Search Results
Found 2 results
510(k) Data Aggregation
(287 days)
XSmart iQ is a cordless motor hand-piece with torque control used for driving files in both reciprocating and continuous mode during an endodontic procedure.
The XSmart iQ is a cordless motor hand-piece having a power stage (motor hand-piece, gears and power supplier), a console based on an Apple iPad Mini® and a communication system (Bluetooth). The motor parameters can be transferred, stored and displayed via a dedicated iPad Mini® application (Endo iQ app), connected to the cordless motor hand-piece through a Bluetooth communication protocol.
Device components (including accessories):
- Motor hand-piece support and case ●
- Motor hand-piece
- Contra angle
- Spray nozzle (used for lubrication)
- iPad Mini® Protective case (Rubber Plugs and Protective case opener) ●
- AC adapter ●
- Getting started guide
- iPad Mini® with Endo iQ iOS App (not part of the kit) ●
The provided document is a 510(k) premarket notification for the Dentsply Sirona XSmart iQ, a cordless motor hand-piece for endodontic procedures. It details the device's characteristics and compares it to a predicate device, the e3 Torque Control Motor (K103653). The document primarily focuses on demonstrating substantial equivalence through non-clinical performance data and risk analysis.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly present a single table labeled "Acceptance Criteria." Instead, it lists various tests performed and their successful outcomes. The "acceptance criteria" are implied by the statement "Test results indicated that the XSmart iQ complies with the applicable Standard" or "XSmart iQ passed the test" for each study.
Here's a consolidated table based on the non-clinical performance data:
Acceptance Criteria (Implied by Standard Compliance/Test Pass) | Reported Device Performance |
---|---|
Biocompatibility: | |
ISO 10993-5:2009, EN ISO 10993-1:2009, ISO 7405:2008, DIN EN ISO 10993-5, 1999 (Cytotoxicity for contra angle) | All patient contact components (including contra angle) demonstrated biocompatibility. Test results indicated that the XSmart iQ contra angle complies with the applicable standards. |
Electromagnetic Compatibility & Electrical Safety: | |
IEC 60601-1:2005/A1:2012 (3rd Ed.) (Basic safety & essential performance) | Test results indicated that the XSmart iQ complies with the applicable Standard. |
IEC 80601-2-60:2012 (Basic safety & essential performance of dental equipment) | Test results indicated that the XSmart iQ complies with the applicable Standard. |
EN 60601-1-2:2007 (EMC - Requirements and tests) | Test results indicated that the XSmart iQ complies with the applicable Standard. |
EN 60601-1-6:2010 (General requirements for basic safety & essential performance) | Test results indicated that the XSmart iQ complies with the applicable Standard. |
IEC 60601-1-8:2012 (Alarm systems) | Test results indicated that the XSmart iQ complies with the applicable Standard. XSmart iQ passed the alarm test. |
Performance Testing (Implied by successful completion of tests outlined in DIN EN ISO 14457-2012): | |
Expected Life / Endurance (Micro motor, contra angle connection, impact on electronics, contra angle play, battery capacity) | XSmart iQ passed the test. Endurance testing passed successfully and no major impact from stress is expected in the lifetime of device. |
Micro-Motor Performance (Withstand load until 16 mNm, 0% failure rate, no motor defects after 500 hrs) | The device withstood load (at least until 16 milli-Newton meters) with a failure rate of 0% and no impact or motor defects observed after a 500 hours of runtime. |
Motor Hand-Piece and Contra Angle Endurance (500 hrs, 1,600 application/removal cycles, 25 lubrication cycles) | Endurance testing passed successfully and no major impact from stress is expected in the lifetime of device. |
Contra Angle Connection Test (1,600 application/removal cycles, 500 hrs runtime, 25 lubrication cycles) | Endurance testing passed successfully and no major impact on the expected lifetime of device is expected through stress. |
Battery Capacity Evaluation (Minor changes after 500 hrs runtime, 300 charge cycles, no degradation) | Battery capacity showed minor changes after 500 hours of runtime and approximately 300 charge cycles. No degradation of the cells was observed and the test was passed successfully. |
Oil-Ingress into Motor Test (No ingress observed after 750 hrs runtime, 25 lubrication cycles) | After 750 hours of runtime and 25 contra angle lubrication cycles no ingress of oil could be observed. This test was passed successfully. |
Performance Tests Motor Hand-Piece (Speed, torque, reciprocating movement, warning beeps) | XSmart iQ motors undergoing performance test passed the defined examination. No device failure occurred during test phase. |
Apex Pass Through (Resistance, housing isolation, capacity requirements, no influence on external apex locators) | Test has shown the resistance the housing isolation and capacity requirements are met by XSmart iQ. Also the display locations of two apex locators are not influence by measurement through XSmart iQ internal contact. |
Bluetooth Simultaneous Connection Test (No interference during simulated normal use) | No interference between the devices was observed. |
Marking Tests (No degradation in readability after rubs with water, ethanol, isopropanol) | No degradation in the readability of marking was observed. |
Endurance Battery Test (No major impact on lifetime/function after 300 charge/discharge cycles) | All test passed. No major impact of charging/discharging on lifetime and function of batteries. |
Button Test (Buttons functional after 50,000 push cycles) | All buttons were functional after applying the defined push cycles. |
Contra Angle Tests (ISO 14457:2012: Backlash, lifespan, sterilization resistance, thermodisinfectability) | The backlash fluctuated within the stipulated tolerance of ±6°. The contact quality when transferring the signal on the sliding contact was good at all times and remained within the stipulated range, above 895, on the VDW Gold drive unit. The resistance value remained below 5 ohms at all time (below 15 ohms required). The sterilization did not reveal any decisive negative influences that would affect the use of the test objects in practice. The thermo washer disinfection does not affect the usability of the contra angle heads. There were no incidents during the test affecting use to an extent that would raise concerns about releasing the product. |
Effectiveness of Contra Angle Test (Efficiency > 80%, ultimate moment within expected range) | The efficiency levels measured (90%–95%) are significantly over the 80% required. At 66 and 70 Ncm (Newton Centimeter), the ultimate moments fell within the same range on the drive side. |
Dimensional Benchmark (Smallest diameter among cordless benchmarks, passes test criteria) | The maximum diameter of the complete IQ hand- piece was the smallest of the cordless benchmark and XSmart iQ passed the dimensional benchmark test as per test criteria. |
Docking Station Disinfection Test (Resists disinfection after 250 spraying cycles per ISO 21530) | Docking station passed the test after 250 spraying cycles according to ISO 21530 standard. |
Resistance to Closing Opening Cycle Test (No wear problems during two years usage) | XSmart iQ docking station will not suffer any wear problems during two year usage. |
Bluetooth Working Distance (Passes test criteria) | XSmart iQ passed Bluetooth working distance test as per test criteria. |
Packaging Resistance Test (No visible cracks, deformation, mechanical or electrical damage after transport/climatic conditioning) | No visible cracks, color change and deformation on docking station. No mechanical damage visible and no electrical damage detected on the devices contained in the packaging, after the transportation test and the climatic conditioning. |
Usability Test (Complies with IEC 62366:2007) | Test results indicated that the XSmart iQ complies with the applicable Standard. |
Software Validation: | |
EN 62304:2006 (Software validation requirements) | Both devices (presumably XSmart iQ and its predicate) meet software validation requirement per EN 62304:2006. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size:
- For "Expected Life" testing, the sample size was n=15 for the first pass and n=5 for the second pass for the micro motor and contra angle.
- For "Micro-Motor Performance" and "Button Test", the specific sample size isn't explicitly stated, but the Button Test mentions visual inspection after various push cycles, implying testing on some sample units.
- For "Contra Angle Tests" and "Effectiveness of Contra Angle Test", specific numbers of instruments or units are mentioned (e.g., two prototype instruments for ultimate moment), but a general sample size for these tests is not provided beyond the contra angle specific tests.
- For other tests like Biocompatibility, EMC/Electrical Safety, Marking Tests, Docking Station Disinfection, Packaging Resistance, Usability, and Bluetooth tests, a specific sample size is not explicitly stated in the document. However, performing such tests implies they were done on a representative number of devices.
- Data Provenance: The document does not specify the country of origin of the data. All studies are non-clinical bench testing, meaning they were performed in a lab setting rather than on human subjects. They are inherently prospective in nature, as they were conducted specifically to evaluate the XSmart iQ device's performance against predefined standards and criteria.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This category is not applicable to the provided document. The studies described are non-clinical bench tests and do not involve human interpretation or a "ground truth" derived from expert consensus in the clinical sense. The "ground truth" for these tests is defined by objective engineering and safety standards (e.g., ISO, IEC, EN standards) and the physical properties of the device and its accessories.
4. Adjudication Method for the Test Set:
Not applicable. As the studies are non-clinical bench tests evaluated against objective standards, there is no need for an adjudication method involving multiple human readers or experts. The results are based on measurement and observation against established criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states "Clinical Performance Data: Not applicable." The studies are non-clinical bench tests comparing the device's technical specifications and performance to a predicate device and relevant standards.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The device described, the XSmart iQ, is a physical dental hand-piece with integrated software (Endo iQ app on an iPad Mini). It is not an "algorithm only" device. The performance tests evaluate the device as a whole, including its mechanical, electrical, and software components, in its intended operational context. Therefore, a "standalone algorithm only" performance test, as might be done for AI-driven diagnostic software, is not relevant and was not performed.
7. The Type of Ground Truth Used:
The "ground truth" for the non-clinical tests is based on:
- Established International Standards: Such as ISO 10993, IEC 60601, EN 60601, ISO 14457, ISO 21530. Compliance with these standards is the primary "ground truth."
- Defined Performance Parameters: Explicit mechanical or electrical limits (e.g., "withstood load until 16 milli-Newton meters," "efficiency levels measured (90%–95%) are significantly over the 80% required," "backlash fluctuated within the stipulated tolerance of ±6°").
- Absence of Adverse Events: (e.g., "no impact or motor defects observed," "no degradation of the cells," "no ingress of oil could be observed," "no device failure occurred," "no visual cracks").
There is no mention of pathology, expert consensus (in a clinical sense), or outcomes data used to establish ground truth for the evaluations presented.
8. The Sample Size for the Training Set:
This question is not applicable. The XSmart iQ is a hardware device with associated software, not a machine learning or AI model that requires a "training set" in the conventional sense. The "software validation" mentioned refers to standard software engineering verification and validation, not training of a learned model.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as there is no "training set" for this type of medical device's evaluation.
Ask a specific question about this device
(164 days)
The e3 Torque Control Motor is a medical device designed for use by dentists for use with dental root canal instruments in continuous rotation with torque control or in reciprocating movement.
The e3 Torque Control Motor is an electric motor-driven handpiece intended for the endodontic industry. It works in both reciprocating mode and continuous rotation. This motor is provided with pre-programmed settings for the most popular nickel titanium rotary file systems sold by DENTSPLY. The dentist is also able to program 15 settings of their choice for speed and torque control. The device consists of a control unit with an LCD screen for selection of settings. A foot pedal is connected to the control unit so the dentist can selectively activate the motor. A power supply is also connected to the control unit to allow the dealer to charge the battery. The motor is connected to the control unit and a 6:1 contra angle. Files attach to the contra angle.
This submission K103653 for the e3 Torque Control Motor is a 510(k) premarket notification claiming substantial equivalence to a predicate device, the X-Smart™ Easy (K092614). It does not contain a study that demonstrates the device meets specific performance acceptance criteria in the way a clinical trial or a detailed scientific study would for a novel medical device.
Instead, the submission focuses on demonstrating substantial equivalence through:
- Technological Characteristics Comparison: Highlighting similarities and minor differences between the subject device and the predicate device.
- Compliance with Recognized Standards: Showing that the device meets safety and performance standards (e.g., Biocompatibility, EMC, Electrical Safety, Software Validation).
Therefore, the requested information regarding acceptance criteria and a study proving the device meets them, in the context of performance metrics like accuracy, sensitivity, or specificity, is not applicable to this type of submission. There are no performance acceptance criteria defined in terms of quantifiable device output that would typically be fulfilled by a comparative effectiveness study or a standalone algorithm performance study.
Here's a breakdown of what is available in the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Not applicable in the traditional sense of a performance study. The "acceptance criteria" here are compliance with recognized standards and the "performance" is demonstrated by meeting these standards and showing comparable technological characteristics to a legally marketed predicate device.
Acceptance Criterion (Standard Compliance) | Reported Device Performance (Status) |
---|---|
Biocompatibility (ISO 10993) | All patient contact components demonstrated biocompatibility. |
Electromagnetic Compatibility (EN 60601-1-2) | Conforms to EN 60601-1-2:2007. |
Electrical Safety (EN 60601-1, EN 60601-1-8) | Conforms to EN 60601-1:2007 and EN 60601-1-8:2007. |
Software Validation (EN 62304) | Meets software validation requirements per EN 62304. |
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. This submission did not involve a "test set" of patient data or samples for performance evaluation in the way a diagnostic or AI-driven device would. The testing was primarily against engineering and safety standards.
3. Number of Experts Used to Establish Ground Truth and Their Qualifications
Not applicable. Ground truth, in the context of clinical or diagnostic performance, was not established for this device as it's an electromechanical tool rather than an interpretive one.
4. Adjudication Method
Not applicable. No adjudication was performed as there was no assessment of diagnostic output or subjective interpretation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices where human readers or interpreters are involved (e.g., radiologists interpreting images). The e3 Torque Control Motor is an electromechanical handpiece for dentists.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. There is no "algorithm" in the sense of an AI or diagnostic algorithm whose standalone performance needs to be evaluated. The device has software for control, but its performance is assessed through compliance with standards like EN 62304 for software validation, not through clinical output metrics.
7. Type of Ground Truth Used
Not applicable. The "ground truth" here is compliance with established engineering, safety, and biocompatibility standards, rather than clinical outcomes or pathology reports.
8. Sample Size for the Training Set
Not applicable. There was no "training set" in the context of machine learning or algorithms that learn from data.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
In summary: The K103653 submission for the e3 Torque Control Motor demonstrates substantial equivalence by comparing its technological characteristics to a predicate device and by showing compliance with relevant international and national standards for safety, biocompatibility, electromagnetic compatibility, electrical safety, and software validation. It does not involve performance studies in the way an AI/ML or diagnostic device would, and therefore, many of the requested points are not relevant to this submission.
Ask a specific question about this device
Page 1 of 1