K Number
K250250

Validate with FDA (Live)

Date Cleared
2025-10-17

(262 days)

Product Code
Regulation Number
866.5870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XP system.

Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and/or Graves' disease.

Device Description

The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) consists of:

  • aTPOII ReadyPack® primary reagent pack (Lite Reagent, Solid Phase)
  • aTPOII CAL
    Devices sold separately and included in the ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) are:
  • ADVIA Centaur aTPOII MCM (MCM 1, MCM 2–4)
  • ADVIA Centaur aTPOII QC
  • ADVIA Centaur aTPOII DIL ReadyPack ancillary reagent pack
AI/ML Overview

N/A

FDA 510(k) Clearance Letter - ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

October 17, 2025

Siemens Healthcare Diagnostics, Inc.
Mey Vasquez
Regulatory Affairs Professional
511 Benedict Avenue
Tarrytown, New York 10591

Re: K250250
Trade/Device Name: ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII)
Regulation Number: 21 CFR 866.5870
Regulation Name: Thyroid Autoantibody Immunological Test System
Regulatory Class: Class II
Product Code: JZO
Dated: September 15, 2025
Received: September 16, 2025

Dear Mey Vasquez:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250250 - Mey Vasquez Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250250 - Mey Vasquez Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying Mao -S

Ying Mao, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250250

Device Name: ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII)

Indications for Use (Describe)

The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XP system.

Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and/or Graves' disease.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Siemens Healthcare Diagnostics 510(k) Summary

Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 1 of 14
October 2025

510(k) Summary of Safety and Effectiveness

Introduction: According to the requirements of SMDA 1990 and 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.

The assigned 510(k) Number: K250250

1. Date Prepared

October 17, 2025

2. Applicant Information

Siemens Healthcare Diagnostics Inc.
511 Benedict Avenue,
Tarrytown, NY 10591 USA

Contact: Mey Vasquez
Regulatory Affairs Professional
E-mail: mey.vasquez@siemens-healthineers.com

3. Regulatory Information

Assay
Trade NameADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII)
DeviceThyroid autoantibody immunological test system
DefinitionA thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain autoimmune thyroid disorders.
FDA ClassificationClass II
Review PanelImmunology
Product CodeJZO

Page 6

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 2 of 14
October 2025

Regulation Number21 CFR 866.5870

4. PREDICATE DEVICE

Name of Device: Architect Anti-TPO
510(k): K052407

5. DEVICE DESCRIPTION

The following devices are included in the ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII):

Material Description
aTPOII ReadyPack® primary reagent pack
Lite Reagent
10.0 mL/reagent pack
Recombinant TPO (~90 ng/mL) complexed with mouse monoclonal anti-TPO antibody (~30 ng/mL) labeled with acridinium ester in phosphate buffer; blocker (bovine and mouse); surfactant; sodium azide (< 0.1%); preservatives
Solid Phase
20.0 mL/reagent pack
Streptavidin-coated paramagnetic microparticles (~0.3 mg/mL) with biotinylated mouse monoclonal anti-TPO antibody (~6 μg/mL) in phosphate buffer; blocker (bovine and mouse); surfactant; sodium azide (< 0.1%); preservatives Tg ReadyPack® primary reagent pack
aTPOII CAL
1.0 mL/vial
After reconstitution, low and high levels of anti-TPO (human) in defibrinated human plasma; sodium azide (< 0.1%); preservatives

Page 7

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 3 of 14
October 2025

The following devices are sold separately:

Material Description
ADVIA Centaur aTPOII MCM:
MCM 1:
1.0 mL/vial
After reconstitution, defibrinated human plasma; sodium azide (< 0.1%); preservatives
MCM 2–4:
1.0 mL/vial
After reconstitution, various levels of anti-TPO (human) in defibrinated human plasma; sodium azide (< 0.1%); preservatives
ADVIA Centaur aTPOII QC
2.0 mL/vial
After reconstitution, low and high levels of anti-TPO (human) in defibrinated human plasma; sodium azide (< 0.1%); preservatives
ADVIA Centaur aTPOII DIL ReadyPack ancillary reagent pack
2 x 5.0 mL/reagent pack
Defibrinated human plasma negative for anti-TPO (human); sodium azide (< 0.1%); preservatives

6. INTENDED USE/ INDICATIONS FOR USE

The ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system.

Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and Graves' Disease.

7. Special Conditions for Use Statement

For Prescription Use

8. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Page 8

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 4 of 14
October 2025

ItemCandidate DevicePredicate
ADVIA Centaur aTPOIIArchitect Anti-TPO (K052407)
Intended Use/ Indications for UseThe ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system. Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and Graves' Disease.Architect® Anti-TPO is a chemiluminescent microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma (EDTA and Heparin) on the Architect® i System. The Architect® Anti-TPO assay is intended for use as an aid in the diagnosis of autoimmune thyroid disease.
InstrumentADVIA Centaur XP SystemARCHITECT® i System
Assay Range4.6–400.0 IU/mL0.5 to 1000.0 IU/mL
LoB2.3 IU/mL0.159 IU/mL
LoD4.6 IU/mLNone
LoQ4.6 IU/mLNone
MeasurementQuantitativeSame
TechnologyChemiluminescentChemiluminescent microparticle (CMIA)
Operating PrincipleFully automated 1-step competitive immunoassayChemiluminescent microparticle immunoassay (CMIA) referred as Chemiflex
Sample typeSerum, EDTA plasma, Lithium Heparin PlasmaSame

Page 9

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 5 of 14
October 2025

ItemCandidate DevicePredicate
ADVIA Centaur aTPOIIArchitect Anti-TPO (K052407)
StandardizationThe assay standardization is traceable to the NIBSC (National Institute for Biological Standards and Control) 66/387 Research Standard, Human Anti-Thyroid Microsome Serum. Assigned values for calibrators and controls are traceable to this standardization.The anti-TPO reference calibrator is traceable to the NIBSC 66/387 reference preparation.
Calibration2 levels6 levels
Clinical Cut-Off10.4 IU/mL5.6 IU/mL
Sample Volume30 µLThe minimum sample volume is calculated by the system.

9. PERFORMANCE CHARACTERISTICS DATA

9.1. Detection Capability

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2.

The ADVIA Centaur aTPOII assay has a LoB of 2.3 IU/mL, a LoD of 4.6 IU/mL, and a LoQ of 4.6 IU/mL.

9.2. Precision

Precision was determined using the ADVIA Centaur XP system in accordance with CLSI document EP05‑A3. Samples were assayed in replicates with 2 runs per day using a 20‑day protocol. The following results are representative of the performance of the assay:

SampleN^aMean (IU/mL)RepeatabilityWithin-Laboratory Precision
SD^b (IU/mL)CV^c (%)SD (IU/mL)CV (%)
Serum A806.00.569.30.8614.2

Page 10

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 6 of 14
October 2025

SampleN^aMean (IU/mL)RepeatabilityWithin-Laboratory Precision
SD^b (IU/mL)CV^c (%)SD (IU/mL)CV (%)
Serum B8012.81.179.11.189.2
Serum C8017.11.7610.31.9711.5
Serum D8026.21.676.41.847.0
Serum E80124.45.734.66.215.0
Serum F80170.08.995.39.745.7
Serum G80362.422.126.122.126.1
Control 18025.31.355.31.756.9
Control 280175.48.514.911.996.8

^a Number of measurements.
^b Standard deviation.
^c Coefficient of variation.

9.3. Reproducibility

Reproducibility was determined using the ADVIA Centaur XP system in accordance with CLSI Document EP05‑A3. Testing was performed using 3 instruments and 3 reagent lots. Samples were assayed in replicates of 5 with 1 run per day using a 5‑day protocol. (Number of measurements per sample = 225.) The following results are representative of the performance of the assay:

SampleMean (IU/mL)RepeatabilityBetween DayBetween InstrumentBetween LotReproducibility
SD^a (IU/mL)CV^b (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)
Serum A5.20.7013.50.6813.10.000.00.305.81.0219.6
Serum B10.20.797.80.686.70.898.80.222.21.3913.7
Serum C16.11.147.10.855.31.167.20.422.61.8811.7
Serum D23.41.566.70.954.10.512.20.411.81.948.3
Serum E28.91.876.50.923.20.602.12.167.53.0710.6
Serum F131.96.204.73.162.42.331.811.238.513.4110.2
Serum G172.09.385.54.252.53.642.113.417.817.3010.1
Serum H358.617.965.015.564.39.982.827.587.737.7510.5

Page 11

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 7 of 14
October 2025

SampleMean (IU/mL)RepeatabilityBetween DayBetween InstrumentBetween LotReproducibility
SD^a (IU/mL)CV^b (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)
Control 128.71.766.11.184.10.772.70.411.42.298.0
Control 2201.08.064.05.582.81.460.73.711.810.585.3

^a Standard deviation.
^b Coefficient of variation.

9.4. Linearity

Linearity testing was performed in accordance with CLSI Document EP06‑ed2. The assay is linear for the measuring interval of 4.6–400.0 IU/mL.

9.5. Specimen Equivalence

Specimen equivalency was determined with the Passing-Bablok regression model using the ADVIA Centaur XP system in accordance with CLSI Document EP09c‑Ed3. Agreement of the specimen types may vary depending on the study design and population tested.

Tube (y) vs. Serum (x)Regression EquationSample IntervalN^ar^b
Plasma, dipotassium EDTAy = 0.99x - 0.87 IU/mL6.40–380.9 IU/mL560.975
Plasma, lithium hepariny = 0.94x + 0.1 IU/mL6.40–398.0 IU/mL560.987

^a Number of samples tested.
^b Correlation coefficient.

9.6. Interferences

Hemolysis, Icterus, Lipemia (HIL)

Interference testing was performed using the ADVIA Centaur XP system in accordance with CLSI Document EP07‑Ed3. The following substances do not interfere with the assay when present in serum at the concentrations indicated. Bias due to these substances does not exceed ± 4.0 IU/mL at anti-TPO concentrations of 10.4–40.0 IU/mL or ± 10.0% at anti-TPO concentrations of 130.0–270.0 IU/mL.

SubstanceSubstance Test Concentration
Hemoglobin1000 mg/dL (10 g/L)
Bilirubin, conjugated60 mg/dL (712 µmol/L)

Page 12

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 8 of 14
October 2025

SubstanceSubstance Test Concentration
Bilirubin, unconjugated60 mg/dL (1026 µmol/L)
Lipemia (Intralipid)3500 mg/dL (35 g/L)

Other Substances

Interference testing was performed using the ADVIA Centaur XP system in accordance with CLSI Document EP07‑Ed3. The following substances do not interfere with the assay when present in serum at the concentrations indicated. Bias due to these substances does not exceed ± 4.0 IU/mL at anti-TPO concentrations of 10.4–40.0 IU/mL or ± 10.0% at anti-TPO concentrations of 130.0–270.0 IU/mL.

SubstanceSubstance Test ConcentrationSubstanceSubstance Test Concentration
Acetaminophen20 mg/dL (1323 µmol/L)Liothyronine0.0075 mg/dL (0.115 µmol/L)
Acetylcysteine15 mg/dL (919.1 µmol/L)Losartan potassium0.2 mg/dL (4.73 µmol/L)
Acetylsalicylic acid (aspirin)65 mg/dL (3608 µmol/L)Methimazole8 mg/dL (700.7 µmol/L)
Ampicillin sodium7.5 mg/dL (201.9 µmol/L)Methyldopa2.25 mg/dL (106.5 µmol/L)
Ascorbic acid5.25 mg/dL (298.1 µmol/L)Methylprednisolone0.783 mg/dL (20.9 µmol/L)
Atorvastatin calcium0.075 mg/dL (1.34 µmol/L)Metoprolol tartrate0.15 mg/dL (5.61 µmol/L)
Biotin3500 ng/mL (14.3 µmol/L)Metronidazole12.3 mg/dL (702.9 µmol/L)
Carbimazole3 mg/dL (161.1 µmol/L)Octreotide0.03 mg/dL (0.294 µmol/L)
Cefoxitin660 mg/dL (15.44 mmol/L)Omeprazole0.84 mg/dL (24.32 µmol/L)
Cholesterol400 mg/dL (10.4 mmol/L)Phenylbutazone32.1 mg/dL (1.04 mmol/L)
Cyclosporine0.18 mg/dL (1.5 µmol/L)Prednisone9.9 µg/dL (0.276 µmol/L)
Dexamethasone1.2 mg/dL (30.6 µmol/L)Propranolol Hydrochloride24 mg/dL (811.4 µmol/L)
Doxycycline1.8 mg/dL (40.5 µmol/L)Propylthiouracil30 mg/dL (1.76 mmol/L)
EDTA, tripotassium425 mg/dL (9.60 mmol/L)Rheumatoid factor (RF)750 IU/mL
Human anti-mouse antibody (HAMA)67 ng/mL (67 µg/L)Rifampicin4.8 mg/dL (58.3 µmol/L)
Ibuprofen50 mg/dL (2424 µmol/L)Silwet L 720 (Octamethylcyclotetrasiloxane)0.03 mg/mL (101 µmol/L)
Immunoglobulin G (IgG)6 g/dL (60 g/L)Theophylline6 mg/dL (333.0 µmol/L)

Page 13

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 9 of 14
October 2025

SubstanceSubstance Test ConcentrationSubstanceSubstance Test Concentration
Iodide38 mg/dL (2.99 mmol/L)Thyroxine1000 µg/mL (1.29 mmol/L)
Levodopa0.75 mg/dL (38.0 µmol/L)Total Protein12 g/dL (120 g/L)

9.7. Cross-Reactivity

Cross-reactivity was determined using the ADVIA Centaur XP system in accordance with CLSI Document EP07‑Ed3. Cross-reactivity of samples spiked with various substances does not exceed ± 4.0 IU/mL at anti-TPO concentrations of 10.4–40.0 IU/mL or ± 5.0% at anti-TPO concentrations of 130.0–270.0 IU/mL.

SubstanceSubstance Test Concentration
Anti-thyroglobulin antibody (TgAb)2000 IU/mL
Thyroglobulin (Tg)700 ng/mL
Thyroid-stimulating immunoglobulin (TSI)113 IU/L

9.8. Reagent Stability

The shelf-life stability of the ADVIA Centaur aTPOII reagents (ReadyPack and calibrator) and ADVIA Centaur aTPOII diluent was determined to be 24 months.

The on-board stability of the ADVIA Centaur aTPOII reagents (ReadyPack and calibrator) was determined to be 42 days on the ADVIA Centaur XP system with a calibration interval of 42 days.

The on-board stability of the ADVIA Centaur aTPOII calibrators on the ADVIA Centaur XP system was determined to be 8 hours.

The on-board stability of the ADVIA Centaur aTPOII diluent on the ADVIA Centaur XP system was determined to be 28 days.

The ADVIA Centaur aTPOII calibrators when reconstituted were determined to be stable at 2-8°C and room temperature for 28 days and 8 hours, respectively.

9.9. Sample Stability

Page 14

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 10 of 14
October 2025

  • This specimen types recommended for this assay are serum and plasma. Sample stability studies were conducted using SST, K2 EDTA Plasma, and Lithium Heparin tube types.
  • After centrifugation, specimens in the primary collection device are stable for up to 8 hours at room temperature. Samples in the primary collection device include plasma stored on packed red cells, and serum samples processed and stored in gel-barrier blood collection tubes.
  • Separated samples are stable for up to 8 hours at room temperature, and for up to 7 days at 2–8°C.
  • Separated samples are stable at ≤ -20°C for up to 6 months. Avoid more than 2 freeze- thaw cycles. Do not store in a frost-free freezer. Thoroughly mix thawed samples and centrifuge them before using.

9.10. Expected Values

A reference interval for apparently healthy adults was established non‑parametrically using the ADVIA Centaur XP system in accordance with CLSI Document EP28‑A3c. Samples were collected prospectively from euthyroid male and female adult subjects with normal TSH levels. The reference interval was determined by calculating the 95th percentile of the distribution of values. A small percentage of samples (3.8%) were above the clinical cut-off of 10.4 IU/mL. Some level of positivity may be expected in the normal population.

GroupN^aMedian IU/mLReference Interval IU/mL
Adults (23–96 years)261< 4.6< 10.4

^a Number of samples tested.

As with all in vitro diagnostic assays, each laboratory should determine its own reference interval for the diagnostic evaluation of patient results. Consider these values as guidance only.

9.11. Clinical Performance

The performance of the ADVIA Centaur aTPOII assay was determined by testing a total of 666 samples using the ADVIA Centaur XP system in accordance with CLSI Document EP12‑A2.

A prospective multi‑center clinical study was conducted to evaluate the ADVIA Centaur aTPOII assay in the detection of autoimmune thyroiditis across 6 sites in the United States. A total of 364 samples were prospectively collected from patients presenting to endocrinologists with signs and symptoms of AITD. This population included patients with previously diagnosed Graves' disease and Autoimmune thyroiditis (formerly known as Hashimoto's thyroiditis) who

Page 15

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 11 of 14
October 2025

were treatment‑naïve or had been on treatment ≤ 6 months, and female patients who were pregnant or postpartum. Definitive diagnosis of AITD or non-autoimmune thyroid disease (NAITD) was made by the endocrinologist according to American Thyroid Association criteria.

Clinical sensitivity and specificity estimates were determined for the 364 prospectively collected samples:

Table 1: Graves' sensitivity and specificity estimates

ParameterNEstimate95% CI
Sensitivity10060.0% (60/100)(50.2, 69.1%)
Specificity15789.2% (140/157)(83.3, 93.1%)

Table 2: Autoimmune thyroiditis sensitivity and specificity estimates

ParameterNEstimate95% CI
Sensitivity10782.2% (88/107)(73.9, 88.3%)
Specificity15789.2% (140/157)(83.3, 93.1%)

Table 3: Overall sensitivity and specificity estimates

ParameterNEstimate95% CI
Sensitivity20771.5% (148/207)(65.0, 77.2%)
Specificity15789.2% (140/157)(83.3, 93.1%)

Additionally, 302 samples were retrospectively or prospectively collected from subjects with other thyroid conditions, disease states, or pregnancy, which may be expected to have elevated thyroid antibodies, to evaluate the specificity of the ADVIA Centaur aTPOII assay. The 302 samples were not evaluated for the presence or absence of autoimmune thyroid disease.

The sensitivity and specificity estimates were determined for the prospectively and retrospectively collected samples:

Page 16

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 12 of 14
October 2025

DiagnosisAutoimmune thyroiditis (Hashimoto's Disease)Graves' DiseaseControls^aTotal
ADVIA Centaur aTPOII
Positive608884231
Negative4019375435
Total100107459666

Autoimmune thyroiditis (Hashimoto's Disease)
Sensitivity: 60.0% (60/100) 95 CI 50.2–69.1%
Specificity: 81.7% (375/459) 95 CI 77.9,85.0%

Graves'
Sensitivity: 82.2% (88/107) 95 CI 73.9–88.3%
Specificity: 81.7% (375/459) 95 CI 77.9-85.0%

^a The controls cohort includes 302 samples from subjects with other thyroid conditions, disease states, or pregnancy and 157 samples from prospectively collected samples diagnostically confirmed negative for autoimmune thyroid disease; see the distribution of samples table below.

The overall sensitivity and specificity estimates were determined to be 71.5% for sensitivity and 81.7% for specificity.

ParameterNEstimate95% CI
Sensitivity20771.5% (148/207)(65.0, 77.2%)
Specificity45981.7% (375/459)(77.9, 85.0%)

For each clinical subgroup, the percent positivity of the ADVIA Centaur aTPOII assay was calculated using an assay cut‑off of 10.4 IU/mL.

The percent agreement between the ADVIA Centaur aTPOII assay and a commercially available anti-thyroid peroxidase (aTPO) assay was evaluated in accordance with CLSI EP12-A2 guidelines. The study included:

  • 364 AITD and NAITD prospectively collected samples
  • 570 NAITD retrospectively collected samples
  • 24 apparently healthy individual samples

Page 17

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 13 of 14
October 2025

Samples outside the assay's measuring interval were excluded, resulting in a final analysis set of 280 samples. The percent agreement estimates are summarized in the accompanying table below.

Comparative Assay
PositiveNegativeTotal
ADVIA Centaur aTPOII Assay
≥ 10.4 IU/mL1991200
< 10.4 IU/mL285280
Total22753280

PPA: 87.7% (199/227)[95% Confidence Interval: (82.8,91.3%)]
NPA: 98.1% (52/53)[95% Confidence Interval: (90.1,99.7%)]

Agreement of the assays may vary depending on the study design, comparative assay, and population tested.

Distribution of target and differential disease samples and antibody positivity rates:

ADVIA Centaur aTPOIIPredicate
Target conditionsNn POS (%)
Graves' Disease10060 (60%)
Hashimoto's Thyroiditis10788 (82.2%)
Total207148 (71.5%)
Differential diagnosis controls
Multinodular goiter622 (3.2%)
Non-autoimmune thyroid disease (NAITD)9515 (15.8%)
Thyroid Carcinoma204 (20.0%)
Silent Painless Thyroiditis207 (35.0%)
Subacute Thyroiditis100 (0.0%)
Hepatitis C virus (HCV)200 (0.0%)
Hepatitis B virus (HBV)207 (35.0%)
Human immunodeficiency virus (HIV)201 (5.0%)

Page 18

Siemens Healthcare Diagnostics 510(k) Summary
Siemens Healthcare Diagnostics, Inc
AINN Response – ADVIA Centaur aTPOII (K250250) Page 14 of 14
October 2025

ADVIA Centaur aTPOIIPredicate
Diabetes Type 1102 (20.0%)
Sjogren's Syndrome101 (10.0%)
Primary Biliary Cholangitis109 (90.0%)
Systemic Sclerosis104 (40.0%)
Pernicious Anemia101 (10.0%)
Rheumatoid Arthritis100 (0.0%)
Systemic Lupus Erythematosus (SLE)101 (10.0%)
Addison's Disease100 (0.0%)
Miscarriage2020 (100.0%)
Pregnancy – 1st Trimester311 (3.2%)
Pregnancy – 2nd Trimester307 (23.3%)
Pregnancy – 3rd Trimester312 (6.5%)
Total45984 (18.3%)

10. CONCLUSION

The ADVIA Centaur aTPOII assay is substantially equivalent to the predicate device, the Architect Anti-TPO (K052407).

§ 866.5870 Thyroid autoantibody immunological test system.

(a)
Identification. A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.(b)
Classification. Class II (performance standards).