(246 days)
Not Found
No
The document describes ECG leads, which are passive components for connecting electrodes to a monitor. There is no mention of any computational or analytical functions that would involve AI/ML.
No
This device is an ECG lead intended for connecting electrodes to an ECG monitoring device, specifically for use in X-ray imaging environments. Its function is to facilitate the monitoring of physiological signals, not to provide therapeutic treatment itself.
No
The device, RT Carbon ECG Leads, connects ECG electrodes to an ECG monitoring device. It is a lead wire that transmits physiological signals, not a device that diagnoses conditions.
No
The device description clearly states it is a physical lead wire connecting electrodes to a monitor, and the performance studies focus on physical properties like tensile strength and dielectric withstand voltage, indicating it is a hardware device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the leads are for connecting ECG electrodes to an ECG monitoring device in X-ray imaging environments. This is a device used on the patient to acquire physiological signals, not a device used to examine specimens from the body.
- Device Description: The description reinforces that the leads connect electrodes on the patient to a physiological detection module or monitor.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.) or providing diagnostic information based on such analysis. The device's function is to facilitate the acquisition of electrical signals from the heart.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device does not fit that definition.
N/A
Intended Use / Indications for Use
The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECG trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT Carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.
Product codes
DSA
Device Description
The RT Carbon ECG Leads are intended to connect electrodes placed on the patient to a physiological detection module or monitor (CTM-300, CTM-400, CTM-500, 7000 Series and 3000 Series). The design includes 3 and 4 lead designs with overall lengths of 24", 30" and 36".
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray imaging environments
Anatomical Site
Not Found
Indicated Patient Age Range
adult, pediatric and neonatal patient populations.
Intended User / Care Setting
trained healthcare professional
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The RT Carbon ECG Leads were assessed, thoroughly tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900. The following bench testing was completed at Ivy in accordance to EC53:
- Patient Leadwire to Trunk Cable Interconnection
- Flex life
- Tensile strength
- Connector mating/unmating cycles
- Connector retention force
- Contact resistance
- Leadwire resistance
- Dielectric withstand voltage
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2900 Patient transducer and electrode cable (including connector).
(a)
Identification. A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2024
Ivy Biomedical Systems, Inc. Felicia Piel Vice President, QA/RA 11 Business Park Drive Branford, Connecticut 06405
Re: K240087
Trade/Device Name: RT Carbon ECG Leads Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA, DRX Dated: August 16, 2024 Received: August 16, 2024
Dear Felicia Piel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
1
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
2
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
510(k) Number (if known) K240087
Device Name RT Carbon ECG Leads
Indications for Use (Describe)
The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECC trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT Carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.
Type of Use (Select one or both, as applicable) | |
---|---|
[x] Prescription Use (Part 21 CFR 801 Subpart D) | [ ] Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/0 description: The image shows the logo for Ivy Biomedical. The logo consists of the word "Ivy" in a stylized font, with a red triangle above the "I". To the right of "Ivy" is the word "BIOMEDICAL" in a smaller, sans-serif font. The logo is simple and modern, and the use of color makes it eye-catching.
510(k) SUMMARY
1 Applicant
Ivy Biomedical Systems, Inc. 11 Business Park Drive Branford, CT 06405 Phone: (203) 481-4183 Fax: (203) 481- 8734
ನ Primary Contact
Felicia Piel Vice President, QA/RA Direct: (203) 643-9933 felicia.piel@ivybiomedical.com
3 Preparation Date: November 7th, 2023
4 Device Name
Proprietary Name: | RT Carbon ECG Leads |
---|---|
Common/Usual Name: | RT Carbon ECG Leads |
5 Device Classification
Establishment Registration Number: | 1221108 |
---|---|
Classification Name: | Patient transducer and electrode cable (including connector) |
Classification Regulation: | 21 CFR 870.2900 |
Device Classification: | Class II |
Product Code: | DSA |
6 Device Description
The RT Carbon ECG Leads are intended to connect electrodes placed on the patient to a physiological detection module or monitor (CTM-300, CTM-400, CTM-500, 7000 Series and 3000 Series). The design includes 3 and 4 lead designs with overall lengths of 24", 30" and 36".
7 Intended Use
The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECG trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT
5
Image /page/5/Picture/0 description: The image shows the logo for Ivy Biomedical. The logo consists of the word "Ivy" in a stylized font, with a red triangle above the "I". To the right of "Ivy" is the word "BIOMEDICAL" in a simple, sans-serif font.
carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.
Predicate Device 8
Predicate | Clearance 510(k) |
---|---|
Vital Connections Stealthshield | [K970755] |
9 Technical Characteristics
The RT Carbon ECG Leads have the same technical characteristics as the predicate Vital Connections Stealthshield [K970755]. It is a 3 or 4 lead, non-magnetic, carbon lead set.
The RT Carbon ECG Lead sets primary differences from the predicate are listed below:
- The Stealthshield [K970755] is sold in a 40" variation
10 Comparison of Technological Characteristics with the Predicate Device
ECG monitoring is the technological principle for both the subject and the predicate devices. At a high level, the subject and predicate devices are based on the following same technological elements:
- Connect to ECG electrodes
- Intended for use in X-ray imaging environments ●
- Shieleded ECG Plug EC53 compliant connector ●
- Grabber clip leads ●
- RF shielded ●
- 3 or 4 lead wires
11 Performance Data
a. Biocompatibility Testing
The biocompatibility evaluation for the RT Carbon ECG Leads was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part I: Evaluation and Testing'" May 1, 1995, and Internation Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part I: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. This includes the following tests:
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
6
Image /page/6/Picture/0 description: The image shows the logo for Ivy Biomedical. The logo consists of the word "Ivy" in a stylized font, with a red triangle above the "i". To the right of "Ivy" is the word "BIOMEDICAL" in a simple, sans-serif font. The logo is clean and modern, and the use of color is eye-catching.
- Systemic Toxicity ●
- Pyrogen Testing .
The Ivy Biomedical RT Carbon ECG Lead Wires categorization by duration of contact under section 5.3 is Limited Exposure (A). The use of the RT Carbon ECG Leads during prep and scan is limited and less than 24 hours.
b. Performance Testing
The RT Carbon ECG Leads were assessed, thoroughly tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900. The following bench testing was completed at Ivy in accordance to EC53:
- Patient Leadwire to Trunk Cable Interconnection ●
- Flex life
- Tensile strength ●
- Connector mating/unmating cycles ●
- Connector retention force ●
- Contact resistance ●
- Leadwire resistance ●
- Dielectric withstand voltage .
12 Animal Study
Animal studies were not performed.
13 Clinical Test Conclusion
Clinical trials were not performed.
14 Non-Clinical Test Conclusion
The performance data support the safety of the device and hardware verification and validation demonstrate that the RT Carbon ECG Leads should perform as intended in the specified use conditions and is substantially equivalent to the predicate.