510(k) Data Aggregation

    K Number
    K251998

    Validate with FDA (Live)

    Date Cleared
    2026-02-27

    (245 days)

    Product Code
    Regulation Number
    862.1110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica CH Diazo Total Bilirubin (D_TBil) assay is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates using the Atellica CH Analyzer. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).

    Device Description

    Atellica CH Diazo Total Bilirubin is a photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. A specific mixture of detergents enables the determination of the total bilirubin.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251543

    Validate with FDA (Live)

    Date Cleared
    2026-02-06

    (262 days)

    Product Code
    Regulation Number
    862.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica® IM TSH3‑Ultra II (TSH3ULII) assay is for in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in human serum and plasma (EDTA and lithium heparin) using the Atellica® IM Analyzer. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

    Device Description

    This assay is a third-generation assay that employs anti-FITC monoclonal antibody covalently bound to paramagnetic particles, an FITC-labeled anti-TSH capture mouse monoclonal antibody, and a tracer consisting of a proprietary acridinium ester and an anti‑TSH mouse monoclonal antibody conjugated to bovine serum albumin (BSA) for chemiluminescent detection.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251630

    Validate with FDA (Live)

    Date Cleared
    2026-01-05

    (222 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    50 - 120
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Atellica IM total PSA II (tPSAII) assay is for in vitro diagnostic use in the quantitative measurement of total prostate-specific antigen (PSA) in human serum and plasma (EDTA and lithium-heparin) using the Atellica IM Analyzer.

    This assay is indicated as an aid in the detection of prostate cancer in conjunction with a digital rectal exam (DRE) in men aged 50 years and older. Prostate biopsy is required for diagnosis of prostate cancer. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer.

    Device Description

    The Atellica IM total PSA II (tPSAII) assay consists of:

    tPSAII ReadyPack® primary reagent pack

    • Lite Reagent (10.0 mL/reagent pack): Unlabeled monoclonal mouse anti-fPSA antibody (~250 ng/mL); monoclonal mouse anti-PSA antibody (~180 ng/mL) labeled with acridinium ester; buffer; bovine serum albumin (BSA); preservative.
    • Solid Phase (20.0 mL/reagent pack): Monoclonal mouse anti-PSA antibody (~3.5 μg/mL) labeled with biotin and bound to streptavidin paramagnetic particles; buffer; BSA; bovine gamma globulin (BGG); sodium azide (< 0.1%); preservative.
      • Storage: Unopened at 2–8°C (Until expiration date on product), Onboard (42 days).

    tPSAII CAL (2.0 mL/vial): Purified PSA from human seminal fluid in buffer; BSA; NaN3 (< 0.1%).

    • Storage: Unopened at 2–8°C (Until expiration date on product), Opened at 2–8°C (30 days), On the system at room temperature (8 hours).

    The tPSAII assay will have two configurations: 100 tests kit and 500 tests kit (5 x 100T in the carton).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250816

    Validate with FDA (Live)

    Date Cleared
    2025-12-04

    (261 days)

    Product Code
    Regulation Number
    866.5870
    Age Range
    22 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur Anti-Thyroglobulin II (aTgII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroglobulin in human serum and plasma (EDTA, lithium heparin, sodium heparin) using the ADVIA Centaur XP system.

    Anti-thyroglobulin (aTg) measurements are used, in conjunction with clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and Graves' disease.

    Device Description
    ComponentVolumeIngredients
    ADVIA Centaur aTgII Primary Reagent ReadyPack (included in assay kit)
    ADVIA Centaur aTgII Lite Reagent10.0 mL/packHuman thyroglobulin labeled with acridinium ester (~1.2 µg/mL); buffered saline; bovine gamma globulin;
    ADVIA Centaur aTgII Solid Phase Reagent20.0 mL/packBiotinylated human thyroglobulin bound to streptavidin-coated paramagnetic particles (~0.6 mg/mL); buffered saline; bovine gamma globulin; BSA; sodium azide (< 0.1%); preservative
    ADVIA Centaur aTgII Ancillary Reagent ReadyPack (included in assay kit)
    ADVIA Centaur aTgII Ancillary Reagent17.5 mL/packGoat serum; mouse serum; sodium azide (< 0.1%); preservative
    ADVIA Centaur aTgII Calibrator (included in assay kit)
    ADVIA Centaur aTgII Low and High Calibrators1.0 mL/vial LyophilizedAfter reconstitution, low or high levels of monoclonal mouse anti-human thyroglobulin; goat serum; mouse serum; sodium azide (< 0.1%); preservative
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250250

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (262 days)

    Product Code
    Regulation Number
    866.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XP system.

    Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and/or Graves' disease.

    Device Description

    The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) consists of:

    • aTPOII ReadyPack® primary reagent pack (Lite Reagent, Solid Phase)
    • aTPOII CAL
      Devices sold separately and included in the ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) are:
    • ADVIA Centaur aTPOII MCM (MCM 1, MCM 2–4)
    • ADVIA Centaur aTPOII QC
    • ADVIA Centaur aTPOII DIL ReadyPack ancillary reagent pack
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242981

    Validate with FDA (Live)

    Date Cleared
    2025-06-20

    (267 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica IM Thyroglobulin (Tg) assay is for in vitro diagnostic use in the quantitative measurement of thyroglobulin in human serum and plasma (EDTA and lithium heparin) using the Atellica IM Analyzer.

    Thyroglobulin measurements are used as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.

    Device Description

    The Atellica IM Thyroglobulin (Tg) assay includes:

    • Tg ReadyPack primary reagent pack:
      • Lite Reagent: mouse monoclonal anti-human Tg antibody labeled with acridinium ester (~1.13 μg/mL); bovine serum albumin (BSA); mouse IgG; buffer; stabilizers; preservatives (7.5 mL/reagent pack).
      • Solid Phase: streptavidin-coated paramagnetic microparticles preformed with biotinylated mouse monoclonal antihuman Tg antibody (~267 μg/mL); BSA; mouse IgG; buffer; stabilizers; preservatives (15.0 mL/reagent pack).
    • Ancillary Well Reagent: BSA; bovine gamma globulin; buffer; preservatives (6.0 mL/reagent pack).
    • Tg CAL: After reconstitution, human thyroglobulin; BSA; buffer; stabilizers; preservatives (2.0 mL/vial).

    The following devices are sold separately:

    • Atellica IM Tg MCM:
      • MCM 1: After reconstitution, bovine serum albumin (BSA); buffer; stabilizers; preservatives (1.0 mL/vial).
      • MCM 2–5: After reconstitution, various levels of human thyroglobulin; BSA; buffer; stabilizers; preservatives (1.0 mL/vial).
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Atellica IM Thyroglobulin (Tg) assay:

    Device: Atellica IM Thyroglobulin (Tg) Assay
    Purpose: Quantitative measurement of thyroglobulin in human serum and plasma as an aid in monitoring differentiated thyroid cancer patients who have undergone thyroidectomy with or without radioiodine ablation.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes various performance characteristics, which serve as acceptance criteria for the device. The reported performance is directly from the summary.

    Acceptance Criteria CategorySpecific Acceptance Criteria (implicit from study design)Reported Device Performance
    Detection CapabilityLoB, LoD, LoQ determined per CLSI EP17-A2LoB: 0.039 ng/mL (0.059 pmol/L) LoD: 0.044 ng/mL (0.067 pmol/L) LoQ: 0.050 ng/mL (0.076 pmol/L)
    PrecisionPrecision determined per CLSI EP05-A3 (within-laboratory and repeatability)Repeatability (CV%): 1.2% - 6.4% across various concentrations Within-Laboratory Precision (CV%): 2.3% - 9.0% across various concentrations
    ReproducibilityReproducibility determined per CLSI EP05-A3 (across sites, runs, days)Reproducibility (CV%): 1.9% - 5.8% across various concentrations
    LinearityLinearity determined per CLSI EP06-ed2 within stated assay rangeLinear for 0.050–150 ng/mL (0.076–227 pmol/L)
    Specimen EquivalencePerformance equivalence across serum, EDTA plasma, lithium heparin plasmaPerformance confirmed equivalent across serum, EDTA plasma, lithium heparin plasma, and associated gel barrier tubes.
    Interferences (HIL)Bias < 10% for Hemoglobin, Bilirubin, Lipemia at specified concentrationsNo bias > 10% observed for tested HIL substances.
    Interferences (Other Substances)Bias < 10% for various common substances/medications/biomarkers at specified concentrationsNo bias > 10% observed for tested other substances.
    Cross-ReactivityCross-reactivity < 1.0% for specified substances (T3, T4, TSH, Galectin-3, T2)Cross-reactivity < 1.0% for tested substances.
    Reagent StabilityDefined on-board and reconstituted calibrator stability28 days on-board; Calibrators stable 45 days (2-8°C) / 60 days (≤ -20°C, thaw once).
    Sample StabilityDefined stability for various sample types and storage conditionsStable 3-4 days (2-8°C), 4 days (RT), 12-24 months (frozen); ≤ 4 freeze-thaw cycles.
    High Dose Hook EffectNo hook effect within a specified concentration rangeNo hook effect up to 80,000 ng/mL (121,200 pmol/L).
    Expected ValuesReference intervals established per CLSI EP28-A3cHealthy Adults: 2.44–74.9 ng/mL Post-thyroidectomy adults: < 1.27 ng/mL
    Clinical PerformanceSensitivity and specificity calculated by comparing assay results to structural disease (SD) at a defined cut-off (0.2 ng/mL). Confidences intervals for these parameters.Sensitivity: 98.2% (95% CI: 94.6%, 100.0%) Specificity: 53.4% (95% CI: 47.8%, 58.0%) PPV: 10.0% (95% CI: 8.7%, 11.2%) NPV: 99.8% (95% CI: 99.5%, 100.0%)

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Performance Test Set Sample Size: 291 serum samples collected from 189 subjects.
    • Data Provenance:
      • The document states "A prospective, multi-center study was conducted." This indicates prospective data collection across multiple sites.
      • The country of origin is not explicitly stated in the provided text.
      • All samples were from subjects diagnosed with differentiated thyroid cancer, 6 or more weeks following thyroidectomy or radioiodine ablation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • The document does not specify the number of experts or their qualifications for establishing the ground truth (structural disease). It simply states: "SD [Structural Disease] was established and classified as either positive or negative by cross-sectional or functional imaging results."
    • This suggests that the ground truth was derived from standard clinical imaging reports rather than a consensus of independent expert readers specifically for this study.

    4. Adjudication Method for the Test Set

    • The document does not describe an adjudication method for the test set's ground truth (structural disease). It implies that the imaging results themselves provided the classification. This means there was no adjudication process as typically seen with multiple human readers reviewing images.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done.
    • This study is for an in vitro diagnostic (IVD) assay (a lab test), not an AI-assisted imaging or diagnostic tool where human readers work with or without AI. The performance metrics presented are for the analytical and clinical performance of the assay itself, comparing its results to a ground truth (structural disease status), not to human reader performance or improvement with AI.

    6. If a Standalone Performance Study Was Done

    • Yes, this is effectively a standalone (algorithm only) performance study.
    • The Atellica IM Tg assay is an automated in vitro diagnostic device. Its performance characteristics (sensitivity, specificity, precision, linearity, etc.) are evaluated intrinsically, independent of human interpretation of the assay result values. The output is a quantitative measurement of thyroglobulin.

    7. The Type of Ground Truth Used

    • Ground truth for clinical performance: Structural disease (SD) status obtained from "cross-sectional or functional imaging results."
    • Ground truth for analytical performance (LoB, LoD, LoQ, Precision, etc.): Established through laboratory protocols and reference materials (e.g., CLSI guidelines, certified reference materials like BCR CRM 457, spiked samples, control materials).

    8. The Sample Size for the Training Set

    • The document does not specify a separate training set or its sample size for the Atellica IM Tg assay.
    • For IVD assays like this, the "training" is typically inherent in the assay's development and optimization process (e.g., reagent formulation, calibration curve development), which uses various known samples and standards, rather than a distinct, labeled "training dataset" as would be seen for a machine learning algorithm. The performance characteristics studies presented are akin to a "verification/validation set."

    9. How the Ground Truth for the Training Set Was Established

    • As a traditional IVD assay, there isn't a "training set" in the sense of a machine learning model.
    • Ground truth for assay development and calibration: This would have been established using reference materials (like BCR CRM 457), characterized control samples, and potentially a large panel of clinically characterized patient samples used during the assay's development and optimization phases. These activities are part of the broader product development lifecycle rather than a distinct "training set" with ground truth generated by experts in the context of a clinical study for submission. Standardization is explicitly noted as traceable to BCR CRM 457, which serves as a primary standard for establishing the quantitative accuracy of the assay.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243570

    Validate with FDA (Live)

    Date Cleared
    2025-04-25

    (158 days)

    Product Code
    Regulation Number
    862.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TSHL method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone (TSH, thyrotropin) in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of TSH are used in the diagnosis and monitoring of thyroid disease.

    The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.

    Device Description

    The Dimension® LOCI® Thyroid Stimulating Hormone Flex® reagent cartridge (TSHL) and Dimension® LOCI® Free Thyroxine Flex® reagent cartridge (FT4L) assays were cleared under K081074 and K073604, respectively. The components of the cleared assays were modified to reduce biotin interference.

    The modified Assays are comprised of the following components:

    Dimension® LOCI® Thyroid Stimulating Hormone Flex® reagent cartridge (TSHL): prepackaged liquid reagents in a plastic eight-well cartridge. Wells 1-2 contain Biotinylated TSH antibody (7.5 µg/mL mouse monoclonal), wells 3-4 contain TSH antibody coated Chemibeads (200 µg/mL mouse monoclonal), and wells 5-6 contain Streptavidin Sensibeads (1400 µg/mL recombinant E. coli). Wells 1-6 contain buffers, stabilizers and preservatives. Wells 7-8 are empty.

    Dimension® LOCI® Free Thyroxine Flex® reagent cartridge (FT4L): prepackaged liquid reagents in a plastic eight-well cartridge. Wells 1-2 contain Streptavidin Sensibeads (225 µg/mL recombinant E. coli), wells 3-4 contain T3 Chemibeads (200 µg/mL), and wells 5-6 contain FT4 Biotinylated antibody (50 ng/mL mouse monoclonal). Wells 1-6 contain buffers, stabilizers and preservatives. Wells 7-8 are empty.

    Test Principle: Both devices use a homogeneous chemiluminescent immunoassay based on LOCI® technology.
    For TSHL, it's a sandwich immunoassay where sample is incubated with biotinylated antibody and Chemibeads to form bead-TSH-biotinylated antibody sandwiches. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination at 680 nm generates singlet oxygen from Sensibeads which diffuses into Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 nm and is a direct function of TSH concentration.
    For FT4L, it's a sequential immunoassay where sample is incubated with biotinylated antibody. T3 Chemibeads are added and form bead/biotinylated antibody immunocomplexes with the non-saturated fraction of the biotinylated antibody. Sensibeads are then added and bind to the biotin to form bead pair immunocomplexes. Illumination at 680 nm generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 nm and is an inverse function of FT4 concentration.

    AI/ML Overview

    The document provided is a 510(k) clearance letter from the FDA for two in-vitro diagnostic (IVD) devices: Dimension® LOCI® Thyroid Stimulating Hormone Flex® reagent cartridge (TSHL) and Dimension® LOCI® Free Thyroxine Flex® reagent cartridge (FT4L). It describes the devices, their intended use, and the performance characteristics tested to demonstrate substantial equivalence to previously cleared predicate devices.

    However, it's crucial to understand that this document describes a reagent cartridge, which is a laboratory assay, not an AI/ML-driven device or an imaging device. Therefore, many of the requested criteria (e.g., sample size for training/test sets for AI, data provenance like country of origin for AI, ground truth establishment by experts, adjudication methods, MRMC studies, standalone AI performance) are not applicable to this type of device. The document details the performance of the assay itself in measuring biomarker concentrations, not an AI's ability to interpret images or assist human readers.

    I will interpret the request based on the information provided for this specific IVD device, noting where certain requested details are not relevant to the nature of the device.


    Acceptance Criteria and Study to Prove Device Meets Criteria (for an IVD Reagent Cartridge)

    The device in question, a reagent cartridge for quantitative measurement of TSH and FT4, is a laboratory assay, not an AI/ML or imaging interpretation device. Therefore, the "acceptance criteria" and "study" are focused on analytical performance characteristics (accuracy, precision, linearity, interference, detection limits, etc.) compared to a predicate device, rather than diagnostic accuracy metrics of an AI.

    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVD reagent cartridge, "acceptance criteria" are typically defined by ranges, limits, or statistical agreementsdemonstrating analytical performance comparable or superior to the predicate device and meeting relevant clinical or analytical standards (e.g., CLSI guidelines). The reported performance demonstrates that the modified devices meet these standards.

    Performance CharacteristicAcceptance Criteria (Implicit from CLSI Guidelines/Predicate Comparison)Reported Device Performance (TSHL)Reported Device Performance (FT4L)
    Detection LimitsMeet/Be comparable to predicate; within acceptable analytical ranges.LoB: 0.003 µIU/LLoD: 0.005 µIU/LLoQ: 0.007 µIU/LLoB: 0.03 ng/dLLoD: 0.05 ng/dLLoQ: 0.06 ng/dL
    Linearity / Measuring IntervalLinear across the claimed measuring range with acceptable bias.0.007 – 100 µIU/mL0.1 – 8.0 ng/dL
    Method Comparison (vs. Predicate)High correlation (r close to 1), slope close to 1, small y-intercept.N=145 Serum samplesy = 0.99x + 0.039 µIU/mL(Correlation (r) implicitly high, as regression equation suggests strong agreement)N=146 Serum samplesy = 1.02x + 0.03 ng/dL(Correlation (r) implicitly high, as regression equation suggests strong agreement)
    Precision (Repeatability)Within-run and total precision (SD/CV) within acceptable clinical laboratory limits.TSHL: Levels 0.110-88.676 µIU/mLWithin-Run %CV: 2.6-4.4%Total %CV: 1.1-3.0% (Note: Table 5 "Total" %CV for Level 1 is 2.6%, matching within-run %CV, but for others, it's lower. This might be a typo in the table, typically Total CV > Within-Run CV).FT4L:Levels 0.81-6.41 ng/dLWithin-Run %CV: 2.2-2.6%Total %CV: 0.9-1.1%
    Precision (Reproducibility)Total reproducibility (SD/CV) across lots and systems within acceptable clinical laboratory limits.TSHL:Levels 0.094-81.372 µIU/mLReproducibility %CV: 4.6-7.6%FT4L:Levels 0.70-6.49 ng/dLReproducibility %CV: 1.8-2.4%
    Recovery (Dilution)For TSHL, diluted samples should show recovery close to 100% of the true value.TSHL:Recovery ranged from 100% to 106% for various samples diluted 5x.N/A (FT4L not described for dilution recovery)
    Interference (Biotin)Modified assay shows significantly reduced interference compared to predicate.TSHL & FT4L: Specimens with biotin up to 1200 ng/mL demonstrate ≤10% change in results (significant improvement from predicate's 250 ng/mL for TSHL and 100 ng/mL for FT4L).TSHL & FT4L: Specimens with biotin up to 1200 ng/mL demonstrate ≤10% change in results.
    Reference Range VerificationResults from healthy samples confirm the established reference intervals.TSHL: Verified for adults (0.358-3.74 µIU/mL) and pediatric populations.FT4L: Verified for adults (0.76-1.46 ng/dL) and pediatric populations.
    Matrix ComparisonComparable performance across different sample matrices.Comparable values to serum samples for lithium heparin, sodium heparin, and K2-EDTA plasma.Same as TSHL.
    Hook EffectNo significant hook effect within specified range.No hook effect observed up to 30,000 µIU/mL.N/A (FT4L not described for hook effect)

    2. Sample Sizes and Data Provenance for the Test Set

    The concept of a "test set" in the context of an IVD reagent cartridge refers to the set of samples used for various analytical performance studies. These are not typically split into "training" and "test" sets as in AI/ML.

    • Method Comparison:
      • TSHL: 145 patient samples (serum)
      • FT4L: 146 patient samples (serum)
    • Precision (Repeatability): 5 serum samples (TSHL), 3 serum samples (FT4L)
    • Precision (Reproducibility): 5 serum samples (TSHL), 3 serum samples (FT4L)
    • Linearity: Low and high human serum pools used to create dilution series (TSHL: 12 levels, FT4L: 10 levels)
    • Interference (Biotin and HIL): Samples spiked with interferents, specific TSH/FT4 levels tested.
    • Dilution Recovery: 7 samples (TSHL)
    • Reference Range Verification: "Apparently healthy samples" (specific N not provided, but typically a statistically significant number for verification per CLSI EP28-A3C).
    • Matrix Comparison: Samples of various tube types (Serum, lithium heparin, sodium heparin, K2-EDTA plasma)

    Data Provenance: The document does not specify the country of origin of the patient samples. The studies are explicitly described as analytical performance studies rather than clinical outcome studies, and they are retrospective (samples tested in the lab, not followed prospectively).

    3. Number of Experts and Qualifications for Ground Truth

    This is not applicable as the device is a quantitative IVD assay (reagent cartridge), not an AI/ML device requiring expert interpretation of complex clinical data or images. The "ground truth" for this device is the actual concentration of TSH or FT4 in the sample, typically established either by:

    • Reference methods (e.g., mass spectrometry, although not explicitly stated as the ground truth method here).
    • The predicate device itself (as used in method comparison studies, where the predicate is the "comparison assay").
    • Spiking known concentrations into matrices.

    4. Adjudication Method for the Test Set

    This is not applicable for a quantitative IVD reagent. Adjudication methods (e.g., 2+1, 3+1) are typically used in scenarios where human experts interpret data (like medical images), and their disagreements need to be resolved to establish a definitive ground truth for AI model evaluation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. An MRMC study is designed to evaluate the diagnostic performance of human readers, often with and without AI assistance, on a set of cases. This device is a reagent cartridge that provides a quantitative measurement, not an AI that assists human interpretation.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)

    This is not applicable. This device is a reagent cartridge that runs on an automated system, providing a quantitative result. It's inherently "standalone" in providing the measurement, but it's not an "algorithm only" in the sense of an AI interpreting complex data. The performance metrics listed (precision, accuracy relative to predicate, linearity, etc.) are its "standalone" performance.

    7. Type of Ground Truth Used

    The "ground truth" for this type of quantitative diagnostic test is based on:

    • Comparison to a legally marketed predicate device: The current, FDA-cleared versions of the TSHL and FT4L assays (K081074 and K073604) acted as the "gold standard" or comparison method for the method comparison studies.
    • Known concentrations: For linearity, recovery, and interference studies, samples were prepared with known concentrations or spiked with known amounts of analytes or interferents.
    • Analytically verified samples: Samples used for precision studies have mean values derived from repeated measurements.

    8. Sample Size for the Training Set

    This is not applicable as the device is a non-AI/ML IVD reagent cartridge. There is no concept of a "training set" for this type of product. The development and optimization of the reagent formulation are internal processes, but they don't involve "training" a model on a dataset in the AI sense.

    9. How Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242685

    Validate with FDA (Live)

    Date Cleared
    2024-12-04

    (89 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica® CH Creatinine_3 (Crea3) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin, dipotassium EDTA, and sodium heparin), and urine using the Atellica® CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

    Device Description

    The Atellica CH Crea3 assay is based on the reaction of picrate with creatinine in an alkaline medium to produce a red chromophore creatinine picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Crea3 assay is a modification of the Jaffe method, using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because non-specific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 µmol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 µmol/L) from each result.

    AI/ML Overview

    The provided text describes the performance characteristics and studies for the Atellica® CH Creatinine_3 (Crea3) assay, a new in vitro diagnostic device for quantitative determination of creatinine. It compares this new device to a predicate device, the Atellica® CH Creatinine_2 (Crea_2) assay.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: The document focuses on establishing substantial equivalence for an in vitro diagnostic (IVD) test, which primarily relies on analytical performance characteristics rather than clinical outcome studies or multi-reader multi-case (MRMC) comparative effectiveness studies typically seen with imaging AI devices. Therefore, some of your requested information (like number of experts for ground truth, adjudication methods, MRMC studies, and training set details for an AI model) are not directly applicable or provided in this type of submission.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are established through various analytical performance studies, primarily comparing it to a legally marketed predicate device (Atellica® CH Creatinine_2). The acceptance criteria are implicitly defined by the successful demonstration of equivalence or meeting pre-defined performance goals for each characteristic.

    Here's a table summarizing the acceptance criteria (inferred from the "designed to have" or "determined in accordance with" statements and the reported results meeting these) and the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Atellica® CH Creatinine_3 (Crea3))
    Detection CapabilityLoB: $\le$ LoD for serum and urine samples. LoD: $\le$ 0.15 mg/dL for serum/plasma; $\le$ 3.00 mg/dL for urine. LoQ: $\le$ 0.15 mg/dL for serum/plasma with $\le$ 0.10 mg/dL total analytical error; $\le$ 3.00 mg/dL for urine with $\le$ 1.50 mg/dL total analytical error.Serum/plasma: LoB: 0.05 mg/dL LoD: 0.10 mg/dL LoQ: 0.15 mg/dL Urine: LoB: 0.50 mg/dL LoD: 1.00 mg/dL LoQ: 3.00 mg/dL (All results meet the stated design goals/acceptance criteria).
    PrecisionDetermined in accordance with CLSI Document EP05-A3 (indicates adherence to specific statistical targets for repeatability and within-lab precision, implicitly accepted if within CLSI guidelines for the assay's use).Serum Samples (n=80 each): - Serum 1 (0.38 mg/dL): Repeatability SD 0.006, CV 1.6%; Within-Lab SD 0.012, CV 3.2% - Serum 2 (0.73 mg/dL): Repeatability SD 0.023, CV 3.2%; Within-Lab SD 0.029, CV 4.0% - Serum 3 (0.73 mg/dL): Repeatability SD 0.006, CV 0.8%; Within-Lab SD 0.019, CV 2.6% - Serum 4 (1.18 mg/dL): Repeatability SD 0.007, CV 0.6%; Within-Lab SD 0.019, CV 1.6% - Serum QC 1 (1.85 mg/dL): Repeatability SD 0.007, CV 0.4%; Within-Lab SD 0.024, CV 1.3% - Serum QC 2 (6.21 mg/dL): Repeatability SD 0.011, CV 0.2%; Within-Lab SD 0.067, CV 1.1% - Serum 5 (17.39 mg/dL): Repeatability SD 0.035, CV 0.2%; Within-Lab SD 0.189, CV 1.1% - Serum 6 (28.54 mg/dL): Repeatability SD 0.056, CV 0.2%; Within-Lab SD 0.317, CV 1.1% Urine Samples (n=80 each): - Urine 1 (56.74 mg/dL): Repeatability SD 0.102, CV 0.2%; Within-Lab SD 0.746, CV 1.3% - Urine 2 (135.80 mg/dL): Repeatability SD 0.206, CV 0.2%; Within-Lab SD 1.601, CV 1.2% - Urine QC 1 (195.79 mg/dL): Repeatability SD 0.253, CV 0.1%; Within-Lab SD 2.376, CV 1.2% (All results demonstrate low CVs, indicating good precision).
    ReproducibilityDetermined in accordance with CLSI Document EP05-A3 (implies meeting specific statistical targets for variability components across different days, lots, and instruments).Serum Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.0% to 5.0%. Urine Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.4% to 1.6%. (All results demonstrate good reproducibility across conditions).
    Assay ComparisonSerum: Correlation coefficient $\ge$ 0.950 and slope of 1.00 $\pm$ 0.05, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression. Urine: Correlation coefficient $\ge$ 0.950 and slope of 0.000 $\pm$ 3.00, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression.Serum (n=151): Regression equation y = 1.00x - 0.04 mg/dL, correlation coefficient (r) = 1.000. Sample range 0.44 to 28.64 mg/dL. Urine (n=113): Regression equation y = 1.00x + 0.14 mg/dL, correlation coefficient (r) = 1.000. Sample range 12.60 to 237.06 mg/dL. (Both serum and urine results meet the acceptance criteria for correlation and slope).
    Specimen EquivalenceDetermined using Weighted Deming regression (implicitly, the regression line should demonstrate equivalence, i.e., close to y=x, with high correlation coefficient).Sodium Heparin (n=50): y = 1.00x + 0.00 mg/dL, r=0.999. Lithium Heparin (n=50): y = 0.99x + 0.06 mg/dL, r=0.999. Dipotassium EDTA (n=50): y = 0.98x + 0.04 mg/dL, r=0.998. (All demonstrate strong equivalence to serum reference).
    Interferences (HIL)$\le$ 10% interference from hemoglobin, bilirubin, and lipemia. Bias > 10% or 0.15 mg/dL (whichever is greater for serum/plasma) is considered interference.Reported biases for Hemoglobin (1000 mg/dL), Conjugated Bilirubin (40-45 mg/dL), Unconjugated Bilirubin (45-60 mg/dL), and Lipemia (2250-3000 mg/dL) are all within the $\pm$10% or $\pm$0.15 mg/dL threshold for the tested analyte concentrations, demonstrating acceptable interference profiles.
    Interfering SubstancesBias $\le$ 10% or $\pm$0.15 mg/dL for Serum/plasma samples. Bias $\le$ 10% for Urine samples (for listed substances).Most tested substances (e.g., Acetaminophen, Ascorbic Acid, etc.) show negligible bias, meeting the criteria. Substances showing bias beyond acceptance criteria for Serum: - Cefoxitin: Significant interference (e.g., 243.6% and 947.9% bias at high concentrations). - Cephalothin: Shows significant bias (e.g., 44.0% bias at 180 mg/dL). - Glucose: Shows bias beyond 10% at higher concentrations (e.g., 11.5% at 500 mg/dL and 22.5% at 1000 mg/dL). - Total Protein: Shows bias beyond 0.15 mg/dL at 15 g/dL (0.45 mg/dL). - Acetohexamide: Shows bias beyond 10% at 2.0 mg/dL (10.4%). - Hydroxocobalamin (Cyanokit): Shows significant bias (e.g., 14.5% and 49.3% at higher concentrations). Substances showing bias beyond acceptance criteria for Urine: - Cefoxitin: Shows bias beyond 10% at higher concentrations (e.g., 11.3% and 15.4%). (The document explicitly lists these substances under "Interference beyond $\pm$10% for Serum" and "Interference beyond $\pm$10% for Urine," indicating that they failed the non-interference criteria at the tested concentrations. This is typical for IVD submissions, where known interferences are identified for labeling purposes).
    StandardizationThe assay shall be traceable to the reference material SRM967, from the National Institute of Standards and Technology (NIST).Statement confirms the assay is traceable to NIST SRM967.

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set Sample Sizes:
        • Detection Capability: Not explicitly stated as "sample size" but data points obtained according to CLSI EP17-A2.
        • Precision: 80 data points per serum/urine sample type (duplicate runs for 20 days, 2 runs/day).
        • Reproducibility: 225 data points per serum/urine sample type (n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots).
        • Assay Comparison: 151 serum samples and 113 urine samples.
        • Specimen Equivalence: 50 samples for each plasma type (Sodium Heparin, Lithium Heparin, Dipotassium EDTA) compared to serum.
        • Interference (HIL & Non-Interfering Substances): Not explicitly stated as a total sample size, but experiments are designed to test specific analyte concentrations with and without interferents, following CLSI EP07-ED3.
      • Data Provenance: Not explicitly stated in terms of country of origin. Given the manufacturer (Siemens Healthcare Diagnostics Inc. in Tarrytown, New York, USA) and FDA submission, it's highly probable the studies were conducted in the US or in compliance with US regulatory standards. The studies described are retrospective in the sense that they use pre-collected or prepared samples to assess the analytical performance of the device under controlled conditions, not prospective in tracking patient outcomes in a clinical trial.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • For an in vitro diagnostic (IVD) device measuring a quantitative analyte like creatinine, "ground truth" is typically established by reference methods or established laboratory standards and calibrators, not by human expert consensus or labeling of medical images.
      • The "ground truth" for creatinine concentration in this context is based on traceable reference materials (NIST SRM 967) and established laboratory measurement principles, and the performance is compared against a legally marketed predicate device.
      • Therefore, this question (relevant for AI/imaging devices) does not directly apply to this type of IVD submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical trials or image labeling pipelines where there's human interpretation involved and a need to resolve disagreements among multiple readers; this is an analytical performance study of an IVD assay.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/imaging device. It's an in vitro diagnostic assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is an automated IVD assay performed on the Atellica® CH Analyzer. Its intended use is quantitative determination of creatinine. Therefore, the performance described (precision, accuracy, interference, etc.) is its standalone performance without a human in the loop for the analytical measurement itself, though a human still interacts with the instrument and interprets the results in a clinical context.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary "ground truth" for the Atellica® CH Creatinine_3 assay's performance is traceability to NIST SRM 967 (a certified reference material for creatinine) and comparison to a legally marketed predicate device (Atellica® CH Creatinine_2) using method comparison validated against CLSI guidelines. This is a form of analytical reference standard and comparative performance to an established method.
    7. The sample size for the training set:

      • This device is an analytical chemistry assay, not a machine learning/AI algorithm that requires a "training set" in the computational sense. The "development" or "optimization" of the assay would involve various experimental data, but it's not codified as a "training set" for an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the AI/ML context for this type of device. The assay development would rely on scientific principles of analytical chemistry, reagent formulation, and instrument calibration against known standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240315

    Validate with FDA (Live)

    Date Cleared
    2024-10-10

    (251 days)

    Product Code
    Regulation Number
    864.7295
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INNOVANCE Anti-Xa assay in combination with INNOVANCE Heparin Calibrator is an In-vitro diagnostic automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of UFH and LMWH can be performed on the BCS XP System, CS-2500 System, CS-5100 System and the CA-660 System. For use with plasma from patients undergoing anticoagulant therapy with either UFH or LMWH.

    INNOVANCE Anti-Xa assay in combination with INNOVANCE Apixaban Standard provides quantitative determination of the concentration of apixaban in human plasma collected from venous blood samples in 3.2 % sodium citrated tubes in the clinical laboratory. The quantitative determination of apixaban can be performed on CS-2500 System. For use with plasma from patients undergoing anticoagulant therapy with apixaban in situations where quantification of apixaban levels may be indicated:

    · Patient with bleeding,

    · Patient with risk for bleeding (e.g. during perioperative management),

    · Patient with conditions affecting pharmacokinetics (e.g. deteriorating renal function, extremes of body weight, treatment with other drugs known to affect pharmacokinetics of apixaban).

    The performance of this device has not been established in neonate and pediatric patient populations.

    Device Description

    INNOVANCE Anti-Xa assay is a one stage chromogenic assay. The reagent kit consists of two components. One component (Reagent) contains Xa. the other (Substrate) a chromogenic substrate specific for Xa. Upon mixing of INNOVANCE Anti-Xa Reagent and INNOVANCE Anti-Xa Substrate, Xa converts the chromogenic substrate into two products, one of them is paranitroaniline. The formation of paranitroaniline can be quantified by the coaqulation analyzer employing light absorption at a specific wavelength (405 nm).

    In the presence of a heparin containing sample the formation of paranitroaniline will be reduced in a time dependent manner. This is due to inhibition of Xa by the heparin/AT complex is formed in the patient's plasma and competes with the substrate conversion by Xa. The concentration of the complex is not only dependent on the concentration of heparin but also on the availability of the patient's endogenous antithrombin. By comparison to a reference curve the heparin activity of the sample can be quantified.

    To reduce the influence from heparin antagonists, such as platelet factor 4 (PF4), dextran sulfate is included in the reaction mixture.

    In the presence of an apixaban containing sample factor Xa is inhibited directly by this inhibitor. Comparison to an inhibitor specific reference curve allows quantification of the inhibitor concentration in the sample.

    AI/ML Overview

    The provided text describes the performance data for the INNOVANCE Anti-Xa assay, a device for quantitative determination of anticoagulants. As the request is about acceptance criteria and study proving it, and this document is an FDA 510(k) summary, specific acceptance criteria would be internal to the manufacturer's validation plan and not explicitly stated as "acceptance criteria" in this public summary. However, the performance data presented (Precision, Reproducibility, LoB, LoD, LoQ, Linearity, Measuring Interval, Interference, and Method Comparison) implicitly demonstrate that the device meets the necessary performance characteristics for its intended use, making it "substantially equivalent" to a predicate device.

    Here's an attempt to extract and infer the information based on the provided document:

    1. Table of Acceptance Criteria (Inferred from Performance Data) and Reported Device Performance

    Since explicit acceptance criteria are not stated as pass/fail thresholds in this summary, they are inferred based on the presented results which led to substantial equivalence. The predicate device's performance often sets the benchmark for substantial equivalence.

    Performance MetricImplied Acceptance Criteria (Inferred from Industry Standards/Predicate)Reported Device Performance (INNOVANCE Anti-Xa)
    Precision (Within-Device/Lab CV%)Generally low CV% (e.g., < 10% for clinical assays)Plasma Pool 1: CV% not reported, SD = 2.0 ng/mLPlasma Pool 2: CV% not reported, SD = 1.1 ng/mLINNOVANCE Apixaban Control 1: 2.48%Plasma Pool 3: 2.03%INNOVANCE Apixaban Control 2: 2.57%Plasma Pool 3: 2.09%
    Reproducibility (Combined sites CV%)Generally low CV% across sites< 65 ng/mL: ≤ 4.6 CV% (SD) / CV% not reported≥ 65 ng/mL: ≤ 6.68%
    Limit of Blank (LoB)Very low, ideally close to 0 ng/mL1.0 ng/mL
    Limit of Detection (LoD)Low enough to detect clinically relevant concentrations3.3 ng/mL
    Method LinearityAcceptable linearity across the measuring rangeLinear over 10.5 ng/mL to 378 ng/mL
    Measuring Interval (Range)Clinically appropriate range for therapeutic monitoring20 to 350 ng/mL (extendable up to 700 ng/mL with dilution)
    InterferenceNo significant interference from common endogenous/exogenous substances at specified concentrationsNo interference observed up to stated concentrations for Bilirubin, Hemoglobin, Lipids, Danaparoid sodium, Edoxaban, Fondaparinux, LMWH, UFH, Rivaroxaban, Atorvastatin, Ticagrelor, Acetylsalicylic acid, Warfarin, Isosorbide dinitrate.
    Method Comparison (Correlation (r) and Regression equation)High correlation (e.g., r > 0.95) and close to y=x for agreement with predicate.Overall (n=301): r = 0.989, y = 1.026x - 8.590 ng/mL (across 3 sites)

    2. Sample Size and Data Provenance:

    • Test Set Sample Size:
      • Precision: 240 measurements for each of the 6 samples (INNOVANCE Apixaban Control 1 & 2, and 4 plasma pools).
      • LoB, LoD, LoQ: Not explicitly stated as a single number, but calculation involves measurements of multiple samples (e.g., "samples with an assigned value of 0 ng/mL apixaban" for LoB; "samples with an assigned value of 2, 4, 6, 8 and 10 ng/mL apixaban" for LoD; "samples with an assigned value of 16, 18, 20, 22 and 24 ng/mL apixaban" for LoQ).
      • Linearity & Measuring Interval: Not explicitly stated as a single number, but likely involves multiple spike levels and measurements.
      • Interference: Not explicitly stated, but involves testing various interfering substances.
      • Method Comparison: 301 fresh and frozen plasma samples (multicentric study across 3 sites).
    • Data Provenance: The document does not specify the country of origin for the data. It refers to "internally conducted studies" and a "multicentric study" for reproducibility and method comparison, implying clinical laboratories. Since the applicant is Siemens Healthcare Diagnostics Products GmbH, Marburg, Germany, it's highly probable that some or all studies were conducted in Germany or other European countries, though this is not explicitly stated. The studies are described as conforming to CLSI (Clinical and Laboratory Standards Institute) guidelines, which are international standards. The samples for method comparison were "fresh and frozen plasma samples." It is a prospective study in terms of testing the device performance against established samples/methods.

    3. Number of Experts and Qualifications for Ground Truth:

    This device is an in-vitro diagnostic assay that measures the concentration of substances (heparin, apixaban) in blood plasma directly. Therefore, the "ground truth" is typically defined by:

    • Reference Methods: For method comparison, it refers to the results obtained from the predicate device (HemosIL Liquid Anti-Xa) on the ACL TOP system.
    • Assigned Values: For calibrators and controls, the "assigned value" is based on the manufacturer's preparation and testing, often traceable to internal reference preparations or qualified methods like LC-MS/MS for substances like apixaban, as stated in the traceability sections for calibrators and controls.

    There is no mention of human "experts" (e.g., radiologists, pathologists) establishing ground truth, as this is a quantitative laboratory assay, not an imaging device requiring expert interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable. This is a quantitative laboratory assay, not a device where interpretation or human reading requires adjudication. The ground truth for method comparison is the measurement from the predicate device (HemosIL Liquid Anti-Xa).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is an in-vitro diagnostic device for quantitative chemical analysis, not an imaging device or AI algorithm with human-in-the-loop performance component.

    6. Standalone (Algorithm Only) Performance:

    This device is a standalone diagnostic assay (reagent kit used with specific automated analyzers). Its performance data, as detailed, represents its standalone performance. There is no separate "algorithm" in the sense of AI that needs to be evaluated independently of the assay system.

    7. Type of Ground Truth Used:

    • Reference Method Comparison: For method comparison, the ground truth was established by the predicate device (HemosIL Liquid Anti-Xa), which is a legally marketed device, providing a comparative ground truth.
    • Internal Reference/Assigned Values: For precision, LoB, LoD, LoQ, linearity, and interference studies, the ground truth for samples (e.g., controls, plasma pools, spiked samples) was established by their assigned values, internal reference preparations, or spiked concentrations. The traceability section explicitly mentions that for apixaban, calibration values are "traceable to apixaban supplied by the manufacturer and quantitated in plasmas assayed by Liquid Chromatography - tandem Mass Spectrometry (LC-MS/MS)" for the predicate, and for the subject device, "calibrated against an internal reference preparation."

    8. Sample Size for the Training Set:

    This document describes the analytical validation of a diagnostic assay (a chemical reagent system) rather than an AI/ML algorithm. Therefore, the concept of a "training set" for an algorithm doesn't directly apply in the traditional sense. The development of such assays involves extensive R&D, formulation, and preliminary testing, but these are not referred to as "training sets" in the context of typical regulatory submissions for IVDs. The closest equivalent would be the R&D samples used during the development and optimization phase of the reagent and its performance on the instruments. No specific numbers are provided for this.

    9. How the Ground Truth for the Training Set was Established:

    As above, the concept of a "training set" is not relevant here. For the analytical studies presented, "ground truth" (or reference values) for various samples (calibrators, controls, patient samples, spiked samples) were established through recognized methods:

    • Predicate Device Measurements: For comparative studies, the measurements from the legally marketed predicate device served as a comparative reference.
    • LC-MS/MS (Liquid Chromatography - tandem Mass Spectrometry): This highly accurate analytical technique is explicitly mentioned for establishing traceability and quantifying apixaban in calibrators for the predicate device, and the subject device uses an "internal reference preparation" which would likely also be validated against similar high-accuracy methods.
    • Gravimetric/Volumetric Spiking: For linearity, LoD, and LoQ studies, samples are often prepared by precisely adding known concentrations of the analyte (apixaban) to a matrix, making the spiked concentration the ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241800

    Validate with FDA (Live)

    Date Cleared
    2024-07-26

    (35 days)

    Product Code
    Regulation Number
    862.1475
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview