(219 days)
NobelActive® S TiUltra™, NobelActive® S TiUnite®
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function.
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications.
This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
NobelReplace® S TiUltra™, NobelReplace® S TiUnite®
NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are indicated for single or multiple unit restorations.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite®
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function.
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications.
This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved.
The Subject Device, Nobel Biocare S Series Implants, comprises of six Device Lines NobelActive S TiUltra™, NobelParallel™ S TiUltra™, NobelReplace S TiUltra™, NobelActive S TiUnite, NobelParallel™ S TiUnite, NobelReplace S TiUnite.
Nobel Biocare S Series Implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental prostheses to restore chewing function. The Subject Device features a consistent diameter at the implant-abutment interface, regardless of the implant size or type. This means that the prosthetic components, such as abutments and restorative elements, have uniform dimensions and connections across all implant sizes within the system. The connection used is the Narrow Platform (NP) conical connection.
The implant bodies of the Subject Devices and Predicate devices are the same. This allows use of the same drills and drilling protocols, rescue procedures and rescue tools. By using the existing Narrow Platform connection, the same prosthetic workflows, prosthetic components, and lab components can be used.
The implants are available in different sizes as listed in Table 1.
The Nobel Biocare S Series Implants are manufactured from commercially pure titanium and incorporate a conical connection (Size Narrow Platform (NP)) with a hex interface in the collar region, which combines with Nobel Biocare's existing prosthetic Abutment portfolio cleared in K071370, K083100, K132746, K233208, K202344, K200040, K132749, K161416, K202452, K240346, K041236, K133731, K061003 and K161435.
Table 2 outlines which single-unit restoration abutments with an NP connection can be used with the subject device.
Table 3 outlines which multi-unit restoration abutments and bridges with an NP connection can be used with the subject devices.
Table 4 outlines the temporary and healing restorations with an NP connection that can be used with the subject devices.
Table 5 lists the corresponding submissions where compatible Class II accessories with an NP connection were previously identified.
The On1™ Base/On1™ Base Xeal must not be used in combination with NobelActive® S, NobelParallel™ S, NobelReplace® S implants.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown are indicated for maxillary lateral and mandibular central/lateral incisors only.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown up to a maximum angulation of 20°
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with GoldAdapt Engaging CC NP abutments in straight (0°) configurations only.
No angulation is allowed for Universal Base Conical Connection Abutments when mated with the NobelActive® S, NobelParallel™ S, NobelReplace® S implants with diameters greater than 3.5mm.
N/A
Ask a specific question about this device
(255 days)
NobelZygoma TiUltra implants
The NobelZygoma TiUltra implants are endosseous dental implants intended to be surgically placed in the zygomatic bone to support a dental prosthesis in the upper arch, in order to restore patient esthetics and chewing function. The NobelZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Multi-Unite Abutment Xeal Zygoma
The Multi-unit Abutments Xeal Zygoma are indicated to support the placement of multiple unit, screw-retained prosthetic restorations in the maxilla including full arch dentures.
NobelZygoma Multi-Unite Abutment Xeal Zygoma Screws
The NobelZygoma Multi-Unit Abutment Screws are indicated for use to secure a dental abutment or framework to a dental implant for supporting tooth replacements and are indicated as an aid in prosthetic rehabilitation.
The Subject Device NobelZygoma TiUltra Implant system is composed of three device lines: NobelZygoma TiUltra Implants, Multi-unit Abutments Xeal Zygoma and NobelZygoma Multi-unit Abutment Xeal Zygoma Screws.
The NobelZygoma TiUltra Implants are endosseous dental implants surgically placed in the zygomatic bone to provide support for prosthetic devices. They are used with patients with severe bone resorption in the maxilla to provide multi-point support of the dental reconstruction. The NobelZygoma TiUltra Implants are used to support the rehabilitation of a fully or a partially edentulous maxilla. The Subject Devices are only intended for extra-maxillary placement.
The subject Implants are made from unalloyed titanium grade 4 (ASTM F67) and present a two-level anodized surface with a soluble salt (protective) layer. Two variants of the Implants are available based on the collar angulation: the NobelZygoma 0° and the NobelZygoma 45° TiUltra Implants. The NobelZygoma 0° CC TiUltra Implants feature a conical connection platform with an internal hex (size RP). The NobelZygoma 45° Ext Hex TiUltra Implants include an external hexagonal platform (size RP).
The NobelZygoma TiUltra implants are available in lengths ranging from 30 mm to 60 mm, in 2.5 mm increments. They feature a partially threaded design. The neck portion, measuring 3.6 mm from the platform, and the shaft portion which varies depending on the implant length, are unthreaded. For all implant lengths, threading consistently begins at the apex and extends 18 mm coronally.
The NobelZygoma TiUltra Implants are co-packed with an implant mount, which is attached via a pre-assembled implant mount screw to the platform of the implant.
The Multi-unit Abutments Xeal Zygoma are premanufactured prosthetic components intended to be connected to the compatible NobelZygoma TiUltra Implants to support the placement of a dental prosthesis.
The Multi-unit Abutment (MUA) Xeal Zygoma are made from Titanium alloy (Ti-6Al-4V (90% titanium, 6% aluminum, 4% vanadium, ASTM F136) and feature a non-porous oxide layer with a soluble salt (protective) layer. The Multi-unit Abutments Xeal Zygoma are attached to the respective NobelZygoma TiUltra Implants by the compatible NobelZygoma Multi-unit Abutment Xeal Zygoma Screw.
The Multi-unit Abutment Xeal Zygoma are available in four different angulations (straight or 0°, 17°, 45° and 60°) and in different heights ranging from 3 to 9mm (2mm increase). The 45° and 60° angulations are used with the NobelZygoma 0° TiUltra implants, while the Straight (or 0°) and 17° angulations are compatible with the NobelZygoma 45° TiUltra Implants. The 17° MUA is only available in two heights (3 and 5mm).
The Multi-unit Abutments Xeal Zygoma are provided sterile and co-packed with the respective abutment handle and NobelZygoma Multi-Unit Abutments Xeal Zygoma Screw.
The NobelZygoma Multi-unit Abutment Xeal Zygoma Screws are dental implant screws designed to fasten dental implant system components to a dental implant or to another component.
The subject screws are made of titanium alloy (Ti-6Al-4V (90% titanium, 6% aluminum, 4% vanadium, ASTM F136) and are partially DLC-coated. They are available in different design and sizes to fit the different Multi-unit Abutments Xeal Zygoma. The Straight NobelZygoma Multi-unit Abutment Xeal Zygoma Screws are available in four sizes to fit the respective Straight Multi-Unit Abutment heights. The screws for the 45°/60° Multi-Unit Abutments and the 17° Multi-Unit Abutment are only available in one size that fits all abutment sizes.
The NobelZygoma Multi-unit Abutment Xeal Zygoma Screws are co-packed with the compatible Mult-unit Abutment Xeal Zygoma and are also available separately.
The provided document is a 510(k) Clearance Letter from the FDA for the "NobelZygoma TiUltra Implant system." This type of document primarily focuses on establishing substantial equivalence to previously cleared predicate devices, rather than explicitly detailing acceptance criteria and presenting a single, comprehensive study proving the device meets those criteria.
However, the document does contain information about non-clinical and clinical testing performed to support the substantial equivalence claim. I will extract the relevant information to answer your questions to the best of my ability, acknowledging that not all requested details may be explicitly present in this type of regulatory submission summary.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Note: The document does not explicitly list acceptance criteria in a formal table with pass/fail values. Instead, it describes various tests performed and compares the subject device's performance to predicate devices or established standards. The "Reported Device Performance" column reflects the summary of findings from the validation studies described.
| Acceptance Criterion (Inferred from testing) | Reported Device Performance (Summary of Findings) |
|---|---|
| Mechanical Fatigue Performance | Performed at least as well as the defined benchmark reference system in dynamic loading tests (ISO 14801 adapted for zygomatic implants and FDA Guidance). |
| Implant Insertion Torque | Site preparation protocols allow seating implants with equal or less torque than the benchmark reference system. |
| Magnetic Resonance (MR) Safety | Labeled as MR Conditional according to ASTM F2503, based on testing in accordance with relevant ASTM standards and FDA Guidance. |
| Biocompatibility | Verified in accordance with ISO 10993-1 series and FDA Guidance. |
| Packaging Performance | Validated as intended (maintains integrity after simulated transportation and distribution) in accordance with ISO 11607-1, ISO 11607-2, ASTM D4169, ASTM D4332, ASTM F1886, ASTM F2096. |
| Shelf-life | Confirmed at 5 years in accordance with ASTM F1980. |
| Sterilization Validation | Conducted in accordance with ISO 11137-1, ISO 11137-2, ISO 11137-3, AAMI TIR29, and AAMI TIR 35. Achieved SAL 10-6. |
| Endotoxin Pyrogenicity | Endotoxin limit respected through LAL test according to United States Pharmacopeial Convention; periodic monitoring done as per AAMI ST72. |
| Surface Modification/Roughness | Comparative roughness verification (WLI) confirmed the subject devices exhibit a rougher surface at the collar/shaft compared to Reference Device #3. Within roughness range of Reference Device #7 (which has a similar surface technology). |
| Clinical Performance / Implant Survival | Multiple clinical studies and RWE show implant survival rates generally above 90% and up to 100% for various zygomatic implant configurations (including predicate/reference devices). Post-market surveillance data for subject devices showed similar performance in osseointegration complaint rates compared to NobelZygoma TiUnite implants. TiUltra implants overall showed a mean survival rate of 99.5%. |
| Clinical Adverse Events | Reported adverse events (e.g., implant failure, soft tissue recession, inflammation, sinusitis, hematomas, paresthesia, orbital cavity penetration, screw/prosthesis fracture) were managed with appropriate treatments and did not raise new questions of safety and effectiveness, or were comparable to known predicates. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not provide specific sample sizes for the non-clinical tests (fatigue, torque, MR, biocompatibility, packaging, sterilization, endotoxin, surface roughness). These are typically determined by relevant standards.
- For clinical published literature, the sample sizes varied:
- Three studies on non-full arch applications involved 176 similar zygomatic implants.
- Systematic literature review on quad configuration reported data for a mean implant survival rate (implies a larger, unspecified number of cases).
- Two clinical studies involving Nobel Biocare zygomatic implants: n=15 (minimum 79 months follow-up) and n=302 (mean 7.9 years follow-up).
- A total of 40 clinical publications were provided to support the established history of safe clinical use (number of implants/patients unspecified in this summary).
- Seven clinical cases with a total of 16 NobelZygoma TiUltra implants were presented.
- 36 clinical publications on NobelReplace and TiUltra implant family (number of implants/patients unspecified in this summary).
- For clinical published literature, the sample sizes varied:
- Data Provenance:
- Clinical Studies: Published literature (implies peer-reviewed, multi-center, potentially international data).
- Real-world evidence (RWE): Post-market surveillance for the Subject Devices from July 2024 to May 2025. This is prospective data specific to the subject device after initial market introduction (likely under a different clearance or within a broader market).
- Clinical Case Studies: Seven cases presented directly by the submitter.
- The document implies that data from various regions would be included in "published literature" and "post-market surveillance." No specific countries of origin are mentioned beyond "Nobel Biocare" (Sweden/Switzerland) and "Southern Implants (Pty) Ltd" (implies South Africa). The nature of most clinical data is retrospective (published studies, post-market surveillance).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not specify the number of experts used to establish ground truth for the test set (clinical studies). Clinical studies typically rely on diagnoses and assessments made by treating clinicians (dental surgeons, prosthodontists) based on established medical practices and diagnostic criteria, rather than a separate "ground truth" panel for the study itself, unless it's a specific adjudication or consensus study. Such details are usually found within the full study protocols and reports, which are not part of this 510(k) summary.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
The document does not specify any adjudication methods for the clinical data presented. This information would typically be detailed in the methodology sections of the individual clinical studies or systematic reviews referenced, which are not provided in this summary.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document describes a dental implant system (hardware), not an AI-powered diagnostic or assistive tool. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study focusing on human reader improvement with/without AI assistance is not applicable and was not performed.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is not applicable as the device is a dental implant system, not an algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- Non-Clinical Tests: Ground truth is based on established engineering and materials science principles, international standards (ISO, ASTM), and FDA guidance documents. Performance is evaluated against these benchmarks.
- Clinical Performance Data: Ground truth is primarily based on outcomes data, specifically:
- Implant survival rate: Directly reported as percentage (e.g., 97.4%, 100%, 89.9%, 99.5%).
- Reported adverse events and complications: Clinical observations and diagnoses by treating dentists/surgeons. Resolution of these events are considered outcomes.
- Osseointegration: Inferred from success rates and lack of failure/complaints related to implant stability.
- Patient esthetics and chewing function: These are indications, and clinical data indirectly supports their achievement through successful implant function.
8. The Sample Size for the Training Set
The concept of a "training set" is typically associated with machine learning or AI algorithms. Since this is a physical medical device (dental implant system), it primarily relies on engineering design, materials science, and clinical validation. Therefore, a "training set" in the context of an algorithm is not applicable. The development and testing would involve design iterations and validation studies, but not an algorithmic training process with a distinct training set.
9. How the Ground Truth for the Training Set Was Established
As noted in point 8, the concept of a "training set" in the context of an AI/ML algorithm is not applicable to this physical device. Therefore, the establishment of ground truth for such a set is also not applicable. The "ground truth" for the device's design and performance validation is derived from established scientific principles, pre-clinical testing, and clinical outcomes for the device itself and its predicates.
Ask a specific question about this device
(333 days)
The NobelProcera® Zirconia Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.
The NobelProcera Zirconia Implant Bridge (Dental Bridge) is a patient-specific, dental implant supported, screw-retained dental implant bridge which is connected to compatible Nobel Biocare root-form endosseous dental implants (Dental Implants) or Multi-unit Abutments and is intended to restore chewing function in partially and fully edentulous patients. The Dental Bridge is available as either a Framework requiring veneering in a dental lab or as a Full Contour design requiring minimum laboratory processing. The Dental Bridge is made from yttria-stabilized tetragonal zirconia. It is designed in a dental laboratory, hospital or dental practice by scanning, designing and ordering the restoration using dental CAD/CAM software and approved dental scanner. The finished design is sent to Nobel Biocare manufacturing facility for industrial production. After production, the Dental Bridge is sent to the laboratory for finishing. The Dental Bridge is available for use with Nobel Biocare's Dental Implants having Internal Conical Connection (CC), External Hex Connection, Internal Tri-Channel Connection and Nobel Biocare's Multi-unit Abutment connections. One Dental Bridge can feature connections to 2 up to 10 Dental Implants. All Dental Bridges are provided with the required Clinical and/or Prosthetic Screw: The Clinical Screw connects the Dental Bridge with the Dental Implant. The Prosthetic Screw connects the Dental Bridge with a Multi-unit Abutment. Dental Bridge connections on the Internal Conical Connection Dental Implant require use of Clinical Metal Adapters. Clinical Metal Adapters are provided with the Dental Bridge. No adapter is needed for the External Hex, Internal Tri-Channel, or Multi-unit Abutment connections. Clinical and Prosthetic Screws are placed from the occlusal side of the restoration. To facilitate access to the connection of the restoration to the implant/abutment a screw channel must be designed into the restoration. The Dental Bridge feature an Angulated Screw Channel (ASC) when connected to a Dental Implant with Internal Conical Connection and has an ASC feature to connect to Multi-unit Abutments. NobelProcera Zirconia Implant Bridges ASC's angulation can be defined by the customer in an angulation (to the implant's or Multi-unit Abutment's axis) between 0° and 25°. The ASC allows access through a screw channel which is not in line with the implant.
The provided text is a 510(k) summary for the NobelProcera Zirconia Implant Bridge. It describes the device, its indications for use, and a comparison to a predicate device.
However, the document states: "Clinical data is not required to establish substantial equivalence for the Subject Device. Non-clinical test data was conducted in accordance with FDA Guidance 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments'. This testing involved examining finished assembled implant/abutment systems of the worst-case scenario through fatigue testing."
This indicates that a clinical study with acceptance criteria, sample sizes for test and training sets, expert ground truth, adjudication methods, or MRMC studies for AI performance was not part of the submission for this particular device. The evaluation relied on non-clinical (bench) testing, specifically fatigue testing, to demonstrate substantial equivalence to a predicate device.
Therefore, I cannot provide the requested information from the provided text as it pertains to clinical studies or AI performance. The document only mentions "fatigue performance testing" as the evidence for the revised extension length.
To answer your request, here's what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: While not explicitly listed as a "table of acceptance criteria" in the format usually requested for clinical studies, the document implies that the device's performance regarding "extension length" was evaluated through fatigue testing. The revised extension length of 16mm (for the subject device) versus 10mm (for the predicate device) suggests an acceptance criterion related to the ability to withstand fatigue at this increased length. The wording "fatigue performance testing was performed to demonstrate that the difference does not raise different questions of safety and effectiveness" implies that the device successfully met the fatigue requirements for the 16mm extension.
- Reported Device Performance: The document states that "The revised extension length is supported by the fatigue performance testing provided in the premarket notification." and "fatigue testing was performed to demonstrate that the difference does not raise different questions of safety and effectiveness." This indicates that the subject device, with its maximum extension length of 16mm, passed the non-clinical fatigue tests, demonstrating performance equivalent to or better than the predicate device.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not applicable for a non-clinical fatigue test. The testing involved "examining finished assembled implant/abutment systems of the worst-case scenario through fatigue testing." The number of units tested is not specified in this summary. Data provenance is not relevant for bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. This was a non-clinical, benchtop fatigue test, not a study requiring expert ground truth for clinical interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This was a non-clinical, benchtop fatigue test.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a dental implant bridge, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This device is a physical medical device, not a software algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- Not applicable for non-clinical fatigue testing. The "ground truth" for fatigue testing would be the physical measurement of failure or deformation under specified loading conditions, and comparison against a defined standard or predicate performance.
8. The sample size for the training set:
- Not applicable. This document describes a physical medical device and non-clinical testing, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established:
- Not applicable. See point 8.
Ask a specific question about this device
(118 days)
NobelProcera® Titanium ASC Abutments:
NobelProcera® Abutment Titanium is a patient-matched CAD/CAM prosthetic component directly connected to endosseous dental implants and is indicated for use as an aid in prosthetic rehabilitation.
Omnigrip Clinical Screw Titanium:
Clinical and Abutment Screws are indicated for use to secure a dental abutment or framework to a dental implant in the maxilla or mandible for supporting tooth replacements and are indicated as an aid in prosthetic rehabilitation.
The Subject Device NobelProcera® Titanium ASC Abutment is composed of two device lines: NobelProcera® Titanium ASC Abutment and the Clinical Screw Omnigrip Titanium.
NobelProcera® Titanium ASC Abutment is a patient matched CAD/CAM dental prosthesis which is connected to the Nobel Biocare implants featuring an internal conical and/or internal tri-channel connection, is available in the platform sizes NP, RP WP and 6.0 (for internal tri-channel connection only) and is intended for use as an aid in prosthetic rehabilitation to restore chewing function and esthetic appearance.
NobelProcera® Titanium ASC Abutments undergo patient matched customization at a Nobel Biocare production facility for the final, finished abutment device manufacturing.
NobelProcera® Titanium ASC Abutment is connected to the implant with a clinical screw and features an angulated screw channel which can be defined by the customer in an angulation (to the implant's axis) between 0° and 25°, in addition the abutment can be angulated to a maximum of 30°. The clinical screw features the Omnigrip Interface which allows tightening up to 25°.
NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium are composed of titanium vanadium alloy Ti6Al4V ELI (ISO 5832-3, ASTM F136) and the surface of the abutments are provided with and without anodization and the Omnigrip Clinical Screw Titanium are provided with and without DLC coating.
The finished NobelProcera® Titanium ASC Abutment supports the placement of a cement-retained dental prosthesis.
Omnigrip Clinical Screw Titanium is available for the NP, RP, WP and 6.0 (for internal tri-channel connection only) platform, the devices connect the NobelProcera Titanium ASC Abutments to the dental implants. The devices feature an Omnigrip interface.
Here's an analysis of the provided text regarding acceptance criteria and device performance.
Important Note: The provided document is an FDA 510(k) Premarket Notification letter. This type of document is a submission seeking regulatory clearance, not a standalone study report. As such, it outlines the basis for demonstrating substantial equivalence to a predicate device, which includes non-clinical testing. It does not typically contain detailed acceptance criteria and performance data for a study proving a device meets those criteria in the way a clinical trial report or a comprehensive validation study would.
The document primarily focuses on demonstrating that the new device is substantially equivalent to legally marketed predicate devices, meaning it is as safe and effective. It refers to non-clinical tests that were performed, but does not present the specific acceptance criteria or the numerical results of those tests in a "performance table" format.
Therefore, many of the requested points below cannot be fully extracted from this document, as it's not the type of report that contains that level of detail for acceptance criteria and specific study outcomes. I will highlight what can be inferred or directly stated from the provided text, and what cannot.
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner as one might find in a detailed engineering validation report or a clinical study. Instead, it states that "non-clinical tests demonstrate that the device is substantial equivalent."
The document mentions several tests and standards the device was subjected to:
- Packaging system performance testing: per ASTM D4169. (Specific acceptance criteria and reported performance not detailed.)
- Dynamic loading testing: conducted according to ISO 14801 and FDA Guidance Document ("Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" (May 12, 2004)). (Specific acceptance criteria and reported performance not detailed.)
- Magnetic Resonance compatibility testing: according to ASTM F2052, ASTM F2213, ASTM F2119, and ASTM F2182. Status: "MR Conditional." (This is a performance outcome, but the specific metrics and acceptance thresholds for each ASTM standard are not detailed.)
- Verification of biocompatibility: in accordance with ISO 10993-1. (The conclusion is that "no new issues regarding biocompatibility were raised," implying acceptance, but specific criteria and test results not detailed.)
- End user cleaning and sterilization validation: in accordance with ISO 17665-1. (Implied acceptance, but specific criteria and results not detailed.)
The "acceptance criteria" here are generally referred to as meeting the requirements of the standards and demonstrating substantial equivalence to the predicate device. The "reported device performance" is summarized as favorable results sufficient for substantial equivalence determination.
2. Sample sizes used for the test set and the data provenance
- Sample sizes: Not specified in the provided text for any of the non-clinical tests mentioned (packaging, dynamic loading, MR compatibility, biocompatibility, sterilization validation).
- Data provenance: Not explicitly stated (e.g., country of origin). The testing is non-clinical/bench testing, not patient data. It is implied to be internal testing by the manufacturer, Nobel Biocare AB (Sweden).
- Retrospective or prospective: N/A, as these are non-clinical/bench tests, not clinical studies involving patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Applicability: This question is not directly applicable to the type of non-clinical, bench testing described for the NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium. "Ground truth" established by experts (e.g., radiologists for image interpretation) is relevant for AI/ML device evaluations, particularly in diagnostics. This document refers to physical device testing (mechanical, biocompatibility, etc.).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Applicability: Not applicable. Adjudication methods are typically used in clinical studies or expert reviews to resolve disagreements in interpretations or diagnoses for establishing "ground truth," which is not the nature of the non-clinical tests described here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Applicability: Not applicable. This device is a dental abutment and screw, not an AI/ML-enabled diagnostic device for image interpretation. Therefore, no MRMC study or AI assistance evaluation would have been performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Applicability: Not applicable. See point 5. This is a physical medical device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Applicability: The concept of "ground truth" as typically applied in AI/ML validation (e.g., for diagnostic accuracy) does not directly apply to the non-clinical device testing described. For these tests, the "ground truth" is typically defined by the specified parameters and performance requirements of the relevant international standards (e.g., ISO 14801 for dynamic loading, ISO 10993-1 for biocompatibility). The devices are expected to meet the performance criteria defined by these standards.
8. The sample size for the training set
- Applicability: Not applicable. This refers to the training of an AI/ML algorithm. The document describes a physical medical device.
9. How the ground truth for the training set was established
- Applicability: Not applicable. See point 8.
Ask a specific question about this device
(163 days)
DTX Studio Clinic is a software program for the acquisition, management, transfer and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental restorative solutions.
It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations.
It can be used to support guided implant surgery whereby the results can be exported. DTX Studio Clinic is a computer assisted detection (CADe) device that analyses intraoral radiographs to identify and localize dental findings, which include caries, calculus, periapical radiolucency, root canal filling deficiency, discrepancy at the margin of an existing restoration and bone loss. The DTX Studio Clinic CADe functionality is indicated for use by dentists for the concurrent review of bitewing and periapical radiographs of permanent teeth in patients 15 years of age or older.
DTX Studio™ Clinic is a software interface for dental/medical practitioners used to analyze 2D and 3D imaging data, in a timely fashion, for the treatment of dental, craniomaxillofacial and related conditions. DTX Studio Clinic displays and processes imaging data from different devices (i.e. Intra/Extra Oral X-Rays, (CB)CT scanners, intraoral and extraoral cameras).
DTX Studio Clinic features an Al-powered Focus Area Detection algorithm which analyzes 2D intraoral radiographs for potential dental findings or image artifacts. The detected focus areas can be converted afterwards to diagnostic findings after approval by the user.
The following dental findings can be detected by the device:
• Caries
· Discrepancy at marqin of an existing restoration
· Periapical radiolucency
- Root canal filling deficiency
• Bone loss
· Calculus
The provided text describes the 510(k) summary for the DTX Studio Clinic (4.0) device, focusing on its substantial equivalence to predicate devices, particularly regarding its AI-powered "Focus Area Detection" functionality. While the document mentions verification and validation activities, it does not provide a detailed breakdown of the acceptance criteria and performance study results as requested by the prompt.
Specifically, it states:
- "A comparative analysis between the output of the focus area detection algorithm of the output of the primary predicate device (K221921) was performed. The test results show that for each intraoral x-ray image, the focus area detection is executed, and the same number of focus areas were detected compared to the predicate device. All detected bounding boxes were identical and the acceptance criteria was met."
This statement indicates that the new device's AI performance for "Focus Area Detection" was evaluated based on its identity to the primary predicate device (K221921), not against independent clinical performance criteria like sensitivity, specificity, or AUC based on a ground truth established by experts. It implies a non-inferiority or equivalence study where the benchmark is the previously cleared AI, rather than a de novo clinical performance evaluation.
Therefore, many of the requested details about the study are not present in this document. I will fill in the table and address the questions based on the information available, highlighting what is not provided.
Acceptance Criteria and Device Performance Study Details for DTX Studio Clinic (4.0) Focus Area Detection
Based on the provided 510(k) summary, the evaluation of the DTX Studio Clinic (4.0)'s AI-powered "Focus Area Detection" functionality was primarily a comparative analysis against its primary predicate device (DTX Studio Clinic 3.0, K221921), rather than a standalone clinical performance study establishing specific diagnostic metrics against a human-expert ground truth. The acceptance criterion appears to be identical output to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Metric | Acceptance Criteria (as per document) | Reported Device Performance (as per document) |
|---|---|---|
| Focus Area Detection Output | For each intraoral x-ray image, the focus area detection must be executed, and the same number of focus areas must be detected compared to the predicate device (K221921). All detected bounding boxes must be identical to the predicate. | "The test results show that for each intraoral x-ray image, the focus area detection is executed, and the same number of focus areas were detected compared to the predicate device. All detected bounding boxes were identical and the acceptance criteria was met." Additionally, "There are no functional or technical differences between the Focus Area detection in DTX Studio Clinic 3.0 (primary predicate - K221921) and the current subject device." |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated. The document refers to "each intraoral x-ray image" without specifying the total number of images used in the comparative analysis.
- Data Provenance: Not explicitly stated (e.g., country of origin). The document implies the use of intraoral x-ray images, but their source, whether retrospective or prospective, or geographical origin, is not detailed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable / Not provided. The ground truth for the evaluation appears to be the output of the predicate AI device (K221921), not a human expert consensus. Therefore, no human experts were explicitly stated to establish a direct ground truth for the DTX Studio Clinic (4.0)'s performance itself in this comparative test. The predicate device's original clearance would have been based on such studies.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable / None specified. Since the comparison was between the subject device's AI output and the predicate AI device's output, there's no mention of human adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC comparative effectiveness study was not described for the DTX Studio Clinic (4.0) for this submission. The evaluation was a technical comparison of the AI output to an existing cleared AI, not a study on human-AI augmented performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, in spirit, but not as a de novo clinical performance study. The described test was a standalone comparison of the algorithm's output (DTX Studio Clinic 4.0) against another algorithm's output (predicate K221921) to confirm identical functionality. It was not a performance study measuring the algorithm's diagnostic accuracy (e.g., sensitivity, specificity) against a clinical gold standard. The document emphasizes that "There are no functional or technical differences between the Focus Area detection in DTX Studio Clinic 3.0 (primary predicate - K221921) and the current subject device."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The output of the predicate AI device (K221921). The "ground truth" for verifying the DTX Studio Clinic (4.0)'s "Focus Area Detection" functionality was the identical output of the legally marketed predicate device (K221921), which previously underwent its own validation.
8. The sample size for the training set
- Not provided. The document describes the "Focus Area Detection" as an "AI-powered" algorithm using "supervised machine learning," but it does not specify the sample size of the training set used for this algorithm. This information would typically be part of the original K221921 submission.
9. How the ground truth for the training set was established
- Not provided. The document states "supervised machine learning" was used, which implies a labeled training dataset. However, the method for establishing the ground truth (e.g., expert annotations, pathology confirmation) for that training data is not detailed in this 510(k) summary. This information would likely have been part of the K221921 submission when the AI algorithm was first cleared.
Ask a specific question about this device
(141 days)
Titanium Abutment Blank Nobel Biocare N1™ TCC is a premanufactured prosthetic component directly connected to an endosseous dental implant and is indicated for use as an aid in prosthetic rehabilitation for single units and multiple units of up to three units.
The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories.
The Titanium Abutment Blank Nobel Biocare N1™ TCC is a premanufactured titanium abutment which can be customized via a validated CAD/CAM workflow in the dental office or dental laboratory to meet patient-specific anatomical requirements. The customization of the subject device is designed using a dental laboratory software and milled in the dental laboratory, using a Computer Aided Design (CAD)/Computer Aided Manufacturing (CAM) machine.
All digitally designed CAD/CAM customizations for the Titanium Abutment Blank Nobel Biocare N1™ TCC are only intended to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories.
The subject device has a premanufactured connection for the Nobel Biocare N1 ™ TCC TiUltra implants on one end and a premanufactured connection for the milling blank holder on the other end. These connections are not patient-specific.
The subject device is available for NP and RP implant platforms.
The subject device is used by dental healthcare professionals in dental offices and dental laboratories.
The subject device is composed of titanium vanadium alloy Ti6Al4V ELI (ISO 5832-3, ASTM F136) and features a surface with the same anodization already cleared in K211109.
It is an implantable single use device. The device is provided non-sterile and intended to be sterilized by the user prior to placement in the patient.
The Titanium Abutment Blank Nobel Biocare N1™ TCC is packaged with a Clinical Screw NB N1 TCC.
This document is a 510(k) summary for the Titanium Abutment Blank Nobel Biocare N1™ TCC. It is a premarket notification for a medical device seeking clearance from the FDA, asserting substantial equivalence to previously cleared predicate devices.
Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly provide a table of "acceptance criteria" for the overall device. Instead, it details various non-clinical performance tests conducted and indicates that the device met established performance specifications. The "reported device performance" is described through the outcomes of these tests.
However, based on the non-clinical test data section and the comparison table (Table 2), we can infer certain performance aspects that were evaluated against reference points (predicates or standards).
| Characteristic / Test | Acceptance Criteria (Inferred from Predicate/Standards) | Reported Device Performance |
|---|---|---|
| End-User Cleaning and Sterilization | In accordance with FDA Guidance: "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance... issued March 17, 2015" | Validation was conducted and met the guidance specifications. |
| Biocompatibility | In accordance with ISO 10993 | Testing was conducted, and no new issues regarding biocompatibility were raised. |
| Fatigue Testing | In accordance with FDA Guidance: "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments", issued May 12, 2004, and demonstrated substantial equivalence to predicate/reference devices. | Testing conducted on the worst-case system demonstrated substantial equivalence to the predicate and reference devices. |
| Software Verification and End-to-end Workflow Validation | Applicable restrictions are in place and cannot be modified by the user; avoidance of damage or modification of implant-abutment connection geometry during milling; locking of restriction zones from user editing. | Validation was completed, demonstrating that applicable restrictions are in place and cannot be modified by the user, and showing avoidance of damage/modification of connection geometry, and locking of restriction zones. |
| MRI Compatibility | MR Conditional classification (leveraged from K212125). | Demonstrated that the subject device is MR Conditional. |
| Minimum Wall Thickness | Similar to predicate (0.4-0.5mm) and existing reference device #2 (0.3mm). | NP 0.38mm; RP 0.49mm, deemed substantially equivalent as demonstrated by fatigue testing. |
| Minimum Gingiva Height | Similar to predicate (0.5mm) and existing reference device #1 (0.3mm). | 0.335mm, deemed substantially equivalent as demonstrated by fatigue testing. |
| Minimum Diameter | Similar to predicate (3.0-6.0mm) and reference device #1 (2.52-6.0mm). | NP Ø3.21mm; RP Ø3.49mm, deemed similar. |
| Minimum Post Height | Similar to predicate (4mm) and reference device #1 (4mm). | 4.05mm, deemed similar. |
| Maximum Abutment Angulation | Similar to predicate (30°/20°) and reference device #1 (30°). | 30°, deemed identical. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the numerical sample sizes for each non-clinical test (e.g., number of abutments for fatigue testing, or number of software test cases). It refers to tests being conducted on "the worst-case system" for fatigue testing.
The data provenance is internal testing conducted by the manufacturer, Nobel Biocare AB. The document does not specify a country of origin for the data itself, but the submitter is based in Sweden. These are retrospective tests conducted to support the premarket notification.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not applicable as the document describes non-clinical engineering and performance testing of a physical and software device, not a diagnostic or AI-driven decision support system requiring expert-established ground truth from patient data.
4. Adjudication Method for the Test Set
This information is not applicable for the same reason as point 3.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study was not done. This device is a component for dental implants and does not involve human readers or AI in the context of comparative effectiveness for diagnostic or interpretive tasks.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This concept is primarily relevant for AI/ML-enabled diagnostic or decision-support software. While the device utilizes CAD/CAM software, the "standalone performance" in this context refers to the software's ability to accurately design and a milling machine's ability to accurately produce the customized abutment according to the design parameters and restrictions. The "Software verification and End-to-end workflow validation" addresses this by demonstrating that the software ensures applicable restrictions are in place and cannot be modified by the user, and that damage to the connection geometry is avoided during milling. It implicitly assesses the algorithm's performance within the manufacturing workflow.
7. The Type of Ground Truth Used
For the non-clinical tests described:
- Cleaning and Sterilization: The "ground truth" is adherence to established FDA guidance and validated protocols.
- Biocompatibility: The "ground truth" is compliance with ISO 10993 standards.
- Fatigue Testing: The "ground truth" is compliance with FDA guidance for dental implants and abutments and demonstrated substantial equivalence to the mechanical properties of predicate devices. This involves mechanical testing to failure or for a specified number of cycles under defined loads.
- Software Verification and End-to-end Workflow Validation: The "ground truth" is the established design parameters, manufacturing specifications, and the functionality requirements for the software (e.g., locking restriction zones, avoiding damage to connection geometry). It's based on engineering specifications.
- MRI Compatibility: The "ground truth" is the established MR Conditional specifications from the referenced K212125 clearance.
8. The Sample Size for the Training Set
This information is not applicable. This device is a physical medical device manufactured using a CAD/CAM workflow, not an AI/ML system that requires a "training set" in the conventional sense. The CAD/CAM software would have been developed and validated through its own software lifecycle, but that is distinct from an AI model training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as point 8.
Ask a specific question about this device
(270 days)
DTX Studio Clinic is a computer assisted detection (CADe) device that analyses intraoral radiographs to identify and localize dental findings, which include caries, calculus, periapical radiolucency, root canal filling deficiency, discrepancy at margin of an existing restoration and bone loss.
The DTX Studio Clinic CADe functionality is indicated for the concurrent review of bitewing and periapical radiographs of permanent teeth in patients 15 years of age or older.
DTX Studio Clinic features an AI-powered Focus Area Detection algorithm which analyzes intraoral radiographs for potential dental findings or image artifacts. The detected focus areas can be converted afterwards to diagnostic findings after approval by the user. The following dental findings can be detected by the device: Caries, Discrepancy at margin of an existing restoration, Periapical radiolucency, Root canal filling deficiency, Bone loss, Calculus.
The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the DTX Studio Clinic 3.0.
Here's the breakdown:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" as a pass/fail threshold, but rather presents the performance results from the standalone (algorithm only) and clinical (human-in-the-loop) studies. The acceptance is implicitly based on these results demonstrating clinical benefit and safety.
Standalone Performance (Algorithm-Only)
| Dental Finding Type | Metric | Reported Performance (95% CI) |
|---|---|---|
| Caries | Sensitivity | 0.70 [0.65, 0.75] |
| Mean IoU | 58.6 [56.2, 60.9]% | |
| Mean Dice | 71.9 [69.9, 74.0]% | |
| Periapical Radiolucency | Sensitivity | 0.68 [0.59, 0.77] |
| Mean IoU | 48.9 [44.9, 52.9]% | |
| Mean Dice | 63.7 [59.9, 67.5]% | |
| Root Canal Filling Deficiency | Sensitivity | 0.95 [0.91, 0.99] |
| Mean IoU | 51.9 [49.3, 54.6]% | |
| Mean Dice | 66.9 [64.3, 69.4]% | |
| Discrepancy at Restoration Margin | Sensitivity | 0.82 [0.77, 0.87] |
| Mean IoU | 48.4 [46.0, 50.7]% | |
| Mean Dice | 63.5 [61.3, 65.8]% | |
| Bone Loss | Sensitivity | 0.78 [0.75, 0.81] |
| Mean IoU | 44.8 [43.4, 46.3]% | |
| Mean Dice | 60.1 [58.7, 61.6]% | |
| Calculus | Sensitivity | 0.80 [0.76, 0.84] |
| Mean IoU | 55.5 [53.7, 57.3]% | |
| Mean Dice | 70.1 [68.4, 71.7]% | |
| Overall | Sensitivity | 0.79 [0.74, 0.84] |
| Precision | 0.45 [0.40, 0.50] |
Clinical Performance (Human-in-the-Loop)
| Metric | Reported Performance (95% CI) |
|---|---|
| Overall AUC Increase (Aided vs. Unaided) | 8.7% [6.5, 10.9]% |
| Caries AUC Increase | 6.1% |
| Periapical Radiolucency AUC Increase | 10.2% |
| Root Canal Filling Deficiency AUC Increase | 13.5% |
| Discrepancy at Restoration Margin AUC Increase | 10.1% |
| Bone Loss AUC Increase | 5.6% |
| Calculus AUC Increase | 7.2% |
| Overall Instance Sensitivity Increase | 22.4% [20.1, 24.7]% |
| Caries Sensitivity Increase | 19.6% |
| Bone Loss Sensitivity Increase | 23.5% |
| Calculus Sensitivity Increase | 18.1% |
| Discrepancy at Restoration Margin Sensitivity Increase | 28.5% |
| Periapical Radiolucency Sensitivity Increase | 20.6% |
| Root Canal Filling Deficiency Sensitivity Increase | 27.4% |
| Overall Image Level Specificity Decrease | 8.7% [6.6, 10.7]% |
2. Sample Size and Data Provenance
-
Test Set (Standalone Performance):
- Sample Size: 452 adult intraoral radiograph (IOR) images (bitewings and periapical radiographs).
- Data Provenance: Not explicitly stated, but implicitly retrospective as they were "assembled" and "ground-truthed" for the study.
-
Test Set (Clinical Performance Assessment - MRMC Study):
- Sample Size: 216 periapical and bitewing IOR images.
- Data Provenance: Acquired in US-based dental offices by either sensor or photostimulable phosphor plates. This suggests retrospective collection from real-world clinical settings in the US.
3. Number of Experts and Qualifications for Ground Truth
-
Test Set (Standalone Performance):
- Number of Experts: A group of 10 dental practitioners followed by an additional expert review.
- Qualifications: "Dental practitioners" and "expert review" (no further details on experience or specialized qualifications are provided for this set).
-
Test Set (Clinical Performance Assessment - MRMC Study):
- Number of Experts: 4 ground truthers.
- Qualifications: All ground truthers have "at least 20 years of experience in reading of dental x-rays."
4. Adjudication Method for the Test Set
-
Test Set (Standalone Performance): "ground-truthed by a group of 10 dental practitioners followed by an additional expert review." - The specific consensus method (e.g., majority vote) is not explicitly stated.
-
Test Set (Clinical Performance Assessment - MRMC Study): Ground truth was defined by 4 ground truthers with a 3 out of 4 consensus. This is an explicit 3+1 (or simply 3 out of 4) adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Yes, a MRMC comparative effectiveness study was done.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:
- The primary endpoint, overall Area Under the Curve (AUC), showed a statistically significant increase of 8.7% (CI [6.5, 10.9], p<0.001) for all dental finding types combined in the aided reading arm compared to the control (unaided) arm.
- Instance-level sensitivity (reader's ability to detect existing dental findings) improved substantially by 22.4% (CI [20.1, 24.7]) overall.
- Specific AUC increases per finding type ranged from 5.6% (bone loss) to 13.5% (root canal filling deficiency).
- Specific sensitivity increases per finding type ranged from 18.1% (calculus) to 28.5% (discrepancy at restoration margin).
- A mild decrease in image-level specificity of 8.7% (CI [6.6, 10.7]) was observed.
6. Standalone (Algorithm Only) Performance
- Yes, a standalone performance assessment was done.
- This assessment measured the performance of the AI-powered Focus Area Detection algorithm "by itself, in the absence of any interaction with a dentist." The results (Overall sensitivity, precision, and Mean IoU/Dice scores per finding type) are detailed in the table in section 1 above.
7. Type of Ground Truth Used
- Expert Consensus:
- For the standalone performance study, ground truth was established by "a group of 10 dental practitioners followed by an additional expert review."
- For the clinical performance (MRMC) study, ground truth was defined by "4 ground truthers with a 3 out of 4 consensus," all having "at least 20 years of experience in reading of dental x-rays."
8. Sample Size for the Training Set
- The document does not specify the sample size for the training set. It only describes the test sets.
9. How Ground Truth for the Training Set Was Established
- The document does not describe how the ground truth for the training set was established. It only refers to the training process generically as "supervised machine learning" and focuses on the ground truthing of the test sets.
Ask a specific question about this device
(136 days)
DTX Studio Clinic is a software program for the acquisition, management, transfer and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental restorative solutions. It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations.
DTX Studio Clinic is a software interface for dental/medical practitioners used to analyze 2D and 3D imaging data, in a timely fashion, for the treatment of dental, cramomaxillofacial and related conditions. DTX Studio Clinic displays and processes imaging data from different devices (i.e. intraoral X-Rays, (CB)CT scanners, intraoral scanners, intraoral and extraoral cameras).
This document is a 510(k) Premarket Notification for the DTX Studio Clinic 3.0. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed technical study report with specific acceptance criteria and performance metrics for novel functionalities.
Therefore, the requested information regarding detailed acceptance criteria, specific performance data (e.g., accuracy metrics), sample sizes for test sets, data provenance, expert qualifications, and ground truth establishment for the automatic annotation of mandibular canals is not explicitly detailed in the provided text.
The document states that "Automatic annotation of the mandibular canals" is a new feature in DTX Studio Clinic 3.0, and it is compared to the reference device InVivoDental (K123519) which has "Creation and visualization of the nerve manually or by using the Automatic Nerve feature." However, it does not provide the specific study details for validating this new feature within DTX Studio Clinic 3.0. It only broadly states that "Software verification and validation testing was conducted on the subject device."
Based on the provided text, I cannot fulfill most of the requested information directly because it is not present. The document's purpose is to establish substantial equivalence based on the overall device function and safety, not to detail the rigorous validation of a specific AI/ML component with numerical acceptance criteria.
However, I can extract the available information and highlight what is missing.
Acceptance Criteria and Study for DTX Studio Clinic 3.0's Automatic Mandibular Canal Annotation (Information extracted from the document):
Given the provided text, the specific, quantitative acceptance criteria and detailed study proving the device meets these criteria for the automatic annotation of the mandibular canal are not explicitly described. The document focuses on a broader claim of substantial equivalence and general software validation.
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature/Metric | Acceptance Criteria | Reported Device Performance | Source/Methodology (if available in text) |
|---|---|---|---|
| Automatic annotation of mandibular canals | Not explicitly stated in quantitative terms. Implied acceptance is that the functionality is "similar as in the reference device InVivoDental (K123519)" and the user can "manually indicate or adjust the mandibular canal." | No specific performance metrics (e.g., accuracy, precision, recall, Dice coefficient) are provided. The text states: "The software automatically segments the mandibular canal based on the identification of the mandibular foramen and the mental foramen. This functionality is similar as in the reference device InVivoDental (K123519). The user can also manually indicate or adjust the mandibular canal." | Comparison to reference device and user adjustability. Software verification and validation testing was conducted, but details are not provided. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified for the automatic mandibular canal annotation feature. The document states "Software verification and validation testing was conducted on the subject device," but provides no numbers.
- Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: Not mentioned or detailed. The document primarily makes a substantial equivalence claim based on the device's overall functionality and features, not a comparative effectiveness study involving human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not explicitly detailed. The document describes the automatic segmentation functionality and mentions that the user can manually adjust, implying a human-in-the-loop scenario. No standalone performance metrics are provided.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Type of Ground Truth: Not specified for the automatic mandibular canal annotation. Given the context of a dental/maxillofacial imaging device, it would likely involve expert annotations on CBCT scans, but this is not confirmed in the text.
8. The sample size for the training set:
- Training Set Sample Size: Not specified. This document is a 510(k) submission, which focuses on validation, not the development or training process.
9. How the ground truth for the training set was established:
- Ground Truth Establishment for Training Set: Not specified.
Summary of what can be inferred/not inferred from the document regarding the mandibular canal annotation:
- New Feature: Automatic annotation of mandibular canals is a new feature in DTX Studio Clinic 3.0 that was not present in the primary predicate (DTX Studio Clinic 2.0).
- Comparison to Reference Device: This new feature's "functionality is similar as in the reference device InVivoDental (K123519)", which has "Creation and visualization of the nerve manually or by using the Automatic Nerve feature."
- Human Oversight: The user has the ability to "manually indicate or adjust the mandibular canal," suggesting that the automatic annotation is an aid to the diagnostic process, not a definitive, unreviewable output. This is typical for AI/ML features in medical imaging devices that are intended to support, not replace, clinical judgment.
- Validation Claim: The submission states that "Software verification and validation testing was conducted on the subject device and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices'." This implies that the validation was performed in accordance with regulatory guidelines, but the specific details of that validation for this particular feature are not disclosed in this public summary.
Ask a specific question about this device
(98 days)
The NobelProcera® Zirconia Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.
NobelProcera Zirconia Implant Bridges (Dental Bridges) are patient-specific, dental implant supported, screw-retained dental implant bridges which are connected to compatible Nobel Biocare root-form endosseous dental implants or Multi-unit abutments and are intended to restore chewing function in partially and fully edentulous patients. The Dental Bridge is available as either a Framework requiring veneering in a dental lab or as a Full Contour design requiring minimum laboratory processing.
NobelProcera Zirconia Implant Bridges are made from 'Nacera Pean' (yttriastabilized tetragonal zirconia), Reference Device #1, K143071. The Dental Bridges are designed in a dental laboratory, hospital or dental practice by scanning, designing and ordering the restoration using dental CAD/CAM software and a Nobel Biocare/KaVo-approved dental scanner.
The finished design is sent to Nobel Biocare manufacturing facility for industrial production. After production, the Dental Bridge is sent to the laboratory for finishing.
NobelProcera Zirconia Implant Bridges are available for use with Nobel Biocare's root-form endosseous dental implants (Dental Implants) havinq Internal Conical Connection (CC), External Hex Connection, Internal Tri-Channel Connection and Nobel Biocare's Multi-unit Abutment Connections (for MUA and MUA Plus). One Dental Bridge can feature connections to 2 to 10 Dental Implants.
All NobelProcera Zirconia Implant Bridges are provided with the required clinical and/or Prosthetic Screw: The clinical screw connects the Dental Bridge with the Dental Implant. The Prosthetic Screw connect the Dental Bridge with a Multi-unit Abutment.
Dental Bridge connections on the Internal Conical Connection Dental Implant require use of Clinical Metal Adapters. Clinical Metal Adapters are provided with the Dental Bridge. No adapter is needed for the external hex, internal tri-channel, or Multi-unit abutment connections.
The provided text is a 510(k) Summary for a medical device (NobelProcera Zirconia Implant Bridge) seeking substantial equivalence to a predicate device. It describes performance testing but does not present the specific acceptance criteria or the reported device performance in a table, nor does it detail a study designed to directly prove the device meets pre-defined acceptance criteria in the manner requested.
Instead, the document focuses on demonstrating substantial equivalence to a predicate device through a comparison of technological characteristics and performance data. The performance data mentioned are from fatigue testing and biocompatibility evaluation, suggesting these are the key areas where the device's performance needs to be comparable to or better than the predicate.
Here's an analysis of the available information in the context of your request:
1. A table of acceptance criteria and the reported device performance:
The document does not provide a table explicitly outlining acceptance criteria and reported device performance. It only states that "The results of the testing were used to address questions related to substantial equivalence based on difference in design between the Subject and Predicate Devices." and that "The performance testing results provided in this submission supports that the Subject Device performs as well as the Predicate Devices for its intended use."
The closest to "acceptance criteria" are the standards used for testing:
- Fatigue Performance: Modified version of ISO 14801 (in saline solution) to reflect clinical loading. The specific acceptance criteria (e.g., minimum fatigue limit, number of cycles) are not detailed. It only states that the fatigue limit was determined.
- Biocompatibility: ISO 10993-1:2018. Acceptance is implied by the statement "The results demonstrated that biocompatibility testing... is appropriate to assess the biological safety of the subject device." and "The results demonstrate the biocompatibility of the subject device."
- Sterilization: ISO 17665-1, ISO 17665-2, ANSI/AAMI ST79, and ANSI/AAMI TIR 12. Acceptance is implied by "Validation for the cleaning and sterilization... was conducted."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: Not explicitly stated for performance testing.
- Data Provenance: Not specified. It's likely from testing conducted by the manufacturer, Nobel Biocare AB (Sweden), or a contracted lab. The document does not specify if the data is retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable as the evaluation did not involve human interpretation or a "ground truth" derived from expert consensus in the sense of image analysis or diagnostic accuracy. The performance testing is engineering/laboratory-based.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable for the type of performance testing described. Adjudication methods are typically used in clinical studies or expert reviews.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable as the device is a dental implant bridge, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This information is not applicable as the device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the performance testing is based on:
- Engineering standards: ISO 14801 for fatigue.
- Biocompatibility standards: ISO 10993-1.
- Sterilization standards: ISO 17665-1, ISO 17665-2, ANSI/AAMI ST79, and ANSI/AAMI TIR 12.
- The performance is compared to a predicate device rather than an absolute clinical "ground truth."
8. The sample size for the training set:
This information is not applicable as the device is a physical medical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established:
This information is not applicable as the device is a physical medical device.
Summary of what is available regarding the study:
The study described is a performance testing study aimed at demonstrating that the NobelProcera Zirconia Implant Bridge is substantially equivalent to a predicate device.
- Objective: To show that new technological characteristics (material Nacera Pearl, MUA Plus connection, DLC coating on prosthetic screw, maximum of 10 implants instead of 14) do not raise new concerns and that its performance is comparable to the predicate.
- Key Tests Conducted:
- Fatigue testing: Using a modified version of ISO 14801 (in saline solution) to reflect clinical loading. "Worst-case assessment" was followed.
- Biocompatibility evaluation: According to ISO 10993-1:2018, specifically cytotoxicity and chemical characterization using GC-MS analysis.
- Sterilization validation: According to ISO 17665-1, ISO 17665-2, ANSI/AAMI ST79, and ANSI/AAMI TIR 12.
- Comparison Basis: The subject device and predicate device were tested under identical conditions. The results "support that the Subject Device performs as well as the Predicate Devices for its intended use."
- Clinical Data: "No clinical data was used to support the decision of Substantial Equivalence."
In conclusion, the document provides evidence of engineering and biological performance testing against relevant standards and in comparison to a predicate device, rather than a study with explicit acceptance criteria for diagnostic accuracy or human performance improvement.
Ask a specific question about this device
(29 days)
DTX Studio Clinic is a software program for the acquisition, management, transfer and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental restorative solutions. It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations.
DTX Studio Clinic is a software interface for dental/medical practitioners used to analyze 2D and 3D imaging data, in a timely fashion, for the treatment of dental, craniomaxillofacial and related conditions. DTX Studio Clinic displays and processes imaging data from different devices (i.e. Intraoral and extraoral X-rays, (CB)CT scanners, intraoral scanners, intraoral and extraoral cameras).
Here's a breakdown of the acceptance criteria and study information for the DTX Studio Clinic device, based on the provided text:
Important Note: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, not necessarily a comprehensive clinical study report. Therefore, some information requested (like specific sample sizes for test sets, the number and qualifications of experts for ground truth, adjudication methods, MRMC study effect sizes, and detailed information about training sets) is not explicitly stated in this document. The focus here is on software validation and verification.
Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical "acceptance criteria" in the format of a table with specific metrics (e.g., sensitivity, specificity, accuracy thresholds). Instead, the acceptance is based on demonstrating that the DTX Studio Clinic software performs its intended functions reliably and safely, analogous to the predicate and reference devices, as verified through software validation and engineering testing.
The "reported device performance" is primarily described through the software's functionality and its successful verification and validation.
| Feature/Criterion | Reported Device Performance (DTX Studio Clinic) | Comments (Based on 510(k) Summary) |
|---|---|---|
| Clinical Use | Supports diagnostic and treatment planning for craniomaxillofacial anatomical area. | "Primarily the same" as the predicate device CliniView (K162799). Differences in wording do not alter therapeutic use. |
| Image Data Import & Acquisition | Acquires/imports DICOM, 2D/3D images (CBCT, OPG/panorex, intra-oral X-ray, cephalometric, clinical pictures). Also imports STL, NXA, PLY files from intraoral/optical scanners. Directly acquires images from supported modalities or allows manual import. Imports from 3rd party PMS systems via VDDS or OPP protocol. | Similar to CliniView, with additional capabilities (STL, NXA, PLY, broader PMS integration). Subject device does not control imaging modalities directly for acquisition settings, distinguishing it from CliniView. |
| Data Visualization & Management | Displays and enhances digital images. Provides image filters, annotations, distance/angular measurements, volume and surface area measurements (for segmentation). Stores data locally or in DTX Studio Core database. Comparison of 3D images and 2D intraoral images in the same workspace. | Core functionality is similar to CliniView. Enhanced features include volume/surface area measurements and comparison of different image types within the same workspace. |
| Airway Volume Segmentation | Allows volume segmentation of indicated airway, volume measurements, and constriction point determinations. | Similar to reference device DentiqAir (K183676), but specifically limited to airway (unlike DentiqAir's broader anatomical segmentation). |
| Automatic Image Sorting (IOR) | Algorithm for automatic sorting of acquired or imported intra-oral X-ray images to an FMX template. Detects tooth numbers (FDI or Universal). | This is a workflow improvement feature, not for diagnosis or image enhancement. |
| Intraoral Scanner Module (ioscan) | Dedicated intraoral scanner workspace for acquisition of 3D intraoral models (STL, NXA, PLY). Supports dental optical impression systems. | Classified as NOF, 872.3661 (510(k) exempt). Does not impact substantial equivalence. |
| Alignment of Intra-oral/Cast Scans with (CB)CT Data | Imports 3D intraoral models or dental cast scans (STL/PLY) and aligns them with imported CB(CT) data for accurate implant planning. | Similar to reference device DTX Studio Implant (K163122). |
| Implant Planning | Functionality for implant planning treatment. Adds dental implant shapes to imported 3D data, allowing user definition of position, orientation, type, and dimensions. | Similar to reference device DTX Studio Implant (K163122), which also adds implants and computes surgical templates. |
| Virtual Tooth Setup | Calculates and visualizes a 3D tooth shape for a missing tooth position based on indicated landmarks and loaded intra-oral scan. Used for prosthetic visualization and input for implant position. | A new feature not explicitly present in the predicate devices but supported by the overall diagnostic and planning workflow. |
| Software Validation & Verification | Designed and manufactured under Quality System Regulations (21 CFR § 820, ISO 13485:2016). Conforms to EN IEC 62304:2006. Risk management (ISO 14971:2012), verification testing performed. Software V&V testing conducted as per FDA guidance for "Moderate Level of Concern." Requirements for features have been met. | Demonstrated through extensive software engineering and quality assurance processes, not clinical performance metrics. |
Study Information
-
Sample sizes used for the test set and the data provenance:
- Not explicitly stated in the provided text. The document mentions "verification testing" and "validation testing" but does not detail the specific sample sizes of images or patient cases used for these tests.
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It focuses on the software's functionality and its comparison to predicate devices, rather than the performance on specific clinical datasets.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated in the provided text. The 510(k) summary primarily addresses software functionality verification and validation, not a diagnostic accuracy study involving expert ground truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not explicitly stated in the provided text.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done or reported. The document explicitly states: "No clinical data was used to support the decision of substantial equivalence." This type of study would involve clinical data and human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, in spirit. The software validation and verification described are for the algorithm and software functionalities operating independently. While the device does not make autonomous diagnoses (it "supports the diagnostic process and treatment planning"), its individual features (like airway segmentation, image sorting, virtual tooth setup) are tested in a standalone manner in terms of their computational correctness and adherence to specifications. However, this is distinct from standalone clinical performance (e.g., an AI algorithm making a diagnosis without human input). The document focuses on the technical performance of the software.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the software verification and validation, the implicit "ground truth" would be the software's functional specifications and requirements. For features like measurements or segmentation, this would likely involve mathematical correctness checks or comparison to pre-defined anatomical models or manually delineated reference segmentations. It is not based on expert consensus, pathology, or outcomes data in a clinical diagnostic sense, as no clinical data was used for substantial equivalence.
-
The sample size for the training set:
- Not explicitly stated in the provided text. The document describes a medical device software for image management and analysis, not a machine learning model that typically requires a large 'training set' in the deep learning sense. If any features (like the automatic image sorting or virtual tooth setup) utilize machine learning, the details of their training (including sample size) are not provided in this 510(k) summary.
-
How the ground truth for the training set was established:
- Not explicitly stated in the provided text. As mentioned above, details about training sets are absent. If machine learning is involved in certain features, the ground truth would typically be established by expert annotation or curated datasets, but this is not detailed here.
Ask a specific question about this device