Search Filters

Search Results

Found 109 results

510(k) Data Aggregation

    Why did this record match?
    510k Summary Text (Full-text Search) :

    NIBP Cuff (U1684S-C51N); Disposable Neonatal NIBP Cuff (U1685S-C51N)
    Regulation Number: 21 CFR 870.1120
    U1684S-C51N, U1685S-C51N
    Classification: Medical Specialty: Cardiovascular
    Regulation: 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimed disposable blood pressure cuff (Models U1682S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (4–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The Unimed Disposable Neonatal NIBP Cuffs (Models U1682S-C51N, U1683S-C51N, U1684S-C51N, U1685S-C51N) are single-patient-use accessories for non-invasive blood pressure (NIBP) measurement. Each cuff is designed for neonatal limb circumferences ranging from 4 cm to 15 cm (Neonate 2 through Neonate 5) as follows.

    • U1682S-C51N: 4 cm - 8 cm.
    • U1683S-C51N: 6 cm - 11 cm.
    • U1684S-C51N: 7 cm - 13 cm.
    • U1685S-C51N: 8 cm - 15 cm.
      The cuffs are intended to be used with compatible blood pressure monitors employing oscillometry. They are non-sterile, and are disposed of after single-patient use to reduce cross-contamination risks.
    AI/ML Overview

    This 510(k) clearance letter is for a Disposable Neonatal NIBP Cuff. It does not describe an AI/ML powered medical device, nor does it detail a study involving AI/ML.

    Therefore, many of the requested criteria related to AI/ML device validation (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, etc.) are not applicable to this document.

    The document primarily focuses on establishing substantial equivalence to a predicate device through non-clinical testing and material biocompatibility.

    Here's the information that can be extracted from the provided text, primarily related to the non-clinical testing performed to meet acceptance criteria for a physical medical device:


    Acceptance Criteria and Device Performance for Disposable Neonatal NIBP Cuff

    The acceptance criteria for this device are primarily based on established international standards for non-invasive blood pressure cuffs and biocompatibility. The performance is demonstrated through non-clinical testing, confirming adherence to these standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Compliance)Reported Device Performance (Test Results)
    ISO 81060-1:2007 (Non-invasive sphygmomanometers – Requirements and test methods for non-automated measurement type)The device met all design specifications based on testing conducted in accordance with this standard.
    ISO 80601-2-30:2018 (Medical electrical equipment — Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers)The device met all design specifications based on testing conducted in accordance with this standard.
    Biocompatibility - ISO 10993-5 (Tests for in vitro cytotoxicity)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Biocompatibility - ISO 10993-10 (Tests for skin sensitization)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Biocompatibility - ISO 10993-23 (Tests for irritation)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Pressure LimitsThe device operates within the accepted range of 0-300 mmHg, which is the same as the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in the provided document. The non-clinical testing likely involved a representative sample of each cuff model (U1682S-C51N, U1683S-C51N, U1684S-C51N, U1685S-C51N).
    • Data Provenance: The document states "Non-clinical tests were conducted..." but does not specify the country of origin where the testing was performed. Given the manufacturer's location (Shenzhen, China), it's plausible the testing was conducted there or at a certified lab supporting the manufacturer. The testing is described as non-clinical, implying it was laboratory-based product verification and validation, not retrospective or prospective human clinical data collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable. This document describes the clearance of a physical medical device (NIBP cuff) via substantial equivalence, not an AI/ML algorithm. Therefore, there is no mention of "ground truth" derived from expert consensus on medical images or data. The "ground truth" for this device would be the defined specifications and performance requirements established by the relevant ISO standards.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical trial settings, especially for AI/ML performance evaluation where multiple human readers might disagree on findings. For non-clinical device testing like this, adherence to predefined test protocols and specifications is the primary method of evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI assistance on human reader performance, which is not applicable to a non-AI medical device like a blood pressure cuff.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance evaluation of an algorithm was not done. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by compliance with recognized consensus standards (ISO 81060-1, ISO 80601-2-30, and ISO 10993 series for biocompatibility). The device's performance metrics are evaluated against the specific requirements and test methods outlined in these standards.

    8. The Sample Size for the Training Set

    This is not applicable. There is no training set mentioned as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. There is no training set mentioned as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250757
    Manufacturer
    Date Cleared
    2025-05-29

    (78 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    DXN

    • 21 CFR 880.2910/FLL
    • 21 CFR 880.2400/KMI
    • 21 CFR 870.1025/DSI
    • 21 CFR 870.1425/DQK
    • 21 CFR 870.1120
      CFR 880.2910/ FLL
      21 CFR 870.1025/ DSI
      21 CFR 870.1425/ DQK
      21 CFR 880.2400/ KMI
      21 CFR 870.1120
      CFR 880.2910/ FLL
      21 CFR 870.1025/ DSI
      21 CFR 870.1425/ DQK
      21 CFR 880.2400/ KMI
      21 CFR 870.1120
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.

    The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.

    The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.

    The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.

    The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.

    The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

    Devices with Masimo technology are only indicated for use with Masimo accessories.

    Radius VSM Accessories:

    Radius VSM ECG Electrodes are disposable, single-patient use ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.

    Radius VSM Blood Pressure Cuffs are accessories intended to be used with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.

    Device Description

    The Radius VSM and accessories are an FDA cleared (K223498), wearable, battery-operated, multi-modular patient monitoring platform that allows for the ability to scale and tailor the use of different monitoring technologies based upon the hospital and clinician's assessment of what technologies are appropriate.

    As part of this submission, a MAP feature is being added to the Radius VSM. The feature is a software feature that uses the previously cleared systolic and diastolic measurement capabilities to automate the calculation of MAP using the following formula: MAP = 1/3* Systolic + 2/3*Diastolic.

    The MAP is calculated by the Radius VSM NIBP Module and displayed on the Radius VSM Wearable Monitor. There were no other features added as part of this submission.

    AI/ML Overview

    The provided 510(k) clearance letter and summary discuss the addition of a Mean Arterial Pressure (MAP) feature to the previously cleared Radius VSM and Accessories device. The primary focus of the performance data section is on validating this new MAP feature.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document states that the acceptance criterion for Blood Pressure (including MAP) is:

    "Meets ISO 81060-2 (Mean difference of ≤5 mmHg with a standard deviation of ≤8 mmHg)"

    The document directly states that the results of the clinical testing supported the clinical performance of the MAP in accordance with ISO 81060-2. While specific numerical results (e.g., the exact mean difference and standard deviation achieved) are not explicitly provided in the summary table, the clearance implies that these metrics fell within the specified ISO 81060-2 limits for the MAP feature.

    Table 1: Acceptance Criteria and Reported Device Performance for MAP Feature (as inferred from the document)

    FeatureAcceptance CriteriaReported Device Performance
    Mean Arterial Pressure (MAP)Meets ISO 81060-2: Mean difference of ≤5 mmHg with a standard deviation of ≤8 mmHgPerformance met ISO 81060-2 (i.e., mean difference and standard deviation were within the specified limits).

    Study Details for MAP Feature Validation

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: The document does not explicitly state the numerical sample size (number of subjects/patients) used for the clinical test set. It only mentions "clinical study data."
      • Data Provenance: The document does not specify the country of origin. It indicates it was a "clinical study" and implies it was prospective ("clinical testing is provided to support its performance" for the added feature).
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

      • Not applicable as the ground truth was established by an objective reference device, not human experts.
    3. Adjudication Method for the Test Set:

      • Not applicable, as the method for ground truth establishment was comparison to a reference device.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No, an MRMC study was not done. The study was a comparison of the device's calculated MAP to invasively measured MAP from a reference device. This is a technical performance validation, not a study assessing human reader improvement with AI assistance.
    5. If a Standalone Performance (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this was a standalone performance study. The Radius VSM automatically calculates the MAP based on the NIBP measurements (Systolic and Diastolic Pressure). The clinical testing validated the accuracy of this calculation against a reference standard, without human intervention in the MAP calculation or interpretation for the test itself.
    6. The Type of Ground Truth Used:

      • Reference Ground Truth: Invasively measured MAP values from a 510(k) cleared reference device (K171801). This reference device is identified as "IntelliVue Multi-Measurement Module X3." This constitutes a device-based reference standard or instrument-based ground truth.
    7. The Sample Size for the Training Set:

      • The document does not provide information about a training set since the MAP feature appears to be a direct calculation using a standard formula (MAP = 1/3* Systolic + 2/3*Diastolic) rather than a machine learning model that requires a training phase. While the device as a whole (Radius VSM) likely had training and validation phases for its other parameters, the specific "addition of a Mean Arterial Pressure (MAP) feature" is described as a software feature that "automates the calculation" using a known formula. Therefore, a separate training set for this specific MAP feature is unlikely to have been required or used in the conventional machine learning sense.
    8. How the Ground Truth for the Training Set was Established:

      • As inferred above, a specific training set and ground truth establishment for this isolated MAP calculation feature are not described, given its nature as a direct formulaic calculation.

    Summary of Key Information:

    The core of this submission revolves around adding a simple, formula-based calculation for MAP. The primary study presented is a clinical validation confirming that the device's computed MAP aligns with a known industry standard (ISO 81060-2) when compared against an invasive reference device. This is a technical performance validation rather than a complex AI-driven diagnostic study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242623
    Date Cleared
    2024-10-31

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: Disposable Blood Pressure cuff: Reusable Blood Pressure cuff Regulation Number: 21 CFR 870.1120
    |
    | Classification | 21CFR 870.1120
    |
    | Classification
    Regulation | 21CRF 870.1120
    | 21CRF 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate size. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in neonatal, infant, child and adult sizes. The subject device includes disposable blood pressure cuff and reusable blood pressure cuff and contains Cuff with/without bladder and single/dual tubes Air Hose.

    AI/ML Overview

    The provided document is a 510(k) summary for a blood pressure cuff and does not contain detailed information about acceptance criteria, reported device performance for specific metrics, or any study proving such performance. The document explicitly states:

    "No clinical study is included in this submission"

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them because such information is not present in the provided text. The document focuses on demonstrating substantial equivalence to a predicate device primarily through non-clinical data and compliance with voluntary standards for design specifications.

    However, I can extract the information that is present:

    8. Non-clinical data

    • Performance testing: Non-clinical tests were conducted to verify that the proposed devices met all design specifications.
    • Compliance with standards: The test results demonstrated that the proposed device complies with ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.

    Based on this, here's what can be stated about the available information:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document mentions compliance with ISO 81060-1, which would contain performance requirements for non-automated sphygmomanometers (blood pressure cuffs). However, the specific acceptance criteria used for the device and the reported device performance against those criteria are not detailed.
    2. Sample sized used for the test set and the data provenance: Not provided. The document only mentions "non-clinical tests" and "performance testing" but does not give details about sample sizes or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study or expert-based ground truth establishment is mentioned. The testing focused on technical specifications and standards compliance.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. The document explicitly states "No clinical study is included in this submission."
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a blood pressure cuff, which is a physical accessory, not an algorithm.
    7. The type of ground truth used: Not applicable for performance evaluation as no clinical data or comparative performance study is provided. For demonstrating substantial equivalence, the ground truth indirectly relies on the established performance and safety of the predicate device and the standards.
    8. The sample size for the training set: Not applicable. This is a non-AI/ML device.
    9. How the ground truth for the training set was established: Not applicable. This is a non-AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240165
    Date Cleared
    2024-03-20

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Aneroid Sphygmomanometer with Stethoscope; Aneroid Sphygmomanometer Regulation Number: 21 CFR 870.1120
    STETHOSCOPE |
    | Common name: | Blood Pressure Cuff |
    | Regulation Number: | 870.1120
    stethoscope |
    | Common name: | Blood Pressure Cuff |
    | Regulation Number: | 870.1120
    | Same |
    | Regulation
    Number | 21 CFR 870.1120
    | 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on new borns, infants, children, young adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.

    Device Description

    The ANEROID SPHYGMOMANOMETER and ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE are manual blood pressure monitor used to measure blood presure.It consists of a cuff,a inflation bulb,a pressure gauge, an air valve and a stethoscope(optional).

    Cuff: The cuff is made of a durable fabric and is wrapped around the upper arm of the patient. It inflates and deflates during the measurement process. The material of outside layer is nylon cloth.

    Inflation Bulb: The inflation bulb is a rubber bulb that connects to the cuff and is used to manually inflate the cuff with air. The material is Neoprene or Silicon rubber or Nature latex or PVC.

    Pressure Gauge: The pressure gauge is a circular dial with a needle that indicates the pressure within the cuff. It is calibrated in millimeters of mercury (mmHg) and shows the systolic and diastolic blood pressure readings. The material is Aluminum or ABS.

    Air Valve: The valve is a control mechanism that allows for the release of air from the cuff once the measurement is complete.

    Stethoscope: Although not a part of the device itself, a stethoscope is typically used in conjunction with an Aneroid Sphygmomanometer to listen for the blood flow sounds, known as Korotkoff sounds, in the brachial artery during the measurement. Three model are single head Stethoscope(Model: TY-S01),Dual Head Stethoscope(Model: TY-S05) and Spraque Rappaport Stethoscope(Model: TY-S04) individually.These stethoscopes are listed in FDA database.

    AI/ML Overview

    The provided submission describes an Aneroid Sphygmomanometer. Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Reference Standard)
    Accuracy± 3 mmHg (ISO 81060-1:2007)
    BiocompatibilityComplies with ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021
    Safety & PerformanceComplies with ISO 81060-1:2007

    The document explicitly states that the "Accuracy" of the device is ± 3 mmHg, which is also the performance reported. For "Biocompatibility" and "Safety & Performance", compliance with the specified ISO standards is the stated performance.

    Note: The submission describes an aneroid sphygmomanometer, which is a mechanical device. The concept of an "AI device" with human-in-the-loop performance or standalone performance, sample sizes for test/training sets, adjudication methods, and expert qualifications for ground truth establishment, are not applicable to this type of device. The provided text is a 510(k) summary for a traditional medical device, not an AI/ML-powered device.

    Therefore, the following points are answered with "Not Applicable" (N/A) due to the nature of the device.

    2. Sample size used for the test set and the data provenance
    N/A - This device is a mechanical blood pressure monitor, not an AI/ML-powered device that uses a "test set" in the context of data. The performance testing refers to compliance with a standard (ISO 81060-1), which involves testing the device itself, not an algorithm's performance on a dataset.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    N/A - See explanation above.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    N/A - See explanation above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    N/A - This is a mechanical device, not an AI-assisted one.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    N/A - This is a mechanical device, not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
    N/A - The ground truth for this device's accuracy would be established by direct comparison to a reference standard, as defined by ISO 81060-1, which likely involves a calibrated pressure standard.

    8. The sample size for the training set
    N/A - This is a mechanical device, not an AI/ML-powered device that requires a "training set."

    9. How the ground truth for the training set was established
    N/A - See explanation above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233276
    Date Cleared
    2023-12-21

    (83 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    215222 China

    Re: K233276

    Trade/Device Name: Reusable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120
    Information Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number:870.1120
    |
    | Regulation
    Number | 21 CFR 870.1120
    | 21 CFR 870.1120
    | 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in adult size.

    Device Description

    The proposed device is an accessory used in conjunction with non-invasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable. The model difference of the proposed device includes bladder type, tube, color, joint model and size.

    AI/ML Overview

    The provided text is a 510(k) summary for a Reusable Blood Pressure Cuff. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device's performance against specific acceptance criteria for an AI/ML medical device.

    Therefore, the document does not contain the information required to describe the acceptance criteria and the study that proves an AI/ML device meets those criteria.

    Specifically, the document does not provide:

    • A table of acceptance criteria and reported device performance: The document lists standards the device complies with (IEC 80601-2-30, ISO 81060-1, ISO 10993-5, ISO 10993-10), but these are general safety and performance standards for a mechanical device, not AI/ML acceptance criteria like sensitivity, specificity, or AUC. No specific performance metrics are reported against quantitative acceptance criteria.
    • Sample size and data provenance for a test set: No mention of a test set, data provenance, or sample size for evaluating AI/ML algorithm performance.
    • Number and qualification of experts for ground truth: Not applicable, as this is not an AI/ML device requiring expert-established ground truth.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: Not applicable.
    • Standalone (algorithm only) performance: Not applicable.
    • Type of ground truth used: Not applicable.
    • Sample size for the training set: Not applicable.
    • How ground truth for the training set was established: Not applicable.

    The document states:

    • "6. Non-Clinical Test Conclusion" and "6.1 Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 80601-2-30: 2018..., ISO 81060-1:2007..., ISO 10993-5: 2009..., ISO 10993-10: 2010..."
    • "6.2 Human Factors / Usability Testing Usability testing is not required per the risk profile of this product."
    • "7. Clinical Test Conclusion No clinical study is included in this submission."

    This unambiguously indicates that the submission is for a physical medical device (a blood pressure cuff), assessed through non-clinical bench testing for compliance with manufacturing and biocompatibility standards, and not an AI/ML driven diagnostic or assistive device that would require the specified performance evaluations.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    1201W2, 12L2W2, 1202W2, 12L3W2, 1203W2, 1204W2, 1205W2, 1206W2 and 1207W2) Regulation Number: 21 CFR 870.1120
    |
    | Regulation Number | 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system.

    The subject device is categorized into two types of models according to its usage.

    Reusable Blood Pressure Cuff models:

    1201B1,12L2B1,1202B1,12L3B1,1204B1,1205B1,1206B1,1207B1-1#,1207B1-2#,1207B1-

    3#,1207B1-4#,1207B15#,1201W2,12L2W2,1202W2,12L3W2,1203W2,

    1204W2,1205W2,1206W2 and 1207W2

    Disposable Blood Pressure Cuff models:

    1201D1,12L2D1,1202D1,12L3D1,1204D1,1205D1,1204D1,1207D1-1#,1207D1-2#,1207D1-

    3#,1207D1-4#,1207D1-5#

    Connector models: PLINMA-03-127-003, PLINMA-03-T, PLINMA-03-Z3

    AI/ML Overview

    The provided text is a 510(k) summary for a Blood Pressure Cuff. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics. The information about a study proving device performance is limited.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily refers to meeting ISO standards and does not present a table of specific acceptance criteria (e.g., specific accuracy ranges for blood pressure measurements) versus reported device performance in a clinical setting. It states that the device was "tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type". However, it does not provide the results of this testing in a quantifiable form or against specific, pre-defined acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states, "The subject and predicate devices are exactly the same. Since the blood pressure cuffs are identical, no further clinical testing is necessary." This implies that no separate test set with human subjects was used for this 510(k) submission to evaluate the performance of the device in terms of blood pressure measurement accuracy. The data provenance is assumed to be from the predicate device's testing, but details are not provided here.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Since no separate clinical performance testing was conducted for this submission, no experts were used to establish ground truth from newly collected data for this specific device.

    4. Adjudication Method for the Test Set:

    Not applicable, as no dedicated clinical performance test set was used for this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable, as this is a blood pressure cuff and not an AI-powered diagnostic imaging device. The document does not describe any human reader involvement or AI assistance.

    6. Standalone Performance Study:

    A standalone performance study focused on technical compliance with ISO 81060-1 was mentioned ("The subject device has been tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type"). However, the results of this testing, which would quantify its performance against the ISO standard's requirements, are not provided in this summary. The summary only attests that the device "met the requirements."

    7. Type of Ground Truth Used:

    For the "performance data" mentioned, the ground truth would be the reference measurements and methods defined within the ISO 81060-1 standard for non-invasive sphygmomanometers. For the biocompatibility testing, the ground truth would be the established scientific methods and criteria for evaluating cytotoxicity, sensitization, and irritation.

    8. Sample Size for the Training Set:

    Not applicable, as this is a medical device (blood pressure cuff) and not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    Summary of Acceptance Criteria and Study (based on available information):

    Acceptance Criteria (Implied)Reported Device Performance (Summary)
    Compliance with ISO 81060-1 for non-automated measurement type"The subject device has been tested according to ISO 81060-1... and met the requirements."
    Biocompatibility (Cytotoxicity, Sensitization, Irritation)"The test results show that the proposed device has no issue of cytotoxicity, sensitization, or skin irritation."
    Functional equivalence to predicate devicesStated that "The subject and predicate devices are exactly the same."
    Limb Circumference Range (as per ISO 81060-1 and labeling)"the size of limb circumference of subject device met the requirements of ISO 81060-1, and the limb circumference was marked in the labeling."
    Tube Configuration (meeting ISO 81060-1)"The subject device met the requirements of ISO 81060-1."

    Study Details:

    • Test Set Sample Size: Not applicable for clinical performance testing, as it was deemed unnecessary due to the device being "exactly the same" as predicates. For technical testing to ISO 81060-1, the sample size is implicitly defined by the standard's requirements for testing a physical device.
    • Data Provenance: Not specified for the ISO 81060-1 technical testing, but typically conducted in a laboratory setting.
    • Experts for Ground Truth: Not applicable for clinical performance as no new human data was collected. Ground truth for technical testing is established by standard operating procedures and reference equipment.
    • Adjudication Method: Not applicable for clinical performance.
    • MRMC Study: No.
    • Standalone Study: Yes, a technical standalone study was performed to ISO 81060-1 and biocompatibility standards, with results indicating compliance, although specific performance metrics are not listed in this summary.
    • Type of Ground Truth: ISO standard requirements and established biocompatibility testing protocols.
    • Training Set Sample Size: Not applicable (not an AI/ML device).
    • Training Set Ground Truth Establishment: Not applicable.

    The document primarily relies on demonstrating substantial equivalence to predicate devices and adherence to relevant ISO standards, rather than presenting a detailed clinical performance study specific to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231961
    Device Name
    NIBPCuff
    Date Cleared
    2023-08-30

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K231961

    Trade/Device Name: Reusable NIBP Cuff, Disposable NIBP Cuff Regulation Number: 21 CFR 870.1120
    |
    | Classification | 21CFR 870.1120
    | / |
    | Classification
    Regulation | 21CRF 870.1120
    | 21CRF 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes.

    The proposed device includes disposable blood pressure cuff and reusable blood pressure cuff. Both disposable and reusable blood pressure cuff have the same structure, which contains Cuff with bladder and single-tube Air Hose.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a non-invasive blood pressure (NIBP) cuff. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report of the device's accuracy or efficacy as one might find for a more complex AI-driven medical device.

    Therefore, the requested information, particularly concerning "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-driven device (like an effect size of human readers improving with AI, or standalone algorithm performance), is largely not applicable to this document. This document primarily describes the physical characteristics, intended use, and manufacturing standards of a blood pressure cuff.

    However, I can extract the relevant information where available, focusing on performance data related to blood pressure cuffs.

    Here's an analysis based on the provided text:

    Key Takeaway: This 510(k) submission primarily relies on physical and material testing and comparison to a predicate device to demonstrate substantial equivalence, not a clinical study proving its performance against specific physiological accuracy criteria in humans, nor an AI-driven system's performance.


    1. A table of acceptance criteria and the reported device performance

    The document doesn't present a table of acceptance criteria and corresponding reported device performance in the same way one would for an AI model's diagnostic accuracy. Instead, it refers to compliance with established standards for blood pressure cuffs.

    • Acceptance Criteria (as implied by standards compliance):
      • Max. Leakage:
    Ask a Question

    Ask a specific question about this device

    K Number
    K230859
    Date Cleared
    2023-06-28

    (91 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Zhejiang 325011 China

    Re: K230859

    Trade/Device Name: Aneroid Sphygmomanometer Regulation Number: 21 CFR 870.1120
    Sphygmomanometer Regulation Name: Blood Pressure Cuff Product Code: DXQ,LDE Regulation Number: 21 CFR 870.1120
    Regulation Name: Blood Pressure Cuff 510(k) number: K212416 Product Code: DXQ Regulation Number: 21 CFR 870.1120
    | Same |
    | Regulation Number | 21 CFR 870.1120
    | 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aneroid sphygmomanometer is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on children or adults. The device is intended to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.

    Device Description

    The product is a manual non-invasive aneroid sphygmomanometer which respectively uses an inflation cuff wrapped around the upper arm. The cuff is inflated and deflated by a manual inflation bulb. Besides a manometer (Aneroid gauge), the accessories of MODEL Ks-1222,Ks-1621,Ks-3201,Ks-2203 also includes one selected cuff, one inflation bulb, instruction manual and a vinyl bag. It is conjunction with Stethoscope when use.

    AI/ML Overview

    The provided document is a 510(k) summary for an Aneroid Sphygmomanometer. It details the device's characteristics and its comparison to a predicate device to establish substantial equivalence. However, this document does not describe a study involving an AI-powered device or a multi-reader multi-case (MRMC) comparative effectiveness study.

    The device in question, an "Aneroid Sphygmomanometer," is a manual, non-automated blood pressure meter. The performance testing cited is against ISO 81060-1:2007 "Non-invasive sphygmomanometersPart 1: Requirements and test methods for non-automated measurement type." This standard governs the accuracy of manual blood pressure devices, not AI performance.

    Therefore, many of the requested details, such as acceptance criteria for AI performance, sample sizes for test/training sets for AI, ground truth establishment for AI, expert involvement for AI, MRMC studies, and standalone AI performance, are not applicable to this submission.

    The document focuses on demonstrating that the new manual device performs similarly to a legally marketed predicate manual device and meets relevant performance and biocompatibility standards for non-automated devices.

    However, I can extract the general "acceptance criteria" for the accuracy of the blood pressure measurement as stated in the comparison table and the performance test standard:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for Blood Pressure Measurement Accuracy)Reported Device Performance (Subject Device)
    Pressure: +/- 3 mmHg of reading (per ISO 81060-1)Pressure: +/- 3 mmHg of reading
    Compatibility with Environment
    Can be used from 50°F to 104°F (10°C to 40°C) and 15% ~ 85% RH humidityCan be used from 50°F to 104°F (10°C to 40°C) and 15% ~ 85% RH humidity
    Biocompatibility
    Conforms to the requirements of ISO 10993 series standards (specifically ISO 10993-1, -5, -10, -23)Conforms to the requirements of ISO 10993 series standards

    2. Sample size used for the test set and the data provenance:

    • Sample Size for performance testing: The document states "full performance testing representing the current devices currently marketed" and mentions compliance with ISO 81060-1. While the specific number of subjects/measurements is not detailed in this summary, ISO 81060-1 typically requires clinical validation with a certain number of subjects (e.g., at least 85 subjects for blood pressure accuracy evaluation). The document does not specify data provenance (country, retrospective/prospective).
    • Data Provenance: Not explicitly stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for AI-driven performance. For a manual blood pressure device assessed against ISO 81060-1, the "ground truth" for blood pressure typically comes from a reference measurement method (e.g., intra-arterial measurement or a highly accurate clinical standard measurement performed by trained personnel). The document does not specify the number or qualifications of these individuals involved in the ground truth establishment for clinical validation, as it's a standard process dictated by the ISO standard.

    4. Adjudication method for the test set:

    • Not applicable for AI-driven performance. For a manual blood pressure device, adjudication would not be in the typical sense of expert review of AI outputs.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This is a manual, non-AI device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a manual device; there is no algorithm in the sense of AI. The performance testing is for the physical device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For blood pressure measurement accuracy (referenced by ISO 81060-1), the ground truth is typically established by simultaneous measurements using a highly accurate reference method (e.g., intra-arterial pressure or a validated mercury sphygmomanometer measurement by trained observers). This is a technical ground truth related to physical measurement.

    8. The sample size for the training set:

    • Not applicable. This is a non-AI device; there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. This is a non-AI device.

    In summary, the provided FDA 510(k) summary is for a traditional, non-automated medical device (Aneroid Sphygmomanometer). Therefore, it does not contain information related to AI-specific study designs, acceptance criteria, or performance metrics. The "study" proving the device meets acceptance criteria refers to non-clinical performance testing against established international standards for manual blood pressure devices and biocompatibility.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223498
    Manufacturer
    Date Cleared
    2023-06-01

    (192 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CFR 880.2910/ FLL
    21 CFR 880.2400/ KMI
    21 CFR 870.1025/ DSI
    21 CFR 870.1425/ DQK
    21 CFR 870.1120
    CFR 880.2910/ FLL
    21 CFR 870.1025/ DSI
    21 CFR 870.1425/ DQK
    21 CFR 880.2400/ KMI
    21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Radius VSM:

    The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.

    The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.

    The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.

    The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.

    The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.

    The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVI as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PV i measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

    Devices with Masimo technology are only indicated for use with Masimo accessories.

    Radius VSM Accessories:

    Radius VSM ECG Electrodes are disposable, single-patient ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.

    Radius VSM Blood Pressure Cuffs are accessories intended to be use with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.

    Device Description

    The Radius VSM and Accessories is a wearable, multi-modular patient monitoring platform that allows for the ability to scale and tailor the use of monitoring technologies based upon the hospital's and clinician's assessment of what technologies are appropriate. The purpose of this submission is the premarket notification for the introduction of Masimo Radius VSM and Accessories, including its use with the previously cleared Root (K191882) and Masimo Patient SafetyNet (K071047).

    The Radius VSM and Accessories system comprises of the Radius VSM Wearable Monitor, Radius VSM ECG Module and Electrodes, and the Radius VSM NiBP Module and Cuff.

    AI/ML Overview

    The provided text describes the acceptance criteria and study results for the Masimo Radius VSM and Accessories device, focusing specifically on the Non-invasive Blood Pressure (NiBP) feature.

    1. Acceptance Criteria and Reported Device Performance (NiBP Feature):

    The clinical performance analysis for the NiBP feature supported by the Masimo Radius VSM device had the following acceptance criteria and reported values:

    ParameterAcceptance CriteriaReported Device PerformancePass/Fail
    Mean value of the differences ( $\bar{x}_n$ )$ \bar{x}_n \le 5$ mmHgSystolic: -1.23 mmHgPass
    Diastolic: -2.67 mmHgPass
    Standard deviation of differences ( $s_n$ )$s_n \le 8$ mmHgSystolic: 7.32 mmHgPass
    Diastolic: 7.13 mmHgPass
    Standard deviation of differences per subject (sm)Systolic: ≤ 6.82 mmHgSystolic: 6.17 mmHgPass
    Diastolic: ≤ 6.39 mmHgDiastolic: 6.26 mmHgPass

    The device met all specified acceptance criteria for the NiBP feature.

    2. Sample Size and Data Provenance for the Test Set:

    • Sample Size:
      • NiBP Feature: 89 subjects.
      • ECG Waveform Comparison: 31 subjects.
      • Patient Posture, Position, and Activity: 20 subjects.
      • Aggregate Respiration Rate (First Study): 48 subjects.
      • Aggregate Respiration Rate (Second Study): The number of healthy volunteer subjects is not explicitly stated, but it's implied to be a separate group for validation of integration.
    • Data Provenance: The document does not explicitly state the country of origin. The studies are described as "clinical studies," implying prospective data collection for the purpose of validating the device. The term "healthy volunteer subjects" used in the fifth study further suggests prospective, controlled data collection.

    3. Number of Experts and Qualifications for Ground Truth:

    The document does not specify the number or qualifications of experts used to establish ground truth for any of the studies mentioned.

    4. Adjudication Method for the Test Set:

    The document does not describe any specific adjudication method for the test set data.

    5. MRMC Comparative Effectiveness Study:

    No mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or human readers improving with AI assistance is made in the provided text. The studies focus on device performance against reference measurements or previously cleared monitors/algorithms, not on human-AI collaboration.

    6. Standalone Performance (Algorithm Only):

    • For the NiBP feature, the study was conducted to validate the clinical performance of the Radius VSM's NiBP feature against reference blood pressure measurements, implying standalone performance of the algorithm integrated into the device.
    • For the ECG waveform comparison, the device's ECG output was compared to an existing FDA-cleared ECG monitor, indicating standalone performance of the device's ECG functionality.
    • For the patient posture, position, and activity feature, the testing supported the "correct integration of the algorithm that was previously cleared," suggesting a focus on the device's implementation of an existing standalone algorithm.
    • For the Aggregate Respiration Rate, the algorithm's performance was evaluated against manually annotated capnography data, indicating standalone algorithm performance.

    7. Type of Ground Truth Used:

    • NiBP: Clinical performance was validated through comparison against "reference blood pressure measurements."
    • ECG: Comparison against an "FDA cleared ECG monitor."
    • Patient Posture, Position, and Activity: Based on the "correct integration of the algorithm that was previously cleared." The original ground truth for this algorithm (K191882) is not detailed here, but the study validates its implementation in the new device.
    • Aggregate Respiration Rate: "Reference respiration rate derived from manual annotated capnography data."

    8. Sample Size for the Training Set:

    The document does not provide information on the sample size used for training sets for any of the algorithms or features. The studies described are validation (test set) studies.

    9. How Ground Truth for the Training Set was Established:

    As no information regarding training sets is provided, there is no detail on how their ground truth was established. The document focuses on the validation of integrated features, some of which (like PVi, RRa, and position monitoring) leverage previously cleared Masimo technologies, implying that their development and training (if applicable) occurred prior to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230145
    Device Name
    ONE-CUF
    Date Cleared
    2023-03-14

    (55 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Drive Wauwatosa, Wisconsin 53226

    Re: K230145

    Trade/Device Name: One-cuf Regulation Number: 21 CFR 870.1120
    |
    | Regulation: | 21 CFR 870.1120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The blood pressure cuff is an accessory used in con-invasive blood pressure measurement systems. It is non-sterile, single-patient use. It is available in pediatric and adult sizes. The cuff is not designed, sold or intended for use except as indicated.

    Device Description

    ONE-CUF blood pressure cuffs are accessories used in conjunction with noninvasive blood pressure (NIBP) measurement systems to obtain a blood pressure measurement. Blood pressure cuffs are non-sterile and single-patient use and may not be reprocessed for use on additional patients. They are available in pediatric and adult sizes. The devices are not designed, sold or intended for use except as indicated. Blood pressure cuffs do not include serviceable parts or components. Non-Invasive Blood Pressure Cuffs incorporate an inflatable non-distensible bladder, sized to encircle a patient's limb. The cuff includes one or two flanges for attaching flexible tubing. This allows air to flow in and out of the cuff bladder for inflation and deflation. Inflation allows for occlusion of an artery to facilitate the measurement of automated noninvasive blood pressure (NIBP). The cuff tubes are terminated with connectors that allow for attachment to a blood pressure hose. This cuff will be offered with two different connection systems, bayonet and DINACLICK.

    AI/ML Overview

    The information provided describes the ONE-CUF blood pressure cuff and its path to substantial equivalence, primarily based on non-clinical testing and comparison to a predicate device. It explicitly states that no clinical or animal studies were required to support substantial equivalence. Therefore, there is no study described that proves the device meets specific acceptance criteria related to a model's performance in a diagnostic or predictive context (e.g., accuracy, sensitivity, specificity for an AI algorithm).

    The "acceptance criteria" referred to in this document are primarily related to product performance specifications and compliance with voluntary standards for a mechanical medical device, rather than the performance metrics of an AI model.

    Here's an attempt to structure the answer based on the provided text, recognizing the limitations regarding AI-specific criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (ONE-CUF)
    Physical PerformanceLead Rate (Maximum allowable pressure loss rate)0.6 sccm (standard cubic cm per minute)
    Air LeakageLess than 4mmHg/min
    Pressure Range0-300 mmHg
    Mechanical StrengthData provided in support of substantial equivalence (specific values not detailed in summary)
    Tensile TestingData provided in support of substantial equivalence (specific values not detailed in summary)
    EnvironmentalOperating Temperature0° C to 40° C
    Storage Temperature-20° C to 55° C
    Relative Humidity (Operating)15% to 90% humidity, non-condensing
    Relative Humidity (Storage)0% to 95% humidity, non-condensing
    BiocompatibilityIrritation, Sensitization, Cytotoxicity (evaluated per ISO 10993-1:2018; chemical characterization also performed)Evaluated per ISO10993-1:2018
    Standards ComplianceIEC 80601-2-30:2018 (Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers)Compliant (applicable sections)
    ISO 81060-1:2007 (Non-invasive sphygmomanometers Part 1: Requirements and test methods for non-automated measurement type)Compliant (applicable sections)
    OtherEnvironmental ConditioningData provided in support of substantial equivalence (specific details not in summary)
    Pressure CapacityData provided in support of substantial equivalence (specific details not in summary)
    Human Factors Validation (Summative Usability)Performed to support substantial equivalence (specific outcomes not detailed)

    2. Sample size used for the test set and the data provenance

    Not applicable. The reported studies are non-clinical, focusing on device performance and safety specifications rather than a diagnostic model's performance on a test set of patient data. There is no mention of a "test set" in the context of an AI algorithm or diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a blood pressure cuff and does not involve AI or diagnostic interpretation requiring expert ground truth for a test set.

    4. Adjudication method for the test set

    Not applicable. There is no test set for diagnostic performance or AI algorithm output.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a non-AI blood pressure cuff. No MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a non-AI blood pressure cuff. No standalone algorithm performance was evaluated.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is established by adherence to documented test procedures, engineering specifications, and established international standards (e.g., ISO, IEC) for medical device performance and biocompatibility.

    8. The sample size for the training set

    Not applicable. This device is a non-AI blood pressure cuff. There is no training set for an algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This device is a non-AI blood pressure cuff. There is no training set for an algorithm.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The “study” proving the ONE-CUF meets its acceptance criteria is a compilation of non-clinical performance testing and biocompatibility testing, along with compliance with relevant voluntary standards. The documentation states:

    • Non-Clinical Testing: This included evaluations for Environmental Conditioning, Leak, Pressure Capacity, Mechanical Strength, Tensile Testing, and other performance aspects. While specific results are not detailed in the summary, these tests were submitted to support substantial equivalence.
    • Biocompatibility Testing: This was performed in accordance with ISO 10993-1:2018 ("Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process") and included Chemical Characterization, Cytotoxicity, Sensitization, and Irritation/Intracutaneous Reactivity.
    • Compliance with Voluntary Standards: The device was designed and tested for compliance with IEC 80601-2-30:2018 (for automated noninvasive sphygmomanometers) and ISO 81060-1:2007 (for non-automated sphygmomanometer test methods).
    • Human Factors Validation: Summative usability testing was performed.

    The conclusion is that based on these non-clinical performance data, biocompatibility results, and compliance with standards, the device is considered substantially equivalent to a legally marketed predicate device (SOFT-CUF K120125). No clinical or animal studies were required or performed to support this determination.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 11