(536 days)
Medicon Hellas Albumin: Reagent for the quantitative measurement of albumin in serum. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Medicon Hellas Calcium: Reagent for the quantitative measurement of calcium in serum or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Medicon Hellas Creatinine: Reagent for the quantitative measurement of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Medicon Hellas Glucose: Reagent for the quantitative measurement of glucose in serum and urine. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Medicon Hellas Direct Bilirubin; Reagent for the quantitative measurement of direct bilirubin (conjugated) in serum. Measurements of the level of direct bilirubin is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall blader block.
Medicon Hellas Total Bilirubin: Reagent for the quantitative measurements of total bilirubin in serum. Measurements of the levels of total bilirubin is used in the diagnosis and treatment of liver. hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Medicon Hellas Urea Nitrogen: Reagent is for the quantitative measurement of urea nitrogen in serum and urine. Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
The Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Glucose, Medicon Hellas Direct Bilirubin, Medicon Hellas Total Bilirubin, and Medicon Hellas Urea Nitrogen are reagents for use with Diatron Pictus 500 Clinical Chemistry Analyzers. They are test systems for the quantitative measurement of albumin, calcium, creatinine, glucose, direct and total bilirubin, and urea nitrogen in human serum and urine where clinically applicable. The methods employed are photometric, utilizing reactions between the sample and reagents to produce a colored chromophore or a change in absorbance that is proportional to the concentration of the analyte. The analyzer photometer reads the absorbances at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically.
The provided text describes the performance of several Medicon Hellas assays (Albumin, Calcium, Creatinine, Glucose, Direct Bilirubin, Total Bilirubin, and Urea Nitrogen) when run on the Diatron Pictus 500 Clinical Chemistry Analyzer, demonstrating their substantial equivalence to predicate devices (Beckman Coulter AU reagents on AU2700 analyzer, and Abbott Architect Direct Bilirubin on Architect c8000 analyzer).
Here's an analysis of the provided information, structured to address your specific points regarding acceptance criteria and study details:
1. A Table of Acceptance Criteria and the Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a single, overarching table with pass/fail remarks. Instead, it describes each performance characteristic and then presents the results. The "Summary" sections for each study type imply that the results met the pre-defined acceptance criteria for demonstrating substantial equivalence. For instance, for accuracy, it states "Accuracy studies completed on at least three lots of each candidate reagent confirm that Medicon albumin... are substantially equivalent to the related predicate devices." This implies that the statistical analyses (Deming regression, R2, slope, intercept) fell within acceptable ranges. Similarly, for precision, it states "All lots passed acceptance criteria for each applicable sample type at each level."
Since explicit acceptance criteria are not presented, they are inferred from the demonstrated performance and the statement that the devices "passed acceptance criteria" or "met statistical acceptance criteria." Below is a table summarizing the reported device performance for each analyte. The "Acceptance Criteria" column will reflect the general statements of success or the implied ranges from the results themselves, as explicit numerical targets for individual tests are not given.
Implied Acceptance Criteria and Reported Device Performance
| Analyte | Performance Characteristic | Implied/General Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Medicon Hellas Albumin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence (>0.98 is generally good) | R2 = 0.9862 |
| Slope | Values close to 1 | Slope = 1.0180 | |
| Intercept | Values close to 0 | Intercept = 0.05 | |
| Reportable Range (Serum) | Comparable to predicate device | 1.50 - 6.00 g/dL | 1.50 - 6.00 g/dL |
| LOD (Serum) | Acceptable for clinical use | 0.40 g/dL | 0.40 g/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.50 g/dL | 0.50 g/dL |
| Precision (Serum) | CV% within acceptable clinical limits (e.g., typically <10% for these analytes) | Repeatability CV (Level 1,2,3): 2.21%, 1.45%, 1.37% | Between Run CV (Level 1,2,3): 5.25%, 3.80%, 2.36% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations (e.g., Hemoglobin 500 mg/dL, Triglycerides 3000 mg/dL) | (See document for detailed thresholds per interferent) |
| Medicon Hellas Calcium | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9949, Urine R2 = 0.9965 |
| Slope | Values close to 1 | Serum Slope = 1.0099, Urine Slope = 0.9888 | |
| Intercept | Values close to 0 | Serum Intercept = -0.3, Urine Intercept = -0.8 | |
| Reportable Range | Comparable to predicate device | Serum: 4.0 - 18.0 mg/dL, Urine: 2.0 - 40.0 mg/dL | Serum: 4.0 - 18.0 mg/dL, Urine: 2.0 - 40.0 mg/dL |
| LOD | Acceptable for clinical use | Serum: 0.5 mg/dL, Urine: 1.3 mg/dL | Serum: 0.5 mg/dL, Urine: 1.3 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 0.5 mg/dL, Urine: 1.5 mg/dL | Serum: 0.5 mg/dL, Urine: 1.5 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.16%, 1.06%, 0.82% | Between Run CV (L1,2,3): 3.37%, 1.51%, 1.95% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Creatinine | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9989, Urine R2 = 0.9992 |
| Slope | Values close to 1 | Serum Slope = 1.0207, Urine Slope = 0.9904 | |
| Intercept | Values close to 0 | Serum Intercept = -0.10, Urine Intercept = -0.81 | |
| Reportable Range | Comparable to predicate device | Serum: 0.3 - 25.0 mg/dL, Urine: 1.2 - 300.0 mg/dL | Serum: 0.3 - 25.0 mg/dL, Urine: 1.2 - 300.0 mg/dL |
| LOD | Acceptable for clinical use | Serum: 0.2 mg/dL, Urine: 1.0 mg/dL | Serum: 0.2 mg/dL, Urine: 1.0 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 0.2 mg/dL, Urine: 1.1 mg/dL | Serum: 0.2 mg/dL, Urine: 1.1 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 2.41%, 1.08%, 1.04% | Between Run CV (L1,2,3): 3.63%, 4.58%, 2.30% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Direct Bilirubin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | R2 = 0.9978 |
| Slope | Values close to 1 | Slope = 0.9656 | |
| Intercept | Values close to 0 | Intercept = -0.01 | |
| Reportable Range (Serum) | Comparable to predicate device | 0.2 - 15.0 mg/dL | 0.2 - 15.0 mg/dL |
| LOD (Serum) | Acceptable for clinical use | 0.1 mg/dL | 0.1 mg/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.2 mg/dL | 0.2 mg/dL |
| Precision (Serum) | CV% within acceptable clinical limits | Repeatability CV (L1,2,3): 3.11%, 2.46%, 2.48% | Between Run CV (L1,2,3): 2.31%, 3.14%, 2.29% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Glucose | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9992, Urine R2 = 0.9989 |
| Slope | Values close to 1 | Serum Slope = 0.9715, Urine Slope = 1.0222 | |
| Intercept | Values close to 0 | Serum Intercept = 2.7, Urine Intercept = -0.9 | |
| Reportable Range | Comparable to predicate device | Serum: 10 - 700 mg/dL, Urine: 10 - 660 mg/dL | Serum: 10 - 700 mg/dL, Urine: 10 - 660 mg/dL |
| LOD | Acceptable for clinical use | Serum: 1.7 mg/dL, Urine: 2.4 mg/dL | Serum: 1.7 mg/dL, Urine: 2.4 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 4.0 mg/dL, Urine: 6.0 mg/dL | Serum: 4.0 mg/dL, Urine: 6.0 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.72%, 0.99%, 0.96% | Between Run CV (L1,2,3): 1.35%, 1.48%, 2.08% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Total Bilirubin | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | R2 = 0.9996 |
| Slope | Values close to 1 | Slope = 1.0125 | |
| Intercept | Values close to 0 | Intercept = -0.06 | |
| Reportable Range (Serum) | Comparable to predicate device | 0.10 - 30.00 mg/dL | 0.10 - 30.00 mg/dL |
| LOD (Serum) | Acceptable for clinical use | 0.01 mg/dL | 0.01 mg/dL |
| LOQ (Serum) | Acceptable for clinical use | 0.09 mg/dL | 0.09 mg/dL |
| Precision (Serum) | CV% within acceptable clinical limits | Repeatability CV (L1,2,3): 0.92%, 0.38%, 0.57% | Between Run CV (L1,2,3): 1.69%, 1.38%, 1.79% |
| Interferences (Serum) | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
| Medicon Hellas Urea Nitrogen | |||
| Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence | Serum R2 = 0.9983, Urine R2 = 0.9972 |
| Slope | Values close to 1 | Serum Slope = 1.0001, Urine Slope = 0.9844 | |
| Intercept | Values close to 0 | Serum Intercept = -0.2, Urine Intercept = 21.9 | |
| Reportable Range | Comparable to predicate device | Serum: 3 - 100 mg/dL, Urine: 24 - 1300 mg/dL | Serum: 3 - 100 mg/dL, Urine: 24 - 1300 mg/dL |
| LOD | Acceptable for clinical use | Serum: 2 mg/dL, Urine: 21 mg/dL | Serum: 2 mg/dL, Urine: 21 mg/dL |
| LOQ | Acceptable for clinical use | Serum: 3 mg/dL, Urine: 24 mg/dL | Serum: 3 mg/dL, Urine: 24 mg/dL |
| Precision | CV% within acceptable clinical limits | Serum: Repeatability CV (L1,2,3): 1.94%, 2.14%, 1.07% | Between Run CV (L1,2,3): 2.24%, 2.56%, 3.16% |
| Interferences | Recovered values within ±10% of neat samples | Insignificant interference up to noted concentrations | (See document for detailed thresholds per interferent) |
2. Sample sizes used for the test set and the data provenance:
-
Sample Size for Test Set:
- Accuracy (Method Comparison): "A minimum of 70 clinical specimens, spanning the dynamic ranges, were assayed." Specific numbers are provided per analyte:
- Medicon Hellas Albumin: 112 samples (Serum)
- Medicon Hellas Calcium: 94 samples (Serum), 81 samples (Urine)
- Medicon Hellas Creatinine: 126 samples (Serum), 98 samples (Urine)
- Medicon Hellas Direct Bilirubin: 77 samples (Serum)
- Medicon Hellas Glucose: 99 samples (Serum), 100 samples (Urine)
- Medicon Hellas Total Bilirubin: 95 samples (Serum)
- Medicon Hellas Urea Nitrogen: 116 samples (Serum), 81 samples (Urine)
- Reportable Range (Linearity): "At least nine levels of each sample types were tested." (N=4 per level on Pictus P500)
- Sensitivity (LOD/LOQ):
- LoB/LoD: "5 Blank samples and 5 Low Levels samples respectively which were measured 4 times each day for a total of 60 measurements in 3 days."
- LoQ: "10 samples that span the low end of linearity were measured 5 times each day for a total of 150 measurements in 3 days."
- Interferences: "Serum and urine sample pools at low and high levels were prepared." The exact number of individual samples forming these pools is not specified beyond being "pools."
- Precision: "Precision study results from running applicable serum and urine samples (Level 1, Level 2 and Level 3) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level."
- Accuracy (Method Comparison): "A minimum of 70 clinical specimens, spanning the dynamic ranges, were assayed." Specific numbers are provided per analyte:
-
Data Provenance (e.g., country of origin of the data, retrospective or prospective):
- The document implies that the studies were conducted by Medicon Hellas, S.A. in Greece, given their address on the first page.
- The data appears to be prospective as it describes experiments conducted ("studies were performed," "testing confirmed," "protocol followed"). It references the collection and analysis of clinical specimens specifically for these validation studies. It does not mention retrospective analysis of existing patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable in the context of this 510(k) submission. This type of submission is for in vitro diagnostic (IVD) reagents, which measure specific analytes in bodily fluids. The "ground truth" for these measurements is typically established by comparative analysis against predicate devices and well-characterized reference methods (e.g., those detailed in CLSI guidelines for accuracy, linearity, precision). These are not image-based AI models requiring human expert interpretation for ground truth.
The "experts" involved would be the laboratory personnel performing the assays according to established clinical laboratory standards and the statistical analysis, rather than medical experts providing subjective interpretations. The document does not specify the number or qualifications of the laboratory personnel who conducted the tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. This concept (e.g., 2+1, 3+1 reader adjudication) is primarily used in studies where human readers provide subjective interpretations (e.g., radiology studies). For IVD devices, the "ground truth" is based on the analytical performance against established reference methods or predicate devices, which involves quantitative measurements and statistical analysis, not human adjudication of subjective interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI-assisted diagnostic imaging device. It is a chemical reagent intended for quantitative measurement of analytes in bodily fluids. Therefore, MRMC studies and the effect size on human readers are not relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The device (reagent) essentially operates in a "standalone" fashion in terms of its chemical reaction and measurement, independent of human interpretive intervention for the measurement itself. The performance data (accuracy, precision, linearity, etc.) presented is the standalone performance of the reagent on the specified analyzer. Human involvement is in operating the instrument, quality control, and interpreting the numerical results in a clinical context, but not in the measurement process being tested for substantial equivalence.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" in this context is established by:
- Comparative Method: The primary method for proving substantial equivalence is by demonstrating strong correlation and agreement with legally marketed predicate devices (Beckman Coulter AU reagents and Abbott Architect Direct Bilirubin reagent) using real patient clinical specimens. This acts as the standard for comparison.
- CLSI Guidelines: Various performance characteristics (linearity, sensitivity, precision, interferences) were evaluated according to Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., CLSI EP09c for accuracy, CLSI EP06-A for linearity, CLSI EP17-A2 for sensitivity, CLSI EP07-A and EP37 for interferences, CLSI EP05-A3 for precision). These guidelines represent accepted industry standards for validating in vitro diagnostic devices, thereby defining the "ground truth" for these analytical measurements.
- Reference Values: For linearity studies, "samples were assigned their reference values arithmetically from serial dilutions of the high-level sample," indicating a quantitatively derived reference for linearity.
Therefore, the ground truth is based on a combination of comparison to predicate devices and adherence to established analytical reference methods and industry standards (CLSI guidelines).
8. The sample size for the training set:
Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense. It is a chemical reagent. The "training" here would be the chemical formulation and manufacturing process, which is established through R&D and QA/QC, not data input to an algorithm.
9. How the ground truth for the training set was established:
Not applicable, as there is no "training set" for a chemical reagent. The "ground truth" for the development of the reagent itself would be the established chemical principles and desired analytical performance characteristics (e.g., reactivity, specificity, stability, sensitivity) based on scientific literature and previous experience with similar assays.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 21, 2021
Medicon Hellas, S.A. Sotia Mitropoulos Regulatory Manager Melitona 5-7, 153 44, Gerakas, Greece
Re: K200898
Trade/Device Name: Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Direct Bilirubin, Medicon Hellas Glucose, Medicon Hellas Total Bilirubin, Medicon Hellas Urea Nitrogen Regulation Number: 21 CFR 862.1035 Regulation Name: Albumin Test System Regulatory Class: Class II Product Code: CIX, CDQ, CJY, CGX, CIG, CFR Dated: February 24, 2021 Received: February 24, 2021
Dear Sotia Mitropoulos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Marianela Perez-torres -S
Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200898
Device Name
Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Total Bilirubin, Medicon Hellas Direct Bilirubin and Medicon Hellas Urea Nitrogen.
Indications for Use (Describe)
Medicon Hellas Albumin: Reagent for the quantitative measurement of albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Medicon Hellas Calcium: Reagent for the quantitative measurement of calcium in serum or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Medicon Hellas Creatinine: Reagent for the quantitative measurement of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Medicon Hellas Glucose. Reagent for the quantitative measurement of glucose in serum and urine. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Medicon Hellas Direct Bilirubin; Reagent for the quantitative measurement of direct bilirubin (conjugated) in serum. Measurements of the level of direct bilirubin is used in the diagnosis and treatment of liver, hematological, and metabolic disorders, including hepatitis and gall blader block.
Medicon Hellas Total Bilirubin; Reagent for the quantitative measurements of total bilirubin in serum. Measurements of the levels of total bilirubin is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Medicon Hellas Urea Nitrogen: Reagent for the quantitative measurement of urea nitrogen in serum and urine. Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510k (K200898) safety and effectiveness information is re-submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 807.92 (a)(1): Name: | MEDICON HELLAS S.A. |
|---|---|
| Address: | Melitona 5-7, 153 44 Gerakas, Greece |
| Phone: | +302.1.0660.6000 |
| FAX: | +302.1.0661.2666 |
| Contact: | Ms. Sotia Mitropoulos |
| Email: | mitropoulos@mediconsa.com |
| US Agent: | Diatron US, Inc.1.833.228.7931Mr. Frank Matuszakfrank.matuszak@diatron.com |
807.92 (a)(2): Device name- trade name and common name, and classification
Trade Name: Medicon Hellas Albumin. Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Glucose, Medicon Hellas Direct Bilirubin, Medicon Hellas Total Bilirubin, and Medicon Hellas Urea Nitrogen, for use with Diatron Pictus 500 Clinical Chemistry Analyzers.
Common Name: test systems for albumin, calcium, creatinine, glucose, direct and total bilirubin and urea nitrogen for testing human serum and urine where clinically applicable.
Classification Name(s):
21 CFR § 862.1035 - Albumin Test system - Class II- Product Code CIX
21 CFR § 862.1145 - Calcium Test system - Class II- Product Code CJY
21 CFR § 862.1225 - Creatinine Test system - Class II- Product Code CGX
21 CFR & 862.1110 - Direct Bilirubin Test system - Class II- Product Code CIG
21 CFR § 862.1345 - Glucose Test system - Class II- Product Code CFR
21 CFR § 862.1110 - Total Bilirubin Test system - Class II- Product Code CIG
21 CFR § 862.1770 - Urea Nitrogen Test system - Class II- Product Code CDQ
{4}------------------------------------------------
807.92 (a)(3): Identification of the legally marketed predicate devices
- 21 CFR § 862.1035 Albumin Test system Class II-Beckman Coulter Albumin (K924368)
- 21 CFR § 862.1145 Calcium Test system Class II-Beckman Coulter Calcium (K061575)
- 21 CFR & 862.1225 Creatinine Test system- Class II-Beckman Coulter Creatinine (K934361)
- 21 CFR § 862.1345 Glucose Test system Class II-Beckman Coulter Glucose (K944406)
- 21 CFR & 862.1110 Direct Bilirubin Test system Class II-Abbott Architect Direct Bilirubin (K022180)
- 21 CFR § 862.1110 Total Bilirubin Test system Class II-Beckman Coulter Total Bilirubin (K924964)
- 21 CFR & 862.1770 Urea Nitrogen Test system Class II-Beckman Coulter Urea Nitrogen (BUN) (K944250)
807.92 (a)(4): Device Description
Medicon Hellas Albumin:
The BCG method is applied on serum samples, without serum blank. The bi-chromatic photometric determination of albumin is based on albumin binding the anionic dye bromocresol green in mildly acidic pH. The raise in absorbance at 620/750 nm is proportional to the concentration of albumin in the sample. The sample volume used for the assay is 2uL.
Medicon Hellas Calcium:
Calcium ions react with Arsenazo III dye in acidic pH, to form a violet complex. The absorbance at 650/750 nm is proportional to the concentration of calcium in serum or urine samples. The analyzer photometer reads the absorbance at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically. The sample volume used by the assay is 2uL for serum or urine samples.
Medicon Hellas Creatinine:
Creatinine forms a yellow-orange complex with picrate anions in alkaline medium. The rate of change of absorbance of the complex at 505/700 nm is proportional to creatinine concentration in serum or urine samples. The sample volume used by the assay is 22uL for serum and urine samples.
{5}------------------------------------------------
Medicon Hellas Glucose: Glucose (Hexokinase)
The determination of glucose is according to the hexokinase method based on the following reactions using 2uL of sample for serum and urine samples. The concentration of glucose is determined bi-chromatically by the absorbance change at 340/380 nm.
Hexokinase Glucose + ATP 6-phosphate glucose + ADP 6-phosphate glucose + NAD
Medicon Hellas Direct Bilirubin :
The modified Malloy-Evelyn method is applied on serum samples. Direct (conjugated) bilirubin reacts with diazotised sulfanilic acid in acidic pH to form azobilirubin, a colored chromophore. The absorbance of chromophore at 550/650 nm is proportional to the concentration of direct bilirubin in the sample. The free (unbound) bilirubin fraction does not react. The reaction is an end-point type. The sample volume used by the assay is 8uL.
Medicon Hellas Total Bilirubin:
The DPD method is applied, whereby 3,5-dichloropheny|diazonium tetrafluoroborate (DPD) binds directly to total bilirubin in acidic medium, forming azobilirubin, whose rate of absorbance at 550/700 nm is proportional to the concentration of total bilirubin in serum samples. The analyzer photometer reads the absorbance at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically. The sample volume used by the assay is 20uL.
Medicon Hellas Urea Nitrogen:
Medicon Hellas Urea Nitrogen is an in vitro diagnostic assay for the quantitative determination of urea nitrogen in human serum and urine. Urea Nitrogen is determined according to the following reactions, using 7uL of sample for serum and urine samples:
Urease Urea Nitrogen + H2O ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+ 2NH3 + CO2 GLDH 2NH4+ + 2-a-Ketoglutarate + 2NADH -2-L-Glutamate + 2NAD+ 2H2O 1
The rate of absorbance change at 340/380 nm is proportional to the urea nitrogen concentration in the sample.
For all assays the analyzer photometer reads the absorbances at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically.
{6}------------------------------------------------
807.92 (a)(5): Intended Use
Medicon Hellas Albumin: Reagent for the quantitative measurement of albumin in serum. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Medicon Hellas Calcium: Reagent for the quantitative measurement of calcium in serum or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Medicon Hellas Creatinine: Reagent for the quantitative measurement of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Medicon Hellas Glucose: Reagent for the quantitative measurement of glucose in serum and urine. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Medicon Hellas Direct Bilirubin; Reagent for the quantitative measurement of direct bilirubin (conjugated) in serum. Measurements of the level of direct bilirubin is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall blader block.
Medicon Hellas Total Bilirubin: Reagent for the quantitative measurements of total bilirubin in serum. Measurements of the levels of total bilirubin is used in the diagnosis and treatment of liver. hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Medicon Hellas Urea Nitrogen: Reagent is for the quantitative measurement of urea nitrogen in serum and urine. Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
{7}------------------------------------------------
| Characteristic | Candidate- Medicon Hellas(Medicon) reagents | Predicate- Beckman Coulter (BC) AUReagents run on the BC AU2700analyzer | Predicate- Abbott Diagnostics (AD)Direct Bilirubin run on the Architectc8000 analyzer |
|---|---|---|---|
| Mode of Detection | Photometric | Same | Same |
| Photometric Detector | Photodiode | Same | Same |
| Optical Modes | Monochromatic, Bi-chromatic,Turbidimetric | Same | Same |
| Analytical Measure | Endpoint and kinetic with sampleblanking | Same | Same |
| Wavelengths | 12 (340 - 750nm) | 13 (340 - 800nm) | 16 (340 - 804nm) |
| Linear Absorbance Range | -0.1 to 3.6 | 0.0 to 3.0 | -0.1 to 3.2 |
| Device Class, Regulation Code, ProductCode | Class I, 21 CFR 862.2160 (discretephotometric chemistry analyzer), JJE | Same | Same |
| Reagent Platform | Candidate- Medicon Hellas reagents | Predicate - Beckman Coulter AUReagents | Predicate - Abbott Diagnostics DirectBilirubin reagent |
| Albumin | Albumin (K924368) | ||
| Calcium | Calcium (K061575) | ||
| Creatinine (picrate based) | Creatinine (K934361) | ||
| Test Systems | Glucose | Glucose (K944406) | |
| Direct Bilirubin | N/A | Direct Bilirubin (K022180) | |
| Total Bilirubin | Total Bilirubin (K924964) | ||
| Urea Nitrogen | Urea Nitrogen (K944250) | ||
| Albumin- 21 CFR § 862.1035, CIX | Same | ||
| Calcium- 21 CFR § 862.1145, CJY | Same | ||
| Creatinine- 21 CFR § 862.1225, CGX | Same | ||
| Device Class II, Regulation Code,Classification Product Code | Glucose- 21 CFR § 862.1345, CFR | Same | |
| Direct Bilirubin- 21 CFR § 862.1110, CIG | Same | ||
| Total Bilirubin-21 CFR § 862.1110, CIG | Same | ||
| Urea Nitrogen - 21 CFR § 862.1770, CDQ | Same | ||
| Intended Use | Quantitative determinations of albumin,calcium, creatinine, direct bilirubin,glucose, total bilirubin and urea withserum or urine samples where clinicallyapplicable. | Quantitative determinations of albumin,calcium, creatinine, glucose, total bilirubinand urea with serum or urine sampleswhere clinically applicable. | Quantitative determinations of directbilirubin on human serum samples. |
| Testing Environment | Clinical labs | Same | Same |
| Method Principle | Photometry | Same | Same |
| Specimen Type | Human serum, and urine if applicable | Same | Same |
807.92 (a)(6): Technological Similarities and Differences to the Predicates
{8}------------------------------------------------
807.92 (b)(1,2): Brief Description of Nonclinical and Clinical Data
Medicon Hellas (Medicon) has performed a series of studies to confirm the substantial equivalence of their test systems to the predicate devices. Results from each candidate test system were found to be substantially equivalent to their related predicate device. The predicate device used for Medicon Albumin, Medicon Calcium, Medicon Creatinine, Medicon Total Bilirubin and Medicon Urea Nitrogen (BUN) was the Beckman Coulter (BC) AU 2700 analyzer using the corresponding AU reagents. The predicate device for Medicon Direct Bilirubin was the Abbott Diagnostics Architect c8000 analyzer using its corresponding reagent.
Two to three lots of each candidate device were used for the studies. Testing confirmed equivalent performance of the candidate systems to the corresponding predicate devices.
Studies included accuracy (method comparison), reportable range (linearity), sensitivity (limit of quantification), precision including: repeatability, within run, between run and within lab (total), interfering endogenous and exogenous substances, reagent on-board stability and calibration frequency. The following provide performance summary information from the studies for each candidate reagent. The following summarize the type and scope of testing completed for each candidate device. Following this section, the data for each of the candidate devices is provided.
Method Comparison- The protocol followed was according to CLSI EP09c: accuracy testing was run simultaneously in blind fashion with the candidate and the predicate devices. A minimum of 70 clinical specimens, spanning the dynamic ranges, were assayed in singleton and in a blinded fashion on the at least three lots of each candidate system and on a representative lot of each predicate device. Results were tabulated and evaluated using Analyse-it software to generate Deming regression statistics.
Summary: Accuracy studies completed on at least three lots of each candidate reagent confirm that Medicon albumin, calcium, creatinine, glucose, total bilirubin and urea nitrogen are substantially equivalent to the related predicate devices - Beckman Coulter AU reagents used on the AU2700 analyzer. Medicon direct bilirubin reagent used with a Diatron Pictus 500 analyzer is substantially equivalent to its predicate device - Abbott Architect c8000 direct bilirubin used with the Architect c8000 analyzer.
Reportable Range - Linearity - Statistical analysis is performed at a 5% level of significance and the evaluation is according to CLSI EP06-A. At least nine levels of each sample types were tested with at least three lots of the corresponding Medicon reagents. The samples were assigned their reference values arithmetically from serial dilutions of the high-level sample. Each level was tested N=4 on the Pictus P500. results from each level were averaged. All statistics were calculated using Analyse-it Software.
Summary: the combined data from various lots of each of the Medicon candidate reagents used with the Pictus 500 chemistry analyzers met statistical acceptance criteria. Reportable ranges were comparable to the predicate device reportable ranges.
{9}------------------------------------------------
Sensitivity (LOD/LOQ) - Medicon completed sensitivity studies following CLSI EP17-A2. LoB was calculated using non-parametrical statistics and LoD was calculated from LoB following equation LoD = LoB+1.654*SD:
LoB/LoD were determined using: 5 Blank samples and 5 Low Levels samples respectively which were measured 4 times each day for a total of 60 measurements in 3 days.
LoQ: 10 samples that span the low end of linearity were measured 5 times each day for a total of 150 measurements in 3 days. LoQ was calculated until a specific concentration generated a CV of <= 20%.
Summary - The lots of each of the candidate reagents of demonstrated acceptable levels of blank, limit of detection and defined the limit of quantification similar to the corresponding predicate devices.
Interferences- at least two lots of each candidate device were tested for interferences from endogenous and exogenous compounds using serum and urine samples, where applicable, Medicon performed comprehensive studies after searching in the bibliography and the FDA database for similar recently cleared devices, to assess and define a list of exogenous substances. The study plan and protocol followed CLSI EP07-A and EP37 guidelines. Serum and urine sample pools at low and high levels were prepared, then spiked with increasing concentrations of the interference compounds. The results define the level at which interferences of the noted compounds will affect results by more than +/- 10%.
Summary - for each of the candidate devices, interferences studies determined the levels at which specific substances may interfere with results. These are tabulated in each of the candidate devices' data tables.
Stability/Calibration frequency:
Different lots of each candidate device were evaluated for on-board stability and calibration frequency covering specific date intervals and the related sample types. The results were tabulated and the data were analyzed for equivalence between lots. Results means, standard deviations and percent coefficient of variation (CV%).
Stability - Reagents were loaded on to the Pictus 500 analyzer's cooled reagent tray. Each reagent was calibrated at the start of each study. Two fresh serum pools and two control pools were assayed in triplicate on Day 0. Samples were frozen from -20℃ to - 70℃, and then retested at fixed intervals until the reagent failed the acceptance criteria. When a reagent failed acceptance criteria, it was recalibrated. If recalibration did not bring the results within the acceptance criteria, the reagent's on-board useful life limit was established.
Calibration frequency: The un-capped on-board reagent stability claim is defined as the number of days between the start testing date and the last test day that results were within the acceptance limits without re-calibration. Calibration frequency is defined as the required time interval between calibrations that ensures reagents recover values within the acceptance criteria.
{10}------------------------------------------------
Summary - stability -Each lot of each candidate device, with the corresponding sample types passed acceptance criteria. Studies confirmed the on-board stability and calibration frequency limits defined within each candidate device's Instructions for Use document (product insert).
Precision CV% - (repeatability, between run and within lab (total) precision) - testing precisely followed protocol CLSI EP05-A3. Precision study results from running applicable serum and urine samples (Level 1, Level 2 and Level 3) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level on two or three lots of each candidate device. The three precision elements were calculated from the total data for each sample type and lot generated over the evaluation period. Statistics were calculated using Analyse-it software.
Summary - The results were tabulated and the data were analyzed for means, standard deviations and percent coefficient of variation (CV%). All lots passed acceptance criteria for each applicable sample type at each level.
Tabulated results – for each of the candidate devices follow:
| Reportable Range | Low Value High Value | ||
|---|---|---|---|
| serum | 1.50 | 6.00 | g/dL |
| LOD (limit of detection) - serum | 0.40 | g/dL | |
| LOQ (limit of quantification) -serum | 0.50 | g/dL | |
| Precision (g/dL) -serum | Level 1 | SD | CV |
| Repeatability | 3.1 | 0.07 | 2.21% |
| Between Run | 3.1 | 0.16 | 5.25% |
| Within Lab | 3.1 | 0.18 | 5.70% |
| Level 2 | SD | CV | |
| Repeatability | 4.0 | 0.06 | 1.45% |
| Between Run | 4.0 | 0.15 | 3.80% |
| Within Lab | 4.0 | 0.19 | 4.80% |
| Level 3 | SD | CV | |
| Repeatability | 5.2 | 0.07 | 1.37% |
| Between Run | 5.2 | 0.12 | 2.36% |
| Within Lab | 5.2 | 0.24 | 4.61% |
| Interferences - serum | Recovered values within ±10% of neat samples -insignificant up to | ||
| Unconjugated Bilirubin | insignificant up to | 20 mg/dL | |
| Conjugated Bilirubin | insignificant up to | 20 mg/dL | |
| Hemoglobin | insignificant up to | 500 mg/dL | |
| Triglycerides | insignificant up to | 3000 mg/dL | |
| L-ascorbic acid | insignificant up to | 3 mg/dL | |
| Acetametophen | insignificant up to | 16 mg/dL | |
| Acetylsalicylic Acid | insignificant up to | 3 mg/dL | |
| Aminosalicylic Acid | insignificant up to | 46 mg/dL | |
| Ibuprofen | insignificant up to | 21.9 mg/dL |
Medicon Hellas Albumin - code # 1419-0192
Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a BECKMAN COULTER AU 2700 analyzer, each run with their corresponding reagents. Calculated results were as follows:
| Range | # Samples | R2 Corr. | Slope | Intercept | |
|---|---|---|---|---|---|
| Serum | 1.6 to 5.9 | 112 | 0.9862 | 1.0180 | 0.05 |
{11}------------------------------------------------
| Reportable Range | Low Value | High Value | ||||||
|---|---|---|---|---|---|---|---|---|
| Serum | 4.0 | 18.0 | mg/dL | |||||
| Urine | 2.0 | 40.0 | mg/dL | |||||
| Serum | Urine | |||||||
| LoD (limit of detection) | 0.5 | mg/dL | 1.3 | mg/dL | ||||
| LoQ (limit of quantification) | 0.5 | mg/dL | 1.5 | mg/dL | ||||
| Precision Serum (mg/dL) | Level 1 | SD | CV | |||||
| Repeatability | 5.9 | 0.07 | 1.16% | |||||
| Between Run | 5.9 | 0.20 | 3.37% | |||||
| Within Lab (Total) | 5.9 | 0.32 | 5.44% | |||||
| Level 2 | SD | CV | ||||||
| Repeatability | 8.6 | 0.09 | 1.06% | |||||
| Between Run | 8.6 | 0.13 | 1.51% | |||||
| Within Lab (Total) | 8.6 | 0.35 | 4.07% | |||||
| Level 3 | SD | CV | ||||||
| Repeatability | 12.7 | 0.10 | 0.82% | |||||
| Between Run | 12.7 | 0.25 | 1.95% | |||||
| Within Lab (Total) | 12.7 | 0.44 | 3.49% | |||||
| Precision Urine (mg/dL) | Level 1 | SD | CV | |||||
| Repeatability | 6.2 | 0.16 | 2.59% | |||||
| Between Run | 6.2 | 0.09 | 1.48% | |||||
| Within Lab (Total) | 6.2 | 0.23 | 3.73% | |||||
| Level 2 | SD | CV | ||||||
| Repeatability | 12.9 | 0.25 | 1.90% | |||||
| Between Run | 12.9 | 0.49 | 3.82% | |||||
| Within Lab (Total) | 12.9 | 0.67 | 5.21% | |||||
| Level 3 | SD | CV | ||||||
| Repeatability | 30.6 | 0.40 | 1.31% | |||||
| Between Run | 30.6 | 0.90 | 2.95% | |||||
| Within Lab (Total) | 30.6 | 1.07 | 3.51% | |||||
| Interferences | Recovered values within ±10% of neat samples - insignificant up to noted values | |||||||
| Serum | Urine | |||||||
| Lv1 | Lv2 | Up to | Lv1 | Lv2 | Up to | |||
| Bilirubin | 7.0 | 13.6 | 20 mg/dL | N/A | N/A | N/A | ||
| Hemoglobin | 7.2 | 13.6 | 500 mg/dL | 6.2 | 11.9 | 500 mg/dL | ||
| Triglycerides | 7.3 | 12.8 | 3000 mg/dL | N/A | N/A | N/A | ||
| L-ascorbic acid | 7.3 | 14.2 | 3 mg/dL | 8 | 12.1 | 500 mg/dL | ||
| Acetaminophen | 7.2 | 11.8 | 15.6 mg/dL | N/A | N/A | N/A | ||
| Acetylsalycilic Acid | 7.7 | 12.8 | 3 mg/dL | N/A | N/A | N/A | ||
| Albumin | N/A | 6.5 | 12.7 | 500 | ||||
| Boric Acid | N/A | 7.2 | 15.6 | 500 | ||||
| Conjugated Bilirubin | 8.3 | 12.6 | 20 mg/dL | 7.6 | 16.5 | 50 mg/dL | ||
| Glucose | N/A | 6.8 | 11.9 | 10 mg/dL | ||||
| Ibuprofen | 8.2 | 13.1 | 21.9 mg/dL | 9.7 | 15.7 | 21.9 mg/dL | ||
| IgG | N/A | 6.6 | 15.2 | 500 | ||||
| Magnesium | N/A | 7.4 | 12.8 | 4 | ||||
| pH | N/A | 6.1 | 12.7 | 4 | ||||
| Specific gravity | N/A | 6.1 | 9.7 | 1.01-1.03 |
Medicon Hellas Calcium – code # 1419-0201
Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a BECKMAN COULTER AU 2700 analyzer, each run with their corresponding reagents. Representive results as follows;
| Sample type | Low Value | High Value | # Samples | R2 Corr. | Slope | Intercept |
|---|---|---|---|---|---|---|
| Serum (mg/dL) | 4.0 | 18.0 | 94 | 0.9949 | 1.0099 | -0.3 |
| Urine (mg/dL) | 2.3 | 39.9 | 81 | 0.9965 | 0.9888 | -0.8 |
N/A 500 mg/dL N/A mg/dL 500 mg/dL N/A N/A 500 mg/dL 500 mg/dL 50 mg/dL 10 g/dL 21.9 mg/dL 500 mg/dL 4 mmol/L ნ
{12}------------------------------------------------
Medicon Hellas Creatinine – code # 1419-0030
| Reportable Range | Low Value | High Value | ||||
|---|---|---|---|---|---|---|
| Serum | 0.3 | 25.0 | mg/dL | |||
| Urine | 1.2 | 300.0 | mg/dL | |||
| Serum | Urine | |||||
| LoD (limit of detection) | 0.2 | mg/dL | 1.0 | mg/dL | ||
| LoQ (limit of quantification) | 0.2 | mg/dL | 1.1 | mg/dL | ||
| Precision Serum (mg/dL) | Level 1 | SD | CV | |||
| Repeatability | 1.3 | 0.03 | 2.41% | |||
| Between Run | 1.3 | 0.05 | 3.63% | |||
| Within Lab (Total) | 1.3 | 0.08 | 6.40% | |||
| Level 2 | SD | CV | ||||
| Repeatability | 4.9 | 0.05 | 1.08% | |||
| Between Run | 4.9 | 0.22 | 4.58% | |||
| Within Lab (Total) | 4.9 | 0.23 | 4.71% | |||
| Level 3 | SD | CV | ||||
| Repeatability | 9.6 | 0.10 | 1.04% | |||
| Between Run | 9.6 | 0.22 | 2.30% | |||
| Within Lab (Total) | 9.6 | 0.28 | 2.92% | |||
| Precision Urine (mg/dL) | Level 1 | SD | CV | |||
| Repeatability | 63.4 | 1.31 | 2.07% | |||
| Between Run | 63.4 | 1.68 | 2.66% | |||
| Within Lab (Total) | 63.4 | 2.15 | 3.39% | |||
| Level 2 | SD | CV | ||||
| Repeatability | 100.6 | 1.39 | 1.38% | |||
| Between Run | 100.6 | 1.84 | 1.83% | |||
| Within Lab (Total) | 100.6 | 2.31 | 2.29% | |||
| Level 3 | SD | CV | ||||
| Repeatability | 308.6 | 3.81 | 1.23% | |||
| Between Run | 308.6 | 6.16 | 2.00% | |||
| Within Lab (Total) | 308.6 | 7.25 | 2.35% | |||
| Interferences Recovered values within ±10% of neat samples - insignificant | ||||||
| Serum | Insignificant up to | Urine | Insignificant up to | |||
| Bilirubin | 1.5 to 5.2 | 10.0 mg/dL | 84 to 267 | 40 mg/dL | ||
| Hemoglobin | 1.3 to 6.6 | 220 mg/dL | 88 to 269 | 5.0 g/L | ||
| Triglycerides | 1.4 to 7.0 | 1200 mg/dL | N/A | |||
| L-ascorbic acid | 1.4 to 6.8 | 3.0 mg/dL | 84 to 268 | 50 mg/dL | ||
| Acetaminophen | 1.3 to 6.3 | 15.6 mg/dL | N/A | |||
| Acetylsalicylic acid | 1.2 to 5.6 | 3.0 mg/dL | N/A | |||
| Cefazolin | 1.4 to 6.0 | 40 mg/dL | N/A | |||
| Cefoxitin | 1.5 to 5.9 | 6.0 mg/dL | N/A | |||
| Cephalothin | 1.5 to 5.6 | 12 mg/dL | N/A | |||
| Chloramphenicol | 1.0 to 6.0 | 7.8 mg/dL | N/A | |||
| Dopamine | 0.9 to 6.0 | 0.062 mg/dL | N/A | |||
| Etamsylate | 1.6 to 5.6 | 6.0 mg/dL | N/A | |||
| Flucytosine | 1.2 to 5.7 | 20.4 mg/dL | N/A | |||
| Furosemide | 1.6 to 5.6 | 1.59 mg/dL | N/A | |||
| Glucose | N/A | 86 to 273 | 3000 mg/dL | |||
| Ibuprofen | 1.4 to 6.1 | 23.9 mg/dL | N/A | |||
| Streptomycin | 1.3 to 6.1 | 25.8 mg/dL | N/A | |||
| Boric Acid | N/A | 58 to 200 | 500 mg/dL | |||
| pH | N/A | 59 to 137 | pH 2 to pH 12 | |||
| Specific Gravity | N/A | 36 to 202 | 1.01-1.03 | |||
| Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a BECKMAN | ||||||
| COULTER AU 2700 analyzer, each run with their corresponding reagents. Representive results were; | ||||||
| Sample type | Low Value | High Value | # Samples | R2 Corr. | Slope | Intercept |
| Serum (mg/dL) | 0.3 | 25.0 | 126 | 0.9989 | 1.0207 | -0.10 |
| Urine (mg/dL) | 1.9 | 286.9 | 98 | 0.9992 | 0.9904 | -0.81 |
{13}------------------------------------------------
Medicon Hellas Direct Bilirubin – code # 1419-0152
| Reportable Range | Low Limit | High Limit | |
|---|---|---|---|
| serum | 0.2 | 15.0 | mg/dL |
| LOD (limit of detection) -serum | 0.1 | mg/dL | |
| LOQ (limit of quantification) -serum | 0.2 | mg/dL | |
| Precision -serum | Level 1 | SD | CV |
| Repeatability | 0.7 | 0.02 | 3.11% |
| Between Run | 0.7 | 0.02 | 2.31% |
| Total | 0.7 | 0.03 | 4.98% |
| Level 2 | SD | CV | |
| Repeatability | 1.8 | 0.04 | 2.46% |
| Between Run | 1.8 | 0.06 | 3.14% |
| Total | 1.8 | 0.08 | 4.70% |
| Level 3 | SD | CV | |
| Repeatability | 7.5 | 0.19 | 2.48% |
| Between Run | 7.5 | 0.17 | 2.29% |
| Total | 7.5 | 0.34 | 4.52% |
| Serum- Interferences | Recovered values within ±10% of neat samples - insignificant | ||
| Intereferent Substances | Sample Range | Insignificant up to | |
| Hemoglobin | 0.74 to 2.10 | 40 mg/dL | |
| Triglycerides | 0.49 to 1.99 | 3000 mg/dL | |
| L-ascorbic acid | 0.56 to 2.03 | 2.7 mg/dL | |
| Acetaminophen | 0.64 to 1.63 | 15.6 mg/dL | |
| Acetylsalicylic Acid | 0.64 to 1.53 | 3.0 mg/dL |
lbuprofen
Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a Abbott Labs Architect c8000 analyzer, each run with their corresponding reagents. Representative results from the studies are as follows:
0.64 to 1.52
21.9 mg/dL
| Serum (mg/dL) | Range | # Samples | R2 Corr. | Slope | Intercept |
|---|---|---|---|---|---|
| 0.2 to 14.7 | 77 | 0.9978 | 0.9656 | -0.01 |
{14}------------------------------------------------
| Medicon Hellas Glucose - code # 1419-0018 | ||||
|---|---|---|---|---|
| -- | -- | -- | -- | ------------------------------------------- |
Interferences
| Reportable Range | Low Value | High Value | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Serum | 10 | 700 | mg/dL | ||||||
| Urine | 10 | 660 | mg/dL | ||||||
| Serum | Urine | ||||||||
| LoD (limit of detection) | 1.7 | mg/dL | 2.4 | mg/dL | |||||
| LoQ (limit of quantification) | 4.0 | mg/dL | 6.0 | mg/dL | |||||
| Precision - Serum (mg/dL) | Level 1 | SD | CV | Level 2 | SD | CV | Level 3 | SD | CV |
| Repeatability | 66 | 1.1 | 1.72% | 101 | 1.0 | 0.99% | 303 | 2.9 | 0.96% |
| Between Run | 66 | 0.9 | 1.35% | 101 | 1.5 | 1.48% | 303 | 6.3 | 2.08% |
| Within Lab (Total) | 66 | 1.9 | 2.88% | 101 | 1.8 | 1.78% | 303 | 9.5 | 3.13% |
| Precision - Urine (mg/dL) | Level 1 | SD | CV | Level 2 | SD | CV | Level 3 | SD | CV |
| Repeatability | 14 | 0.5 | 3.63% | 134 | 2.2 | 1.67% | 532 | 7.9 | 1.49% |
| Between Run | 14 | 0.2 | 1.12% | 134 | 1.2 | 0.93% | 532 | 7.6 | 1.42% |
| Within Lab (Total) | 14 | 1.0 | 6.86% | 134 | 3.2 | 2.43% | 532 | 17.9 | 3.36% |
Recovered values within ±10% of neat samples - insignificant
| Intereferent Substances | SerumSample Range | SerumInsignificant up to | UrineSample Range | UrineInsignificant up to |
|---|---|---|---|---|
| Albumin | 85-139 mg/dL | 7.5 g/dL | 33-252 mg/dL | 500 mg/dL |
| Bilirubin | 79-186 mg/dL | 20 mg/dL | N/A | N/A |
| Calcium | N/A | N/A | 25-241 mg/dL | 50 mg/dL |
| Conjugated Bilirubin | 88-153 mg/dL | 20 mg/dL | 29-269 mg/dL | 50 mg/dL |
| Acetaminophen | 79-142 mg/dL | 15.6 mg/dL | N/A | N/A |
| Acetylsalicylic Acid | 80-155 mg/dL | 3 mg/dL | N/A | N/A |
| Ibuprofen | 84-157 mg/dL | 21.9 mg/dL | N/A | N/A |
| Creatinine | N/A | N/A | 29-255 mg/dL | 300 mg/dL |
| Hemoglobin | 77-185 mg/dL | 500 mg/dL | 24-238 mg/dL | 500 mg/dL |
| IgG | 71-132 mg/dL | 7500 mg/dL | 31-252 mg/dL | 500 mg/dL |
| L-ascorbic acid | 76-190 mg/dL | 3 mg/dL | 21-264 mg/dL | 500 mg/dL |
| Triglycerides | 85-192 mg/dL | 3000 mg/dL | N/A | N/A |
| Urea | N/A | N/A | 25-229 mg/dL | 6000 mg/dL |
| Uric Acid | 79-137 mg/dL | 20 mg/dL | 27-216 mg/dL | 250 mg/dL |
| pH | N/A | N/A | 30-315 mg/dL | pH 2-12 |
| Specific gravity | N/A | N/A | 27-232 mg/dL | 1.01-1.03 |
Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a BECKMAN COULTER AU 2700 analyzer, each run with their corresponding results were determined as;
| Sample type | Low Value | High Value | # Samples | R2 Corr. | Slope | Intercept |
|---|---|---|---|---|---|---|
| Serum (mg/dL) | 13 | 696 | 99 | 0.9992 | 0.9715 | 2.7 |
| Urine (mg/dL) | 10 | 656 | 100 | 0.9989 | 1.0222 | -0.9 |
{15}------------------------------------------------
Medicon Hellas Total Bilirubin – code # 1419-0142
| Reportable Range | Low Value High Value | |||
|---|---|---|---|---|
| Serum | 0.10 | 30.00 | mg/dL | |
| LOD (limit of detection) -serum | 0.01 | mg/dL | ||
| LOQ (limit of quantification) -serum | 0.09 | mg/dL | ||
| Precision -serum (mg/dL) | Level 1 | SD | CV | |
| Repeatability | 1.4 | 0.01 | 0.92% | |
| Between Run | 1.4 | 0.02 | 1.69% | |
| Total | 1.4 | 0.04 | 2.97% | |
| Level 2 | SD | CV | ||
| Repeatability | 5.5 | 0.02 | 0.38% | |
| Between Run | 5.5 | 0.08 | 1.38% | |
| Total | 5.5 | 0.20 | 3.56% | |
| Level 3 | SD | CV | ||
| Repeatability | 9.6 | 0.05 | 0.57% | |
| Between Run | 9.6 | 0.17 | 1.79% | |
| Total | 9.6 | 0.27 | 2.86% | |
| Interferences -serum | Recovered values within ±10% of neat samples - insignificant | |||
| Hemoglobin | insignificant up to | 75 | mg/dL | |
| Triglycerides | insignificant up to | 1800 | mg/dL | |
| L-ascorbic acid | insignificant up to | 3 | mg/dL | |
| Acetametophen | insignificant up to | 15.6 | mg/dL | |
| Acetylsalicylic Acid | insignificant up to | 3 | mg/dL | |
| Ibuprofen | insignificant up to | 21.9 | mg/dL |
Accuracy- A comparison was performed between lots of reagent on a Pictus® 500 analyzer, and a BECKMAN COULTER AU 2700 analyzer, each run with their corresponding reagents. Representative results from the studies are as follows:
| Range | # Samples | R2 Corr. | Slope | Intercept | |
|---|---|---|---|---|---|
| Serum (mg/dL) | 0.3 to 29.1 | 95 | 0.9996 | 1.0125 | -0.06 |
{16}------------------------------------------------
| Reportable Range | Low Value | High Value | |||
|---|---|---|---|---|---|
| Serum | 3 | 100 | mg/dL | ||
| Urine | 24 | 1300 | mg/dL | ||
| Serum | Urine | ||||
| LoD (limit of detection) | 2 | mg/dL | 21 mg/dL | ||
| LoQ (limit of quantification) | 3 | mg/dL | 24 mg/dL | ||
| Precision Serum (mg/dL) | Level 1 | SD | CV | ||
| Repeatability | 10 | 0.2 | 1.94% | ||
| Between Run | 10 | 0.2 | 2.24% | ||
| Within Lab (Total) | 10 | 0.3 | 2.97% | ||
| Level 2 | SD | CV | |||
| Repeatability | 19 | 0.4 | 2.14% | ||
| Between Run | 19 | 0.5 | 2.56% | ||
| Within Lab (Total) | 19 | 0.6 | 3.34% | ||
| Level 3 | SD | CV | |||
| Repeatability | 51 | 0.5 | 1.07% | ||
| Between Run | 51 | 1.6 | 3.16% | ||
| Within Lab (Total) | 51 | 1.7 | 3.34% | ||
| Precision Urine (mg/dL) | Level 1 | SD | CV | ||
| Repeatability | 371 | 6.7 | 1.81% | ||
| Between Run | 371 | 8.7 | 2.35% | ||
| Within Lab (Total) | 371 | 11.5 | 3.10% | ||
| Level 2 | SD | CV | |||
| Repeatability | 700 | 11.8 | 1.69% | ||
| Between Run | 700 | 20.4 | 2.92% | ||
| Within Lab (Total) | 700 | 23.6 | 3.37% | ||
| Interferences | Recovered values within ±10% of neat samples | ||||
| Serum | Urine | ||||
| Intereferent Substance | Sample Range Insignificant up to | Sample |
Medicon Hellas Urea Nitrogen – code # 1419-0022
| Interferences | Recovered values within ±10% of neat samples - insignificant | |||
|---|---|---|---|---|
| Serum | Urine | |||
| Intereferent Substance | Sample Range | Insignificant up to | Sample Range | Insignificant up to |
| Bilirubin | 13-39 | 20 mg/dL | N/A | |
| Conjugated Bilirubin | 15-39 | 20 mg/dL | 412-871 | 40 mg/dL |
| Hemoglobin | 14-38 | 250 mg/dL | 439-975 | 500 mg/dL |
| Triglycerides | 13-34 | 3000 mg/dL | N/A | |
| L-ascorbic acid | 13-39 | 3 mg/dL | 411-900 | 50 mg/dL |
| Acetaminophen | 14-36 | 15.6 mg/dL | N/A | |
| Acetylsalicylic acid | 13-39 | 3 mg/dL | N/A | |
| Albumin | N/A | 464-879 | 500 mg/dL | |
| Aminosalicylic Acid | 12-33 | 45.5 mg/dL | N/A | |
| Boric Acid | N/A | 450-925 | 500 mg/dL | |
| Cefoxitin | 13-33 | 220 mg/dL | N/A | |
| Chloramphenicol | 12-34 | 7.8 mg/dL | N/A | |
| Furosemide | N/A | 362-822 | 1.59 mg/dL | |
| Glucose | N/A | 461-887 | 5000 mg/dL | |
| Ibuprophen | 14-38 | 23.9 mg/dL | N/A | |
| Methyldopa | 13-36 | 2.25 mg/dL | N/A | |
| pH | N/A | 391-694 | pH 2 to 12 | |
| Specific Gravity | N/A | 410-857 | 1.01-1.03 | |
| Streptomycin | 14-34 | 25.8 mg/dL | N/A |
BECKMAN COULTER AU 2700 analyzer, each run with their corresponding reagents. Representive results were determined as;
| Sample type | Low Value | High Value | # Samples | R2 Corr. | Slope | Intercept |
|---|---|---|---|---|---|---|
| Serum (mg/dL) | 3 | 93 | 116 | 0.9983 | 1.0001 | -0.2 |
| Urine (mg/dL) | 31 | 1244 | 81 | 0.9972 | 0.9844 | 21.9 |
{17}------------------------------------------------
807.92 (b)(2): Brief Description of Clinical Data
Not applicable
807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing
Comprehensive study results using multiple lots of each candidate reagent, on serum and or urine samples establish that Medicon Hellas reagents used with the Diatron Pictus 500 analyzer are substantially equivalent to the Beckman Coulter AU and Abbott Diagnostics predicate devices assayed with their associated instruments.
§ 862.1035 Albumin test system.
(a)
Identification. An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.(b)
Classification. Class II.