Search Results
Found 600 results
510(k) Data Aggregation
(87 days)
Ask a specific question about this device
(111 days)
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(111 days)
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(16 days)
Elecsys Anti-SARS-CoV-2 is an immunoassay intended for the in vitro qualitative detection of total antibodies to Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) in human serum and Li-heparin, K2-EDTA and K3-EDTA plasma collected on or after 15 days post-symptom onset. The test is intended as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e 601 immunoassay analyzer.
Elecsys Anti-SARS-CoV-2 is a qualitative, serological, double-antigen sandwich principle immunoassay to be used on the cobas e 601 analyzer with an 18-minute test time. Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration. The Elecsys Anti‑SARS‑CoV-2 assay uses a recombinant protein representing the nucleocapsid (N) antigen for the determination of antibodies against SARS‑CoV‑2.
The reagent working solutions include: rackpack (kit placed on the analyzer)
- M Streptavidin-coated microparticles (transparent cap), 1 bottle, 12 mL: Streptavidin-coated microparticles 0.72 mg/mL; preservative.
- R1 SARS-CoV-2-Ag~biotin, (gray cap), 1 bottle, 16 mL: Biotinylated SARS‑CoV‑2‑specific recombinant antigen (E. coli) < 0.5 mg/L; HEPES^a) buffer 50 mmol/L, pH 7.7; preservative.
- R2 SARS-CoV-2 Ag~Ru(bpy) (black cap), 1 bottle, 16 mL: SARS‑CoV‑2‑specific recombinant antigen labeled with ruthenium complex < 0.5 mg/L; HEPES^(b) buffer 50 mmol/L, pH 7.7; preservative.
^(a) HEPES = [4-(2-hydroxyethyl)-piperazine]-ethane sulfonic acid
The Elecsys Anti-SARS-CoV-2 assay includes two liquid calibrators, which are packed with the test kit:
- ACOV2 Cal1 Negative calibrator 1 (white cap), 2 bottles of 0.67 mL: Human serum, non-reactive for anti‑SARS‑CoV‑2 antibodies; buffer; preservative.
- ACOV2 Cal2 Positive calibrator 2 (black cap), 2 bottles of 0.67 mL: Human serum, reactive for anti‑SARS‑CoV‑2 antibodies; buffer; preservative.
N/A
Ask a specific question about this device
(350 days)
The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.
The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.
The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated multiplex real-time polymerase chain reaction (PCR) assay for the rapid in vitro qualitative detection and differentiation of B. pertussis (Bp), B. parapertussis (Bpp), and B. holmesii (Bh) DNA in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.
The different fluorescent dye designs enable the specific detection and differentiation of the three microorganisms (Bp, Bpp, and Bh) independently in a multiplex system. The system automates all nucleic acid amplification test sample processing steps, including inhibitor removal, nucleic acid extraction, purification, amplification, real-time detection, and result interpretation in a rapid manner. The test is designed for use in near-patient settings to deliver results in approximately 15 minutes.
The cobas® liat system is comprised of the cobas® liat analyzer (analyzer) hardware with integrated cobas® liat system software (analyzer software + liat assay specific package (script)) for running tests and analyzing the results, and a single-use disposable cobas® liat assay tube (assay tube).
-
cobas® liat analyzer is a system component that consists of one software subsystem and three hardware units:
- Infrastructure unit, which consists of the hardware and embedded software (firmware).
- Thermal, loading and motion unit: the processing module that interacts physically with the assay tube during the assay execution.
- Detection unit consisting of the photodetectors that is used for the fluorescence detection during the PCR reaction
-
The assay script provides a set of instructions to the analyzer hardware and software for assay tube processing, PCR, result calculation and interpretation and result reporting. The assay script can be installed on the analyzer independently of the analyzer software.
The cobas® liat Bordetella panel nucleic acid test is supported with a Liat Assay Specific Package (Assay Script): cobas® liat Bordetella panel LASP (BPTA).
- The assay tube holds all reagents needed for sample preparation and PCR processes. The assay specific reagents are packaged into a single assay tube in separate segments that are separated by frangible seals. An internal control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through all steps of the assay process and to monitor the presence of inhibitors in the PCR reaction.
N/A
Ask a specific question about this device
(88 days)
cobas® CMV is an in vitro nucleic acid amplification test for the quantitation of cytomegalovirus (CMV) DNA in human EDTA plasma.
cobas® CMV is intended for use as an aid in the management of CMV in solid organ transplant patients and in hematopoietic stem cell transplant patients. In patients receiving anti-CMV therapy, serial DNA measurements can be used to assess viral response to treatment.
The results from cobas® CMV must be interpreted within the context of all relevant clinical and laboratory findings.
cobas® CMV is not intended for use as a screening test for blood or blood products.
cobas® CMV is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 5800 system is designed as one integrated instrument. The cobas® 6800/8800 systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 5800 or cobas® 6800/8800 system software which assigns test results for all tests as either target not detected, CMV DNA detected < LLoQ (lower limit of quantitation), CMV DNA detected > ULoQ (upper limit of quantitation), or a value in the linear range LLoQ < x < ULoQ. Results can be reviewed directly on the system screen, exported, or printed as a report.
Nucleic acid from patient samples and added lambda DNA-QS molecules is simultaneously extracted. In summary, viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors are removed with subsequent wash reagent steps and purified nucleic acid is eluted from the glass particles with elution buffer at elevated temperature.
Selective amplification of target nucleic acid from the sample is achieved by the use of target virus-specific forward and reverse primers which are selected from highly-conserved regions of the CMV DNA polymerase (UL54) gene. Selective amplification of DNA-QS is achieved by the use of sequence-specific forward and reverse primers which are selected to have no homology with the CMV genome. A thermostable DNA polymerase enzyme is used for amplification. The target and DNA-QS sequences are amplified simultaneously utilizing a universal PCR amplification profile with predefined temperature steps and number of cycles. The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythimidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicon from previous PCR runs is eliminated by the AmpErase enzyme, which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicon are not eliminated since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.
The cobas® CMV master mix contains one detection probe specific for CMV target sequences and one for the DNA-QS. The probes are labeled with target-specific fluorescent reporter dyes allowing simultaneous detection of CMV target and DNA-QS in two different target channels. The fluorescent signal of the intact probes is suppressed by the quencher dye. During the PCR amplification step, hybridization of the probe to the specific single-stranded DNA templates results in cleavage by the 5'-to-3' nuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probes are generated and the cumulative signal of the reporter dye is concomitantly increased. Real-time detection and discrimination of PCR products is accomplished by measuring the fluorescence of the released reporter dyes for the viral targets and DNA-QS.
N/A
Ask a specific question about this device
(88 days)
cobas® HCV is an in vitro nucleic acid amplification test for both the detection and quantitation of hepatitis C virus (HCV) RNA, in human EDTA plasma or serum, of HCV antibody positive or HCV-infected individuals. Specimens containing HCV genotypes 1 to 6 are validated for detection and quantitation in the assay.
cobas® HCV is intended for use as an aid in the diagnosis of HCV infection in the following populations: individuals with antibody evidence of HCV with evidence of liver disease, individuals suspected to be actively infected with HCV antibody evidence, and individuals at risk for HCV infection with antibodies to HCV. Detection of HCV RNA indicates that the virus is replicating and therefore is evidence of active infection.
cobas® HCV is intended for use as an aid in the management of HCV-infected patients undergoing anti-viral therapy. The assay can be used to measure HCV RNA levels at baseline, during treatment, at the end of treatment, and at the end of follow up of treatment to determine sustained or non-sustained viral response. The results must be interpreted within the context of all relevant clinical and laboratory findings.
cobas® HCV has not been approved for use as a screening test for the presence of HCV in blood or blood products.
Assay performance characteristics have been established for individuals treated with certain direct-acting antiviral agents (DAA) regimens. No information is available on the assay's predictive value when other DAA combination therapies are used.
cobas® HCV is a quantitative test performed on the cobas® 5800 system, cobas® 6800 system and cobas® 8800 system. cobas® HCV enables the detection and quantitation of HCV RNA in EDTA plasma or serum of infected patients. Dual probes are used to detect and quantify, but not discriminate genotypes 1–6. The viral load is quantified against a non-HCV armored RNA quantitation standard (RNA-QS), which is introduced into each specimen during sample preparation. The RNA-QS also functions as an internal control to assess substantial failures during the sample preparation and PCR amplification processes. In addition, the test utilizes three external controls: a high titer positive, a low titer positive, and a negative control.
N/A
Ask a specific question about this device
(28 days)
The CoaguChek XS Plus System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.
The CoaguChek XS Plus system is a portable coagulation monitoring system to monitor prothrombin time (PT) in patients receiving oral anticoagulant therapy. The system uses the amperometric detection of thrombin in the blood sample. A test strip is used to determine a PT value from 8 µL of whole blood. Onboard quality control is available on every test strip and the system also features an optional external quality control material (CoaguChek XS PT Control).
N/A
Ask a specific question about this device
(90 days)
Elecsys Phospho-Tau (181P) Plasma is an in vitro electrochemiluminescence immunoassay (ECLIA) intended for the measurement of the phosphorylated Tau 181 protein in human plasma on cobas e immunoassay analyzers.
The Elecsys Phospho-Tau (181P) Plasma assay result is intended to be used as an aid in the initial assessment for Alzheimer's disease and other causes of cognitive decline in adult patients aged 55 years and older, presenting with signs, symptoms, or complaints of cognitive decline. The result should be interpreted in conjunction with other clinical information.
A negative test result is consistent with a negative amyloid positron emission tomography (PET) scan result and reduced likelihood that a patient's cognitive impairment is due to amyloid pathology. These patients should be investigated for other causes of cognitive decline.
A positive test result may not be consistent with a positive amyloid PET scan result. Patients with an initial positive result should be further investigated to determine whether the amyloid pathology can be a cause of cognitive impairment.
In vitro electrochemiluminescence immunoassay (ECLIA) intended for the measurement of the phosphorylated Tau 181 protein (pTau181p) in human plasma.
Elecsys Phospho-Tau (181P) Plasma utilizes a sandwich test principle and has a total duration time of 18 minutes.
- 1st incubation: 30 µL of sample, biotinylated monoclonal antibody specific for phosphorylation at threonine 181, and a monoclonal tau-specific antibody labeled with a ruthenium complex^a) react to form a sandwich complex.
- 2nd incubation: After addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell II M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the cobas link.
N/A
Ask a specific question about this device
(266 days)
The cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems is an automated, multiplex, nucleic acid test that utilizes real-time polymerase chain reaction (PCR) technology intended for simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) in nasopharyngeal swab specimens obtained from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. Conversely, positive results do not rule out coinfection with other organisms, and the agent(s) detected by the cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems may not be the definite cause of disease.
cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems (cobas® Respiratory 4-flex) is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 5800 System is designed as one integrated instrument. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 5800 System or cobas® 6800/8800 Systems software(s), which assigns results for all tests. Results can be reviewed directly on the system screen and printed as a report.
Nucleic acid from patient samples and added Internal Control RNA (RNA IC) molecules are simultaneously extracted. Nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors, are removed with subsequent wash steps and purified nucleic acid is eluted from the magnetic glass particles with elution buffer at elevated temperature. External controls (positive and negative) are processed in the same way.
Selective amplification of target nucleic acid from the sample is achieved by the use of target-specific forward and reverse primers detecting conserved viral genome regions as shown in Table 1.
Selective amplification of RNA IC is achieved by the use of non-competitive, sequence specific forward and reverse primers, which have no homology with the viral-target specific genomes. Amplified target is detected by the cleavage of fluorescently labeled oligonucleotide probes. Roche's temperature assisted generation of signal (TAGS) technology, short TAGS technology, is introduced to differentiate up to three targets per fluorescence channel, enabling the detection of up to14 targets, and the Internal Control, per well. A thermostable DNA polymerase enzyme is used for amplification.
Multiplicity of target detection is enabled with temperature-dependent quenching of cleaved fluorescent target-specific probes. This is achieved by separating signals from probes into introduced thermal channels, where fluorescence is acquired at two additional fixed temperatures for each amplification cycle.
During the PCR amplification step, hybridization of the probes to the specific single-stranded DNA template results in cleavage of the probe by the 5' to 3' exonuclease activity of the DNA polymerase, resulting in separation of the reporter and quencher dyes, and the generation of a fluorescent signal. Conventional probes release fluorescence signal immediately upon separation of reporter from quencher. TAGS probes rely on temperature dependent fluorescence activation, requiring both nuclease cleavage during the extension phase, as well as an increase in reaction temperature, to activate the otherwise dormant fluorophore. For this reason, during each PCR cycle the TAGS technology captures fluorescence in five available fluorescence channels in combination with three thermal channels (detection of fluorescence at three defined temperatures T1, T2 and T3).
The cobas® Respiratory 4-flex master mix contains detection probes which are specific for influenza A virus, influenza B virus, RSV, SARS-CoV-2 and the RNA Internal Control (RNA IC) nucleic acid, which enables simultaneous detection and differentiation of influenza A virus, influenza B virus, RSV, and SARS-CoV-2 viral targets and the RNA IC.
The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythymidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicons from previous PCR runs are destroyed by the AmpErase enzyme [uracil-N-glycosylase], which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not destroyed since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.
The RESP-4FLEX ASAP enables the system to differentiate and report the qualitative results of the four targets influenza A virus, influenza B virus, RSV and SARS-CoV-2. For each specimen the customer can test for any combination of the four enabled virus targets. Also, additional target calculation (digital reflex) can be ordered for the four enabled virus targets (influenza A virus, influenza B virus, RSV and SARS-CoV-2) on the cobas® 5800 System.
Here's an analysis of the acceptance criteria and study detailed in the provided FDA clearance letter for the cobas Respiratory 4-flex, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA clearance letter does not explicitly state pre-defined acceptance criteria for the clinical performance. Instead, it reports the observed performance metrics (PPA and NPA) from the clinical studies. For the purpose of this table, I will present the reported clinical performance as the "met acceptance criteria," assuming these values were deemed acceptable by the FDA for clearance.
| Analyte | Acceptance Criteria (Reported PPA CI) | Device Performance (PPA %) | Acceptance Criteria (Reported NPA CI) | Device Performance (NPA %) |
|---|---|---|---|---|
| Influenza A | ||||
| Fresh Pros. | (90.5%, 98.5%) | 96.2 | (99.3%, 99.9%) | 99.7 |
| Frozen Pros. | (89.0%, 99.1%) | 96.8 | (99.4%, 99.8%) | 99.7 |
| Combined Pros. | (92.3%, 98.3%) | 96.4 | (99.5%, 99.8%) | 99.7 |
| Retrospective | (94.1%, 100.0%) | 100.0 | (97.6%, 99.4%) | 98.8 |
| Influenza B | ||||
| Fresh Pros. | (94.5%, 100.0%) | 100.0 | (99.6%, 100.0%) | 99.9 |
| Frozen Pros. | (79.8%, 99.3%) | 95.8 | (99.8%, 100.0%) | 100.0 |
| Combined Pros. | (94.0%, 99.8%) | 98.9 | (99.8%, 100.0%) | 99.9 |
| Retrospective | (87.1%, 99.6%) | 97.5 | (98.3%, 99.7%) | 99.3 |
| RSV | ||||
| Fresh Pros. | (77.4%, 94.7%) | 88.7 | (99.8%, 100.0%) | 100.0 |
| Frozen Pros. | (81.5%, 95.3%) | 90.4 | (99.8%, 100.0%) | 100.0 |
| Combined Pros. | (83.1%, 93.9%) | 89.7 | (99.9%, 100.0%) | 100.0 |
| Retrospective | (96.4%, 100.0%) | 100.0 | (98.2%, 99.7%) | 99.3 |
| SARS-CoV-2 | ||||
| Fresh Pros. | (92.4%, 98.6%) | 96.7 | (98.0%, 99.1%) | 98.6 |
| Frozen Pros. | (95.3%, 98.9%) | 97.7 | (97.8%, 98.8%) | 98.4 |
| Combined Pros. | (95.4%, 98.5%) | 97.3 | (98.0%, 98.8%) | 98.5 |
Study Proving Device Meets Criteria: The detailed clinical performance evaluation described in "5. CLINICAL PERFORMANCE EVALUATION" (pages 23-26) demonstrates that the cobas Respiratory 4-flex achieves the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) values against a U.S. FDA-cleared molecular comparator assay. These reported percentage agreements, along with their 95% confidence intervals, serve as the evidence that the device meets the performance expectations for clinical use.
2. Sample Size Used for the Test Set and Data Provenance
-
Prospective Clinical Study:
- Total NPS specimens enrolled: 4,475
- Total NPS specimens tested: 4,378 (1,832 fresh, 2,546 frozen)
- Total NPS specimens evaluable: 4,341 (1,827 fresh, 2,514 frozen)
- Data Provenance: Fresh prospective specimens collected at eleven collection sites during the 2023-2024 respiratory viral season. Frozen prospective specimens collected during parts of the 2022-2023 respiratory viral season at seven sites and the 2023-2024 respiratory viral season from 14 collection sites. The specific country of origin is not explicitly stated but implies U.S. based on the "U.S. testing sites" mention for retrospective samples, and the FDA clearance context. The data is prospective.
-
Retrospective Clinical Study:
- Total NPS specimens enrolled: 770
- Total NPS specimens evaluable:
- Influenza A: 657
- Influenza B: 647
- RSV: 659
- Data Provenance: Archived NPS specimens collected between 2014 and 2022 from individuals with signs and symptoms of respiratory viral infection. Tested at three (3) U.S. testing sites. The data is retrospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of human experts to establish ground truth for the test set. The ground truth was established by a "U.S. FDA-cleared molecular assay" as the comparator method.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method involving experts for discrepant results. Instead, it states that an "FDA 510(k) cleared comparator" method was used to establish the "ground truth." For the influenza A target in the prospective study, "three (3) additional specimens (two (2) fresh and one (1) frozen) were excluded from analysis due to inconclusive results obtained from the comparator test," implying a reliance on the comparator's definitive result rather than external adjudication. The same pattern is noted for retrospective samples, with exclusions for failed or invalid comparator test results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance
This section is Not Applicable to the provided document. The cobas Respiratory 4-flex is an automated, multiplex nucleic acid detection test, not an AI-powered diagnostic tool requiring human reader interpretation or assistance. Therefore, an MRMC study or analysis of human reader improvement with AI is not relevant.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, this was a standalone performance evaluation. The device is an automated in vitro diagnostic (IVD) test. The clinical performance evaluation directly compares the cobas Respiratory 4-flex's results (algorithm only, as it's an automated system) against a predicate FDA-cleared molecular assay. There is no human interpretation or intervention in the generation of the primary test result from the cobas system itself.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The ground truth for the clinical performance evaluation was established by a "U.S. FDA-cleared molecular assay." This means another legally marketed and validated molecular diagnostic test was used as the reference standard.
8. The Sample Size for the Training Set
The document does not specify a separate training set or its sample size. This is typical for in vitro diagnostic (IVD) devices that use established laboratory techniques (like PCR) rather than machine learning algorithms which require explicit training data. The development of such assays involves extensive analytical validation (LoD, precision, specificity, inclusivity, etc.) and then clinical validation with independent samples.
9. How the Ground Truth for the Training Set Was Established
Since no explicit training set for a machine learning algorithm is mentioned, the concept of "ground truth for the training set" is not applicable in the context of this device's validation as described. The analytical studies (LoD, inclusivity, specificity, etc.) characterize the inherent performance of the assay's chemical and hardware components.
Ask a specific question about this device
Page 1 of 60