K Number
K243753

Validate with FDA (Live)

Date Cleared
2025-11-20

(350 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.

The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.

Device Description

The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated multiplex real-time polymerase chain reaction (PCR) assay for the rapid in vitro qualitative detection and differentiation of B. pertussis (Bp), B. parapertussis (Bpp), and B. holmesii (Bh) DNA in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.

The different fluorescent dye designs enable the specific detection and differentiation of the three microorganisms (Bp, Bpp, and Bh) independently in a multiplex system. The system automates all nucleic acid amplification test sample processing steps, including inhibitor removal, nucleic acid extraction, purification, amplification, real-time detection, and result interpretation in a rapid manner. The test is designed for use in near-patient settings to deliver results in approximately 15 minutes.

The cobas® liat system is comprised of the cobas® liat analyzer (analyzer) hardware with integrated cobas® liat system software (analyzer software + liat assay specific package (script)) for running tests and analyzing the results, and a single-use disposable cobas® liat assay tube (assay tube).

  • cobas® liat analyzer is a system component that consists of one software subsystem and three hardware units:

    • Infrastructure unit, which consists of the hardware and embedded software (firmware).
    • Thermal, loading and motion unit: the processing module that interacts physically with the assay tube during the assay execution.
    • Detection unit consisting of the photodetectors that is used for the fluorescence detection during the PCR reaction
  • The assay script provides a set of instructions to the analyzer hardware and software for assay tube processing, PCR, result calculation and interpretation and result reporting. The assay script can be installed on the analyzer independently of the analyzer software.

The cobas® liat Bordetella panel nucleic acid test is supported with a Liat Assay Specific Package (Assay Script): cobas® liat Bordetella panel LASP (BPTA).

  • The assay tube holds all reagents needed for sample preparation and PCR processes. The assay specific reagents are packaged into a single assay tube in separate segments that are separated by frangible seals. An internal control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through all steps of the assay process and to monitor the presence of inhibitors in the PCR reaction.
AI/ML Overview

N/A

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

November 20, 2025

Roche Molecular Systems, Inc.
Khushvanreep Singh
Regulatory Affairs Specialist
4300 Hacienda Drive
Pleasanton, California 94588

Re: K243753
Trade/Device Name: cobas liat Bordetella panel nucleic acid test
Regulation Number: 21 CFR 866.3980
Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay
Regulatory Class: Class II
Product Code: OZZ, OOI
Dated: December 04, 2024
Received: December 05, 2024

Dear Khushvanreep Singh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243753 - Khushvanreep Singh Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243753 - Khushvanreep Singh Page 3

Sincerely,

Bryan M. Grabias -S
2025.11.20 10:25:01 -05'00'

Bryan Grabias
Acting Branch Chief
Bacterial Respiratory and Medical Countermeasures Branch
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K243753

Device Name: cobas liat Bordetella panel nucleic acid test

Indications for Use (Describe)

The cobas liat Bordetella panel nucleic acid test (cobas liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.

The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

cobas® liat Bordetella panel nucleic acid test (K243753)

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameRoche Molecular Systems, Inc.
Address4300 Hacienda Drive, Pleasanton, CA 94588-2722
ContactKhushvanreep SinghPhone: (908) 253-7864Email: Khushvanreep.singh@roche.com
Date PreparedNovember 19, 2025
Proprietary Namecobas® liat Bordetella panel nucleic acid test
Common Namecobas® liat Bordetella panel
Classification21 CFR 866.3980Respiratory viral panel multiplex nucleic acid assay
Product CodesOZZOOI
Predicate DevicesVerigene® Respiratory Pathogens Flex Nucleic Acid Test (Verigene® RP Flex) (K143653)
Establishment RegistrationRoche Molecular Systems, Inc. (2243471)

Page 6

1. DEVICE DESCRIPTION

The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated multiplex real-time polymerase chain reaction (PCR) assay for the rapid in vitro qualitative detection and differentiation of B. pertussis (Bp), B. parapertussis (Bpp), and B. holmesii (Bh) DNA in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.

The different fluorescent dye designs enable the specific detection and differentiation of the three microorganisms (Bp, Bpp, and Bh) independently in a multiplex system. The system automates all nucleic acid amplification test sample processing steps, including inhibitor removal, nucleic acid extraction, purification, amplification, real-time detection, and result interpretation in a rapid manner. The test is designed for use in near-patient settings to deliver results in approximately 15 minutes.

The cobas® liat system is comprised of the cobas® liat analyzer (analyzer) hardware with integrated cobas® liat system software (analyzer software + liat assay specific package (script)) for running tests and analyzing the results, and a single-use disposable cobas® liat assay tube (assay tube).

  • cobas® liat analyzer is a system component that consists of one software subsystem and three hardware units:

    • Infrastructure unit, which consists of the hardware and embedded software (firmware).
    • Thermal, loading and motion unit: the processing module that interacts physically with the assay tube during the assay execution.
    • Detection unit consisting of the photodetectors that is used for the fluorescence detection during the PCR reaction
  • The assay script provides a set of instructions to the analyzer hardware and software for assay tube processing, PCR, result calculation and interpretation and result reporting. The assay script can be installed on the analyzer independently of the analyzer software.

Page 7

The cobas® liat Bordetella panel nucleic acid test is supported with a Liat Assay Specific Package (Assay Script): cobas® liat Bordetella panel LASP (BPTA).

  • The assay tube holds all reagents needed for sample preparation and PCR processes. The assay specific reagents are packaged into a single assay tube in separate segments that are separated by frangible seals. An internal control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through all steps of the assay process and to monitor the presence of inhibitors in the PCR reaction.

2. INDICATIONS FOR USE

The cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.

The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.

Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.

3. TECHNOLOGICAL CHARACTERISTICS

The primary technological characteristics and intended use of the RMS cobas® liat Bordetella panel are substantially equivalent to other legally marketed nucleic acid amplification tests intended for the qualitative detection of pertussis respiratory infection.

Page 8

As indicated in Table 1, the RMS cobas® liat Bordetella panel is substantially equivalent to significant characteristics of the identified predicate device, the currently cleared Verigene® RP Flex) (K143653).

Table 1: Comparison of the cobas® liat Bordetella panel and the Predicate Device

Submitted Device: cobas® liat Bordetella panel nucleic acid testPredicate Device: Verigene® RP Flex (K143653)
Regulation Name866.3980866.3980
Product CodeOZZOOIOZZOOIOCCOEMOEPOOUOZE
Intended UseThe cobas® liat Bordetella panel nucleic acid test (cobas® liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.The Verigene® Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection. The test is performed on the automated Verigene System utilizing reverse transcription (RT), polymerase chain reaction (PCR), and microarray hybridization to detect gene sequences of the following organism types and subtypes:Viruses:AdenovirusHuman MetapneumovirusInfluenza AInfluenza A (Subtype H1)Influenza A (Subtype H3)Influenza BParainfluenza 1Parainfluenza 2Parainfluenza 3Parainfluenza 4Respiratory Syncytial Virus ARespiratory Syncytial Virus BRhinovirusBacteria:Bordetella parapertussis/bronchiseptica,Bordetella holmesii,Bordetella pertussis
Sample TypeNasopharyngeal swab (NPS)Same
Analyte TargetsBordetella pertussis (BP)Bordetella parapertussis (BPP)Bordetella holmesii (BH)Bordetella parapertussis/bronchiseptica,Bordetella holmesii,Bordetella pertussisAdenovirusHuman Metapneumovirus

Page 9

Submitted Device: cobas® liat Bordetella panel nucleic acid testPredicate Device: Verigene® RP Flex (K143653)
Influenza AInfluenza A (Subtype H1)Influenza A (Subtype H3)Influenza BParainfluenza 1Parainfluenza 2Parainfluenza 3Parainfluenza 4Respiratory Syncytial Virus ARespiratory Syncytial Virus BRhinovirus
Ancillary Collection KitsFlocked swab with Universal Transport Media (UTM)Same
Time to resultAbout 15 minutes~ 2 hours
Amplification TechnologyReal-time PCRSame
Detection ChemistryTaqMan® probes with fluorescent dyesMicroarray hybridization, Gold nanoparticle probe with Light Scattering
Controls UsedSample processing control (IC)Positive and negative controlSame
Instrumentationcobas® liat SystemVerigene System
User ComplexityCLIA-WaivedCLIA-waived

4. NON-CLINICAL PERFORMANCE EVALUATION

4.1. Analytical sensitivity (Limit of Detection)

Analytical sensitivity was determined by analyzing a dilution series of two representative strains of Bordetella pertussis (Bp, Strains A639 and E431), Bordetella parapertussis (Bpp, Strains E838 and A747), and Bordetella holmesii (Bh, Strains F061 and ATCC 51541). The Bp, Bpp, and Bh cultures were co-spiked and diluted in pooled negative nasopharyngeal swab (NPS) clinical specimens collected in UTM to seven (7) concentration levels in 2-fold dilutions. All levels were tested with at least twenty (20) replicates for each of three (3) unique lots of reagents. The limit of detection (LoD) for each specimen type is shown in colony forming units per mL (CFU/mL) for Bp (Table 2), Bpp (Table 3), and Bh (Table 4) respectively as the target concentration which can be detected in ≥ 95% of the replicates for the worst-performing of the reagent lots.

Page 10

Table 2: Bp concentration levels with at least 95% observed hit rate for worst-performing lot tested

StrainLoD (CFU/mL)Hit rate (Mean Ct)
A6394395.2% (34.6)
E4313695.2% (34.3)

Table 3: Bpp concentration levels with at least 95% observed hit rate for worst-performing lot tested

StrainLoD (CFU/mL)Hit rate (Mean Ct)
E8383295.2% (33.5)
A7473695.2% (32.6)

Table 4: Bh concentration levels with at least 95% observed hit rate for worst-performing lot tested

StrainLoD (CFU/mL)Hit rate (Mean Ct)
F06134100% (29.1)
515412195.2% (31.1)

A study was performed to demonstrate that the LoD with target co-spiked samples is comparable to single target spiked samples.

4.2. Reactivity/Inclusivity

Inclusivity testing was performed for 12 Bp, 6 Bpp, and 3 Bh strains using one lot of reagents. Testing was performed using Bp, Bpp, and Bh cultures that were diluted into pools of negative clinical specimens. Three replicates per dilution level were tested for each strain. The strains tested and the lowest concentrations detected are listed in Table 5, Table 6, and Table 7 for Bp, Bpp, and Bh respectively.

Table 5: Inclusivity testing for Bp strains

StrainCFU/mL
ATCC 84670.12
ATCC BAA-5890.006

Page 11

StrainCFU/mL
ATCC 514450.12
ATCC 538940.05
ATCC 93400.04
ATCC 103800.48
ATCC BAA-13350.3
ATCC 93060.3
ATCC 127430.12
ATCC 97970.12
ATCC 127425.0
ATCC 847840.0

Table 6: Inclusivity testing for Bpp strains

StrainCFU/mL
ATCC 153112.5
E59540.0
ATCC 1523710.0
ATCC BAA-58715.0
ATCC 930511.9
C51047.6

Table 7: Inclusivity testing for Bh strains

StrainCFU/mL
ATCC 7000531.75
ATCC 7000520.48
C69012.5

4.3. Cross reactivity and microbial interference

A panel of bacteria, fungi and viruses, including those commonly found in patient specimens, were tested to assess analytical specificity. High titer stocks of the potentially cross-reacting microorganisms listed in Table 8 were tested in negative nasopharyngeal swab matrix for cross-reactivity with cobas® liat Bordetella panel, and also tested for microbial interference in the presence of Bp, Bpp, and Bh cultures at ~3x LoD concentrations. Three (3) replicates in target positive background and three (3) replicates in target negative background were tested for each

Page 12

non-target organism. The testing concentrations for potentially interfering organisms were ≥ 1 x 10⁶ units/mL for bacteria or fungi and ≥ 1 x 10⁵ units/mL for viruses. Results indicated that none of the non-target organisms tested generated false positive or false negative results due to cross-reactivity or interference.

Table 8: Microorganisms tested for analytical specificity/cross-reactivity

Bacteria / FungiVirus
Bordetella bronchiseptica¹Legionella pneumophila
Bordetella aviumMoraxella catarrhalis
Bordetella hinziiMycobacterium tuberculosis
Bordetella petriiMycoplasmoides pneumoniae
Bordetella trematumNeisseria gonorrhoeae
Achromobacter denitrificans¹Neisseria meningitidis
Chlamydia pneumoniaePseudomonas aeruginosa
Chlamydia trachomatisMethicillin-resistant Staphylococcus aureus
Corynebacterium diphtheriaeStaphylococcus epidermidis
Escherichia coliStenotrophomonas maltophilia
Haemophilus influenzaeStreptococcus pneumoniae
Klebsiella pneumoniaeStreptococcus pyogenes
Lactobacillus acidophilusStreptococcus salivarius
Lactiplantibacillus plantarumCandida albicans (fungus)
----
----

¹Three different strains were tested
²Tested at highest concentration possible per stock concentration

4.4. Competitive inhibition

To assess competitive inhibition between Bp, Bpp, and Bh, a total of three different combinations of low concentration of target (~3x LoD) were mixed with high concentrations (1E7 CFU/mL) of the other two targets in pooled clinical specimen matrix. Each combination was tested in replicates of 5 using one lot of reagents. Testing results indicated that when two target microorganisms were present at high concentrations, no interference was observed for microorganisms that were present at low concentrations.

Page 13

4.5. Endogenous and exogenous interference

Potentially interfering substances that may be commonly encountered in nasopharyngeal swab clinical specimens were evaluated. Testing was executed using pooled clinical specimens spiked with potential interferents. Potential interferents were tested with and without Bp, Bpp, and Bh at ~3x LoD using one lot of reagents. Five replicates each of Bp/Bpp/Bh negative sample and Bp/Bpp/Bh positive sample were tested with each substance.

The substances listed in Table 9 at the concentrations tested did not interfere in the detection of Bp, Bpp, or Bh. Flonase fluticasone nasal spray resulted in false negative results in one replicate when tested at concentrations higher than 5% v/v, and may interfere with the assay performance at concentrations greater than 5% v/v.

Table 9: List of products tested at concentrations that do not show interference

Potential InterferentConcentration Tested
Human Cells (PBMC)1E6 cells/mL
Mucin Bovine Type 1-S10 mg/mL
Human Whole Blood10% (v/v)
Afrin Pump Mist10.0 % (v/v)
Flonase5.0 % (v/v)
Chloraseptic Sore Throat Lozenges10.0 mg/mL
Mupirocin10.0 mg/mL
Relenza10.0 mg/mL
Tamiflu7.5 mg/mL
Tobramycin0.6 mg/mL*
Albuterol Sulfate5.0 % (w/v)
Sucrets Sore Throat & Cough Lozenges– Vapor Cherry5.0 % (w/v)
Ampicillin powder10.0 mg/mL
Azithromycin tablet0.25 mg/mL*
Beconase AQ Nasal Spray8.5 mg/mL
Children's Dimetapp Cold and Cough5.0 % (v/v)
Ciprofloxacin1.3 mg/mL
Erythromycin0.47 mg/mL*
Neo-Synephrine Nasal Spray15.0 % (v/v)
Rifampin3.75 mg/mL*
Robitussin Cough + Chest Congestion DM liquid gel capsule10.0 % (v/v)
Ocean Saline Nasal Spray (sodium chloride)10.0 % (v/v)
Flumist6.3 % (v/v)

Page 14

*Note: For these substances, the concentration reflects the maximum amount soluble under the conditions tested.

4.6. Reproducibility study

A reproducibility study assessed the total variability of the cobas® liat Bordetella panel across study sites, operators, analyzers, testing days, and cobas® liat Bordetella panel reagent lots. The cobas® liat Bordetella panel was evaluated at three (3) sites. Two (2) operators at each site tested a three (3) member reproducibility panel in triplicate on five (5) non-consecutive days across three (3) reagent lots, for a total of 810 runs (3 panel members ⋅ 3 replicates ⋅ 2 operators ⋅ 5 days ⋅ 3 sites ⋅ 3 reagent lots). A total of nine (9) cobas® liat analyzers, three (3) at each study site, were used. The reproducibility panel, prepared by spiking Bordetella bacteria into a UTM-based human clinical matrix, comprised a negative, a low positive (1x-2x LOD), and a moderate positive (3x-5x LOD) for each of B. pertussis, B. parapertussis and B. holmesii bacteria. For a given bacteria, the expected result for the negative panel member is "Not Detected," while the expected result for the low positive and moderate positive panel member is "Detected".

The percent agreement with expected results is summarized in Table 10.

Table 10: cobas® liat Bordetella panel reproducibility: percentage agreement by target analyte and expected concentration

Target AnalyteExpected ConcentrationValid Tests (N)Results in Agreement With Target Analyte (n)Percent Agreement n/N x 10095% Score CI
Negative0264264100.0(98.6, 100.0)
Bp1x-2x LOD25825799.6(97.8, 99.9)
Bp3x-5x LOD265265100.0(98.6, 100.0)
Bpp1x-2x LOD25825799.6(97.8, 99.9)
Bpp3x-5x LOD265265100.0(98.6, 100.0)
Bh1x-2x LOD25825799.6(97.8, 99.9)
Bh3x-5x LOD265265100.0(98.6, 100.0)

Note: Results were in agreement when a positive panel member had a valid result of "Detected" for the target analyte or when the negative panel member had a valid result of "Not Detected" for the target analyte.

The variance component analyses for Ct values by target analyte and site are shown in the tables below respectively.

Page 15

Table 11: cobas® liat Bordetella panel reproducibility: variance component analysis for Ct values by target analyte

Target AnalyteExpected Concentrationn/N^aMean CtSite SDSite CV%Lot SDLot CV%Day SDDay CV%Run SDRun CV%Within-Run (Residual) SDWithin-Run (Residual) CV%Total SDTotal CV%
Bp1x-2x LOD257/25833.20.050.10.351.10.230.70.150.50.591.80.742.2
Bp3x-5x LOD265/26532.00.200.60.491.50.000.00.220.70.451.40.732.3
Bpp1x-2x LOD257/25831.90.170.50.311.00.150.50.000.00.571.80.682.1
Bpp3x-5x LOD265/26530.50.210.70.291.00.140.50.050.20.511.70.642.1
Bh1x-2x LOD257/25827.80.000.00.271.00.140.50.110.40.562.00.652.3
Bh3x-5x LOD265/26526.80.140.50.291.10.040.10.140.50.411.50.542.0

Ct: cycle threshold; LOD: Limit of Detection; SD: standard deviation; CV%: percent coefficient of variation.
^a n is the number of positive tests, which contribute Ct values to the analysis. N is the total number of valid tests for the panel member.

Table 12: cobas® liat Bordetella panel reproducibility: variance component analysis for Ct values by site

SiteTarget AnalyteExpected Concentrationn/N^aMean CtLot SDLot CV%Day SDDay CV%Run SDRun CV%Within-Run (Residual) SDWithin-Run (Residual) CV%Total SDTotal CV%
Site ABp1x-2x LOD85/8533.10.551.70.190.60.190.60.581.80.852.6
Site ABp3x-5x LOD88/8831.80.662.10.000.00.210.70.401.30.802.5
Site BBp1x-2x LOD85/8533.20.341.00.000.00.060.20.652.00.732.2
Site BBp3x-5x LOD89/8931.80.280.90.000.00.140.50.461.50.561.8
Site CBp1x-2x LOD87/8833.30.371.10.060.20.110.30.531.60.652.0
Site CBp3x-5x LOD88/8832.20.531.60.000.00.190.60.461.40.732.3
Site ABpp1x-2x LOD85/8531.90.270.90.000.00.000.00.662.10.722.3
Site ABpp3x-5x LOD88/8830.40.280.90.230.80.000.00.451.50.581.9
Site BBpp1x-2x LOD85/8531.80.220.70.260.80.000.00.521.60.622.0
Site BBpp3x-5x LOD89/8930.30.160.50.000.00.080.30.531.70.561.8
Site CBpp1x-2x LOD87/8832.10.421.30.120.40.000.00.491.50.652.0
Site CBpp3x-5x LOD88/8830.70.411.30.100.30.120.40.531.70.692.2
Site ABh1x-2x LOD85/8527.80.250.90.170.60.000.00.521.90.602.2

Page 16

SiteTarget AnalyteExpected Concentrationn/N^aMean CtLot SDLot CV%Day SDDay CV%Run SDRun CV%Within-Run (Residual) SDWithin-Run (Residual) CV%Total SDTotal CV%
Site ABh3x-5x LOD88/8926.70.250.90.150.50.000.00.411.50.501.9
Site BBh1x-2x LOD85/8527.80.200.70.230.80.000.00.612.20.682.4
Site BBh3x-5x LOD89/8926.70.210.80.000.00.200.80.401.50.491.8
Site CBh1x-2x LOD87/8827.80.291.00.030.10.271.00.531.90.662.4
Site CBh3x-5x LOD88/8826.90.381.40.090.30.080.30.411.50.572.1

Ct: cycle threshold; LOD: Limit of Detection; SD: standard deviation; CV%: percent coefficient of variation.
Note: Site A = MCR; Site B = APR; Site C = URM.
^a n is the number of positive tests, which contribute Ct values to the analysis. N is the total number of valid tests for the panel member.

5. CLINICAL PERFORMANCE EVALUATION

The clinical performance of cobas® liat Bordetella panel (BP, BPP, and BH) was evaluated in a multi-site prospective study. Over 800 patients with suspected pertussis respiratory infection were enrolled between July 2023 – February 2024, and nasopharyngeal specimens were prospectively collected from patients presenting to point-of-care (POC) settings (e.g. emergency rooms, outpatient clinics, etc.). Subject demographics for evaluable, prospective clinical subjects are provided in Table 13. Archived clinical specimens and contrived samples were supplemented to fulfil the sample size requirement. The archived positive specimens were previously collected from symptomatic patients and characterized using an FDA-cleared Nucleic Acid Amplification Test ( NAAT). Specimens were tested using cobas® liat Bordetella panel, the reference comparator methods per respective IFU and, if required, with a validated sequencing method to confirm the comparator result.

Table 13: Subject Demographics – Evaluable Prospective Population

CharacteristicsStatisticEstimates
TotalN769
Age (Years)
Mean ± SD33.9 ± 22.33
Median35.0
Min , Max0.1 , 85.0
Age Category
< 1n (%)27 ( 3.51%)
1 to <5n (%)89 ( 11.57%)

Page 17

CharacteristicsStatisticEstimates
5 to <12n (%)69 ( 8.97%)
12 to <18n (%)32 ( 4.16%)
18 to <40n (%)219 ( 28.48%)
40 to <65n (%)270 ( 35.11%)
>=65n (%)63 ( 8.19%)
Sex at Birth
Malen (%)338 ( 43.95%)
Femalen (%)431 ( 56.05%)
Ethnicity
Hispanic / Latinon (%)194 ( 25.23%)
Not Hispanic / Latinon (%)568 ( 73.86%)
Not Reportedn (%)2 ( 0.26%)
Unknownn (%)5 ( 0.65%)
Race
American Indian / Alaska Nativen (%)22 ( 2.86%)
Asiann (%)27 ( 3.51%)
Black / African Americann (%)164 ( 21.33%)
Whiten (%)527 ( 68.53%)
Mixed-Race ^an (%)22 ( 2.86%)
Othern (%)2 ( 0.26%)
Not Reportedn (%)5 ( 0.65%)

Note: Unknown category indicates subjects for whom the corresponding information is not available.
^a Mixed-Race combinations include: American Indian or Alaska Native/White (n=3), Asian/Black or African American (n=1), Asian/White (n=1), Asian/White/Other (n=1), Black or African American/White (n=15), Native Hawaiian or Other Pacific Islander/White (n=1).

The cobas® liat Bordetella panel's clinical performance was assessed by comparing results to FDA-cleared target specific NAAT. For Bpp and Bh, NAAT was used as the reference comparator. For Bp, the reference method was a composite of NAAT and a validated bi-directional sequencing method to confirm the presence of Bp.

The cobas® liat Bordetella Panel clinical performance results, PPA and NPA versus the comparator reference method, are summarized in Table 14.

Table 14: Summary Clinical Performance cobas® liat Bordetella panel by Target and Specimen Type

TargetSpecimen typeTotal (N)PPAPPA LCL 95% Score CIPPA UCL 95% Score CINPANPA LCL 95% Score CINPA UCL 95% Score CI
BpProspective743NCNCNC100.0% (743/743)99.5%100.0%
BpArchived160100% (42/42)91.6%100.0%99.2% (117/118)95.4%99.9%

Page 18

TargetSpecimen typeTotal (N)PPAPPA LCL 95% Score CIPPA UCL 95% Score CINPANPA LCL 95% Score CINPA UCL 95% Score CI
BpContrived32798.8% (80/81)93.3%99.8%99.6% (245/246)97.7%99.9%
BpOverall123099.2% (122/123)95.5%99.9%99.8% (1105/1107)99.3%100.0%
BppProspective7430.0% (0/1)0.0%79.3%100.0% (742/742)99.5%100.0%
BppArchived170100.0% (28/28)87.9%100.0%100.0% (142/142)97.4%100.0%
BppContrived327100.0% (108/108)96.6%100.0%99.5% (218/219)97.5%99.9%
BppOverall124099.3% (136/137)96.0%99.9%99.9% (1102/1103)99.5%100.0%
BhProspective702NCNCNC99.9% (701/702)99.2%100.0%
BhContrived328100.0% (139/139)97.3%100.0%98.9% (187/189)96.2%99.7%
BhOverall1030100.0% (139/139)97.3%100.0%99.7% (888/891)99.0%99.9%

Abbreviations: Bp = B.pertussis, Bpp = B. parapertussis, Bh = B. holmesii, CI = confidence interval, PPA = Positive Percent Agreement, NPA = Negative Percent Agreement, LCL = Lower Confidence Limit, UCL = Upper Confidence Limit, NC = Not Calculable.

Note: For prospective and archived specimens, the reference method is the respective comparator results. For contrived specimens the reference method is the expected result.

6. CONCLUSIONS

A comparison of the intended use, technological characteristics, and the results of non-clinical analytical and clinical performance studies demonstrate that cobas® liat Bordetella panel nucleic acid test is substantially equivalent to the predicate device.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.