Search Filters

Search Results

Found 68 results

510(k) Data Aggregation

    K Number
    K232542

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-16

    (268 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wedge XL Delivery Catheter is intended to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature.

    Device Description

    The Wedge XL Delivery Catheter is a single lumen catheter designed to be introduced over a steerable quidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Radiopaque markers at the distal end facilitate fluoroscopic visualization. A larger diameter distal segment helps provide stability for navigation. The distal 110 cm outer surface of the Wedge XL Delivery Catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the Wedge XL Delivery Catheter hub is used for the attachment of accessories.

    AI/ML Overview

    This document describes the regulatory approval of the MicroVention, Inc. Wedge XL Delivery Catheter (K232542) and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and performance data in the typical format of a clinical trial for an AI/software device.

    The information provided confirms that the device is a percutaneous catheter, not an AI/software as a medical device (SaMD). Therefore, the questions regarding acceptance criteria and performance studies specifically for AI/Software (e.g., test set, ground truth, expert adjudication, MRMC studies) are not directly applicable in the format requested.

    However, I can extract the information relevant to the device's performance testing and general acceptance:

    The submission demonstrates the device meets acceptance criteria through a series of bench testing and an animal study, alongside biocompatibility evaluation and sterilization information. The overall acceptance criterion is substantial equivalence to the predicate device (Wedge Microcatheter K172014) and a reference device (Delivery Catheter of the Route 92 Medical 088 Access System K200121).

    Here's a breakdown of the provided information within the context of a medical device submission:


    Acceptance Criteria and Reported Device Performance (Table Format - Adapted for a Non-AI Device):

    Acceptance Criteria CategorySpecific Test/EvaluationAcceptance Criteria Met / Reported Performance
    Bench PerformancePhysical AttributesAll tests passed pre-determined acceptance criteria.
    Surface ContaminationAll tests passed pre-determined acceptance criteria.
    Coating Lubricity and DurabilityAll tests passed pre-determined acceptance criteria.
    Simulated Use (Preparation/Ease of Assembly, Introducer Sheath Insertion, Tracking With/Without Guidewire, Delivery Catheter/Guidewire Lock Up, Catheter Ovalization, Lubricity/Durability of Hydrophilic Coating, Guide Catheter Tracking over Wedge XL Delivery Catheter, Particle Testing)All tests passed pre-determined acceptance criteria.
    Dynamic Burst PressureAll tests passed pre-determined acceptance criteria.
    Freedom from Air LeakageAll tests passed pre-determined acceptance criteria.
    Freedom from Liquid LeakageAll tests passed pre-determined acceptance criteria.
    Static Burst PressureAll tests passed pre-determined acceptance criteria.
    Force at BreakAll tests passed pre-determined acceptance criteria.
    Catheter Flexural FatigueAll tests passed pre-determined acceptance criteria.
    Catheter Particle TestingAll tests passed pre-determined acceptance criteria.
    Kink ResistanceAll tests passed pre-determined acceptance criteria.
    Torque StrengthAll tests passed pre-determined acceptance criteria.
    Luer-Connector Dimensions and Performance (ISO 80369-7)All tests passed pre-determined acceptance criteria.
    RadiopacityYes (Previously conducted on an equivalent test article).
    Corrosion ResistanceYes (Previously conducted on an equivalent test article).
    BiocompatibilityCytotoxicity (ISO MEM Elution Test)Non-cytotoxic (slight reactivity, grade 1).
    Sensitization (ISO Guinea Pig Maximization Test)Non-sensitizer (no sensitization observed).
    Irritation Reactivity (ISO Intracutaneous Reactivity Test)Non-irritant (no irritation observed).
    Material Mediated Pyrogenicity (ISO/USP)Non-pyrogenic (no temperature rise).
    Systemic Toxicity (ISO Acute Systemic Toxicity Test)No systemic toxicity (no weight loss, mortality, or evidence of systemic toxicity).
    Hemocompatibility (Hemolysis Assay - Extract and Direct Methods, ASTM)Non-hemolytic (1.2% hemolysis in direct, 0.0% in indirect).
    Hemocompatibility (Complement Activation Assay, ISO)Non-activator (no activation of SC5b-9).
    Hemocompatibility (Partial Thromboplastin Time (PTT) Test, ASTM)Non-activator (no significant difference from control).
    Hemocompatibility (Heparinized Blood Platelet and Leukocyte Count Assay, ASTM)Non-activator (no significant difference from control).
    Hemocompatibility (In-Vitro Blood Loop Assay, ISO)Thromboresistant (minimal thrombus formation observed).
    Animal StudySafety & Performance (Porcine model, tracking with SOFIA 88 Catheter vs. predicate with SOFIA 6F Catheter)Performed equally; no remarkable gross findings; comparable histological impact; results consistent with routine catheterization and did not raise safety concerns. Deemed equivalent.

    For the points specifically pertaining to AI/ML device studies, the following is direct or inferred from the document:

    1. Sample sizes used for the test set and the data provenance:

      • This is not an AI/ML device. For bench testing, samples are typically physical units. For the animal study, it mentions a "porcine model," implying a number of animals were used, but the specific count is not provided. The study followed GLP (Good Laboratory Practice) Regulation (21 CFR Part 58), indicating a controlled, prospective study. The origin is a "porcine model," which is not human data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable as this is not an AI/ML device. Ground truth for device performance would be established through physical measurements, chemical analyses, and histopathological evaluation by trained personnel (e.g., lab technicians, pathologists) in the animal study, but the specific number and qualifications are not detailed beyond "Good Laboratory Practice (GLP)".
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not an AI/ML device.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable as this is not an an AI-driven or diagnostic device that involves human reader interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

      • For bench testing: Physical measurements, performance metrics based on engineering and material science standards.
      • For biocompatibility: Laboratory assays with established controls and reference standards.
      • For the animal study: Pathology (histological impact on arterial sections, postmortem examination for gross findings), and outcomes data (safety and performance of catheter tracking, vascular response to device use).
    7. The sample size for the training set:

      • Not applicable as this is not an AI/ML device that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230775

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-14

    (177 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA EX Intracranial Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA EX Intracranial Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA EX Intracranial Support Catheter is not intended for use in coronary arteries.

    Device Description

    The SOFIA™ EX Intracranial Support Catheter is a single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal segment is designed to facilitate vessel selection with 55-65cm of distal shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. The catheter is placed in a dispenser tube (HDPE) and is placed on a packaging card (polyethylene) that is provided in a sterile barrier Tyvek pouch and placed in a carton box.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the MicroVention, Inc. SOFIA EX Intracranial Support Catheter:

    Summary of Acceptance Criteria and Device Performance for SOFIA EX Intracranial Support Catheter (K230775)

    The provided document describes the SOFIA EX Intracranial Support Catheter as being substantially equivalent to its predicate device (SOFIA EX Intracranial Support Catheter, K182602). This means that the acceptance criteria and performance are largely benchmarked against the predicate device, implying similar performance expectations.

    The document focuses on bench testing to demonstrate this equivalence.


    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied by equivalence to predicate)Reported Device Performance
    Simulated Use and Physician Simulated UseSuccessful tracking and adequate hydrophilic coating lubricity in a tortuous anatomical model. Performance comparable to predicate device.Device met acceptance criteria.
    Kink ResistanceAbility to withstand bending experienced in tortuous anatomy without kinking or failure. Performance comparable to predicate device.Device met acceptance criteria.
    Particulate TestingNumber and size of particulates generated during simulated use should be comparable to the predicate device.Number of particulates generated are comparable to the predicate device.
    BiocompatibilityShould be biocompatible for limited exposure (≤ 24 hours), externally communicating device with circulating blood contact, per ISO 10993-1, with identical materials, processing, and sterilization as the predicate device.No additional biocompatibility testing required as materials, processing, and sterilization are identical to the predicate which previously met these criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific bench test (e.g., number of catheters tested for simulated use, kink resistance, or particulate testing). The data provenance is bench testing, meaning it was conducted in a laboratory setting, not on patient data. No country of origin for the data is specified, but it can be assumed to be related to MicroVention, Inc.'s operations, likely within the US, given the FDA submission. The nature of the testing is prospective in the sense that new samples of the subject device were manufactured and tested according to a predefined protocol.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not Applicable (N/A) for this submission. The ground truth for these bench tests is based on objective measurements and predefined engineering specifications, not expert interpretation of diagnostic data.

    4. Adjudication Method for the Test Set

    Not Applicable (N/A). Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving human interpretation or agreement on outcomes. Bench tests rely on precise measurements and adherence to specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "No animal or clinical studies were conducted because bench testing was determined sufficient for verification and validation purposes." Therefore, no MRMC study was performed, and no human reader improvement effect size is available.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

    Not Applicable (N/A). This device is a medical catheter, a physical product, not an AI algorithm. Therefore, the concept of "standalone algorithm performance" does not apply.

    7. Type of Ground Truth Used

    The ground truth for the bench tests is based on objective engineering measurements and predefined specifications. For example:

    • Simulated Use/Lubricity: Successful navigation through a tortuous model, visual inspection for damage, and measurement of friction/lubricity metrics.
    • Kink Resistance: Mechanical testing to observe and quantify resistance to kinking under specified bending conditions.
    • Particulate Testing: Quantitative measurement of particle count and size using standard laboratory methods.
    • Biocompatibility: Demonstrated through adherence to ISO 10993-1 standards, which were previously met by the predicate device (same materials, processing, sterilization).

    8. Sample Size for the Training Set

    Not Applicable (N/A). This is a medical device, not an AI/ML algorithm. There is no "training set" in the context of this submission. The device's design is based on engineering principles and validated through bench testing, not machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable (N/A). As there is no training set for an AI/ML algorithm, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222694

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-27

    (233 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EmPro™ EPS/Nanoparasol™ EPS is indicated for use as a guidewire to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the filter placement should be between 3.0 and 6.5 mm.

    Device Description

    MicroVention's Embolic Protection System (EPS) is marketed under two names: EmPro™ Embolic Protection System and Nanoparasol™ Embolic Protection System. The Embolic Protection System (EPS) is designed to capture and remove dislodged debris during a carotid interventional procedure. It consists of three basic components and additional accessories:

    1. An Embolic Protection Device (EPD) consisting of a nitinol braided mesh filter with an atraumatic distal tip built on an integrated .014" PTFE coated stainless steel capture delivery wire.
    2. A 3.5F delivery catheter with 150 cm length.
    3. A 5F retrieval catheter with 150 cm working length. Accessories include a wire introducer, EPD loading cover, sheath introducer and a torque device. Catheters are provided in two separate dispenser coils.
    AI/ML Overview

    The provided text describes the acceptance criteria and the study proving the device meets these criteria for the EmPro™ EPS/Nanoparasol™ EPS. This device is an embolic protection system indicated for use as a guidewire to contain and remove embolic material during angioplasty and stenting procedures in carotid arteries.

    Here's the breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (EmPro™/Nanoparasol™ EPS)
    Primary Endpoint: Major Adverse Event (MAE) rate (Death, Stroke, or MI within 30 days of index procedure + ipsilateral stroke 31-365 days after procedure) upper limit of 95% exact binomial CI below PG of 13.9%MAE rate: 5.9% (95% exact binomial CI: 3.89, 10.69); p=0.0014. Upper limit of 95% CI was 9.22%, which was below the PG of 13.9%.
    Embolic Protection System (EPS) technical success98.8% (253/256) of subjects.
    EPS successfully inserted99.6% (255/256)
    EPS successfully deployed in subject98.8% (253/256)
    EPS successfully retrieved99.6% (255/256)
    Vessel dissection at EPS filter site0
    Neurological assessments (in animal study)All animals met the acceptance criteria for neurological assessments at both timepoints (30 and 180 days post-operatively).
    Histological analysis of local tissues (in animal study)Showed negligible vessel injury, inflammation, and neointimal response where the EPD was deployed.
    Performance scores (in animal study)Met or exceeded acceptance criteria in both intervention groups.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Clinical Study (Test Set): 256 patients (n=256) in the Intent-To-Treat (ITT) population.
    • Data Provenance: The study was a "multicenter, single-arm, interventional study" called the CONFIDENCE study (IDE G140249). While the specific countries are not mentioned, FDA submissions typically involve studies conducted in the US or under equivalent international standards. The study design is prospective.
    • Sample Size for Animal Study: 6 animals (porcine model).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the clinical study's endpoint definitions (MAE, technical success). Clinical trials typically involve a clinical events committee (CEC) composed of expert physicians to independently adjudicate events like death, stroke, and MI, but this level of detail is not provided in this regulatory summary.

    For the animal study, the document mentions "histological analysis" and "neurological assessments" were performed, implying expert evaluation, but the number and specific qualifications of the experts are not detailed.

    4. Adjudication method for the test set

    The document does not explicitly detail an adjudication method (e.g., 2+1, 3+1) for the clinical study's primary and secondary endpoints. For a multi-center clinical study with composite endpoints like MAE, it is standard practice to have a Clinical Events Committee (CEC) adjudicate events, often with multiple readers and a pre-defined adjudication process, but the specifics are not provided here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is an embolic protection system, not an AI-powered diagnostic tool, so such a study would not be applicable. The performance evaluated relates to the mechanical and clinical outcomes of the device itself and its ability to capture emboli, not human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical device (an embolic protection system), not an algorithm or AI system. Its performance is evaluated in vivo (animal and human clinical studies) and in vitro (bench testing), not as a standalone algorithm.

    7. The type of ground truth used

    • Clinical Study (CONFIDENCE study): The ground truth for the primary endpoint (MAE) was based on clinical outcomes: death, stroke, and myocardial infarction (MI), confirmed by medical records and follow-up. For secondary endpoints like technical success (e.g., EPS successfully inserted, deployed, retrieved), the ground truth would be based on procedural observations and documentation. This represents outcomes data and procedural success data.
    • Animal Study: Ground truth was established through direct observation during intervention (e.g., tracking, deployment, retrieval), post-intervention assessment (thrombus formation, particulate capture, device damage, neurological dysfunction), and post-explantation histopathology/histology to evaluate tissue response.

    8. The sample size for the training set

    This document does not describe a "training set" in the context of an AI/ML model. The studies described are for the validation and performance evaluation of a physical medical device. Therefore, a training set sample size, as understood in AI/ML development, is not applicable or provided.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" for an AI/ML model, this question is not applicable. The device's design and engineering would have been informed by a vast amount of existing medical knowledge, material science, and prior device development, rather than a specific "training set" with established ground truth in the AI/ML sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222115

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-10

    (176 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ISAAC™ Neurovascular Navigation Catheter is indicated for use in facilitating advancement of catheters through the neuro and peripheral vasculature and introduction of diagnostic agents. The ISAAC™ Neurovascular Navigation Catheter is not intended for use in the coronary vasculature.

    Device Description

    The ISAAC™ Neurovascular Navigation Catheter) is a braid-reinforced variable stiffness catheter with a pre-shaped distal segment. The distal end of the catheter is coated with a hydrophilic coating around the curve of the pre-shaped section of the Simmons (SIM) configuration. It is a single lumen catheter with a radiopaque distal coiled section and a Luer hub on the proximal end. The ISAAC Catheter is sterile, non-pyrogenic and intended for single use only.

    AI/ML Overview

    The FDA 510(k) summary for the MicroVention, Inc. ISAAC Neurovascular Navigation Catheter (K222115) indicates that a range of performance testing was conducted to demonstrate the device's substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria / Performance ClaimReported Device Performance
    Physical AttributesMeasurement of usable length, proximal/distal outer diameters, distal length, and inner diameters.Pass
    Tip FlexibilityDevice shall have less force to bend at the distal tip than the comparator reference device.Pass
    Simulated UsePerformance rated during simulated-use testing in benchtop vessel model.Pass
    Radio DetectabilityCatheter must be visible under X-ray fluoroscopy.Pass
    Kink ResistanceKink resistance measured by subjecting the device to bending in simulated tortuous anatomy.Pass
    Static BurstFluid injected into the lumen at increasing pressure until catheter burst while distal tip occluded.Pass
    Liquid LeakagePressure maintained for 30 seconds with occluded distal tip; device inspected for leakage per ISO 10555-1.Pass
    Leakage and Damage Under High Static Pressure ConditionsDyed fluid injected until rated burst pressure reached and held for 10 seconds while distal tip occluded.Pass
    Air LeakageVacuum applied; observed for air bubbles per ISO 80369-7 and ISO 10555-1.Pass
    Dynamic BurstCatheter injected with fluid at a set pressure and inspected for damage.Pass
    Force at Break (Hub/Distal)Hub/distal sections secured into tensile test machine; pulled until catheter broke, pull force recorded.Pass
    Particulate TestingCatheter underwent simulated-use testing in a benchtop model and evaluated for particulates.Pass
    Surface ContaminationDevice must be free from visible surface defects.Pass
    Corrosion ResistanceCatheter tested per ISO 10555-1 and ISO 11070.Pass
    Coating Durability and LubricityDevice secured to tensile machine, hydrated, force to slide through clamp recorded (20 cycles for lubricity, 100 cycles for durability).Pass
    Catheter Flexural FatigueTensile strength and pressure characteristics measured per ISO 10555-1.Pass
    Hub and Luer ConnectorLuer connector tested to dimensional and performance requirements per ISO 80369-7.Pass
    StiffnessCatheter stiffness profile compared to the reference device.Pass
    Torque StrengthEvaluated by measuring number of catheter rotations until failure after tracking through a tortuous anatomical model.Pass
    Biocompatibility (Cytotoxicity)Non-cytotoxicNon-cytotoxic
    Biocompatibility (Irritation Reactivity)Non-irritantNon-irritant
    Biocompatibility (Maximization - Sensitization)Non-sensitizingNon-sensitizing
    Biocompatibility (Systemic Toxicity)Non-acute systemically toxicNon-acute systemically toxic
    Biocompatibility (Pyrogenicity)Non-pyrogenicNon-pyrogenic
    Biocompatibility (Hemocompatibility In-Vitro Blood Loop Assay)Thrombogenic risk potential similar to the predicateSimilar to predicate
    Biocompatibility (Hemocompatibility Hemolysis Assay)Non-hemolyticNon-hemolytic
    Biocompatibility (Hemocompatibility Complement Activation Assay)Non-activator of complement systemNon-activator of complement system
    Biocompatibility (Hemocompatibility Partial Thromboplastin Time (PTT) Assay)No effect on the PTTNo effect on PTT
    Biocompatibility (Hemocompatibility Heparinized Blood Platelet and Leukocyte Count Assay)PassPass
    Animal Study (Tracking Performance, Support, Safety)Perform comparably to the control article (Chaperon Guiding Catheter) in terms of tracking, support, and safety.Performed comparably
    Animal Study (Device-associated Complications)No significant device-associated complications (dissection, perforation, embolic debris, thrombus, hemorrhage, ischemia, necrosis, fibrin deposition, IEL rupture, EEL rupture, mineralization, neointimal maturation, medial/adventitial injury/fibrosis).No significant complications

    2. Sample Size Used for the Test Set and Data Provenance:

    • Benchtop and Biocompatibility Testing: The specific sample sizes for each individual benchtop and biocompatibility test are not explicitly detailed in this summary. However, these tests are generally conducted on a statistically significant number of device units or material samples to ensure representativeness and reproducibility.
    • Animal Study: The animal study was conducted using a "porcine model." The exact number of animals involved is not specified, but it's referred to as an "acute animal testing" in a "porcine model."
    • Data Provenance:
      • Benchtop Testing: Likely laboratory-generated data from MicroVention, Inc., performed under controlled conditions.
      • Biocompatibility Testing: Performed on material extracts or the device itself, likely by specialized laboratories following ISO standards.
      • Animal Study: Conducted in accordance with FDA Good Laboratory Practice (GLP) Regulation (21 CFR Part 58), indicating a prospective study specifically designed to assess the device. The provenance is internal to the controlled study environment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Benchtop and Biocompatibility Testing: The summary does not specify the use of "experts" to establish a ground truth in the traditional sense (e.g., for image interpretation). Instead, these tests rely on objective, measurable criteria defined by established international standards (e.g., ISO, ASTM). The "ground truth" would be the direct measurement or observation of the device's physical, mechanical, and biological properties against predefined limits or comparison with a predicate.
    • Animal Study: The evaluation of the animal study results regarding complications and performance comparability would involve veterinarians and potentially pathologists who are experts in animal physiology and disease interpretation. Their specific number and qualifications are not detailed in this summary.

    4. Adjudication Method for the Test Set Without Explanation:

    • Benchtop and Biocompatibility Testing: An "adjudication method" in the sense of resolving discrepancies between multiple expert interpretations is generally not applicable to the objective measurements performed in these tests. Results are typically pass/fail based on predetermined specifications or direct comparison to controls/predicates.
    • Animal Study: While not explicitly stated, observations in animal studies, especially pathological findings, often involve review by multiple experts (e.g., veterinary pathologists). If discrepancies occur, an adjudication process involving consensus or a tie-breaking expert would likely be in place, though this is not detailed in the provided text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    • Not Applicable. The ISAAC Neurovascular Navigation Catheter is a physical medical device (catheter) and not an AI-powered diagnostic or assistive tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    • Not Applicable. As mentioned above, this is a physical medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    • Benchtop Testing: Objective physical and mechanical measurements, adherence to industry standards, and comparison to predicate device characteristics.
    • Biocompatibility Testing: Results obtained from standardized biological assays (e.g., cell cultures, animal models for irritation/sensitization) providing objective data on biological reactions, evaluated against established acceptance criteria (e.g., "non-cytotoxic," "non-irritant").
    • Animal Study: Direct observation of device performance (tracking, support) and histological/pathological assessment for complications (e.g., dissection, thrombus, tissue injury) in a living porcine model, compared against the control article.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is a physical medical device, and the concept of a "training set" for an AI algorithm is not relevant here. The device design and manufacturing process would involve extensive engineering development and iterative testing (developmental testing) which is distinct from an AI training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. As there is no "training set" in the context of an AI algorithm for this device, a method for establishing its ground truth is not relevant. The device development relies on engineering principles, material science, and performance testing against established standards and predicate device characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K214024

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-09-12

    (264 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.

    Device Description

    The SOFIA 88 Catheter is a non-tapered, single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal seqment is designed to facilitate vessel selection with 60 cm of distal-shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (SOFIA 88 Catheter), demonstrating its substantial equivalence to a predicate device. This type of regulatory submission primarily relies on performance (benchtop) and pre-clinical (animal) testing rather than clinical studies with human subjects or AI algorithm performance.

    Therefore, many of the requested criteria (e.g., acceptance criteria for AI performance, sample size for test sets (human data), number of experts for ground truth, MRMC studies, standalone AI performance, training set details) are not applicable to this specific document as it pertains to a physical medical device (catheter) and not an AI/ML powered device.

    However, I can extract the relevant information regarding the performance and equivalency testing conducted for the SOFIA 88 Catheter.


    Device: SOFIA 88 Catheter

    Type of Device: Percutaneous Catheter (physical medical device)


    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally implied by the "Pass" result for each test, indicating that the device met the pre-defined standards for each physical and functional characteristic. The document doesn't explicitly state quantitative acceptance limits for each test but rather confirms successful completion.

    Test TitleTest Method SummaryReported Device Performance
    Dimensional VerificationMeasured catheter length, proximal and distal outer diameters, and distal inner diameter.Pass
    Coating Lubricity and DurabilityEvaluated hydrophilic coating for frictional force and durability.Pass
    Simulated UseEvaluated performance under simulated use in a tortuous anatomical model to assess preparation, introduction, tracking, and support of the device.Pass
    Dynamic and Static Burst PressureSubjected device to rated burst pressures to test catheter integrity.Pass
    Air LeakageTested device for air leakage requirements.Pass
    Liquid LeakageTested device for liquid leakage requirements.Pass
    Tensile StrengthTensile tested device to failure, recorded the force at breakage.Pass
    Flexural FatigueSubjected device to flexural fatigue from repeated bending during simulated use testing and from repeated hoop stress on the catheter from pressure testing.Pass
    Particulate TestingMeasured number and size of particulates generated during simulated use in a tortuous anatomical model and compared results with the predicate device.Pass
    Kink ResistanceEvaluated device for kink resistance after simulated use testing.Pass
    Torque StrengthEvaluated device for torque strength by measuring the number of catheter rotations until failure after tracking through a tortuous anatomical model.Pass
    Radio DetectabilityEvaluated device visibility under fluoroscopy.Pass
    Corrosion ResistanceEvaluated device corrosion after immersion in saline.Pass
    Small-bore ConnectorEvaluated luer connector for dimensional and performance requirements per reference standard.Pass
    Stiffness ProfileCompared catheter stiffness profile along proximal and distal sections to the reference device.Pass
    Biocompatibility Tests
    CytotoxicityL-929 mouse fibroblast cells (MEM) with 6.0 cm²/mL at 37°C for 24 hrs.Non-cytotoxic
    Irritation ReactivityNormal saline and sesame seed oil (SSO) on New Zealand White Rabbits with 6.0 cm²/mL at 50°C for 72 hrs.Non-irritant
    MaximizationNormal saline and sesame seed oil (SSO) on Guinea Pigs with 6.0 cm²/mL at 50°C for 72 hrs.Non-sensitizing
    Systemic ToxicityNormal saline and sesame seed oil (SSO) on Albino outbred strain (ND4) mice with 6.0 cm²/mL at 50°C for 72 hrs.Non-acute systemically toxic
    PyrogenicityNormal saline on New Zealand White Rabbits with 6.0 cm²/mL at 50°C for 72 hrs.Non-pyrogenic
    Hemocompatibility (Blood Loop)Loop system circulated with freshly drawn sheep blood; test article exposed to circulating blood at 37°C for 4 hrs.Thromboresistant
    Hemolysis AssayPBS – Phosphate Buffered Saline / Blood from Three New Zealand White Rabbits; 6.0 cm²/mL at 50°C for 72 hrs.Non-hemolytic
    Complement Activation AssayNormal Human Serum (NHS); 6.0 cm²/mL at 37°C for 60 min.Non-activator of complement system
    Partial Thromboplastin Time (PTT) AssayHuman Plasma / Freshly Drawn Human Plasma; 6.0 cm²/mL at 37°C for 15 min.No effect on the PTT
    Heparinized Blood Platelet and Leukocyte Count AssayHuman Blood / Freshly Drawn Human Blood; 12 cm²/mL at 37°C for 60 min.Pass
    Animal Study
    Acute Animal TestingComparison of SOFIA 88 Catheter to SOFIA 6F DAC for catheter tracking and tip stability in a porcine model. GLP-compliant. Examined for dissection/perforation, vasospasm, hemorrhage, necrosis, edema.Performed equally (SOFIA 88 Catheter and SOFIA 6F DAC). Found comparable. No issues noted.

    2. Sample size used for the test set and the data provenance

    • Benchtop Testing: While specific sample sizes for each benchtop test are not provided in this summary, these tests generally involve a statistically significant number of device units (e.g., n=3, 5, or more depending on the test and standard requirements) to ensure reproducibility and reliability. The provenance is internal laboratory testing by MicroVention, Inc. and its contracted labs.
    • Biocompatibility Testing: Sample sizes are mentioned for some tests (e.g., New Zealand White Rabbits for irritation/pyrogenicity, Guinea Pigs for maximization, ND4 mice for systemic toxicity, sheep blood for blood loop assay, three rabbits for hemolysis, human serum/plasma for others). The provenance is laboratory testing conducted according to ISO and USP standards.
    • Animal Study: The document refers to "a porcine model" without specifying the exact number of animals. It's an acute animal study. The provenance is internal or contracted laboratory animal testing (conducted in accordance with FDA GLP Regulation 21 CFR Part 58).

    All data provenance is internal R&D and pre-clinical testing, not human clinical retrospective or prospective data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the ground truth for a physical medical device like a catheter is established through objective physical and chemical measurements (benchtop tests) and observed physiological responses in animal models, not through expert human interpretation of medical images or data requiring "ground truth" establishment in the context of an AI algorithm.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no human interpretation or subjective judgment that would require an adjudication method in the context of this device's testing. Measurements are objective and pass/fail criteria are pre-defined.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a physical catheter, not an AI application, so no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is a physical device, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established via objective engineering and material science measurements (e.g., dimensions, pressure, force), validated laboratory biological assays (e.g., cytotoxicity, hemolysis), and observed physiological responses and pathological findings in an animal model. It is not derived from expert consensus, pathology, or outcomes data in the sense of clinical diagnostic accuracy.

    8. The sample size for the training set

    This is not applicable as this is a physical medical device, not an AI/ML powered device that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable for the reason stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193607

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-21

    (208 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BOBBY Balloon Guide Catheter is intended: For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The BOBBY Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter with an external hydrophilic coating. The BOBBY Balloon Guide Catheter incorporates a compliant balloon, radiopaque markers, and a bifurcated luer hub on the proximal end. The BOBBY Balloon Guide Catheter has an inner lumen through which a guidewire and catheter can be inserted, and a co-axial outer lumen that is used to inflate and deflate the balloon with a syringe filled with contrast media. A bifurcated luer hub is attached to the proximal end of the balloon guide catheter to provide access to both the inner and outer lumens. In addition, a hydrophilic coating is applied to the distal end of the balloon guide catheter to provide a lubricious outer surface for catheter advancement in the vasculature. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. The balloon incorporates a distal air-purging system to purge air from the inflation lumen prior to use. The balloon catheter also incorporates radiopaque markers to facilitate fluoroscopic visualization and indication of the balloon position.

    AI/ML Overview

    The provided text describes the acceptance criteria and study proving the device meets those criteria for the BOBBY Balloon Guide Catheter. It's important to note that this is a medical device (catheter), not an AI/algorithm-based device, so the usual metrics for AI performance (like sensitivity, specificity, MRMC studies, training/test sets for AI) are not applicable here.

    The "acceptance criteria" for this device are established through various performance and biocompatibility tests designed to ensure its safety and effectiveness for its intended use. The "study that proves the device meets the acceptance criteria" refers to the results of these performance and biocompatibility tests.

    Here's the breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comprehensive table under "Performance Testing Summary" and "Biocompatibility Testing Summary." The "Acceptance Criteria" for each test are implied by the "Reference Standard" and the "Results" section confirming that "All samples met the pre-determined acceptance criteria" or similar statements of compliance.

    Key:

    • Acceptance Criteria (Implied): Meeting the requirements of the stated reference standard or pre-determined specifications.
    • Reported Device Performance: "Pass" indicating that all samples tested met the criteria.
    Performance Test Study NameDescriptionReference Standard / Implied Acceptance CriteriaReported Device Performance (Results)
    Conditioning, Distribution, and Shelf Life Aging VerificationTo demonstrate packaging strength and integrity following accelerated aging to 1-year shelf life equivalent.ASTM F88 Seal Strength, ASTM F2096 Gross LeaksPass (All samples met pre-determined acceptance criteria)
    Packaging Visual InspectionTo demonstrate product meets packaging visual inspection requirements.N/A (Internal visual inspection requirements implied)Pass (All samples met pre-determined acceptance criteria)
    Visual Surface RequirementsTo demonstrate product satisfies visual surface requirements.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Dimensional/Physical Attributes InspectionTo demonstrate product meets dimensional specifications.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Inflation Volume vs Balloon DiameterTo demonstrate product meets inflation volume vs balloon diameter specifications.In consideration of ISO 10555-4:2013Pass (All samples met pre-determined acceptance criteria)
    Balloon Burst VolumeTo demonstrate Balloon is capable of withstanding an injection volume above the maximum fill volume.In consideration of ISO 10555-4:2013Pass (All samples met pre-determined acceptance criteria)
    Tip StiffnessTo demonstrate stiffness of distal end is similar to other marketed devices.N/A (Comparative to predicate implied)Pass (All samples met pre-determined acceptance criteria)
    Torque TestingTo demonstrate product is capable of 720 degrees of rotation about central lumen axis without failure.N/A (Internal functional requirement implied)Pass (All samples met pre-determined acceptance criteria)
    Force at BreakTo demonstrate product satisfies force at break requirements.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Small Bore Connector Compliance with StandardTo demonstrate product meets requirements.ISO 80369-7 2016Pass (All samples met pre-determined acceptance criteria)
    RadiopacityTo determine radiopaque characteristics.ISO 10555-1:2013, ASTM F640-12Pass (All samples met pre-determined acceptance criteria)
    Particulate, Coating IntegrityTo determine quantity and size of particles generated during simulated use.USP <788> Particulate Matter in InjectionsPass (All samples met pre-determined acceptance criteria)
    Freedom from Liquid LeakageTo demonstrate product meets liquid leakage requirements.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Hub Aspiration Air LeakageTo demonstrate product meets hub aspiration air leakage requirements.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Balloon Fatigue TestTo demonstrate no degradation of Balloon after repetitive inflation cycles.In consideration of ISO 10555-4:2013Pass (All samples met pre-determined acceptance criteria)
    Simulated UseSimulated use under in vitro conditions in a cerebral vascular model.In consideration of ISO 10555-4:2013Pass (All samples met pre-determined acceptance criteria)
    Flexural FatigueTo demonstrate product does not lose structural integrity when used in tortuous path model.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Kink ResistanceTo demonstrate device has similar kink resistance compared to predicate device.N/A (Comparative to predicate implied)Pass (All samples met pre-determined acceptance criteria)
    Lubricity and durability of the hydrophilic coatingTo demonstrate hydrophilic coating is lubricious and durable.N/A (Internal functional requirement implied)Pass (All samples met pre-determined acceptance criteria)
    Guidewire lumen burst pressure (Static, dynamic)To demonstrate device does not burst below rated burst pressure.ISO 10555-1:2013Pass (All samples met pre-determined acceptance criteria)
    Balloon Deflation TimeTo demonstrate device has similar balloon deflation time compared to predicate device.ISO10555-4:2013Pass (All samples met pre-determined acceptance criteria)
    Lumen CollapseTo demonstrate guidewire lumen does not collapse under aspiration.N/A (Internal functional requirement implied)Pass (All samples met pre-determined acceptance criteria)

    Biocompatibility Tests:

    Test NameTest Method / Implied Acceptance CriteriaReported Device Performance (Results)
    CytotoxicityISO 10993-5 (Noncytotoxic)Pass (Noncytotoxic according to predetermined acceptance criteria)
    SensitizationISO 10993-10, Kligman Maximization Test (No sensitization response)Pass (Did not elicit a sensitization response according to predetermined acceptance criteria)
    Intracutaneous IrritationISO 10993-10 (Test requirements for intracutaneous reactivity met)Pass (Test requirements for intracutaneous reactivity were met according to predetermined acceptance criteria)
    Systemic Toxicity: Systemic Injection TestISO 10993-11 (Test requirements for systemic toxicity met)Pass (Test requirements for systemic toxicity were met according to predetermined acceptance criteria)
    Systemic Toxicity: Material Mediated PyrogenicityISO 10993-11, USP NF 36:2018 <151> Pyrogen Test (Non-pyrogenic)Pass (Non-pyrogenic, met the predetermined acceptance criteria)
    Hemocompatibility: HemolysisASTM F756-17, ISO 10993-4 (Non-hemolytic)Pass (Non-hemolytic, met the predetermined acceptance criteria)
    Hemocompatibility: Complement ActivationISO 10993-4, SC5b-9 Complement Activation (Does not activate complement system)Pass (Does not activate the complement system, met the predetermined acceptance criteria)
    Hemocompatibility: ThrombogenicityISO 10994-4, ASTM F2888-19 (Demonstrates similar thromboresistance characteristics as control device)Pass (Demonstrates similar thromboresistance characteristics as the control device, met the predetermined acceptance criteria)
    Hemocompatibility: In Vitro HemocompatibilityISO 10993-4 (Not expected to result in adverse effects in vivo)Pass (Not expected to result in adverse effects in vivo, met the predetermined acceptance criteria)
    Hemocompatibility: Partial Thromboplastin TimeISO 10994-4, ASTM F2382-18 (Does not have an effect on coagulation of human plasma)Pass (Does not have an effect on coagulation of human plasma, met the predetermined acceptance criteria)

    2. Sample size used for the test set and the data provenance

    The document states "All samples met the pre-determined acceptance criteria" implying multiple samples were tested for each performance test. However, the specific sample sizes for each test are not explicitly provided in this summary.

    Data Provenance: The tests are described as in vitro (benchtop testing) and in vitro biocompatibility tests, some conducted in a "cerebral vascular model." It is retrospective in the sense that the results are being reported after the tests were conducted. The country of origin for the data is not specified but would typically be where MicroVention, Inc. conducts its R&D and testing, likely in the USA given their address is Tustin, California.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device submission. The "ground truth" for this medical device is established by physical and chemical properties and performance characteristics measured against recognized international standards (e.g., ISO, ASTM, USP) and internal specifications, rather than by human expert review of images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable as there is no human-in-the-loop assessment requiring adjudication for this device's performance testing. The tests are objective measurements against set standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is relevant for AI-powered diagnostic devices that assist human readers. The BOBBY Balloon Guide Catheter is a physical medical device, not an AI or imaging device, and does not involve human readers interpreting data assisted by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is not an algorithm-only or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance and biocompatibility is based on established engineering and material science standards (ISO, ASTM, USP), verified physical/chemical measurements, and the device's ability to consistently meet predefined specifications for functionality and safety in a laboratory setting. For comparison, the predicate device's characteristics also serve as a comparative ground.

    8. The sample size for the training set

    This question is not applicable. This device is not an AI/machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no AI training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192684

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-22

    (118 days)

    Product Code
    Regulation Number
    876.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroPearl® Microspheres are intended for the embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH).

    Device Description

    The HydroPearl TM Microspheres are a pre-formed, compressible, precisely calibrated, spherical embolic agent consisting of a biocompatible hydrogel. The HydroPearl TM Microspheres are offered in a variety of diameters ranging from 75-1100µm and are provided in a sterile syringe pre-filled with microspheres in phosphate buffered saline. The pre- filled syringe is packaged in a sealed sterile dispenser tray. The HydroPearl TM Microspheres are delivered to the treatment site through a delivery catheter.

    AI/ML Overview

    The provided document describes the acceptance criteria and study for the HydroPearl Microspheres, specifically for the expanded indication of prostatic artery embolization (PAE) for symptomatic benign prostatic hyperplasia (BPH).

    Here's an organized breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria with specific numerical thresholds (e.g., "IPSS score reduction must be X%"). Instead, it reports the observed clinical outcomes and concludes that these results support safe and effective use. The "acceptance criteria" here are implied through the positive clinical outcomes and the comparison to established understanding of PAE effectiveness.

    Acceptance Criteria (Implied)Reported Device Performance (HydroPearl Microspheres)
    Primary Endpoint: Reduction in IPSS score during follow-upMean baseline IPSS: 20.7 (range; 4-30) Mean IPSS at latest follow-up (8.6 months): 9.1 (range; 2-18) (p<0.05)
    Secondary Endpoint: Improvement in Quality of Life (QoL)Mean baseline QoL score: 3.7 (range; 1-6) Mean QoL score at follow-up: 1.2 (range; 0-3) (p<0.05)
    Technical Success: Bilateral prostatic artery embolizationAchieved in all 17 patients (100%)
    Clinical Success (based on symptom improvement/medication reduction)Achieved in 16/17 patients (94%) One patient considered a "clinical failure" (had LUTS well controlled on medication, wanted to cease medication, which was achieved, but symptoms mildly worsened from IPSS 4 to 10). Half of patients on BPH medications at baseline no longer required them after PAE; two additional patients reduced daily medications. Patient relying on CIC: Catheterized 5 times/day before PAE, 0 or 1 time/day after PAE.
    Safety: Low rate of adverse eventsOne major complication: hospitalization for acute prostatitis (n=1), resolved without residual effects. Minor complications: hematuria (n=1), hematospermia (n=1), considered expected for PAE and likely due to mild non-target embolization.
    Equivalence to Predicate DeviceComparative mechanical testing showed equivalent performance to Embosphere Microspheres. Pre-clinical animal testing showed similar performance to Embosphere Microspheres with favorable embolization results.

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): 17 patients
    • Data Provenance: Retrospective analysis from a single US center.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts for establishing ground truth for the clinical data. The data appears to be derived from standard clinical assessments (IPSS, QoL scores, observation of complications) and procedural records (technical success of embolization), rather than an explicit "ground truth" established by independent expert panel review.

    4. Adjudication method for the test set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical outcomes in the retrospective study. Clinical outcomes were reported based on collected patient data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focuses on the device's clinical performance in patients, not on comparing AI assistance to human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the HydroPearl Microspheres are a medical device (embolization agent), not an AI algorithm. The study assesses the performance of the physical device when used by clinicians.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the clinical study's effectiveness was established through patient-reported outcomes data and objective clinical assessments:

    • International Prostate Symptom Score (IPSS) and Quality of Life (QoL) scores (patient-reported).
    • Observation of BPH medication usage reduction/cessation.
    • Objective assessment of technical success (bilateral PAE).
    • Adverse event reporting.

    8. The sample size for the training set

    The document does not describe a "training set" in the context of an AI algorithm. For the clinical performance anaysis, it refers to a retrospective analysis of existing data.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI algorithm study requiring a training set with established ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192135

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-24

    (139 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIA 21, 27, 33 - The VIA® Microcatheter is intended for the introduction of interventional devices (such as the WEB device/stents/flow diverters) and infusion of diagnostic agents (such as contrast media) into the neuro, peripheral, and coronary vasculature.

    VIA 17, 17 Preshaped 45°, 17 Preshaped 90° - The VIA® Microcatheter is intended for the introduction of interventional devices (such as coils/stents) and infusion of diagnostic agents (such as contrast media) into the neuro, peripheral, and coronary vasculature.

    Device Description

    The VIA® Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature. The physician inserts the catheter into the vein or artery through the skin (percutaneous) using a sheath or guidewire. The device can then be navigated to the treatment site. Navigation is aided by the coated surface of the catheter which assists with manipulation while in the vasculature. Throughout the procedure the physician can obtain the position of the catheter by the radiopaque marker bands using fluoroscopic techniques (VIA17 Microcatheter has 2 radiopaque marker bands. VIA 21, 27 and 33 Microcatheters have one radiopaque tip marker band). Diagnostic and interventional devices can be delivered through the lumen of the catheter to the treatment site. The proximal end of the catheter incorporates a standard luer adapter to facilitate attachment of accessories.

    AI/ML Overview

    The provided FDA 510(k) summary for the VIA® Microcatheter does not contain the specific information required to answer all parts of your request, as it pertains to a medical device's 510(k) clearance rather than a diagnostic AI algorithm study. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing (biocompatibility and bench testing), not on a clinical effectiveness study with human readers and AI assistance.

    Therefore, for aspects related to AI performance, human expert adjudication, effect size with AI assistance, and large-scale training/test sets for an AI algorithm (questions 2-9), the information is not applicable or not provided in this document.

    However, I can extract the acceptance criteria and reported performance for the device itself (the VIA® Microcatheter) based on the provided bench testing data.


    Acceptance Criteria and Reported Device Performance (Based on Bench Testing)

    Acceptance Criteria / TestReported Device Performance
    Visual and Dimensional InspectionPass
    Hub ISO 80369-7Pass
    WEB RetractionPass
    Kink ResistancePass
    TensilePass
    Tip BucklingPass
    Steam Shaping / Shape RetentionPass
    Catheter Leakage and Static BurstPass
    Coating Friction & DurabilityPass
    Coating ParticulatePass
    Tracking ForcePass

    Note: Some tests (Coating Integrity, Catheter Torque Strength, Flow Rate, Dead Space) were reported "For characterization only," implying they were not subject to specific pass/fail acceptance criteria for this submission but were measured to understand device properties.


    Study Details (for the VIA® Microcatheter, not an AI study)

    1. A table of acceptance criteria and the reported device performance:

      • See table above.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: The document does not specify exact sample sizes for each bench test. For biocompatibility tests, specific numbers of animals (e.g., 17 male Hartley Guinea Pigs, 3 female New Zealand White Rabbits, 20 female Albino Swiss Mice, 2 Canine) are mentioned, but these are for biological testing, not for evaluating an AI algorithm on a medical dataset.
      • Data Provenance: Not applicable for device performance bench testing in the context of AI. The tests are laboratory-based.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. This is not an AI study requiring expert ground truth for medical images.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is not an AI study requiring expert adjudication.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This document outlines the premarket notification for a physical medical catheter, not an AI-powered diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For biocompatibility testing, the "ground truth" is defined by biological responses in test organisms and in vitro systems, evaluated against established ISO standards and criteria for non-toxicity, non-sensitization, non-irritation, non-hemolysis, etc.
      • For bench testing, the "ground truth" is adherence to predefined engineering and functional specifications or standards (e.g., ISO 10555-1:2014, ISO80369-7:2016).
    8. The sample size for the training set:

      • Not applicable. There is no training set for an AI algorithm in this document specific to the device's performance evaluation.
    9. How the ground truth for the training set was established:

      • Not applicable. There is no AI training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192625

    Validate with FDA (Live)

    Date Cleared
    2019-11-21

    (59 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.

    Device Description

    The PG Pro Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the catheter hub is used for the attachment of accessories.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for the "PG Pro Microcatheter". This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific performance acceptance criteria for an AI/ML type of medical device through a clinical study.

    Therefore, the text does not contain the information requested regarding:

    1. A table of acceptance criteria and reported device performance for an AI/ML device.
    2. Sample size and data provenance for a test set.
    3. Number of experts and their qualifications for ground truth establishment.
    4. Adjudication method for a test set.
    5. MRMC comparative effectiveness study results.
    6. Standalone performance of an algorithm.
    7. Type of ground truth used (expert consensus, pathology, outcomes data).
    8. Sample size for a training set.
    9. How ground truth for the training set was established.

    Instead, the document describes:

    • Device Description: The PG Pro Microcatheter is a single lumen catheter designed for introducing over a guidewire to access small, tortuous vasculature for infusing diagnostic and therapeutic agents in the peripheral vasculature.
    • Predicate Device: Headway 27 Microcatheter (K142449). A reference device (Headway 17 Microcatheter, K101542) is also listed.
    • Comparisons: The document includes a detailed "TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE" (Table 1) comparing the PG Pro Microcatheter to its predicate and reference devices across properties like Indications For Use, Principle of Operation, Material(s), Proximal ID/OD, Distal ID/OD, Number of Marker bands, Coating, Effective Length, Tip Configuration, Guidewire Compatibility, Method of Supply, and Accessories. The table highlights differences such as the PG Pro being intended only for peripheral vasculature, using a Tungsten coil vs. stainless steel, having a different range of effective lengths, being straight configuration only, being compatible with larger guidewires, and not including a shaping mandrel.
    • Performance Data (Non-Clinical):
      • Biocompatibility testing: Conducted according to ISO10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, and hemocompatibility, as it is considered a limited (<24 hour) circulating blood contacting device.
      • Bench testing: A comprehensive list of tests was performed, including surface contamination, physical attributes, force at break, leakage (liquid, air), static and dynamic burst, durability and lubricity of hydrophilic coating, simulated use in flow model, catheter flow rate, kink resistance, stiffness, flexural fatigue, particulate testing, catheter dead space, torque strength, hub testing, corrosion testing, and radio detectability.
      • Animal Study: An acute large animal study was conducted with n=3 animals. The study assessed the device's safety profile, compatibility with therapeutic agents, and tissue response, concluding that the PG Pro Microcatheter performed similarly to the predicate device (Headway 27 Microcatheter).
    • Conclusion: The PG Pro Microcatheter is considered substantially equivalent to the identified predicate based on performance, intended use, design, materials, principle of operation, and overall technological characteristics. The nonclinical data, verification, and validation testing support this conclusion and demonstrate the device should perform as intended.

    In summary, this document is a 510(k) submission for a physical medical device (microcatheter) and relies on benchmarking against a predicate device, along with non-clinical (bench and animal) testing to demonstrate substantial equivalence, not the evaluation of an AI algorithm against specific performance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173200

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-06-11

    (252 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA Plus Aspiration Catheter with the Gomco 405 Aspiration Tubing Kit is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Device Description

    The SOFIA Plus Aspiration Catheter is a single lumen catheter designed to be introduced over a steerable guidewire to access the neurovasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. A single radiopaque marker at the distal end facilitates fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic coating to reduce friction during navigation in the vasculature. A luer fitting on the microcatheter hub is used for the attachment of accessories. The strain relief at the hub provides kink resistance for the proximal end. A steam shaping mandrel and introducer sheath are also packaged with the catheter. The SOFIA Plus Aspiration Catheter is used to remove thrombus/embolus from the neurovasculature using aspiration tubing and pump.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the SOFIA Plus Aspiration Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials. Therefore, the information provided does not align with the typical structure of an AI/ML medical device study demonstrating algorithmic performance against acceptance criteria.

    The document primarily details the pre-clinical testing on the device itself and an animal study to evaluate its aspiration performance and safety. It does not describe an AI/ML algorithm or a study of its performance.

    However, I can extract and structure the information about the acceptance criteria and performance of the physical device based on the provided text, while acknowledging that this is not an AI/ML study.

    Here's an interpretation of the request using the available information:


    The SOFIA Plus Aspiration Catheter underwent extensive pre-clinical testing to demonstrate its performance and substantial equivalence to a predicate device. The acceptance criteria for these tests were typically defined by established specifications, industry standards (e.g., ISO), or direct comparison to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility
    Cytotoxicity (ISO 10993-5)Non-cytotoxicCell culture treated with test article exhibited slight reactivity (Grade 1), concluded as non-cytotoxic. PASSED
    Sensitization/Irritation (ISO 10993-10) - Guinea pig maximization sensitizationWeak allergic potential or sensitizing capacityExtracts of the test article elicited no reaction at the challenge (0% sensitization) following the induction phase (Grade 1). PASSED
    Sensitization/Irritation (ISO 10993-10) - Intracutaneous reactivityNon-irritantExtracts of the test article did not show a significantly greater biological reaction than the sites injected with the control article. PASSED
    Hemocompatibility – Rabbit Blood Direct and Indirect Contact (ISO 10993-4)Non-hemolyticHemolysis index was 0.13% (direct contact) and 0.0% (indirect contact). PASSED
    Hemocompatibility – Unactivated Partial Thromboplastin Time Assay Direct Contact (ISO 10993-4)No effect on coagulationNo statistically significant difference found between the Unactivated Partial Thromboplastin Time (UPTT) of the plasma exposed to the test article and that of the plasma exposed to either the negative control or the untreated control. PASSED
    Hemocompatibility - Complement Activation Assay (ISO 10993-4)No effect on complement activationC3a and SC5b-9 levels < negative and untreated controls. PASSED
    Hemocompatibility - Thrombogenicity Study in Dogs (ISO 10993-4)No significant thrombosisMinimal thrombosis observed with a Grade 0 in two out of two test sites and two out of two control sites. PASSED
    Systemic Toxicity - Systemic Injection Test (ISO 10993-11)No toxic effectsExtracts of test article did not induce a significantly greater biological reaction than the control extracts when injected in Swiss Albino mice. PASSED
    Systemic Toxicity - Rabbit Pyrogen Test (ISO 10993-11)Non-pyrogenicThe temperature increases (maximum) was 0.03°C from baseline. PASSED
    Bench Testing
    Simulated UseDevice performs as intended, demonstrates equivalency to predicate deviceAchieved a rating $\geq$ 3 for preparation/ease of assembly, introducer sheath interaction, introducer peel away, tracking with guidewire/microcatheter, microcatheter/guidewire lockup, lubricity and durability of hydrophilic coating, microcatheter/guidewire removal, removal/aspiration of clot. PASSED
    Dimensional and Physical AttributesDevice meets specified dimensional requirements, substantially equivalent to predicate deviceMet specified dimensional requirements for catheter OD, catheter ID, overall working length, length of distal section, length of distal tip to marker band and total length of hub/strain relief. PASSED
    Kink ResistanceDevice resistant to kinking around small radii turns, same as predicate deviceNo kinks when wrapped around 0.030 and 0.040-inch pin gauges. No kinks noted during simulated use testing. PASSED
    Durability/Lubricity of Hydrophilic CoatingDevice tracks easily with no coating cracking or separationAchieved a rating of $\ge$ 3 during simulated use testing for coating durability and lubricity. PASSED
    Catheter StiffnessDevice tracks in tortuous anatomy while advancing to target site, equivalent to predicate and competitive devicesDevice stiffness equivalent to predicate and competitive devices. PASSED
    Catheter Flexural FatigueDevice integrity suitable for intended clinical useNo flexural fatigue following repeated bending during simulated use testing and repeated hoop stress following pressure and air aspiration testing. PASSED
    Force at Break (Shaft and hub)Tensile strength test results meet acceptance criteria and equivalent to predicate and competitive devicesCatheter force at break $\ge$ 15N for shaft section and hub/catheter junction. PASSED
    Force at Break (After tip shaping)Device meets acceptance criteria and no difference between pre- and post- force at breakCatheter force at break $\ge$ 15N for distal section assessed pre- and post-tip shaping. PASSED
    Static Burst PressureDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No damage of pressurized catheter at 46 psi. PASSED
    Fluid Leakage at > 46 psiDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No liquid leakage from hub and catheter shaft at 46 psi for 30 seconds. PASSED
    Air LeakageDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No air leakage at hub into syringe for 15 seconds. PASSED
    Dynamic BurstDevice met labeled maximum infusion pressure of 300 psiTest articles did not burst at or below 300 psi. PASSED
    Catheter CollapseNo observation of device collapse, equivalent to predicate deviceTest articles were pressurized for a duration of 6 minutes and no observation of catheter collapse. PASSED
    Vacuum PressureNo difference was measured between device and predicateVacuum pressure testing of catheter at distal tip was compared to vacuum pressure of source. No difference detected. PASSED
    Torque ResponseEquivalent or better than predicateCatheter torque response was assessed in vascular model and met acceptance criteria. PASSED
    Torque StrengthMet acceptance criteriaCatheter was torqued and did not break. PASSED
    Advance/RetractMet acceptance criteria, equivalent or better than predicateCatheter was subjected to advance/retraction test in a vascular model and was equivalent or better than predicate. PASSED
    RadiopacityMarker band is visible under fluoroscopyMarker band was detectable under fluoroscopy. PASSED
    Corrosion ResistanceCatheter is corrosion resistantMet ISO 10555-1. PASSED
    Particulate TestingCatheter does not generate particulateMet USP <788> criteria. PASSED

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for each bench test. Instead, it refers to "test articles" and general results (e.g., "Test articles achieved a rating...") and in some cases, "two out of two test sites" for the animal study.

    • Test Set Provenance: The tests are pre-clinical bench tests performed by the manufacturer (MicroVention, Inc., Tustin, California, USA) and an animal study (swine model). The information does not specify the country of origin of data in terms of retrospective/prospective human patient data, as this is not a human clinical trial. All tests would have been performed prospectively as part of the device development and submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable as the document describes pre-clinical physical device testing and an animal study, not an AI/ML algorithm study that requires expert adjudication of image-based ground truth. For the simulated-use tests, the performance ratings (e.g., $\geq3$) would be based on qualitative assessment by trained personnel, but they are not referred to as "experts" in the context of clinical ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. As this is not an AI/ML algorithm study involving interpretation of clinical data by multiple readers, there is no need for an adjudication method like 2+1 or 3+1.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, this was not done. This document pertains to the 510(k) clearance of a physical medical device. MRMC studies are typically performed for AI-driven diagnostic or assistive technologies to assess their impact on human reader performance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, this was not done. This is not an AI/ML algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Engineering Specifications and Standards: For dimensional, mechanical, and safety properties (e.g., ISO standards, specified psi limits, force at break criteria).
    • Comparative Performance to Predicate Devices: Performance attributes (e.g., stiffness, vacuum pressure, torque response) were compared to the legally marketed predicate device (Penumbra Reperfusion Catheter ACE64 and ACE68).
    • Animal Study Observations: For aspiration performance and histopathological evaluations in a swine model.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML study, therefore there is no training set for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI/ML algorithm training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 7