K Number
K172014

Validate with FDA (Live)

Date Cleared
2018-04-05

(276 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wedge Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices, such as the SOFIA 6F Catheter, in the neurovasculature.

Device Description

The Wedge Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The microcatheter has a semi-rigid proximal section with an outer shaft made of Grilamid nylon. The catheter shaft transitions to progressively softer durometers and different lengths of Polyether block amide (Pebax). The distal-most length of the microcatheter, beyond the enlarged segment, consists of a softer, atraumatic polyurethane.

The enlarged segment on the distal end of the Wedge is designed to reduce the gap between the OD of the guidewire and ID of the Sofia 6F. The tapered bulb section, approximately 1 cm length and located approximately 1.5 cm from the distal tip, can be identified on fluoroscopy between the two radiopaque proximal marker bands of the Wedge Microcatheter. The bulb OD (0.068") is sized specifically to work with the lumen ID (0.070") of the Sofia 6F allowing for continuous flush of saline through the Sofia.

Three radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the microcatheter is coated with a hydrophilic polymer coating to reduce friction during navigation in the vasculature. The lubricious inner liner is made from polytetrafluoroethylene (PTFE). A luer fitting on the Microcatheter hub is used for the attachment of accessories. The hub/strain relief provides for the kink resistance at the proximal end. The microcatheter has a straight tip that is designed to be steam shaped by the physicians at the time of the use. A steam shaping mandrel and introducer sheath (accessories) are packaged with the catheter.

AI/ML Overview

The provided document describes the MicroVention, Inc. Wedge Microcatheter and its substantial equivalence to predicate devices, rather than a standalone study with defined acceptance criteria and detailed device performance metrics in the way a clinical trial or a specific performance study for a diagnostic device might.

However, based on the Verification Test Summary and the Biocompatibility Evaluation sections, we can infer the acceptance criteria and reported device performance for various engineering and biological aspects. The "Results" column in these tables implicitly represents the device's performance against an unstated "Pass" acceptance criterion.

Here's an attempt to structure the information as requested, interpreting the "Pass" result as meeting the acceptance criteria.


1. Table of Acceptance Criteria and Reported Device Performance

Test DescriptionAcceptance Criteria (Inferred)Reported Device Performance
Verification Test Summary
Surface Contamination"Pass" standardPass
Physical Attributes"Pass" standardPass
Force at Break (Catheter Distal Section)"Pass" standardPass
Force at Break (Catheter Hub Junction)"Pass" standardPass
Freedom from Leakage (Low Pressure, Long Duration)"Pass" standardPass
Freedom from Leakage (High Pressure, Short duration)"Pass" standardPass
Freedom from Leakage - Air"Pass" standardPass
Static Burst Pressure"Pass" standardPass
Dynamic Burst Pressure"Pass" standardPass
Coating Durability/Lubricity"Pass" standardPass
Tip Shape and Tip Retention"Pass" standardPass
Simulated Use"Pass" standardPass
Flow Rate"Pass" standardPass
Kink Resistance"Reference Only" standardReference Only
Catheter Stiffness"Pass" standardPass
Catheter Flexural Fatigue"Pass" standardPass
Catheter Particle Testing"Pass" standardPass
Dead Space"Reference Only" standardReference Only
Torque Strength"Pass" standardPass
Corrosion Resistance*"Pass" standardPass
Gauging Test*"Pass" standardPass
Separation Force*"Pass" standardPass
Unscrewing Torque*"Pass" standardPass
Resistance to Overriding*"Pass" standardPass
Stress Cracking*"Pass" standardPass
Radiopacity* (Visibility under fluoroscopy)"Pass" standardPass
Pyrogenicity*"Pass" standardPass
Ship Testing*"Pass" standardPass
Shelf Life Testing"Pass" standardPass
Biocompatibility Evaluation
Cytotoxicity - Medium Eluate MethodNon-cytotoxicNon-cytotoxic
Sensitization: Maximization Test in Guinea PigsNon-sensitizerNon-sensitizer
Intracutaneous ReactivityNon-irritatingNon-irritating
Systemic Injection Test in MiceSystemically non-toxicSystemically non-toxic
Rabbit Pyrogen TestNonpyrogenic (rise < 0.5°C)Nonpyrogenic
ASTM Blood Compatibility - Direct and Indirect Contact HemolysisNon-hemolyticNon-hemolytic
Unactivated Partial Thromboplastion Time TestNo effect on clottingNo effect on clotting
Complement ActivationNon-activatedNon-activated
ThrombogenicityNon-thrombogenicNon-thrombogenic

Note: For the tests marked with an asterisk (), the document states: "Testing was previously conducted on test article that was equivalent in all aspects relevant to the testing performed, therefore it was deemed unnecessary to repeat the testing for the Wedge Microcatheter." This implies these tests were accepted based on prior results on an equivalent device.*

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample sizes for each specific verification test mentioned. The "Verification Test Summary" refers to tests conducted on the "subject device," implying direct testing of the Wedge Microcatheter.

  • Animal Testing Summary: While not a "test set" in the sense of human patient data, the animal study compared the Wedge Microcatheter with the predicate Headway 21. It doesn't specify the number of animals used.
  • Biocompatibility Evaluation: This section describes various in vitro and in vivo tests (e.g., Guinea pigs for sensitization, mice for systemic injection, rabbits for pyrogenicity). The numerical sample sizes for these animal models are not provided, nor is the country of origin. The data is prospective, generated specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document does not describe a scenario where "experts" are establishing ground truth for a test set in the context of clinical images or diagnostic outcomes. The tests performed are primarily engineering performance tests and biological reactivity tests. Therefore, this question is not applicable in the traditional sense for this type of device submission. The "ground truth" for these tests is established by standardized testing methods and validated experimental procedures.

4. Adjudication Method for the Test Set

Not applicable. This is not a study involving human interpretation of data requiring adjudication. The device performance is assessed through objective measurements and standard laboratory protocols.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device submission for a microcatheter, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm or AI device. The tests performed are standalone in the sense that they evaluate the device itself, but not an algorithm.

7. The Type of Ground Truth Used

  • Engineering/Performance Tests: The "ground truth" is based on established engineering standards and specifications for catheter performance (e.g., burst pressure, flow rate, force at break). These are objective, measurable outcomes.
  • Biocompatibility Tests: The "ground truth" is based on the results of validated in vitro and in vivo biological assays comparing the device's interaction with biological systems against established thresholds for toxicity, irritation, sensitization, etc., as per ISO 10993-1:2009 and FDA guidelines.
  • Animal Testing: The "ground truth" for the animal study (comparative performance and histopathology) was direct observation of the animal's vasculature and subsequent histopathological examination.

8. The Sample Size for the Training Set

Not applicable. This is not an AI or machine learning device that requires a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable. As this is not an AI or machine learning device, there is no training set or associated ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath.

April 5, 2018

MicroVention, Inc. Marina Emond Manager, Regulatory Affairs 1311 Valencia Avenue Tustin, California 92780

Re: K172014

Trade/Device Name: Wedge Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 1, 2018 Received: March 5, 2018

Dear Marina Emond:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/4 description: The image shows the text "Carlos L. Pena -S" in a simple, sans-serif font. The text is arranged horizontally, with a clear distinction between the first name, last name, and the suffix. The background is plain and does not distract from the text.

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172014

Device Name Wedge Microcatheter

Indications for Use (Describe)

The Wedge Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices, such as the SOFIA 6F Catheter, in the neurovasculature.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary - K172014

Trade Name:Wedge Microcatheter
Generic Name:Percutaneous Catheter
Classification:II, 21 CFR 870.1250, DQY
Submitted By:MicroVention, Inc.1311 Valencia AveTustin, California, USA
Contact:Marina EmondManager, Regulatory AffairsMarina.Emond@Microvention.com(714) 247-8296
Date:April 1, 2018
Predicate Device:Headway 21 Microcatheter (K093160)
Headway 17 Microcatheter (K083343)
AXS Offset Delivery Assist Catheter (K163259)

Device Description:

The Wedge Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The microcatheter has a semi-rigid proximal section with an outer shaft made of Grilamid nylon. The catheter shaft transitions to progressively softer durometers and different lengths of Polyether block amide (Pebax). The distal-most length of the microcatheter, beyond the enlarged segment, consists of a softer, atraumatic polyurethane.

The enlarged segment on the distal end of the Wedge is designed to reduce the gap between the OD of the guidewire and ID of the Sofia 6F. The tapered bulb section, approximately 1 cm length and located approximately 1.5 cm from the distal tip, can be identified on fluoroscopy between the two radiopaque

{4}------------------------------------------------

proximal marker bands of the Wedge Microcatheter. The bulb OD (0.068") is sized specifically to work with the lumen ID (0.070") of the Sofia 6F allowing for continuous flush of saline through the Sofia.

Three radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the microcatheter is coated with a hydrophilic polymer coating to reduce friction during navigation in the vasculature. The lubricious inner liner is made from polytetrafluoroethylene (PTFE). A luer fitting on the Microcatheter hub is used for the attachment of accessories. The hub/strain relief provides for the kink resistance at the proximal end. The microcatheter has a straight tip that is designed to be steam shaped by the physicians at the time of the use. A steam shaping mandrel and introducer sheath (accessories) are packaged with the catheter.

Indications for Use:

The Wedge Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices, such as the SOFIA 6F Catheter, in the neurovasculature.

Technological Characteristics and Product Feature Comparison:

The subject device, Wedge Microcatheter is substantially equivalent to the predicate devices in terms of:

  • Intended use
  • Scientific technology
  • · Fundamental design
  • · Materials and processes for packaging and sterilization of devices

A tabular comparison of the technological characteristics between the predicate devices and subject device is provided below.

{5}------------------------------------------------

DeviceCharacteristicsHeadway 17 Microcatheter(K083343)Headway 21 Microcatheter(K093160)AXS Offset Delivery AssistCatheter (K163259)Wedge Microcatheter(Proposed)
DeviceClassification/Product CodeClass II/ DQY(Percutaneous catheters)Class II/ DQY(Percutaneous catheters)Class II/ DQY(Percutaneous catheters)Class II/ DQY(Percutaneous catheters)
Intended UseThe Headway Microcatheteris intended for generalintravascular use, includingthe peripheral, coronary andneuro vasculature for theinfusion of diagnosticagents, such as contrastmedia, and therapeuticagents, such as occlusioncoils.Intended for generalintravascular use, includingthe peripheral, coronary andneuro vasculature for theinfusion of diagnostic agents,such as contrast media, andtherapeutic agents, such asocclusion coils.The AXS Offset DeliveryAssist Catheter is intended toassist in the delivery ofinterventional devices, suchas distal access catheters, inthe neurovasculature.The Wedge Microcatheteris intended for generalintravascular use,including the peripheral,coronary and neurovasculature for theinfusion of diagnosticagents, such as contrastmedia, and to assist in thedelivery of interventionaldevices, such as theSOFIA 6F Catheter, inthe neurovasculature.
Catheter OD.025" -.031".028" - .034".050".028" - .034"Equivalent to predicates
Catheter ID.017".021".021".021"Equivalent to predicates
TipShapeable15 cmShapeable15 cmStraight2 cmShapeable, same aspredicates Headway 17and Headway 21
DeviceHeadway 17 MicrocatheterHeadway 21 MicrocatheterAXS Offset Delivery AssistWedge Microcatheter
Characteristics(K083343)(K093160)Catheter (K163259)(Proposed)
Distalsegment/bulbN/AN/AThe distal outer diameter(OD) of the AXS OffsetCatheter gradually increasesfrom 0.036in at the ROMarker to 0.050in, 2 cmproximal to the RO marker.The bulb section with a0.050in OD is maintained for28 cm, then graduallydecreases towards theproximal section. The overalldistal profile of the AXSOffset catheter acts as asmooth transition andreduces the gap between theouter diameter of a steerableguidewire and inner diameterof a DAC while allowing forcontinuous saline flushthrough the DAC.Using the same principleas the predicate AXSOffset, the slightlyenlarged segment (bulb)of the Wedge reduces thegap between the outsidediameter of the guidewireand the inside lumen ofthe Sofia allowing forcontinuous flush of salinethrough the Sofia.
EffectiveLength$150 cm \pm 2$$150 cm \pm 2$150 cm158-160 cmEquivalent to predicates
CoatingHydrophilic CoatingHydrophilic CoatingHydrophilic CoatingHydrophilic coating ofthe same composition aspredicates Headway 17and Headway 21 toreduce friction during use.
HydrophilicCoating Length100-105 cm100-105 cm80 cm110-115 cmEquivalent to predicates
DeviceCharacteristicsHeadway 17 Microcatheter(K083343)Headway 21 Microcatheter(K093160)AXS Offset Delivery AssistCatheter (K163259)Wedge Microcatheter(Proposed)
PackagingMaterialDispenser hoop:PolyethyleneMounting card:PolyethylenePouch: TyvekCarton Box:Bleached SulfateSame as Headway 17Catheter is placed in adispenser coil, then insertedinto a pouch and placedinside a carton box.Same aspredicates Headway 17and Headway 21;equivalent to AXS Offset.
PackageConfig.Microcatheter isplaced in adispenser hoopand accessories ona mounting cardthat is theninserted into thepouch. The pouchis then placedinside a cartonbox.
Method ofSupplyingSterile and single useSterile and single useSterile and single useSame
Method ofSterilizationEthylene oxideEthylene oxideEthylene oxideSame

Product Feature Comparison of Subject Device with Predicate Devices (K083343, K093160, K163259)

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Verification Test Summary:

The results of verification and validation testing conducted on the subject device demonstrate that it performs as intended and are summarized as follows:

Test DescriptionResult
Surface ContaminationPass
Physical AttributesPass
Force at Break (Catheter Distal Section)Pass
Force at Break (Catheter Hub Junction)Pass
Freedom from Leakage (Low Pressure, Long Duration)Pass
Freedom from Leakage (High Pressure, Short duration)Pass
Freedom from Leakage - AirPass
Static Burst PressurePass
Dynamic Burst PressurePass
Coating Durability/LubricityPass
Tip Shape and Tip RetentionPass
Simulated UsePass
Flow RatePass
Kink ResistanceReference Only
Catheter StiffnessPass
Catheter Flexural FatiguePass
Catheter Particle TestingPass
Dead SpaceReference Only
Torque StrengthPass
Corrosion Resistance*Pass
Gauging Test*Pass
Separation Force*Pass
Unscrewing Torque*Pass
Resistance to Overriding*Pass
Test DescriptionResult
Stress Cracking*Pass
Radiopacity* (Visibility under fluoroscopy)Pass
Pyrogenicity*Pass
Ship Testing*Pass
Shelf Life TestingPass

{9}------------------------------------------------

" Testing was previously conducted on test article that was equivalent in all aspects relevant to the testing performed, therefore it was deemed unnecessary to repeat the testing for the Wedge Microcatheter.

Animal Testing Summary:

The acute performance/efficacy and safety parameters analyzed (insertion of introducer sheath into RHV, peel away introducer sheath from catheter, track test with guidewire, track test with guidewire/SOFIA 6F and overall performance) were comparable between the Wedge Microcatheter and predicate Headway 21 with no dissection, perforation, luminal narrowing, thrombus formation or distal emboli were noted for both test articles.

Histopathology Results: Morphometric measurements showed no neointimal growth or stenosis in both, the Wedge Microcatheter and predicate Headway 21, with largely identical endothelial loss and no incidence of vessel wall thinning or aneurysmal dilatation. There were no disruptions of the internal elastic lamina or medial layers in any vessel regardless of treatment device. Injury was limited to occasionally marked endothelial cell denudation without remarkable thrombus deposition in any treated vessel segment. Overall, the results of the study demonstrated substantial equivalence of the Wedge Microcatheter to the predicate.

Biocompatibility Evaluation:

The in vitro and in vivo biocompatibility safety studies performed on the Wedge Microcatheter have demonstrated the biocompatibility of the Wedge Microcatheter and support compliance with the ISO 10993-1:2009 and FDA guidelines. The device was determined to be non-cytotoxic, non-sensitizing, intracutaneously non-irritating, systemically non-toxic, non-pyrogenic (material-mediated), non-hemolytic, have no effect on clotting, non-complement activating, and non-thrombogenic. The results of biocompatibility evaluation are summarized as follows:

{10}------------------------------------------------

TestTest SummaryConclusions
Cytotoxicity - MediumEluate MethodThe test article extract exhibited between no cell lysis(grade 0) to slight reactivity (grade 1).Non-cytotoxic
Sensitization:Maximization Test inGuinea PigsNo irritation was present on any of the test or negativecontrol (0% sensitized) guinea pigs.Non-sensitizer
Intracutaneous ReactivityNo evidence of irritation (score 0.0).Non-irritating
Systemic Injection Test inMiceNo weight loss, mortality, or evidence of systemictoxicity from the extract exposure to the mice wasobserved.Systemicallynon-toxic
Rabbit Pyrogen TestThe rise of rabbit temperatures during the tree hours ofobservation did not exceed 0.5°C.Nonpyrogenic
ASTM BloodCompatibility - Direct andIndirect Contact HemolysisThe test article demonstrated 0.59% hemolysis in directcontact and 1.25% hemolysis in indirect contact.Non-hemolytic
Unactivated PartialThromboplastion TimeTestAn average clotting time of the test article showed nosignificant difference from the control.No effect onclotting
Complement ActivationThe plasma exposed to the test article for 90 minutes wasfound to exhibit no statistically significant increase inC3a.Non-activated
ThrombogenicityBoth animals exhibited no signs of toxicity during thestudy (score 0)Non-thrombogenic

Summary of Substantial Equivalence:

The information presented in this 510(k) demonstrates the substantial equivalence between the predicates Headway 21 Microcatheter (K093160), Headway 17 (K083343), AXS Offset (K163259), and the Wedge Microcatheter in regard to the design, construction materials, operating principle, and intended use.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).