K Number
K233728

Validate with FDA (Live)

Device Name
SIGNA Champion
Date Cleared
2024-01-19

(59 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Device Description

SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).

AI/ML Overview

The provided text does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/human reader performance study. The document is a 510(k) premarket notification for a Magnetic Resonance Diagnostic Device (SIGNA™ Champion).

The relevant sections state:

  • "The subject of this premarket submission, the SIGNA™ Champion, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission."
  • "The sample clinical images demonstrate acceptable diagnostic image performance of the SIGNA™ Champion in accordance with the FDA Guidance 'Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices' issued on October 10, 2023. The image quality of the SIGNA™ Champion is substantially equivalent to that of the predicate device."

This indicates that the FDA clearance for the SIGNA™ Champion MR system was based on demonstrating substantial equivalence to a previous predicate device (SIGNA™ Voyager) through non-clinical testing and the review of sample clinical images, rather than a prospective clinical study involving human readers and AI assistance for diagnostic tasks.

Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI-assisted diagnostic device performance study because such a study was not conducted or reported in this 510(k) submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 19, 2024

GE Healthcare (Tianjin) Company Limited % Glen Sabin Director - Regulatory Affairs, MR Strategy GE Medical Systems, LLC 3200 N Grandview Blvd. Waukesha, Wisconsin 53188

Re: K233728

Trade/Device Name: SIGNA™ Champion Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI Dated: November 21, 2023 Received: November 21, 2023

Dear Glen Sabin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K233728

Device Name

SIGNA™ Champion

Indications for Use (Describe)

The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with a white GE monogram inside, followed by the words "GE HealthCare" in purple. The font is sans-serif and the overall design is clean and modern.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

DateJanuary 19, 2024
SubmitterGE Healthcare (Tianjin) Company LimitedNo. 266 Jingsan Road, Tianjin Airport Economic AreaTianjin, P.R. China 300308
PrimaryContact PersonXinyu SongLead Specialist, Regulatory Affairs, MRGE HealthCarePhone: +86 186 1188 4503E-mail: Xinyu.Song@ge.com
SecondaryContact PersonGlen SabinDirector - Regulatory Affairs, MR StrategyGE HealthCarePhone: 262 894-4968E-mail: glen.sabin@ge.com
Device TradeNameSIGNA™ Champion
Common/UsualNameMagnetic Resonance Diagnostic Device
ClassificationNamesMagnetic Resonance Diagnostic Device per 21 CFR 892.1000
Product CodeLNH, LNI
PredicateDeviceSIGNA™ Voyager (K192426)
ReferenceDevice(1) SIGNA™ Victor (K223439)(2) SIGNA™ Hero (K213668)

Reason for Submission:

This 510(k) is being submitted due to the introduction of SIGNA™ Champion, a new 1.5T MR system from GE HealthCare.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem features a stylized "GE" monogram inside a circle with decorative flourishes. The text "GE HealthCare" is written in a clean, sans-serif font and is colored in the same purple as the emblem.

Device Description:

SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).

A high-level summary of significant hardware and software modifications is provided below:

Hardware changes (compared with Predicate Device SIGNA™ Voyager):

  • Magnet Enclosure -
  • -Patient Table and related support components
  • Host Computer -
  • RF Transmit Chain -
  • Wireless Gating -

Software Changes (compared with Reference Device 2 SIGNA™ Hero)

  • Sonic DL -
  • AIR Recon DL -

Indications for Use

The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.

Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance.

These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. On the left is the GE monogram in purple, which is a stylized version of the letters G and E intertwined. To the right of the monogram is the text "GE HealthCare" in purple. The text is in a simple, sans-serif font.

Technology

The SIGNA™ Champion employs the same fundamental scientific technology as its predicate device.

SIGNA™ Champion is built with superconducting magnet, RF transmit architecture, RF receive chain and software application suite.

Comparison of Indications for Use

The changes in technology do not impact the indications for use.

The indications for use have not been changed, other than to reflect the SIGNA™ Champion product name.

Therefore, the intended use is the same as the predicate device in accordance with the FDA's guidance document "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", dated 28 July 2014.

Comparison of Technological Characteristics

Overall, the SIGNA™ Champion employs the same fundamental scientific technology and operating principles as the predicate device.

There are some differences in characteristics between the proposed device and the predicate device, as summarized below:

Subsystem orComponentPredicate DeviceSIGNA™ Voyager(K192426)Proposed DeviceSIGNA™ Champion(K233728)Comments
Magnet1.5T Superconducting Magnet with activeshielding1.5T Superconducting Magnet with activeshieldingSIGNA™ Champion uses thesame magnet as thepredicate SIGNA™ Voyager,with a modified enclosuredesign.
GradientSubsystemWater cooled gradient coil with activeshielding.Water cooled gradient coil with activeshielding.SIGNA™ Champion uses thesame gradient subsystem asthe predicate SIGNA™Voyager.
Subsystem orComponentPredicate DeviceSIGNA™ Voyager(K192426)Proposed DeviceSIGNA™ Champion(K233728)Comments
RF TransmitSubsystemTransmit with integrated body coil and localT/R coil.Transmit with integrated body coil and localT/R coil.SIGNA™ Champion uses thesame integrated body coil asthe predicate SIGNA™Voyager, but othercomponents in the transmitchain are different. TheSIGNA™ Champion uses theIn Scan Room transmitarchitecture found inReference Device 1 (SIGNA™Victor).
RF ReceiveSubsystemDigitize-Per-Pin (DPP) receive chainarchitecture.Digitize-Per-Pin (DPP) receive chainarchitecture.SIGNA™ Champion uses thesame DPP architecture asthe predicate SIGNA™Voyager.
RF CoilsComprehensive suite of detachable coils forimaging all anatomiesComprehensive suite of detachable coils forimaging all anatomiesCoils used by SIGNA™Champion are also used bythe predicate SIGNA™Voyager and/or ReferenceDevice 1 (SIGNA™ Victor).
SoftwareFeaturesComprehensive suite of software features,pulse sequences, and image processingapplications to support MR imaging of allanatomies.Comprehensive suite of software features,pulse sequences, and image processingapplications to support MR imaging of allanatomies.Both SIGNA™ Champion andthe predicate SIGNA™Voyager are fully capable MRsystems with a wide range ofsoftware features. SIGNA™Champion includes somenew and enhanced featuressuch as AIR Recon DL andSonic DL that were notincluded in the predicateK192426 submission. TheSIGNA™ Champion softwarefeatures are similar to thosefound on Reference Device 2(SIGNA™ Hero).
GatingAccessoriesRespiratory,peripheral, andcardiac gating.Respiratoryperipheral andcardiac gating withwireless connectionoption.In addition to SIGNA™Voyager's gating solution,SIGNA™ Champion offers anadditional option to usewireless connection.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a purple circle on the left, followed by the words "GE HealthCare" in purple. The text is in a sans-serif font and is slightly larger than the monogram.

SIGNA Champion

510(k) Premarket Notification

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized "GE" monogram. The text "GE HealthCare" is also purple.

These differences do not raise any different questions regarding safety and effectiveness. Both devices must address questions of whether they provide an adequate level of image quality appropriate for diagnostic use. The performance data described in this submission include results of both bench testing and clinical testing that show the image quality performance of SIGNA™ Champion compared to the predicate device.

Determination of Substantial Equivalence

Summary of Non-Clinical Tests:

The SIGNA™ Champion and the predicate device were subject to similar risk management testing to demonstrate substantial equivalence of safety and performance.

Testing to the following voluntary standards included:

  • ANSI AAMI ES60601-1 ●
  • IEC 60601-1-2
  • IEC 60601-2-33
  • IEC 62304
  • IEC 60601-1-6 ●
  • IEC 62366-1
  • . ISO 10993-1

In addition, the SIGNA™ Champion complies with applicable NEMA MS standards for MRI and NEMA PS3 standard for DICOM, as does the predicate device.

Both the SIGNA™ Champion and the predicate device have a successful biocompatibility track record, as demonstrated by ISO 10993 testing and by their history of use in previously cleared devices.

The following quality assurance measures were applied to the development of the subject device, as they were for the predicate device:

  • Risk Analysis ●
  • Requirements Reviews
  • . Design Reviews
  • Testing on unit level (Module verification)
  • Integration testing (System verification)
  • . Performance testing (Verification)
  • . Simulated use testing (Validation)

Summary of Clinical Tests:

The subject of this premarket submission, the SIGNA™ Champion, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with a white GE monogram inside. To the right of the circle is the text "GE HealthCare" in purple.

SIGNA Champion

510(k) Premarket Notification

The sample clinical images demonstrate acceptable diagnostic image performance of the SIGNA™ Champion in accordance with the FDA Guidance "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" issued on October 10, 2023. The image quality of the SIGNA™ Champion is substantially equivalent to that of the predicate device.

Substantial Equivalence Conclusion:

The indications for use of the proposed device are comparable to the claimed predicate device. The SIGNA™ Champion employs equivalent technology to the claimed predicate device. Additionally, the results from the above non-clinical tests demonstrate that the device performs as intended. Therefore, the SIGNA™ Champion is substantially equivalent to the predicate device to which it has been compared.

Conclusion

In conclusion, GE HealthCare considers the SIGNA™ Champion to be as safe, as effective, with performance that is substantially equivalent to the predicate device.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.