Search Filters

Search Results

Found 205 results

510(k) Data Aggregation

    K Number
    K242962
    Date Cleared
    2025-05-05

    (222 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . |
    | DSA | 21 CFR 870.2900 Patient transducer and electrode cable (including connector). |
    | MSX, DRT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in a hospital environment and during patient transport inside hospitals.

    Not intended for home use. Intended for use by health care professionals.

    Device Description

    The Telemetry Monitor 5500 is a battery-operated patient worn monitor with a touchscreen display. The Telemetry Monitor 5500 is intended to monitor and record, and to generate alarms for, multiple physiological parameters like ECG, SpO2, and respiration rate. It is equipped with a 1.4 GHz radio to enable wireless bi-directional data flow with Philips' Patient Information Center (PIC) iX.

    AI/ML Overview

    This 510(k) clearance letter pertains to the Philips Telemetry Monitor 5500 Release 4.0. The provided document focuses on demonstrating substantial equivalence to a predicate device and includes information regarding performance testing against established standards. However, it does not contain specific details about acceptance criteria for particular performance metrics, nor does it describe studies proving the device meets those specific acceptance criteria in the format often associated with AI/ML device clearances (e.g., sensitivity, specificity, AUC).

    Instead, the document primarily cites compliance with general medical device standards and internal testing to support its claims of performance. Therefore, many of the requested sections below cannot be fully populated as the information is not present in the provided text.


    Acceptance Criteria and Device Performance

    The document does not explicitly state quantitative acceptance criteria for device performance (e.g., a specific minimum sensitivity or accuracy percentage). Instead, it states that the device was assessed for conformity with relevant standards and that "results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards."

    Performance MetricAcceptance Criteria (from document)Reported Device Performance (from document)
    SpO₂ AccuracyRange between ± 1-3 % for measurements between 70-100 % SpO₂ (implied by predicate, identical claim for subject device)Accuracy claims (results) are sensor specific and range between ± 1-3 % for measurements between 70-100 % SpO₂.
    Electrical SafetyConformity with IEC 60601-1 Edition 3.2 2020-08Found to comply
    EMCConformity with ANSI AAMI IEC 60601-1-2:2014 [Including AMD1:2021]Found to comply
    UsabilityConformity with IEC 60601-1-6 Edition 3.2 2020-07 and ANSI AAMI IEC 62366-1:2015+AMD1:2020Found to comply; Summative usability testing has shown that the performance of the subject device does not raise questions of safety and effectiveness relative to the predicate device.
    Alarm SystemsConformity with ANSI AAMI IEC 60601-1-8:2006 and A1:2012 [Including AMD 2:2021]Found to comply
    ECG Monitoring EquipmentConformity with ANSI AAMI IEC 60601-2-27:2011(R) 2016/IEC 60601-2-27:2011-03Found to comply
    Multifunction Patient MonitoringConformity with IEC 80601-2-49Found to comply
    Pulse Oximeter EquipmentConformity with ISO 80601-2-61 Second edition 2017-12Found to comply
    Software Life Cycle ProcessesConformity with ANSI AAMI IEC 62304:2006/A1:2016 and FDA's Guidance "Content of Premarket Submissions for Device Software Functions"Development and testing conducted in accordance with these standards/guidance.

    Study Details:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Test Set Sample Size: Not explicitly stated for any specific performance metric beyond general statements of testing compliance. For the "Clinical Studies" section related to SpO2 accuracy, it only notes that studies were conducted to "support accuracy performance" and "meet the acceptance criteria laid out in the associated protocols." No sample size for patients or data points is provided.
      • Data Provenance: Not specified. The document mentions "Philips conducted clinical studies" but does not detail the country of origin of the data or whether it was retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable. The document describes performance testing against established standards and internal "accuracy specifications". For the SpO2 accuracy claims, ground truth would typically be established by invasive measurements (e.g., CO-oximetry of arterial blood samples), not by expert review of device output. The document does not provide details on how ground truth was established for the clinical studies supporting SpO2 accuracy.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable/Not Specified. Since expert review for ground truth is not indicated, adjudication methods are not relevant in the context described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No. The provided text does not mention any MRMC comparative effectiveness study. This device is a telemetry monitor for physiological parameters, not an AI-assisted diagnostic tool for human readers. "Summative usability testing" was done, indicating human interaction with the device, but not in a comparative effectiveness study involving AI assistance for human "readers."
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Standalone algorithm performance is implied for certain functions, but not detailed. The device uses "Philips proprietary monitoring algorithms for ECG arrhythmia monitoring" (EASI, Hexad, ST/AR Arrhythmia Monitoring, ST/AR ST Analysis Algorithm, ST/AR QT/QTc Interval Monitoring). The performance of these algorithms would constitute "standalone" performance, but the document only states "Functional and system level testing... was performed. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards." No specific performance metrics for these algorithms are provided.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • Not explicitly stated for all performance metrics.
        • For SpO₂ accuracy, ground truth is typically established by comparing the device's SpO₂ readings to invasive arterial blood gas analysis (CO-oximetry). The document only states "Philips conducted clinical studies... to support accuracy performance."
        • For other performance metrics related to compliance with standards (e.g., electrical safety, EMC, usability), the "ground truth" is adherence to the standard's requirements, demonstrated through specific tests.
    7. The sample size for the training set

      • Not Applicable/Not Specified. The document mentions "Philips proprietary monitoring algorithms," which would have been developed using training data. However, the size or nature of any training set is not disclosed in this 510(k) summary, as it's not a primary requirement for demonstrating substantial equivalence for this type of device.
    8. How the ground truth for the training set was established

      • Not Applicable/Not Specified. The document does not provide information on how the ground truth for any potential training sets used in the development of "Philips proprietary monitoring algorithms" was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242986
    Date Cleared
    2025-05-02

    (218 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    102401
    China

    Re: K242986
    Trade/Device Name: SpO2 Extension Cable
    Regulation Number: 21 CFR 870.2900
    Including Connector)
    Classification: II
    Product Code: DSA, DQA
    Regulation Number: 21 CFR 870.2900
    name | SpO2 Extension Cable | SpO2 Extension Cable K222370 | / |
    | Classification Regulation | 21 CFR 870.2900
    | 21 CFR 870.2900 | SAME |
    | Classification | II | II | SAME |
    | Product Code | DSA, DQA | DSA | SAME

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SpO2 Extension Cables are intended to be used to connect SpO2 sensors, placed at appropriate sites on the patient to a monitoring device for SpO2 monitoring by health care professional.

    Device Description

    SpO2 Extension Cable consists of two parts: the SpO2 patient cable and the adapter. The SpO2 patient cable is used for connecting the monitors, and the adapter is used for connecting the SpO2 patient cable and the SpO2 sensor.

    The SpO2 Extension Cable is the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors.

    By using the same types of construction and technological characteristics to the compatible patient monitor and SpO2 sensor, the SpO2 Extension Cable can avoid measured data corrupted.

    The proposed device was modification from the legally marketed (existing) device "SpO2 Extension Cable"/K222370, manufactured by "Beijing Rongrui-Century Science & Technology Co., Ltd.", same as the sponsor of this submission.

    Compared to the initial product (K222370), only the structure of the SpO2 Extension Cable has changed. Originally a non-separable complete extension cable, now an extension cable can be separated into two parts, called the SpO2 patient cable and the adapter.

    The type of instrument connector (which connects one end of the monitoring device) for the SpO2 patient cable is 20pin, there are four types pairing interfaces between SpO2 patient cable and adapter, the pairing interfaces name for the SpO2 patient cable and adapter connection is DB9F-LNCS, MLNCS-MLNCS, LNOP-LNOP, RD-RD and the other end of the adapter connecting to SpO2 sensor is DB9F. Adapters are divided into two categories: with cable and without cable. Adapters for each type of interface are available with and without cables.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for the "SpO2 Extension Cable" is for a Class II medical device, specifically an extension cable used to connect SpO2 sensors to monitoring devices. Based on the content of the letter, the submission is about demonstrating substantial equivalence to a predicate device, not necessarily proving that an AI/ML algorithm meets specific performance criteria.

    Therefore, many of the requested points, such as AI/ML specific acceptance criteria, sample sizes for training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI/ML, are not applicable to this type of device clearance. This clearance focuses on electrical safety, EMC, biocompatibility, and functional compatibility of a hardware component.

    Here's a breakdown of the requested information based on the provided document, highlighting what is applicable and what is not:


    Analysis of Acceptance Criteria and Study for SpO2 Extension Cable (K242986)

    This submission (K242986) pertains to an SpO2 Extension Cable, which is a hardware component. The clearance is based on demonstrating substantial equivalence to a previously cleared predicate device (K222370), rather than establishing performance of an AI/ML algorithm. Therefore, many of the typical acceptance criteria and study details associated with AI/ML devices are not relevant here.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this type of hardware device are primarily based on compliance with recognized consensus standards and functional compatibility. The "performance" here refers to meeting these engineering and safety specifications, rather than clinical accuracy metrics typically seen for diagnostic algorithms.

    Acceptance Criterion (Standard Compliance / Test)Reported Device Performance (Compliance)
    Biocompatibility:
    ISO 10993-5: In Vitro CytotoxicityMeets standard
    ISO 10993-10: Irritation and Skin SensitizationMeets standard
    Electrical Safety & Performance:
    IEC 60601-1 (Edition 3.2): General requirementsComplies
    ISO 80601-2-61: Pulse Oximeter Equipment specificsComplies
    ANSI AAMI EC53: ECG trunk cables/leadwires (likely for general cable integrity/materials, given context)Complies
    Electromagnetic Compatibility (EMC):
    IEC 60601-1-2 (Edition 4.1): EMC requirementsComplies
    Functional Performance:
    Compatibility testing (no measured data corrupt during communication between SpO2 sensors and host monitors)Performance accuracy evaluated and proven

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in terms of a "test set" like in AI/ML validation. Compliance is shown through testing of engineering samples of the device itself and its components against the specified standards.
    • Data Provenance: Not applicable in the context of clinical data for an AI/ML algorithm. The "data" here comes from laboratory testing of the physical cable. The device manufacturer is Beijing Rongrui-Century Science & Technology Co., Ltd. in China. The testing would have been conducted in a controlled lab environment.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not Applicable. Ground truth, in the sense of expert annotation for clinical data, is not established for this type of hardware device. The device's performance is verified against engineering standards and functional specifications.

    4. Adjudication Method for the Test Set

    • Not Applicable. As there's no "ground truth" derived from expert consensus on clinical cases, no adjudication method is relevant.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not Applicable. An MRMC study is designed to assess the impact of an AI algorithm on human reader performance. This device is an extension cable, not an AI algorithm.

    6. Standalone (Algorithm Only) Performance

    • Not Applicable. This is a physical cable, not a software algorithm.

    7. Type of Ground Truth Used

    • Not Applicable. The "ground truth" for this device's acceptance is its adherence to recognized engineering, electrical, biocompatibility, and functional performance standards. This is established through non-clinical testing results, not clinical outcomes data or expert consensus on patient cases.

    8. Sample Size for the Training Set

    • Not Applicable. No training set exists as this is a hardware product, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. No training set or associated ground truth for training.

    Summary of the K242986 Submission's Core Argument:

    The core of the K242986 submission is to demonstrate that the modified structure of the SpO2 Extension Cable (now separable into two parts) does not negatively impact its safety and effectiveness when compared to the predicate device (K222370), which was a non-separable cable. This is achieved by showing continued compliance with relevant performance, safety, and compatibility standards. The primary difference noted is a structural change, and the manufacturer successfully demonstrated that this change "will not affect the safety and effectiveness of the proposed device" through non-clinical testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242798
    Device Name
    Airmod
    Date Cleared
    2025-02-28

    (165 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    : BZQ / 21 CFR
    868.2375
    • Secondary: DQA / 21
    CFR 870.2700
    • Secondary: DSA / 21 CFR
    870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Airmod, when used in conjunction with Accursound Electronic Stethoscope AS-101, is a software as medical device intended to be used for the continuous, non-invasive monitoring of respiratory rate (RR) in adult patients who are subjected to procedural sedation and/or anesthesia.

    Airmod is intended for use by healthcare professionals in hospitals and healthcare facilities who are legally credentialed to perform procedural sedation and/or anesthesia.

    Airmod is intended for Android-based devices only.

    Device Description

    Airmod 114 is an Android-based software application designed to aid healthcare professionals by monitoring a sedated and/or anesthetized patient's breathing in real time. The device has an AIbased algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals of breathing sounds collected by AccurSound Electronic Stethoscope AS-101. Airmodid for the continuous monitoring of respiratory rate (RR) in adults who are subjected to procedural sedation. Airmod™ is designed for use in hospitals and healthcare facilities performing procedural sedation/anesthesia. The device is not intended for patients who are not anesthetized/sedated.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Airmod device, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance (Airmod™)
    Root Mean Square Error (RMSE)
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    5 Leadset, Grabber, Chest, AAMI, OR; M1680A, 6 Lead Set Grabber AAMI, ICU Regulation Number: 21 CFR 870.2900
    Classification Name | Panel & Name: Cardiovascular Devices
    Subpart & Division: 21 CFR §870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Philips IntelliVue reusable ECG lead sets/trunk cables and Philips OR ECG trunk cables are indicated for continuous monitoring of cardiac signals for both diagnostic and monitoring purposes. These devices are limited by the indications for use of the connected monitoring and diagnostic equipment in healthcare facilities. These devices are intended to interact with patient intact skin only.

    Device Description

    Philips Reusable ECG Lead Sets and Trunk Cables are non-invasive reusable medical devices designed to transfer ECG signals from the patient to the monitor used in the healthcare facilities, which includes intensive care unit (ICU) for both adults and neonates and the operating room (OR). The ECG cables and lead sets are used to forward an electrical cardiac signal via electrical wires from the electrode attached on patient skin to the ECG measurement hardware in the patient monitor and to support other outputs as needed. The portfolio consists of two types of trunk cables and two types of lead sets: ICU and OR, including various electrode connectors with AAMI and IEC coloring. The ECG Trunk cables and lead sets are suitable for multiple patient use. The Philips Reusable ECG Lead Sets and Trunk Cables have a service life of 18 months. The Philips Reusable ECG Lead Sets and Trunk Cables can be used with any Philips monitors including Philips monitor/defibrillator product lines and Philips monitoring or diagnostic equipment which claim compatibility to Philips Reusable ECG Lead Sets and Trunk Cables.

    AI/ML Overview

    This document is a 510(k) premarket notification for Philips Reusable ECG Lead Sets and Trunk Cables. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a standalone study with clinical accuracy metrics. Therefore, many of the requested details about acceptance criteria, clinical study design, and ground truth are not applicable or explicitly stated in this type of submission.

    However, I can extract the information that is present regarding performance testing and standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Instead of specific quantitative clinical acceptance criteria and device performance metrics, this submission relies on compliance with recognized harmonized standards. The "performance" is demonstrated through successful completion of tests outlined in these standards.

    Acceptance Criteria (Compliance with Standard)Reported Device Performance
    IEC 60601-1:2005+A1:2012+A2:2020 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    ANSI AAMI EC53:2013/(R)2020 (ECG trunk cables and patient lead wires)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-2-25:2011/(R)2016 (Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-2-27:2011(R)2016 (Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic equipment)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-1-12 Edition 1.1 2020-07 CONSOLIDATED VERSION (Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. (Specifically, ICU Lead Sets and Trunk cables comply with this additional standard).
    IEC 62366-1:2015+AMD1:2020: (Part 1: Application of usability engineering to medical devices)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    Reprocessing Standards: ISO 17664-2:2021, AAMI TIR12 (2020)/(R)2023, and ANSI AAMI ST98:2022, and FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"The subject device adheres to these standards, and its proposed reprocessing instructions are as effective as the predicate device, meeting regulatory requirements for reusable devices.
    Biocompatibility: (ISO 10993-1, as inferred from comparison to predicate)Both the subject and predicate devices meet standards for biocompatibility.
    Electrical Conductivity, Mechanical Strength: (as inferred from comparison to predicate)Both the subject and predicate devices meet standards for electrical conductivity and mechanical strength.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    This document explicitly states: "The subject devices, like the primary predicate devices, did not require clinical trials." Therefore, there is no "test set" in the context of a clinical study for external validation of performance. Performance is primarily demonstrated through compliance with recognized standards via non-clinical bench testing. The data provenance would be from internal lab testing, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
    Not applicable. No clinical test set requiring expert ground truth was used as per the statement above.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
    Not applicable. No clinical test set requiring expert adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an ECG cable and lead set, a hardware accessory. It is not an AI algorithm for interpreting medical images or signals, nor does it involve "human readers" or "AI assistance" in that sense.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
    Not applicable. This is a hardware accessory, not an algorithm. Performance is assessed through its physical and electrical properties, not algorithmic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable in the sense of clinical ground truth. For non-clinical bench testing, the "ground truth" would be established by the specifications and measurement techniques defined within the harmonized standards (e.g., electrical resistance, mechanical durability measured against stated tolerances, biocompatibility testing results).

    8. The sample size for the training set
    Not applicable. This is a hardware accessory, not a machine learning model, so there is no training set.

    9. How the ground truth for the training set was established
    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240087
    Date Cleared
    2024-09-13

    (246 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Branford, Connecticut 06405

    Re: K240087

    Trade/Device Name: RT Carbon ECG Leads Regulation Number: 21 CFR 870.2900
    transducer and electrode cable (including connector) |
    | Classification Regulation: | 21 CFR 870.2900
    tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECC trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT Carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.

    Device Description

    The RT Carbon ECG Leads are intended to connect electrodes placed on the patient to a physiological detection module or monitor (CTM-300, CTM-400, CTM-500, 7000 Series and 3000 Series). The design includes 3 and 4 lead designs with overall lengths of 24", 30" and 36".

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Ivy Biomedical RT Carbon ECG Leads. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing rather than presenting novel clinical efficacy or diagnostic performance data. Therefore, many of the requested categories related to AI performance studies are not applicable.

    Here's the information extracted and categorized as requested:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria alongside reported device performance in the format of specific thresholds and measured values for the RT Carbon ECG Leads. Instead, it states that the device was "thoroughly tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900" and in accordance with EC53. The reported performance is summarized as:

    "The performance data support the safety of the device and hardware verification and validation demonstrate that the RT Carbon ECG Leads should perform as intended in the specified use conditions and is substantially equivalent to the predicate."

    This implies that the device met the acceptance criteria derived from the listed tests.

    Acceptance Criterion (Implicit from tests performed)Reported Device Performance
    Biocompatibility:
    CytotoxicityMet (implies no cytotoxicity)
    SensitizationMet (implies no sensitization)
    IrritationMet (implies no irritation)
    Systemic ToxicityMet (implies no systemic toxicity)
    Pyrogen TestingMet (implies no pyrogenicity)
    Performance Testing (per EC53):
    Patient Leadwire to Trunk Cable InterconnectionMet (implies proper interconnection)
    Flex lifeMet (implies satisfactory durability)
    Tensile strengthMet (implies satisfactory strength)
    Connector mating/unmating cyclesMet (implies satisfactory durability)
    Connector retention forceMet (implies satisfactory retention)
    Contact resistanceMet (implies satisfactory electrical contact)
    Leadwire resistanceMet (implies satisfactory electrical resistance)
    Dielectric withstand voltageMet (implies satisfactory insulation)

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for individual tests conducted on the RT Carbon ECG Leads. The testing was described as "conducted at Ivy" (implying in-house lab testing) and does not involve human data, thus the provenance (country of origin, retrospective/prospective) is not applicable in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/diagnostic device that requires expert ground truth establishment for a test set. The tests are engineering and biological evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/diagnostic device that requires adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an ECG lead, not an AI system or diagnostic tool for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an ECG lead, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The 'ground truth' for the performance testing is based on established engineering standards (EC53) and biocompatibility standards (ISO 10993-1). Compliance with these standards serves as the benchmark against which the device's performance is measured.

    8. The sample size for the training set

    Not applicable. This is a medical device (ECG lead), not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a medical device (ECG lead), not an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240377
    Date Cleared
    2024-05-07

    (90 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Redmond, Washington 98052

    Re: K240377

    Trade/Device Name: LIFEPAK® ECG Cables Regulation Number: 21 CFR 870.2900
    ---|--------------|
    | Cable, Transducer and Electrode, Patient,
    (Including Connector), Class II (870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIFEPAK ECG cables are reusable electrode cable systems intended to transmit signals from patient electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes. LIFEPAK ECG cables are intended for use in outdoor and indoor emerqency care settings.

    LIFEPAK ECG Cables:

    • LIFEPAK 3-wire ECG Cable -
    • LIFEPAK 5-wire ECG Cable i
    • -LIFEPAK 4-wire ECG cable
    • LIFEPAK 3-wire extended precordial ECG Cable l
    • -LIFEPAK 6-wire expandable precordial ECG cable
    Device Description

    The LIFEPAK ECG cables are an accessory designed to be used with Physio-Control LIFEPAK 35 monitor/defibrillator. The LIFEPAK ECG cables are intended for use by Advanced Life Support (ALS) and Basic Life Support (BLS) providers to transmit signals from patient electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes.

    The LIFEPAK ECG Cables are reusable ECG trunk cables and lead wires that connect to the LIFEPAK 35 monitor/defibrillator for ECG monitoring. There are several types or configurations of ECG cables: 3-wire, a 4-wire cable assembly to be used with the LIFEPAK 6-wire precordial cable and/or the LIFEPAK 3-wire precordial cable to perform a 12 or 15-lead ECG Cable assembly. When the LIFEPAK 4-wire ECG cable is configured with a 6-wire expandable precordial cable, a 12-lead ECG is constructed. Likewise, when the LIFEPAK 4-wire ECG cable is configured with a 6-wire expandable precordial cable and a 3-wire extended precordial ECG cable a 15-lead ECG is constructed.

    The 4- wire ECG cable is available in two lengths, 5 feet or 8 feet. Additionally, the 3-, 4-, and 5-wire ECG Cables and the 6-wire precordial ECG cable are available with color coding according to the American Heart Association (AHA) or International Electrotechnical Commission (IEC) standards.

    The LIFEPAK ECG cables are a reusable electrode cable system intended to transmit signals from patient electrodes to the LIFEPAK 35 monitor/defibrillator for both diagnostic and monitoring purposes. These accessories consist of conductor wires that are shielded and insulated by a polyurethane jacket material. The lead wires are color coded according to which electrode location they are attached to on the patient.

    The LIFEPAK ECG cables are not intended for use with other manufacturers' defibrillators. The LIFEPAK ECG cables are intended only for use with LIFEPAK 35 monitor/defibrillator.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for LIFEPAK® ECG Cables. This is a device for transmitting ECG signals, not an AI/ML-based medical device. Therefore, many of the requested criteria (such as MRMC studies, AI/ML-specific ground truth, training set information, etc.) are not applicable to this submission.

    The document focuses on demonstrating substantial equivalence to a predicate device (Multi-Link™ X2 ECG Cable and Leadwire System) primarily through performance testing of the hardware components and comparison of technical characteristics.

    Here's an attempt to answer the questions based only on the provided document, acknowledging that much of it is not relevant for an AI/ML device:


    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests conducted against applicable standards:

    Test PerformedApplicable StandardAcceptance Criteria (Implied)Reported Device Performance
    Connector mating/unmating, Retention force60601-2-25:2011, 60601-2-27:2011, ANSI AAMI EC53:2013/(R)2020Meet standard requirementsMet
    Inspection of Air Clearance60601-2-25:2011, 60601-2-27:2011Meet standard requirementsMet
    Cable Noise TestingANSI AAMI EC53:2013/(R)2020Meet standard requirementsMet
    Flex/Twist Life TestingANSI AAMI EC53:2013/(R)2020Meet standard requirementsMet
    Tensile Strength TestingANSI AAMI EC53:2013/(R)2020Meet standard requirementsMet
    Defibrillation Protection and Energy ReductionIEC 60601-1:2005/AMD2:2020, IEC 60601-2-25:2011, IEC 60601-2-27:2011Meet standard requirementsMet
    Dielectric Withstand TestingIEC 60601-1:2005/AMD2:2020, ANSI AAMI EC53:2013/(R)2020Meet standard requirementsMet
    Leakage Current TestingIEC 60601-1:2005/AMD2:2020Meet standard requirementsMet
    Storage Conditioning and Drop TestIEC 60601-1:2005/AMD2:2020, IEC 60601-2-27:2011Meet standard requirementsMet
    Material Resistance for Cleaning and Disinfection StressIEC 60601-1:2005/AMD2:2020Meet standard requirementsMet
    Electromagnetic CompatibilityIEC 60601-1-2:2014Meet standard requirementsMet
    Shock, Vibration and IP ClassificationIEC 60601-1:2005/AMD2:2020, IEC 60601-2-27:2011, IEC 60529: 2013Meet standard requirementsMet
    Biocompatibility AssessmentISO 10993-1Meet standard requirementsYes - Assessed
    ASCA Test ReportsN/A (General term)Meet standard requirementsYes
    Design Verification TestingN/A (General term)Meet standard requirementsYes

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each physical test. It confirms that "Performance testing has been completed to demonstrate that the proposed LIFEPAK ECG Cables meet the safety and performance requirements established in the design specifications." These are typically laboratory-based engineering tests on device samples, not studies on patient data. The provenance for these tests is the manufacturer (Physio-Control Inc.). The tests are "non-clinical performance testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a hardware device undergoing engineering and safety performance testing, not an AI/ML algorithm requiring expert interpretation of medical images or data for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically for human-read or AI-generated interpretations of medical data, not for physical performance tests of a cable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML-based device. No human clinical studies were submitted as part of this 510(k) Premarket Notification.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML-based device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is typically adherence to established engineering, electrical, and biocompatibility standards (e.g., IEC 60601, ANSI AAMI EC53, ISO 10993-1). Compliance with these standards confirms the device's technical specifications and safety profile, rather than a clinical ground truth like a pathology report.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML-based device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML-based device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Lifesciences TruWave Invasive Pressure Transducers (or equivalent)) (11230-000024) Regulation Number: 21 CFR 870.2900
    ------|------------------------|--------------|
    | Patient transducer and electrode cable
    (21 CFR 870.2900
    |
    | Regulation Number | 21 CFR 870.2900
    | 21 CFR 870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIFEPAK Invasive Pressure Adapter Cable is intended to connect invasive pressure transducers from the patient to the LIFEPAK monitor/defibrillator for invasive pressure monitoring purposes and/or aid in diagnostic evaluation by a health care professional. The LIFEPAK Invasive Pressure Adapter Cable is intended for use in outdoor and indoor emergency care settings.

    Device Description

    The LIFEPAK Invasive Pressure Adapter Cable is a reusable, insulated, shielded, electrical cord (trunk cable) with a proximal connector (to connect with the LIFEPAK 35 device) and a main yoke that houses three distal leans with connectors (to connect with invasive pressure transducers). The LIFEPAK Invasive Pressure Adapter Cable is designed to transmit an electrical signal (data) between devices (i.e., to connect invasive pressure transducers from the patient to the LIFEPAK 35 monitor/defibrillator). The invasive pressure cable is designed to connect up to three pressure transducers to the front panel of the LIFEPAK 35 monitor/defibrillator for invasive pressure monitoring. Three channels are available for invasive pressure (IP) monitoring, with labels P1, P2, and P3. Each channel also has a user-selectable label for the line type through the LIFEPAK 35 monitor/defibrillator.

    Invasive Pressure monitoring involves the conversion of fluid pressure into an electrical signal. The conversion is accomplished with a pressure transducer (i.e., IP Probe). The invasive pressure cable passes the electrical signal from the transducers to the LIFEPAK 35 monitor/defibrillator.

    The LIFEPAK Invasive Pressure Adapter Cable is available in two different models that are compatible with either ICU Medical Transpac® IV Disposable Pressure Transducers or Edwards Lifesciences TruWave® Disposable Pressure Transducers. These pressure transducers provide industry standard sensitivity and defibrillation protection of at least 360 joules.

    The LIFEPAK Invasive Pressure Adapter Cable is not intended for use with other manufacturers' defibrillators and/or monitors. The LIFEPAK Invasive Pressure Adapter Cable is intended only for use with LIFEPAK 35 monitor/defibrillator.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the "LIFEPAK® Invasive Pressure Adapter Cable." It does not contain information about acceptance criteria for a device's performance, nor does it detail a study proving the device meets specific performance criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device through general performance testing.

    Therefore, most of the requested information cannot be extracted from this document. However, I can provide the available details:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The document states "Performance testing has been completed to demonstrate that the proposed LIFEPAK Invasive Pressure Adapter Cable meets the safety and performance requirements established in the design specifications." However, the specific acceptance criteria and the numerical results of these tests are not detailed.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The document only mentions "Biocompatibility Evaluation" and "Design Verification Testing" were completed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. The document does not describe a study involving expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. The document does not describe a study with adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The device is an adapter cable, not an AI-powered diagnostic tool, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The device is a physical cable, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    Not applicable. The document does not describe a study that uses ground truth in this context. The "performance data" refers to engineering and biocompatibility testing.

    8. The sample size for the training set:

    Not applicable. The device is a physical cable, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. The device is a physical cable, not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222712
    Date Cleared
    2023-11-22

    (440 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    80026

    Re: K222712

    Trade/Device Name: Fogg System Patient Monitoring Cables Regulation Number: 21 CFR 870.2900
    transducer and electrode cable (including connector) |
    | Classification: | Class II per 21 CFR §870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Patient Monitoring Cables are intended to be used with SpO2, Temperature, BP, ECG, Cardiac Output, and related monitoring devices. The Patient Monitoring Cables are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring devices, or between monitoring devices, for general monitoring and/or diagnostic evaluation by health care professionals.

    Device Description

    The Patient Monitoring Cables are replacements for similar accessory cables manufactured or specified by Original Equipment Manufacturers (OEM) for their respective transducers, monitors, and peripheral equipment. These are non-patient-contact, insulated, electrical cables with a connector (plug or receptacle) at each end, designed to transmit electrical power and/or signal (data) between medical devices (e.g., to connect SpO2 sensor, IBP transducer to a patient monitor). They are not intended to connect to the mains (i.e., not mains power cables), do not generate any type of power and/or signal, and have no additional non-electrical conducting or processing functionality.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for "Fogg System Patient Monitoring Cables." This document focuses on the substantial equivalence of the cables to a predicate device based on bench testing of physical and electrical properties, not on a clinical "study" involving AI or human-in-the-loop performance. Therefore, many of the questions related to AI device performance evaluation (such as MRMC studies, training set details, and expert ground truth establishment for AI) are not applicable to this submission.

    Here's a breakdown of the acceptance criteria and the "study" (nonclinical test data) presented in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of acceptance criteria with corresponding performance results. Instead, it states that the Fogg System Patient Monitoring Cables "met all acceptance criteria from the testing and analyses". The testing and analyses performed were:

    Acceptance Criteria (Implied)Reported Device Performance
    Cable assembly resistance and continuityConformed to standards
    Cable material signal integrityConformed to standards
    Electromagnetic interference and safety analysisConformed to standards
    Conformance to applicable sections of standardsMet IEC 60601 and AAMI/ANSI BP22

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document mentions "worst case representative cable assemblies and cable material" were used for performance testing. It does not provide a specific number for the sample size.
    • Data Provenance: The testing was "nonclinical" and performed on the physical devices/materials. The document does not specify a country of origin for the data, but given it's a US FDA submission, it's implied that the testing was conducted in a manner acceptable for US regulatory purposes. The testing is not retrospective or prospective clinical data; it's laboratory bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The "ground truth" for this type of device (patient monitoring cables) is established by engineering specifications, international standards (IEC 60601, AAMI/ANSI BP22), and established testing methodologies for electrical and material properties, not by human expert review of clinical cases or images.

    4. Adjudication method for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical imaging studies, often involving expert readers. For physical device testing, the "ground truth" is determined by whether the device's electrical and material properties meet established engineering and safety standards. There is no human adjudication process described for this type of nonclinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This submission is for patient monitoring cables, which are passive accessories, not an AI-powered diagnostic device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This submission is for patient monitoring cables, not a standalone AI algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is based on engineering specifications and adherence to recognized international standards such as IEC 60601 and AAMI/ANSI BP22 for patient transducer and electrode cables. It is a technical ground truth related to electrical continuity, signal integrity, electromagnetic compatibility, and material properties, not clinical outcomes or expert consensus on clinical data.

    8. The sample size for the training set

    This question is not applicable. There is no AI component requiring a training set for this device.

    9. How the ground truth for the training set was established

    This question is not applicable. As there is no AI component, no training set or its ground truth was established for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230571
    Date Cleared
    2023-05-30

    (90 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Canada

    Re: K230571

    Trade/Device Name: RFP-100A Connector Cable (Single Use) Regulation Number: 21 CFR 870.2900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the RFP-100A Connector Cable (Single Use) is to connect the RFP-100A Generator to separately cleared compatible RF devices.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called the "RFP-100A Connector Cable (Single Use)". It indicates that the device has been found substantially equivalent to a legally marketed predicate device.

    Crucially, this document does not include information about acceptance criteria, performance studies, or ground truth establishment for an AI/ML device.

    The device described, a connector cable, is a passive component used to connect two other cleared medical devices (a generator and RF devices). Its function is likely to transmit electrical signals, and its acceptance criteria would focus on electrical safety, biocompatibility, mechanical integrity, and compatibility with the specified devices, rather than the kind of performance metrics (sensitivity, specificity, AUC) typically associated with AI/ML-driven diagnostic or prognostic tools.

    Therefore, I cannot extract the requested information (table of acceptance criteria, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details) from the provided text because it is not relevant to the type of device being discussed.

    To provide the information you've requested, I would need a 510(k) summary or a similar document for an AI/ML-enabled medical device that performs a diagnostic, prognostic, or therapeutic function.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222370
    Date Cleared
    2023-04-02

    (240 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Beijing 102401 China

    Re: K222370

    Trade/Device Name: SpO2 Extension Cable Regulation Number: 21 CFR 870.2900
    Electrode, Patient, (Including Connector) Classification: II Product Code: DSA Regulation Number: 21 CFR 870.2900
    Patient Monitoring Cables (SpO2
    Extension Cable)
    K192404 | / |
    | Classification | 21 CFR 870.2900
    | 21 CFR 870.2900 | SAME |

    6

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SpO2 Extension Cables are intended to be used to connect SpO2 sensors, placed at appropriate sites on the patient to a monitoring device for SpO2 monitoring by health care professional.

    Device Description

    SpO2 Extension Cables are intended to be used to connect SpO2 sensors, placed at appropriate sites on the patient to a monitoring device for SpO2 monitoring by health care professional.

    The SpO2 Extension Cable is the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after manufacturers for their respective monitors.

    The SpO2 Extension Cable is comprised of Plug, Cable/Leadwires and Connector. It's intended to plug into monitoring device and connect with SpO2 sensor, for transmitting signals which generated by SpO2 sensor to monitoring device.

    By using the same types of construction and technological characteristics to the compatible patient monitor and SpO2 sensor, the SpO2 Extension Cable can avoid measured data corrupted.

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for an SpO2 Extension Cable. It confirms that the device is substantially equivalent to a predicate device and is subject to general controls. It does not contain information about clinical studies with AI or human readers, or detailed acceptance criteria for an AI-based device.

    Therefore, I cannot answer your specific questions related to acceptance criteria, study design for AI evaluation, sample sizes, expert involvement, or ground truth establishment as this information is not present in the provided text for this specific device.

    The document primarily focuses on the device's classification, general controls, and non-clinical testing to demonstrate substantial equivalence to a predicate device, which is a different type of evaluation than what you typically find for AI/ML-driven medical devices.

    Summary of available information (and absence of requested information):

    • 1. A table of acceptance criteria and the reported device performance: Not explicitly stated as pass/fail criteria from a performance study. The "performance" is demonstrated through compliance with electrical safety, performance, and EMC standards, and biocompatibility, as well as functional compatibility.

      • Acceptance Criteria (Inferred from compliance):
        • Biocompatibility: Meet ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Sensitization and Irritation).
        • Electrical Safety & Performance: Meet IEC 60601-1, ISO 80601-2-61 (Pulse Oximeter equipment), and ANSI AAMI EC53 (ECG trunk cables and patient leadwires).
        • EMC: Meet IEC 60601-1-2.
        • Compatibility: No measured data corruption during communication between SpO2 sensors and host monitors.
      • Reported Device Performance (Inferred from conclusions):
        • "The test results demonstrated that the proposed device complies with the following standards..." (listed above).
        • "SpO2 Extension Cable has also been evaluated the performance accuracy through Compatibility testing, which proves that no measured data corrupt during communication between SpO2 sensors and host monitors."
    • 2. Sample sized used for the test set and the data provenance: Not applicable. This is a hardware device, not an AI model. Testing involved standards compliance and compatibility, not a data-driven test set for an algorithm.

    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI model is not relevant here. Device performance is assessed against engineering standards and functional requirements.

    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this is not an AI device.

    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is not an AI device.

    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for this device's performance is compliance with established international and national standards for medical electrical equipment, biocompatibility, and functional compatibility.

    • 8. The sample size for the training set: Not applicable. This is a hardware device.

    • 9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 21