Search Results
Found 1 results
510(k) Data Aggregation
(199 days)
The SpiderFX Embolic Protection Device is indicated for use as a guidewire and embolic protection system to contain and remove embolic material in conjunction with the TurboHawk, either during standalone procedures or together with PTA and/or stenting, in the treatment of severely calcified lesions in arteries of the lower extremities.
The SpiderFX® Embolic Protection Device is a percutaneously delivered distal embolic protection system that can be delivered over any 0.014" or 0.018" guidewire. The SpiderFX Embolic Protection Device contains a Capture Wire composed of a nitinol mesh filter mounted on a 190 cm or a convertible 320/190 cm PTFE-coated 0.014" stainless steel guidewire and a dualended SpiderFX Catheter for delivery and recovery. The SpiderFX® Embolic Protection Device uses the following materials: pebax, grilamid, platinum/iridium, nitinol, stainless steel, PTFE coating, gold tungsten, and hydrophilic coating.
Here's a breakdown of the acceptance criteria and study information for the SpiderFX® Embolic Protection Device, based on the provided 510(k) summary:
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Specific Acceptance Criteria (if quantifiable) | Reported Device Performance |
---|---|---|
Primary Safety Endpoint (Clinical Study) | 30-day freedom from Major Adverse Event (MAE) rate performance goal = 85.5% | 93.1% (122/131) 30-day freedom from MAE rate. 95% lower confidence limit was 88.3%. |
Biocompatibility | Meets requirements for biocompatibility testing outlined in ISO 10993-1 Part 1: 2003 | All leveraged tests (cytotoxicity, sensitization, intracutaneous injection, systemic injection, hemolysis, pyrogen, complement activation, and thrombogenicity) met specified acceptance criteria. |
Stent Compatibility | Not specified in detail, implied to be functionally compatible | Not explicitly quantifiable, but "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
Filter Efficiency | Not specified in detail, implied to be functionally efficient | Not explicitly quantifiable, but "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
Radial Outward Force | Not specified in detail, implied to meet functional requirements | Not explicitly quantifiable, but "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
Simulated Use | Not specified in detail, implied to meet functional requirements | "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
Deployment/Retrieval Forces | Not specified in detail, implied to meet functional requirements | "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
In Vivo Animal Studies | Not specified, implied to demonstrate safety and effectiveness for proposed use | "Tests were performed" and results "demonstrate that the technological characteristics and performance criteria are comparable". |
Embolic Capture Efficiency and Retrieval Ability | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Filter Capacity | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Resistance to Filter Rupture During Removal of a Fully Loaded Filter | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Flow Characteristics | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Tip Flexibility | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Tensile Strength | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Torque Strength | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Torque Response | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Kink Resistance | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Dimensional Verification | Not specified, implied to meet functional requirements | "Test results met the specified acceptance criteria". |
Package Integrity | Not specified, implied to maintain sterility and device integrity | "Test results met the specified acceptance criteria". |
Sterilization | Not specified, implied to meet sterility assurance level | "Test results met the specified acceptance criteria". |
Shelf Life | Not specified, implied to maintain device integrity and function over shelf life | "Test results met the specified acceptance criteria". |
Study Details
Clinical Study (DEFINITIVE Cat+)
-
Sample size used for the test set and data provenance:
- Sample Size: 131 subjects.
- Data Provenance: Prospective, multi-center, non-randomized, single-arm study. The country of origin is not explicitly stated, but it's typically a multi-national or US-based study for FDA submissions. The study involved comparison to a performance goal derived from an observational multi-center registry (TALON).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Number of Experts: Not explicitly stated, but a "clinical events committee (CEC)" was used for adjudication. Specific number and qualifications are not provided in this summary.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: "as adjudicated by the clinical events committee (CEC)". The specific method (e.g., majority vote, consensus) for the CEC is not detailed.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, this was not an MRMC comparative effectiveness study involving human readers and AI assistance. This was a clinical study evaluating the device's safety and effectiveness.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not applicable. This device is a physical embolic protection system, not an AI algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth: For the primary safety endpoint, the "ground truth" was defined by the occurrence of Major Adverse Events (MAE) through 30 days post-procedure, as adjudicated by a Clinical Events Committee (CEC). This is essentially outcomes data, interpreted and confirmed by expert consensus within the CEC.
-
The sample size for the training set:
- Training Set Sample Size: Since this is a physical medical device and not an AI algorithm, there isn't a "training set" in the conventional machine learning sense for the clinical study. The study population of 131 subjects served as the test set for the device's performance.
-
How the ground truth for the training set was established:
- Training Set Ground Truth Establishment: Not applicable, as there is no traditional "training set" for an AI model.
Pre-Clinical Performance Testing
For the extensive list of performance tests (Stent Compatibility, Filter Efficiency, Radial Outward Force, Simulated Use, Deployment/Retrieval Forces, In Vivo Animal Studies, Embolic Capture Efficiency and Retrieval Ability, Filter Capacity, Resistance to Filter Rupture During Removal of a Fully Loaded Filter, Flow Characteristics, Tip Flexibility, Tensile Strength, Torque Strength, Torque Response, Kink Resistance, Dimensional Verification, Package Integrity, Sterilization, Shelf Life):
- Sample Size, Data Provenance, Experts for Ground Truth, Adjudication, MRMC, Standalone, Ground Truth Type, Training Set: These details are not provided in the 510(k) summary for these specific pre-clinical tests. They are generally performed by engineers and technicians according to validated test methods, and the "ground truth" is typically the measured physical properties and performance against pre-defined engineering specifications or industry standards. The summary indicates that "Test results met the specified acceptance criteria" for these tests, leveraging data from predicate device submissions (K063204 or K052659).
Ask a specific question about this device
Page 1 of 1