Search Filters

Search Results

Found 644 results

510(k) Data Aggregation

    K Number
    K251805
    Date Cleared
    2025-10-15

    (125 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250662
    Device Name
    Bunkerhill MAC
    Manufacturer
    Date Cleared
    2025-10-03

    (212 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250766
    Device Name
    LungQ 4
    Manufacturer
    Date Cleared
    2025-10-02

    (203 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K251645
    Date Cleared
    2025-09-26

    (120 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Self-Propelled CT Scan Base Kit, CGBA-035A:
    The movable gantry base unit allows the Aquilion ONE (TSX-308A) system to be installed in the same procedure room as the INFX-8000C system, enabling coordinated clinical use within a shared workspace. This configuration provides longitudinal positioning along the z-axis for image acquisition.

    Alphenix, INFX-8000C/B, INFX-8000C/S, V9.6 with Calculated DAP:
    This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. The Calculated Dose Area Product (DAP) feature provides an alternative method for determining dose metrics without the use of a physical area dosimeter. This function estimates the cumulative reference air kerma, reference air kerma rate, and cumulative dose area product based on system parameters, including X-ray exposure settings, beam hardening filter configuration, beam limiting device position, and region of interest (ROI) filter status. The calculation method is calibration-dependent, with accuracy contingent upon periodic calibration against reference measurements.

    Device Description

    The Alphenix 4DCT is composed of the INFX-8000C interventional angiography system and the dynamic volume CT system, Aquilion ONE, TSX-308A. This combination enables patient access and efficient workflow for interventional procedures. Self-Propelled CT Scan Base Kit, CGBA-035A, is an optional kit intended to be used in conjunction with an Aquilion ONE / INFX-8000C based IVR-CT system. This device is attached to the Aquilion ONE CT gantry to support longitudinal movement and allow image acquisition in the z-direction (Z-axis), both axial and helical. When this option is installed, the standard CT patient couch is replaced with the fixed catheterization table utilized by the interventional x-ray system, INFX-8000C. The Self-Propelled CT Scan Base Kit, CGBA-035A, will be used as part of an Aquilion ONE / INFX-8000C based IVR-CT system. Please note, the intended uses of the Aquilion ONE CT System and the INFX-8000C Interventional X-Ray System remain the same. There have been no modifications made to the imaging chains in these FDA cleared devices and the base system software remains the same. Since both systems will be installed in the same room and to prevent interference during use, system interlocks have been incorporated into the systems.

    The Alphenix, INFX-8000C/B, INFX-8000C/S, V9.6 with Calculated DAP, is an interventional x-ray system with a ceiling suspended C-arm as its main configuration. Additional units include a patient table, x-ray high-voltage generator and a digital radiography system. The C-arms can be configured with designated x-ray detectors and supporting hardware (e.g. x-ray tube and diagnostic x-ray beam limiting device). The INFX-8000C system incorporates a Calculated Dose Area Product (DAP) feature, which provides an alternative method for determining dose metrics without the use of a physical area dosimeter. This function estimates the cumulative reference air kerma, reference air kerma rate, and cumulative dose area product based on system parameters, including X-ray exposure settings, beam hardening filter configuration, beam limiting device position, and region of interest (ROI) filter status. The calculation method is calibration-dependent, with accuracy contingent upon periodic calibration against reference measurements.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250928
    Date Cleared
    2025-09-08

    (164 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniTom Elite with PCD computed tomography (CT) system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck.

    The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast. The PCD CT system has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI).

    Device Description

    The subject OmniTom Elite with PCD Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K233767). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    OmniTom Elite system with photon counting detector (PCD) provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251484
    Device Name
    CT:VQ
    Manufacturer
    Date Cleared
    2025-08-28

    (106 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CT:VQ software is a non-invasive image post-processing technology, using CT lung images to provide clinical decision support for thoracic disease diagnosis and management in adult patients. It utilizes two non-contrast chest CT studies to quantify and visualize ventilation and perfusion.

    Quantification and visualizations are provided as DICOM images. CT:VQ may be used when Radiologists, Pulmonologists, and/or Nuclear Medicine Physicians need a better understanding of a patient's lung function and/or respiratory condition.

    Device Description

    CT:VQ is a Software as a Medical Device (SaMD) technology, which can be used in the analysis of a paired (inspiratory/expiratory) non-contrast Chest CT. It is designed to measure regional ventilation (V) and regional perfusion (Q) in the lungs.

    The Device provides visualization and quantification to aid in the assessment of thoracic diseases. These regional measures are derived from the lung tissue displacement, the lung volume change, and the Hounsfield Units of the paired (inspiratory/expiratory) chest CT.

    The Device outputs DICOM images containing the ventilation output and perfusion output, consisting of a series of image slices generated with the same slice spacing as the expiration CT. In each slice the intensity value for each voxel represents either the value of ventilation or the value for perfusion, respectively, at that spatial location. Additional Information sheet is also generated containing quantitative data, such as lung volume.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the CT:VQ device, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Device Performance

    The acceptance criteria for CT:VQ are implicitly demonstrated through its strong performance in clinical studies, showing agreement with established gold standards. While explicit numerical acceptance criteria are not provided in a table format within the summary, the narrative describes the goals of the study:

    • Consistency/Agreement with Nuclear Medicine Imaging (SPECT/CT): The device's regional ventilation and perfusion measurements should align well with SPECT/CT findings.
    • Correlation with Pulmonary Function Tests (PFTs): CT:VQ metrics should statistically correlate with standard PFTs like DLCO and FEV1/FVC ratio.
    • Interpretability and Clinical Actionability: The outputs should be clear, understandable, and useful for clinicians.
    • Safety and Effectiveness Profile: The device should have a safety and effectiveness profile similar to the primary predicate device.

    Table of Acceptance Criteria and Reported Device Performance (as inferred from the text):

    Acceptance Criterion (Inferred)Reported Device Performance
    Strong regional agreement with SPECT VQ (Ventilation)CT:VQ showed strong regional agreement with SPECT VQ across lobar distributions of ventilation. In the Reader Performance Study, clinicians consistently rated CT:VQ outputs as having good to excellent agreement with SPECT across all lung regions.
    Strong regional agreement with SPECT VQ (Perfusion)CT:VQ showed strong regional agreement with SPECT VQ across lobar distributions of perfusion. In the Reader Performance Study, clinicians consistently rated CT:VQ outputs as having good to excellent agreement with SPECT across all lung regions.
    Correlation with Gas Transfer Impairment (DLCO)Quantitative perfusion heterogeneity metrics derived from CT:VQ demonstrated stronger associations with gas transfer impairment (DLCO) than those derived from SPECT, suggesting improved physiological sensitivity. There was a statistically significant correlation between the CT:VQ and PFT outputs.
    Correlation with Airway Obstruction (FEV1 and FEV1/FVC % predicted)Ventilation heterogeneity metrics from CT:VQ correlated well with FEV1 and FEV1/FVC % predicted. There was a statistically significant correlation between the CT:VQ and PFT outputs.
    Interpretability and Clinical Actionability by Intended UsersThe Reader Performance Study affirmed that CT:VQ outputs are interpretable and clinically actionable by intended users.
    Inter-reader variability similar to SPECTInter-reader variability was not significantly different for CT:VQ than for SPECT.
    Feasibility of generating reliable and consistent dataThe clinical studies successfully demonstrated the feasibility of generating valid data that is reliable and consistent with Nuclear Medicine Ventilation imaging results.
    Safety and effectiveness profile similar to predicate deviceBased on the clinical performance, CT:VQ was found to have a safety and effectiveness profile that is similar to the primary predicate device. It also demonstrated the capability to provide information without contrast agents (unlike some alternative perfusion methods).
    Robustness across various CT inputsVerification testing demonstrated that the Device was robust within acceptable performance limits across the entire range of inputs (CT scanners, institutions, varying lung volumes, image properties affecting voxel size and SNR). Specific performance limits are not quantified in the summary, but the general claim of robustness is made.

    Study Details

    Here's a breakdown of the specific information requested about the studies:

    1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Sizes:
      • Reader Performance Study: n=77
      • Standalone Performance Assessment: n=58 (a subset of the overall clinical studies data)
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submission is from 4DMedical Limited in Australia, and the FDA clearance is in the US. The description mentions "clinically-acquired data included paired chest CTs acquired on CT scanners across a range of manufacturers and models and at different institutions, across a diverse range of patients." This suggests multi-institutional data, potentially from various geographic locations, but this is not confirmed.
      • Retrospective or Prospective: Not explicitly stated whether the studies were retrospective or prospective. The description "clinical studies were also conducted to demonstrate the safety and efficacy... in the context of clinical care" and comparing with "gold-standard and best practice measures for respiratory diagnosis" often implies retrospective analysis of existing data combined with prospective data collection, but this is not definitive in the text.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not explicitly stated for establishing ground truth, although for the Reader Performance Study, "clinicians with expertise in thoracic imaging and pulmonary care" were involved in rating the outputs. The implication is that these experts, along with SPECT/CT and PFT results, contributed to the ground truth.
    • Qualifications of Experts: "Clinicians with expertise in thoracic imaging and pulmonary care." No specific number of years of experience or board certifications (e.g., radiologist with 10 years of experience) is provided.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not explicitly stated. The summary mentions "Inter-reader variability was not significantly different for CT:VQ than for SPECT," which implies multiple readers, but the method for resolving discrepancies or establishing a final ground truth from multiple readers is not detailed.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: A "Reader Performance Study" was conducted with n=77 cases, involving "clinicians with expertise in thoracic imaging and pulmonary care." This aligns with the characteristics of an MRMC study.
    • Effect Size of Human Reader Improvement with AI vs. without AI assistance: The summary does not provide an effect size or direct comparison of human reader performance with CT:VQ assistance versus without it. The study focused on assessing:
      • Agreement between CT:VQ outputs and SPECT.
      • Interpretability and clinical actionability of CT:VQ outputs.
      • Inter-reader variability of CT:VQ vs. SPECT.
        It does not quantify an improvement in reader accuracy or efficiency due to AI assistance.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Yes, a "Standalone Performance Assessment" was performed with a subset of 58 cases. The findings indicated strong regional agreement between CT:VQ and SPECT VQ measurements and stronger associations of CT:VQ perfusion metrics with DLCO compared to SPECT.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: A combination of established clinical diagnostics was used:
      • Nuclear Medicine Imaging (Single photon emission computed tomography, SPECT/CT): Used as a "gold-standard and best practice measure" for regional ventilation and perfusion.
      • Pulmonary Function Tests (PFTs): Specifically Diffusing capacity of the lung for carbon monoxide (DLCO) and FEV1/FVC ratio, used to correlate with CT:VQ outputs.
      • Clinical Diagnosis/Findings: Implied through "Case Studies further illustrated key advantages of CT:VQ... successfully replicated the diagnostic findings of SPECT."

    7. The sample size for the training set:

    • Training Set Sample Size: Not explicitly stated in the provided text. The summary only mentions the sample sizes for the clinical validation studies (test sets).

    8. How the ground truth for the training set was established:

    • Training Set Ground Truth Establishment: Not explicitly stated how the ground truth for the training set was established, as the training set size and characteristics are not detailed. Typically, it would involve similar rigorous processes (e.g., expert annotation, gold-standard imaging modalities, clinical outcomes) as the test set, but this information is absent in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250941
    Device Name
    Revolution Vibe
    Date Cleared
    2025-08-01

    (126 days)

    Product Code
    Regulation Number
    892.1750
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

    This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results.

    The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.

    The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

    If the spectral imaging option is included on the system, the system can acquire CT images using different kV levels of the same anatomical region of a patient in a single rotation from a single source. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. This approach enables images to be generated at energies selected from the available spectrum to visualize and analyze information about anatomical and pathological structures.

    GSI provides information of the chemical composition of renal calculi by calculation and graphical display of the spectrum of effective atomic number. GSI Kidney stone characterization provides additional information to aid in the characterization of uric acid versus non-uric acid stones. It is intended to be used as an adjunct to current standard methods for evaluating stone etiology and composition.

    The CT system is indicated for low dose CT for lung cancer screening. The screening must be performed within the established inclusion criteria of programs/ protocols that have been approved and published by either a governmental body or professional medical society.

    Device Description

    This proposed device Revolution Vibe is a general purpose, premium multi-slice CT Scanning system consisting of a gantry, table, system cabinet, scanner desktop, power distribution unit, and associated accessories. It has been optimized for cardiac performance while still delivering exceptional imaging quality across the entire body.

    Revolution Vibe is a modified dual energy CT system based on its predicate device Revolution Apex Elite (K213715). Compared to the predicate, the most notable change in Revolution Vibe is the modified detector design together with corresponding software changes which is optimized for cardiac imaging providing capability to image the whole heart in one single rotation same as the predicate.

    Revolution Vibe offers an accessible whole heart coverage, full cardiac capability CT scanner which can deliver outstanding routine head and body imaging capabilities. The detector of Revolution Vibe uses the same GEHC's Gemstone scintillator with 256 x 0.625 mm row providing up to 16 cm of coverage in Z direction within 32 cm scan field of view, and 64 x 0.625 mm row providing up to 4 cm of coverage in Z direction within 50 cm scan field of view. The available gantry rotation speeds are 0.23, 0.28, 0.35, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, and 1.0 seconds per rotation.

    Revolution Vibe inherits virtually all of the key technologies from the predicate such as: high tube current (mA) output, 80 cm bore size with Whisper Drive, Deep Learning Image Reconstruction for noise reduction (DLIR K183202/K213999, GSI DLIR K201745), ASIR-V iterative recon, enhanced Extended Field of View (EFOV) reconstruction MaxFOV 2 (K203617), fast rotation speed as fast as 0.23 second/rot (K213715), and spectral imaging capability enabled by ultrafast kilovoltage(kv) switching (K163213), as well as ECG-less cardiac (K233750). It also includes the Auto ROI enabled by AI which is integrated within the existing SmartPrep workflow for predicting Baseline and monitoring ROI automatically. As such, the Revolution Vibe carries over virtually all features and functionalities of the predicate device Revolution Apex Elite (K213715).

    This CT system can be used for low dose lung cancer screening in high risk populations*.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Revolution Vibe CT system does not include detailed acceptance criteria or a comprehensive study report to fully characterize the device's performance against specific metrics. The information focuses more on the equivalence to a predicate device and general safety/effectiveness.

    However, based on the text, we can infer some aspects related to the Auto ROI feature, which is the only part of the device described with specific performance testing details.

    Here's an attempt to extract and describe the available information, with clear indications of what is not provided in the document.


    Acceptance Criteria and Device Performance for Auto ROI

    The document mentions specific performance testing for the "Auto ROI" feature, which utilizes AI. For other aspects of the Revolution Vibe CT system, the submission relies on demonstrating substantial equivalence to the predicate device (Revolution Apex Elite) through engineering design V&V, bench testing, and a clinical reader study focused on overall image utility, rather than specific quantitative performance metrics meeting predefined acceptance criteria for the entire system.

    1. Table of Acceptance Criteria and Reported Device Performance (Specific to Auto ROI)

    Feature/MetricAcceptance Criteria (Implicit)Reported Device Performance
    Auto ROI Success Rate"exceeding the pre-established acceptance criteria"Testing resulted in "success rates exceeding the pre-established acceptance criteria." (Specific numerical value not provided)

    Note: The document does not provide the explicit numerical value for the "pre-established acceptance criteria" or the actual "success rate" achieved for the Auto ROI feature.

    2. Sample Size and Data Provenance for the Test Set (Specific to Auto ROI)

    • Sample Size: 1341 clinical images
    • Data Provenance: "real clinical practice" (Specific country of origin not mentioned). The images were used for "Auto ROI performance" testing, which implies retrospective analysis of existing clinical data.

    3. Number of Experts and Qualifications to Establish Ground Truth (Specific to Auto ROI)

    • Number of Experts: Not specified for the Auto ROI ground truth establishment.
    • Qualifications of Experts: Not specified for the Auto ROI ground truth establishment.

    Note: The document mentions 3 readers for the overall clinical reader study (see point 5), but this is for evaluating the diagnostic utility and image quality of the CT system and not explicitly for establishing ground truth for the Auto ROI feature.

    4. Adjudication Method for the Test Set (Specific to Auto ROI)

    • Adjudication Method: Not specified for the Auto ROI test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes, a "clinical reader study of sample clinical data" was carried out. It is described as a "blinded, retrospective clinical reader study."

    • Effect Size of Human Readers Improvement with AI vs. without AI assistance: The document states the purpose of this reader study was to validate that "Revolution Vibe are of diagnostic utility and is safe and effective for its intended use." It does not report an effect size or direct comparison of human readers' performance with and without AI assistance (specifically for the Auto ROI feature within the context of reader performance). The study seemed to evaluate the CT system's overall image quality and clinical utility, possibly implying that the Auto ROI is integrated into this overall evaluation, but a comparative effectiveness study of the AI's impact on human performance is not described.

      • Details of MRMC Study:
        • Number of Cases: 30 CT cardiac exams
        • Number of Readers: 3
        • Reader Qualifications: US board-certified in Radiology with more than 5 years' experience in CT cardiac imaging.
        • Exams Covered: "wide range of cardiac clinical scenarios."
        • Reader Task: "Readers were asked to provide evaluation of image quality and the clinical utility."

    6. Standalone (Algorithm Only) Performance

    • Was a standalone study done? Yes, for the "Auto ROI" feature, performance was tested "using 1341 clinical images from real clinical practice," and "the tests results in success rates exceeding the pre-established acceptance criteria." This implies an algorithm-only evaluation of the Auto ROI's ability to successfully identify and monitor ROI.

    7. Type of Ground Truth Used (Specific to Auto ROI)

    • Type of Ground Truth: Not explicitly stated for the Auto ROI. Given the "success rates" metric, it likely involved a comparison against a predefined "true" ROI determined by human experts or a gold standard method. It's plausible that this was established by expert consensus or reference standards.

    8. Sample Size for the Training Set

    • Sample Size: Not provided in the document.

    9. How Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not provided in the document.

    In summary, the provided documentation focuses on demonstrating substantial equivalence of the Revolution Vibe CT system to its predicate, Revolution Apex Elite, rather than providing detailed, quantitative performance metrics against specific acceptance criteria for all features. The "Auto ROI" feature is the only component where specific performance testing (standalone) is briefly mentioned, but key details like numerical acceptance criteria, actual success rates, and ground truth methodology for training datasets are not disclosed. The human reader study was for general validation of diagnostic utility, not a comparative effectiveness study of AI assistance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251061
    Date Cleared
    2025-07-28

    (115 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

    The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.

    This CT system can be used for low dose lung cancer screening in high risk populations*.

    *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    Device Description

    Siemens intends to update the software version syngo CT VB20 (update) for the following NAEOTOM Alpha class CT systems:

    Dual Source NAEOTOM CT scanner systems:

    • NAEOTOM Alpha (trade name ex-factory CT systems: NAEOTOM Alpha.Peak; trade name installed base CT systems with SW upgrade only: NAEOTOM Alpha)

    For simplicity, the product name of NAEOTOM Alpha will be used throughout this submission instead of the trade name NAEOTOM Alpha.Peak.

    • NAEOTOM Alpha.Pro

    Single Source NAEOTOM CT scanner system:

    • NAEOTOM Alpha.Prime

    The subject devices NAEOTOM Alpha (trade name ex-factory CT systems: NAEOTOM Alpha.Peak) and NAEOTOM Alpha.Pro with software version SOMARIS/10 syngo CT VB20 (update) are Computed Tomography X-ray systems which feature two continuously rotating tube-detector systems, denominated as A- and B-systems respectively (dual source NAEOTOM CT scanner system).

    The subject device NAEOTOM Alpha.Prime with software version SOMARIS/10 syngo CT VB20 (update) is a Computed Tomography X-ray system which features one continuously rotating tube-detector systems, denominated as A-system (single source NAEOTOM CT scanner system).

    The detectors' function is based on photon-counting technology.

    In this submission, the above-mentioned CT scanner systems are jointly referred to as subject devices by "NAEOTOM Alpha class CT scanner systems".

    The NAEOTOM Alpha class CT scanner systems with SOMARIS/10 syngo CT VB20 (update) produce CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors. The CT images can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions. The radiation therapy planning support includes, but is not limited to, Brachytherapy, Particle Therapy including Proton Therapy, External Beam Radiation Therapy, Surgery. The computer system delivered with the CT scanner is able to run optional post-processing applications.

    Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.

    The platform software for the NAEOTOM Alpha class CT scanner systems is syngo CT VB20 (update) (SOMARIS/10 syngo CT VB20 (update)). It is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. The software platform provides plugin software interfaces that allow for the use of specific commercially available post-processing software algorithms in an unmodified form from the cleared stand-alone post-processing version.

    Software version syngo CT VB20 (update) (SOMARIS/10 syngo CT VB20 (update)) shall support additional software features compared to the software version of the predicate devices NAEOTOM Alpha class CT systems with syngo CT VB20 (SOMARIS/10 syngo CT VB20) cleared in K243523.

    Software version SOMARIS/10 syngo CT VB20 (update) will be offered ex-factory and as optional upgrade for the existing NAEOTOM Alpha class systems.

    The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study that proves the device meets those criteria for the NAEOTOM Alpha CT Scanner Systems. However, the document primarily focuses on demonstrating substantial equivalence to a predicate device and safety and effectiveness based on non-clinical testing and adherence to standards, rather than detailing a specific clinical performance study with defined acceptance criteria for a diagnostic aid.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics in the way one would typically find for a diagnostic AI device (e.g., sensitivity, specificity, AUC). Instead, it states that:

    • Acceptance Criteria for Software: "The test specification and acceptance criteria are related to the corresponding requirements." and "The test results show that all of the software specifications have met the acceptance criteria."
    • Acceptance Criteria for Features: "Test results show that the subject devices...is comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices."
    • Performance Claim: "The conclusions drawn from the non-clinical and clinical tests demonstrate that the subject devices are as safe, as effective, and perform as well as or better than the predicate devices."

    The closest the document comes to defining and reporting on "performance criteria" for a specific feature, beyond basic safety and technical functionality, are for the HD FoV 5.0 and ZeeFree RT algorithms.

    Acceptance Criteria (Implied)Reported Device Performance
    HD FoV 5.0 algorithm: As safe and effective as HD FoV 4.0.HD FoV 5.0 algorithm: Bench test results comparing it to HD FoV 4.0 based on physical and anthropomorphic phantoms. Performance was also evaluated by board-approved radio-oncologists and medical physicists via a retrospective blinded rater study. No specific metrics (e.g., image quality scores, diagnostic accuracy) are provided in this summary.
    ZeeFree RT reconstruction:ZeeFree RT reconstruction:
    - No relevant errors in CT values and noise in homogeneous water phantom.- Bench test results show it "does not affect CT values and noise levels in a homogenous water phantom outside of stack-transition areas compared to the non-corrected standard reconstruction."
    - No relevant errors in CT values in phantoms with tissue-equivalent inserts (even with metals and iMAR).- Bench test results show it "introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm."
    - No relevant geometrical distortions in a static torso phantom.- Bench test results show it "introduces no relevant geometrical distortions in a static torso phantom."
    - No relevant deteriorations of position or shape in a dynamic thorax phantom (spherical shape with various breathing motions).- Bench test results show it "introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom when moving a spherical shape according to regular, irregular, and patient breathing motion." Also states it "can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device."
    - Successfully applied to 4D respiratory-gated images (Direct i4D).- Bench test results show it "can successfully be applied to 4D respiratory-gated sequence images (Direct i4D)."
    - Enables optional reconstruction of stack artifact-corrected images which reduce misalignment artifacts where present in standard images.- Bench test results show it "enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts, if such stack alignment artefacts are identified in non-corrected standard images."
    - Does not introduce relevant new artifacts not present in non-corrected standard reconstruction.- Bench test results show it "does not introduce relevant new artefacts, which were previously not present in the non-corrected standard reconstruction." Also states it "does not introduce new artifacts, which were previously not present in the non-corrected standard reconstruction, even in presence of metals."
    - Independent from physical detector width of acquired data.- Bench test results show it "is independent from the physical detector width of the acquired data."

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "physical and anthropomorphic phantoms" for HD FoV 5.0 and "homogeneous water phantom" and "phantom with tissue-equivalent inserts," and "dynamic thorax phantom" for ZeeFree RT. It also refers to "retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions" for ZeeFree RT, but does not specify the sample size (number of cases/patients) or the country of origin for these real-world examination datasets. The data provenance (retrospective/prospective) is stated for the rater study for ZeeFree RT as retrospective, but not for the HD FoV 5.0 rater study (though implied by "retrospective blinded rater study").

    3. Number of Experts and Qualifications for Ground Truth

    For the HD FoV 5.0 and ZeeFree RT rater studies, the experts were "board-approved radio-oncologists and medical physicists." The number of experts is not specified, nor is their specific years of experience.

    4. Adjudication Method for the Test Set

    The document explicitly states "retrospective blinded rater study" for HD FoV 5.0 and ZeeFree RT. However, it does not specify the adjudication method (e.g., 2+1, 3+1, none) if there were multiple raters and disagreements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document states that for HD FoV 5.0 and ZeeFree RT, "the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater study." This indicates a reader study, which is often a component of an MRMC study.

    However, the study described does not appear to be comparing human readers with AI assistance vs. without AI assistance. Instead, for HD FoV 5.0, it's comparing the new algorithm's results to its predecessor, HD FoV 4.0. For ZeeFree RT, it's comparing the reconstruction to "Standard reconstruction" and assessing if it introduces errors or new artifacts. It's an evaluation of the algorithm's output, not necessarily a direct measure of human reader improvement with AI assistance. Therefore, no effect size for human reader improvement with AI vs. without AI assistance is reported because this specific type of comparative effectiveness study was not described.

    6. Standalone (Algorithm Only) Performance Study

    Yes, standalone (algorithm only) performance was conducted. The bench testing described for both HD FoV 5.0 and ZeeFree RT involves detailed evaluations of the algorithms' outputs using phantoms and comparing them to established standards or previous versions. For example, for ZeeFree RT, the bench test objectives include demonstrating that it "introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom" and "does not introduce relevant new artefacts." This is an assessment of the algorithm's direct output.

    7. Type of Ground Truth Used

    The ground truth used primarily appears to be:

    • Phantom-based measurements: For HD FoV 5.0 (physical and anthropomorphic phantoms) and ZeeFree RT (homogeneous water phantom, tissue-equivalent inserts, static torso phantom, dynamic thorax phantom). These phantoms have known properties which serve as ground truth for evaluating image quality metrics.
    • Expert Consensus/Interpretation: For HD FoV 5.0 and ZeeFree RT, it involved "board-approved radio-oncologists and medical physicists" in "retrospective blinded rater studies." This suggests the experts' interpretations (potentially comparing image features or diagnostic quality) formed a part of the ground truth or served as the primary evaluation method. The text doesn't specify if there was a pre-established "true" diagnosis or condition for these clinical cases, or if the experts were rating image quality or agreement with a reference standard.

    8. Sample Size for the Training Set

    The document does not specify the sample size for the training set for any of the algorithms or software features. This document is a 510(k) summary, which generally focuses on justification for substantial equivalence rather than detailed algorithm development specifics.

    9. How the Ground Truth for the Training Set Was Established

    The document does not describe how the ground truth for the training set was established, as it does not provide information about the training set itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250970
    Device Name
    Marie
    Manufacturer
    Date Cleared
    2025-07-25

    (116 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device.

    Device Description

    The Marie System is intended to acquire CT images and enable the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device.

    The Marie System is intended to be used by healthcare professionals to image patients in an upright position rather than conventional supine treatments, to enable precise treatment planning and patient positioning for radiotherapy.

    Specifically, it is intended to:

    • Image the patient to provide image-guided radiation therapy
    • Image the patient to acquire images for the purpose of treatment planning
    • Immobilize patients in an upright position for upright radiotherapy.

    The Marie System is comprised of two major sub-systems: a computed tomography (CT) imaging system that performs pretreatment imaging and treatment simulation in the upright positions and a beam agnostic, patient positioning system that supports the patient in the upright positions.

    The Marie Imaging System is used with compatible devices for treatment delivery and patient immobilization. The positioning system is designed with six degrees of freedom of motion and a patient positioning system to provide the desired posture for each cancer site to achieve accurate, reproducible patient setups, while the imaging system acquires helical scans by translating and rotating up the patient.

    AI/ML Overview

    The provided text solely describes the Leo Cancer Care Marie System as a Computed Tomography X-ray System with its features, safety, and performance details. It outlines the regulatory clearance (FDA 510(k)) based on substantial equivalence to a predicate device (P-ARTIS K160611). It describes the device's characteristics and the non-clinical tests performed to demonstrate its performance and functionality against design and risk management requirements.

    Crucially, the provided text does not contain any information about acceptance criteria for AI performance, clinical study design, sample sizes for test or training sets, ground truth establishment using experts, or any MRMC comparative effectiveness studies. The document is a 510(k) clearance summary for a medical device (a CT imaging and patient positioning system), not for an AI diagnostic or assistive software. The "performance testing" section refers to hardware and software system performance, not AI model performance.

    Therefore, I cannot answer most of your questions based on the provided input. The questions you've asked are typically relevant for AI/ML-based medical devices or diagnostics that involve image interpretation and require rigorous validation against human expert performance. The Marie System, as described, is a physical imaging system for patient positioning and CT image acquisition, not an AI software.

    If this were an AI device, the answers would need to be extracted from sections detailing clinical performance studies, AI model validation, or similar. Since those sections are absent, I am unable to provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250822
    Date Cleared
    2025-07-03

    (107 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.

    The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.

    This CT system can be used for low dose lung cancer screening in high risk populations*.

    *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    Device Description

    Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:

    a) Single Source CT Scanner systems (SOMATOM go. Platform):

    • SOMATOM go.Now
    • SOMATOM go.Up
    • SOMATOM go.All
    • SOMATOM go.Top
    • SOMATOM go.Sim
    • SOMATOM go.Open Pro

    In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" CT scanner systems.

    b) Dual Source CT Scanner system:

    • SOMATOM Pro.Pulse

    The above listed subject devices with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one (Single Source) or two (Dual Source) continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go. Platform and the SOMATOM Pro.Pulse with software SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for the SOMATOM CT Systems focuses heavily on establishing substantial equivalence to predicate devices through comparisons of technological characteristics, hardware, and software. It generally asserts that the device has met performance criteria through verification and validation testing, but it does not provide a detailed "Acceptance Criteria Table" with specific quantitative metrics and reported device performance. Similarly, it describes the types of studies performed (e.g., bench testing, retrospective blinded rater study), but it lacks the specific details requested regarding sample sizes, data provenance, expert qualifications, and effect sizes that would typically be found in a detailed study report.

    Therefore, I will extract and synthesize the information that is available in the document and explicitly state where the requested information is not provided.


    Understanding the Device and its Changes

    The devices under review are Siemens SOMATOM CT Systems (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, and SOMATOM Pro.Pulse) with a new software version, SOMARIS/10 syngo CT VB20. This new software version builds upon the previous VB10 version cleared in K233650 and K232206.

    The submission focuses on modifications and new features introduced with VB20, including:

    • Eco Power Mode: New feature for reduced energy consumption during idle times (not supported on go.Now and go.Up).
    • Oncology Exchange: New feature for transferring prescription information from ARIA Oncology Information System.
    • myExam Contrast: New feature for exchanging contrast injection parameters.
    • FAST 3D Camera/FAST Integrated Workflow: Modifications including retrained algorithms, collision indication, and Centerline/Grid Overlay.
    • FAST Planning: Extended to detect additional body regions.
    • myExam Companion (myExam Compass/myExam Cockpit): Clinical decision trees now available for child protocols.
    • HD FoV 5.0: New extended field of view reconstruction algorithm (for go.Sim and go.Open Pro only).
    • CT guided intervention – myAblation Guide interface: New interface.
    • Flex 4D Spiral: Modifications regarding dynamic tube current modulation.
    • ZeeFree RT: New stack artifact reduced reconstruction for respiratory-related examinations (for go.Open Pro only).
    • DirectDensity: Modified to include stopping-power ratio (Kernel St).
    • DirectLaser: Patient Marking workflow improvement.
    • Respiratory Motion management - Open Online Interface: New interface for respiratory gating.
    • DirectSetup Notes: Enabled for certain SOMATOM go. Platform systems.

    The core argument for clearance is substantial equivalence to predicate devices. This means that, despite modifications, the device is as safe and effective as a legally marketed device (the predicates).


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not contain a specific table of quantitative acceptance criteria with corresponding reported device performance values. Instead, it describes general acceptance criteria related to verification and validation tests and then provides qualitative statements about the test results demonstrating comparability or improvement over predicate devices.

    Here's a summary of the described performance evaluations:

    Feature/MetricAcceptance Criteria (Qualitative)Reported Device Performance (Qualitative)
    OverallMeet acceptance criteria for all software specifications. Enable safe and effective integration. Perform as intended in specified use conditions."All software specifications have met the acceptance criteria." "Verification and validation support the claims of substantial equivalence." "Perform(s) as intended in the specified use conditions." "As safe, as effective, and perform as well as or better than the predicate devices."
    FAST 3D Camera Accuracy (Isocentering, Range, Direction)Comparable or better accuracy to predicate device for adults; extend support to adolescents."Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents."
    FAST Planning CorrectnessHigh fraction (percentage) of ranges calculated correctly and without needing change. Meets interactive requirements (fast calculation time)."For more than 90% of the ranges no editing action was necessary to cover standard ranges." "For more than 95%, the speed of the algorithm was sufficient."
    HD FoV 5.0 Performance (vs. HD FoV 4.0)As safe and effective as HD FoV 4.0."Results obtained with the new HD FoV 5.0 algorithm are compared with its predecessor, the HD FoV 4.0 algorithm, based on physical and anthropomorphic phantoms...This comparison is conducted to demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm." (No quantitative metrics provided in this document excerpt regarding this comparison's outcome).
    Flex 4D Spiral Functionality & Image QualityProper function and acceptable image quality."The performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral." (No quantitative metrics provided)
    ZeeFree RT Reconstruction PerformanceNo relevant errors in CT values and noise in homogeneous phantoms. No relevant errors in CT values in tissue-equivalent phantoms. No relevant geometrical distortions in static phantoms. No relevant deteriorations of position/shape in dynamic phantoms. No relevant new artifacts. Maintain performance with iMAR. Independent of detector width."introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom" "introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm" "introduces no relevant geometrical distortions in a static torso phantom" "introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom" "does not introduce relevant new artefacts" "can be successfully applied in combination with metal artifact correction (iMAR)" "is independent from the physical detector width"
    DirectDensity Performance (iBHC variants)Reduced dependence on tube voltage and filtration for non-water-like tissues. Image values aligned with material properties."reduced dependence on tube voltage and filtration compared to the corresponding quantitative kernel (Qr) with iBHC Bone for non-water-like tissues, such as adipose and bone." "generate image value closely aligned with the respective material properties." "has been validated."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document provides very limited, qualitative information:

    • FAST 3D Camera: Optimized using "additional data from adults and adolescence patients." No specific number of patients or images mentioned.
    • FAST Planning: Evaluated on "patient data." No specific number of patients or images mentioned.
    • HD FoV 5.0: Evaluated with "physical and anthropomorphic phantoms."
    • Flex 4D Spiral: No specific sample size or data type mentioned for performance assessment.
    • ZeeFree RT: Evaluated with "homogeneous water phantom," "phantom with tissue-equivalent inserts," "static torso phantom," and "dynamic thorax phantom." Also, "retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions." No specific number of phantoms, images per phantom, or patient cases mentioned.
    • DirectDensity: Evaluated on "SOMATOM CT scanner models." No specific sample size or data type mentioned.

    Data Provenance:

    • Country of Origin: Not specified for the patient data used for algorithm optimization/validation.
    • Retrospective or Prospective:
      • FAST 3D Camera: Implied retrospective as it uses "additional data."
      • FAST Planning: Implied retrospective as it uses "patient data."
      • HD FoV 5.0: Retrospective for the blinded rater study.
      • ZeeFree RT: Retrospective for the blinded rater study of clinical cases. The phantom tests are by nature not retrospective/prospective.

    3. Number of Experts and Qualifications for Ground Truth

    • HD FoV 5.0: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.
    • ZeeFree RT: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.

    For other tests, ground truth appears to be established by phantom measurements or internal engineering verification, rather than human expert reads validating clinical ground truth.


    4. Adjudication Method for the Test Set

    The document mentions "retrospective blinded rater study" for HD FoV 5.0 and ZeeFree RT. However, it does not specify the adjudication method used (e.g., 2+1, 3+1, none) for these studies. It only states they were "blinded."


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned for HD FoV 5.0 and ZeeFree RT. Both were "retrospective blinded rater studies."
    • Effect Size: The document does not report specific effect sizes (e.g., how much human readers improve with AI vs. without AI assistance). It only states that the purpose of the comparison was to "demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm" and for ZeeFree RT that it "enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts." This implies an assessment of non-inferiority or improvement in image quality, but specific quantitative results for reader performance are not provided in this excerpt.

    6. Standalone (Algorithm Only) Performance

    The document describes tests for several algorithms (FAST 3D Camera, FAST Planning, HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, DirectDensity) using phantoms and "patient data." These evaluations seem to be focused on the algorithm's performance in generating images or calculations, independent of human interpretation in some cases (e.g., accuracy of FAST 3D Camera, correctness percentage of FAST Planning).

    However, it does not explicitly use the term "standalone performance" to differentiate these from human-in-the-loop assessments. The mention of "retrospective blinded rater studies" for HD FoV 5.0 and ZeeFree RT indicates a human-in-the-loop component for that specific evaluation, but the phantom testing mentioned alongside them would be considered standalone.


    7. Type of Ground Truth Used

    • Phantom Data: For HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, and DirectDensity, physical and/or anthropomorphic phantoms were used, implying the ground truth is precisely known physical characteristics or pre-defined phantom configurations.
    • Expert Consensus/Reads: For HD FoV 5.0 and ZeeFree RT, board-approved radio-oncologists and medical physicists performed retrospective blinded rater studies, implying their interpretations/ratings served as a form of ground truth or evaluation metric. It's not explicitly stated if this was against a clinical gold standard (e.g., pathology) or if it was a comparative assessment of image quality and clinical utility.
    • Internal Verification: For FAST 3D Camera, FAST Planning, accuracy was assessed, likely against internal system metrics or pre-defined ideal outcomes.

    8. Sample Size for the Training Set

    The document does not provide any specific information about the sample size used for training the algorithms (e.g., for FAST 3D Camera, FAST Planning, HD FoV 5.0, ZeeFree RT). It only states that FAST 3D Camera was "optimized using additional data" and FAST Planning's algorithm had "product development, validation, and verification on patient data."


    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any specific information on how the ground truth for the training set was established. It only mentions the data types used for validation/verification (phantoms, patient data from two institutions, expert raters).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 65