General requirements for safety)
    • ANSI/AAMI EC53:1995/(R) 2001 (ECG cables and leadwires)
    • FDA 21CFR Part 898 Final rule
    • DIN 42-802 (for specific interconnection requirements)

    These standards define the objective performance and safety benchmarks that the device must meet.


    8. The Sample Size for the Training Set

    This section is not applicable. The device is a hardware component (cable/lead-wire), not an AI/ML algorithm that requires a "training set."


    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083010

    Validate with FDA (Live)

    Date Cleared
    2009-06-10

    (244 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Shenzhen Med-Linket™ Adult SpO2 Finger Sensor is indicated for use in continuous, non-invasive arterial oxygen saturation and pulse rate monitoring.

    Model S0044D-S (S0012A) is reusable and for patients weighing more than 40 kg.

    The Shenzhen Med-Linket™ Adult SpO2 Finger Sensor is contraindicated for use on active patients, or during conditions of motion or low perfusion.

    Device Description

    Shenzhen Med-Linket™ Adult SpO2 Finger Sensors (Med-Linket Sensors) are compatible sensors for use with major brands of patient monitors and oximeter devices.

    Med-Linket Sensors are electro-optical sensors which function without skin penetration, electrical contact, or heat transfer. The sensors use optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the patient monitor or oximeter device. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The LED and photodiode are contained in silicon rubber pads.

    AI/ML Overview

    Acceptance Criteria and Study for Shenzhen Med-Linket™ Adult SpO2 Finger Sensors

    As per the provided 510(k) summary, the Shenzhen Med-Linket™ Adult SpO2 Finger Sensors (Model S0044D-S) underwent various performance tests to demonstrate substantial equivalence to predicate devices. The primary focus for demonstrating clinical performance was on pulse rate accuracy and SpO2 accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    Pulse Rate AccuracyMeets specifications defined in ISO 9919: 2005, Medical electrical equipment — Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use."Test results indicated that the sensors meet the published specifications for pulse rate accuracy over the specified range." (Specific numerical criteria of ISO 9919:2005 not provided in the document for direct comparison, but compliance is stated).
    Clinical (SpO2 Accuracy)Meets published specifications for accuracy over the specified range (as determined by comparison to co-oximetry under controlled hypoxia)."Test results indicated that the sensors meet the published specifications for accuracy over the specified range." (Specific numerical accuracy criteria not provided, but compliance is stated). This implies that the bias and precision within the specified SpO2 range met pre-defined limits typical for pulse oximeters, such as an Arms (Accuracy Root Mean Square) value below a certain threshold (e.g., 2-3% SpO2).
    BiocompatibilityPatient contact materials are non-toxic, non-sensitizing, and non-irritating, in accordance with ISO 10993-1: 2003."Test results indicated that the patient contact materials were non-toxic, non-sensitizing and non-irritating."
    Electrical SafetyComplies with applicable clauses of IEC 60601-1:1988; Am1: 1991; A2: 1995 and ISO 9919: 2005."Test results indicated that the sensors comply with the applicable clauses of the Standards."
    Electromagnetic CompatibilityComplies with applicable clauses of IEC 60601-1-2:2001."Test results indicated that the sensors comply with the applicable clauses of the Standard."
    CleaningCleaning instructions do not damage sensor labeling, degrade materials, and sensors remain functional after cleaning."Test results indicated that the cleaning instructions do not damage sensor labeling, degrade materials, and that the sensors remain functional after cleaning."
    Excessive TemperatureSensors do not exceed 41°C at the sensor-tissue interface at an ambient temperature of 35°C, in accordance with ISO 9919: 2005."Test results indicated that the sensors do not exceed 41°C at the sensor-tissue interface at an ambient temperature of 35°C."
    EnvironmentalOperative and storage temperatures, and random vibration do not damage sensor functionality (tested in accordance with internal protocols)."Test results indicated that the operative and storage temperatures, and random vibration do not damage sensor functionality."

    2. Sample Size for the Test Set and Data Provenance

    The document does not specify the exact sample size (number of subjects/patients) used for the clinical performance testing related to SpO2 accuracy. However, it states that "All testing was performed under an institutionally approved protocol with subject informed consent." This indicates a prospective study involving human subjects. The country of origin for the clinical data is not explicitly stated.

    For other performance tests like biocompatibility, electrical safety, etc., the "sample size" would refer to the number of devices or material samples tested, which is not detailed in the summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The clinical ground truth for oxygen saturation (SpO2) was established using co-oximetry, which is an objective measurement from arterial blood samples, not through expert interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth for clinical SpO2 accuracy was established by co-oximetry, which is an objective measurement and does not require an adjudication method among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This device is a sensor component for a pulse oximeter, which provides direct physiological measurements (SpO2 and pulse rate), rather than an interpretive system requiring human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the performance testing for SpO2 and pulse rate accuracy evaluates the standalone performance of the sensor in conjunction with a patient monitor/oximeter device. The clinical testing conducted under controlled hypoxia directly assesses the sensor's ability to accurately measure physiological parameters without human interpretation in the loop.

    7. The Type of Ground Truth Used

    For SpO2 accuracy, the ground truth used was arterial oxygen saturation as determined by co-oximetry. This is considered the gold standard for measuring oxygen saturation in blood. The study involved "controlled hypoxia," where subjects' oxygen levels were intentionally lowered to create a range of SpO2 values that the device would be challenged to measure accurately.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware sensor, not a machine learning or AI-driven system that requires a "training set" for algorithm development. Its functionality is based on established electro-optical principles for pulse oximetry.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1