Search Results
Found 60 results
510(k) Data Aggregation
(255 days)
Atellica® CH Diazo Direct Bilirubin (D DBil) Trade/Device Name: Regulation Number: 21 CFR 862.1110
Bilirubin FDA Classification: Class II Review Panel: Chemistry Product Code: CIG Regulation Number: 21 CFR 862.1110
The Atellica® CH Diazo Direct Bilirubin (D DBil) assay is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma using the Atellica® CH Analyzer. Measurement of direct bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic-hematological, and metabolic disorders, including hepatitis and gall bladder block.
Atellica® CH Diazo Direct Bilirubin is a Photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. Absorbance is measured at 545/658 nm.
The provided text describes the performance characteristics and acceptance criteria for the Atellica® CH Diazo Direct Bilirubin (D DBil) assay. Here's a breakdown of the requested information:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance:
| Performance Characteristic | Acceptance Criteria (Design Goal) | Reported Device Performance |
|---|---|---|
| Detection Capability | LoQ ≤ 0.10 mg/dL | LoQ = 0.10 mg/dL |
| Assay Comparison | Correlation coefficient (r) ≥ 0.950 Slope: 1.00 ± 0.10 | r = 0.993 Slope y = 0.95x - 0.03 mg/dL (0.95, within 1.00 ± 0.10) |
| Interferences (HIL) | ≤ 10% bias from hemoglobin, bilirubin (presumably total bilirubin as an icteric substance), and lipemia. Bias > 10% is considered interference. | Hemoglobin: Interference observed above 12.5 mg/dL. Lipemia: No interference ≤ 1000 mg/dL |
| Non-Interfering Substances | Bias due to these substances ≤ 10% | All tested substances showed ≤ 10% bias at specified concentrations. |
Note: Specific acceptance criteria for precision and reproducibility are not explicitly listed as single values but are implied by the comprehensive presentation of the data, demonstrating acceptable variability for a diagnostic assay. The document states that the results "support that the Candidate Device... is substantially equivalent."
2. Sample size used for the test set and the data provenance:
- Assay Comparison: N = 100 samples
- Specimen Equivalency (Plasma vs. Serum): N = 53 samples for each plasma type (Lithium heparin, Sodium heparin, K2(EDTA)).
- Precision: N = 80 for each serum level (4 serum levels tested, total 320 measurements).
- Reproducibility: N = 225 for each serum level (4 serum levels tested, total 900 measurements).
- Interferences (HIL and Non-interfering Substances): The number of samples for interference testing is not explicitly stated as a single 'N' for the test set. However, the tables indicate specific analyte concentrations tested (e.g., for Hemoglobin, Lipemia, Acetaminophen, etc.), implying multiple measurements were performed for each interference level.
Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable as the device is an in vitro diagnostic assay for quantitative determination of direct bilirubin. The "ground truth" in this context refers to the measured concentration of direct bilirubin, which is established by established laboratory methods, standard reference materials, and comparison to a predicate device, rather than expert interpretation of images or clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation (e.g., medical imaging interpretation) where multiple readers assess cases and discrepancies are resolved by a super-reader. For a quantitative diagnostic assay, the "ground truth" is determined by objective measurement rather than expert consensus on subjective findings.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. The device is an in vitro diagnostic assay, not an AI-assisted diagnostic tool that would involve human readers interpreting cases. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is not relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device is a standalone in vitro diagnostic assay. Its performance is measured independently of human interpretation in the clinical setting, although laboratory personnel operate the analyzer and interpret the numerical results in the context of a patient's overall clinical picture. The studies described (Precision, Reproducibility, Assay Comparison, Specimen Equivalency, Interferences) all reflect the standalone performance of the assay on the Atellica CH Analyzer.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device's performance studies is established by:
- Reference Methods/Predicate Device: The "Assay Comparison" section uses the Wako Direct Bilirubin V assay as the comparative method.
- Internal Reference Standards: The assay's traceability is to internal reference standards manufactured by gravimetric methods.
- Control Samples/Spiking: For precision, reproducibility, and interference studies, samples are prepared with known concentrations of analyte or interferents.
8. The sample size for the training set:
This information is not provided in the document. This type of detail is typically associated with machine learning or AI algorithm development, which is not the primary focus of this in vitro diagnostic device submission. The device involves a chemical reaction and photometric measurement, not a "training set" in the machine learning sense.
9. How the ground truth for the training set was established:
This information is not provided and is not applicable as the device does not involve a "training set" in the context of machine learning. The assay mechanism is based on a defined chemical reaction (diazo colorimetry) rather than a trained algorithm.
Ask a specific question about this device
(245 days)
: K222104
Trade/Device Name: Atellica® CH Diazo Total Bilirubin (D_TBil) Regulation Number: 21 CFR 862.1110
Bilirubin FDA Classification: Class II Review Panel: Chemistry Product Code: CIG Regulation Number: 21 CFR 862.1110
The Atellica® CH Diazo Total Bilirubin (D TBil) assay is for in vitro diagnostic use in the quantitative determination of total bilirubin in adults and children (non-neonates) in human serum and plasma using the Atellica® CH Analyzer. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Atellica CH Diazo Total Bilirubin is a photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. A specific mixture of detergents enables the determination of the total bilirubin.
The provided document describes the Siemens Atellica® CH Diazo Total Bilirubin (D_TBil) assay, an in vitro diagnostic device, and its performance characteristics to demonstrate substantial equivalence to a predicate device (Dimension TBI Flex reagent cartridge).
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this in-vitro diagnostic device are generally defined by demonstrating performance within established statistical limits or comparison to a predicate device, as per CLSI (Clinical and Laboratory Standards Institute) guidelines. The "acceptance criteria" themselves are not always explicitly stated as pass/fail thresholds for each performance characteristic in a simple numerical format, but rather as meeting the objectives of the study design (e.g., correlation coefficient of ≥ 0.950).
| Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Detection Capability | LoQ < 0.10 mg/dL | LoB: 0.01 mg/dL, LoD: 0.02 mg/dL, LoQ: 0.10 mg/dL |
| Precision (Repeatability) | Not explicitly stated as a single value, but typically refers to a low %CV. | Serum 1 (1.02 mg/dL): SD 0.015 mg/dL (1.5%CV)Serum 2 (13.40 mg/dL): SD 0.053 mg/dL (0.4%CV)Serum 3 (22.39 mg/dL): SD 0.067 mg/dL (0.3%CV) |
| Precision (Within-Lab) | Not explicitly stated as a single value. | Serum 1 (1.02 mg/dL): SD 0.034 mg/dL (3.3%CV)Serum 2 (13.40 mg/dL): SD 0.140 mg/dL (1.0%CV)Serum 3 (22.39 mg/dL): SD 0.189 mg/dL (0.8%CV) |
| Reproducibility (Total) | Not explicitly stated as a single value. | Sample 1 (1.03 mg/dL): SD 0.023 mg/dL (2.2%CV)Sample 2 (13.24 mg/dL): SD 0.091 mg/dL (0.7%CV)Sample 3 (22.14 mg/dL): SD 0.146 mg/dL (0.7%CV) |
| Assay Comparison (Method) | Correlation coefficient (r) ≥ 0.950 and slope of 1.00 ± 0.10 compared to predicate. | r = 0.997 with predicate (Dimension TBI). Regression Equation: y = 1.02x + 0.08 mg/dL. The slope of 1.02 is within the 1.00 ± 0.10 range. |
| Specimen Equivalency | High correlation (r) and a close to 1.00 slope for different plasma types vs. serum. | Plasma (Lithium Heparin) vs. Serum: r = 0.997, y = 0.98x + 0.05 mg/dLPlasma (Sodium Heparin) vs. Serum: r = 0.998, y = 1.00x + 0.02 mg/dLPlasma (K2 EDTA) vs. Serum: r = 0.998, y = 0.99x + 0.03 mg/dL |
| Interferences (HIL) | ≤ 10% bias from hemoglobin and lipemia at specified concentrations. | Hemoglobin (1000 mg/dL): -9.3% bias (at 1.08 mg/dL Bilirubin), -7.1% bias (at 13.86 mg/dL Bilirubin)Lipemia (1000 mg/dL Triglyceride): -7.8% bias (at 0.90 mg/dL Bilirubin), 0.5% bias (at 12.94 mg/dL Bilirubin)All observed biases are ≤ 10%. |
| Non-Interfering Substances | Bias ≤ 10% for listed substances at specified concentrations (often with acceptance criteria of | All listed substances (e.g., Acetaminophen, Carbenicillin, Ascorbic acid, Ibuprofen, etc.) showed observed % bias ≤ 10% or within the specified acceptance criteria for the observed analyte concentration. |
| Expected Values | Verification of reference interval. | Verified reference interval: 0.3 - 1.2 mg/dL (5.13 - 20.52 µmol/L). |
2. Sample Size Used for the Test Set and Data Provenance
- Detection Capability: Not explicitly stated as a "test set" in the context of patient samples, but the determination was in accordance with CLSI Documents EP17-A2.
- Precision:
- Repeatability/Within-Lab Precision: 80 measurements for each of the three serum levels (Serum 1, 2, 3), assayed in duplicate over 20 days.
- Reproducibility: 225 measurements for each of the three serum levels (Sample ID 1, 2, 3), assayed with 5 replicates per run for 5 days using 3 instruments/sites and 3 reagent lots.
- Assay Comparison (Method Comparison): N = 103 patient samples (Serum)
- Specimen Equivalency: N = 57 patient samples for each plasma type (Lithium Heparin, Sodium Heparin, K2 EDTA) compared to serum.
- Interferences (HIL and Non-interfering Substances): The studies were performed with specific interferent concentrations and two bilirubin levels. The sample size refers to the number of spiked samples tested, but not explicitly the number of unique patient samples that may have been used to create these spiked materials.
- Data Provenance: The document does not specify the country of origin for the samples or whether the studies were retrospective or prospective. Given the nature of in-vitro diagnostic development, these are typically prospective studies using a mix of spiked and potentially remnant clinical samples.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable and therefore not provided in the document. For an in-vitro diagnostic assay that quantitatively measures an analyte (total bilirubin), the "ground truth" is established by the reference method or comparison method (the predicate device in this case) and the inherent accuracy and traceability of the calibrators, rather than by human expert review of images or clinical cases. The ground truth for quantitative assays typically relies on metrological traceability to certified reference materials (NIST Standard Reference Material 916 in this case) and established laboratory protocols.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1, 3+1) are common in studies involving human interpretation of medical images or clinical data, especially for AI applications where consensus among experts establishes the ground truth. For quantitative in-vitro diagnostics, the "ground truth" is determined by established analytical methods and reference standards, not by human adjudication of qualitative results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This is not applicable. An MRMC study is designed to assess the performance of a diagnostic system that involves human readers (e.g., radiologists interpreting images) and typically compares the effectiveness of human readers with and without AI assistance. The Atellica CH Diazo Total Bilirubin assay is an automated in-vitro diagnostic test, meaning it does not involve human readers for interpretation beyond the initial sample collection and analysis setup. It is a standalone analytical device.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This is essentially what the entire submission describes. The performance characteristics presented (Detection Capability, Precision, Reproducibility, Assay Comparison, Specimen Equivalency, Interference) are all standalone performance studies of the medical device itself. There is no human-in-the-loop component for the interpretation or output of the assay. The device directly measures and reports total bilirubin levels.
7. Type of Ground Truth Used
The ground truth for this quantitative in-vitro diagnostic assay is established through:
- Metrological Traceability: The assay is traceable to the NIST Standard Reference Material 916. This provides a fundamental basis for accuracy.
- Comparison to a Predicate Device: The performance of the new device is compared against a legally marketed predicate device (Dimension TBI assay), which serves as the "truth" or reference in method comparison studies.
- Controlled Samples: Precision, reproducibility, and interference studies use manufactured control materials, internal validations, and spiked samples with known concentrations or interferent levels, whose 'truth' values are established through rigorous analytical methods and often validated against reference methods.
8. The Sample Size for the Training Set
This information is not provided in the document, and it's generally not applicable in the typical sense of "training set" for traditional in-vitro diagnostic devices. These devices are developed, validated, and optimized through a series of analytical studies using various types of samples (e.g., precision materials, linearity samples, spiked samples, patient samples). The "training" isn't a machine learning training phase, but rather the process of optimizing the reagent formulation, reaction kinetics, and instrument parameters. The data used for these optimizations would be proprietary and extensive, but not typically referred to as a "training set" in regulatory submissions for IVDs. The analytical performance studies (like those detailed above) serve as the validation and verification of the final, optimized product.
9. How the Ground Truth for the Training Set Was Established
As explained in point 8, the concept of a "training set" with a defined "ground truth" for a traditional IVD like this is generally not applicable in the same way as for AI/ML devices. The "ground truth" for the development and optimization of such an assay would be established through:
- Reference Methods: Using established, highly accurate reference methods to determine the true concentration of bilirubin in control materials and patient samples used during development.
- Certified Reference Materials: Calibrating and verifying the assay against NIST or other internationally recognized certified reference materials.
- Known Spiking: Creating samples with precisely known additions of bilirubin or interferents to test linearity, recovery, and interference effects.
In summary, the provided document details a comprehensive set of analytical studies to demonstrate the performance and substantial equivalence of the Atellica® CH Diazo Total Bilirubin assay as a standalone in-vitro diagnostic device. The evaluation follows a different paradigm than AI/ML algorithms, thus many questions related to expert review, adjudication, and MRMC studies are not relevant.
Ask a specific question about this device
(59 days)
Longford, Ireland
Re: K223324
Trade/Device Name: Total Bilirubin2 Regulation Number: 21 CFR 862.1110
Governing Regulation Number: 21 CFR §862.1110 Product Code: CIG
{4}---------------------------------
The Total Bilirubin2 assay is used for the quantitation of total bilirubin in human serum or plasma, of adults and neonates, on the ARCHITECT c System.
Measurement of total bilirubin, an organic compound formed during the normal destruction of red blood cells, is used in the diagnosis and treatment of liver, hematological, and metabolic disorders, including hepatitis and disorders of the biliary tract. In newborn infants, the Total Bilirubin2 assay is intended to measure the levels of total bilirubin (conjugated and unconjugated) in serum or plasma to aid in the diagnosis and management of neonatal jaundice and hemolytic disease of the newborn.
The Total Bilirubin2 assay (subject device) is an automated clinical chemistry assay for the quantitation of total bilirubin in human serum or plasma, of adults and neonates, on the ARCHITECT c System. Total (conjugated and unconjugated) bilirubin couples with a diazo reagent in the presence of a surfactant to form azobilirubin. The diazo reaction is accelerated by the addition of surfactant as a solubilizing agent. The increase in absorbance at 548 nm due to azobilirubin is directly proportional to the total bilirubin concentration. The methodology is Diazonium salt.
The provided text describes a 510(k) premarket notification for a medical device called "Total Bilirubin2", an in vitro diagnostic assay. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving clinical effectiveness through the extensive studies typically associated with AI/ML diagnostic tools. Therefore, the questions related to AI/ML specific criteria (like MRMC studies, number of experts for ground truth, sample size for training sets, etc.) are not applicable in this context.
The document primarily details the analytical performance of the Total Bilirubin2 assay.
Here's an analysis based on the information provided, adhering to the request:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this in vitro diagnostic device are typically defined by ranges of acceptable analytical performance, following established CLSI (Clinical and Laboratory Standards Institute) guidelines. The reported device performance is compared against these internal acceptance criteria.
| Performance Metric | Acceptance Criteria (Implicit from CLSI Guidelines/Industry Standards) | Reported Device Performance (as stated) |
|---|---|---|
| Reportable Interval (Range) | Established analytical measuring interval, extended measuring interval, and reportable interval. | Analytical Measuring Interval (AMI): 0.1 – 25.0 mg/dL Extended Measuring Interval (EMI): 25.0 – 125.0 mg/dL Reportable Interval: 0.1 – 125.0 mg/dL |
| Within-Laboratory Precision (SD/CV%) | Specific maximum acceptable SD and %CV for different concentrations, as per CLSI EP05-A3 guidelines. | Control Level 1 (1.1 mg/dL): SD: 0.04 (Range 0.02-0.04), %CV: 3.4 (Range 1.8-3.4) Control Level 2 (4.2 mg/dL): SD: 0.09 (Range 0.09-0.10), %CV: 2.1 (Range 2.0-2.2) Panel A (0.3 mg/dL): SD: 0.00 (Range 0.00-0.03), %CV: 0.0 (Range 0.0-9.2) Panel B (13.3 mg/dL): SD: 0.11 (Range 0.09-0.12), %CV: 0.8 (Range 0.7-0.9) Panel C (22.3 mg/dL): SD: 0.16 (Range 0.16-0.18), %CV: 0.7 (Range 0.7-0.8) |
| System Reproducibility (SD/CV%) | Specific maximum acceptable SD and %CV for different concentrations, as per CLSI EP05-A3 guidelines. | Control Level 1 (1.1 mg/dL): SD: 0.02, %CV: 2.2 Control Level 2 (4.5 mg/dL): SD: 0.16, %CV: 3.5 Panel B (13.4 mg/dL): SD: 0.57, %CV: 4.3 Panel C (22.4 mg/dL): SD: 1.12, %CV: 5.0 |
| Accuracy (Bias) | Bias within an acceptable range, relative to a reference method (Doumas). | Bias ranged from -0.1% to 3.7%. |
| Lower Limits of Measurement | Defined LoB, LoD, and LoQ based on CLSI EP17-A2 guidelines. | LoB: 0.02 mg/dL LoD: 0.04 mg/dL LoQ: 0.07 mg/dL |
| Linearity | Linearity across the specified analytical measuring interval. | Linear across the analytical measuring interval of 0.1 to 25.0 mg/dL. |
| Interference (Endogenous) | Interference within ± 10% for specified substances at given concentrations. | Hemoglobin (1000 mg/dL), Total protein (15 g/dL), Triglycerides (1500 mg/dL): No significant interference (within ± 10%). Indican (1 mg/dL): No significant interference. Indican (2 mg/dL): 17% interference (beyond ±10%). |
| Interference (Exogenous) | Interference within ± 10% for specified substances at given concentrations. | Variety of common drugs tested; no significant interference for most. Indocyanine green (10 mg/L): 9% interference. |
| Method Comparison (Correlation) | High correlation coefficient and acceptable slope/intercept when compared to predicate device. | Serum: Correlation Coefficient: 1.00, Intercept: -0.03, Slope: 1.03 (Range 0.1–22.5 mg/dL) Neonatal serum: Correlation Coefficient: 1.00, Intercept: 0.00, Slope: 1.00 (Range 0.2–22.8 mg/dL) |
| Tube Type Suitability | Acceptable performance across specified tube types. | Serum tubes, Serum separator tubes, Dipotassium EDTA tubes, Lithium heparin tubes, Lithium heparin separator tubes, Sodium heparin tubes were acceptable. |
| Dilution Verification (% Recovery & %CV) | % recovery within 100% ± 10%; imprecision ≤ 7 %CV for automated dilution, ≤ 8 %CV for manual dilution. | Automated Dilution: 96.3% to 104.4% recovery, 1.6% to 2.5% CV. Manual Dilution: 95.0% to 106.7% recovery, 2.2% to 4.9% CV. |
Study Details:
-
Sample size used for the test set and the data provenance:
- Precision (Within-Laboratory): 80 replicates for each control/panel (on a representative combination out of 3 multi-lot/instrument combinations).
- Reproducibility (System): 84 replicates for each control/panel.
- Lower Limits of Measurement: ≥ 60 replicates for LoB and LoD for each of 3 lots on 2 instruments.
- Interfering Substances: Not explicitly stated, but "Each substance was tested at 2 levels of the analyte."
- Method Comparison:
- Serum: 167 samples
- Neonatal serum: 163 samples
- Tube Type: Samples collected from a minimum of 40 donors.
- Dilution Verification: 5 samples prepared with varying concentrations.
- Data Provenance: Not explicitly stated regarding country of origin or whether retrospective/prospective. However, given the nature of in vitro diagnostic analytical studies, samples are typically acquired prospectively or from biobanks for specific analytical testing purposes.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For in vitro diagnostic devices like this bilirubin assay, "ground truth" is established by reference methods or highly characterized calibrators/control materials, not by expert human readers. The accuracy study, for example, compares results to material standardized to the Doumas Total Bilirubin reference method, which represents the "ground truth" for bilirubin measurement. Therefore, expert readers/adjudicators as typically seen in imaging AI studies are not applicable here.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This is an in vitro diagnostic assay, and its performance is evaluated against analytical measurements, not human interpretations requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an in vitro diagnostic test, not an AI/ML-driven imaging or diagnostic algorithm designed to assist human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is an assay performed on an automated system, providing a quantitative result. Its "performance" is inherently "standalone" in generating the numerical value, but it's not an AI algorithm in the sense of image interpretation or complex diagnostic inference.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For analytical performance studies, the "ground truth" for bilirubin concentration is established by reference methods (e.g., the Doumas method for accuracy) or by using certified reference materials and calibrators with known concentrations. This is the gold standard for quantitative in vitro diagnostic measurements.
-
The sample size for the training set:
- Not applicable. This is not an AI/ML device that requires a "training set" in the computational sense. The device's performance is a function of its reagents, instrument, and established methodology, not a learned algorithm.
-
How the ground truth for the training set was established:
- Not applicable. See above.
Ask a specific question about this device
(536 days)
Product Code CJY
21 CFR § 862.1225 - Creatinine Test system - Class II- Product Code CGX
21 CFR & 862.1110
II- Product Code CIG
21 CFR § 862.1345 - Glucose Test system - Class II- Product Code CFR
21 CFR § 862.1110
K934361)
Medicon Hellas Albumin: Reagent for the quantitative measurement of albumin in serum. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Medicon Hellas Calcium: Reagent for the quantitative measurement of calcium in serum or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Medicon Hellas Creatinine: Reagent for the quantitative measurement of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Medicon Hellas Glucose: Reagent for the quantitative measurement of glucose in serum and urine. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Medicon Hellas Direct Bilirubin; Reagent for the quantitative measurement of direct bilirubin (conjugated) in serum. Measurements of the level of direct bilirubin is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall blader block.
Medicon Hellas Total Bilirubin: Reagent for the quantitative measurements of total bilirubin in serum. Measurements of the levels of total bilirubin is used in the diagnosis and treatment of liver. hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Medicon Hellas Urea Nitrogen: Reagent is for the quantitative measurement of urea nitrogen in serum and urine. Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
The Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Glucose, Medicon Hellas Direct Bilirubin, Medicon Hellas Total Bilirubin, and Medicon Hellas Urea Nitrogen are reagents for use with Diatron Pictus 500 Clinical Chemistry Analyzers. They are test systems for the quantitative measurement of albumin, calcium, creatinine, glucose, direct and total bilirubin, and urea nitrogen in human serum and urine where clinically applicable. The methods employed are photometric, utilizing reactions between the sample and reagents to produce a colored chromophore or a change in absorbance that is proportional to the concentration of the analyte. The analyzer photometer reads the absorbances at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically.
The provided text describes the performance of several Medicon Hellas assays (Albumin, Calcium, Creatinine, Glucose, Direct Bilirubin, Total Bilirubin, and Urea Nitrogen) when run on the Diatron Pictus 500 Clinical Chemistry Analyzer, demonstrating their substantial equivalence to predicate devices (Beckman Coulter AU reagents on AU2700 analyzer, and Abbott Architect Direct Bilirubin on Architect c8000 analyzer).
Here's an analysis of the provided information, structured to address your specific points regarding acceptance criteria and study details:
1. A Table of Acceptance Criteria and the Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a single, overarching table with pass/fail remarks. Instead, it describes each performance characteristic and then presents the results. The "Summary" sections for each study type imply that the results met the pre-defined acceptance criteria for demonstrating substantial equivalence. For instance, for accuracy, it states "Accuracy studies completed on at least three lots of each candidate reagent confirm that Medicon albumin... are substantially equivalent to the related predicate devices." This implies that the statistical analyses (Deming regression, R2, slope, intercept) fell within acceptable ranges. Similarly, for precision, it states "All lots passed acceptance criteria for each applicable sample type at each level."
Since explicit acceptance criteria are not presented, they are inferred from the demonstrated performance and the statement that the devices "passed acceptance criteria" or "met statistical acceptance criteria." Below is a table summarizing the reported device performance for each analyte. The "Acceptance Criteria" column will reflect the general statements of success or the implied ranges from the results themselves, as explicit numerical targets for individual tests are not given.
Implied Acceptance Criteria and Reported Device Performance
| Analyte | Performance Characteristic | Implied/General Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Medicon Hellas Albumin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence (>0.98 is generally good) | R2 = 0.9862 |
| Slope | Values close to 1 | Slope = 1.0180 | |
| Intercept | Values close to 0 | Intercept = 0.05 | |
| Reportable Range (Serum) | Comparable to predicate device | 1.50 - 6.00 g/dL | 1.50 - 6.00 g/dL |
| LOD (Serum) | Acceptable for clinical use | 0.40 g/dL | 0.40 g/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.50 g/dL | 0.50 g/dL |
| Precision (Serum) | CV% within acceptable clinical limits (e.g., typically <10% for these analytes) | Repeatability CV (Level 1,2,3): 2.21%, 1.45%, 1.37% | Between Run CV (Level 1,2,3): 5.25%, 3.80%, 2.36% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations (e.g., Hemoglobin 500 mg/dL, Triglycerides 3000 mg/dL) | (See document for detailed thresholds per interferent) |
| Medicon Hellas Calcium | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9949, Urine R2 = 0.9965 |
| Slope | Values close to 1 | Serum Slope = 1.0099, Urine Slope = 0.9888 | |
| Intercept | Values close to 0 | Serum Intercept = -0.3, Urine Intercept = -0.8 | |
| Reportable Range | Comparable to predicate device | Serum: 4.0 - 18.0 mg/dL, Urine: 2.0 - 40.0 mg/dL | Serum: 4.0 - 18.0 mg/dL, Urine: 2.0 - 40.0 mg/dL |
| LOD | Acceptable for clinical use | Serum: 0.5 mg/dL, Urine: 1.3 mg/dL | Serum: 0.5 mg/dL, Urine: 1.3 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 0.5 mg/dL, Urine: 1.5 mg/dL | Serum: 0.5 mg/dL, Urine: 1.5 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.16%, 1.06%, 0.82% | Between Run CV (L1,2,3): 3.37%, 1.51%, 1.95% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Creatinine | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9989, Urine R2 = 0.9992 |
| Slope | Values close to 1 | Serum Slope = 1.0207, Urine Slope = 0.9904 | |
| Intercept | Values close to 0 | Serum Intercept = -0.10, Urine Intercept = -0.81 | |
| Reportable Range | Comparable to predicate device | Serum: 0.3 - 25.0 mg/dL, Urine: 1.2 - 300.0 mg/dL | Serum: 0.3 - 25.0 mg/dL, Urine: 1.2 - 300.0 mg/dL |
| LOD | Acceptable for clinical use | Serum: 0.2 mg/dL, Urine: 1.0 mg/dL | Serum: 0.2 mg/dL, Urine: 1.0 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 0.2 mg/dL, Urine: 1.1 mg/dL | Serum: 0.2 mg/dL, Urine: 1.1 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 2.41%, 1.08%, 1.04% | Between Run CV (L1,2,3): 3.63%, 4.58%, 2.30% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Direct Bilirubin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | R2 = 0.9978 |
| Slope | Values close to 1 | Slope = 0.9656 | |
| Intercept | Values close to 0 | Intercept = -0.01 | |
| Reportable Range (Serum) | Comparable to predicate device | 0.2 - 15.0 mg/dL | 0.2 - 15.0 mg/dL |
| LOD (Serum) | Acceptable for clinical use | 0.1 mg/dL | 0.1 mg/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.2 mg/dL | 0.2 mg/dL |
| Precision (Serum) | CV% within acceptable clinical limits | Repeatability CV (L1,2,3): 3.11%, 2.46%, 2.48% | Between Run CV (L1,2,3): 2.31%, 3.14%, 2.29% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Glucose | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9992, Urine R2 = 0.9989 |
| Slope | Values close to 1 | Serum Slope = 0.9715, Urine Slope = 1.0222 | |
| Intercept | Values close to 0 | Serum Intercept = 2.7, Urine Intercept = -0.9 | |
| Reportable Range | Comparable to predicate device | Serum: 10 - 700 mg/dL, Urine: 10 - 660 mg/dL | Serum: 10 - 700 mg/dL, Urine: 10 - 660 mg/dL |
| LOD | Acceptable for clinical use | Serum: 1.7 mg/dL, Urine: 2.4 mg/dL | Serum: 1.7 mg/dL, Urine: 2.4 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 4.0 mg/dL, Urine: 6.0 mg/dL | Serum: 4.0 mg/dL, Urine: 6.0 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.72%, 0.99%, 0.96% | Between Run CV (L1,2,3): 1.35%, 1.48%, 2.08% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Total Bilirubin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | R2 = 0.9996 |
| Slope | Values close to 1 | Slope = 1.0125 | |
| Intercept | Values close to 0 | Intercept = -0.06 | |
| Reportable Range (Serum) | Comparable to predicate device | 0.10 - 30.00 mg/dL | 0.10 - 30.00 mg/dL |
| LOD (Serum) | Acceptable for clinical use | 0.01 mg/dL | 0.01 mg/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.09 mg/dL | 0.09 mg/dL |
| Precision (Serum) | CV% within acceptable clinical limits | Repeatability CV (L1,2,3): 0.92%, 0.38%, 0.57% | Between Run CV (L1,2,3): 1.69%, 1.38%, 1.79% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Urea Nitrogen | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9983, Urine R2 = 0.9972 |
| Slope | Values close to 1 | Serum Slope = 1.0001, Urine Slope = 0.9844 | |
| Intercept | Values close to 0 | Serum Intercept = -0.2, Urine Intercept = 21.9 | |
| Reportable Range | Comparable to predicate device | Serum: 3 - 100 mg/dL, Urine: 24 - 1300 mg/dL | Serum: 3 - 100 mg/dL, Urine: 24 - 1300 mg/dL |
| LOD | Acceptable for clinical use | Serum: 2 mg/dL, Urine: 21 mg/dL | Serum: 2 mg/dL, Urine: 21 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 3 mg/dL, Urine: 24 mg/dL | Serum: 3 mg/dL, Urine: 24 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.94%, 2.14%, 1.07% | Between Run CV (L1,2,3): 2.24%, 2.56%, 3.16% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
2. Sample sizes used for the test set and the data provenance:
-
Sample Size for Test Set:
- Accuracy (Method Comparison): "A minimum of 70 clinical specimens, spanning the dynamic ranges, were assayed." Specific numbers are provided per analyte:
- Medicon Hellas Albumin: 112 samples (Serum)
- Medicon Hellas Calcium: 94 samples (Serum), 81 samples (Urine)
- Medicon Hellas Creatinine: 126 samples (Serum), 98 samples (Urine)
- Medicon Hellas Direct Bilirubin: 77 samples (Serum)
- Medicon Hellas Glucose: 99 samples (Serum), 100 samples (Urine)
- Medicon Hellas Total Bilirubin: 95 samples (Serum)
- Medicon Hellas Urea Nitrogen: 116 samples (Serum), 81 samples (Urine)
- Reportable Range (Linearity): "At least nine levels of each sample types were tested." (N=4 per level on Pictus P500)
- Sensitivity (LOD/LOQ):
- LoB/LoD: "5 Blank samples and 5 Low Levels samples respectively which were measured 4 times each day for a total of 60 measurements in 3 days."
- LoQ: "10 samples that span the low end of linearity were measured 5 times each day for a total of 150 measurements in 3 days."
- Interferences: "Serum and urine sample pools at low and high levels were prepared." The exact number of individual samples forming these pools is not specified beyond being "pools."
- Precision: "Precision study results from running applicable serum and urine samples (Level 1, Level 2 and Level 3) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level."
- Accuracy (Method Comparison): "A minimum of 70 clinical specimens, spanning the dynamic ranges, were assayed." Specific numbers are provided per analyte:
-
Data Provenance (e.g., country of origin of the data, retrospective or prospective):
- The document implies that the studies were conducted by Medicon Hellas, S.A. in Greece, given their address on the first page.
- The data appears to be prospective as it describes experiments conducted ("studies were performed," "testing confirmed," "protocol followed"). It references the collection and analysis of clinical specimens specifically for these validation studies. It does not mention retrospective analysis of existing patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable in the context of this 510(k) submission. This type of submission is for in vitro diagnostic (IVD) reagents, which measure specific analytes in bodily fluids. The "ground truth" for these measurements is typically established by comparative analysis against predicate devices and well-characterized reference methods (e.g., those detailed in CLSI guidelines for accuracy, linearity, precision). These are not image-based AI models requiring human expert interpretation for ground truth.
The "experts" involved would be the laboratory personnel performing the assays according to established clinical laboratory standards and the statistical analysis, rather than medical experts providing subjective interpretations. The document does not specify the number or qualifications of the laboratory personnel who conducted the tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. This concept (e.g., 2+1, 3+1 reader adjudication) is primarily used in studies where human readers provide subjective interpretations (e.g., radiology studies). For IVD devices, the "ground truth" is based on the analytical performance against established reference methods or predicate devices, which involves quantitative measurements and statistical analysis, not human adjudication of subjective interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI-assisted diagnostic imaging device. It is a chemical reagent intended for quantitative measurement of analytes in bodily fluids. Therefore, MRMC studies and the effect size on human readers are not relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The device (reagent) essentially operates in a "standalone" fashion in terms of its chemical reaction and measurement, independent of human interpretive intervention for the measurement itself. The performance data (accuracy, precision, linearity, etc.) presented is the standalone performance of the reagent on the specified analyzer. Human involvement is in operating the instrument, quality control, and interpreting the numerical results in a clinical context, but not in the measurement process being tested for substantial equivalence.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" in this context is established by:
- Comparative Method: The primary method for proving substantial equivalence is by demonstrating strong correlation and agreement with legally marketed predicate devices (Beckman Coulter AU reagents and Abbott Architect Direct Bilirubin reagent) using real patient clinical specimens. This acts as the standard for comparison.
- CLSI Guidelines: Various performance characteristics (linearity, sensitivity, precision, interferences) were evaluated according to Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., CLSI EP09c for accuracy, CLSI EP06-A for linearity, CLSI EP17-A2 for sensitivity, CLSI EP07-A and EP37 for interferences, CLSI EP05-A3 for precision). These guidelines represent accepted industry standards for validating in vitro diagnostic devices, thereby defining the "ground truth" for these analytical measurements.
- Reference Values: For linearity studies, "samples were assigned their reference values arithmetically from serial dilutions of the high-level sample," indicating a quantitatively derived reference for linearity.
Therefore, the ground truth is based on a combination of comparison to predicate devices and adherence to established analytical reference methods and industry standards (CLSI guidelines).
8. The sample size for the training set:
Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense. It is a chemical reagent. The "training" here would be the chemical formulation and manufacturing process, which is established through R&D and QA/QC, not data input to an algorithm.
9. How the ground truth for the training set was established:
Not applicable, as there is no "training set" for a chemical reagent. The "ground truth" for the development of the reagent itself would be the established chemical principles and desired analytical performance characteristics (e.g., reactivity, specificity, stability, sensitivity) based on scientific literature and previous experience with similar assays.
Ask a specific question about this device
(28 days)
Trade/Device Name: VITROS XT Chemistry Products TBIL-ALKP Slides Regulation Number: 21 CFR 862.1110
Rx Only For in vitro diagnostic use only
The TBIL test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Measurements of the levels of bilirubin, an organic compound formed during the normal destruction of red blood cells, are used in the diagnosis and treatment of liver, hematological and metabolic disorders, including hepatitis and gall bladder block.
The ALKP test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measure alkaline phosphatase (ALKP) activity in serum and plasma using VITROS XT 7600 Integrated Systems. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Not Found
This is an FDA 510(k) clearance letter for an in vitro diagnostic (IVD) device, specifically for VITROS XT Chemistry Products TBIL-ALKP Slides. The provided text is a regulatory communication and does not contain the acceptance criteria or study details for the device's performance.
To answer your request, I would need access to the actual 510(k) summary, often referred to as a "510(k) Premarket Notification." This document typically includes the performance data, acceptance criteria, and study designs to demonstrate substantial equivalence to a predicate device.
The information you've provided only states:
- Device Name: VITROS XT Chemistry Products TBIL-ALKP Slides
- Intended Use: Quantitative measurement of total bilirubin (TBIL) and alkaline phosphatase (ALKP) in serum and plasma using VITROS XT 7600 Integrated Systems.
- Regulatory Class: Class II
- Product Code: CIG (Bilirubin (total or direct) test system), CJE (Alkaline phosphatase test system)
- Date of Clearance: April 26, 2019
Without the 510(k) summary or a similar technical document, I cannot extract the specific acceptance criteria, study details, sample sizes, or ground truth information you've requested.
Therefore, I cannot provide the requested table and study details based solely on the provided text.
Ask a specific question about this device
(90 days)
| Clinical Chemistry |
| TBIL | CIG | Class II | 21 CFR 862.1110
- VITROS Chemistry Products CRBM Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5.1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy.
- VITROS Chemistry Products CREA Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Product CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
- VITROS Chemistry Products TBIL Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products TBIL Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hematological and metabolic disorders, including hepatitis and gall bladder block.
- VITROS XT 7600 Integrated System: Rx Only. For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest.
The VITROS XT 7600 Integrated System is a fully automated, computer controlled, clinical chemistry and immunodiagnostic analyzer intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, drugs of abuse, proteins, infectious diseases, as well as cardiac, metabolic, thyroid, anemia, and oncology markers in biological fluids such as serum, plasma, urine and cerebral spinal fluid. The System operates in conjunction with reagents, calibrators and controls designed for use with the System in the MicroSlide, MicroTip or MicroWell format.
The VITROS Chemistry MicroSlide range of products (in this case VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, and VITROS Chemistry Products TBIL Slides), are combined with the VITROS XT 7600 Integrated System to perform the VITROS CRBM, CREA, and TBIL assays.
The document describes the performance of the VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, and the VITROS XT 7600 Integrated System. The main purpose of the study is to demonstrate substantial equivalence to legally marketed predicate devices.
Here's an analysis of the acceptance criteria and study details:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a dedicated table format for each performance metric, but rather describes how results were evaluated in the "Specificity" section and implies acceptance based on the comparison to predicate devices and established guidelines. For the method comparison, precision, linearity, and detection limits, the "reported device performance" is the direct result of the testing.
Given the nature of the submission (510(k) for substantial equivalence in an in-vitro diagnostic device), the acceptance criteria would typically revolve around demonstrating comparable performance to the predicate devices and adherence to established clinical laboratory standards (CLSI guidelines).
Below is a table summarizing the reported performance, with implied acceptance criteria based on standard practices for demonstrating substantial equivalence for in-vitro diagnostic devices.
Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Implied Acceptance Criteria (Based on Substantial Equivalence and CLSI Guidelines) | Reported Device Performance (VITROS XT 7600 Integrated System with corresponding slides) |
|---|---|---|
| Method Comparison | Device results should show substantial agreement with the predicate device (e.g., slopes near 1, intercepts near 0, demonstrating agreement across the measuring range). | CRBM Serum: N=118, Deming Regression, Slope=1.00, Intercept=0.12, Test range 3.1-17.8 µg/mL. CREA Serum: N=116, Passing Bablock, Slope=0.99, Intercept=0.00, Test range 0.25-13.4 mg/dL. CREA Urine: N=122, Passing Bablock, Slope=0.99, Intercept=-0.45, Test range 3.7-331.0 mg/dL. TBIL Serum: N=125, Passing Bablock, Slope=0.99, Intercept=0.01, Test range 0.14-23.65 mg/dL. |
| Precision | Within-lab precision (Total %CV and SD) should be acceptable for clinical use and comparable to predicate device specifications (though explicit predicate precision isn't stated here, it's an implied comparison). Lower %CV indicates higher precision. | CRBM (Serum): Within Lab (Total) %CV ranges from 2.41% to 3.98% across 6 concentration levels (3.9 to 17.6 µg/mL). |
| CREA (Serum): Within Lab (Total) %CV ranges from 1.40% to 1.85% across 6 concentration levels (0.82 to 12.65 mg/dL). | ||
| CREA (Urine): Within Lab (Total) %CV ranges from 1.55% to 2.23% across 6 concentration levels (55.6 to 320.9 mg/dL). | ||
| TBIL (Serum): Within Lab (Total) %CV ranges from 1.40% to 6.72% across 5 concentration levels (0.3 to 21.6 mg/dL). | ||
| Linearity | The device should demonstrate linearity across its claimed measuring range. | The linearity studies support the claimed measuring ranges for the VITROS CRBM, VITROS CREA, and VITROS TBIL assays. |
| Detection Limits (LoB, LoD, LoQ) | Calculated detection limits should be at or below the claimed LoQ and support the low end of the claimed measuring range. Acceptance typically involves comparing these values to the claimed LoQ. | CRBM: LoB = 0.6108 µg/mL; LoD = 0.6821 µg/mL; LoQ = 2.6860 µg/mL. Claimed LoQ = 3.0 µg/mL. TBIL: LoB = 0.0378 mg/dL; LoD = 0.0722 mg/dL; LoQ = 0.0616 mg/dL. Claimed LoQ = 0.10 mg/dL. Creatinine (Serum/Plasma): LoB = 0.0933 mg/dL; LoD = 0.0991 mg/dL; LoQ = 0.1119 mg/dL. Claimed LoQ = 0.15 mg/dL. Creatinine (Urine): LoB = 1.9973 mg/dL; LoD = 2.1986 mg/dL; LoQ = 2.0060 mg/dL. Claimed LoQ = 3.2 mg/dL. In all cases, the calculated LoQ is at or below the claimed LoQ, supporting the claimed assay range. |
| Specificity (Interference) | Observed bias due to interferents should be within predetermined Maximum Allowable Interference (MAI) or within the 95% Confidence Limit if exceeding Claimed Bias, demonstrating comparable performance to the predicate for known and potential interferents. | Results demonstrate acceptable bias on the VITROS XT 7600 versus the VITROS 5600 for currently claimed interferents. Two previously untested analyte/interferent levels (3.0 ug/mL CRBM/ 20 mg/dL Bilirubin and 3.0 ug/mL CRBM/ 3.0 mg/dL Ethamsylate on CRBM MicroSlides) yielded new information. One new interfering substance, Tolazamide, was identified for CREA(s) MicroSlides. The bias profiles for these demonstrated equivalent magnitudes to the VITROS 5600. The IFU for CRBM and CREA have been updated to claim the additional interfering levels and the new interfering substance. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
-
Method Comparison Test Set:
- CRBM: 118 human serum samples.
- CREA: 116 human serum samples and 122 human urine samples.
- TBIL: 125 human serum samples.
- Data Provenance: The document states "human serum samples" and "human urine samples," implying these are clinical samples. The country of origin and whether the data is retrospective or prospective is not specified.
-
Precision Test Set: For each assay (CRBM, CREA serum, CREA urine, TBIL), the study involved:
- 80 replicates (N=80) for each of the multiple fluid levels (e.g., 6 for CRBM, 6 for CREA serum, 6 for CREA urine, 5 for TBIL). The total number of analyses is much higher (e.g., 80 replicates x 6 levels = 480 for CRBM).
- The samples used were Quality Control fluids and human-based precision fluids.
- Data Provenance: Not specified.
-
Linearity Test Set: A series of eleven proportionally related admixtures of low and high test fluids. Each sample was tested in triplicate.
- Data Provenance: Not specified.
-
Detection Limits (LoB, LoD, LoQ) Test Set:
- LoB: 4 blank samples, tested in replicates of 6 over 3 days, using 3 lots of reagents, 4 samples every day, for a total of 216 observations (72 results per reagent lot).
- LoD: 4 pools of human samples with analyte concentrations close to the expected detection limit, tested in replicates of 6 over 3 days, using 3 lots of reagents, with the 4 human sample pools every day, for a total of 216 observations (72 results per reagent lot).
- LoQ: 4 pools of low level samples, tested in replicates of 4 over 3 days, using 3 lots of reagents, 4 samples every day, for a total of 144 observations (48 results per reagent lot).
- Data Provenance: The LoD and LoQ studies used "human samples." The country of origin and whether the data is retrospective or prospective is not specified.
-
Specificity (Interference) Test Set: Chemical interferents, common chemical substances, and claimed non-interferents, including hemoglobin, bilirubin, and intralipid. Testing employed "paired-difference" assessment at a minimum of two analyte levels.
- Data Provenance: Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device measures quantitative concentrations of specific analytes (Carbamazepine, Creatinine, Total Bilirubin). Ground truth for these types of in vitro diagnostic devices usually refers to the reference method (predicate device in this case) or a highly accurate reference standard rather than expert interpretation in the way it applies to image analysis or clinical diagnosis. The document does not mention human experts establishing ground truth in the context of radiologists or similar clinical diagnosticians. The ground truth for the method comparison study was established by the predicate device, VITROS 5600 Integrated System.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
None mentioned. Adjudication methods are typically used when there's a subjective element to ground truth establishment, often involving multiple human readers for diagnostic image interpretation. For quantitative measurements in clinical chemistry, the "truth" is established by reference methods, precision, and linearity studies, not by human adjudication of results.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation, particularly in radiology or pathology, and often involves AI assistance. This document describes an automated in-vitro diagnostic device for quantitative chemical measurements, where human interpretation of results is direct measurement rather than subjective assessment. Therefore, the concept of "human readers improving with AI assistance" is not applicable here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies described are, in essence, standalone performance evaluations of the VITROS XT 7600 Integrated System itself, with the VITROS Chemistry Products slides, operating automatically without continuous human intervention during the measurement process. The system performs the tests, generates results, and its performance (method comparison, precision, linearity, detection limits, specificity) is evaluated. The comparison is against a predicate device, which is also an automated system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the test set, especially for the method comparison, was established by comparison to the legally marketed predicate device (VITROS 5600 Integrated System), which itself would have been previously cleared based on demonstrating accuracy against established reference methods or accepted gold standards for each analyte. For precision, linearity, and detection limits, the ground truth is established by the known characteristics of reference materials and statistical analysis.
8. The sample size for the training set
The document does not mention a training set. This is because the device described is not an AI/ML-based diagnostic algorithm that learns from data. It is a traditional in-vitro diagnostic instrument with chemical reagent slides. The studies are validation studies for the performance of the integrated system, not for training an algorithm.
9. How the ground truth for the training set was established
Since there is no training set mentioned or used for this type of device, this question is not applicable.
Ask a specific question about this device
(60 days)
Systems BILIRUBIN TOTAL 4+1, ELITech Clinical Systems BILIRUBIN DIRECT 4+1 Regulation Number: 21 CFR 862.1110
Clinical Chemistry |
| | | 21 CFR 862.1110
Clinical Chemistry |
| | | 21 CFR 862.1110
ELITech Clinical Systems BILIRUBIN TOTAL 4+1 is intended for the quantitative in vitro diagnostic determination of total bilirubin in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers.
ELITech Clinical Systems BILIRUBIN DIRECT 4+1 is intended for the quantitative in vitro diagnostic determination of direct bilirubin in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers.
It is not intended for use in Point of Care settings.
Measurements of the levels of bilirubin (direct or total), an organic compound formed during the normal and abnormal distruction of red blood cells, are used in the diagnosis and treatment of liver, hematological, and metabolic disorders, including hepatitis and gall bladder block.
ELITech Clinical Systems BILIRUBIN TOTAL 4+1 and ELITech Clinical Systems BILIRUBIN DIRECT 4+1 are available as a kit only. Each kit consists of a bi-reagent R1 & R2.
ELITech Clinical Systems BILIRUBIN TOTAL 4+1:
Reagent 1: R1 Sulphanilic acid 29 mmol/L, Cetrimide 29 mmol/L.
Reagent 2: R2 Sodium nitrite 11 mmol/L.
ELITech Clinical Systems BILIRUBIN DIRECT 4+1:
Reagent 1: R1 Sulphanilic acid 29 mmol/L,
Reagent 2: R2 Sodium nitrite 11 mmol/L.
The document describes the analytical performance and comparison studies for the ELITech Clinical Systems BILIRUBIN TOTAL 4+1 and ELITech Clinical Systems BILIRUBIN DIRECT 4+1 devices. These devices are intended for the quantitative in vitro diagnostic determination of total and direct bilirubin, respectively, in human serum and plasma. The studies aim to demonstrate substantial equivalence to predicate devices (ABX Pentra Bilirubin, Total CP & ABX Pentra Bilirubin, Direct CP).
Here's the breakdown of the information based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
For ELITech Clinical Systems BILIRUBIN TOTAL 4+1:
| Acceptance Criteria Category | Specific Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Precision | Within-run CV% | Not explicitly stated as a numerical criterion, but evaluated by comparing with predicate device and clinical requirements for bilirubin assays. | Level 1 (1.15 mg/dL): 1.8%Level 2 (4.08 mg/dL): 0.4%Level 3 (14.61 mg/dL): 0.5% |
| Total CV% | Not explicitly stated as a numerical criterion. | Level 1 (1.15 mg/dL): 5.0%Level 2 (4.08 mg/dL): 3.1%Level 3 (14.61 mg/dL): 2.9% | |
| Linearity/Reportable Range | Measuring Range | Not explicitly stated as a numerical criterion, but evaluated to cover clinically relevant range. | 0.25 - 25 mg/dL (up to 60.00 mg/dL with auto-dilution) |
| Detection Limit | Limit of Detection (LoD) | Not explicitly stated as a numerical criterion. | 0.04 mg/dL (0.7 µmol/L) |
| Quantification Limit | Limit of Quantification (LoQ) | ≤ 0.07 mg/dL Total Error, and ≥ LoD. | 0.15 mg/dL (2.6 µmol/L) |
| Interference | Bias | Acceptance criterion: ±10% bias for nominal activities of 1.00 mg/dL and 15.00 mg/dL. | Triglycerides: No significant interference up to 2100 mg/dLHemoglobin: No significant interference up to 500 mg/dLAcetaminophen: No significant interference up to 30 mg/dLAscorbic acid: No significant interference up to 4 mg/dLAcetylsalicylic acid: No significant interference up to 200 mg/dL |
| Method Comparison | Correlation (r) | Not explicitly stated as a numerical criterion, but assessed for strong correlation with predicate. | r = 0.999 |
| Coefficient of Determination (r²) | Not explicitly stated as a numerical criterion. | r² = 0.999 | |
| Standard error of the estimate (Sy.x) | Not explicitly stated as a numerical criterion. | 0.19 mg/dL | |
| Matrix Comparison | Correlation (r) | Not explicitly stated as a numerical criterion. | r = 0.998 |
| Coefficient of Determination (r²) | Not explicitly stated as a numerical criterion. | r² = 0.997 | |
| Standard error of the estimate (Sy.x) | Not explicitly stated as a numerical criterion. | 0.36 mg/dL |
For ELITech Clinical Systems BILIRUBIN DIRECT 4+1:
| Acceptance Criteria Category | Specific Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Precision | Within-run CV% | Not explicitly stated as a numerical criterion. | Level 1 (0.36 mg/dL): 3.8%Level 2 (1.51 mg/dL): 1.9%Level 3 (3.99 mg/dL): 0.9% |
| Total CV% | Not explicitly stated as a numerical criterion. | Level 1 (0.36 mg/dL): 5.2%Level 2 (1.51 mg/dL): 5.3%Level 3 (3.99 mg/dL): 4.7% | |
| Linearity/Reportable Range | Measuring Range | Not explicitly stated as a numerical criterion. | 0.08 - 10.55 mg/dL (up to 50.00 mg/dL with auto-dilution) |
| Detection Limit | Limit of Detection (LoD) | Not explicitly stated as a numerical criterion. | 0.01 mg/dL (0.2 µmol/L) |
| Quantification Limit | Limit of Quantification (LoQ) | ≤ 0.05 mg/dL Total Error, and ≥ LoD. | 0.08 mg/dL (1.4 µmol/L) |
| Interference | Bias | Acceptance criterion: ±10% bias for nominal activities of 0.40 mg/dL and 4.00 mg/dL. | Triglycerides: No significant interference up to 2000 mg/dLHemoglobin: No significant interference up to 125 mg/dLAcetaminophen: No significant interference up to 30 mg/dLAscorbic acid: No significant interference up to 0.5 mg/dLAcetylsalicylic acid: No significant interference up to 200 mg/dL |
| Method Comparison | Correlation (r) | Not explicitly stated as a numerical criterion. | r = 0.998 |
| Coefficient of Determination (r²) | Not explicitly stated as a numerical criterion. | r² = 0.995 | |
| Standard error of the estimate (Sy.x) | Not explicitly stated as a numerical criterion. | 0.15 mg/dL | |
| Matrix Comparison | Correlation (r) | Not explicitly stated as a numerical criterion. | r = 0.999 |
| Coefficient of Determination (r²) | Not explicitly stated as a numerical criterion. | r² = 0.997 | |
| Standard error of the estimate (Sy.x) | Not explicitly stated as a numerical criterion. | 0.14 mg/dL |
2. Sample Size Used for the Test Set and Data Provenance
-
Precision (Test Set):
- Sample Size: 80 measurements for each of 3 levels of samples (total 240 measurements per device) per instrument (2 instruments used).
- Data Provenance: Not explicitly stated (e.g., country of origin). The samples are referred to as "samples" or "patient samples." The study was prospective in the sense that controlled experiments were performed to gather data for precision.
-
Linearity (Test Set):
- ELITech Clinical Systems BILIRUBIN TOTAL 4+1: 11 levels of mixed samples.
- ELITech Clinical Systems BILIRUBIN DIRECT 4+1: 11 levels of mixed samples.
- Data Provenance: Not explicitly stated. Likely prepared in a laboratory setting.
-
Detection Limit (Test Set):
- Sample Size: 15 measurements of 4 samples for each device.
- Data Provenance: Prepared from patient samples and diluted with Albumin 6 g/dL - NaCl 0.9 %.
-
Interference (Test Set):
- Sample Size: For each potential interferent, 2 serum sample pools (low and high concentration), with aliquots spiked at various interferent concentrations (9 or 7 or 8 different concentrations). Each point was measured in triplicate per run. Two levels of control were also tested. A control sample was prepared from the sample pool diluted in appropriate diluent.
- Data Provenance: "Serum sample pools." Not explicitly stated.
-
Method Comparison (Test Set):
- Sample Size: 100 serum patient samples for each device.
- Data Provenance: "Serum patient samples." Not explicitly stated, but likely from a clinical laboratory or hospital in the country where the studies were conducted (France or USA, as these are locations given for the submitter and contact person). The study was likely prospective in nature for collecting these samples for comparison against the predicate.
-
Matrix Comparison (Test Set):
- Sample Size: 40 plasma patients (lithium heparin samples) for each device.
- Data Provenance: "Plasma patients." Not explicitly stated. Likely from a clinical laboratory or hospital.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
For this type of in vitro diagnostic device (quantitative measurement of bilirubin), the "ground truth" is typically established by reference methods or comparison to legally marketed predicate devices, not by expert interpretation. The predicate devices are considered the "ground truth" for the method comparison studies. The document implicitly uses the performance of the legally marketed predicate device (ABX Pentra Bilirubin, Total CP & ABX Pentra Bilirubin, Direct CP) as the standard against which the new device's performance is compared for substantial equivalence.
There is no mention of experts establishing ground truth in the traditional sense of medical imaging or clinical diagnosis.
4. Adjudication Method for the Test Set
Not applicable. This is an in vitro diagnostic device performing quantitative measurements, not an AI or imaging device requiring human adjudication of results. The performance is assessed by comparing quantitative results against reference or predicate methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is typically for imaging devices where human readers interpret medical images with and without AI assistance. The described device is an in vitro diagnostic assay.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Yes, the studies described are standalone performance evaluations of the device (reagents on an analyzer). The device itself (the reagent system) functions as an "algorithm only" in the sense that it performs the chemical reaction and measurement without human interpretive input for each test result. Clinical laboratory technologists would operate the analyzer and interpret the numerical results, but the analytical performance described is the device's intrinsic capability.
7. The Type of Ground Truth Used
The primary ground truth for demonstrating substantial equivalence is the performance of the legally marketed predicate devices (ABX Pentra Bilirubin, Total CP & ABX Pentra Bilirubin, Direct CP). In the method comparison studies, the results from the new devices are compared against the results from the predicate devices using patient samples.
Additionally, for analytical performance like linearity, detection limit, and interference, the "ground truth" for evaluating the performance of the device is based on prepared samples with known concentrations or expected behaviors (e.g., spiked samples with interferents, serially diluted samples).
8. The Sample Size for the Training Set
Not applicable. This device is an in vitro diagnostic reagent system, not an AI/ML-based device that requires a "training set" in the computational sense. Its performance is based on chemical reactions and instrumental measurements, which are validated through analytical studies.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this type of device.
Ask a specific question about this device
(59 days)
: K170065
Trade/Device Name: ADVIA® Chemistry Total Bilirubin 2 (TBIL 2) Regulation Number: 21 CFR 862.1110
system;Bilirubin (total and unbound) in the neonate system |
| Regulation Section: | 21 CFR §862.1110
system;Bilirubin (total and unbound) in the neonate system |
| Regulation Section: | 21 CFR §862.1110
For in vitro diagnostic use in the quantitative determination of total bilirubin in serum and plasma of adults and neonates on the ADVIA® Chemistry systems. Measurement of total bilirubin, an organic compound formed and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
The ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) reagents are liquid ready to use. They are packaged as a kit with two kit sizes available as follows.
Kit Size – 70 mL Wedge Reagent 1 and 70 mL Reagent 2 Wedge
Reagent 1: 4 wedges x 68 mL
Reagent 2: 4 wedges x 25 mL
Kit Size - 40 mL Reagent 1 and 20 mL Reagent 2 Wedge
Reagent 1: 4 wedges x 38 mL
Each reagent kit consists of reagents of components and concentrations summarized below.
Reagent 1: Citrate buffer, pH 2.9 (0.1 mol/L); Detergent
Reagent 2: Phosphate buffer, pH 7.0 (10mmol/L); Sodium metavanadate (4 mmol/L)
Here's a breakdown of the acceptance criteria and study information for the ADVIA® Chemistry Total Bilirubin 2 (TBIL 2) device, based on the provided document:
Acceptance Criteria and Device Performance
The document doesn't explicitly state "acceptance criteria" for each performance characteristic as a distinct set of pre-defined thresholds. Instead, it presents the results of validation studies for various parameters. However, we can infer the implicit "acceptance criteria" by looking at industry standards (like CLSI guidelines cited) and typical performance expectations for such devices. The "reported device performance" directly comes from the study results.
Note: For some parameters, the "acceptance criteria" are implied by the method and regulatory guidelines (e.g., CLSI EP09-A3 for method comparison, which focuses on demonstrating accuracy through strong correlation and acceptable bias at medical decision points).
| Performance Characteristic | Implicit Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Method Comparison | Strong correlation (r value close to 1), low bias at medical decision levels, demonstrating accuracy compared to a legally marketed comparator. (Based on CLSI EP09-A3) | N: 119Range (ADVIA®): 0.7 – 31.6 mg/dLRange (Comparator): 0.8 – 26.6 mg/dLSlope: 1.06y-intercept: -0.24Correlation coefficient (r): 0.990Bias at MDLs: 1.0 mg/dL (-0.2 mg/dL), 13.0 mg/dL (0.5 mg/dL), 17.0 mg/dL (0.8 mg/dL) |
| Analytical Measuring Range/Linearity | Demonstrated linearity across the claimed measuring range, with a slope close to 1 and an r value close to 1. (Based on CLSI EP06-A) | Slope: 0.999y-intercept: 0.016r: 0.999Number of Levels: 9Observed Sample Range: 0.0-39.2 mg/dLAnalytical Measuring Range: 0.15-35.0 mg/dL |
| Limits of Detection and Quantitation | Documented LoB, LoD, and LoQ based on experimental determination following CLSI guidelines. (Based on CLSI EP17-A2 and EP05-A2) | LoB: 0.02 mg/mLLoD: 0.06 mg/dLLoQ: 0.08 mg/dL |
| Interferences | Bias or recovery of interferent to blank within ±10% for relevant substances. (Based on CLSI EP07-A2) | Acceptable with ≤10% bias or recovery for: - Indican: 10 mg/dL- Cyanokit: 40 ug/mL- HbF: 1000 mg/dL- HbA: 1000 mg/dL |
| Expected Values (Reference Interval) | Reference intervals established or verified in accordance with CLSI guidelines and supported by literature. (Based on CLSI EP28-A3c and Wu AHB. Tietz Clinical Guide) | Verified expected values: - 0-1 day: <8.0 mg/dL- 1-2 days: <12.0 mg/dL- 3-5 days: <16.0 mg/dL- >5 days – 60 years: 0.3-1.2 mg/dL- 60 - 90 years: 0.2-1.1 mg/dL- >90 years: 0.2-0.9 mg/dL (Reference: Wu AHB. Tietz Clinical Guide to Laboratory Tests, 4th edition, 2006:172) |
Detailed Study Information:
The provided document describes analytical performance studies for the ADVIA® Chemistry Total Bilirubin 2 (TBIL 2) device. It is a standalone (algorithm only without human-in-the-loop performance) study, as it evaluates the analytical performance of a clinical chemistry assay, not a diagnostic imaging device with human interpretation.
-
Sample Size Used for the Test Set and Data Provenance:
- Method Comparison: N = 119 patient samples.
- Analytical Measuring Range/Linearity: The document states "9 levels" for linearity but does not specify the number of individual samples tested at each level or overall (implied to be an internally prepared linearity panel).
- Limits of Detection and Quantitation: Not explicitly stated for specific test sets, usually involves multiple replicates of blank and low-concentration samples.
- Interferences: Not explicitly stated for specific test sets; involved samples with low and high concentrations of bilirubin plus various interferents.
- Data Provenance: Not explicitly stated. These are typically laboratory-generated samples or de-identified patient samples obtained for research purposes within the testing laboratory's region. The adult population data were previously cleared under K063845, implying this testing focused on neonatal-specific aspects or reaffirming general performance on the new instrument.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- For this type of in vitro diagnostic device, "ground truth" is established through well-characterized reference methods or highly accurate comparator devices/methods. There are no "experts" in the human interpretation sense (like radiologists) involved in establishing the ground truth for these analytical measurements.
- The "comparator method" for the method comparison study served as the reference for ground truth in that context. Its specifics (e.g., gold standard, reference material) are not detailed beyond being a "legally marketed comparator method."
-
Adjudication Method for the Test Set:
- Not applicable. This is an analytical performance study of a quantitative assay, not a study involving human readers' interpretations of images or clinical reports requiring adjudication.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC study was not done. This document describes the analytical performance of an in vitro diagnostic assay, which traditionally does not involve human readers interpreting "cases" in the way an imaging device might.
-
If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone study was done. The entire document details the analytical performance of the ADVIA® Chemistry Total Bilirubin 2 (TBIL 2) assay on an automated instrument (ADVIA® Chemistry 1800 System) without human interpretive input for the final result beyond loading samples and running the assay.
-
The Type of Ground Truth Used:
- For the Method Comparison study, the ground truth was established by a legally marketed comparator method.
- For Linearity, LoD/LoQ, and Interference studies, the ground truth involves carefully prepared samples (e.g., spiked samples, diluted samples, reference materials) with known concentrations or expected responses, tested against established analytical validation protocols.
- For Expected Values (Reference Interval), the ground truth was established by literature reference (Wu AHB. Tietz Clinical Guide to Laboratory Tests, 4th edition, 2006:172) and verified according to CLSI guidelines.
-
Sample Size for the Training Set:
- Not applicable. This document describes the validation of a finished assay and instrument system. Clinical chemistry assays are developed and optimized through iterative research and development, but there isn't a "training set" in the machine learning sense. The "reagent formulation and method parameters" (mentioned as remaining the same for adult claims from K063845) represent the output of prior development.
-
How the Ground Truth for the Training Set Was Established:
- Not applicable. As a traditional in vitro diagnostic assay, the concept of a "training set" and associated ground truth is not relevant in the machine learning context. The assay's performance is governed by its chemical reaction principle and instrument calibration/characteristics.
Ask a specific question about this device
(181 days)
BRITAIN
February 16, 2016
Re: K152343
| Trade/Device Name: Direct Bilirubin Regulation Number: 21 CFR 862.1110 |
|---|
| JFM |
The Direct Bilirubin test system is a device intended for the quantitative in vitro determination of Direct Bilirubin in serum and plasma. Bilirubin measurements can be used in the diagnosis and treatment of liver, hematological and metabolic disorders including hepatitis and gall bladder block.
This device is for prescription use only.
The Randox Direct Bilirubin kit consists of ready to use reagent solutions.
CATALOGUE NUMBER: BR8308 COMPONENTS: R1. 4 x 20ml, R2. 4 x 8ml
REAGENT COMPOSITION
R1. Direct Bilirubin RI Tartrate buffer, pH2.9 Detergent Antimicrobials and Preservatives Inhibitors Initial Concentration of Solutions 0.1 mol/L
R2. Direct Bilirubin R2 Phosphate buffer, pH 7.0 Sodium Metavanadate Initial Concentration of Solutions 10 mmol/L 4 mmol/L
Here's a breakdown of the acceptance criteria and the study details for the Randox Direct Bilirubin device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document provides performance characteristics but does not explicitly state "acceptance criteria" in a tabulated format derived from a regulatory body or a specific standard with pass/fail thresholds. Instead, it presents the results of various analytical performance studies. However, some sections imply acceptance criteria through their phrasing (e.g., "deviation from linearity is less than 5%" for linearity, "no significant interference" for specificity, and "≤20% CV imprecision" for LoQ).
Here's an interpretation of implied acceptance criteria and reported performance:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Precision/Reproducibility: Repeatability and intermediate precision within acceptable limits. | Serum Pool 1 (Mean 0.65 mg/dl): Within Run CV: 3.0%, Among Run CV: 1.6%, Among Day CV: 2.4%, Total CV: 4.2% Serum Pool 2 (Mean 2.31 mg/dl): Within Run CV: 3.1%, Among Run CV: 0.0%, Among Day CV: Not reported (likely very low, as next cell is blank), Total CV: 3.1% Serum Pool 4 (Mean 8.41 mg/dl): Within Run CV: 1.5%, Among Run CV: 0.8%, Among Day CV: 0.9%, Total CV: 1.9% |
| Linearity/Reportable Range: Linear function to analyte concentration (deviation from linearity < 5%). | Linearity: 0.1 - 12.6 mg/dl. Slope: 1.01, Intercept: -0.07, r: 0.9995, Syx: 0.10. Reportable Range: 0.1 - 12.6 mg/dl (with auto-dilution for >12.6 mg/dl). |
| Detection Limit (LoQ): Lowest concentration detectable with ≤20% CV imprecision. | LoD: 0.064 mg/dl LoB: 0.006 mg/dl LoQ: 0.133 mg/dl (confirmed to be ≤20% CV imprecision). |
| Analytical Specificity/Interference: No significant interference from common interferents at specified levels (Ac-ceptance Criteria: % of Control ± 10%). | Haemoglobin: No significant interference up to 1000 mg/dl Triglycerides: No significant interference up to 750 mg/dl Intralipid®: No significant interference up to 1000 mg/dl Ascorbic Acid: No significant interference up to 25 mg/dl (This suggests the results fell within ± 10% of the control). |
| Method Comparison with Predicate Device: Strong correlation with the predicate device. | Correlation Coefficient (r): 0.997 (for 103 serum patient samples spanning 0.123-12.46 mg/dl). Regression Equation: Y = 1.01x + 0.01. |
| Matrix Comparison: Agreement between serum and lithium heparin plasma samples. | Correlation Coefficient (r): 1.00 (for a minimum of 40 matched patient sample pairs spanning 0.091-12.48 mg/dl). Regression Equation: Y = 0.99x + 0.01. |
2. Sample Size Used for the Test Set and Data Provenance
- Precision/Reproducibility:
- Sample Size: Not explicitly stated as a number of samples but rather as "control material and unaltered human serum samples" divided into pools (Pool 1, 2, 4) and tested over 20 non-consecutive days with 2 replicates per run. This implies a significant number of measurements for each pool.
- Data Provenance: "unaltered human serum samples." The country of origin is not specified, but the submission is from the UK. The study is prospective in nature (testing conducted for the device).
- Linearity/Assay Reportable Range:
- Sample Size: Samples prepared at 11 levels. Each level run in replicates of five.
- Data Provenance: Not specified, but samples were prepared to cover a range of analyte concentrations. Prospective.
- Detection Limit:
- Sample Size: 240 determinations (for LoD) with 4 low-level samples.
- Data Provenance: Not specified. Prospective.
- Analytical Specificity/Interference:
- Sample Size: Not explicitly stated as a number of samples, but "the analytes detailed below were tested up to the levels indicated at Bilirubin concentrations of 0.14mg/dl and 5.03mg/dl."
- Data Provenance: Not specified. Prospective.
- Method Comparison with Predicate Device:
- Sample Size: 103 serum patient samples.
- Data Provenance: "patient samples." The country of origin is not specified. Likely retrospective, as existing patient samples were used, but the testing itself was prospective.
- Matrix Comparison:
- Sample Size: A minimum of 40 matched patient sample pairs.
- Data Provenance: "Patient samples." The country of origin is not specified. Likely retrospective, as existing patient samples were used, but the testing itself was prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This device is an in vitro diagnostic (IVD) test for quantitative determination of Direct Bilirubin. The "ground truth" for such devices is typically established through reference methods or established predicate devices, not through expert consensus in the same way an imaging or pathology AI might.
- Precision, Linearity, Detection Limit, Specificity: Ground truth is inherent in the analytical methods themselves (e.g., gravimetric dilutions for linearity, spiked samples, controlled interferent concentrations). No external experts are mentioned.
- Method Comparison and Matrix Comparison: The ground truth for these studies is the measurement obtained by the predicate device (Wako Direct Bilirubin V, K053132) or the matched serum/plasma results themselves. No external experts are described as establishing this "ground truth."
4. Adjudication Method for the Test Set
Not applicable for this type of IVD device. Adjudication is typically used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is an automated in vitro diagnostic test, not an AI-assisted human reading device.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies presented are all standalone performance evaluations of the Randox Direct Bilirubin assay system, an automated IVD device. The results reported are directly from the instrument measurements.
7. The Type of Ground Truth Used
- For analytical performance studies (Precision, Linearity, Detection Limit, Specificity): The 'ground truth' is established through controlled laboratory preparations (e.g., known concentrations of analytes, spiked samples, dilutions) and comparisons to established analytical methods as described in CLSI guidelines.
- For method comparison: The 'ground truth' is the predicate FDA-cleared device (Wako Direct Bilirubin V, K053132).
- For matrix comparison: The 'ground truth' is the serum sample measurement when comparing to lithium heparin plasma.
8. The Sample Size for the Training Set
Not applicable. This is a chemical assay, not a machine learning model that requires a training set. The "development" or "optimization" phase of such an assay would involve internal R&D, but it's not a "training set" in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for an IVD chemical assay as described here. Parameter settings and reagent formulations are determined through standard chemical and biochemical R&D processes, not through machine learning ground truth establishment.
Ask a specific question about this device
(162 days)
January 28, 2016
Re: K152344
| Trade/Device Name: Total Bilirubin (T BIL) Regulation Number: 21 CFR 862.1110 |
|---|
| JFM |
For the quantitative in vitro determination of Total Bilirubin for serum and plasma. Total Bilirubin measurements are used in the diagnosis and treatment of hemolytic, biliary and liver disorders, including hepatitis and cirrhosis.
This in vitro diagnostic device is intended for prescription use only.
The Total Bilirubin kit assay consists of ready to use reagent solutions.
CATALOGUE NUMBER: BR8307
R1. Total Bilirubin R1 4 x 20 mL
R2. Total Bilirubin R2 4 x 8 mL
REAGENT COMPOSITION
Contents Initial Concentration of Solutions
R1. Total Bilirubin R1
Citrate buffer, pH2.9 0.1 mol/L
Detergent 0.9%
Antimicrobial
R2. Total Bilirubin R2
Phosphate buffer, pH 7.0 10 mmol/L
Sodium Metavanadate 4 mmol/L
MATERIALS REQUIRED BUT NOT PROVIDED
Randox Assayed Multisera Level 2 (Cat. No. HN 1530) and Level 3 (Cat. No. HE 1532); 510(k) # K942458 Randox Calibration Serum Level 3 (Cat. No. CAL 2351); 510(k) # K053153 RX series Saline (Cat. No. SA 8396)
Here's an analysis of the provided text, focusing on the acceptance criteria and the studies conducted to meet them for the Total Bilirubin (T BIL) device.
Acceptance Criteria and Reported Device Performance
| Criteria Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Linearity/Reportable Range | Deviation from linearity less than 5% across the analytical range. | Linear regression correlation (r) = 0.9999 for the range 0.21 – 26.3 mg/dL. The reported range is 0.21 – 26.3 mg/dL. (Implies linearity within this range was met). |
| Limit of Detection (LoD) | Not explicitly stated as an acceptance criterion for the study, but determined. | LoD = 0.08 mg/dL (based on 240 determinations, 4 low-level samples). |
| Limit of Blank (LoB) | Not explicitly stated as an acceptance criterion for the study, but determined. | LoB = 0.06 mg/dL. |
| Limit of Quantitation (LoQ) | Not explicitly stated as an acceptance criterion for the study, but determined. | LoQ = 0.21 mg/dL (lowest concentration at which precision is still met). |
| Analytical Specificity (Interference) | Recovery within ±10% of the initial value of Total Bilirubin concentration (0.99 mg/dL and 15.03 mg/dL) for specified interferents. | Haemoglobin: No significant interference up to 1000 mg/dL. Triglycerides: No significant interference up to 2000 mg/dL. Intralipid®: No significant interference up to 1000 mg/dL. Ascorbic Acid: No significant interference up to 25.0 mg/dL. (All met the ±10% recovery implicitly). |
| Method Comparison (with predicate device) | Not explicitly stated as an acceptance criterion (e.g., a specific agreement or bias limit), but "substantial equivalence" is the overall goal. | Linear regression equation: Y = 1.02x - 0.02. Correlation coefficient (r) = 0.9999. (This high correlation supports substantial equivalence). |
| Matrix Comparison (Serum vs. Plasma) | Not explicitly stated as an acceptance criterion, but the goal is for method accuracy with plasma to be equivalent to serum and no interference. | Linear regression equation: Y = 0.99x + 0.04. Correlation coefficient (r) = 0.9999. (This high correlation suggests equivalence). |
| Expected/Reference Values | Verified using NCCLS C28-A3 guidelines; all values from 30 normal donors fall within the quoted range for healthy individuals (0.3 – 1.2 mg/dL). | All values from the 30 normal donors tested on the RX Daytona plus fell within the quoted ranges for Healthy Individuals (0.3 – 1.2 mg/dL). |
| Precision/Reproducibility | Not explicitly stated as an acceptance criterion (e.g., a maximum CV%), but detailed results are provided. | See Table 2 (page 6) for detailed SD and CV values across different concentrations for Within Run, Among Run, Among Day, and Total precision. For example, for a mean of 25.0 mg/dL, Total CV was 1.7%; for 0.3 mg/dL, Total CV was 7.4%. These values are typically considered acceptable for clinical assays. |
Study Details
-
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Precision/Reproducibility: 80 determinations per sample type (LIN Pool, LOQ Pool, QC 1, QC 2, Serum Pool 1-4) across two lots, two RX Daytona plus systems, over 20 non-consecutive days with 2 replicates per run. Data provenance is not specified, but the study design suggests prospective testing of control materials and human serum samples (spiked or diluted).
- Linearity/Assay Reportable Range: 11 levels, each run in replicates of five across two lots of reagent on one RX Daytona plus system. Data provenance is not specified.
- Detection Limit: 240 determinations (for LoD) using 4 low-level samples. Data provenance is not specified.
- Analytical Specificity (Interference): Not explicitly stated how many samples or replicates per interferent. Samples were spiked with interferents and compared to control samples. Data provenance is not specified.
- Method Comparison: 106 serum patient samples spanning 0.21 to 26.9 mg/dL. Data provenance is not specified, but these are "patient samples," suggesting retrospective or prospective clinical samples.
- Matrix Comparison: A minimum of 40 matched patient sample pairs (serum and lithium heparin plasma). Data provenance is not specified, but these are "patient samples," suggesting retrospective or prospective clinical samples.
- Expected values/Reference range: Human serum from 30 normal donors. Data provenance is not specified. The study was prospective in nature, testing new samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This is an in vitro diagnostic device for quantitative chemical analysis. The "ground truth" is established by the analytical reference methods or established values of control materials, or comparison to a predicate device. There is no mention of human experts defining ground truth through consensus in the way a radiological study might. For the method comparison, the predicate device (Siemens Healthcare Diagnostic Inc, Total Bilirubin 2 reagent, K063845) serves as the reference, which itself would have undergone rigorous analytical validation.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is an analytical chemistry device, not an imaging device requiring human adjudicated interpretations. The performance is assessed by quantitative analytical metrics.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is an analytical chemistry device, not an AI-assisted diagnostic tool involving human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This is an in vitro diagnostic assay, which by nature operates "standalone" in terms of measurement. The results are then interpreted by a clinician, but the device itself generates a quantitative result without human-in-the-loop performance influencing the measurement. Performance studies like precision, linearity, and analytical specificity are inherently "standalone" evaluations of the device's analytical function.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Reference Methods/Materials: For linearity and precision, the ground truth is based on gravimetrically prepared samples, known concentrations of control materials, or dilutions with expected values.
- Predicate Device: For method comparison, the results from the legally marketed predicate device (Siemens Healthcare Diagnostic Inc, Total Bilirubin 2 reagent, K063845) serve as the comparative ground truth.
- Literature/Guidelines: For reference range verification, established normal ranges from scientific literature (e.g., "Tietz Clinical Guide to laboratory Tests") and guidelines (NCCLS C28-A3) are used.
-
The sample size for the training set
- Not applicable. This is an analytical chemistry device, not a machine learning model that requires a training set.
-
How the ground truth for the training set was established
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 6