(122 days)
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages.
Computed Tomography (CT) is an indispensable imaging modality in clinical diagnostics due to its ability to provide detailed cross-sectional images of anatomical structures. However, achieving high spatial resolution remains a persistent challenge, particularly in applications requiring the visualization of fine structural details such as in inner auditory canal imaging, vascular studies, lung imaging, and bone microarchitecture analysis. Reconstruction techniques that attempt to boost spatial resolution typically also would amplify the noise which often result in tradeoffs to be made between resolution and noise.
As part of the continuous innovation to solve the above challenges in CT imaging, GEHC developed a deep learning-based CT reconstruction algorithm specifically designed for high contrast lung and bone imaging, which is the subject of this premarket notification. This reconstruction algorithm, marketed under the name True Definition DL (TDDL), is an additional user-selectable recon option specifically designed for lung and bone imaging. It aims to enhance spatial resolution for both in-plane and cross-plane directions. This algorithm is incorporated into the reconstruction chain of the Revolution CT /Apex Family CT systems including Revolution CT/Revolution CT ES, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select all cleared under (K213715) and Revolution Vibe (K250941).
True Definition DL offers three strengths that the user can choose depending on enhancement preferences. The benefits provided by this enhancement include improved spatial resolution measured by MTF.
Here's a breakdown of the acceptance criteria and study details for the "True Definition DL" device, as extracted from the provided FDA 510(k) clearance letter:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Enhance spatial resolution for both in-plane and cross-plane directions in lung and bone imaging. | Improves the in-plane and cross-plane resolution for body (lung and bone) scans as well as bony structures in the head without significantly increasing the image noise, in comparison with a sharper resolution kernel. |
| Provide diagnostic value by improving high contrast spatial resolution in routine bone (including inner ears and spines) and lung exams. | Validated through a reader study; improves confidence in the assessment of high contrast structures in these types of images. |
| Improve confidence in the assessment of high contrast structures in lung and bone images. | Validated through a reader study; improves confidence in the assessment of high contrast structures in these types of images. |
| Not present algorithm-induced artifacts or enhancements not supported by underlying anatomies. | Readers confirmed that the images enhanced by the subject device do not present algorithm-induced artifacts or enhancements not supported by underlying anatomies. |
| Consistently enhance anatomical edges without introducing spurious structures. | Results indicated that True Definition DL consistently enhanced anatomical edges without introducing spurious structures. |
| Preserve noise texture under realistic dose conditions. | Results indicated that True Definition DL preserved noise texture under realistic dose conditions. |
| Maintain strong correlation with input and target images across all frequency bands. | Results indicated that True Definition DL maintained strong correlation with input and target images across all frequency bands. |
Study Details
-
Sample size used for the test set and the data provenance:
- Sample size: Not explicitly stated for the clinical reader study, only referred to as "sample clinical data."
- Data provenance: Not explicitly stated (e.g., country of origin or retrospective/prospective).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of experts: Not explicitly stated.
- Qualifications: "US board-certified radiologists."
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated. The document mentions "readers confirmed," implying a consensus or majority opinion, but the specific adjudication method is not described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A reader study was performed ("clinical testing was carried out in the form of a reader study").
- The document states that the study "validated that True Definition DL provides diagnostic value by improving high contrast spatial resolution... and improves confidence in the assessment of high contrast structures."
- However, the document does not explicitly state that it was an MRMC comparative effectiveness study comparing human readers with AI vs. without AI assistance. It primarily focuses on the diagnostic value and confidence with the use of the images enhanced by True Definition DL.
- No effect size or specific metrics quantifying improvement with AI assistance are provided.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone evaluation of the algorithm's performance was indicated through "robustness of the True Definition DL algorithm" testing. This included assessment of hallucination risks, enhancement of anatomical edges, preservation of noise texture, and correlation with input/target images, suggesting an algorithm-level assessment.
- Bench testing on phantoms also measured the performance of True Definition DL against industry-standard IQ metrics and traditional metrics without direct human interaction.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the clinical reader study, the "ground truth" was established by expert opinion/consensus of US board-certified radiologists who evaluated the images and confirmed diagnostic value and lack of artifacts.
- For bench testing, "well established test methods using industry-standard IQ metrics" provided quantifiable reference points for performance.
-
The sample size for the training set:
- Not explicitly stated.
-
How the ground truth for the training set was established:
- Not explicitly stated. The document only mentions that True Definition DL is a "deep learning-based CT reconstruction algorithm," implying a training phase, but provides no details on how the training data was curated or its ground truth established.
FDA 510(k) Clearance Letter - True Definition DL
Page 1
March 23, 2026
GE Healthcare Japan Corporation
℅ Yonghui Han
Senior Regulatory Affairs Leader
7-127, Asahigaoka, 4-Chome
Hino-Shi, TOKYO, 191-8503
JAPAN
Re: K253686
Trade/Device Name: True Definition DL
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed tomography x-ray system
Regulatory Class: Class II
Product Code: JAK
Dated: February 20, 2026
Received: February 20, 2026
Dear Yonghui Han:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K253686 - Yonghui Han Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K253686 - Yonghui Han Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253686
Device Name
True Definition DL
Indications for Use (Describe)
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) Premarket Notification Submission - True Definition DL
510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
Date: February 17, 2026
Submitter: GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha, Wisconsin 53188
Primary Contact: Yonghui Han
Sr. Regulatory Affairs Leader
Phone: 1 (262) 225-1914
Email: Yonghui.han@gehealthcare.com
Secondary Contacts: Helen Peng
Sr. Regulatory Affairs Director
Phone: +1 (262) 424-8222
Email: hong.peng@gehealthcare.com.
Device Trade Name: True Definition DL
Device Classification Class II
Regulation Number/ Product Code: 21 CFR 892.1750 Computed tomography x-ray system / JAK
Predicate Device Information
Device Name: Deep Learning Image Reconstruction
Manufacturer: GE Medical Systems, LLC
510(k) Number: K213999
Regulation Number/ Product Code: 21 CFR 892.1750 Computed tomography x-ray system / JAK
Device Description
Computed Tomography (CT) is an indispensable imaging modality in clinical diagnostics due to its ability to provide detailed cross-sectional images of anatomical structures. However, achieving high spatial resolution remains a persistent challenge, particularly in applications requiring the visualization of fine structural details such as in inner auditory canal imaging, vascular studies, lung imaging, and bone microarchitecture analysis. Reconstruction techniques that attempt to boost spatial resolution typically also would amplify the noise which often result in tradeoffs to be made between resolution and noise.
Page 6
As part of the continuous innovation to solve the above challenges in CT imaging, GEHC developed a deep learning-based CT reconstruction algorithm specifically designed for high contrast lung and bone imaging, which is the subject of this premarket notification. This reconstruction algorithm, marketed under the name True Definition DL (TDDL), is an additional user-selectable recon option specifically designed for lung and bone imaging. It aims to enhance spatial resolution for both in-plane and cross-plane directions. This algorithm is incorporated into the reconstruction chain of the Revolution CT /Apex Family CT systems including Revolution CT/Revolution CT ES, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select all cleared under (K213715) and Revolution Vibe (K250941).
True Definition DL offers three strengths that the user can choose depending on enhancement preferences. The benefits provided by this enhancement include improved spatial resolution measured by MTF.
Intended Use
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement.
Indications for Use
True Definition DL is a deep learning based CT reconstruction method intended for high contrast spatial resolution enhancement for bone and lung imaging. True Definition DL may be used for patients of all ages.
Technological Characteristics
Since the True Definition DL is a recon option for the Revolution CT/Apex system, it utilizes the same hardware and software platform for reconstruction. The table below summarizes the substantive feature/technological differences between the predicate device, and the proposed device. The changes described below do not change the fundamental technology of the predicate device or host system and do not raise any new questions of safety and effectiveness.
Page 7
| Specification | Predicate DeviceDeep Learning Image Reconstruction(K213999) | Proposed DeviceTrue Definition DL |
|---|---|---|
| Patient Population | Patients of all ages | Same |
| Reconstruction Matrix | 512 x 5121024 x 1024 | 1024 x 1024 |
| Anatomies | Head, Body (including bone and lung), Cardiac | Bone and lung |
| Hardware | DLIR is a recon option implemented on the operator console for Revolution Apex systems. | Same |
| Clinical workflow | Built into routine clinical workflow with 3 strengths to select from (Low, Medium, High) | Same |
Determination of Substantial Equivalence
True Definition DL has successfully completed the design control testing per GE HealthCare's quality system. No additional hazards were identified, and no unexpected test results were observed. True Definition DL was designed under GE HealthCare's QMS per the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the successfully completed design verification & validation, as well as the extensive bench testing is sufficient for FDA's substantial equivalence determination.
The following quality assurance measures have been applied to the development of the system:
- Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- Code Review
- Software Unit Implementation
- Software Integrations and Integration Testing
- System Testing
- Safety Testing (Verification)
- Image Performance Testing (Verification)
- Simulating Use Testing (Validation)
- Software Release
The testing and results did not raise different questions of safety and effectiveness than associated with predicate device.
Summary of Non-Clinical Testing
A range of bench testing on phantoms was performed to measure the performance of True Definition DL. The tests include well established test methods using industry-standard IQ metrics. The test data demonstrated acceptable overall image quality and substantially equivalent performance of True Definition DL to the predicate measured on the Revolution CT/Apex scanner's traditional metrics.
Page 8
The result of the bench testing showed that True Definition DL improves the in-plane and cross-plane resolution for body ( lung and bone)scans as well as bony structures in the head acquired from Revolution CT/Apex without significantly increasing the image noise, in comparison with a sharper resolution kernel.
To evaluate the algorithm robustness of the True Definition DL algorithm, GEHC conducted qualitative and quantitative test utilizing the entire image reconstruction pipeline. The testing also included assessment of hallucination risks and their mitigations with the results indicating that True Definition DL consistently enhanced anatomical edges without introducing spurious structures, preserved noise texture under realistic dose conditions, and maintained strong correlation with input and target images across all frequency bands.
Summary of Clinical Testing
The clinical testing was carried out in the form of a reader study of sample clinical data processed using True Definition DL software. The images were evaluated by US board-certified radiologists.
The result of this reader study validated that True Definition DL provides diagnostic value by improving high contrast spatial resolution in routine bone (including inner ears and spines) and lung exams and improves confidence in the assessment of high contrast structures in in these types of images.
In addition, the readers confirmed that the images enhanced by the subject device do not present algorithm-induced artifacts or enhancements not supported by underlying anatomies.
Substantial Equivalence
True Definition DL was developed under GEHC's quality system. Design verification, along with bench testing and the clinical reader study provided in this submission demonstrates that True Definition DL is substantially equivalent to the predicate device. Verification & Validation testing and risk management processes did not raise any new questions of safety or effectiveness, or unexpected results, or adverse effects stemming from the changes. GE HealthCare believes that True Definition DL is substantially equivalent to the predicate, hence it is safe and effective for its intended use.
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.