K Number
K203677

Validate with FDA (Live)

Device Name
ViewPoint 6
Date Cleared
2021-01-15

(29 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.

ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint 6 displays to the user, can be used for diagnostic purposes.

ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

Device Description

ViewPoint 6 is an image archiving and reporting software for medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data. It provides different calculations and tools to allow for the assessment of the images and data.

ViewPoint 6 is for professional use only and enables quick diagnostic reporting with standardized terminology. It has an intuitive graphical user interface (GUI) and is based on Microsoft Windows® with defined hardware requirements for the user to install on their computer.

ViewPoint 6 provides exam type specific reporting forms for various medical care areas. Forms are composed of different sections with data entry fields. The documentation can include measurements, exam findings, images, and graphs. All data is saved in the ViewPoint 6 database and can be compiled to a professional report. Images and image sequences can be reviewed in the ViewPoint 6 display area based on user preference. ViewPoint 6 supports both a single workstation and a client - server setup. The number of user licenses determines how many workstations in the network have concurrent access to the database. Access can be limited to read - only functionality

AI/ML Overview

The document provided is a 510(k) Premarket Notification Submission for the GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC's ViewPoint 6 device (K203677). This submission primarily focuses on establishing substantial equivalence to a predicate device (ViewPoint 6, K192917) and a reference device (EchoPAC Software Only, K200852).

Based on the provided information, the 510(k) submission states that clinical studies were not required or performed to support the substantial equivalence of ViewPoint 6. Therefore, there is no detailed study data, acceptance criteria, or performance metrics within this document that would typically be associated with a clinical trial or performance evaluation study specifically proving the device meets certain acceptance criteria through direct testing.

The conclusion explicitly states:
"The subject of this premarket submission, ViewPoint 6, did not require clinical studies to support substantial equivalence."

"GE Healthcare considers the ViewPoint 6 to be as safe, as effective, and performance is substantially equivalent to the predicate device."

This means that the manufacturer is asserting the new ViewPoint 6 is functionally equivalent to the previously cleared ViewPoint 6 (K192917) and EchoPAC Software Only (K200852) based on its intended use, technological characteristics, and non-clinical performance (e.g., meeting voluntary standards, software development quality assurance measures).

Therefore, it is not possible to provide the requested information regarding acceptance criteria, reported device performance, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details from the provided text, as such a study was not conducted or reported in this submission.

The document highlights the device's adherence to voluntary standards and internal quality assurance measures in lieu of clinical performance testing for this specific 510(k) submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.

GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226

Re: K203677

Trade/Device Name: ViewPoint 6 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 16, 2020 Received: December 17, 2020

Dear Tracey Ortiz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

January 15, 2021

{1}------------------------------------------------

2

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203677

Device Name ViewPoint 6

Indications for Use (Describe)

ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purpostic interpretation of images, electronic documentations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.

ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images, The medical images, which ViewPoint 6 displays to the used for diagnostic purposes.

ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

GE Healthcare

510(k) Premarket Notification Submission

Image /page/3/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around its perimeter, giving it a classic and recognizable appearance.

510(k) Summary

K203677

In accordance with 21 CFR 807.92 the following summary of information is provided: Date: December 16, 2020

Submitter:GE Medical Systems Ultrasound and Primary Care Diagnostics
9900 Innovation Drive
Wauwatosa, WI 53226
  • Primary Contact Person: Tracey Ortiz Regulatory Affairs Director GE Healthcare T:(262)470-1003
    Secondary Contact Bryan Behn Regulatory Affairs Leader Person: GE Healthcare
Trade Name:ViewPoint 6
Common/Usual Name:PACS-Picture archiving and communications system
Classification Names:Class II
Product Code:LLZ, Picture archiving and communications system, 21 CFR892.2050, 90-LLZ
Predicate Device:ViewPoint 6 (K192917)
Product Code:LLZ, Picture archiving and communications system, 21 CFR892.2050, 90-LLZ
Reference Device:EchoPAC Software Only (K200852)
Product Code:LLZ, Picture archiving and communications system, 21 CFR892.2050, 90-LLZ

Device Description:

ViewPoint 6 is an image archiving and reporting software for medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data. It provides different calculations and tools to allow for the assessment of the images and data.

ViewPoint 6 is for professional use only and enables quick diagnostic reporting with standardized terminology. It has an intuitive graphical user interface (GUI) and is based on Microsoft Windows® with defined hardware requirements for the user to install on their computer.

ViewPoint 6 provides exam type specific reporting forms for various medical care areas. Forms are composed of different sections with data entry fields. The documentation

{4}------------------------------------------------

GE Healthcare 510(k) Premarket Notification Submission

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

can include measurements, exam findings, images, and graphs. All data is saved in the ViewPoint 6 database and can be compiled to a professional report. Images and image sequences can be reviewed in the ViewPoint 6 display area based on user preference. ViewPoint 6 supports both a single workstation and a client - server setup. The number of user licenses determines how many workstations in the network have concurrent access to the database. Access can be limited to read - only functionality

Intended Use/Indication for Use:

ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.

ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint 6 displays to the user, can be used for diagnostic purposes.

ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

Technology:

ViewPoint 6 employs the same fundamental scientific technology as its predicate device.

Determination of Substantial Equivalence:

Comparison to Predicate Devices

The proposed ViewPoint 6 system is substantially equivalent to the ViewPoint 6 (K192917) with regards to intended use, capabilities, technological characteristics, safety and effectiveness.

FeaturePredicate DeviceViewPoint6 cleared(K192917)Proposed DeviceViewPoint 6Discussion ofDifferences
Intended useIdenticalIdenticaln/a
Indications for UseIdenticalIdenticaln/a
Contra-indicationsIdenticalIdenticaln/a
Patient populationIdenticalIdenticaln/a
Environment of useIdenticalIdenticaln/a
Human FactorsIdenticalIdenticaln/a
DesignIdenticalIdenticaln/a
PerformanceIdenticalIdenticaln/a
Standards metIdenticalIdenticaln/a
Compatibility withother devices/ConnectivityAdded:• Full Image Export CVUS - SendImages for IACEL Accreditation• Migrate patient and study datafrom EchoPAC to ViewPoint 6• Encapsulated PDF DocumentExport• Export to standard DICOM DIRSimilar function has beencleared in referencedevice EchoPACsoftware only (K200852)The difference has noimpact to ViewPoint 6safety and effectiveness.Other capabilities aresimilar to predicate.
Image Editing andManagementAdded:• Export of anonymizedmulti-frames available• Message upon openingseveral exams in EchoPACregarding DICOM SR dataEquivalentThe difference hasno impact toViewPoint 6 safetyand effectiveness.
Report editing andmanagementAdded:• User configured units ofmeasurements• User Masks - New ControlTypes• Keep list item and fieldvisibility during update• Additional Exam LockedStates• For radio buttons, 'hide inreport' shall work individually• Option to calculate and printnext appointment date• Enhancements forEchocardiography• Extend OB/Gyn ReportingContent• Enhanced Fetal Anatomy andPlacenta Section• Enhanced Reporting inMultiple Pregnancies• Vascular EnhancementsEquivalentThe difference hasno impact toViewPoint 6 safetyand effectiveness.
CalculationIdenticaln/a
Data MiningIdenticaln/a
Coding forBillingAdded:• Unify and simplify medicalcodingEquivalentThe difference has noimpact to ViewPoint 6safety and effectiveness.
Data Privacyand SecurityIdenticaln/a
SchedulerIdenticaln/a

{5}------------------------------------------------

GE Healthcare 510(k) Premarket Notification Submission

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance.

{6}------------------------------------------------

Image /page/6/Picture/22 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by a swirling, wave-like pattern, also in blue. The logo is simple, clean, and easily recognizable.

510(k) Premarket Notification Submission

Summary of Non-Clinical Tests:

ViewPoint 6 and its applications comply with voluntary standards:

  • IEC 62366-1:2015 Medical devices-Application of usability engineering to medical devices
  • IEC 62304:2006+A1:2015, Medical device software-Software life cycle process ●
  • NEMA PS 3.1-3.20 (2016), Digital Imaging and Communications in Medicine ● (DICOM) Set. (Radiology)
  • ISO 14971:2019 Medical Devices-Application of risk management to medical devices
  • IEC 82304-1:2016, Health software-General requirements for product safety ●

The following quality assurance measures are applied to the development of the system:

  • Risk Analysis
  • Requirements Reviews ●
  • Design Reviews ●
  • o Testing on unit level (Module verification)
  • Integration testing (System verification) 0
  • o Performance testing (Verification & Validation)
  • Safety testing (Verification)

Summary of Clinical Tests:

The subject of this premarket submission, ViewPoint 6, did not require clinical studies to support substantial equivalence.

Conclusion:

GE Healthcare considers the ViewPoint 6 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

N/A