Search Filters

Search Results

Found 253 results

510(k) Data Aggregation

    K Number
    K253686

    Validate with FDA (Live)

    Date Cleared
    2026-03-23

    (122 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    0 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages.

    Device Description

    Computed Tomography (CT) is an indispensable imaging modality in clinical diagnostics due to its ability to provide detailed cross-sectional images of anatomical structures. However, achieving high spatial resolution remains a persistent challenge, particularly in applications requiring the visualization of fine structural details such as in inner auditory canal imaging, vascular studies, lung imaging, and bone microarchitecture analysis. Reconstruction techniques that attempt to boost spatial resolution typically also would amplify the noise which often result in tradeoffs to be made between resolution and noise.

    As part of the continuous innovation to solve the above challenges in CT imaging, GEHC developed a deep learning-based CT reconstruction algorithm specifically designed for high contrast lung and bone imaging, which is the subject of this premarket notification. This reconstruction algorithm, marketed under the name True Definition DL (TDDL), is an additional user-selectable recon option specifically designed for lung and bone imaging. It aims to enhance spatial resolution for both in-plane and cross-plane directions. This algorithm is incorporated into the reconstruction chain of the Revolution CT /Apex Family CT systems including Revolution CT/Revolution CT ES, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select all cleared under (K213715) and Revolution Vibe (K250941).

    True Definition DL offers three strengths that the user can choose depending on enhancement preferences. The benefits provided by this enhancement include improved spatial resolution measured by MTF.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the "True Definition DL" device, as extracted from the provided FDA 510(k) clearance letter:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Enhance spatial resolution for both in-plane and cross-plane directions in lung and bone imaging.Improves the in-plane and cross-plane resolution for body (lung and bone) scans as well as bony structures in the head without significantly increasing the image noise, in comparison with a sharper resolution kernel.
    Provide diagnostic value by improving high contrast spatial resolution in routine bone (including inner ears and spines) and lung exams.Validated through a reader study; improves confidence in the assessment of high contrast structures in these types of images.
    Improve confidence in the assessment of high contrast structures in lung and bone images.Validated through a reader study; improves confidence in the assessment of high contrast structures in these types of images.
    Not present algorithm-induced artifacts or enhancements not supported by underlying anatomies.Readers confirmed that the images enhanced by the subject device do not present algorithm-induced artifacts or enhancements not supported by underlying anatomies.
    Consistently enhance anatomical edges without introducing spurious structures.Results indicated that True Definition DL consistently enhanced anatomical edges without introducing spurious structures.
    Preserve noise texture under realistic dose conditions.Results indicated that True Definition DL preserved noise texture under realistic dose conditions.
    Maintain strong correlation with input and target images across all frequency bands.Results indicated that True Definition DL maintained strong correlation with input and target images across all frequency bands.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample size: Not explicitly stated for the clinical reader study, only referred to as "sample clinical data."
      • Data provenance: Not explicitly stated (e.g., country of origin or retrospective/prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of experts: Not explicitly stated.
      • Qualifications: "US board-certified radiologists."
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not explicitly stated. The document mentions "readers confirmed," implying a consensus or majority opinion, but the specific adjudication method is not described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • A reader study was performed ("clinical testing was carried out in the form of a reader study").
      • The document states that the study "validated that True Definition DL provides diagnostic value by improving high contrast spatial resolution... and improves confidence in the assessment of high contrast structures."
      • However, the document does not explicitly state that it was an MRMC comparative effectiveness study comparing human readers with AI vs. without AI assistance. It primarily focuses on the diagnostic value and confidence with the use of the images enhanced by True Definition DL.
      • No effect size or specific metrics quantifying improvement with AI assistance are provided.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone evaluation of the algorithm's performance was indicated through "robustness of the True Definition DL algorithm" testing. This included assessment of hallucination risks, enhancement of anatomical edges, preservation of noise texture, and correlation with input/target images, suggesting an algorithm-level assessment.
      • Bench testing on phantoms also measured the performance of True Definition DL against industry-standard IQ metrics and traditional metrics without direct human interaction.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the clinical reader study, the "ground truth" was established by expert opinion/consensus of US board-certified radiologists who evaluated the images and confirmed diagnostic value and lack of artifacts.
      • For bench testing, "well established test methods using industry-standard IQ metrics" provided quantifiable reference points for performance.
    7. The sample size for the training set:

      • Not explicitly stated.
    8. How the ground truth for the training set was established:

      • Not explicitly stated. The document only mentions that True Definition DL is a "deep learning-based CT reconstruction algorithm," implying a training phase, but provides no details on how the training data was curated or its ground truth established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K253738

    Validate with FDA (Live)

    Date Cleared
    2026-01-23

    (60 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 3.0T AIR™ 32CH HNA is a receive-only RF Coil designed for use with GE HealthCare 3.0T MRI systems. The coil is indicated for high-resolution magnetic resonance imaging (MRI) of the head and brain. When used with a Posterior Array in the MRI System patient table it also includes neck, cervical spine, neurovascular structures, upper thoracic spine, and brachial plexus imaging. The nucleus detected is hydrogen.

    Device Description

    The 3.0T AIR 32CH HNA is a receive-only radio frequency coil engineered to deliver optimal signal-to-noise ratio, uniform anatomical coverage, and high acceleration capabilities, including multiband imaging. It is intended for high-resolution magnetic resonance imaging (MRI) of the head and brain. When used in conjunction with the Posterior Array in the MRI system's patient table, it also supports imaging of the neck, cervical spine, neurovascular structures, upper thoracic spine, and brachial plexus. The nucleus detected is hydrogen. This coil is compatible with GE HealthCare 3.0T MRI systems.

    The 3.0T AIR 32CH HNA is designed to be used by a Registered MRI Technologist in a hospital or clinical setting. The Registered MRI Technologist will operate the scanner from the control room. If the patient is claustrophobic another MRI Technologist or clinical staff member may stay in the magnet room with the patient.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253639

    Validate with FDA (Live)

    Device Name
    View
    Manufacturer
    Date Cleared
    2026-01-08

    (50 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    View is a software application that displays, processes, and analyzes medical image data and associated clinical reports to aid in diagnosis for healthcare professionals. It streamlines standard and advanced medical imaging analysis by providing a complete suite of measurement tools intended to generate relevant findings automatically collected for export and save purposes.

    Typical users of this system are authorized healthcare professionals.

    Mammography images may only be interpreted using a monitor compliant with requirements of local regulations and must meet other technical specifications reviewed and accepted by the local regulatory agencies. Lossy compressed mammographic images and digitized film screen images should not be reviewed for primary image interpretations with use of the View.

    Device Description

    View is a cloud-native software application designed to support healthcare professionals in the display, processing, and analysis of medical image data. It enhances diagnostic workflows by integrating intelligent tools, streamlined accessibility, and advanced visualization capabilities, including specialized support for breast imaging.

    View brings together 2D imaging, basic 3D visualization and advance image analysis in a single, intuitive interface. This simplifies information access, improves workflow efficiency, and reduces the need for multiple applications.

    Key features include:

    • Smart Reading Protocol (SRP) which uses machine learning for creating and applying hanging protocols (HP)
    • AI workflow to support both DICOM Secondary Capture Object & DICOM Structured Report for displaying AI findings and enabling rejection/modification of the AI findings.
    • Displays 2D, 3D, and historical comparison exams in customizable layouts.
    • Enables smooth transitions between 2D and 3D views, either manually or as part of hanging protocols.
    • Advantage Workstation integration for deeper analysis through dedicated 3D applications.
    • Offers a full suite of measurement, annotation and segmentation tools for DICOM images.
    • Captures all measurements and annotations in a centralized findings panel.
    • Enhanced access for DICOM images stored in the cloud server.
    • Easy way to integrate with external systems using FHIRcast.
    • Better user experience by having a native MIP/MPR/Smart Segmentation/Volume Rendering
    • Seamless access to breast images through cloud with specific tools for Mammo images.
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "View" device (K253639) offer limited details regarding specific acceptance criteria and the studies conducted to prove device performance. The information is high-level and generalized.

    Based on the available text, here's the breakdown of what can and cannot be extracted:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria (e.g., minimum accuracy, sensitivity, specificity) or specific reported performance metrics for the device. It focuses on functional equivalence and verification/validation testing without presenting performance data.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "The Smart Reading Protocol function, which uses machine learning for creating and applying hanging protocols, was tested on various imaging modality datasets representative of the clinical scenarios where View is intended to be used."

    • Sample Size: "various imaging modality datasets" – The exact sample size (number of images, cases, or patients) is not specified.
    • Data Provenance: "representative of the clinical scenarios" – The country of origin and whether the data was retrospective or prospective are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide any information on the number or qualifications of experts used to establish ground truth for the test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not specify any adjudication method used for the test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document mentions a "comparison was performed between the predicate device (Universal Viewer) and the subject device (View) and showed that the devices are equivalent" for the Smart Reading Protocol function. However, this is a comparison between devices, not an MRMC study designed to assess human reader improvement with AI assistance. Therefore:

    • A specific MRMC comparative effectiveness study involving human readers with and without AI assistance is not described.
    • Consequently, an effect size for human reader improvement is not provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document implies that the Smart Reading Protocol (SRP) function, which uses machine learning, was tested. The statement "A comparison was performed between the predicate device (Universal Viewer) and the subject device (View) and showed that the devices are equivalent" regarding SRP suggests an evaluation of the algorithm's output. However, whether this testing was strictly standalone performance (algorithm only) versus integrated system performance is not explicitly detailed. Given the context of a 510(k) summary for a "Medical Image Management And Processing System," the testing likely covers the integrated system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document does not specify the type of ground truth used for its testing.

    8. The sample size for the training set

    The document does not provide any information on the sample size used for the training set.

    9. How the ground truth for the training set was established

    The document does not provide any information on how the ground truth for the training set was established.


    Summary of available information regarding acceptance criteria and study data:

    Unfortunately, the provided FDA 510(k) summary is very high-level and primarily focuses on justifying substantial equivalence based on technological characteristics and general verification/validation processes. It does not contain the detailed performance metrics, sample sizes, ground truth establishment methods, or reader study details typically found in more comprehensive clinical study reports. The approval is based on the device having "substantial equivalent technological characteristics" and being "as safe and as effective" as the predicate device.

    The only specific functional testing mentioned is for the "Smart Reading Protocol" using machine learning, which was compared to the predicate device to show equivalence. However, the details of this comparison (e.g., performance metrics, specific acceptance criteria for equivalence, or study design) are not included.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242925

    Validate with FDA (Live)

    Device Name
    MR Contour DL
    Manufacturer
    Date Cleared
    2025-04-01

    (189 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MR Contour DL generates a Radiotherapy Structure Set (RTSS) DICOM with segmented organs at risk which can be used by trained medical professionals. It is intended to aid in radiation therapy planning by generating initial contours to accelerate workflow for radiation therapy planning. It is the responsibility of the user to verify the processed output contours and user-defined labels for each organ at risk and correct the contours/labels as needed. MR Contour DL is intended to be used with images acquired on MR scanners, in adult patients.

    Device Description

    MR Contour DL is a post processing application intended to assist a clinician by generating contours of organ at risk (OAR) from MR images in the form of a DICOM Radiotherapy Structure Set (RTSS) series. MR Contour DL is designed to automatically contour the organs in the head/neck, and in the pelvis for Radiation Therapy (RT) planning of adult cases. The output of the MR Contour DL is intended to be used by radiotherapy (RT) practitioners after review and editing, if necessary, and confirming the accuracy of the contours for use in radiation therapy planning.

    MR Contour DL uses customizable input parameters that define RTSS description, RTSS labeling, organ naming and coloring. MR Contour DL does not have a user interface of its own and can be integrated with other software and hardware platforms. MR Contour DL has the capability to transfer the input and output series to the customer desired DICOM destination(s) for review.

    MR Contour DL uses deep learning segmentation algorithms that have been designed and trained specifically for the task of generating organ at risk contours from MR images. MR Contour DL is designed to contour 37 different organs or structures using the deep learning algorithms in the application processing workflow.

    The input of the application is MR DICOM images in adult patients acquired from compatible MR scanners. In the user-configured profile, the user has the flexibility to choose both the covered anatomy of input scan and the specific organs for segmentation. The proposed device has been tested on GE HealthCare MR data.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for MR Contour DL:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: MR Contour DL

    MetricOrgan Anatomy RegionAcceptance CriteriaReported Performance (Mean)Outcome
    DICE Similarity Coefficient (DSC)Small Organs (e.g., chiasm, inner-ear)≥ 50%67.4% - 98.8% (across all organs)Met
    Medium Organs (e.g., brainstem, eye)≥ 65%79.6% - 95.5% (across relevant organs)Met
    Large Organs (e.g., bladder, head-body)≥ 80%90.3% - 99.3% (across relevant organs)Met
    95th percentile Hausdorff Distance (HD95) ComparisonAll OrgansImproved or Equivalent to Predicate DeviceImproved or Equivalent in 24/28 organs analyzed; average HD95 of 4.7 mm (< predicate average)Met
    Likert Score (Reader Study)All OrgansMean Likert Score ≥ 3.0 (where 3 = good, some correction needed)3.0 - 4.5 (across all organs)Met

    Note: The HD95 values for specific organs are provided in Table 4 of the document, showing individual comparisons (Improved, Not-Improved, Equivalent, N/A). The overall performance for HD95 is summarized as met based on the text "improved or equivalent HD95 value in 24/28 of the organs analyzed and an average HD95 performance of 4.7 mm, which is smaller than the average corresponding HD95 values of the predicate device."

    2. Sample Sizes and Data Provenance

    • Test Set (Non-Clinical/Bench Testing):
      • Total Cases: 105 retrospectively collected exams.
        • Head/Neck: 50 cases (23 from independently collected cohorts, 27 separated from development data)
        • Pelvis: 55 cases (32 from independently collected cohorts, 23 separated from development data)
      • Data Provenance:
        • Country of Origin: USA (72% Head/Neck, 58% Pelvis) and Europe (NL 28% Head/Neck, UK 42% Pelvis)
        • Retrospective/Prospective: Retrospectively collected
    • Test Set (Clinical/Reader Study):
      • Total Cases: 70 cases (a subset of the non-clinical test data).
        • Head/Neck: 30 cases
        • Pelvis: 40 cases
      • Data Provenance: Same as non-clinical testing, as it was a subset.
    • Training Set: Not explicitly stated. The document mentions "separated from the development data cohorts before the models were trained," implying a training set existed but its size is not given.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Number of Experts: Three (3) board-certified radiation oncologists.
    • Qualifications: Two (2) from the USA, one (1) from Europe. All certified radiation oncologists. Experience level (e.g., "10 years experience") is not specified beyond "board certified."

    4. Adjudication Method (Test Set)

    • Non-Clinical/Bench Testing (Ground Truth Generation):
      1. Manual contours delineated by GEHC operators trained using international guidelines (DAHANCA, RTOG).
      2. Manual contours were revised (corrected and approved) by the three board-certified radiation oncologists.
      3. All three independently validated ground-truth contours were incorporated in the performance evaluation. This suggests a form of consensus or a voting mechanism, but the exact "adjudication" (e.g., 2+1, averaging) is not detailed. It implies that the final ground truth was derived from the combination of all three expert reviews.
    • Clinical/Reader Study:
      • Automated contours were scored by the three certified radiation oncologists.
      • Readers completed their assessments independently and were blinded to the results of other readers' assessments.
      • All three independently provided Likert Scores were incorporated in the performance evaluation. Similar to ground truth generation, the exact method of combining scores beyond "incorporating" is not specified, but the final reported value is a "Likert score MEAN."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was it done? Yes, a multi-reader study was conducted to assess the adequacy of the contours. The readers were radiologists providing an assessment (Likert score) of the AI-generated contours.
    • Effect Size of Human Readers' Improvement with AI vs. without AI Assistance: This study was structured to evaluate the adequacy of AI-generated contours for use in RT planning, with human readers providing an assessment of these pre-generated AI contours. It was not a comparative effectiveness study designed to measure the improvement in human reader performance when assisted by AI versus unassisted human performance (e.g., human-only contouring vs. AI-assisted human contouring). The study aimed to show that the AI's output is acceptable for human review and correction, not how much faster or more accurate humans become with the AI.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Was it done? Yes, the "Non-Clinical Testing" or "Bench Testing" section directly assesses the algorithm's standalone performance using DSC and HD95 metrics, comparing its output to expert-generated ground truth. The algorithm generates the initial contours, which are then evaluated for accuracy against the established ground truth.

    7. Type of Ground Truth Used

    • Non-Clinical/Bench Testing: Expert consensus (manual contours by trained operators, revised and approved by three board-certified radiation oncologists).
    • Clinical/Reader Study: Expert opinion/assessment (Likert scores provided by three board-certified radiation oncologists on the adequacy of the AI-generated contours).

    8. Sample Size for the Training Set

    • The sample size for the training set is not explicitly provided in the document. It only states that the test data cases (27 head/neck, 23 pelvis) were "separated from the development data cohorts before the models were trained."

    9. How the Ground Truth for the Training Set was Established

    • The method for establishing ground truth for the training set is not explicitly detailed. It can be inferred that it followed a similar process to the test set ground truth (manual contouring by trained operators, potentially reviewed by experts), as it mentions "development data cohorts," but the specifics are absent.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241300

    Validate with FDA (Live)

    Device Name
    ViewPoint 6
    Manufacturer
    Date Cleared
    2024-07-02

    (54 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images and generation of medical reports primarily for diagnostic ultrasound.

    ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint 6 displays to the user, can be used for diagnostic purposes.

    ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system.

    ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

    Device Description

    ViewPoint 6 is an image archiving and reporting software for medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data. ViewPoint 6 is for professional use only and enables quick diagnostic reporting with standardized terminology. It is designed with intuitive graphical user interfaces (GUIs) and is based on Microsoft Windows® with defined hardware requirements.

    Viewpoint 6 provides exam type specific reporting forms for various medical care areas. Forms are composed of different sections with data entry fields. The documentation can include measurements, exam findings, images and graphs. All data is saved in the ViewPoint 6 database and can be compiled to a professional report. Images and image sequences can be reviewed in the ViewPoint 6 display area based on user preference.

    ViewPoint 6 supports both a single workstation and a client/browser - server setup. The number of user licenses determines how many workstations in the network have concurrent access to the database. Access can be limited to read-only functionality.

    ViewPoint 6 software is a server-based application with client-server architecture, accessed via client computers or mobile devices as well as browser-based systems. Viewpoint 6 is installed on client provided servers within a hospital network.

    The software comes with features to view, annotate, measure, calculate, save and retrieve clinical data (including images via DICOM format) to support patient documentation and record keeping related to ultrasound image scans. Additionally, the software is available for patient administrative tasks such as appointment scheduling and exam billing.

    This product does not control or alter any of the medical devices providing data across the hospital network.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "ViewPoint 6." This type of document focuses on demonstrating substantial equivalence to a predicate device and typically does not contain detailed primary study results with acceptance criteria and specific performance metrics as would be found in a clinical study report. It primarily outlines the scope of V&V activities and voluntary standards adhered to.

    Based on the provided text, the following information can be extracted regarding the device performance and acceptance criteria:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding reported device performance values in terms of clinical accuracy (e.g., sensitivity, specificity). Instead, it states that:

    "Successful completion of design verification and validation testing was performed to confirm that software and user requirements have been met."

    This implies that the acceptance criteria are tied to the fulfillment of software and user requirements, which are assessed through various testing activities, but the specific numerical performance metrics are not detailed in this summary. The general statement about "Performance dependent on customer hardware but minimum hardware requirements for acceptable performance are defined in the System Requirements" suggests that performance acceptance is related to system specifications rather than clinical efficacy metrics.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify any sample sizes for a clinical test set, nor does it provide data provenance (e.g., country of origin, retrospective or prospective nature). The V&V activities mentioned (Risk Analysis, Requirements Reviews, Design Reviews, Testing on unit level, Integration testing, Performance testing, Safety testing) are typically software engineering and system-level tests and do not involve a clinical test set with patient data for performance evaluation in the context of diagnostic accuracy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    No clinical test set is described, and therefore, there is no information about the number or qualifications of experts used to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    No clinical test set is described, so no adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document states:

    "The similarities and differences between the subject device and the predicate device, were determined not to have a significant impact on the device's performance, the clinical performance, and the actual use scenarios. Therefore, the subject of this premarket submission, ViewPoint 6, did not require clinical studies to support substantial equivalence."

    This explicitly indicates that no MRMC comparative effectiveness study, or any clinical study, was conducted or deemed necessary for this 510(k) submission. Therefore, there is no information about AI assistance or its effect size on human reader improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is described as "an image archiving and reporting software for medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data." The indications for use state it "is not intended to be used as an automated diagnosis system." This confirms it's a tool to assist clinicians, not a standalone diagnostic algorithm. No standalone algorithmic performance study was conducted or mentioned.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    As no clinical studies were performed for diagnostic accuracy, no specific ground truth type (expert consensus, pathology, outcomes data, etc.) is mentioned. The V&V activities focused on software and system requirements.

    8. The sample size for the training set

    The document does not refer to any AI/ML components that would require a training set. Therefore, no training set sample size is provided.

    9. How the ground truth for the training set was established

    Since no training set is mentioned for AI/ML, there is no information on how its ground truth would be established.

    In summary:

    This 510(k) submission for ViewPoint 6 primarily relies on demonstrating substantial equivalence to a predicate device (ViewPoint 6 v6.12 K203677) through software validation and verification activities, adherence to voluntary standards, and a comparison of technological characteristics. It explicitly states that clinical studies were not required because the changes from the predicate device were not deemed to have a significant impact on clinical performance. Therefore, detailed information regarding clinical performance acceptance criteria, sample sizes for test or training sets, expert adjudication, or AI performance metrics is not present in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233698

    Validate with FDA (Live)

    Device Name
    True Enhance DL
    Date Cleared
    2024-04-11

    (146 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    True Enhance DL is a deep learning-based image processing method trained to estimate monochromatic, 50 keV GSI images. The algorithm is intended to improve the contrast of 120 kVp, single energy images of the body.

    This device is intended to provide non-quantitative, adjunct information and should not be interpreted without the original 120 kVp image.

    True Enhance DL may be used for patients of all ages.

    Device Description

    True Enhance DL is a deep learning-based image processing method for contrast enhanced images of the body obtained using the Revolution Ascend Family (K213938), which consists of multiple commercial configurations: Revolution Ascend Elite, Revolution Ascend Plus, and Revolution Ascend Select. True Enhance DL is intended to post-process single energy, 120 kVp images to output nonquantitative, adjunctive information with better contrast than single energy input data.

    True Enhance DL brings four deep leaning models that the user can choose depending on different contrast enhancement phases. These four models are CT Angiography, Arterial, Portal/Venous, and Delayed True Enhance DL.

    True Enhance DL is not intended to replace hardware based Monochromatic Imaging by Gemstone Spectral Imaging (GSI) technology or replicate GSI dual energy acquisitions. The device was trained to estimate monochromatic, 50 keV GSI images, and only enhances images from 120 kVp acquisitions on non-GSI Revolution Ascend systems.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implicit)Reported Device Performance
    Primary Goal: Improve the contrast of 120 kVp, single energy images of the body."The result of this reader study and head-to-head material comparison validated that True Enhance DL software provides additional benefit by improving contrast in the True Enhance output when compared to the original 120 kVp single energy images."
    Provide non-quantitative, adjunct information.The device's indication explicitly states it "is intended to provide non-quantitative, adjunct information."
    Not replace hardware-based Monochromatic Imaging by Gemstone Spectral Imaging (GSI) technology or replicate GSI dual energy acquisitions."True Enhance DL is not intended to replace hardware based Monochromatic Imaging by Gemstone Spectral Imaging (GSI) technology or replicate GSI dual energy acquisitions."
    Output estimable as 50 keV GSI images."The device was trained to estimate monochromatic, 50 keV GSI images."
    No new or different questions of safety or effectiveness compared to the predicate device."GE's quality system's design verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate."
    Achieve adequate image quality."The changes associated with True Enhance DL do not create a new Intended Use and represent technological characteristics that produce images that have demonstrated adequate image quality..."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated as a number of cases. The text mentions "sample clinical data" and "Additional representative clinical cases and anthropomorphic phantom cases."
    • Data Provenance: Retrospective. The study used "retrospectively collected representative clinical cases." The country of origin is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Number of Experts: Four.
    • Qualifications of Experts: "Four board certified radiologists." Specific years of experience are not mentioned.

    4. Adjudication Method for the Test Set:

    • The text does not explicitly state a formal adjudication method (e.g., 2+1, 3+1). It indicates that the four radiologists each provided a comparative assessment of image quality related to diagnostic use. This suggests individual reader assessment rather than a consensus-building adjudication process for ground truth. However, they were asked to "rate the contrast enhancement in the True Enhance DL series vs the native image series," which implies a comparative evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    • A MRMC-like study was done, as "four board certified radiologists" read the images.
    • Effect Size: The text states, "the readers were asked to rate the contrast enhancement in the True Enhance DL series vs the native image series" and "validated that True Enhance DL software provides additional benefit by improving contrast." However, a quantitative effect size of human readers' improvement with AI vs. without AI assistance is not provided in this summary. The focus was on the software's ability to improve contrast rather than a comparative effectiveness of human performance with and without the tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Yes, a standalone evaluation was conducted to assess image characteristics. The text mentions "Additional representative clinical cases and anthropomorphic phantom cases from a GSI system generating both single energy 120 kVp and 50 keV monochromatic images were evaluated for CT number in various anatomical regions to study image characteristics for different materials of the device output compared to 50 keV and 120 kVp reference images." This assesses the algorithm's output properties directly against a reference, which constitutes a standalone performance aspect.

    7. The Type of Ground Truth Used:

    • Expert Consensus / Reader Assessment: For the image quality and contrast improvement aspects, the subjective assessment of "four board certified radiologists" served as the ground truth.
    • Reference Images / Clinical Data: For the standalone evaluation, "50 keV and 120 kVp reference images" (likely derived from GSI systems with known energy characteristics) were used to study the algorithm's output. Clinical cases with "disease/pathology" were used, implying the presence of known conditions, although how these conditions served as "ground truth" for the AI's performance beyond simply being present in the data is not fully detailed.

    8. The Sample Size for the Training Set:

    • The sample size for the training set is not provided in this document. The text only states, "The device was trained to estimate monochromatic, 50 keV GSI images."

    9. How the Ground Truth for the Training Set Was Established:

    • The document implies that the training was based on "to estimate monochromatic, 50 keV GSI images." This suggests that 50 keV monochromatic GSI images (likely acquired from dual-energy CT scans, which serve as a form of ground truth for spectral decomposition) were used as the target output for the deep learning model during training. The process of generating these reference 50 keV GSI images themselves would involve the CT system's physics and reconstruction algorithms.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233728

    Validate with FDA (Live)

    Device Name
    SIGNA Champion
    Date Cleared
    2024-01-19

    (59 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

    Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/human reader performance study. The document is a 510(k) premarket notification for a Magnetic Resonance Diagnostic Device (SIGNA™ Champion).

    The relevant sections state:

    • "The subject of this premarket submission, the SIGNA™ Champion, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission."
    • "The sample clinical images demonstrate acceptable diagnostic image performance of the SIGNA™ Champion in accordance with the FDA Guidance 'Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices' issued on October 10, 2023. The image quality of the SIGNA™ Champion is substantially equivalent to that of the predicate device."

    This indicates that the FDA clearance for the SIGNA™ Champion MR system was based on demonstrating substantial equivalence to a previous predicate device (SIGNA™ Voyager) through non-clinical testing and the review of sample clinical images, rather than a prospective clinical study involving human readers and AI assistance for diagnostic tasks.

    Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI-assisted diagnostic device performance study because such a study was not conducted or reported in this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232346

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-27

    (84 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Expert Access with Remote Scanning is intended as a remote collaboration tool to view and review MR images, to remotely control MR Imaging Devices and to initiate MRI scans remotely.

    Digital Expert Access with Remote Scanning is a remote scan assistance solution which allows remote control of an MR Imaging Device including the ability to initiate a scan remotely. This access provides real time communication mechanisms between the remote and onsite users to facilitate the acquisition occurring on the device. Access must be granted by the onsite user operating the system. Images reviewed remotely are not for diagnostic use.

    Device Description

    Digital Expert Access is a variation of the Customer Remote Console cleared under K150193. It is a remote scan assistance solution designed to address the skill variability in technologists and their need for ondemand support by allowing them to interact directly with a remote expert connected to the hospital network. By enabling the collaboration between an Onsite Technologist and Remote Expert, Digital Expert Access helps the onsite technologist to seek guidance and real time support on scanning related queries including but not limited to training, procedure assessment, and scanning parameter management.

    Digital Expert Access with Remote Scanning introduces a feature that enables the Remote Expert to initiate a scan and make changes in real time during the scanning session. This remote scan feature is only available when Digital Expert Access is connected to a compatible GE HealthCare MRI system.

    Digital Expert Access with Remote Scanning enables the following capabilities for the Onsite Technologist and the Remote Expert:

      1. Collaborative session between an Onsite Technologist and Remote Expert
      1. Real-time scanner screen share and live annotation
    • ന് Remote console access and control
      1. Remote Scan Initiation

    Digital Expert Access with Remote Scanning is not intended for diagnostic use or patient safety-related management. This solution is not intended to be used by individuals who are not properly trained in the operation of GE HealthCare Medical Imaging systems. Digital Expert Access with Remote Scanning does not directly interface with any patients and requires the Onsite Technologist to be continuously present throughout the scanning procedure. Digital Expert Access with Remote Scanning does not acquire any MRI images, nor does it do any post image processing. All image acquisition and image processing is conducted by the GE HealthCare MRI system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a device called "Digital Expert Access with Remote Scanning." This document outlines the device's purpose, intended use, and substantial equivalence to predicate devices, but it does not contain information about specific acceptance criteria or an analytical study proving the device meets those criteria in terms of performance metrics like accuracy, sensitivity, or specificity.

    The document emphasizes that the device did not require clinical studies to support substantial equivalence. Instead, the focus was on non-clinical tests, primarily design verification and validation testing, to ensure proper implementation of design requirements and that no new safety or effectiveness issues were introduced.

    Therefore, many of the requested details about acceptance criteria for performance, sample sizes, ground truth establishment, expert involvement, and MRMC studies are not present in this document because the submission strategy relied on substantial equivalence based on technological similarity and non-clinical testing rather than performance studies against defined quantitative acceptance criteria.

    However, based on the information provided, here's what can be extracted and inferred regarding the "acceptance criteria" and "study":

    Acceptance Criteria (Inferred from Non-Clinical Testing and Device Functionality):

    Since no performance metrics are given, the acceptance criteria are implicitly related to the successful functionality and safety of the remote scanning feature. These would likely be qualitative or functional:

    Acceptance Criterion (Inferred)Reported Device Performance (Summary)
    1. Successful Remote Control of MR Imaging DeviceDemonstrated through non-clinical verification and validation testing on a subset of GE HealthCare MRI systems.
    2. Successful Remote Scan InitiationDemonstrated through non-clinical verification and validation testing on a subset of GE HealthCare MRI systems.
    3. Real-time Communication Mechanisms (Onsite & Remote Users)Device described as providing real-time communication mechanisms; functionality tested during design verification.
    4. Access Granting Mechanism by Onsite UserFunctionality described and tested: "Access must be granted by the onsite user circulating the system."
    5. No New Potential Safety Risks IntroducedConcluded based on design verification and validation testing; no unexpected test results or safety issues observed.
    6. Proper Implementation of Design RequirementsVerified through design control testing per GE HealthCare's quality system (e.g., Requirement Definition, Risk Analysis, Technical Design Reviews).
    7. Maintenance of Safety & EffectivenessConfirmed by design verification and validation testing, concluding it has not been affected.

    Study Information (Based on Non-Clinical Testing):

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: Not explicitly stated as a number of "cases" or "patients" in a clinical study. The testing was conducted on a "subset of GE HealthCare MRI systems" on the bench. This implies a sample of hardware configurations rather than patient data.
      • Data Provenance: Not applicable as it was non-clinical "bench" testing, not involving patient data or a specific country of origin for such data. It was performed internally by GE HealthCare.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not applicable. As this was non-clinical, functional testing, there was no "ground truth" in the clinical diagnostic sense established by experts. Testing would have involved engineers and quality assurance personnel verifying the system's intended functionality.
    3. Adjudication Method for the Test Set:

      • Not applicable for functional verification testing. Adjudication methods are typically used when comparing human interpretations or algorithmic outputs against a gold standard in diagnostic or clinical performance studies.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

      • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, Digital Expert Access with Remote Scanning did not require clinical studies to support substantial equivalence..." The device is a remote control and collaboration tool, not an AI-assisted diagnostic device meant to improve human reader performance.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

      • Not applicable in the context of performance. The device is inherently a "human-in-the-loop" collaboration tool. While its technical components were likely tested in isolation (standalone components), the overall device's "performance" is tied to its functional remote control and communication capabilities, not an algorithmic diagnostic output.
    6. The Type of Ground Truth Used:

      • For the non-clinical testing, the "ground truth" was the pre-defined design requirements and expected functional behavior of the device. This is typical for engineering verification and validation testing, where the "truth" is whether the system performs as designed and specified (e.g., does the remote expert successfully initiate a scan?).
    7. The Sample Size for the Training Set:

      • Not applicable. The document does not describe the device as employing machine learning or AI that would require a "training set" of data in the sense of a deep learning model. It is a control and communication software system.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable. As no training set for a machine learning model is mentioned, ground truth establishment for such a set is irrelevant to this submission.

    In summary, the FDA 510(k) clearance for this device was based on a demonstration of "substantial equivalence" to existing predicate devices through non-clinical design verification and validation testing, focusing on ensuring the new remote scanning feature did not introduce new safety or effectiveness issues, rather than requiring quantitative performance studies with clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223212

    Validate with FDA (Live)

    Device Name
    Precision DL
    Manufacturer
    Date Cleared
    2023-04-27

    (192 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Precision DL is a deep learning-based image processing method intended to enhance image quality of non-ToF PET images for clinical oncology purpose, using F-18 FDG. Precision DL may be used for patients of all ages.

    Device Description

    Precision DL is a deep learning-based image processing method intended for PET oncology 18F-FDG images obtained using the predicate device Omni Legend PET/CT system. Precision DL enhances the non-ToF Q.Clear images to have image quality performance similar to PET images obtained using ToF capable PET systems, including enhancement in image Contrast Recovery (CR), Contrast to Noise Ratio (CNR), and quantitation accuracy. Precision DL's training used clinical data from diverse clinical sites, accounting for relevant variations in patients and sites' protocols.

    Precision DL brings three deep learning models to provide users the choice between different strengths of contrast enhancement and noise reduction. The three models, Low, Medium, and High Precision DL, are trained such that the High Precision DL brings the highest contrast enhancement and lowest noise reduction, while the Low Precision DL brings the lowest contrast enhancement and highest noise reduction. Medium Precision DL brings contrast-noise tradeoff in between High and Low Precision DL.

    Precision DL is deployed within the acquisition and processing software of Omni Legend, for processing images produced using non-ToF Q.Clear image reconstruction.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study for Precision DL, based on the provided FDA 510(k) summary:

    Device: Precision DL (Deep Learning-based image processing for non-ToF PET images)

    Intended Use: Enhance image quality of non-ToF PET images for clinical oncology using F-18 FDG.


    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary describes performance improvements over non-ToF Q.Clear reconstruction. It does not explicitly state discrete acceptance criteria values but rather demonstrates general improvements in imaging metrics and equivalence to ToF images.

    Metric / Acceptance CriteriaReported Device Performance (Precision DL vs. non-ToF Q.Clear)
    Quantitation AccuracyImproved accuracy. Performance similar to ToF images.
    Contrast Recovery (CR)Enhanced. Performance similar to ToF images.
    Contrast-to-Noise Ratio (CNR)Enhanced. Performance similar to ToF images.
    Lesion DetectabilityExplicitly tested, and implied improvement given CR and CNR enhancements.
    Dose / Time ReductionExplicitly tested. (Specific results not detailed, but likely aims to show maintenance of quality with reduced dose/time, or enhanced quality at standard dose/time).
    Overall Image Quality (Clinical Assessment)Acceptable diagnostic results by board-certified radiologists, demonstrating acceptable image quality.
    Preference (Clinical Assessment)Readers preferred Precision DL images over unassisted images, and found them similar to Discovery MI ToF images.
    Safety and Effectiveness (Regulatory Acceptance)No new questions of safety or effectiveness, hazards, unexpected results, or adverse effects were identified compared to the predicate device. Substantially Equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Data for Bench Testing): 80 PET-CT exams.
      • 40 exams from an Omni Legend system.
      • 40 exams from Discovery MI systems (with hardware-based ToF).
    • Sample Size (Clinical Reader Study): Not explicitly stated precisely for the number of cases and images. It mentions "clinical cases of the same patients obtained on Discovery MI and Omni Legend with Precision DL."
    • Data Provenance: Multiple clinical sites in North America, Europe, and Israel. The data was "segregated, completely independent, and not used in any stage of the algorithm development, including training." This indicates prospective or retrospectively collected data used for testing only.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts (for Ground Truth related to phantom data): Not applicable for phantom data, as ground truth is known from inserted lesions.
    • Number of Experts (for Clinical Reader Study): "Board certified radiologists." The exact number is not explicitly stated in the summary, nor are their specific years of experience. However, the study involved reviews and preference questions by these experts.

    4. Adjudication Method

    • The summary mentions a "clinical reader study" where "board certified radiologists... answered blinded preference questions comparing clinical cases." This suggests individual reader assessments were aggregated, but it does not explicitly state an adjudication method like 2+1 or 3+1 for resolving discrepancies in diagnostic findings. The focus appears to be on overall image quality and preference rather than a specific diagnostic consensus for each case.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Yes, a clinical reader study was performed, which included "board certified radiologists" reviewing "clinical cases." They answered "blinded preference questions comparing clinical cases of the same patients obtained on Discovery MI and Omni Legend with Precision DL."
    • Effect Size of Human Readers with AI vs. Without AI Assistance: The summary states, "The results of the reader study and preference questions support the determination of substantial equivalence. All readers attested that their assessments of Precision DL demonstrated acceptable diagnostic results." While it indicates positive results and physician acceptance, it does not quantify an effect size of how much human readers improved their performance (e.g., in diagnostic accuracy, confidence, or reduced read time) with AI assistance compared to reading without AI assistance (i.e., using only non-ToF Q.Clear images). The study primarily focused on image quality acceptability and preference.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, a standalone performance assessment was conducted as part of the "additional engineering bench testing." This included quantitative assessments using both clinical and phantom data for metrics such as Quantitation Accuracy, Contrast Recovery, Contrast-to-Noise Ratio, Lesion Detectability, and Dose/Time Reduction. This part of the testing directly evaluated the algorithm's output (processed images) against established ground truths/references.

    7. Type of Ground Truth Used

    • For Bench Testing (Quantitative Metrics):
      • Phantom Data: Known quantitation from inserted lesions of known size, location, and contrast.
      • Clinical Data: Discovery MI's ToF PET images served as a reference for comparison, implying they are considered the gold standard for high-quality images that Precision DL aims to emulate.
    • For Clinical Reader Study: The "ground truth" here is implied to be the expert consensus on acceptable diagnostic quality and preference rather than a definitive diagnosis based on pathology or long-term outcomes for each case. The goal was to confirm that the enhanced images retained or improved diagnostic acceptability.

    8. Sample Size for the Training Set

    • The summary states, "Precision DL's training used clinical data from diverse clinical sites, accounting for relevant variations in patients and sites' protocols."
    • However, the specific sample size for the training set is not provided in the given text.

    9. How the Ground Truth for the Training Set Was Established

    • The summary indicates that Precision DL "is trained to enhance non-ToF images to have IQ performance similar to ToF images." This implies that the ground truth for training would likely be high-quality ToF PET images (potentially from a system like Discovery MI) that the algorithm was designed to mimic or achieve certain quality metrics aligned with ToF performance.
    • The text doesn't detail the process of establishing ground truth for individual images within the training set, but it's reasonable to infer a reference standard from ToF images was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230807

    Validate with FDA (Live)

    Date Cleared
    2023-04-20

    (28 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Deep Learning Image Reconstruction software is a deep learning based reconstruction method intended to produce cross-sectional images of the head and whole body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, (Volumetric), and Cardiac acquisitions, for all ages.

    Deep Learning Image Reconstruction software can be used for head, whole body, cardiac, and vascular CT applications.

    Device Description

    Deep Learning Image Reconstruction is an image reconstruction method that uses a dedicated Deep Neural Network (DNN) that has been designed and trained specifically to generate CT Images to give an image appearance, as shown on axial NPS plots, similar to traditional FBP images while maintaining the performance of ASiR-V in the following areas: image noise (pixel standard deviation), low contrast detectability, high-contrast spatial resolution, and streak artifact suppression.

    The images produced are branded as "TrueFidelity™ CT Images". Reconstruction times with Deep Learning Image Reconstruction support a normal throughput for routine CT.

    The deep learning technology is integrated into the scanner's existing raw data-based image reconstruction chain to produce DICOM compatible "TrueFidelity™ CT Images".

    The system allows user selection of three strengths of Deep Learning Image Recon: Low, Medium or High. The strength selection will vary with individual users' preference for the specific clinical need.

    The DLR algorithm was modified on the Revolution CT/Apex platform for improved reconstruction speed and image quality and cleared in February 2022 with K213999. The same modified DLIR is now being ported to Revolution EVO (K131576) /Revolution Maxima (K192686), Revolution Ascend (K203169, K213938) and Discovery CT750 HD family CT systems including Discovery CT750 HD, Revolution Frontier and Revolution Discovery CT (K120833).

    AI/ML Overview

    The provided text describes that the Deep Learning Image Reconstruction software was tested for substantial equivalence to a predicate device (K213999). The study performed was largely an engineering bench testing, comparing various image quality metrics between images reconstructed with Deep Learning Image Reconstruction (DLIR) and ASiR-V from the same raw datasets.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The text indicates that the device aims to maintain the performance of ASiR-V in specific areas while offering an image appearance similar to traditional FBP images. The "acceptance criteria" can be inferred from the list of image quality metrics evaluated, with the performance goal being comparable or improved relative to ASiR-V.

    Acceptance Criteria (Implied Goal: Performance comparable to or better than ASiR-V)Reported Device Performance (Implied: Met acceptance criteria, no adverse findings)
    Image noise (pixel standard deviation)DLIR maintains ASiR-V performance.
    Low contrast detectability (LCD)Evaluation performed. (Implied: Met acceptance criteria)
    High-contrast spatial resolution (MTF)Evaluation performed. (Implied: Met acceptance criteria)
    Streak artifact suppressionDLIR maintains ASiR-V performance.
    Spatial Resolution, longitudinal (FWHM slice sensitivity profile)Evaluation performed. (Implied: Met acceptance criteria)
    Noise Power Spectrum (NPS) and Standard Deviation of noiseEvaluation performed (NPS plots similar to FBP). (Implied: Met acceptance criteria)
    CT Number UniformityEvaluation performed. (Implied: Met acceptance criteria)
    CT Number AccuracyEvaluation performed. (Implied: Met acceptance criteria)
    Contrast to Noise (CNR) ratioEvaluation performed. (Implied: Met acceptance criteria)
    Artifact analysis (metal objects, unintended motion, truncation)Evaluation performed. (Implied: Met acceptance criteria)
    Pediatric Phantom IQ Performance EvaluationEvaluation performed. (Implied: Met acceptance criteria)
    Low Dose Lung Cancer Screening Protocol IQ Performance EvaluationEvaluation performed. (Implied: Met acceptance criteria)
    Image appearance (NPS plots similar to traditional FBP)Designed to give an image appearance, as shown on axial NPS plots, similar to traditional FBP images.
    No additional risks/hazards, warnings, or limitationsNo additional hazards were identified, and no unexpected test results were observed.
    Maintains normal throughput for routine CTReconstruction times with Deep Learning Image Reconstruction support a normal throughput for routine CT.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The text states "the identical raw datasets obtained on GEHC's Revolution Ascend, Revolution Frontier and Discovery CT750 HD systems". However, the number of cases or specific sample size for these datasets is not explicitly stated.
    • Data Provenance: The raw datasets were "obtained on GEHC's Revolution Ascend, Revolution Frontier and Discovery CT750 HD systems". The country of origin is not specified, and it is stated that the study used retrospective raw datasets (i.e., existing data, not newly acquired data for the study).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The provided text focuses on engineering bench testing and image quality metrics. It does not mention the use of experts to establish ground truth for the test set or their qualifications. The evaluation primarily relies on quantitative image quality metrics.

    4. Adjudication Method for the Test Set

    Since experts were not explicitly used to establish ground truth, there is no mention of an adjudication method for the test set in the provided text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    An MRMC comparative effectiveness study was not performed according to the provided text. The study focused on technical image quality comparisons at the algorithm level, not human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done. The study described is primarily a standalone evaluation of the algorithm's image quality output (e.g., noise, resolution, artifacts, detectability) when compared to images reconstructed with ASiR-V from the same raw data.

    7. The Type of Ground Truth Used

    The "ground truth" for the test set was essentially:

    • Quantitative Image Quality Metrics: Performance relative to ASiR-V for metrics like image noise, LCD, spatial resolution, streak artifact suppression, CT uniformity, CT number accuracy, CNR, spatial resolution (longitudinal), NPS, and artifact analysis.
    • Reference Image Appearance: The stated goal was an image appearance similar to traditional FBP images shown on axial NPS plots.

    There is no mention of pathology, expert consensus on clinical findings, or outcomes data being used as ground truth for this particular substantial equivalence study.

    8. The Sample Size for the Training Set

    The text states that the Deep Neural Network (DNN) is "trained on the CT scanner" and models the scanned object using "information obtained from extensive phantom and clinical data." However, the specific sample size for the training set is not provided.

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set is implicitly established through the "extensive phantom and clinical data" mentioned as being used to train the DNN. The text indicates the DNN is trained to model noise propagation and identify noise characteristics to remove it, and to generate images with an appearance similar to traditional FBP while maintaining ASiR-V performance. This suggests the training involves learning from "ground truth" as defined by:

    • Reference Image Quality: Likely images reconstructed with proven methods (e.g., FBP, ASiR-V) or images from phantoms with known properties.
    • Noise Characteristics: The DNN is trained to understand and model noise.

    However, the specific methodology for establishing this ground truth for the training data (e.g., expert annotation, simulated data, pathology confirmed disease) is not detailed in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 26